You are on page 1of 9

REMOVABLE PROSTHODONTICS

SECTION EDITORS
LOUIS BLATTERFEIN 5.. HOWARD PAYNE

Management of loading forces on mandibular


distal-extension prostheses. Part I: Evaluation of
concepts for design
\
Brian D. Monteith, M.Ch.D.*
Medical UniverGty of Southern Africa, Faculty of Dentistry, Medunsa, Republic of South Africa

T he two structures that support a mandibular distal-


extension removable partial denture differ markedly in
tissue-borne denture base. In the curve on the left, the
intrusive range is about 500 pm, which is 25 times
their viscoelastic response to loading. The differential greater than the measurement of 20 pm obtained when
between the 500 pm resilience of the residual ridge the same force was applied to teeth (Fig. 3). However,
tissues and the 20 pm of the teeth permitted by the an important feature common to both curves is that the
periodontal ligament’ presents a disparity of support response to load takes place in two distinct phases. Phase
that is in contrast to the uniform support accorded a 1 occurs instantaneously with the application of the load
tooth-supported removable partial denture (Fig. 1, A). and reveals a typically elastic response. In contrast,
Hence the denture tends to rotate about its most distal Phase 2 is a much stiffer phase characterized by an
abutments, inducing heavy torsional stresses on, and alteration in the shape of the curve to a typically
possible traumatization of the ridges (Fig. 1, B). viscoelastic pattern, which suggests a sudden marked
Design philosophies aimed at minimizing this rota- increase in tissue resistance.
tional movement have appeared regularly in the dental In similar fashion, the recovery of the system also
literature over the past 50 years. The fact that they takes place in two distinct phases. When the load is
continue to do so indicates that there is no unanimity as removed, there is an instantaneous elastic-type recovery,
to what might be considered the best method of solving which spans about 50% of the displacement range. The
the problem. These design concepts can be classified into second phase is of a more viscoelastic nature; and
four basic categories. although abrupt in onset, it suggests a slow and pro-
tracted return to the baseline.
1. The flexible denture base. This concept is represented by
stress-breaker designs. Effect of rate and duration of loading
2. The floating denture base. In this concept a mucostatical-
ly recorded denture base is related to the abutment teeth The curve on the right in Fig. 2 illustrates the effect
under pressure.‘,’ that a change in the rate of a force application of 4 N has
3. The mucofunctional concept. This approach includes the on the tissue displacement pattern. When the rate of
impression techniques that record the tissue surfaces in application is increased from 4 N/set to 100 N/set, the
the shape that the residual ridges assume under function- reduced time element results in a 20% reduction in the
al loading. amount of displacement that occurs and a much quicker
4. The endosseous implant. The proponents of this concept return to the baseline. Increasing the duration of the
seek to eliminate all movement of the denture base by
force has just the opposite effect. The same studies
reconstituting the missing distal abutment.
have shown that a static load applied for 10 minutes
The purpose of this article is to examine these gradually depressed the tissues further into the resis-
concepts in the light of recent research into the viscoelas- tant second phase. This is called tissue creep, which if
tic behavior of the teeth and mucosal tissues, as well as prolonged will not only push the system to its viscoelastic
that of endosseous implants as reported by Wills and limit, but will result in the phase of recovery becoming so
Manderson4 and Picton and Wills.5*6 extended as to require 3 or 4 hours to return to its
baseline.
VISCOELASTIC BEHAVIOR OF TEETH AND
MUCOSA Effect of area of loading
Pattern of tissue response to loading Altering the unit area of load application has a
Fig. 2 illustrates the mucous membrane displacement definite bearing on the amount of displacement that
curve that resulted when a force of 4 N was applied to a takes place. One might expect, for example, that halving
the area of tissue covered by a denture base from 150
*Professor <andHead, Department of Prosthetic Dentistry. mm’ to 75 mm2 would have the arithmetic effect of
THE JOURNAL OF PROSTHETIC DENTISTRY 673
MONTEITH

so0 pu
RECOWZRY 400

500
n I

Fig. 2. Mucous membrane displacement curves that


result from application of a force of 4 N at rates of 4
N/set and 100 N/set, respectively. Note that displace-
ment occurs in two distinct phases and that there is less
displacement when rate of force application is 100
N/second combined with quicker return to baseline
when load is removed. (Redrawn from Manderson,
R.D., et al.‘)

