Insert the sample loop until it touches the back of the connection hole, and
then use a wrench to tighten it.
If the sample loop is not fully inserted and does not reach the back of the
connection hole, a dead volume will be created and may cause a broad peak
or cross contamination.
As shown in the
diagram, to ensure
the band heads face
toward the left side
of the instrument,
twist about
90° counterclockwise
when seen from the
front of the
instrument, and
attach the sample
loop to the hook.
After inspection and replacement, be sure to attach the Z mount cover and
clamp the flat cable.
Failure to do so may result in the cable interfering with another internal part
and causing abnormal operation or a malfunction.
After Z mount cover is installed, check that Z mount cover does not come off
with lightly pulling it forward.
If Z mount cover is not correctly installed, it may drop, resulting in damaging
the parts of the autosampler.
After the instrument starts, input the login pass ID Number and press [OK] to display
the [Replacing the Sample Loop] screen.
Necessary parts
The measuring pump plunger moves all the way back, and the [Measuring pump plunger
seal] screen is displayed.