Auto Urine Analyzer

UA-240
User Manual

BIOBASE GROUP

Version 2020.11
Description

Thank you for purchasing auto urine analyzer.

Before using the product, please reading this manual carefully, and correctly use the
product.

After reading, please properly save user manual, and reading it at any time.

Warning

●Don't use wet hand to pull out the power, it may get an electric shock.

●Don’t use damaged wires and connecting cables. Don’t trample, twist or pull the
wires and cables. If the wires and cables are broken, it may cause a fire.
●The analyzer must be used under good earthing.

●Confirm that the input voltage meets the requirements of the instrument, using the
specified specifications fuse.

●Before connect the power cable, please confirm analyzer switch is in a state of
[0].

●Do not be used in inflammable and explosive environment.

●During the instrument running, please do not contact with moving parts to avoid
accidents.

Attention

●The instrument must be used by medical laboratory professionals or trained


doctors, nurses or or laboratory technicians.

●The hospitals or inspection organizations should prepare a maintenance plan, and


in strict accordance with the it to maintenance, otherwise it may result in equipment
failure.

●Please use and save the test strips refer to the user manual, and ensure to use them
in the period of validity.
Biobase Biodustry (Shandong) Co., Ltd

●Don't use the test strips that exceed of period of validity.

●Don't use turpentine oil, benzene to clean the external of instrument, because it
may cause the change of the color and shape. Scrub with a soft cloth or wet cloth. For
serious besmirch, clean with diluted detergent or alcohol.

●Winter outdoor temperature is lower, the instrument should be save at room


temperature for 24 hours , then use it to test.

Biological hazard mark

● Operators have an obligation to abide the region/ country rules for reagents,
consumables and other emissions, waste liquid and waste sample processing .

●Sample has potential biological risk of infectious, may irritate eyes, skin and
mucous membrane. Operators should comply with the requirements of the
laboratory safety operation, and dressed in personal equipment (e.g., gloves) when
contact related items in the laboratory.

●Correctly handling equipment consumables.

●Don't reuse of disposable product.

3 / 43
Chapter I Auto Urine Analyzer Instruction

1.1 Description

【Name】Auto Urine Analyzer


【Model】UA-240
【Manufacturer】Biobase Biodustry(Shandong)Co.,Ltd
【Address】No.51 South Gongye Road,Jinan,China 250101
【Contact】Tel:+86-531-81219803/01 Fax:+86-531-81219804
Email:service_sd@biobase.cc
【Lifetime】8 years
【 Structural composition 】 Test system, liquid pipe system (including automatic
sampling, dispensation, cleaning cleaning liquid path), test tube rack sample injection
system, strips selection system, operation control system, analysis and processing
software, display system, printer system, and bar code scanning device.
【 Application 】 Match with test strips, according to color change of the reaction
generated by the reagent area of strips and the biochemical components of urine
samples, detecting urobilinogen, bilirubin, ketone body, blood, protein, nitrite, white
blood cell, glucose, PH, specific gravity.
【Function】
a) Self-checking, identify and report an error;
b) The result unit should have at least the international system of units;
c) Should have an output port;
d) Should be able to store test data;
e) Should have correction function.

【Performance】

1 Repeatability
Reflectivity coefficient of variation (CV) ≤1%
2 Accuracy
Test results with the reference solution value differences is no more than one order of
magnitude, and can not appear reverse differ. Positive reference solution can not
appear negative result, negative reference solution can not be a positive result.
Biobase Biodustry (Shandong) Co., Ltd

3 Stability
Reflectivity test results of variation (CV) ≤1% within 8 h after start up.
4 Testing speed
Testing speed≥240 tests /h.
5. Carry pollution
Test the positive samples of the each items’ highest concentration test results (except
for specific gravity and PH), then test negative samples. Negative samples can not
appear positive.
6. Safety
Accord with appendix A (the safety requirements, test methods and inspection rules)
and appendix B,(safety requirements and test methods) requirement.
7. Environment
Comply with the provisions of GB/T14710-2009.
8. Electromagnetic compatibility
Accord with the GB/T18268.1-2010 and GB/T18268.26-2010 requirements.

1.2 Introduction

Auto urine analyzer according to the principle of photoelectric colorimetry, through


the reagent area on the test strip and urine biochemical components of the reaction of
color change, determination of urine biochemical content.
When steeped the urine samples of the strip is placed into a strip holder,
transmission device will deliver the strips to the underface of detector, each reagent
block that has produced a chemical reaction on the test strip is irradiated by the light
source, and the reflected light is received by the detector. The reagent and
corresponding composition is an independent reactions, show different color, the color
of the depth and some components in urine is proportional to the relationship. The
color is deeper, the absorption value is bigger, and the reflected value is smaller; on
the other hand, the color is more shallow, the absorption value is smaller, the reflected
value is bigger. That is to say, the depth of the color is proportional to the
concentrations of the components in urine samples. There is a blank block in the

5 / 43
Biobase Biodustry (Shandong) Co., Ltd

instrument to compensate for errors in urine color and instrument changes. Will each
reagent area reflected light magnitude compared with the blank piece reflected light
quantity, by calculating the reflectivity, instrument according to the reflectivity to
determine the content of biochemical components in urine.
Formula

Tm  C s
R(%)   100 %
Ts  C m

R——Reflectivity
Tm——The reflection intensity of the reagent block for measuring wavelength
Cs—— The reflection intensity of the blank block for reference wavelength
Ts—— The reflection intensity of the reagent block for reference wavelength
Cm——The reflection intensity of the blank block for measuring wavelength
The principle graph of auto urine analyzer as shown in the figure below:

Control test strip selection


Automatic strip
selection device
Central processing unit

Display
Control test strip
Software

delivery Test strip delivery


device
Printer
Control suction, adding
and cleaning Liquid road
device
Waste collection
device Control specimen
supply Automatic sample
A/D conversions
Otical system adding device

1.3 Technical parameter

Test items: Urobilinogen(UBG) Bilirubin(BIL) Ketone body(KET)

6 / 43
Biobase Biodustry (Shandong) Co., Ltd

Blood(BLD) Protein(PRO) Nitrite(NIT)


Leukocyte(LEU) Glucose(GLU)Specific gravity(SG)
PH value(PH)
Monochromatic wavelength 525nm 610nm 660nm
Test principle Strip: Albedometry
Test tube racks volume 5 test tube rack (capacity 50 samples)
Sample volume Not lower than the marked scale line on the test tube rack
Inhalation sample volume < 1.0ml
Data storage 20,000 results
Output RS232
Language English. Other languages on request
Power supply AC110V/220V 60/50Hz
Consumption 300W
Fuse 250V 2A
Dimension(W*D*H) 660*625*581mm
Weight 65Kg
Printer Built-in thermal printer
Transmission method two-way transmission
Transmission rate 9600bps

Note: Auto urine analyzer can connect the bar Code reader, can identify EAN 13,
EAN - 8, Code 39, Code - 128 yards.

