UA-240
User Manual
BIOBASE GROUP
Version 2020.11
Description
Before using the product, please reading this manual carefully, and correctly use the
product.
After reading, please properly save user manual, and reading it at any time.
Warning
●Don't use wet hand to pull out the power, it may get an electric shock.
●Don’t use damaged wires and connecting cables. Don’t trample, twist or pull the
wires and cables. If the wires and cables are broken, it may cause a fire.
●The analyzer must be used under good earthing.
●Confirm that the input voltage meets the requirements of the instrument, using the
specified specifications fuse.
●Before connect the power cable, please confirm analyzer switch is in a state of
[0].
●During the instrument running, please do not contact with moving parts to avoid
accidents.
Attention
●Please use and save the test strips refer to the user manual, and ensure to use them
in the period of validity.
Biobase Biodustry (Shandong) Co., Ltd
●Don't use turpentine oil, benzene to clean the external of instrument, because it
may cause the change of the color and shape. Scrub with a soft cloth or wet cloth. For
serious besmirch, clean with diluted detergent or alcohol.
● Operators have an obligation to abide the region/ country rules for reagents,
consumables and other emissions, waste liquid and waste sample processing .
●Sample has potential biological risk of infectious, may irritate eyes, skin and
mucous membrane. Operators should comply with the requirements of the
laboratory safety operation, and dressed in personal equipment (e.g., gloves) when
contact related items in the laboratory.
3 / 43
Chapter I Auto Urine Analyzer Instruction
1.1 Description
【Performance】
1 Repeatability
Reflectivity coefficient of variation (CV) ≤1%
2 Accuracy
Test results with the reference solution value differences is no more than one order of
magnitude, and can not appear reverse differ. Positive reference solution can not
appear negative result, negative reference solution can not be a positive result.
Biobase Biodustry (Shandong) Co., Ltd
3 Stability
Reflectivity test results of variation (CV) ≤1% within 8 h after start up.
4 Testing speed
Testing speed≥240 tests /h.
5. Carry pollution
Test the positive samples of the each items’ highest concentration test results (except
for specific gravity and PH), then test negative samples. Negative samples can not
appear positive.
6. Safety
Accord with appendix A (the safety requirements, test methods and inspection rules)
and appendix B,(safety requirements and test methods) requirement.
7. Environment
Comply with the provisions of GB/T14710-2009.
8. Electromagnetic compatibility
Accord with the GB/T18268.1-2010 and GB/T18268.26-2010 requirements.
1.2 Introduction
5 / 43
Biobase Biodustry (Shandong) Co., Ltd
instrument to compensate for errors in urine color and instrument changes. Will each
reagent area reflected light magnitude compared with the blank piece reflected light
quantity, by calculating the reflectivity, instrument according to the reflectivity to
determine the content of biochemical components in urine.
Formula
Tm C s
R(%) 100 %
Ts C m
R——Reflectivity
Tm——The reflection intensity of the reagent block for measuring wavelength
Cs—— The reflection intensity of the blank block for reference wavelength
Ts—— The reflection intensity of the reagent block for reference wavelength
Cm——The reflection intensity of the blank block for measuring wavelength
The principle graph of auto urine analyzer as shown in the figure below:
Display
Control test strip
Software
6 / 43
Biobase Biodustry (Shandong) Co., Ltd
Note: Auto urine analyzer can connect the bar Code reader, can identify EAN 13,
EAN - 8, Code 39, Code - 128 yards.
1.4 Structure
7 / 43
Biobase Biodustry (Shandong) Co., Ltd
图 1-1 图 1-2
1 Power switch 2 Emergency port 3 Strips seal 4 Built-in printer
compartment
5 IC card 6 LCD 7 Sample transmission 8 Computer serial
device interface
9 Bar code reader 10 Cooling fan 11 Cleaning fluid level sensor interface
interface
12 Waste Tank Interface (DRAIN) 13Cleaning Tank Interface (WASH)
14 Power supply cooling fan window 15 Power Socket
1.5 Symbol
Symbol Significance
Attention; Emphasis on the important information in the operation steps or need to pay
special attention to the content of the users.
