INSTRUMENT BACKSIDE VIEW
CONNECTOR DESCRIPTION
2
3
Key Function
4
Switching the unit on and off
Increasing gain
Decreasing gain
5
LED
Symbol Description
A Gate alarm
R Function reject is active
D Function dual is active
Initial Start-Up
Power Supply
The TRU-SONIC can be operated with an external power supply through AC/DC adaptor or with built-in Lithium-Ion battery pack.
You can connect the TRU-SONIC to the main power supply if the instrument battery pack less the batteries power.
6
The battery pack is situated at the instrument back. The battery two side built-in sliding guide
Line and locked with screw.
The Lithium pack also can put on the instrument charge or remove charge without instrument.
The battery pack installed a DC charge receptacle connector same as instrument.
See following picture shoed how to remove and install the battery pack.
The instrument screen right bottom side showed symbol of battery power. If the battery-pack power less than 20%, the instrument will show “Low Battery”.
The instrument still can use approx. 20 min. when “Low Battery” alarm appear. The user need to switch off the unit and plug it to the main power supply or
charge the battery.
You can charge the Lithium-Ion Battery either direct in the instrument or by means of an external charge from charger.
Internal Charging
Requirement
- Lithium battery pack, our P/N
- Power supply/charger unit, P/N
7
If a battery is located in the instrument, the charging process is start automatically when you connect the plug-in power supply unit. You can carry out ultrasonic
inspection and charge a battery at the same time.
The charging time is approx. 10 hours with a simultaneous ultrasonic inspection.
If the instrument is not been used for ultrasonic inspectors, the charging time is approx. 8 hours. This charging time applies to ambient temperatures from 25-30
degree C, Please take into consideration that the battery pack are not charged to their full capacity at higher temperatures.
The LED display on the plug-in power supply unit indicates the status of the charging process. First plug the power supply to the main power ( 220 VAC or 110
VAC ), the LED on the charger will turn red into green. Then plug the power supply plug direct to the instrument receipt socket or battery charge socket. When
battery fully charged, the power supply charger LED indicator will turn red into green.
External Charging
Lithium-Ion batteries can be charged by means of an external charged. The Lithium battery installed a power receipt socket same as instrument. Use the same
way as we charged by the previous way..
Connecting UT Probes
The prepare of TRU-SONIC for operation, you have to connected a probe to it. Any brand probe can be use for the TRU-SONIC, provided the appropriate cable
is available and the operating frequency is within an adequate range.
The TRU-SONIC is available with the probe cable connectors Lemo 1.
The probe is connected to the socket at top right on the instrument casing. Both connector socket are equally suitable ( connected in parallel) for connecting
probes equipped with only one ultrasonic element (ultrasonic transducer) so that it does not matter which one of the two sockets is used.
When connected dual-element (TR) probe (having one transmitter element and one receiver element), or two probes (of which one is transmitting and the other
one receiving), attention should be paid to connecting the transmitter element to the right-hand sockets (transmitter, marked with black circle at the rear of the
instrument case) and the receiver element to the left-hand socket (receiver, marked with red circle).
Attention:
If this is not taken into account, the consequence would be a mismatching which may lead to considerable power losses or even to echo waveform distortions.
8
Starting the TRU-SONIC
Switching On TRU-SONIC
The system carry out a self-check and then switches over to stand-by mode.
The setting of all function values and the basic setting are same as before switching-on the instrument.
Reset
If any functions can not be operated after a warm start, or if you want to reset the instrument to the basic setup, then you should carry out a cold start by
simultaneously pressing the
and Key – the instrument will reset to the basic setup and clean all memory file
and key – the instrument will reset to the basic setup and keep all memory file
9
Switch off the keypad Bi-Bi sound
After press turn on TRU-SONIC, press again will turn off keypad sound
Principles of Operation
Operator’s controls
0
LED A: Gate alarm
R: Rejection on
D: Dual on
Screen Display
Note:”
The screen display always shows the gain and the adjusted dB step value.