1.4 Structure

7 / 43
Biobase Biodustry (Shandong) Co., Ltd

图 1-1 图 1-2
1 Power switch 2 Emergency port 3 Strips seal 4 Built-in printer
compartment
5 IC card 6 LCD 7 Sample transmission 8 Computer serial
device interface
9 Bar code reader 10 Cooling fan 11 Cleaning fluid level sensor interface
interface
12 Waste Tank Interface (DRAIN) 13Cleaning Tank Interface (WASH)
14 Power supply cooling fan window 15 Power Socket

1.5 Symbol

Symbol Significance

Attention; personal injury

Attention; Emphasis on the important information in the operation steps or need to pay
special attention to the content of the users.

Biological pollution symbols, the risk of potential biological infectious.

Alternating current (AC) symbol

For in vitro diagnostic use only

Serial number

Grounding mark

Disconnect the power supply

Connect the power supply


8 / 43
Biobase Biodustry (Shandong) Co., Ltd

UP Handle with care

Avoid the rain Avoid the sun

The above symbol information is included in the analyzer and the test strip.

Chapter 2 Installation of auto urine analyzer

2.1 Installation environmental conditions


It is recommended that the instrument be installed by professional personnel;
 Please read this manual carefully and use appropriate wording if installed on your
own;
 Please put the instrument on a steady, flat table, don't put it together with a
centrifuge and other vibratory instruments
 Do not place the instrument in a place where it is chemical substances, corrosive
gases or strong electromagnetic interference;
 Do not place the instrument in a place where it is direct sunlight, humid or high
temperature;
 Before moving the instrument, please split the instrument into two parts: host and
sample conveyor.
 The equipment should be equipped with ventilation and maintenance of the space
required by the operator, and can operate the disconnecting switch and power
plug.
 Do not disassemble or modify the instrument;
 Do not place any liquid bottles on top of the instrument to prevent bottles from
tipping over, liquid leaks into analyzer interior.
 The working temperature range of the instrument is 15 ℃ ~ 35 ℃, the optimum
temperature is 20 ℃ ~ 25 ℃, relative humidity ≤75%.

2.2 Unpacking
Remove the fully automated dry chemical urine analyzer and accessories from the
three boxes A, B, and C, and check with the packing list. If the parts are damaged,
please contact the supplier.
9 / 43
Biobase Biodustry (Shandong) Co., Ltd

2.3 Installation of the machine


2.3.1 Installation of sample conveyor

(1)After unpacking the sample conveyor, first remove the four feet installed on the
bottom of the box.
(2)Remove the wooden box at both ends and remove the sample conveyor cover.
(3)In the bottom board mounting hole to install the foot bolts and tighten the nut,
fixed firmly.
(4)Remove the two nylon ties that fixed the left and right pushers.
(5)Before use, place the bar code reader assembly with the four screws as shown
in the figure.
(6)The sample conveyor placed in front of the analyzer, insert the two plugs on the
host into the corresponding concave socket of the sample conveyor and strictly next to
the host.

Note: After install the sample conveyor, take off the ribbon between the sample
extractor on the host and the pull plate, then remove the tie between the belt and the
torr frame at the left end of the sample conveyor. Otherwise, the instrument can not
run.

2.3.1.2 Connection of waste liquid bucket/cleaning liquid bucket

Note:
Before install the waste pipe and cleaning pipe, remove the waste and cleaning pipe
from the rear panel.

2.3.2.1 Connection of the waste liquid bucket

The other end of the waste pipe directly installed in the waste liquid bottle cap, and
the waste pipe into the bottle length of 40-50mm, and then the cap installed in the
waste bucket; the length of the waste pipe is 2m, if the distance from the waste pipe is
close to the instrument, you can cut off a length of waste pipe. Do not insert the
twining waste pipe into the waste liquid bucket to prevent the pipe from clogging. The
waste liquid bucket is placed vertically on the floor.

2.3.2.2 Connection of the cleaning liquid bucket

Before connecting the cleaning liquid bucket, first unscrew the cleaning liquid bucket
cap, replace the cleaning liquid bottle cap component that comes with it, then tighten
the cap; then connect the cleaning liquid pipe to the cleaning liquid bucket and cover
it.

2.3.3 Installation of printing paper

10 / 43
Biobase Biodustry (Shandong) Co., Ltd

(1) Use of a width of 57 mm, less than 45 mm in diameter thermal printing paper
roll;
(2) After unlocking the printer cover, open the printer cover up;
(3) Load the paper into the printer; let the paper pass under the pulley 2 and roll
the knob 1 at the same time to draw the paper and cover the printer cover.(When the
machine charged work, the printer will automatically output the printer paper)

2.3.4 Connection of power line

Connect one end of the power cable to the power port of the device and the other end
to the dedicated power supply on the hospital to ensure reliable grounding of the
power supply.

Note: The socket connected to the power cable must be grounded reliably.

2.3.5 Connect PC
The instrument can be connected to a computer via a communication cable to transfer
data to a computer. Plug one end of the communication cable into the communication
interface on the back of the instrument and the other end into the serial interface of the
computer.

2.3.6 Connect the dot matrix printer

The instrument can be connected to a dot matrix printer and output test results via an
external pin printer. Connect one end of the printer connection to the printer and the
other end to the external printer port on the back of the instrument

Note:
Auto urine analyzer can be connected to EPSON LQ1600K series, EPSON LQ300K
series of dot matrix printer, PANASONIC1121 series of dot matrix printer.

2.3.7 External bar code reader installation

Auto urine analyzer can be connected with the bar code reader, scanning the bar code
on sample tube. Remove the bar code reader from the box, connect the bar code
reader data cable, and the other end of the data cable (RS-232 interface) and auto
analyzer bar code reader interface connected.

