Chemistry analyzer
service manual
Content
1
Semi-automatic Chemistry Analyzer
Chemistry analyzer
The built-in dot matrix LCD display provided with apheliotropic board ensures the clear show of all test
results, functional status and operational instruction. The English indication system guides users’ operation,
and improves man-machine communication.
1.2 Specification:
A) Lamp-house: 12V/20V W lamp
B) Wavelength range: DR-7000E grating 300~800mm available
There are eight wavelengths in the DR-7000D filter:
340nm、405nm、450nm、492nm、505nm、546nm、578nm、630nm
C)Absorbency range: -0.3~3.0Abs
D)Wavelength accuracy: grating wavelength accuracy±1nm
Filter wavelength accuracy±1nm
E)External output: RS-232 serial port. Parallel interface for printer
F)Memory: QC data, 200 item parameters, 3000 testing results
G)Methodology: end point, kinetic, 2-point kinetic, absorbance,
2
Semi-automatic Chemistry Analyzer
multipoint
H)Measurement time: 0~999 programmable
Delay time: 0~999 programmable
I) Flowcell: quartz flowcell, measuring volume is 32ul, light diameter is 1cm.
J)Temperature control: Peltier elements,25 ,30 ,37 and room temperature, accuracy ±0.1 。
K)Sampling system: the instrument adopt step motor control, sampling volume from 100ul to 9999ul
programmable
Biohazard Mark.
There are two biohazard marks on the equipment , one is on the back-board, another is on the top of the
pump pipette. As shown in Figure 1-1
Biohazard mark
There might be a potential infectivity for the sample & waste reagent, please make
sure to take protection action during the testing and cleaning process and instrument
maintenance.
Samples and waste reagent should be disposed in a suitable package conforming to the local
regulation for laboratory.
3
Semi-automatic Chemistry Analyzer
T=I/I。
A=log(I。/I)
T ____ Transparency
I ____ Transmitted light intensity
I。___ Incident light intensity
A ____ Absorbance
Amplifier
VG
Lamp-house VI Display/Keyboard Printer
Vn
The principle diagram of the electronic system is shown above in Figure 2-2. When passing through the
monochromator the ray of the lamp-house will be changed into monochromic light. The result will be then
received by the photoelectric detector when it passes through the flow colorimetric cuvette, and further
transformed into electric signal. The signal will be amplified and filtered, and then transmitted into A/D
transformer. When the information is transformed into numbers, it will be sent into data collection system
4
Semi-automatic Chemistry Analyzer
of the computer for kinds of calculation and transaction. Users can either read the final result on the
display or alternatively get it printed out.
2.2 Functions
1.Measurement methods:
a)End-point b) Two-point kinetic
c)Kinetic d) Absorbance
e)Multipoint
2. Internal and external printer is available for uses’ options.
3. Sampling volume, test method is programmable
4. Users can determine whether to execute real time print out of the test results.
5. If any operational mistakes or other false results are found during the process of measurement, users can
choose the repeat key to retest, the original serial number of the test remains.
6.Rinsing program is available between two sample tests to avoid cross contamination.
7. Up to 31 days quality control tracking, and the printing results provided.
8. Integrated test report printing is available
9. Reaction curve display:
Reaction curve for sample and standard liquid will be displayed on the screen during the testing, as shown
in Figure 2-3.
ABS LIMIT
Absorbance Absorbance
ABS LIMIT
0 0
t1 time t1 time
Name of part:
Name of part:
7
Semi-automatic Chemistry Analyzer
Name of part:
8
Semi-automatic Chemistry Analyzer
Name of part:
9
Semi-automatic Chemistry Analyzer
Name of part:
10
Semi-automatic Chemistry Analyzer
Name of part:
11
Semi-automatic Chemistry Analyzer
Name of part:
Please note: even though there is fungus or dirty on the grating and reflector, also can’t wipe it ,
if wiping it so that run result is wrong.