10
Tru-sonic Display Sub-Menu Guide
First press GATE Second press GATE First press SET Second press SET
into BASE into a GATE into b GATE into DATE into COLOR into PULSE
RANGE a START b START HORN GROUND DAMP
0250mm 0050mm 0150mm OFF 100Ω
TRIG a WIDTH b WIDTH BRIGHT WORD POWER
S 0100mm 0050mm HIGH 400V
MTL VEL a THRESH b THRESH UNIT ECHO DUAL
5920m/s 88.0% 25.0% mm
D-DELY a LOGIC b LOGIC DATE GRID REF-dB
0000mm POS POS 11-28-06 +00dB
P-DELY ab GATE b GATE TIME SOLID PRF-MOD
0.00us OFF OFF 15:28 OFF 100Hz
11
AND
into PK MEM into SAVE into RECALL into RECV into TRIG into CAL into DAC
PK MEM NO. 000 NO. 000 REJECT ANGLE MEA FP DAC 3C
OFF PUSH 1 PUSH 1 00% 00.0° PEAK OFF
ENVLOP FILE NAM LOAD FREQU X-VALUE S-REF 1 DAC 4C
OFF 0000 PUSH 1 0.5-5M 10.0mm 50mm OFF
SHADOW FILE TYP TRANSIT RF-MODE THICKNE S-REF 2 AWS
OFF A-SCAN PUSH 1 OFF 000mm 50mm OFF
B-SCAN DATE FILE NAM AGB 50% DIAMET a START AVG
OFF 11-26-05 0000 OFF 10.0mm 0050mm OFF
TIME
T-SCAN FILE DEL FINE G b START RESET
34.6℃
OFF 13:48 PUSH 1 00 0150mm PUSH 1
12
The means of the 8 function group key are displayed at the bottom of the screen. The currently selected function group is display on left screen.
You can configure the four positions of the measurement line for set and measured values as required. Example of a measurement line
Ha = amplitude height Gate A(%)
Sa = sound path Gate A
Ra=
Da=
Ala=
Hb = Amplitude height Gate B (%)
Sb = Sound path Gate B
Sab=
Dab=
F = status indicator TOF = Flank
P= peak in gate
13
Selecting Display Unit
In the function group, press key, press unit display unit right < key,
use left side ^ key, change the unit display mm or inch.
Setting the date
In the function group, press key, press unit display date < key, use
left side ^ key, change the unit display date
In the function group, press key, press unit display time < key, use
left side ^ key, change the unit display time
14
The functions of the TRU-SONIC are combined to form function groups of three
operating levels
The function, operated via left 3 keys, enable to quickly and direct set the gain.
- Press the key to select the function showed left side.
- Press the key to increasing the setting
- Press the key to decreasing the setting
The gain control function is always directly available from left up key.
Press select a certain incrementation for setting the gain. , dB setting choice
from 1,2,6,12,20 dB 5 steps corase setting.
For the fine gain control, you can press receiver key , press key on right of
15
Adjusting the Display Range ( Function Group BASE )
The function group BASE enable you to make the basic adjustment of the screen display range. The display on RANGE
the screen must be adjusted for the material to be tested.( function MTLVEL) and for the probe used (function for
0250mm
P-DELAY).
Note: TRIG
In order to accurately adjust the material velocity and the probe delay, please read the section Calibration of S
TRU-SONIC.
MTL VEL
5920m/s
D-DELY
0000mm
P-DELY
0.00us
RANGE (Display Range)
You can adjust the display range if your measurement in range.
The coarse adjustment from 25, 50, 100, 200, 400, 800, 1600, 3200, 6000 mm. Press the RANGE right one time change from 25 change to 50, press
again from 50 change to 100,etc.
Any time when press the left side key, the range display will return back to the 25 mm, user can press range set another display range .
The fine adjustment can press BASE Key , press RANGE right key, then press key select the Fine range display.
16
MTVEL (Sound Velocity)
Use MTVEL to set sound velocity within the test object. You can use sound velocity between 1000-9999m/s
Coarse adjustment 5920 m/s or 3330 m/s. Direct press MTVEL right key to select 5920 m/s or 3340 m/s
17
D-DELAY ( Display Starting Point)
This function allow you to shift the complete screen display and consequently also the display zero. If the display should for example start from the surface of
the test object, the value in D-DELAY must be set to 0.
Press D-DELAY right key to select D-DELAY, Use key adjustment. Adjust the value for the display starting point.