2.4 First-boot check


After the instrument is installed, you should to do the first boot check. If you find a
problem or display an error during this process, please contact the supplier.
2.4.1 When the power switch is turned on, the display shows "Instrument initialization,
please wait ...", the instrument is initialized: (Figure 2-1)
11 / 43
Biobase Biodustry (Shandong) Co., Ltd

图 2-1
2.4.2 After the self-test passes, the display shows the main menu: (Figure 2-2)

图 2-2
2.4.3 Test paper anti-fake IC card use

In Figure 2-2, press the "Settings " button, the pop-up interface shown in
Figure 2-3:

12 / 43
Biobase Biodustry (Shandong) Co., Ltd

图 2-3

Click "IC card " button, pop-up dialog box:

图 2-4
Insert the IC card attached to the test paper, press the "OK" button, the instrument
prompts "check the correct code", you can to test with the normal.
In the process of reading the IC card, there are several cases:
(a)The IC card supplied with each bottle of test paper can only be used once. If you
input again and press the "OK" button, the instrument prompts " The code has been
used";
(b)If the test strip in the test bottle out of shelf life, after reading the IC card, press
the "OK" button, the instrument prompts "expired code".

2.4.4 Use of “OFF”key

13 / 43
Biobase Biodustry (Shandong) Co., Ltd

After the test work is completed every day, you can press the "OFF " button in
the lower right corner of Figure 2-2, the screen prompts "cleaning", the instrument
will automatically clean the pipeline, the screen displays "Please shut down", then you
can turn off power supply.

2.4.5 Precautions for using the instrument:


(a)Place the test paper in the test paper compartment (in order to ensure the accuracy
of the test, please put the remaining test paper in the test paper tube after the end of
the work, and then cover the top of the test paper tube, Test paper can be stored up to
24 hours);
(b)Place the test tube with the sample on the tube rack and place the tube holder on
the sample conveyor;
(c)press the "Start" button after the instrument automatically test, the end of the test,
built-in thermal machine automatically print test report; (in the instrument test process
do not touch any position on the screen, otherwise it will lead to errors)
(d)the instrument during the test process, some function keys on the screen become
"virtual key" (dotted display), the function key at this time is invalid key;
(e)If the instrument is not displayed as expected, or if the printed result is different
from the expected value, refer to the relevant item setting.

Chapter 3 Auto urine analyzer function and settings

3.1Overview
Auto urine analyzer through the touch screen to achieve man-machine dialogue, the
display shows the relevant settings options and settings, the user only need to press
the corresponding touch key to complete the set operation. Do not use a hard or sharp
object to operate the touch screen to avoid damage to the touch screen.

3.2 Key description


3.2.1 Main page
(1)Name: Off key (2) Name: Start key

Function: Start the test.


Function: Turn off the instrument. Location: Shortcut bar at the bottom of the
Location: Shortcut bar at the bottom of the main page.
main page.

14 / 43
Biobase Biodustry (Shandong) Co., Ltd

(3) Name: Emergency key (4) Name: Menu key

Function: Perform an emergency test. Function: Enter the menu page, you can
Location: Short bar at the bottom of the main query results, set the number and other operations.
page. Location: Shortcut bar at the bottom of the
main page.
(5) Name: Setting key

Function: Enter the control panel page, you


can set the time, language, baud rate, test paper
and other operations.
Location: The shortcut bar at the bottom
right of the main page.

3.2.2 Menu
(1) Name: Code setting (2) Name:Latest record query

Function: Make the test number setting Function: Enter the test results page to
Location: Menu page. display the current test results
Location: Menu page.
(3) Name: Code query (4)Name: Range query

Function: For Number query, click pop-up


query dialog box, you can enter the query range. Function: Range Lookup
Location: Menu page. Location: Menu page.

15 / 43
Biobase Biodustry (Shandong) Co., Ltd

(5)Name: Emergency records (6)Name: Emergency range

Function: Query emergency records. Function:Scope of emergency


Location: Menu page. Location: Menu page.
(7) Name: Clear records (8) Name: Barcode

Function: Clear all test records Function: Barcode scanner scan


Location: Menu page. Location: Menu page.

3.3.1 Date setting

Press the main interface (Figure 2-2) "Time setting" button, enter the "Settings Panel"
interface (Figure 3-2)

Figure 3-2
Press the "time setting" button, enter the "time setting" (Figure 3-3) menu, press the
"date" button, the screen displays:

16 / 43
Biobase Biodustry (Shandong) Co., Ltd

Figure 3-3
Press the "OK" button to determine the time, date changes, press the "Cancel" key to close the
interface.
3.3.2 Date Format
Press the "Date Format" button in Figure 3-2 to bring up the dialog box:

Figure 3-5
The system provides "YYYY/MM/DD" and "MM/DD/YYYY" two formats, click the text within
the dialog box to switch, click the "return" key, give up the changes and return to the previous
menu.
3.3.3 Printer setting
Press the "Print Setup" button in Figure 3-2 to bring up the dialog box:

17 / 43
Biobase Biodustry (Shandong) Co., Ltd

Figure 3-6
The system provides "off" and "open" two print settings, click the text within the dialog box to
switch, click the "return" key, give up the changes and return to the previous menu.
3.3.4 Unit system

Press the "Unit Setup " button in Figure 3-2 to bring up the dialog box:

Figure 3-7
The system provides "international", "normal" and "symbol" three units, click the text within the
dialog box to select the format, click the "return" button to give up the changes and return to the
previous menu.
3.3.5 Abnormal value system

Press the "Asterisk " button in Figure 3-2 to bring up the dialog box:

18 / 43
Biobase Biodustry (Shandong) Co., Ltd

Figure 3-8

Click "Close / Open" to switch the system abnormal value on and off. Click to return to the
previous menu.
3.3.6 Symbol system

Press "Symbol System " key in Figure 3-2 to switch between "On" or "Off".
When the "symbol system" is "on":
When the unit system is set to "General", the printed results are shown in Figure 3-9.
When the unit system is set to "International", the printed results are shown in Figure 3-10.
When the "unit system setting" is "Symbol", the printing result is shown in Figure 3-11;