TOOTH

B
Fig. 1. A, Mesial and distal abutments provide uniform
support for tooth-borne removable partial denture or
fixed prosthesis. B, Mixed support arises from loss of
distal abutment. Because residual ridge tissues are 25
times more resilient than periodontal attachment appa- Fig. 3. Typical displacement curve resulting from
ratus, removable partial denture under vertical load will application of axial load of 4 N at a rate of 4 N/set to
tend to rotate around a fulcrum point that coincides tooth. Note that both loading and recovery phases
with its most distal occlusal rest. feature distinct elastic and viscoelastic elements. (Re-
drawn from Manderson, R.D., et al.‘)

doubling the amount of tissue displacement under the sagittal plane is such that the denture base portion, in
same load. Surprisingly the studies under review show effect, becomes a beam that under load behaves like the
that halving the area of support in this way results not in moving radius of a circle. The distal extremity of the
a twofold increase but a geometric fourfold increase in denture base will scribe an arc (Fig. 4), and because it is
tissue displacement. furthest from its center of movement at the hinge will
consequently undergo more movement than any other
FLEXIBLE DENTURE BASE CONCEPT point along the length of the denture base. Because they
Hailed by some and decried by others, stressbreakers are subject to geometric principles, the arc lengths
provide a means of interposing a flexible connection scribed by all points along the length of the denture base
between the tooth-borne retainer portion of a removable will reflect a gradient that is governed by a strictly linear
partial denture and its distally extended tissue-borne differential. As Vahid? demonstrated, the tissue dis-
base. Although many types of stressbreakers are avail- placement along the crest of the mandibular residual
able, the principle that underlies their use is aimed at ridge presents a similar differential; that is, the tissues
ensuring that all forces that fall on the denture base immediately adjacent to the last abutment tooth are far
portion will be directed toward the residual ridges alone less resilient than those in the region of the retromolar
and thus protect the teeth from possible torsional effect. pad. The ridge tissues, however, do not behave according
The vertical movement permitted by a hinge in the to a geometric system, so it is most unlikely that the
674 NOVEMBER 1984 VOLUME 52 NUMBER 5
CONCEPTS FOR DESIGN

Fig. 4. Flexible denture base removable partial denture


demonstrating steep arc differential when location of
stressbreaker hinge results in short-radius beam.

J IV I

Fig. 5. Long-radius design incorporating split major


connector. Locating point of resilience as far anteriorly
as possible reduces arc differential to a minimum.

tissue displacement differential along the length of the


ridge will coincide with the strictly linear differential Fig. 6. Evidence of trauma in region immediately distal
imposed by the denture base. The inevitable result, to last abutment tooth. Because it displays less compress-
therefore, will be an uneven match of stress and resil- ibility than tissues elsewhere, this region bears the
ience, and the areas that are least resilient will be made brunt of functional stresses of prosthesis made from
to bear most of the load. In terms of the “square rule” mucostatic impression.
relating displacement to area, this limitation of the area
under load might induce tissue creep and the risk of occlusal rests to bind on their seats and transmit torque
insidious damage to the underlying alveolar bone. to the abutment teeth, Hindel2 described a technique to
Efforts to reduce the arcing differential to less stressful induce a vertical floating movement in the denture base,
gradients have resulted in “long radius” designs such as in which the rest will contact the base of its seat only
that of Levin.* Long radius designs seek to lengthen the when the ridge tissues that support the denture base are
active radius of the system by ensuring that the flexible maximally compressed. As a result, the masticatory load
joint is located as far anteriorly as possible. This will be divided between the ridges and the abutment
frequently involves the use of a split major connector that teeth. On release of the load, the mucosa will rebound,
is two separate lingual bars joined anteriorly in a the entire structure will move upward, and the occlusal
manner that will permit a degree of resilience between rests buoyed up by this movement will be prevented from
the denture base and the retainer portions of the acting as fulcrums. This technique incorporates a
framework to which they are attached (Fig. 5). Length- mucostatic impression of the ridges related by means of
ening the beam radius in this way could reduce the finger pressure to the abutment teeth.
arcing differential to a gradient more in keeping with the In measuring this philosophy against the biophysical
inherent displa.ceability of the ridge tissues. It is more concepts outlined earlier, it is helpful to examine the
likely, however, that clinical successes attendant on underlying principle of a mucostatic impression related
the use of these devices are to be attributed to the in- under pressure. Applegate9~‘o stated the futility of
herent ability of biologic systems to tolerate the rudely mucostatic impressions for removable partial dentures.
mechanical concepts so frequently imposed on them His reasoning was that the ridge tissues are not uniform-
by dentistry. ly compressible and that a base made from mucostatic
impressions will result in the firmer areas doing all the
FLOATING DENTURE BASE CONCEPT work, and the more compressible areas receiving little or
Recognizing that a distal-extension denture base no work stimuli. Vahidi’ showed conclusively that the
sinking into the underlying ridge mucosa will cause ridge tissue immediately distal to the last abutment tooth
THE JOURNAL OF PROSTHETIC DENTISTRY 675
MONTEITH