18
P-DELAY (Probe Delay)
Every probe has a delay between the transducer element and coupling face. This means that the initial pulse must first pass through this delay line before the
sound wave can enter the test object. You can compensate for this influence of the delay line in the function P-DELAY
If the value for P-DELAY is not known, read the section calibration in order to determine this value
- Select the function P-DELAY
- Adjust the value for the probe delay use key to the right value
- Probe delay measurement, see attached example
19
Adjusting the Pulser ( Function group PULS )
This function serves for matching the probe. You can use it to adjust the damping of the probe’s oscillating circuit and
consequently change the height, width and resolution of the echo display DAMP
100Ω
- Low ( 200Ω), this setting has a lower damping effect and produces higher and broader echoes
POWER
- High (100Ω), this setting reduces the echo height but mostly also produces narrow echoes with higher resolution
- Select the function DAMP, 400V
DUAL
- Set the required value by means of key to the right value. REF-dB
POWER (Intensity) +00dB
Use the function POWER to set the pulser voltage. You can choose between two setting ( 400V and 200V). The setting high PRF-MOD
(400V) is recommended for all 100Hz
Inspections in which maximum sensitivity is important, e.g. for the detection of small flaws. Choose the setting low (200V) for
broadband probes or if narrow echoes are required (better latest resolution).
- Select the function POWER.
20
You can use the function DUAL to activate the pulser receiver separation.
- Select the function DUAL. DAMP
100Ω
POWER
- Use the key to the right value 400V
PRF-MOD ( Pulse Repetition Frequency) DUAL
- The pulse repetition frequency indicates the number of times an initial pulse is triggered per second. You can determine REF-dB
whether you need the highest possible PRF value, or whether you are satisfied with a low value. You 5 step selection +00dB
(100Hz,250Hz,500Hz and 1000Hz) PRF-MOD
- The larger you work piece, the smaller PRF value are needed in order avoid phantom echoes. In the case of the smaller 100Hz
PRF value, however, the A-scan update rate becomes lower, for this reason, high value are required if a work piece
should be scanned fast. The best way to determine the suitable PRF value is by experimenting; start from the highest
step and reduce the value until there are no such phantom echoes.
- Select the function PRF-MOD
If the Dual crystal or dual through-transmission probe is select, the LED D (dual) is on
21
Adjusting the receiver (function group RECV)
You will find all function for the adjustment of the pulser in the function group RECV.
- If required, go to the first operating level REJECT
- Select the function group by Press RECV key 00%
FREQU
REJECT 0.5-5M
RF-MODE
The function REJECT allows you to suppress unwanted echo indications, for example structural noise from your test object. OFF
The % screen height setting indicates the minimum height that the echoes should attain in order for them to be displayed on the AGB 50%
screen at all. The reject setting cannot be higher than the lowest threshold setting ( min 1%) of any gate. OFF
FINE G
You should handle this function with great caution, as it may of course happen that you suppress echoes from flaws as well. Many 00
test specifications expressly forbid using the reject function.
- Select the function REJECT
22
You have a choose between 2 frequency range
- Narrow Band from 0.5-4.0 MHz
- Board Band from 4-15 MHz
- Use the key to turn on the AGB 80% on, The echo amplitude will automatic increase up to 80% full screen
23
FINE G ( Fine adjustment of gain)
This function serves for the fine adjustment of the current gain value. The fine adjustment is possible 0.1db per step.
- - Select the function FINE G
24
Setting The Gates (Function Group aGAT and bGAT)
Display of Gates
All functions for setting the (dual) gate are arranged in the function group aGAT and bGAT.
- If required, go to the first operating level
- Select the function group aGAT or bGAT
TASKS OF THE GATES
It monitors the range of the test object where you expect to detect a flaw. If an echo exceeds or falls bellow the
gate, an alarm signal is output via the LED A.
The gate A and B are independent of one another. Gate A can also have the function of an echo-start gate.
The gate choose the echo for the digital time-of-flight or amplitude measurement. The measured value is indicated
in the measurement line.
Error alarms can be triggered under certain circumstances. These are caused by intermediate conditions in
instrument operation occurring when the
instrument is used. i.e. when function parameters are changed. Possible alarms occurring during instrument
operation (setting of function) are to be ignored.