2010-01-01 11:40 26℃ 2010-01-01 11:40 26℃ 2010-01-01 11:40 26℃


Sample number:0011 Sample number:0011 Sample number:0011
ID ID ID
Shelf number: Shelf number: Shelf number:
*Urobilinogen 3+ >=8mg/dL *Uurobilinogen3+ >=135umol/L *Uurobilinogen 3+
*Bilirubin 3+ >=6mg/dL *Bilirubin3+ >=1.3umol/L *Bilirubin 3+
*Ketone3+ 5mg/dL *Ketone +- 0.5mmol/L *Ketone +-
*Blood 2+ Ca80Ery/uL *Blood 2+ Ca80Ery/uL *Blood 2+
*Protein 3+ >=300mg/dL *Protein 3+ >=3.0g/L *Protein 3+
*Nitrite Pos *Nitrite Pos *Nitrite Pos
*Leukocyte 3+ >=Ca500Leu/uL *Leukocyte 3+ >=Ca500Leu/uL *Leukocyte 3+
*Glucose1+ 100mg/dL *Glucose1+ 5.69mmol/L *Glucose 1+
Proportion >=1.030 Proportion >=1.030 Proportion >=1.030
PH <=5.0 PH <=5.0 PH <=5.0
Figure3-9 Figure3-10 Figure3-11
When the "symbol system" is "off":
When the unit system is set to "General", the printed results are shown in Figure 3-12.
19 / 43
Biobase Biodustry (Shandong) Co., Ltd

When the unit system is set to "International", the print results shown in Figure 3-13;
When the "Unit system setting" is "Symbol", the printed result is shown in Figure 3-14.

2010-01-01 11:40 26℃ 2010-01-01 11:40 26℃ 2010-01-01 11:40 26℃


Sample number:0011 Sample number:0011 Sample number:0011
ID ID ID
Shelf number: Shelf number: Shelf number:
*Urobilinogen >=8mg/dL *Urobilinogen >=135umol/L Urobilinogen 3+
*Bilirubin >=6mg/dL *Bilirubin >=1.3umol/L Bilirubin 3+
* Ketone 5mg/dL *Ketone 0.5mmol/L /Ketone +-
*Blood Ca80Ery/uL *Blood Ca80Ery/uL Blood 2+
*Protein >=300mg/dL *Protein >=3.0g/L Protein 3+
*Nitrite Pos *NitritePos Nitrite Pos
*Leukocyte >=Ca500Leu/uL *Leukocyte >=Ca500Leu/uL Leukocyte 3+
*Glucose1 100mg/dL *Glucose1 5.69mmol/L Glucose1 1+
Proportion >=1.030 Proportion >=1.030 Proportion >=1.030
PH <=5.0 PH <=5.0 PH <=5.0

Figure3-12 Figure3-13 Figure3-14


3.3.7 Language

Press "CN/EN Language " key in Figure 3-2, "Language" to switch between "Chinese"
and "English".

Figure 3-15
Press "CN/EN Language" to confirm the changes to Chinese / English and return to the previous

menu. Click "Back " to discard the changes and return to the previous menu.
3.3.8 Test paper setting

Press the "Test Paper Model " key to toggle between UA14/11.
20 / 43
Biobase Biodustry (Shandong) Co., Ltd

3.3.9 Urobilinogen setting

As shown in Figure 3-2, click"URO Value " key, and "Urobilinogen Primary
Setting" switches between "+3.3", "- 0" or "-3.3".

Figure3-17
Press "+ -3.3 / -0 / -3.3" to confirm the changes and return to the previous menu. Click

"Back " to discard the changes and return to the previous menu.
3.3.10 Instrument calibration
Press the "Adjust" button in Figure 3-2 to pop up the page:

Figure3-18
Press the corresponding item, pop-up dialog box "Please put the standard bar upside down into the
test paper bin to start the instrument calibration". Press OK to start the instrument calibration,
press the cancel key to cancel the instrument calibration and return to the previous page.
3.3.11 Manufacturer settings

Press the "Manufacturer Settings " button in Figure 3-2 to bring up the dialog box:

21 / 43
Biobase Biodustry (Shandong) Co., Ltd

Figure3-19
After you enter the correct password, you can make changes to the manufacturer settings.
3.3.12 Shutdown

Click the " " button to bring up the dialog box:

Figure3-20
Press the "OK" button, enter the shutdown self-test, after the completion of the self-test, you can
turn off the power, press "Cancel" button, return to the previous page.

Chapter 4 Basic Operation


Disclaimer: Failure to use the equipment in accordance with the instructions in this manual
may damage the protection provided by the equipment.
4.1 Preparation before testing

Warning
● The operator should be trained in operating procedures and read the cautions and precautions
in the instructions carefully.
● Test sample capacity: the sample level in the test tube is not below the mark mark line of the
22 / 43
Biobase Biodustry (Shandong) Co., Ltd

test tube rack.


● If the foam in the test sample is higher than 5 mm, the sample is allowed to precipitate the
foam to 5 mm below the test.
●The operator should be away from the probe to take the sample position, so as to avoid damage
when the probe moves.
●Avoid testing viscous samples to avoid clogging the probe.
● If there is gross hematuria, placed in the last position of the last test tube, the instrument after
sucking the hematuria, press the "stop" button on the screen, the instrument will automatically
wash, and then continue to normal test.
● The waste liquid bucket of the instrument should be placed on the ground, the waste liquid
tube as far as possible with the ground perpendicular to the flow of waste liquid to ensure smooth.
4.1.1 Test strip★
a) The test paper seal compartment can hold up to 200 test strips, and the test strip can not be
placed in the test paper seal compartment for more than 24 hours. In order to ensure the accuracy
of the test results, please work every day after the end of the sealed box in the test paper into the
test paper bottle, and then cover the top of the test paper bottle, according to the instrument set the
test strip model to prepare test strips.

Note: To ensure the reliability of the test results, please confirm the type of test paper
before use. The test paper selection function is subject to special authorization by the
supplier.
b) Place the test strip: Open the test strip seal cover and place the test strip in the direction of the
test strip mark in the test paper seal compartment.
c) Place the desiccant: Remove the desiccant (2 bags per cartridge) from the test tube and place
the desiccant in the test paper seal compartment.
4.1.2 Test tube requirements★
a) Specifications: Diameter of 15mm, the length of not more than 100mm of the disposable plastic
tube, to ensure that the height of the test tube does not exceed the lower edge of the left front door
of the instrument.
b) Test tube mouth to be neat, can not have extrusion deformation.
4.1.3 Bar code requirements★
a)Barcode type: CODE128, CODE39.
b) Bar code label size: Bar code printing width requirements of not less than 12mm, in order to
avoid rotating in the test tube may make the instrument can not read bar code, so the printing
width is not less than 12mm. The digital of bar code should be no less than 4 bits, no more than
15.
c) The initial blank and end blank of bar code should not be less than 3mm during shearing
d) Bar code label adhesive requirements:
 Bar code paste should be smooth and no wrinkle;
 Paste the bar code in the correct location (the lower edge of the bar code (including initial
blank) not less than the test tube rack marks line), to ensure the correct reading of bar code,
as shown below (image 6-4). Paste and when putting into the test tube rack in the tube,
confirmed from the longitudinal notch of the test tube rack can see all barcode labels the;
 When the bar code is the CODE39 code in lowercase letters, the screen display after scanning,
and print the report on the ID number are the corresponding capital letters with "+"; record
23 / 43
Biobase Biodustry (Shandong) Co., Ltd

