DEAET01442 The test determines the leak tightness of the system between pump and blank nut. D.07.24[0001] Mandatory for G7167A/B Multisamplers: use 'Blank nut, long' (part number 5067-6127). 1260 Quat Pump
Started at 6/28/2018 2:51:40 PM by DESKTOP-AG16BPK\Laboratorio on DESKTOP-
AG16BPK using Agilent Lab Advisor - Version B.02.10 [253] SP1 - Basic Results Item Value Result Maximum system pressure 600 bar Done Purge Channel A Done Done Remaining purge time 0 min 0 sec of total 1 min Done Resulting flow value 2.21 µl/min System leak 5 µl/min Passed Finished at 6/28/2018 2:55:54 PM