Professional Documents
Culture Documents
UNIT Electrical
Measurements
CONTENTS
Part-1 Measurement System,. 1-2E to 1-17E
. .
Characteristics of Instruments,
Methods of Measurement,
Errors in Measurement and
Measurement Standards,
Measurement Error Combination
1-1E (EC-Sem-5)
1-2E (EC-Sem-5) Electrical Measurements
PART-11
Measurement System, Characteristics of Instruments, Methods of
Measurement, Errors in Measurement and Measurement
Standards, Measurement Error Combination.
Questions-Answers
Long Answer Type and Medium Answer Type Questions
www.
Answer
A measurement system comprises of the following elements:
1. A transducer is one which converts the measurand (quantity) into a
usable electrical output.
2. A signal conditioner, which converts the transducer output into an
electrical quantity suitable for control, recording and/or display.
3. Display or read out devices to display the required information about
the measured, generally in engineering units.
4. An electrical power supply which provides the required excitation to the
transducer and the necessary electrical power to the signal conditioners
and display devices.
5. The output signal may be an analog or digital quantity. The read out or
display device may be in analog or digital form.
Analog
Input Signal display To process
Quantity Transducerj conditioner
recording| control system
(measurand)
devices
|Analog to|
Excitation digital
power converter
supply
Digital
display/
recording
devices
Answer
A Instrument: It is defined as a device for determining the value or
magnitude of a quantity or variable.
B. Various types of electrical instruments are:
a. Absolute instruments
1. These instruments give the magnitude of the quantity under
measurement in terms of physical constants of the instrument.
2. Examples are Tangent Galvanometer and Rayleigh's Current Balance.
b. Secondary instruments:
1. These instruments are so constructed that the quantity being measured
can only be measured by observing the output indicated by the
instrument.
These instruments are calibrated by comparison with an absoiute
2.
instrument or another secondary instrument which has already been
calibrated against an absolute instrument.
Examples are voltmeter, glass thermometer and pressure gauge.
3
C. Deflection type:
In instruments of this type, the deflection of the
instrument provides a
under measurement.
basis for determining the quantity
effect which deflects or
2. The measured quantity produces some physical
produces a mechanical displacement
of the moving system of the
instrument.
d. Null type:
indication leads to
1. In null type of instrument, a zero or null
a
determination of the magnitude of measured quantity.
1-4E (EC-Sem-5) Electrical Measurements
Digital type:
1. The signals which vary in discrete steps and thus take up only finite
different values in a given range are called digital
signals
2. The devices that produce such signals are called digital devices.
Que 1.3. Distinguish between the null type and deflecting type of
instruments. Give examples to support your answer.
Answer
Answer
Generalized instrumentation system:
Quantity to Primary Variable Variable D Data Data
sensing Hconversion manipulation transmissionpresentation|
be measured element element element element element
Answer
Performance characteristics are of
two types:
1. Static characteris ics: It focuses on the measurement of quantities
which are either constantly or
very slowly varying with time. There is
linear relation with time.
2. Dynamic characteristics: It involves the measurement of
that vary with time but maintains quantities
a second order or third order
differential equation with time.
Que 1.6. What are the desirable static characteristics in a
measurement system ?
Answer
They are of different types as follows:
1 Accuracy:
i. Accuracy refers to the closeness of a measured value to a standard or
known value.
. Forexample, if you obtain a weight measurement of 3.2 kg for a given
substance, but the actual or known weight is 10 kg, then your
measurement is not accurate. In this case, your measurement is not
close to the known value.
2. Precision:
i. Precision refers to the closeness of two or more measurements to each
other.
ii. Using the example above, if you weight a given substance five times,
and get 3.2 kg each time, then your measurement is
very precise.
3. Sensitivity:Itis the ratio of change in output with respect to change in
input.
Sensitivity= Changes in Output
Changes in Input
Electronic Instrumentation and Measurements 1-7E (BC-Sem-5)
will be shown on
Reproducibility : It indicates that the same output
the same input conditions.
5. Resolution
change in
a. Changes in output are achieved only after making a
input.