25
To make the assignment easier, the gate are displayed in different colors. You can not vary the colors of gates
because they are fixed as following
Gate A – yellow
Gate B - Green
26
- - Select the function aTHRSH or bTHRSH
27
Setting “TRIG
- ANGLE=probe angle
- X-VALUE = Probe front side to index point distance. ( contact probe = 10 , angle beam probe need measurement by V1 bock )
Setting DAC
28
DAC CURVE SETTING
Prepare a 6 hole test block with different depth. Example drawing showed.
1. Calibrate the probe delay and set screen 175 mm
2. Press DAC enter the DAC submanual
3. move high-light cursor to DAC 3C and press up key turn on the DAC 3C on
4. Put probe on the test block and target to the first hole. Adjust the gain amplitude into 80%. Press * freeze key set point 1
5. Move the probe to second hole. Adjust the gate , press * set point 2
6. Move the probe to third hole. Adjust the gate , press * set point 3
7. Move the probe to no. 4 hole, adjust the gate, press * set point 4
8. Move the probe to no. 5 hole, adjust the gate, press * set point 5
9. Move the probe to no. 6 hole, adjust the gate, press * set point 6
10. press enter key completed the setting and show the setting curve
11. move high cursor set RL and EL dB. Set RL and EL dB
12. Move cursor to 6-point DAC, press up key set on, the DAC curve will display
29
30
Function Group PK.m
- PK MEM
- ENVLOP
- SHADOW
- B-SCAN ( This function is optional accessories, not incl. in the standard package. To do the B-scan, you need encoder
- Scanner. Scanner and connection cable price ref. to the optional accessories )
- T-SCAN
-
PK MEM
Press PK MEM=on can hold the background echo cross the gate
Put probe on 20 mm thick block to get Press PK MEM set it on, when take off the probe, the The MEM background echo will good for reference
back echo cross the gate background echo display and hold on the screen. during the inspection
ENVOLOP
Press ENVOLOP=on can hold the all the echoes cross the gate
31
Put probe on 20 mm thick block to get Press envlop set it on, when move the probe, the all
back echo cross the gate echo cross the gate will hold and display on the
screen.
SHADOW
Press SHADOW=on. The initial and background echo will change color into solid display
Put probe on 20 mm thick block to get Press SHADOW set it on, the echo color will change When move the probe. The show still hold on the
back echo cross the gate into shadow display type display for reference
32
Time B-Scan
1. Select the probes which already calibrated the probe delay. If it’s a new probe, please follow the probe delay measurement procedure calibrate the probe
delay
2. Press BASE key, enter probe delay ( measurement by Peak method )
3. Press CAL set MEAS=Peak
4. Press PULS, set dual=contact
5. Press BASE, press set the screen display range to 50 mm
6. Put the probe on the step bock, make sure the gate can cover the min. and max. range
7. adjust the gain to get 75% amplitude from thick step as following
8. check the min step 20 mm back echo, see Sa=20.0 mm
9. Move the probe on the 20 mm step. Press PK MEM key, press T-SCAN right arrow key set T- scan on and start time-Bcan
Move the probe from 20 mm to 10/5 mm step. After completed of the T-scan,
press enter key and cal (1) key save the screen to the instrument
10. Press COPY recall key, press CAL (1) key recall the T-Bscan data
11. This Time-Bscan map showed scan the step block from 20 mm-10mm-5 mm
33
CALIBRATION:
1. Select the probe type ( contact or angle beam ), frequency, range and velocity
Contact Probe
Use V2 block 20 mm thick. Due to the probe near field resolution, please use second echo calibrate.
The second echo should be 40 mm.
Or, to use V1 block 100 mm
Angle beam probe
Target the probe to V1 R100 mm curve
2. Move gate cross and cover the max. response echo
3. press CAL enter CAL mode, Press MEAS Arrow key key, now the Means/Peak will show double arrow cursor. This mean, the instrument already enter calibration procedure.
