3 Autosampler

12 Fit the male nut and the ferrule


to the sample loop, then use a
6 mm wrench to tighten the male
nut (A) and fit the sample loop to
the joint section on the needle
side.

Insert the sample loop until it touches the back of the connection hole, and
then use a wrench to tighten it.
If the sample loop is not fully inserted and does not reach the back of the
connection hole, a dead volume will be created and may cause a broad peak
or cross contamination.

13 Attach the section (a) of the


sample loop that is positioned by
bands to the innermost hook (b)
attached to the inner left side of
the plastic cover.

As shown in the
diagram, to ensure
the band heads face
toward the left side
of the instrument,
twist about
90° counterclockwise
when seen from the
front of the
instrument, and
attach the sample
loop to the hook.

14 Fix the sample loop to the


remaining 2 hooks (c).

15 Fix the sample loop to the 2


clamps (d) attached on the inside
of the plastic cover.

36 Prominence-i LC-2030 LC-2030C LC-2030C 3D


3.5 Replacing the Sample Loop

16 Align the protrusion (a) on the


top of the Z mount with the slit
(b) on the top of the Z mount
cover and fit them together. Hold
the bottom of the Z mount from
both sides, and fit the right and
left catches (c) on the bottom of
the Z mount cover to the Z
mount bottom.

17 Pass the flat cable (A) through 3


the clamp (B) and close the clamp.

If the clamp (B) is


stiff and difficult to
close, hold the Z
mount cover with
one hand and close
the clamp (B).

After inspection and replacement, be sure to attach the Z mount cover and
clamp the flat cable.
Failure to do so may result in the cable interfering with another internal part
and causing abnormal operation or a malfunction.

After Z mount cover is installed, check that Z mount cover does not come off
with lightly pulling it forward.
If Z mount cover is not correctly installed, it may drop, resulting in damaging
the parts of the autosampler.

18 Fit the male nut and the ferrule


to the sample loop, then use a
6 mm wrench to tighten the male
nut (A) and fit the sample loop to
port 1 of the high-pressure valve.

Prominence-i LC-2030 LC-2030C LC-2030C 3D 37


3 Autosampler

19 Slacken the sample loop fitted to


port 1 as shown in the diagram,
then adjust the tubing
configuration so that the lowest
point is positioned above the
high-pressure valve tray.

20 Move the Z mount by hand


toward the upper part of the
high-pressure valve, and check
that the sample loop does not
interfere with the Z mount
bottom.

21 Attach the front panel.


"3.1 Removing/Attaching the Front Panel" P.12

4 Press (Power button) to turn ON.

After the instrument starts, input the login pass ID Number and press [OK] to display
the [Replacing the Sample Loop] screen.

5 Press the [Next] button.


The next screen is displayed and the usage
value is reset.
The replacement history is recorded in the
maintenance log.

6 Check that the usage value is reset and


then press the [Next] button.
The next screen is displayed.

38 Prominence-i LC-2030 LC-2030C LC-2030C 3D


3.6 Replacing the Measuring Pump Plunger Seal

7 Enter the volume of the fitted sample


loop, and then press [Finish].

When replacing with a sample


loop for 50 µL or below, if the
sample aspiration rate is set to
a value greater than 5.0 µL/sec,
the sample aspiration rate will
be automatically changed to
5.0 µL/sec.

Operation Guide "5.2.2 Autosampler"

3.6 Replacing the Measuring Pump Plunger Seal 3


Necessary tools
Wrench (6 mm and 8 mm)
Allen wrench (M4)
Seal tool (Part No.: S228-25142-01)

Necessary parts

Part Name Part Type Part No.


Measuring Pump Plunger Seal Consumable part S228-35145

1 Press [Plunger Seal] on the [Replacement Parts Selection] screen.


"2 Common Operation" P.7

The measuring pump plunger moves all the way back, and the [Measuring pump plunger
seal] screen is displayed.

2 Replace the measuring pump plunger seal.

1 Open the valve mount door.

Prominence-i LC-2030 LC-2030C LC-2030C 3D 39