It
Drift: Itis undesirable change in output over a period of time.
means
9.
with time.
that with a given input the measured value vary
Answe
follows:
They are of different types as
to which an output changes with respect
1. Fidelity: It indicates a degree
to change in input without considering
dynamic errors.
in response, whenever there
is change
2. Lag:It indicates the delay output
in input parameters.
to rapidity ofinstrument response towards
3. Speed ofresponse : It refers
the changes in input quantities.
indicated by
4. Dynamic errors : It is difference between the value
true value changing with time, if
there is no static
instrument and the
error assumed.
Answer
broad categories
We classify this generally into two
A. Direct measurement:
calibrated
the unknown quantity to some
1. These are made by measuring
standard and comparing the two.
that an error can be
2 It is done by human being, so it is quite possible
introduced during measurement.
Answer
Gross errors:
1. These errors occur due to human mistakes in reading instruments,
recording and calculating results of measurement.
2. These errors can be avoided
by adopting two means
1 immense care should be taken while
taking the reading and recording
the data.
1. Two, three or even more
readings should be taken for the
being measured.
quantity
Systematic errors:
The systematic errors are due to the
shortcoming of the instrument
and the characteristic of the material used in the instrument.
2 These errors cannot be determined by direct and
repetitive observations
of the measurand made each time using same technique.
3. These be located only by having
errors can
repeated measurements
under diferent conditions with different equipment and where
by an entireiy different method.
possible
4. These errors may be instrumental, environmental or observational
errors.
a. instrumental errors
1 These errors arise due to:
i. inherent shortcomings in the instrument.
. Misuse ofthe instrument.
ii. Loading effect of instrument.
2. Instrumental errors are inherent due to construction, calibration or
operations of the instrument. These errors may cause the instruments
reading too low or too high.
Electronic Instrumentation and Measurements 1-9 E (EC-Sem-5)
3. These errors can be avoided by following methods
i. Selecting a proper instrument and planning a proper procedure for
measurement.
. Calibrating the instrument carefully against a standard.
ii. Loading effect can be avoided by using an instrument intelligently and
correetly.
b.
Environmental errors:
1 The environmental errors are due to external conditions to the
measuring device. These errors are much more troublesome than
assembly errors.
2. These errors can be reduced by using the instrument in controlled
conditions of pressure, temperature, humidity in which it was originally
assembled and calibrated.
C.
Observational errors:
1. It can be due to wrong scale reading, wrong recording of data,
inaceurate estimates ofaverage reading, incorrecet conversion of units
in between consecutive readings.
2. To eliminate such observational error, one should use the instrument
with mirror, knife edged pointers etc. Now-a-days, the instruments
with digital display are available which can largely eliminate such
observational error.
Answer
1. Limiting errors/guarantee errors are those errors which are defined by
manufacturer.
2. Such errors exist in instruments due to utilization of different part in
manufacturing an instrument.
33. Each part has some errors, which combined together will be large.
4 Such errors can be found by comparing the instrument against the
standard quantities and accordingly are specified by the manufacturer
5. The value of limiting error largely dependent upon design constraints of
component used in instruments.
Answer
Absolute errors:
1 The difference between the measured value and the true value of
quantity is known as absolute error.
1-10E (EC-Sem-5) Electrical Measurements
4, A-A,
&A, Static error of measurement/absolute error
AMeasured value of quantity
ATruevalue ofquantity
Relative errors:
1. It is defined as the ratio of absolute error to the true value of the
quantity under measurement is known as relative error.
Absolute error dA,
True value
E,Relativevalue
The relative error can be positive or negative.
Que 1.12.| Differentiate between gross errors and systematic
errors. List a few ways of minimizing the effect of errors in
Answer
A Difference:
Que 1.13. A thin wire has a length of 21.7 em and radius 0.46 cm.
Calculate the volume of the wire correct to required significant
= 14.42 cm3
Electronic Instrumentation and Measurements 1-11 E (EC-Sem-5)
Since, the radius of the wire is given in the significant figure of two. So
the result is also rounded off in two digits.