4. Press S-REF1 arrow key, now the cursor will move to S-REF1.
5. Use up and down key set 100 mm ( if use V1 block )
6. Press enter key again. Now the Tru-Sonic will auto-calibrate and the sound path will display 100.0
7. Press BASE key, Now the P-delay showed 2.77 us.
8. Press Freeze key, press enter key again, Save this probe calibrate P-delay.
Example
34
1. Get 4 MHz, 12 mm, 0 degree contact probe and V1 block . Press BASE key set RANGE into125 mm
2. Press MTLVEL RIGHT ARROW KEY, MOVE CURSOR move cursor to MTLVEL. Press arrow key again set MTLVEL 5920 m/s.
3. Press PULS key enter PULS mode
- Set Probe Type- contact
4. Press RECE key enter RECE mode, set FREQU=0.5-5.MHz
5. Press TRIG key, enter TRIG mode
- Set angle 0 degree ( if now showed angel=70, press arrow key SET key select the angle=0.0 degree )
6. Plug probe cable to Tru-Sonic
7. Put couplant on V2, put probe on the block to get response echo.
8. Adjust the gain until second echo raise to 80% height.
9. Press Gate, move Gate A across the second echo and cover it
10. Press CAL key, enter CAL mode, press MEAS right arrow key, the MEAS/Peak showed double arrow.
11. Press S-REF1 right arrow key, the cursor move to S-REF-1. use up/down key adjust the thickness into 100 mm
12. press enter key again. Now the probe sound path will show 100.0 mm.
13. Press BASE key. The P-delay= mm.
14. Now the probe delay already auto-calibrate.
15. Press FREEZE key, Press ENTER key enter SAVE mode. Press CAL/1 key Save the setting
16. Next time when use the same probe, press COPY key, enter recall mod, LOAD right arrow key , move cursor to Load, press CAL/1 key load the previous calibrate setting.
35
1. Get 4 MHz, 8x9 mm, 60 degree contact probe and V2 block . Press BASE key set RANGE into 100 mm
2. Press MTLVEL RIGHT ARROW KEY, MOVE CURSOR move cursor to MTLVEL. Press arrow key again set MTLVEL 5920 m/s.
3. Press PULS key enter PULS mode
Set Probe Type-angle
4. Press RECE key enter RECE mode, set FREQU=0.5-5.MHz
5. Press TRIG key, enter TRIG mode
Set angle 60 degree ( if now showed angel=70, press arrow key select the angle=60.0 degree )
6. Plug probe cable to Tru-Sonic
7. Put couplant on V2, put probe on the block to get response echo.
8. Adjust the gain until second echo raise to 80% height.
9. Press Gate, move Gate A across the second echo and cover it
10. Press CAL key, enter CAL mode, press MEAS right arrow key, the MEAS/Peak showed double arrow.
11. Press S-REF1 right arrow key, the cursor move to S-REF-1. use up/down key adjust the sound path into 100 mm
12. press enter key again. Now the probe sound path will show 100.0 mm.
13. Press BASE key. The P-delay= mm.
14. Now the probe delay already auto-calibrate.
15. Press FREEZE key, Press ENTER key enter SAVE mode. Press CAL/1 key Save the setting
16. Next time when use the same probe, press COPY key, enter recall mod, LOAD right arrow key , move cursor to Load, press CAL/1 key load the previous calibrate setting.
36
4 MHz Contact Probe Manual Delay Measurement
Select one 4 MHz contact probe connected to the cable and plug it to the TRU-SONIC
Get 1 4 step thickness block ( 20, 10, 5 , 2.5mm )
Press TRU-SONIC ON/OFF switch, turn on the TRU-SONIC.
Press pulser key enter pulser sub-manual.
Press DUAL right key, Move cursor to DUAL, use +/- key change the display to contact probe symbol.
37
Use Flank
38
Note:
Use “Peak” or “Flank” have different measurement value display. The probe delay is different. This will case inspection tolerance. Most of time,
PEAK use for flaw detection and FLANK use for thickness measurement.