Serial number
Grounding mark
The above symbol information is included in the analyzer and the test strip.
2.2 Unpacking
Remove the fully automated dry chemical urine analyzer and accessories from the
three boxes A, B, and C, and check with the packing list. If the parts are damaged,
please contact the supplier.
9 / 43
Biobase Biodustry (Shandong) Co., Ltd
(1)After unpacking the sample conveyor, first remove the four feet installed on the
bottom of the box.
(2)Remove the wooden box at both ends and remove the sample conveyor cover.
(3)In the bottom board mounting hole to install the foot bolts and tighten the nut,
fixed firmly.
(4)Remove the two nylon ties that fixed the left and right pushers.
(5)Before use, place the bar code reader assembly with the four screws as shown
in the figure.
(6)The sample conveyor placed in front of the analyzer, insert the two plugs on the
host into the corresponding concave socket of the sample conveyor and strictly next to
the host.
Note: After install the sample conveyor, take off the ribbon between the sample
extractor on the host and the pull plate, then remove the tie between the belt and the
torr frame at the left end of the sample conveyor. Otherwise, the instrument can not
run.
Note:
Before install the waste pipe and cleaning pipe, remove the waste and cleaning pipe
from the rear panel.
The other end of the waste pipe directly installed in the waste liquid bottle cap, and
the waste pipe into the bottle length of 40-50mm, and then the cap installed in the
waste bucket; the length of the waste pipe is 2m, if the distance from the waste pipe is
close to the instrument, you can cut off a length of waste pipe. Do not insert the
twining waste pipe into the waste liquid bucket to prevent the pipe from clogging. The
waste liquid bucket is placed vertically on the floor.
Before connecting the cleaning liquid bucket, first unscrew the cleaning liquid bucket
cap, replace the cleaning liquid bottle cap component that comes with it, then tighten
the cap; then connect the cleaning liquid pipe to the cleaning liquid bucket and cover
it.
10 / 43
Biobase Biodustry (Shandong) Co., Ltd
(1) Use of a width of 57 mm, less than 45 mm in diameter thermal printing paper
roll;
(2) After unlocking the printer cover, open the printer cover up;
(3) Load the paper into the printer; let the paper pass under the pulley 2 and roll
the knob 1 at the same time to draw the paper and cover the printer cover.(When the
machine charged work, the printer will automatically output the printer paper)
Connect one end of the power cable to the power port of the device and the other end
to the dedicated power supply on the hospital to ensure reliable grounding of the
power supply.
Note: The socket connected to the power cable must be grounded reliably.
2.3.5 Connect PC
The instrument can be connected to a computer via a communication cable to transfer
data to a computer. Plug one end of the communication cable into the communication
interface on the back of the instrument and the other end into the serial interface of the
computer.
The instrument can be connected to a dot matrix printer and output test results via an
external pin printer. Connect one end of the printer connection to the printer and the
other end to the external printer port on the back of the instrument
Note:
Auto urine analyzer can be connected to EPSON LQ1600K series, EPSON LQ300K
series of dot matrix printer, PANASONIC1121 series of dot matrix printer.
Auto urine analyzer can be connected with the bar code reader, scanning the bar code
on sample tube. Remove the bar code reader from the box, connect the bar code
reader data cable, and the other end of the data cable (RS-232 interface) and auto
analyzer bar code reader interface connected.
图 2-1
2.4.2 After the self-test passes, the display shows the main menu: (Figure 2-2)
图 2-2
2.4.3 Test paper anti-fake IC card use
In Figure 2-2, press the "Settings " button, the pop-up interface shown in
Figure 2-3:
12 / 43
Biobase Biodustry (Shandong) Co., Ltd
图 2-3
图 2-4
Insert the IC card attached to the test paper, press the "OK" button, the instrument
prompts "check the correct code", you can to test with the normal.