T 27
wherel and g are measured in : 2% and t3 h errors.
1 Given, T 2m
nominal resistance
Que 1.15. A batch of resistors each having a
of 330Q are to be tested and classified as +10% components at 25 °C.
If their temperature coefficient is - 300 ppm/ °C, calculate the
maximum and minimum resistance for these components at 100 °C.
10x330 = t 33 Q
100
2. Largest possible resistance at 25 °C
Max 330 + 33 =363 Q
3. Minimum possible resistance at 25 °C
min 330-33 = 297 2
4 For Rnax
nin
= -0.108 / ° C
Temperature increase 100 °C- 25 °C =75 °C
Total resistance increase, AR =- 0.108 Q/°Cx 75 °C =- 8.12
Maximunm resistance at 100 °C =R+ AR
363 - 8.1 = 354.92
. For Rmin
Resistance change per °C
10
1 Absolute error = t330 x =t 33 2
100
Rmay 330 +0.1 x 330 363 2 lat 25 °C
R in 330-0.1 x 330 297 2 lat 25 °C]
Resistance change per °C:
300
3. AtRmax? AR/°C 363x- -0.10891 2/ °C
10
- 300
At Rmin' AR/ C 297x- =- 0.0891 /°C
10
Iwhere temperature coefficient =- 300 ppm/°C]
4. Temperature increase:
AT 100-25 75°C
5. Total resistanceincrease:
AR
max
a -0.1089 x 75=-8.1675 Q
ARTnin-0.0891 x 75 =- 6.6825 2
6 max 363-8.1675 354.83Q at 100 °Cl
R min 297 6.6828 290.31 Q [at 100 °C] [at 100 °C]
Que 1.17.Define systematie errors in details. A batch ofresistors
that each have a nominal resistance of 330 Q are to be tested and
classified as 10 % components at 25 °C. If their temperature
coefficient is - 300 ppm/°C, calculate the maximum and minimum
Answer
A
Systematicerrors: Refer Q. 1.9, Page 1-8E, Unit-1.
B. Numerical : The procedure is same as Q. 1.15. Page 1-11E,
Unit-1.
(Ans. Ras 357.555 Q, Rin = 292.545 2)
Answer
Standards of measurement:
A standard is a physical representation of unit under measurement.
2. The standards are used for the purpose of obtaining the value of a
physical quantity.
3 egthe fundamental unit of mass in the international standard (SD is
the kilogram.
The types of standards are classified by their function and
applicationas
A. International Standards
1 These are defined on the basis of international agreements.
2. These standards are designed to represent certain units of measurement
to the closest possible
aceuracy.
3. These standards are checked and evaluated
regularly against absolute
measurement in terms of the fundamental units.
4. These are maintained at the international bureau of
and
weights and
measures are not available to ordinary user.
B. Primary Standards
1 These are absolute standards of such a high
accuracy that they can be
used as the ultimate reference standards.
2. The primary standards are maintained in national
standards laboratories.
3. Main function of
primary
the secondary standards.
standard is the verification and calibration
C. Secondary Standards
1. These are the basic reference standards used in
industrial measurement
laboratories.
2. These are maintained by the industries
working on these standards and
checked locally against other reference standards in
that area.
3. A particular industry is responsible for the maintenance and
of secondary standards. calibration
Electronic Instrumentation and Measurements 1-15 E (EC-Sem-5)
Answer
When a quantity is caleulated from measurements made on two
(or more) instruments, it must be assumed that the errors due to
instrument inaccuracy combine in the worst possible way. The resulting
error is then larger than the error in any one instrument.
Sum of Quantities: Where a quantity is determined as the sum oftwo
measurements, the total error is the sum of the absolute errors in each
measurement. It is shown in Fig. 1.19.1.
E =(V, +AV,) + (V2 + AV,)
E=(V+V): (aV, + AV,)
Vt AV
V2t AV2 R2
Difference of quantities:
in which a potential difference is
1. Fig. 1.19.2 illustrates a situation
determined as the difference between two measured voltages. Here
E-
V, AV,R RVAV
Fig. 1.19.2. Error in difference of quantities
equals sum of errors.
