You are on page 1of 2

Patient-ventilator asynchrony reference card

Asynchrony Description On the waveform Waveform example Common possible causes


Trigger asynchronies - during the beginning of inspiration
Flow [l/min]

The time interval between the Flow waveform: look for a longer- Trigger threshold set too high
25 1
Delayed patient’s inspiratory effort and than-normal time interval between Ventilator pneumatics
triggering the delivery of a mechanical the positive deflection in flow 1 and 0 Presence of AutoPEEP
2 Low respiratory drive
breath is increased the delivery of ventilatory support 2
-25 Weak inspiratory effort

Trigger threshold set too high


Flow waveform: look for an abrupt
Flow [l/min] Pressure support too high
change in the steepness of the Set frequency and/or inspiratory time too
The patient’s inspiratory effort 25 1 high (in controlled modes)
Ineffective waveform 1 (decrease in expiratory
fails to trigger the delivery of a Tidal volume set too high
effort flow or increase in inspiratory flow) 0
mechanical breath Presence of AutoPEEP
that is not followed by ventilatory 2
support 2
-25 Low respiratory drive
Weak inspiratory effort
Sedation
Paw [cmH20]

Pressure waveform: look for a 20 Inspiratory trigger sensitivity set too high
A mechanical breath delivered 1 Air leaks in the endotracheal tube cuff,
Auto delivered mechanical breath showing
without an inspiratory effort 10 ventilator circuit or chest tube
triggering no drop in airway pressure 1 at the
beginning of the inspiratory phase Flow oscillations (water or secretion in
0 the circuit, cardiac oscillations)

Flow asynchronies - during the gas delivery

Inappropriate selection of ventilation


mode (more frequent in volume-
Paw [cmH20]
controlled modes)
The delivered flow does not Pressure waveform: look for an
20 High inspiratory effort
Flow
meet the patient’s inspiratory upward concavity 1 preceding the
asynchrony 10 In volume-controlled modes:
flow demands end of the mechanical breath
Inappropriate flow settings
1
0
In pressure-controlled modes:
Inappropriate P-ramp settings

www.hamilton-medical.com/asynchronies
Asynchrony Description On the waveform Waveform example Common possible causes
Termination asynchronies - during the end of inspiration

Cycling criteria (ETS) set too high


Pressure support too low
Flow waveform: look for two assisted Flow [l/min]
P-ramp too short
breaths without expiration between Flow starvation
Two (or more) mechanical 25
Double them or with an expiration interval of High respiratory drive
breaths are delivered during
triggering less than half of the mean inspiratory 0 Time constant too short
one single inspiratory effort
time (often visually displayed as a
waveform with two inspiratory peaks) -25 Double triggering can be an effect of and/
or promoted by reverse triggering or early
cycling

Flow [l/min] In pressure support ventilation:


Flow waveform: look for a small Cycling criteria (ETS) set too high
The duration of the mechanical 25
bump 1 at the beginning of Low levels of ventilator pressure support
Early breath is shorter than the 1 2
expiration (after peak expiratory flow) Time constant too short
cycling duration of the patient’s 0
followed by an abrupt initial reversal 1
inspiratory effort
in the expiratory flow 2 In time-cycled ventilation:
-25
Short inspiratory time setting

In pressure support ventilation:


Cycling criteria (ETS) set too low
Flow [l/min] 2 Pressure support too high
P-ramp too long
Flow waveform: look for a change 25
The duration of the mechanical
in the slope of the inspiratory flow: 1 In pressure control ventilation:
Delayed breath is longer than the
a fast decrease 1 followed by an 0 Cycling criteria (ETS) set too low
cycling duration of the patient’s
exponential (less steep) decline 2 Inspiratory time too long
inspiratory effort
-25
In volume control ventilation:
Low flow
High inspiratory time
High tidal volume

ELO2019041N.01
‘Correct’ waveform, in case of good patient-ventilator synchrony | Patient factors | Ventilator-related factors | Patient-ventilator interface

References
Arellano, D.H., Palliative Medicine & Care. 2017;4(4):1-6. DOI: 10.15226/2374-8362/4/4/00147 ¦ Arnal, J.M., Monitoring Mechanical Ventilation Using Ventilator Waveforms, Springer International Publishing. 2018. ISBN: 9783319586540) ¦ Dres, M., Current opinion in critical care. April 2016;22(3):246-253. DOI:
10.1097/MCC.0000000000000307 ¦ Holanda, M.A., Jornal Brasileiro de Pneumologia. 2018;44(4):321-333. DOI: 10.1590/S1806-37562017000000185 ¦ Mellot, K.G., Critical Care Nurse. 2009 December; 29(6): 41–55. DOI: 10.4037/ccn2009612.

You might also like