query, unable to enter "+". When record query, unable to input "+", so the record only by the
sample number query, query and not by ID.
Such as: ID number is 36Fa, the corresponding number after scanning is 36F+A, "+" also occupies
a character bit.
4.1.4 The cleaning liquid (concentrated) preparation method.★
The solution of concentrated cleaning liquid bottle into the 10L liquid cleaning barrel attached,
and the cleaning liquid bottle is rinsed with distilled water, and the rinsing solution is poured into
the 10L cleaning tank,and then the distilled water is added to 10L Scale line, mix can be used after.

Note: Each barrel cleaning liquid can be tested about 2800 strips in a continuous test
mode.
4.1.5 Check the cleaning liquid bucket, waste liquid bucket and thermal printing paper.★
a) The cleaning liquid bucket: put enough cleaning liquid into the cleaning solution bucket.
b)Waste liquid bucket: correctly install waste liquid bucket.
c) Thermal printing paper: open the cover of the printer, check whether there is printing paper, if
the printing paper color becomes pink, that means printing paper is about to run out, please
change a roll of new printing paper.
4.1.6 Urine samples.★
a)Stick a barcode label on the test tube.
b) Test tube urine sample volume: The test tube sample level is not lower than the test tube rack
mark line.
c)The test tube with urine sample was placed on the test tube rack, and the test tube was vertically
inserted into the rubber base.
d)Place the test tube rack: test tube rack on the load side transfer tray, namely the right side of
emergency port.
Precautions for urine samples:
 Before testing, please fully mix urine samples, but not use centrifuge, otherwise affect the
sensitivity of the results of the examination.
 Use of fresh urine samples. If there is no test within one hour after collection, please seal the
urine and keep it refrigerated, and return it to room temperature before testing.
 Do not add preservatives, disinfectants or detergents to urine samples.
 Urine samples should be kept away from direct sunlight.
 If the urine contains ascorbic acid, the test results of glucose may be lower than its true value.
4.2 Routine test★
(1) After the self-test finished, the screen displays as shown:

24 / 43
Biobase Biodustry (Shandong) Co., Ltd

Figure 4-1

In the picture: "0000" is current ID, "27 ℃" is the current environmental temperature.
(2) Press the "start" button , pop-up dialog box

Figure 4-2
Click "OK" button, the instrument automatically sampling to test tube rack, to test the sample in
turn,then output test report :

25 / 43
Biobase Biodustry (Shandong) Co., Ltd

Figure 4-3

Press the "print " button , print the current query report.

Press the "back " button ,return to the previous page.


4.3 Emergency test★
It is suitable for inserting emergency samples in the normal test,the test methods are as follows:
(1) Putting the urine sample into the test tube, insert the test tube into the test tube rack transfer
tray’s emergency port, push forward the emergency port to the locked position;

(2) Press the "emergency " button, pop-up dialog:

Figure 4-4
The user can input the emergency sample number (input range 1-9999) or use the barcode scanner
to scan the ID number, and press the "OK" key, the instrument will suspend the current test
content for emergency test
26 / 43
Biobase Biodustry (Shandong) Co., Ltd

(3) After the completion of the test, the test results pop-up dialog box:

Figure 4-5

Press the "print " button , print the current query report.

Press the "back " button ,return to the previous page.


4.4 Under the testing★
After the "start" test, instruments start sample testing, and the screen is displayed :

Figure 4-6
4.5 Cancel the test★

Press "cancel test" button, cancel the sample test, return to the main interface.

After cancel the test , the instrument will push all test tube rack, and needle lifting, pipeline
cleaning. After cleaning is completed, complete the test to return to the main page.
27 / 43
Biobase Biodustry (Shandong) Co., Ltd

4.6 The testing process matters needing attention★


 Do not pull the emergency back directly to the emergency port, so as to avoid damage to the
emergency port.
 Don't move the test rack during the test so as to avoid the confusion of the test sequence.
 Do not close to the suction probe during the test, in order to avoid injuries.
 In the process of testing if the cleaning liquid is ran out of, the analyzer will reminds "Please
add cleaning liquid" and stop testing, please add cleaning liquid into the cleaning bucket,
press "confirm" key make the instrument to return to the main screen and then click the
"start" button to continue the test instrument.
 In the testing process, if the "stop" button is used, the analyzer will test the dropped sample
paper and then stop the test. After the test is stopped, the test tube rack will be reset.
 The marking directions of test strips on test paper and test paper seal are consistent, otherwise
the test strip placed against will lead test paper sealed bin stuck, the instrument can not
operate normally.
 The strip should be flat put into test paper sealed bin, otherwise the test paper will be stuck,if
stuck during the test paper, please open the sealed bin’s cover, put the strip again to continue
the test.
 When the sample appears solid precipitation, it is recommended to use artificial mixing
method for mixing.

Note:
 It is a normal phenomenon if the probability of instrument stuck is not more than 1 %, please
remove the test strip and then re-flat on the test paper seal bin.

28 / 43
Biobase Biodustry (Shandong) Co., Ltd

Chapter 5 Record query and cleanup


5.1 Quiry

(1) Press the "menu " button to enter the menu page :

Figure 5-1
Users can query historical records, emergency records, and clean record .
5.2 Number query

Press the "Code setting " button, pop-up dialog:

image 5-2
Move the cursor to the beginning, input the test number to be checked; press the "OK" key to
enter the query result list:

29 / 43
Biobase Biodustry (Shandong) Co., Ltd

image 5-3

Press the "print " button , print the current query report.

Press the "back " button ,return to the previous page.

5.3 Clear records

Press the "clear records " button, then pop-up dialog box:

Figure 5-4
Press "OK" to confirm the empty instrument record, press "Cancel" key to cancel the data empty
operation.