In the process of reading the IC card, there are several cases:
(a)The IC card supplied with each bottle of test paper can only be used once. If you
input again and press the "OK" button, the instrument prompts " The code has been
used";
(b)If the test strip in the test bottle out of shelf life, after reading the IC card, press
the "OK" button, the instrument prompts "expired code".
13 / 43
Biobase Biodustry (Shandong) Co., Ltd
After the test work is completed every day, you can press the "OFF " button in
the lower right corner of Figure 2-2, the screen prompts "cleaning", the instrument
will automatically clean the pipeline, the screen displays "Please shut down", then you
can turn off power supply.
3.1Overview
Auto urine analyzer through the touch screen to achieve man-machine dialogue, the
display shows the relevant settings options and settings, the user only need to press
the corresponding touch key to complete the set operation. Do not use a hard or sharp
object to operate the touch screen to avoid damage to the touch screen.
14 / 43
Biobase Biodustry (Shandong) Co., Ltd
Function: Perform an emergency test. Function: Enter the menu page, you can
Location: Short bar at the bottom of the main query results, set the number and other operations.
page. Location: Shortcut bar at the bottom of the
main page.
(5) Name: Setting key
3.2.2 Menu
(1) Name: Code setting (2) Name:Latest record query
Function: Make the test number setting Function: Enter the test results page to
Location: Menu page. display the current test results
Location: Menu page.
(3) Name: Code query (4)Name: Range query
15 / 43
Biobase Biodustry (Shandong) Co., Ltd
Press the main interface (Figure 2-2) "Time setting" button, enter the "Settings Panel"
interface (Figure 3-2)
Figure 3-2
Press the "time setting" button, enter the "time setting" (Figure 3-3) menu, press the
"date" button, the screen displays:
16 / 43
Biobase Biodustry (Shandong) Co., Ltd
Figure 3-3
Press the "OK" button to determine the time, date changes, press the "Cancel" key to close the
interface.
3.3.2 Date Format
Press the "Date Format" button in Figure 3-2 to bring up the dialog box:
Figure 3-5
The system provides "YYYY/MM/DD" and "MM/DD/YYYY" two formats, click the text within
the dialog box to switch, click the "return" key, give up the changes and return to the previous
menu.
3.3.3 Printer setting
Press the "Print Setup" button in Figure 3-2 to bring up the dialog box:
17 / 43
Biobase Biodustry (Shandong) Co., Ltd
Figure 3-6
The system provides "off" and "open" two print settings, click the text within the dialog box to
switch, click the "return" key, give up the changes and return to the previous menu.
3.3.4 Unit system
Press the "Unit Setup " button in Figure 3-2 to bring up the dialog box:
Figure 3-7
The system provides "international", "normal" and "symbol" three units, click the text within the
dialog box to select the format, click the "return" button to give up the changes and return to the
previous menu.
3.3.5 Abnormal value system
Press the "Asterisk " button in Figure 3-2 to bring up the dialog box:
18 / 43
Biobase Biodustry (Shandong) Co., Ltd
Figure 3-8
Click "Close / Open" to switch the system abnormal value on and off. Click to return to the
previous menu.
3.3.6 Symbol system
Press "Symbol System " key in Figure 3-2 to switch between "On" or "Off".
When the "symbol system" is "on":
When the unit system is set to "General", the printed results are shown in Figure 3-9.
When the unit system is set to "International", the printed results are shown in Figure 3-10.
When the "unit system setting" is "Symbol", the printing result is shown in Figure 3-11;
When the unit system is set to "International", the print results shown in Figure 3-13;
When the "Unit system setting" is "Symbol", the printed result is shown in Figure 3-14.
Press "CN/EN Language " key in Figure 3-2, "Language" to switch between "Chinese"
and "English".
