You are on page 1of 1

Waters - The Science of What's Possible.

Error : Data Incomplete in Empower Project review window - WKB3132

SYMPTOMS
• All injections run properly, but last injection shows "data incomplete" when brought in review
• Recover.log located in C:\Empower\InstrumentServer shows no buffering

ENVIRONMENT
• Empower 2
• Empower 3

CAUSE
Unknown

FIX or WORKAROUND
1. Click on Browse Projects.
2. Open the project and click Injections tab.
3. Highlight the injection, right-click, and then click Review.
4. Click in the chromatogram window, and then right-click.
5. Click Complete Data.If you do not see Complete Data as an option, it indicates that the "Verify incomplete data in
raw data files" privilege is not selected.
6. To select this privilege:Click Configure the System, click User Type, highlight the user type, and click Properties. In
Data Acquisition, select "Verify incomplete data in raw data files". Note: You need administrator privileges to
assign this privilege.

ADDITIONAL INFORMATION

id3132, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC,
EMPOWER2, EMPOWER3, EMPSW, SUP

https://support.waters.com/KB_Inf/Empower_Breeze/WKB3132_Error_Data_Incomplete_in_Empower_Project_review_window
Updated: Tue, 02 May 2023 08:06:46 GMT
Powered by
1

You might also like