30 / 43
Biobase Biodustry (Shandong) Co., Ltd

Chapter 6 Equipment Maintenance


Articles for maintenance: hairbrush、absorbent cotton, protective gloves, distilled
water. Maintenance and inspections are conducted by trained professionals.
Maintenance and inspection of site ventilation facilities and waste generated shall be
dealt with in accordance with relevant state regulations.

Declaration: If the equipment is not used in accordance with the method specified in this
manual, the protection provided by the equipment may be destroyed.
6.1 Cleaning test paper sealing chamber★
After testing 1000 strips and stopping the test, the screen will show "please clean test paper
sealing chamber ". Then press the "sure" button on the screen and hear two buzzes. The screen
will also show "please clean the platform", then press the "sure" button on the screen to return to
the main screen.
When the screen appears "please clean test paper sealing chamber ", the following methods are
used to clean:
 Open the upper cover of test paper sealing chamber;
 Take out the remaining test paper and place them in a clean, dry test paper tube and seal the
upper cover;
 Clean the sealing chamber inside with a brush,to ensure the wheel inside is clean;
 Cover up the upper cover of test paper sealing chamber;
 Open the upper right of the instrument and put down the baffle on the right side of the
instrument; pour out the sealing chamber;
 Clean the inside of the wheel groove with the brush to make sure the groove is clean.
 Righting test paper sealing chamber and the right side bafflet, and cover the upper right of the
instrument.
6.2 Cleaning Waste Box

Waste Box should be cleaned once a week

 Waste box can hold up to 500 pieces of paper


 The urine is potentially contagious, please take protective measures when
cleaning
 Please dispose of waste paper strips according to the clinical laboratory
regulations
Cleaning method:
 Take out the waste box and discard the test paper in the waste box.
 Flush the waste box with water
 Use cloth to dry the waste box or air dry
 Disinfect the waste box
31 / 43
Biobase Biodustry (Shandong) Co., Ltd

 Install the waste box at the left side bottom of the instrument

6.3 Cleaning waste liquid bucket★

Waste liquid tank should be cleaned once a week


 The waste liquid can be poured out of the waste tank in time
 The urine is potentially contagious, please take protective measures when
cleaning
 Please handle the waste liquid according to the clinical laboratory regulations.
6.4 Cleaning fiber optic sensor reflecting mirror★

To avoid dust on the surface of fiber optical sensor so that making the instrument
misjudged, the optical fiber sensor is cleaned once a week.
Cleaning method:After power cut, use the absorbent cotton to wipe fiber sensor
reflecting mirror.

6.5 Cleaning barcode reader scan window★

To avoid stains and dirt on the surface of barcode reader scan window so that making the
barcode reader generate misread or not read barcode phenomenon,barcode reader scan
window is cleaned once a week.
Cleaning method : Wipe the barcode reader scanning window with the absorbent
cotton.

6.6 Cleaning pump replacement★

To ensure the accurate drop of the instrument, it is recommended that the cleaning
pump be changed once a year.
6.9 Accessories list★

A packing box B packing box


1.Main instrument
2.Operation and daily
cleaning representation
3.User Manual 1. Sample conveyor
4.Certification 2.Barcode reader components(optional)
5.Warranty card
6.Cleaning liquid bucket
7.Waste liquid tank

32 / 43
Biobase Biodustry (Shandong) Co., Ltd

C packing box
9. Cleaning liquid bucket cover
1. Test Strip(sample) components
2. Print paper 10. Waste liquid bucket
3. Calibration strip box(1 gray components
strip ) 11. Test tube holder
4. USB cable 12. Bar code reader
5. Fuse 13. Cleaning liquid
6. Hairbrush 14. Cleaning fluid filtered water
7. Power cable mouth components
8. Test tube 15. Cleaning liquid tube
16. Waste liquid tube
Note: Please contact our company if there is any damage of accessories.

6.10 Accessories replacement

a. Fuse replacement
Unplug the host power cord; remove the fuse holder on the power socket with a flat
blade screwdriver; put the fuse in the clip of the fuse holder.
b. Waste liquid bucket replacement
The waste liquid bucket advised changes every three months(the cover can be used
for a long time)。
After the instrument suggests that the liquid bucket is full, the waste liquid needs to be
cleaned out.
Method :Take medical gloves and remove the waste liquid pipe between the main
instrument and the waste liquid bucket bibcock. (One end of the waste liquid barrel
bibcock),then unscrew the waste liquid bucket cover component and pour the waste
liquid in the designated place. Finally, screwed up the cleaning bucket cover and plug
the waste liquid pipe.
c. Printing paper replacement
First click on the front end of the printed cover plate on the left upper part of the machine,
and the cover plate will be automatically flipped open. Take out the discarded paper
cartridge and press the printer tightly to open it. Press the new printer paper to the printer
press roller. Rotate the printer by hand. Turn around 50mm, and lock the printer clamping

33 / 43
Biobase Biodustry (Shandong) Co., Ltd

tightly,thread the exposed print paper out of the corresponding slot in the print cover, and
cover with a print cover.

Chapter 7 Transportation and Storage Condition

7.1Transportation requirement

The instrument should be moisture-proof and waterproof in the transportation,


preventing the violent vibration and extrusion. Handling and handling should be
handled gently.
7.2Storage requirement

Instrument should be stored in the absence of chemicals, corrosive gas, ventilation is


good, health, clean, temperature 15 ℃ ~ 35 ℃ of indoor temperature.

Chapter 8 Electromagnetic Compatibility Specification

The automatic dry chemical urine analyzer of our company belongs to group 1 class A
non-life support equipment according to GB4824,and the equipment should be kept
away from high power radio transmitting station, radar transmitting station, high
frequency current and other equipment.
Statement
a) The equipment conforms to the emission and immunity requirements specified in

GB/T18268.26.
b) The equipment is designed and tested in accordance with class A equipment in GB4824. The
equipment may cause radio interference when under home situations and need to take protective
measures.
c) It is recommended to assess the electromagnetic environment before use of the equipment.

Notice : Prohibit to use the equipment next to the strong radiation source, otherwise it

may interfere with the normal operation of the equipment.

Power line:1.5m

Liquid level sensor connecting line: 1.5m


Electromagnetic Compatibility Testing basic performance:
Accuracy:The test results of all items in gray strip are consistent with the gray strip test values.