Select one 4 MHz, 8x9 mm, 70 degree probe connected to the cable and plug it to the TRU-SONIC
Press TRU-SONIC ON/OFF switch, turn on the TRU-SONIC
Put probe on V1 steel test block, tarket the probe to the 100 mm curve
39
1. Press CAL key, select the MEAS=Peak
2. Press BASE key, set range 200 mm
3. Set TRIG=S
4. Set MTL VEL=3353 m/s
5. Press PULS,set Dual=angle beam
6. Set gain until echo rise 80% full screen
7. Press GATE. Move the gate b across the echo and adjust the width and thresh to 50%
8. Press TRIG set angle =70 degree, s-VALE=12.0 mm, MTL THK = 100 mm
9. Press BASE back BASE setting
10. PressP-DELY right arrow key, press increasing key increasing P DELY until Sa=100.0
11. Move the gate across the echo
12. check Sa=100.0. Now the P-DELY for this probe is 3.87 us
40
4 MHz Dual Crystal Contact Probe Delay Measurement
Select one 4 MHz,12mm, 0 degree dualcrystal probe connected to the cable and plug it to the TRU-SONIC
Press TRU-SONIC ON/OFF switch, turn on the TRU-SONIC
1. Press CAL key, select the MEAS= PEAK
2. Press BASE key, set range 25 mm
3. Set TRIG=S
4. Set MTL VEL=5920 m/s
5. Press PULS,set Dual=angle beam
6. Put probe on 20 mm step block
7. Set gain until echo rise 80% full screen
8. Press TRIG set angle = 0 degree, s-VALE=0.0 mm, MTL THK =20.0 mm
9. Press GATE. Move the gate across the echo and adjust the width and thresh to 50%
10. Press BASE back BASE setting
11. Move the gate across the echo
12. PressP-DELY right arrow key, press increasing key increasing P DELY until Sa=20.0
13. check Sa=20.0. Now the P-DELY for this probe is 3.39 us
41
It’s very important to calibrate the probes delay and record it. If you didn’t calibrate the probes delay and entry to the parameter. The inspection
result will generate big tolerance. Each probes have different probe dealy.
Select one 4 MHz,12mm, 0 degree dual crystal probe connected to the cable and plug it to the TRU-SONIC
Press TRU-SONIC ON/OFF switch, turn on the TRU-SONIC
42
1. Press CAL key, select the MEAS= PEAK
2. Press BASE key, set range 25 mm
3. Set TRIG=S
4. Set MTL VEL=5920 m/s
5. Press PULS,set Dual=angle beam
6. Put probe on 20 mm step block
7. Set gain until echo rise 80% full screen
8. Press TRIG set angle = 0 degree, s-VALE=0.0 mm, MTL THK =20.0 mm
9. Press GATE. Move the gate across the echo and adjust the width and thresh to 50%
10. Press BASE back BASE setting
11. Move the gate across the echo
12. PressP-DELY right arrow key, press increasing key increasing P DELY until Sa=20.0
13. check Sa=20.0. Now the P-DELY for this probe is 3.39 us
14. Move the probe to 10 mm step, move gate across back echo, the Sa=10.0 and Da=10.0
15. Move the probe to 10 mm step, move gate across back echo, the Sa=5.0 and Da5.0
43
Note: When use Flank, the probe delay calibrate is different.
Select one 4 MHz,12mm, 0 degree dual crystal probe connected to the cable and plug it to the TRU-SONIC
Press TRU-SONIC ON/OFF switch, turn on the TRU-SONIC
16. Press CAL key, select the MEAS= PEAK
17. Press BASE key, set range 25 mm
18. Set TRIG=S
19. Set MTL VEL=5920 m/s
20. Press PULS,set Dual=angle beam
21. Put probe on 20 mm step block
22. Set gain until echo rise 80% full screen
23. Press TRIG set angle = 0 degree, s-VALE=0.0 mm, MTL THK =20.0 mm
24. Press GATE. Move the gate across the echo and adjust the width and thresh to 50%
25. Press BASE back BASE setting
26. Move the gate across the echo
27. PressP-DELY right arrow key, press increasing key increasing P DELY until Sa=20.0
28. check Sa=20.0. Now the P-DELY for this probe is 3.39 us
44
29. Move the probe to 10 mm step, move gate across back echo, the Sa=10.0 and Da=10.0
30. Move the probe to 10 mm step, move gate across back echo, the Sa=5.0 and Da5.0
45
Echo to Echo Measurement
Select one 4 MHz contact probe connected to the cable and plug it to the TRU-SONIC
Get 1 4 step thickness block ( 20, 10, 5 , 2.5mm )
Press TRU-SONIC ON/OFF switch, turn on the TRU-SONIC.
Press pulser key enter pulser sub-manual.
Press DUAL right key, Move cursor to DUAL, use +/- key change the display to contact probe symbol.
46
If use “FLANK”, The probe delay calibrate same as before. Just take care set the aGATE and bGATE down cross echo wave first rise line. This will give you
more precision measurement.
Dual contact probe echo-echo do the same way.
47