Discovery RF180 GE Healthcare
LOTO Procedure Consignation 5789495-1EN, Rev 3
Page 2
GE Healthcare Discovery RF180
5789495-1EN, Rev 3 LOTO Procedure Consignation
Revision History
NOTE: This Lockout Tagout procedure is for the Discovery RF180 system only. Others systems could be
installed in the same room, and have a specific LOTO procedure.
Page 3
Discovery RF180 GE Healthcare
LOTO Procedure Consignation 5789495-1EN, Rev 3
The progress of LOTO procedure is described in the following flowchart and is detailed in the steps below.
Page 4
GE Healthcare Discovery RF180
5789495-1EN, Rev 3 LOTO Procedure Consignation
1. Inform users
2. Identify energies
Locate the electrical power distribution box located in the examination room and visualize status of all
magnetic switches:
Page 5
Discovery RF180 GE Healthcare
LOTO Procedure Consignation 5789495-1EN, Rev 3
a. Switch off the table and the generator by pushing down « OFF » button (01) on table remote
console.
b. Switch off the Image processor following the switch off procedure:
.
Turn in OFF position the line switch at the Image processor cabinet when you are in archive/worklist
screen. Wait for the full automatic shutdown.
Page 6
GE Healthcare Discovery RF180
5789495-1EN, Rev 3 LOTO Procedure Consignation
a. Disconnect arrival of electrical main power at all unit by pressing the OFF button in/on the power
distribution box.
OFF
BUTTON
5. Turn in OFF the UPS unit pressing the OFF button at the unit front side:
Page 7
Discovery RF180 GE Healthcare
LOTO Procedure Consignation 5789495-1EN, Rev 3
Remove the line input and output from the UPS unit, set OFF the main line switch S1 at the power
distribution box (Main line switch).
Install LOTO Tools at the electrical power distribution box located in the examination room:
a. Put Red Padlock with Red Label on plug at the front side of the power distribution box as shown in
the following figure:
Page 8
GE Healthcare Discovery RF180
5789495-1EN, Rev 3 LOTO Procedure Consignation
a. At the back side of the UPS unit remove the OUTPUT cables to the OPERA D4000RF image
processor.
b. At the rear side of the Table tower check that the line live lamp is OFF (red square). Using a DVM
check absence of energy at the table line input terminals as shown in the following figure:
Page 9
Discovery RF180 GE Healthcare
LOTO Procedure Consignation 5789495-1EN, Rev 3
c. At the generator cabinet remove the line protection and check, using a DVM, absence of energy at
the line terminals as shown in the following figures:
NOTE: When the system is consigned without any power, all the
movements are blocked.
Page 10
GE Healthcare Discovery RF180
5789495-1EN, Rev 3 LOTO Procedure Consignation
The progress of removal of LOTO procedure is described in the following flowchart and is detailed in the
steps below.
Page 11
Discovery RF180 GE Healthcare
LOTO Procedure Consignation 5789495-1EN, Rev 3
1. Remove Tools
a. Connect behind the UPS the input and output cables coming from the Poxer Box and the
Digital System
4. Inform users
a. Verify all commands are in neutral position by pushing on “OFF” button on console
Page 12
GE Healthcare Discovery RF180
5789495-1EN, Rev 3 LOTO Procedure Consignation
c. Turn in ON the UPS unit pressing the ON button at the unit front side:
d. Enable arrival of electrical main power at all unit by pressing the ON button in the power distribution
box.
ON
BUTTON
Page 13
Discovery RF180 GE Healthcare
LOTO Procedure Consignation 5789495-1EN, Rev 3
7. Switch on equipment
a. Switch on the table and the generator by pushing down « ON » button (02) on table remote console
for 2 seconds.
8. Finalization
Page 14
GE Healthcare Discovery RF180
5789495-1EN, Rev 3 LOTO Procedure Consignation
www.gehealthcare.com