Fig. 8. Accurately fitting clear acrylic resin denture


bases have been adapted to metal framework prior to
Fig. 7. Metal framework constructed on master cast making corrective impression.
obtained by means of irreversible hydrocolloid impres-
sion material.
MUCOFUNCTIONAL CONCEPT
The rationale of this concept is that the soft tissues of
in a distal-extension prosthesis is far less compressible the ridge derive benefit from the massaging effect
than the tissues elsewhere. This means that a mucostati- produced by the rhythmic displacement of blood fluids
tally shaped base would tend to “ride” on this relatively from the superficial vessels during function.9 In terms of
small region. As with stressbreakers, in terms of the the tissue displacement curve (Fig. 2), this vascular
square rule the concentration of functional forces on so displacement coincides with the more elastic first phase
small a region will invite stress of potentially damaging of compression prior to the sudden onset of increased
proportions (Fig. 6). tissue resistance that marks the start of Phase 2. The
The ideal of having the rest contact its seat at the objective of the floating denture base concept was to
precise moment at which the tissues are maximally effect tooth support at the very moment that the ridge
compressed seemsto be an admirable one. Will this ideal tissues become maximally compressed. To take this
state be attainable through the application of arbitrary literally however, would presuppose that this is made to
finger pressure? Masticatory pressures have been ascer- occur at the end of Phase 2, because it is only then that
tained to be approximately 4 N applied at a rate of 100 the tissues are maximally compressed. It is more logical
N/set.’ Therefore, even if the fingers could be schooled to time the seating of rests or attachments to coincide
into consistently applying a force of 4 N, the problem of with the end of Phase 1, because the beneficial massag-
the time factor will remain, that is, the rate and duration ing effect will be retained without placing the event of
of the application. With the floating denture base rest contact beyond the mucocompressive capabilities of
concept, the force is applied through the finger holes in normal function. At first this seems almost impossible to
the impression tray as long as it takes the impression achieve, because the ridge is not uniformly compressible,
material to set. This will involve compression of the and the amount of movement represented by Phase 1 of
mucosal tissues lasting a number of minutes: an imposi- the curve will vary in magnitude from place to place. In
tion that will implicate the added danger of tissue creep. other words, there is not just one curve to be accommo-
The functional implications of this will be that a patient dated, but an infinite number, each of which represents
who exerts compression of 4 N at a rate of 100 N/set the degree of tissue compressibility as it occurs at
will find it impossible to match the amount of compres- different points over the entire ridge. Ideally, if it were
sion gained by finger pressure applied over a number of possible to capture the ridge profile at a stage at which
minutes. The denture, therefore, would remain totally each curve reached the end of Phase 1 and is beginning
and perpetually tissue-borne; and until ridge resorption to feel the resistance at the start of Phase 2, this would be
occurs, the occlusal rests or precision attachments of the the ideal point at which to introduce tooth support.
prosthesis will have little chance of ever contacting their Unlikely as it may seem, this can be achieved simply by
seats. superimposing one kind of elasticity on another through
676 NOVEMBER 1984 VOLUME 52 NUMBER 5
CONCEPTS FOR DESIGN

Fig. 9. Corrective impression of ridges obtained with


soft ZOE impression paste.