34 / 43
Biobase Biodustry (Shandong) Co., Ltd

Appendix A

Guarantee
Dear user:
Thank you for using the automatic dry chemical urine analyzer,our company will provide you
with the following services:
1. Provide technical advice at any time
2. The machine is guaranteed maintenance for one year since purchase.
3. Providing paid services to the following situations;
1) Products that have passed the warranty period;
2) Damage caused by accident and improper use;
3) Damage because the instruction are not to be used according to specifiction and self -
repairing.
With the improvement of technology,our company will provide the fully automatic dry chemical
urine analyzer upgrade service.
For technical support, please contact us according to the following address and phone:
Company Name: Biobase Biodustry(Shandong)Co.,Ltd
Add:No.51 South Gongye Road, Jinan City, China
Tel: +86-531-81219801/03
Fax: +86-531-81219804
E-mail: export@biobase.com
Web: www.meihuatrade.com

35 / 43
Biobase Biodustry (Shandong) Co., Ltd

Appendix B

Connector of automatic dry chemical urine analyzer


and computer

The automatic dry chemical urine analyzer is connected to the computer

via the RS-232 standard interface. The communication agreement is as

follows:

Baud rate 9600,4800,1200、19200 (optional)

Data bits 8-bit

Stop bit 1-bit

Check No

Hardware handshake No

First symbol 02H

Single-byte space 20H

Double-byte space 40H

Line break 0DH0AH

End mark 03H

The urine analyzer is connected with the computer:

36 / 43
Biobase Biodustry (Shandong) Co., Ltd

Data communication format when the ID number is set to open.

(International unit, regular unit, plus system)


1 1 1 1 1 1 1 1 1 1 2 2 2 2 2 2 2
1 2 3 4 5 6 7 8 9
0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6

ST C L

X R F

S C L
SP ⅹ ⅹ ⅹ ⅹ - ⅹ ⅹ - ⅹ ⅹ ⅹ ⅹ : ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P R F

S S S C L
SP M E A S N O . ⅹ ⅹ ⅹ ⅹ
P P P R F

S S S C L
SP I D ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P P P R F

S C L
SP P O R T N O . ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P R F

SP( S S S C L
U B G ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F

SP( S S S C L
B I L ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F

SP( S S S C L
K E T ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F

SP( S S S C L
B L D ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F

SP( S S S C L
P R O ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F

SP( S S S C L
N I T ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F

SP( S S S C L
L E U ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F

SP( S S S C L
G L U ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
*) P P P R F

SP( S S S S C L
S G ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
!) P P P P R F

S S S S C L
SP P H ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P P P P R F

S S S S C L
SP V C ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P P P P R F

S S C L
SP M A L B ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ
P P R F

ET

37 / 43
Biobase Biodustry (Shandong) Co., Ltd

Data communication format when the ID number is set to close.

(International unit, regular unit, plus system)


1 2 3 4 5 6 7 8 9 1 1 1 1 1 1 1 1 1 1 2 2 2 2 2 2 2

0 1 2 3 4 5 6 7 8 9 0 1 2 3 4 5 6

ST C L

X R F

SP ⅹ ⅹ ⅹ ⅹ - ⅹ ⅹ - ⅹ ⅹ S ⅹ ⅹ : ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

P R F

SP M E A S S S N O . S ⅹ ⅹ ⅹ ⅹ C L

P P P R F

SP P O R T S N O . ⅹ ⅹ - ⅹ ⅹ C L

P R F

SP( U B G S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

*) P P P R F

SP( B I L S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

*) P P P R F

SP( K E T S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

*) P P P R F

SP( B L D S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

*) P P P R F

SP( P R O S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

*) P P P R F

SP( N I T S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

*) P P P R F

SP( L E U S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

*) P P P R F

SP( G L U S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

*) P P P R F

SP( S G S S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

!) P P P P R F

SP P H S S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

P P P P R F

SP V C S S S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

P P P P R F

SP M A L B S S ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ ⅹ C L

P P R F

ET

ANNOTATE:
SP(*)---SP OR*
STX=0X02 CR=0X0D LF=0X0A
SP=0X20 ETX=0X03 X=Any ASCLL character
38 / 43
Biobase Biodustry (Shandong) Co., Ltd

Appendix C

Automatic dry chemical urine analyzer output value


Symbol
Short Conventional unit system International system of unit
system
Items Name
plus sign system plus sign system plus sign system plus sign system
open closed open closed
Normal 0.2mg/dL 0.2mg/dL Normal 3.4umol/L 3.4umol/L Normal
Normal 1mg/dL 1mg/dL Normal 17umol/L 17umol/L Normal
Urobilinogen UBG 1+ 2mg/dL 2mg/dL 1+ 34umol/L 34umol/L 1+
2+ 4mg/dL 4mg/dl 2+ 68umol/L 68umol/L 2+
3+ >=8mg/dL >=8mg/dL 3+ >=135umol/L >=135umol/L 3+
Neg Neg Neg Neg Neg
1+ 1mg/dL 1mg/dL 1+ 17umol/L 17umol/L 1+
Bilirubin BIL
2+ 3mg/dL 3mg/dL 2+ 51umol/L 51umol/L 2+
3+ >=6mg/dL >=6mg/dL 3+ >=103umol/L >=103umol/L 3+
Neg Neg Neg Neg Neg
+- 5mg/dL 5mg/dL +- 0.5mmol/L 0.5mmol/L +-
Ketone KET 1+ 15mg/dL 15mg/dL 1+ 1.5mmol/L 1.5mmol/L 1+
2+ 40mg/dL 40mg/dL 2+ 3.9mmol/L 3.9mmol/L 2+
3+ >=80mg/dL >=80mg/dL 3+ >=7.8mmol/L >=7.8mmol/L 3+
Neg Neg Neg Neg Neg
+- Ca10 Ery/UL Ca10 Ery/UL +- Ca10 Ery/UL Ca10 Ery/UL +-
Blood BLD 1+ Ca25 Ery/Ul Ca25 Ery/Ul 1+ Ca25 Ery/Ul Ca25 Ery/Ul 1+
2+ Ca80 Ery/uL Ca80 Ery/Ul 2+ Ca80 Ery/uL Ca80 Ery/uL 2+
3+﹥=Ca200 Ery/uL ﹥=Ca200 Ery/uL 3+ >=Ca200 Ery/uL >=Ca200 Ery/uL 3+
Neg Neg Neg Neg Neg
Trace Trace Trace Trace Trace Trace Trace
Protein PRO 1+ 30mg/dL 30mg/dL 1+ 0.3g/L 0.3g/L 1+
2+ 100mg/dL 100mg/dL 2+ 1.0g/L 1.0g/L 2+
3+ >=300mg/dL >=300mg/dL 3+ >=3.0g/L >=3.0g/L 3+
Neg Neg Neg Neg Neg
Nitrite NIT
Pos Pos Pos Pos Pos
Neg Neg Neg Neg Neg
+- Ca15 Leu/uL Ca15 Leu/uL +- Ca15 Leu/uL Ca15 Leu/uL +-
1+ Ca70 Leu/uL Ca70 Leu/uL 1+ Ca70 Leu/uL Ca70 Leu/uL 1+
Leukocytes LEU
2+ Ca125 Leu/uL Ca125 Leu/uL 2+ Ca125 Leu/uL Ca125 Leu/uL 2+
3+ > = >= Ca500Leu/uL 3+ >= Ca500Leu/uL >= Ca500Leu/uL 3+
Ca500Leu/uL
Neg Neg Neg Neg Neg
1+ 100mg/dL 100mg/dL 1+ 5.6mmol/L 5.6mmol/L 1+
Glucose GLU
2+ 250mg/dL 250mg/dL 2+ 14mmol/L 14mmol/L 2+
3+ 500mg/dL 500mg/dL 3+ 28mmol/L 28mmol/L 3+