Product of quantities:
When a caleulated quantity is the product of two or more quantities, the
percentage error is the sum of the percentage errors in each quantity
Fig. 1.19.3.
P El
= (E t AE) d + A) = El:E al:1AE AE DI
Since AE Al is very small,
P El+ (E Al+ AE)
(EA AE100%
EI 100%
7E*
100
Percentage error in P =(Percentage error in /) + (Percentage error in E)
I+AI
E+AE
Fig. 1.19.3.
+(Percentage error in )
Quantity raised to a power: When a quantity A is raised to a power
B. the percentage error in A" can be shown to be
Electronic Instrumentation and Measurements 1-17E (EC-Sem-5)
PART-2
Review of Indicating and Integrating Instruments PMMC
Instrument, Galvanometer, DC Ammeter,
DC Voltmeter, Series Ohmmeter.
Questions-Answers
Long Answer Type and Medium Answer Type Questions
w.
Answer
A. Constructiom:
A PMMOC meter mechanism is the basic mechanism for most analog
1.
D'Arsonval
instruments. This movement is also called
measuring
movement.
uLLliinluul,
LLLlu,7 scale
Mirror Pointer
S
Permanent
magnet
Radial field- Balancing weight
Answer
A Permanent magnet movingcoil (PMMC) nmeters: Refer Q. 1.20,
Page 1-17E, Unit-1.
B. Proof:
When a current I flows through a one-turn coil situated in a magnetic
field, a force F is exerted on each side of the coil,
F BIl newtons
where, B is the magnetie flux density in tesla, lis the current in amperes,
andl is the length of the coil in meters.
2. Since the force acts on each side of the coil, the total force for a coil of N
turns is
F 2BILN newtons
3. The force on each side acts at a radius r, producing a deflecting torque,
Tp 2BINr(N-m)
= BlIN(2r)
Electrical Measurements
1-20E (EC-Sem-5)
Tp BlIND ..(1.21.1)
where, K is a constant.
5. For a given deflection, the controlling and deflecting torques are equal,
KO BlIND
6. Since all quantities except 0 andI are constant for any given instrument,
the deflection angle is
0 CI ..(1.21.2)
where, C is a constant.
to
7. Eq. (1.21.2) shows that the pointer deflection is always proportional
the coil current. Consequently, the scale of the instrument is linear, or
uniformly divided i.e., if1 mA produces a 1 cm movement ofthe pointer
from zero, 2 mA produces a 2 cm movement, and so on.
Rm
R
Fig. 1.221. A swamping resistance connectedin series withthe coil
Electronic Instrumentation and Measurements 1-21 E (EC-Sem-5)
Answer
A Construction:
1. A galvanometer is essentially a PMMC instrument designed to be
sensitive to extremely low current levels.
2. The simplest galvanometer is a very sensitive instrument with the type
of center-zero-scale as shown in Fig. 1.23.1(a).
2 1
1 2
4 lluluI,. 4
(a) Center-zero scale
Suspension
ribbon
Soft iron Mirror
core Coil
Mirror
Lamp
TT
Permanent Scale 0
magnet
(b) Basic deflection system of a galvanometer using a light beam
Fig. 1.23.1.
1-22 E (EC-Sem-5) Electrical Measurements
V= I (R, +Rm
V
R,+R
35 V
R R, 100 uA
2 k2 = 348 k2
Answer
A Swamping resistance: Refer Q. 1.22, Page 1-20E, Unit-1.
B. Materials used: Refer Q. 1.22, Page 1-20E, Unit-1.
C. Numerical: The procedure is same as Q. 1,.23, Page 1-218, Unit-1.
At 0.1: V = 100 V)
Answer
A Construction:
1. A DC ammeter consists of a PMMC instrument and low resistance
shunt as shown in Fig. 1.25.1.