Figure 3-15
Press "CN/EN Language" to confirm the changes to Chinese / English and return to the previous
menu. Click "Back " to discard the changes and return to the previous menu.
3.3.8 Test paper setting
Press the "Test Paper Model " key to toggle between UA14/11.
20 / 43
Biobase Biodustry (Shandong) Co., Ltd
As shown in Figure 3-2, click"URO Value " key, and "Urobilinogen Primary
Setting" switches between "+3.3", "- 0" or "-3.3".
Figure3-17
Press "+ -3.3 / -0 / -3.3" to confirm the changes and return to the previous menu. Click
"Back " to discard the changes and return to the previous menu.
3.3.10 Instrument calibration
Press the "Adjust" button in Figure 3-2 to pop up the page:
Figure3-18
Press the corresponding item, pop-up dialog box "Please put the standard bar upside down into the
test paper bin to start the instrument calibration". Press OK to start the instrument calibration,
press the cancel key to cancel the instrument calibration and return to the previous page.
3.3.11 Manufacturer settings
Press the "Manufacturer Settings " button in Figure 3-2 to bring up the dialog box:
21 / 43
Biobase Biodustry (Shandong) Co., Ltd
Figure3-19
After you enter the correct password, you can make changes to the manufacturer settings.
3.3.12 Shutdown
Figure3-20
Press the "OK" button, enter the shutdown self-test, after the completion of the self-test, you can
turn off the power, press "Cancel" button, return to the previous page.
Warning
● The operator should be trained in operating procedures and read the cautions and precautions
in the instructions carefully.
● Test sample capacity: the sample level in the test tube is not below the mark mark line of the
22 / 43
Biobase Biodustry (Shandong) Co., Ltd
Note: To ensure the reliability of the test results, please confirm the type of test paper
before use. The test paper selection function is subject to special authorization by the
supplier.
b) Place the test strip: Open the test strip seal cover and place the test strip in the direction of the
test strip mark in the test paper seal compartment.
c) Place the desiccant: Remove the desiccant (2 bags per cartridge) from the test tube and place
the desiccant in the test paper seal compartment.
4.1.2 Test tube requirements★
a) Specifications: Diameter of 15mm, the length of not more than 100mm of the disposable plastic
tube, to ensure that the height of the test tube does not exceed the lower edge of the left front door
of the instrument.
b) Test tube mouth to be neat, can not have extrusion deformation.
4.1.3 Bar code requirements★
a)Barcode type: CODE128, CODE39.
b) Bar code label size: Bar code printing width requirements of not less than 12mm, in order to
avoid rotating in the test tube may make the instrument can not read bar code, so the printing
width is not less than 12mm. The digital of bar code should be no less than 4 bits, no more than
15.
c) The initial blank and end blank of bar code should not be less than 3mm during shearing
d) Bar code label adhesive requirements:
Bar code paste should be smooth and no wrinkle;
Paste the bar code in the correct location (the lower edge of the bar code (including initial
blank) not less than the test tube rack marks line), to ensure the correct reading of bar code,
as shown below (image 6-4). Paste and when putting into the test tube rack in the tube,
confirmed from the longitudinal notch of the test tube rack can see all barcode labels the;
When the bar code is the CODE39 code in lowercase letters, the screen display after scanning,
and print the report on the ID number are the corresponding capital letters with "+"; record
23 / 43
Biobase Biodustry (Shandong) Co., Ltd
query, unable to enter "+". When record query, unable to input "+", so the record only by the
sample number query, query and not by ID.
Such as: ID number is 36Fa, the corresponding number after scanning is 36F+A, "+" also occupies
a character bit.
4.1.4 The cleaning liquid (concentrated) preparation method.★
The solution of concentrated cleaning liquid bottle into the 10L liquid cleaning barrel attached,
and the cleaning liquid bottle is rinsed with distilled water, and the rinsing solution is poured into
the 10L cleaning tank,and then the distilled water is added to 10L Scale line, mix can be used after.