Fig. 11. Metal framework with associated corrective


Fig. 10. Kidge regions are removed from uncorrected impression repositioned on tooth portion of master cast
master cast with saw. prior to altering distal-extension bases.

the choice of an impression material that possesseselastic and the arbitrary finger pressures of the floating denture
properties similar to those of the ridge tissues. The base technique, mucofunctional philosophy appears to
material should be sufficiently stiff to induce tissue offer the only way in which the entire denture-bearing
compliance over the first phase of the displacement region can be loaded in a uniform way. By relying
curve, yet not s,ostiff as to overcome the onset of tissue exclusively on the elastic resistance of the impression
resistance at the start of Phase 2. material to achieve the required amount of tissue dis-
In this manner, the entire ridge mucosa may be placement, there will be less chance of the tissues being
optimally displaced to an extent dictated by the discrete stressed beyond their capabilities. The even nature of the
displaceability of its constituent regions, with the whole resulting displacement and the fact that it is spread over
reaching a state:of resistance to load at exactly the same the entire contacting surface seemsto satisfy the require-
moment. Impression waxes have been formulated to this ments of the square rule relating tissue compression to
end, such as those that form the basis of Applegate’s area in a way that seems unlikely by any other means.
“altered-cast technique.“‘, ‘O Other materials have been
investigated,“, “’ and zinc oxide-eugenol (ZOE) paste, ENDOSSEOUS IMPLANT CONCEPT
once considered to be the basis for essentially mucostatic Another school of thought is that by reconstituting a
impressions,13 h.as viscoelasuc properties that are ideally missing distal abutment, a uniformity of support suitable
suited to this purpose (Figs.. 7-12). for fixed partial dentures can be achieved that eliminates
Of the various philosophies reviewed, only the muco- the need for a removable partial denture. If one accepts
functional concept appears to satisfy the criteria against the precept that an endosseous implant represents a
which all three have been measured. In contrast to the perfect tooth analogue, this concept is attractive, partic-
uneven stressing of isolated ridge regions that might arise ularly to those who consider a fixed prosthesis superior
from the geometric imposition of stressbreaker designs to a removable partial denture. Both these assertions are
THE JOURNAL OF PROSTHETIC DENTISTRY 677
MONTEITH

DEHISCENCE

Fig. 12. Final altered cast with ridges corrected to


conform with functional profile. Fig. 13. Conventional blade vent design demonstrat-
ing tendency of shoulder portion to undergo dehis-
cence.
open to debate; however, they reflect an attitude that
may underlie the enormous increase in the use of
endosseous implants. counter an observed tendency toward bone loss around
In his review of the literature, Kapuri4 has drawn the shoulder regions of blade vents (Fig. 13). By
attention to the scarcity of documented reports aimed at permitting the shoulder to be relocated at a deeper level
establishing success or failure rates in contrast to the within the bone (Fig. 14), the anchor form has the
profusion of articles that deal with matters of technique, arguable result of rendering bone resorption, if not less
design modifications, and isolated successful experiences. likely, at least less apparent.*
It was on the grounds of insufficient sample size and
conflicting results that the Harvard Consensus Confer- Materials research
ence15had to admit that it was unable to assessspecific The efforts of the materials scientists have been
survival estimates of the free-end blade implant, even in directed toward the search for substances sufficiently
its simplest form. It is understandable therefore, that the biocompatible to elicit the least possible reaction when
Conference advised that implants should not replace implanted in the body. Details of this quest have been
conventional dentistry but be used only when conven- furnished by Kapur14 and Lavelle et al.” in their succinct
tional dentistry is unsuitable.15 With its plea for a more yet comprehensive reviews.
properly constituted gathering of data, this reflects a A significant precept of the search for the perfect
more cautious attitude. As Armitage16 pointed out at the implant material is that the attainment of a state of
Conference, “. . . many modifications have taken place to ankylosis between implant and bone is consistently held
improve blade implants. However, all have proven to be the characteristic sign of physiologic compatibili-
inadequate to justify such widespread use of this device ty* ” B Y contrast, the development of a fibrous connec-
today . . . . The rapid increase in the use of this implant tive-tissue sheath at the implant/bone interface is viewed
without documented success is alarming. . . [and its] as evidence of incipient failure and has served as an
continued use . . . requires material and design modifica- incentive to further and greater scientific endeavor.
tions to elicit a more favorable, longstanding inter- However, it is not only compatibility that is at issue in
face.” the occurrence of this fibrous phenomenon, but also the
Significantly, the twin matters of design and materials matter of implant loading. ‘a The appearance of fibrous
research, as well as the nature of the implant/bone tissue at the interface suggests that biocompatibility
interface, have attracted the attention of dental research- alone is not enough to ensure implant viability. Biocom-
ers and material scientists for more than a decade.14 patibility is a passive concept of toleration by the tissues
of something that does nothing to provoke its own
Design development rejection. Consequently, the search has extended into the
Design development has tended toward the empirical;
an example is the introduction of the anchor design to *Mohammed, H.: Personal communication, 1981.