Im
Meter
Input
DC current Ra
Fig. 1.25.1
2. When large current is required to be measured then a resistance is
connected in parallel with the meter to bypass additional current. This
resistance is called a shunt R,j
3. Out of the total input current, I current equal to meter current I passes
flows
through the meter and the balance shunt current =
through shunt resistance as per Kirchhoff's current law (KCL).
h I -1,m
4. Since the shunt resistance R, is in parallel with the meter movement,
the voltage drop across the shunt and meter movement must be same.
1-24 E (EC-Sem-5) Electrical Measurements
V
shunt movement
sh I-1
or R
5.
ah(/I)-1
Therefore ammeter range could be varied by varying the shunt resistance
C. Advantages:
1. Scale is uniformly divided.
2. Very low power consumption.
3. Torque/weight ratio is high so giving high accuracy.
4. Errors due to stray magnetic field are small.
D. Disadvantages:
1. They can be used for DC only. If connected to AC, readings are zero due
to finite inertia of pointer.
2. These instruments are costly.
Answer
A. DC ammeter: Refer Q. 1.25, Page 1-23E, Unit-1.
B. DC voltmeter:
a. Construction:
1. A DC voltmeter is made up of a PMMC instrument and a series multiplier
resistor as shown in Fig. 1.26.1.
2. The D'Arsonval movement can be converted into a DC voltmeter
when a low current flowing through coil is multiplied with its internal
resistance. But the voltage measured is of very low amplitude.
b. Working
1. To measure high voltages, a resistor should be connected in series with
the meter that will drop the excess voltage across it thus the meter
movement does not cross its safe ratings.
2. This series resistor is called as multiplier resistor because it
multiplies
the working ranges of the meter movement.
Electronic Instrumentation and Measurements 1-25 E EC-Sem-5)
Vin R,+R,)
or
R,=-R m
DC input Voltmeter
Vin
(a) 6)
Fig. 1.27.1. (a) Circuit for determination of ammeter resistance,
(b) Circuit for using the ammeter as voltmeter.
1-26 E (EC-Sem-5) Electrical Measurements
4. Let R,, be the internal resistance of the ammeter, then the current
lowing through the circuit is i = EXR + R,), where E is the input
voltage.
5. The voltage drop across R is V, and the current is i, = V,/R. Since
(E-V,)R
R
V,
6. To calculate R, we note that the voltage drop across the ammeter,
showing full seale reading I is V = I x R To make ammeter
full-scale to read full-scale voltage Vo the remaining voltage
V = V- V should drop across R, and from this consideration we
calculate series resistance as,
RR
R mVV
Que 1.28. Describe the working of series ohmmeter.
Answer
Series ohmmeter:
1. The resistance reading is indicated by meter movement. A battery is
connected in the cireuit to produce required current for meter movement.
R, RRR
R+
5. Battery current is required to produce half scale deflection.
E
2R
6. Total current 1, to produce full scale deflection is
psD=2/
E
7.
7. By applying KCL in the circuit
I =l+lpsSD
FsD
8. Using Kirchhoff's voltage law
RspR
1,- psD
R =
IpsD R R
E-1psp R,
2m R-sp R
R R, +Rm E
Ea lRa
..-.
E=1,R4
and -1R
joC)
6. Substituting the value of1, from eq. (3.13.1) in eq. (3.13.2), we have
1+R +jol,)=1,R,+IjoCR
1r+R+ joL -joCRr)= 1,R2 ..3.13.3)
or
2 a , joCR,R+ j»CRgRa|
+R +jol,-joCRr)=I, R
8. Equating the real and the imaginary parts,
3-22 E (EC-Sem-5) Measuring Methods and AC Bridges
RRA
and
L=CrR,
R +R)+RR
Que 3.14. Write the mathematical expression for inductor -factor
and capacitor D-factor. Explain the working of induetance
D-factor of a capacitor:
1. The dissipation factor D defines the quality of the capacitor. D is
simply
the ratio of the component reactance (at a given frequency) to the
resistance measurable at its terminals.
2. Using the parallel equivalent circuit,
1
D
R oCR,
3. Ideally, R, should be very much large then 1/(oC,), giving a very small
dissipation factor.