Note: Each barrel cleaning liquid can be tested about 2800 strips in a continuous test
mode.
4.1.5 Check the cleaning liquid bucket, waste liquid bucket and thermal printing paper.★
a) The cleaning liquid bucket: put enough cleaning liquid into the cleaning solution bucket.
b)Waste liquid bucket: correctly install waste liquid bucket.
c) Thermal printing paper: open the cover of the printer, check whether there is printing paper, if
the printing paper color becomes pink, that means printing paper is about to run out, please
change a roll of new printing paper.
4.1.6 Urine samples.★
a)Stick a barcode label on the test tube.
b) Test tube urine sample volume: The test tube sample level is not lower than the test tube rack
mark line.
c)The test tube with urine sample was placed on the test tube rack, and the test tube was vertically
inserted into the rubber base.
d)Place the test tube rack: test tube rack on the load side transfer tray, namely the right side of
emergency port.
Precautions for urine samples:
Before testing, please fully mix urine samples, but not use centrifuge, otherwise affect the
sensitivity of the results of the examination.
Use of fresh urine samples. If there is no test within one hour after collection, please seal the
urine and keep it refrigerated, and return it to room temperature before testing.
Do not add preservatives, disinfectants or detergents to urine samples.
Urine samples should be kept away from direct sunlight.
If the urine contains ascorbic acid, the test results of glucose may be lower than its true value.
4.2 Routine test★
(1) After the self-test finished, the screen displays as shown:
24 / 43
Biobase Biodustry (Shandong) Co., Ltd
Figure 4-1
In the picture: "0000" is current ID, "27 ℃" is the current environmental temperature.
(2) Press the "start" button , pop-up dialog box
Figure 4-2
Click "OK" button, the instrument automatically sampling to test tube rack, to test the sample in
turn,then output test report :
25 / 43
Biobase Biodustry (Shandong) Co., Ltd
Figure 4-3
Press the "print " button , print the current query report.
Figure 4-4
The user can input the emergency sample number (input range 1-9999) or use the barcode scanner
to scan the ID number, and press the "OK" key, the instrument will suspend the current test
content for emergency test
26 / 43
Biobase Biodustry (Shandong) Co., Ltd
(3) After the completion of the test, the test results pop-up dialog box:
Figure 4-5
Press the "print " button , print the current query report.
Figure 4-6
4.5 Cancel the test★
Press "cancel test" button, cancel the sample test, return to the main interface.
After cancel the test , the instrument will push all test tube rack, and needle lifting, pipeline
cleaning. After cleaning is completed, complete the test to return to the main page.
27 / 43
Biobase Biodustry (Shandong) Co., Ltd
Note:
It is a normal phenomenon if the probability of instrument stuck is not more than 1 %, please
remove the test strip and then re-flat on the test paper seal bin.
28 / 43
Biobase Biodustry (Shandong) Co., Ltd
(1) Press the "menu " button to enter the menu page :
Figure 5-1
Users can query historical records, emergency records, and clean record .
5.2 Number query
image 5-2
Move the cursor to the beginning, input the test number to be checked; press the "OK" key to
enter the query result list:
29 / 43
Biobase Biodustry (Shandong) Co., Ltd
image 5-3
Press the "print " button , print the current query report.
Press the "clear records " button, then pop-up dialog box:
Figure 5-4
Press "OK" to confirm the empty instrument record, press "Cancel" key to cancel the data empty
operation.
30 / 43
Biobase Biodustry (Shandong) Co., Ltd
Declaration: If the equipment is not used in accordance with the method specified in this
manual, the protection provided by the equipment may be destroyed.
6.1 Cleaning test paper sealing chamber★
After testing 1000 strips and stopping the test, the screen will show "please clean test paper
sealing chamber ". Then press the "sure" button on the screen and hear two buzzes. The screen
will also show "please clean the platform", then press the "sure" button on the screen to return to
the main screen.