39 / 43
Biobase Biodustry (Shandong) Co., Ltd

4+ >=1000mg/dL >=1000mg/dL 4+ >=56mmol/L >=56mmol/L 4+


﹤=5.0 ﹤=5.0 ﹤=5.0 ﹤=5.0 ﹤=5.0
5.5 5.5 5.5 5.5 5.5
6.0 6.0 6.0 6.0 6.0
6.5 6.5 6.5 6.5 6.5
PH PH 7.0 7.0 7.0 7.0 7.0
7.5 7.5 7.5 7.5 7.5
8.0 8.0 8.0 8.0 8.0
8.5 8.5 8.5 8.5 8.5
﹥=9.0 ﹥=9.0 ﹥=9.0 ﹥=9.0 ﹥=9.0

40 / 43
Biobase Biodustry (Shandong) Co., Ltd

Appendix D

Fault information table


Fault prompt /
NO. Solution
possible fault
In/out frame unit initializat Please check whether in/out frame unit’s connection and plug-in connection are
1
ion failure fixed
Strip selection box unit Please check whether strip selection box unit’s connection and plug-in
2
initialization failure connection are fixed
Pipeline unit Please check whether pipeline unit’s connection and plug-in connection are
3
initialization failure fixed
Pushing strip unit Please check whether pushing strip unit’s connection and plug-in connection are
4
initialization failure fixed
Detection unit Please check whether detection unit’s connection and plug-in connection are
5
initialization failure fixed
Print unit Please check whether print unit’s connection and plug-in connection are fixed
6
initialization failure
7 Lack of cleaning fluid Add the cleaning fluid and press the "start" button to retest
The test tube shelf is not Rearrange the test tube rack and press the "start" button
8
in place
The direction of the test strip is not correct. Re-place the test strip according to
test strip placement mark instructions;
Check if there is a test paper strip or block stuck in the sealed cabin;
9 The sealed cabin is stuck
Check whether the test paper strips are neatly placed
The probability of occurrence of card is less than 1% is a normal phenomenon.
Please remove the test strip and then place the test paper into sealed cabin
Test paper sealed cabin
10 Add the test strips
without paper
The box of the waste Dump the waste paper
11
paper is full
The probe up-and-down
12
motion failure Please contact the supplier
The probe front and back
13 Please contact the supplier
motion failure
14 The injection pump failure Please contact the supplier
Up and down electrical Please contact the supplier
15 machinery optocoupler
failure
Electrode manipulation Please contact the supplier
16 electrical machinery
optocoupler failure
41 / 43
Biobase Biodustry (Shandong) Co., Ltd

Pushing the test paper Please contact the supplier


17 strip electrical machinery
optocoupler failure
Couldn't find the paper Check whether the test paper strip is stuck in the paper seal cabin
18
strip If you still can not solve, please contact the supplier
Test paper selection box The direction of the test strip is not correct. Re-place the test strip according to
19
failure test strip placement mark instructions
The instrument did not Don't test your instruments in direct sunlight
20
detect the barcode
White reference value Wipe clean white base surface with absorbent cotton. If you still can not solve,
21
error please contact the supplier
Emergency ports are not in Push the emergency port to lock position
22
place
There was no sample for Place the sample in the emergency port
23
emergency port
24 Luminous tube failure Please contact the supplier
Detection electrical Please contact the supplier
25 machinery optocoupler
failure
Temperature exceed the Do not test the instrument in a high temperature environment
26
applicable range
Please clean select strip Use a brush to remove broken paper and debris in the box
27
box Check whether the test paper strips are neatly placed
Lay down the box, remove the bedplate, clean the surface with cleaning fluid.
28 Please clean the bedplate
When you put it back, pay attention to the white base position.
29 Printer without paper Add the paper
30 In/out frame is full Remove the test tube rack from the rack
31 The test strips is used up Use anti-counterfeit IC card to recharge the number of test paper
32 Cleaning tank full Turn off and clean the tank
Test paper strip placement Please contact the supplier
33 electrical machinery
failure
The device is not Please contact the supplier
34
initialized
The main board battery is Replace the main board battery
35
not enough
36 IC card is invalid Please use an effective anti-counterfeiting IC card
Error in IC card insertion Please insert the IC card correctly
37
card position
Insufficient of IC card The number of test paper in the instrument is running out, please add strip
38
number
39 TF card reading failed Please contact the supplier

42 / 43
Biobase Biodustry (Shandong) Co., Ltd

Appendix E

Reagent consumption

Operating categories Unit of cleaning liquid ( ml )


Starting up cleaning 16.5

Start to clean 14

Normal test (1 sample) 3

Washing pipeline 33

Shutdown cleaning 16

43 / 43
BIOBASE GROUP
2# building, No.9 Gangxing Road, High-tech Zone, Jinan City, Shandong Province,
China
Tel: +86-531-81219803/01
Fax: +86-531-81219804
Inquiry: export@biobase.com
Complaints: customer_support@biobase.cc
After-sales service: service_sd@biobase.cc; service_ivd@biobase.cc
Web: www.biobase.cc/www.meihuatrade.com / www.biobase.com