4. When a series equivalent circuit is used, the equation for dissipation
factor can be shown to be
D-oCR
Inductance comparison bridge:
1. The circuit of the inductance comparison bridge is shown in
Fig. 3.14.1.
2 The unknown inductance, represented by its (series equivalent circuit)
L, and R, is measured in terms of a precise standard value of inductor.
Electronic Instrumentation and Measurements 3-23 E (EC-Sem-5)
Z1
R
-D-
ww
R+joL= R+ jol,
R R
..(3.14.1)
RRR,"
5. Equating the real and imaginary components in eq. (3.14.1)
Rand R n d -
R= and L=
R R
Que 3.15. Define Owen's Bridge with phasor diagram. Give its
advantages and disadvantages also.
Answer
This bridge may be used for measurement of an inductance in terms of
1.
capacitance.
L = Unknown self-inductance of resistance R,
2. Let,
R Variable non-inductive resistance,
R = Fixed non-inductive resistance,
C,= Variable standard capacitor,
and C= Fixed standard capacitor.
3-24 E (EC-Sem-5) Measuring Methods and AC Bridges
-E Eg
R, L
QO000-
(D)
E E
(a)
2
aC2 E=E2
1oL --
E=1,Rg
(6) oC
Fig.3.15.1.
3. At balance,
L= R, R,C, and
R=R3 Ca
Advantages:
1. Owen's Bridge can be used over a wide range of measurement of
inductance.
2. The balance equation are simple and do not contain any frequency
component.
Disadvantages
Owen's bridge requires a variable capacitor which is expensive.
Electronic Instrumentation and Measurements 3-25 E (EC-Sem-5)
ler-
E Detector Ic
G1
oC oC3
Fig. 3.16.1. Schering bridge
4. One of the ratio arms consists of a resistance in parallel with a capacitor
and standard arm consist only a capacitor.
5. The standard capacitor is a high quality mica capacitor or an air capacitor
for insulation measurement.
6. The general bridge balance equation is,
Z,2, = 2,23
Y jo
Z Rg,Z=Rz, Z o Ca
Z= R, * 1
joC,
3-26 E (EC-Sem-5) Measuring Methods and AC Bridges
Substituting,
R R, OC, R joc,
R R RoC
oC
Separating real and imaginary terms,
R.: RC ..(3.16.1)
C
1 R
oC, oR,C
...3.16.2)
R
The eq. (3.16.2) gives the value of unknown capacitance C, in terms of
standard capacitor Cg, R, and R
B. Measurement of dielectric loss:
For a series combination of R, and C, the angle between the voltage
across the series combination and voltage across the capacitor C, is
called loss angle (8).
Now, tan = IR
C,)
R,C,
tan = o G ,
oR,C
Que 3.17.Explain capacitance and inductance bridges.
OR
How the capacitance value is measured with the help of De-Sauty's
bridge? This bridge is used for which type of capacitor.
OR
The standard capacitance value in Fig. 3.17.1 is C, = 0.1 F, and
RR, can be set to any ratio between 100: 1 and 1:100. Caleulate
the range of measurements of unknown capacitance C.
Measurements 3-27E (EC-Sem-5)
Electronic Instrumentation and
Answer
Capacitance bridge
This bridge is simplest method of comparing two capacitances.
1
V-V-l,R=
V V
1/oC
1,/oCa
(a) Circuit diagram (6) Phasor diagram
Fig. 3.17.1.
..(3.17.1)
2 At balance
...(3.17.2)
and
have
Dividing eq. (3.19.1) by eq. (3.19.2),
we
3.
R
-jloC-iloC,
After solving, we get
R
C C2
low dielectric losses such as air
4 This bridge is used for the capacitor with
capacitor.
Numerical:
C-CR
100: 1:
For RR, =
10
C =0.1 uFx
1
=10 uF
For R/R, =1: 100:
C = 0.1 uFx =0.001 uF
100
Unit-3.
Inductance bridge : Refer Q. 3.11, Page 3-17E,