When the screen appears "please clean test paper sealing chamber ", the following methods are
used to clean:
Open the upper cover of test paper sealing chamber;
Take out the remaining test paper and place them in a clean, dry test paper tube and seal the
upper cover;
Clean the sealing chamber inside with a brush,to ensure the wheel inside is clean;
Cover up the upper cover of test paper sealing chamber;
Open the upper right of the instrument and put down the baffle on the right side of the
instrument; pour out the sealing chamber;
Clean the inside of the wheel groove with the brush to make sure the groove is clean.
Righting test paper sealing chamber and the right side bafflet, and cover the upper right of the
instrument.
6.2 Cleaning Waste Box
Install the waste box at the left side bottom of the instrument
To avoid dust on the surface of fiber optical sensor so that making the instrument
misjudged, the optical fiber sensor is cleaned once a week.
Cleaning method:After power cut, use the absorbent cotton to wipe fiber sensor
reflecting mirror.
To avoid stains and dirt on the surface of barcode reader scan window so that making the
barcode reader generate misread or not read barcode phenomenon,barcode reader scan
window is cleaned once a week.
Cleaning method : Wipe the barcode reader scanning window with the absorbent
cotton.
To ensure the accurate drop of the instrument, it is recommended that the cleaning
pump be changed once a year.
6.9 Accessories list★
32 / 43
Biobase Biodustry (Shandong) Co., Ltd
C packing box
9. Cleaning liquid bucket cover
1. Test Strip(sample) components
2. Print paper 10. Waste liquid bucket
3. Calibration strip box(1 gray components
strip ) 11. Test tube holder
4. USB cable 12. Bar code reader
5. Fuse 13. Cleaning liquid
6. Hairbrush 14. Cleaning fluid filtered water
7. Power cable mouth components
8. Test tube 15. Cleaning liquid tube
16. Waste liquid tube
Note: Please contact our company if there is any damage of accessories.
a. Fuse replacement
Unplug the host power cord; remove the fuse holder on the power socket with a flat
blade screwdriver; put the fuse in the clip of the fuse holder.
b. Waste liquid bucket replacement
The waste liquid bucket advised changes every three months(the cover can be used
for a long time)。
After the instrument suggests that the liquid bucket is full, the waste liquid needs to be
cleaned out.
Method :Take medical gloves and remove the waste liquid pipe between the main
instrument and the waste liquid bucket bibcock. (One end of the waste liquid barrel
bibcock),then unscrew the waste liquid bucket cover component and pour the waste
liquid in the designated place. Finally, screwed up the cleaning bucket cover and plug
the waste liquid pipe.
c. Printing paper replacement
First click on the front end of the printed cover plate on the left upper part of the machine,
and the cover plate will be automatically flipped open. Take out the discarded paper
cartridge and press the printer tightly to open it. Press the new printer paper to the printer
press roller. Rotate the printer by hand. Turn around 50mm, and lock the printer clamping
33 / 43
Biobase Biodustry (Shandong) Co., Ltd
tightly,thread the exposed print paper out of the corresponding slot in the print cover, and
cover with a print cover.
7.1Transportation requirement
The automatic dry chemical urine analyzer of our company belongs to group 1 class A
non-life support equipment according to GB4824,and the equipment should be kept
away from high power radio transmitting station, radar transmitting station, high
frequency current and other equipment.
Statement
a) The equipment conforms to the emission and immunity requirements specified in
GB/T18268.26.
b) The equipment is designed and tested in accordance with class A equipment in GB4824. The
equipment may cause radio interference when under home situations and need to take protective
measures.
c) It is recommended to assess the electromagnetic environment before use of the equipment.