678 NOVEMBER 1984 VOLUME 52 NUMBER 5


CONCEPTS FOR DESIGN

ENDOSSEOUS IMPLANT

______ RecovERY
;No” viscous

LOAD fSN/secj

Fig. 15. Displacement curve obtained by applying hor-


izontal force of 5 N/set to endosseous implant in vivo.
Elastic nature of displacement is attributable to bending
of implant neck. Absence of viscoelastic element sup-
ports view that endosseous implants should not be
Fig. 14. Anchor design, which by dispensing with regarded as tooth analogues. (Redrawn from Picton,
shoulder, ostensibly obviates possibility of dehiscence. D.C.A., and Wills, D.J.“)

realms of bioreactivity for a material that goes further of 1 to 2 pm (Fig. 15). Puzzled by this anomaly, Picton
than mere toleration by the tissues to one that actively and Wills6 immobilized a similar implant blade in a vise
stimulates a process of biologic integration within the and applied the same loads. The results were identical,
bone. which led them to conclude that the deflection in both
One way in which this criterion has been met involves instances was due to bending of the implant stem, and
coating metal implants with a ceramic tricalcium phos- that it should not be ascribed to displacement of peri-
phate. This substance has been found to react with the implant tissue.
bone within a freshly prepared site in a way that results It is clear from these findings, and can be seen in a
in the implant becoming “locked” within the bone of the comparison of Fig. 15 with Fig. 3, that a blade implant
jaw. ” A similar effect has been achieved with titanium does not react in the same way as a tooth when placed
oxide.” Both methods result in a direct bone/implant under load. Consequently, it cannot be regarded as a
interface with the total exclusion of the unacceptable true tooth analogue.
fibrous tissue layer. The use of bioreactive materials This places the rationale of free-end blade implants in
such as these appears to represent the most promising serious doubt. Intended as a way to avoid the displace-
advance in design and materials research. ment differential between tooth and mucosa that is so
troublesome in distal-extension removable partial den-
Response to loading tures, the free-end blade as a substitute tooth merely
It is possible that the ultimate cause of endosseous results in the creation of a new differential between tooth
implant failure may have little to do with design and and implant. Admittedly, the difference between the 20
materials. The work done by Picton and Wills6 in the km viscoelastic deflection of a tooth and the almost
investigation of the loading response to blade implants negligible elastic deformation of an implant stem is
suggests that it is in the realm of biomechanics that the minute. However, the proportion between the two is
answers should be sought. To put it another way, the almost the same as that which exists between a tooth and
ultimate success or failure of an endosseous implant the tissues of the residual ridge.
might depend not so much on its shape or material, but The matter is further complicated when tooth and
on what the dentist puts on top of it. implant are immovably united by means of a fixed
Picton and Wills6 extended their work on displace- partial denture. Acting as a beam, under functional load
ment determination to include blade implants. A force of it becomes in effect a pernicious lever. At one end it is
several newtons applied vertically to an implant post anchored immovably to a tooth that is capable of an
produced a negligible amount of intrusion. This observa- intrusive movement of 20 pm, while at the other end it is
tion led them to conclude that from a functional point of anchored immovably to an endosseous implant post
view, blade implants can be regarded as being ankylosed capable of no intrusive movement at all. On the applica-
to the surrounding alveolar bone. The application of a tion of vertical load (Fig. 16), the tooth at one end of the
horizontal load, however, resulted in an elastic deflection lever will intrude back into its socket, while at the other
THE JOURNAL OF PROSTHETIC DENTISTRY 679
MONTEITH