12
Semi-automatic Chemistry Analyzer
13
Semi-automatic Chemistry Analyzer
after switch on
Low sucking 1.Check +13.6V and +5V of power 1.Check supply power part or
value or does supply replace power supply
not suck 2.Motor drive chip 3977 is false 2.Re-joint or replace the 3977
sample jointed or broken chip
3.Sucking pipette or pump pipette is 3.Replace sucking pipette or
broken pump pipette
4.The motor is broken 4.Replace the sucking motor
Display 1.The chip 4410 is broken 1.Replace the 4410 chip
temperature 2.Temperature sensor AD590 is 2.Replace the temperature sensor
abnormity broken 3.Reinstall or replace connect
3.The connect wire between the wire
temperature sensor and the main 4.Replace the AD623 chip
board is loose
4.Amplifier AD623 not magnify
multiple
Time can’t 1.Column crystal oscillator 32.768 is 1.Replace column crystal
memorized or broken oscillator
adjust 2.Chip 8563 is broken or false joint 2.Re-joint or replace the 8563
3.3V battery don’t work or socket of chip
battery false joint 3.Replace battery or re-joint
socket of battery
Press sucking 1.Switch of sucking key don’t work 1.Replace switch of sucking key
key can’t suck 2.The connect wire between the 2.Reinstall the connect wire
sample and sucking key and the connecting 3.Replace sucking pipette and
can’t clean board is loose pump pipette
too. 3.Trouble of sucking pipette or 4.Check the connect wire or
pump pipette replace the motor
4.Trouble of sucking motor 5.Re-joint or replace the 3977
5.The 3977 chip of drive board is chip
broken
AD value is 1.Grating motor step with chopping 1.Check motor step
incorrect of wheel motor step is incorrect 2.Check sucking pipette and
each 2.Trouble of liquid circuit pump pipette has leak or not or
wavelength 3.Colorimetric pool has air bubble replace it
4.Trouble of W lamp voltage 3.Use alcohol to clean
colorimetric pool
4.Measure voltage of W lamp
normal range is 11V~12V
Instrument 1.Trouble of 5V of power supply 1.Replace the 4410 chip of the
can’t rise 2.The connect wire between the main board with power board
temperature Peltier and the main board is loose 2.Reinstall the connect wire
3.Trouble of control temperature 3.Replace control temperature
audion audiom
4.Peltier is broken 4.Replace the Piltier
15
Semi-automatic Chemistry Analyzer
16
Semi-automatic
automatic Chemistry Analyzer
Figure 5-3-1
Figure 5-4-1
Pump pipette:
Diameter :2.06mm(inner)
Length:90mm
17
Semi-automatic
automatic Chemistry Analyzer
Figure 5-5-1
(1) Press the cover below the paper exit gently, the cover will flick automatically and be open
(2) Cut the paper as Figure 5-6-1.
(3) Switch on the printer power, insert the paper into the paper entrance, it will enter into the crankshaft
automatically.
Figure 5-6-1
Yes—press enter
No—press exit
Figure 1
Step 4 press 5 系统监测 (system monitoring), the screen will show as following
19
Semi-automatic Chemistry Analyzer
波长 (wavelength) ABS AD
340
405
450
492
505
546
578
630
20
Semi-automatic Chemistry Analyzer
Yes—press enter
No—press exit
Figure 1
Figure 2
Please input “000”. And then the screen will show as follows
If the third group value is 0000. please change it to be 0041 as shown in the picture. And then press “enter”
to save the change
And then from Figure 2. Please press 4.and then go to the next interface
21
Semi-automatic Chemistry Analyzer
Please place the distilled water under the aspiration tube and then press 4 to go to the next interface.
After the values stop changing. Please press “Enter” to save the values. And then the screen will show
Chinese words “存储完成“(save complete)
And then please go back to the initial interface as shown in figure 2. and input “000”. You can see the
middle value has changed. As the screen shows
And then you can click “exit” to leave the current interface and then restart the instrument
22
Semi-automatic Chemistry Analyzer
1
Test report data output format:
(individual data)
1、 Data output format:
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
STX CR LF
SP SP SP SP SP SP SP Report SP Report SP Report SP Report SP Unit CR LF
CR LF
SP date : × × × × - × × - × × SP × × : × × : × × CR LF
SP Sample No : × × × CR LF
SP Item name : × × × × × × × × CR LF
SP Reference value : × × × × × × × × × × × × × × × CR LF
SP Serial No. SP SP SP SP SP Result SP SP SP SP Unit SP SP Reminder info CR LF
SP × × × SP × × × × × × × × × × × × × × × × × SP × × CR LF
CR LF
ETX
Note:
SP------- space character (20h hex)
X--------- virtual value bit
STX----- start character (02h hex)
CR------- return character (0dh hex)
LF-------- line feed character (0ah hex)
ETX------ ending character (03h hex)
(Integrated data)
2、 Data input format:
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
STX CR LF
SP date : × × × × - × × - × × SP × × : × × : × × CR LF
DIRUI INDUSRIAL CO., LTD
SP Sample No : × × × CR LF
SP Item name : × × × × × × × × CR LF
SP Reference value : × × × × × × × × × × × × × × × CR LF
SP Serial No. SP SP SP SP SP Result SP SP SP SP Unit SP SP Reminder info CR LF
SP × × × SP × × × × × × × × × × × × × × × × × SP SP SP SP × CR LF
CR LF
SP Item name : × × × × × × × × CR LF
SP Reference value : × × × × × × × × × × × × × × × CR LF
SP Serial No. SP SP SP SP SP Result SP SP SP SP Unit SP SP Reminder info CR LF
SP × × × SP × × × × × × × × × × × × × × × × × SP × × CR LF
CR LF
……
……
......
......
......
ETX
(QC data )
3、 Data output format:
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30
STX CR LF
SP SP SP SP SP SP SP Report SP Report SP Report SP Report SP Report CR LF
CR LF
SP Date : × × × × - × × - × × SP × × : × × CR LF
SP Item Name : × × × × × × × × CR LF
SP Lots No. : × × × × × × × CR LF
SP Standard SP X : × × × × × × × × × CR LF
SP Standard SP S D : × × × × × × × × × CR LF
SP Result : × × × × × × × CR LF
ETX
31
DIRUI INDUSRIAL CO., LTD
32
DIRUI INDUSRIAL CO., LTD
Note:
1.The above info just for regular items ,for the details, please check the specification on the reagent kits.
2.If needed, user also could program the new parameters according to the specification on the reagent kits.