Notice : Prohibit to use the equipment next to the strong radiation source, otherwise it
Power line:1.5m
34 / 43
Biobase Biodustry (Shandong) Co., Ltd
Appendix A
Guarantee
Dear user:
Thank you for using the automatic dry chemical urine analyzer,our company will provide you
with the following services:
1. Provide technical advice at any time
2. The machine is guaranteed maintenance for one year since purchase.
3. Providing paid services to the following situations;
1) Products that have passed the warranty period;
2) Damage caused by accident and improper use;
3) Damage because the instruction are not to be used according to specifiction and self -
repairing.
With the improvement of technology,our company will provide the fully automatic dry chemical
urine analyzer upgrade service.
For technical support, please contact us according to the following address and phone:
Company Name: Biobase Biodustry(Shandong)Co.,Ltd
Add:No.51 South Gongye Road, Jinan City, China
Tel: +86-531-81219801/03
Fax: +86-531-81219804
E-mail: export@biobase.com
Web: www.meihuatrade.com
35 / 43
Biobase Biodustry (Shandong) Co., Ltd
Appendix B
follows:
Check No
Hardware handshake No
36 / 43
Biobase Biodustry (Shandong) Co., Ltd
ST C L
X R F
S C L
SP ⅹ ⅹ ⅹ ⅹ - ⅹ ⅹ - ⅹ ⅹ ⅹ ⅹ : ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P R F
S S S C L
SP M E A S N O . ⅹ ⅹ ⅹ ⅹ
P P P R F
S S S C L
SP I D ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P P P R F
S C L
SP P O R T N O . ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P R F
SP( S S S C L
U B G ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F
SP( S S S C L
B I L ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F
SP( S S S C L
K E T ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F
SP( S S S C L
B L D ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F
SP( S S S C L
P R O ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F
SP( S S S C L
N I T ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F
SP( S S S C L
L E U ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F
SP( S S S C L
G L U ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F
SP( S S S S C L
S G ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
!) P P P P R F
S S S S C L
SP P H ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P P P P R F
S S S S C L
SP V C ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P P P P R F
S S C L
SP M A L B ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P P R F
ET
37 / 43
Biobase Biodustry (Shandong) Co., Ltd
0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6
ST C L
X R F
SP ⅹ ⅹ ⅹ ⅹ - ⅹ ⅹ - ⅹ ⅹ S ⅹ ⅹ : ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
P R F
SP M E A S S S N O . S ⅹ ⅹ ⅹ ⅹ C L
P P P R F
SP P O R T S N O . ⅹ ⅹ - ⅹ ⅹ C L
P R F
SP( U B G S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
*) P P P R F
SP( B I L S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
*) P P P R F
SP( K E T S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
*) P P P R F
SP( B L D S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
*) P P P R F
SP( P R O S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
*) P P P R F
SP( N I T S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
*) P P P R F
SP( L E U S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
*) P P P R F
SP( G L U S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
*) P P P R F
SP( S G S S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
!) P P P P R F
SP P H S S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
P P P P R F
SP V C S S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
P P P P R F
SP M A L B S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L
P P R F
ET
ANNOTATE:
SP(*)---SP OR*
STX=0X02 CR=0X0D LF=0X0A
SP=0X20 ETX=0X03 X=Any ASCLL character
38 / 43
Biobase Biodustry (Shandong) Co., Ltd
Appendix C
39 / 43
Biobase Biodustry (Shandong) Co., Ltd
40 / 43
Biobase Biodustry (Shandong) Co., Ltd
Appendix D
42 / 43
Biobase Biodustry (Shandong) Co., Ltd
Appendix E
Reagent consumption
Start to clean 14
Washing pipeline 33
Shutdown cleaning 16
43 / 43
BIOBASE GROUP
2# building, No.9 Gangxing Road, High-tech Zone, Jinan City, Shandong Province,
China
Tel: +86-531-81219803/01
Fax: +86-531-81219804
Inquiry: export@biobase.com
Complaints: customer_support@biobase.cc
After-sales service: service_sd@biobase.cc; service_ivd@biobase.cc
Web: www.biobase.cc/www.meihuatrade.com / www.biobase.com