Voltajes Fuente
Voltajes Fuente
TP1 FR9H1304_E.ai
MD-38
3A +15V(F) 1.6A
1.6A Start-piont
-15V detection board
1.5A -15V(H)
Image 1.6A SYN12A
processing board
AUX12V +12V IMG17A/B
+5V(S) 1.6A
TP1
CPU board
CPU91A
1.6A IP leading-edge detection board : SED23A
FR9H1305.ai
MD-39
3.2 Erasure Unit Control 3.2.2 Components of the Erasure Unit Control (#20001 or later)
- The control circuit for the erasure lamp unit is located on the SNS17B board. A
3.2.1 Overview of the Erasure Unit Control control signal from the SNS17B board causes the relay circuit and lamp inverter
circuit of the driver board (INV17A/B) to operate, thereby turning ON or OFF the
(#20001 or later) lamps and fans.
The erasure unit provides idle illumination control, IP erasure control, and fan control. - To enable discrimination between failures of the control board (SNS17A) and driver
board (INV17A/B), LEDs are provided on the SNS17B board to indicate the status
Feature: Erasure unit initialization is effected after verifying that the machine is of the control signal.
emptied of IPs.
{MD:3.3.3_Correspondence between LEDs and Erasure Control Signals and
Idle illumination control FFM Control Signals}
To obtain an adequate lamp light intensity for IP erasure after RU power ON, lamp
(LAMP1 - LAMP5) illumination and erasure unit cooling fan (FANE1) operation are Power supply unit
controlled so as to maintain a constant temperature inside the erasure unit. PSU17A
+5V(A) +24V(E)
CN1 CN2
IP erasure control
To effect IP erasure, the erasure unit pass time is controlled in accordance with the Sensor board CN20
dose received by a processed IP. SNS17B FANE3 FANE2 LAMP2
Safety CN5
TSWE1 FANE1 LAMP1 LAMP3
Fan control thermostat
control circuit CN9 CN7 CN6 CN3 CN4 CN5
Control is exercised so as to operate the erasure unit cooling fan (FANE1) when the
temperature inside the erasure unit is high. LED F4 F1 F2 F3
Temperature adjustment control Lamp
CN2 Relay CN1
- The heat generated during lamp (LAMP1 - LAMP5) illumination is detected by the inverter
Lamp circuit
temperature thermistor (THE1) and used to exercise control of the erasure unit CN16 circuit
cooling fan (FANE1) operation and lamp illumination. temperature
control circuit Inverter board
- While the machine is idle and processing no IP, control is exercised to automatically INV17A CN8
illuminate the lamps for temperature adjustment purposes. Unlit lamp
detection
- Before the lamps are illuminated, they are subjected to preheating. circuit CN6
Inverter board
Temperature INV17B
control fan Lamp
(FANE1) CN2 Relay CN1
circuit inverter
control circuit circuit F1 F2
IP erasure control
Erasure unit pass time is controlled in accordance with the dose received by a read IP,
and the IP is erased.
LAMPB4, 6 LAMPB3, 5