Fig. 17. Unwarranted enthusiasm for a clinical activity,


where more properly extreme caution ought to pre-
vail.
Fig. 16. Movement differential between tooth and
free-end blade is taken up in bending of implant neck.
Resultant torsional forces are expressed as compressive and thus were preserved from the deleterious effects of
stresses in alveolar bone with probable resorptive con- tooth/implant differentials.
sequences in regions indicated. Osseous integration procedures, on the other hand,
bury abutments for an initial period to allow full
ankylosis to develop before risking exposure to function-
end the implant is incapable of similar intrusion and will al forces. The implant philosophy presented by Lavelle
comply by simply bending. In obedience to the third law et al.” makes use of this preparatory measure and seeks
of Newton, the torsional stress set up within the implant to preserve the resulting ankylosis through the use of an
neck by this bending will elicit a reaction at the epimobile polysulphone cone, which should be of value
extremities of the implant blade. The implant will in absorbing differential stresses. Having gone to such
become a force transducer, and the stress built up within lengths to avoid the development of fibrous tissue at the
itself will be transmitted to the surrounding alveolar interface, they are uncompromising in their condemna-
bone in a compressive way. The inevitable consequence tion of it and state, “The notion that a layer of
of compressive stresseson bone is osteoclastic activity, so disorganized connective tissue around an implant is
it is hardly surprising that regions of bone resorption are analogous to the highly organized periodontal ligament
a frequent feature of this modality. is fallacious. Indeed, the presence of connective tissue
The inference is that far from presenting a way to around an implant heralds its demise.”
avoid the problem of mixed support in a distal-extension Unfortunately and ironically, this controversy will
prosthesis, the implant-fixed partial denture-tooth com- probably never be resolved. The Harvard Consensus
bination merely substitutes another form of mixed Conference (recommendation 30) ranked mobility as of
support between the tooth and the implant. first importance in the evaluation of free-end fixed
prostheses and apportioned this criterion a value of 0.8
Clinical comments on a scale of l.*’
The matter of fibrous connective tissue at the interface This is admirable. Unfortunately, in listing its criteria
will remain unresolved as long as there are people who for implant success (recommendation 15), the Confer-
continue to regard it in the light of an “implant- ence deemed the presence of movement less than 1 mm in
periodontal ligament.” Weiss and Rostoker*” reported any direction in an implant to be grounds for self-
the progress of a new type of endosseous implant in congratulation. The proponents of absolute ankylosis
baboons. They welcomed the appearance of a fibrous disagree with this concession. Should the latter be correct
tissue interface on their devices and observe that this in their assertion that any movement in an implant is a
“ . . . allowed some desirable ‘micromovement’ (approxi- harbinger of certain failure, it is unlikely that their
mating that of a natural tooth) without causing bone rectitude will ever be acknowledged. The existence of a
resorption.” It might be argued that the lack of further set of standards against which failure can be rated as
bone change could be ascribed to the fact that none of the successwill give the free-end endosseousblade a respect-
functioning implants was attached to an adjacent tooth ability that it has not earned and promote unwarranted
NOVEMBER 1984 VOLUME 52 NUMBER 5
CONCEPTS FOR DESIGN

