Lab Advisor Diagnostic Result HPLC 1260

G7111B Pump Leak Rate Test Passed


DEAET01442 The test determines the leak rates in the primary and the secondary pump chambers for
D.07.24[0001] component level diagnostic.
1260 Quat Pump

Started at 6/28/2018 2:34:13 PM by DESKTOP-AG16BPK\Laboratorio on DESKTOP-


AG16BPK using Agilent Lab Advisor - Version B.02.10 [253] SP1 - Basic
Results
Item Value Result
Maximum system pressure 400 bar Done
Purge Channel A Done Done
Remaining purge time 0 min 0 sec of total 2 min Done
Leak of primary pump chamber 1.4 µl/min
Leak rate limit 3 µl/min Passed
Leak of secondary pump chamber 1.0 µl/min
Leak rate limit 3 µl/min Passed
Finished at 6/28/2018 2:39:04 PM

Signals

Agilent Technologies Printed at: 6/28/2018 2:39:19 PM Page 1 of 2


Lab Advisor Diagnostic Result HPLC 1260

Signature: ........................................................................

Agilent Technologies Printed at: 6/28/2018 2:39:19 PM Page 2 of 2