You are on page 1of 3

Features

Feature
Dimensions
Weight

Electrical requirements

Water requirements
heat output
Maximum throughput
Sample types
Dead volume
Sample capacity
Sample results storage

Sample barcode types


Sample probe carryover
Sample Clot and bubble detection
Sample interferance mesurement
Sample volume
Reagent type
Reagent on board stability
Reagent pressure mesurement

Calibration frequency
Online error code help

Host interface
Remote diagnostics
On board maintenance records
Alinity i
Chemiluminescence
134 x 119 x 117 cm/1.39 m2
624 kg
SCM: 90–264 V, 16 amp
Each Instrument: 180–264 V, 16 amp
Average: <10 L/hr
Max: <30 L/hr
Average 1634 Btu
Up to 200 TPH
Serum, plasma, whole blood, urine
50 μL (sample cup)
150
200.000
Code 128, Standard Code 39, Interleaved 2 of 5,
Codabar
≤0.1 parts per million†
Yes
No
2–200 μL
100% liquid ready-to-use
15–30 days
Yes

15–30 days

Yes

HL7 or ASTM

AbbottLink
Yes
Architect i1000sr
CHEMIFLEX
124.5 × 149.9 × 76.2 cm
288 kg
C 110–120 V or 200–240 V, ± 10%,
50 or 60 Hz self-adjusting

Purified water to dilute buffer concentrate

2,400 BTU/hr, Running mode


Up to 100 tests/hour
Serum, plasma, whole blood, urine
Yes (50 µL)
65
50.000

Code 128, Standard Code 39, Interleaved 2 of 5, Codabar

≤ 0.1 ppm
Yes
No
10–160 µL (Average: 62 µL)
100% liquid ready-to-use
14–30 days
Yes
Calibrate with new lot number, if controls are out of range or if
specified otherwise within the package insert
Yes
Bi-directional, serial RS-232 interface, host query option
available
AbbottLink
Yes

You might also like