enthusiasm for a clinical activity where, more properly, loading on partial denture movement. J PROSTHETDENT 15:474,
1965.
extreme caution ought to prevail (Fig. 17).
12. De Sousa, V., Pellizzer, A. J., Plese, A., Castleberry, D. J.,
Miller, E. L., and Jackson, E. L.: Support area in lower free end.
SUMMARY
Rev Fat Odont Araptuba 6:21, 1977.
The viscoelastic reaction of ridge mucosa and abut- 13. Kramer, H. M.: Impression technique for removable partial
ment teeth, by virtue of their periodontal ligaments, is a dentures. J PROSTHETDENT 11:84, 1961.
necessary dimension in the evaluation of design concepts 14. Kapur, K. K.: The literature on blade implants. In Schnitman, P.
A., and Shulman, L. B., editors: Dental Implants: Benefit and
for removable partial dentures for patients with distal- Risk. An NIH-Harvard Consensus Development Conference.
extension ridges. Bethesda, Md., 1980, U.S. Department of Health and Human
The possible modalities are (1) removable partial Services, Public Health Service, National Institutes of Health, pp
dentures with flexible denture bases (stressbreakers), (2) 240-260.
use of a fl0atin.g denture base impression technique, (3) 15. Schnitman, P. A., and Shulman, L. B.: Consensus on dental
implants. In Lefkowitz, W., editor: Proceedings of the Second
use of a mucofunctional impression technique to relate International Prosthodontic Congress. St. Louis, 1979, The C.V.
the denture base to the framework, and (4) use of an Mosby Co., pp 217-218.
endosseous implant. An evaluation of each modality has 16. Armitage, J. E.: Risk of blade implants. In Schnitman, P.A., and
been made. Shulman, L.B., editors: Dental Implants: Benefit and Risk. An
NIH-Harvard Consensus Development Conference. Bethseda,
REFERENCES Md., 1980, U.S. Department of Health and Human Services,
1. Manderson, R. D., Wills, D. J., and Picton, D. C. A.: Biome- Public Health Service, National Institutes of Health, pp 294-
chanics of denture-supporting tissues. In Lefkowitz, W., editor: 304.
Proceedings Iof the Second International Prosthodontic Congress. 17. Lavelle, C., Wedgwood, D., and Riess, G.: A new implant
St. Louis, 1979, The C. V. Mosby Co., pp 98-101. philosophy. J PROSTHETDENT 43:71, 1980.
2. Hindels, G. ‘W.: Load distribution in extension partial dentures. 18. Hobkirk, J. A.: Bone behavior around ceramic implants. In
J PROSTHETDENT 2:92, 1952. Lefkowitz, W., editor: Proceedings of the Second International
3. Hindels, G. W.: Stress analysis in distal extension partial Prosthodoniic Congress. St. Louis, 1979, The C. V. Mosby Co.,
dentures. J PROSTHETDENT 7:197, 1957. pp 229-233.
4. Wills, D. J.. and Manderson, R. D.: Biomechanical aspects of 19. Brinemark, P. I., Breine, U., Adell, R. O., Hanson, B. D.,
the support of partial dentures. J Dent 5~310, 1977. Lindstrom, J., and Chisson, A.: Intra-osseous anchorage of
5. Picton, D. C. A., and Wills, D. J.: Viscoelastic properties of the dental prostheses. Stand J Plast Reconstr Surg 3~81, 1969.
periodontal ligament and mucous membrane. J PROSTHETDEHT 20. Weiss, M. B., and Rostoker, W.: Development of a new
40~263, 1978. endosseous dental implant. Part I: Animal studies. J PROSTHET
6. Picton, D. C. A., and Wills, D. J.: The mobility of horizontally DENT 46:646, 1981.
pinned teeth in monkeys and of endosseous blade implants. Br 21. Schnitman, P. A., and Shulman, L. B.: Dental Implants: Benefit
Dent J 139:469, 1975. and Risk. An NIH-Harvard Consensus Development Confer-
7. Vahidi, F.: Vertical displacement of distal extension ridges by ence. Bethseda, Md., 1980, U.S. Department of Health and
dill’erent impression techniques. J PROSTHET DENT 40~374, Human Services, Public Health Service, National Institutes of
1978. Health, pp 326-339.
8. Levin, B.: Stressbreakers: A practical approach. Dent Clin North
Reprint requeststo:
Am 2.3~77, 1979.
DR. BRIAN D. MONTEITH
9. Applegate, 0. (1.: The partial denture base. J PROSTHETDENT
MEDICAL UNIVERSITYOF SOUTHERNAFKICA
5:636. 1955.
FACULTYOF DENTISTRY
10. Applegate, 0. C.: Essentials of Removable Partial Denture
P.O. MEDUNSA 0204
Prosthesis. Philadelphia, 1954, W. B. Saunders Co., pp 166-
REPUBLIC:OF SOUTH AFRICA
199.
1I Holmes, J. B.: Influence of impression procedures and occlusal

Bound volumes available to subscribers


Bound volumes of the JOURNALOF PROSTHETICDENTISTRY are available to subscribers (only) for the 1984 issues from the
publisher at a cost of $39.00 ($49.00 international) for Vol. 51 (January-June) and Vol. 52 (July-December). Shipping
charges are included. Each bound volume contains a subject and author index, and all advertising is removed. Copies are
shipped within 30 days after publication of the last issue in the volume. The binding is durable buckram with the journal
name, volume number, and year stamped in gold on the spine. Payment must accompany all orders. Contact The C. V.
Mosby Co., Circulation Department, 11830 Westline Industrial Drive, St. Louis, MO 63146, USA; phone (800) 325-4177,
ext. 351.
Subscriptions must be in force to qualify. Bound volumes are not available in place of a regular
JOURNAL subscription.

THE TOURNAL OF PROSTHETIC DENTISTRY 681

You might also like