Professional Documents
Culture Documents
Objective: The AnaConda is an anaesthetic gas delivery system for use in Critical Care
alongside the Dräger ventilator to administer isoflurane in a safe and controlled
manner to mechanically ventilated patients.
Associated Documents: Outline other documents that this document should be read in
conjunction with or may be required for implementation of this clinical guideline (If appropriate).
Contraindications
Absolute
• Family history of malignant hyperthermia.
Relative
• Lots of secretions requiring frequent suctioning.
• Raised ICP.
• Pregnancy.
• Profound circulatory shock.
• Thick respiratory secretions.
Critical Incidents
Malignant hyperthermia
• Malignant hyperthermia (MH) is an extremely rare reaction to volatile anaesthetic agents.
It is a medical emergency.
• It manifests as unexplained tachycardia, increased oxygen consumption and CO2
production, fever and rigidity upon exposure to volatile anaesthetic agents.
• Hyperthermia maybe masked. An unexpected and unexplained rise of CO2, or gas sweep
to achieve target CO2, should lead to suspicion of this condition.
• If suspected immediately discontinue the use of sevoflurane, remove the AnaConDa
device, place the patient back on the ICU ventilator on 100% oxygen, start alternate IV
sedation, and refer to the MH guideline (Appendix B).
• MH emergency kit containing Dantrolene is kept in main theatres on the Royal site and in
emergency theatre recovery, as the MH trolley on the Aintree site.
Special Circumstances
Transfers of patients with AnaConDa Device
• AnaConDa can be used with Oxylog Transport Ventilator but at present please use
intravenous sedation (eg. Propofol and opiate) as advised by the Critical Care medical team.
Nebulisation
• Insert high peep/in line nebulisation device between patient and Anaconda device.
• Continuous nebulisation will require more frequent changes of the AnaConDa device.
Change at 12 hours or more before if indication of AnaConDa filter saturation.
• Angle AnaConDa device @ 45 degree (decline towards the patient) to avoid condensation
of frequent nebulisers. Note potential of device saturation due to frequent nebulisers which
may result in increasing circuit resistance (high pressures, low Tidal Volume).
In line Suctioning
• Suctioned Isoflurane quantities are small and unlikely to leak into the ICU environment in
any significant quantity and therefore can be suctioned in the usual way.
Manual Bagging
• Leave the AnaConDa in site when manual bagging is required, disconnection of the
ventilator circuit will only allow a small quantity of Isoflurane (the non-reflected fraction) to
enter the ambient environment.
• Consider an increase in the Isoflurane administration rate in response to increased tidal
volumes during hand ventilation.
Bronchoscopy/Tracheostomy insertion
• Prolonged bronchoscopy and suctioning may decrease alveolar concentration of levels of
isoflurane. Seek medical advice to ensure the maintenance of target MAC level. Consider
conversion to IV sedation during the procedure, then revert back to Isoflurane.
Environmental levels of volatile gas when AnaConDa is used are maximal 1.2ppm meaning
negligible exposure to surrounding patients as well as care givers (50 parts per million are
maximum accepted concentration in the UK).
Spillage of Isoflurane
Every effort should be made to minimise any spillage and breakage. Aim to avoid
spillage/leakage of Isoflurane into the atmosphere at all times with above actions by stopping
infusions, capping syringes and bottles and handling glass bottles with care. Small spillages will
dissipate at normal room temperatures.
Storage of Isoflurane
Royal Site - Isoflurane is stored in the main pharmacy on ITU in the flammables cupboard to
protect from damage. Stock levels is 4x250ml bottles. Order any additional stock required via the
EPMA system or pharmacist. Any isoflurane bottle at the bedspace should be stored in a drawer
or cupboard until replaced in the main pharmacy.
5.
Setting up AnaConDa device and syringe pump
To change AnaConDa:
• Open a new AnaConDa filter.
• Stop syringe pump.
• Disconnect the anaesthetic agent line from the syringe and close the syringe with the
red cap.
• Detach the gas sampling line from the current AnaConDa and seal the gas sampling
port with the attached cap.
• Remove the purple ‘monitor’ label from the new AnaConDa device and attach the gas
sampling line.
• Uncap the syringe and attach the new AnaConDa anaesthetic agent line, and prime
with a bolus of 1.2ml via the pump.
• Remove the used AnaConDa device from the ventilator circuit using ‘manual
disconnect’ option, clamp if needed and disconnect the ventilator side first and then
the patient side.
• Insert the new AnaConDa into the ventilator circuit between the Y-piece and the
patient/nebuliser.
• Connect the anaesthetic agent line of the new AnaConDa to the isoflurane syringe.
• Discard the used AnaConDa and anaesthetic line in general clinical waste as per
wastage SOP.
If COVID-19 patient : please ensure ICU doctor present – as the ETT will need
clamping – and the ventilator will need stopping during change of AnaConDa.
Removal of device
• Stop the syringe pump. Remove syringe and cap securely with red-cap.
• Dispose of capped Isoflurane in a yellow HSW disposal container.
• Replace AnaConDa device with HME or convert to water-humidified circuit.
• Cap AnaConDa device and dispose of in a clinical HSW container.
Image shows
1.AnaConDa mounted at Y-
piece
2.Scavenge hose connecting
2 from Draeger ventilator flow
sensor to Flurabsorb with
1 adaptor and bacterial filter.
Note.
Breathing circuit from Hospital
Bacterial filter from Hospital.
Isoflurane syringe
AnaConDa
Scavenge hose
Gas sample return line
Flurabsorb
Gas sampling line from Data cable exiting Data cable entering back
AnaConDa connecting to Draeger Scio to patient of patient monitor.
water trap on back of monitor and return gas
Draeger Scio. sample line exiting Scio
to port on Flurabsorb
adaptor.
Royal Site: Philips Intellivue MX550 with Draeger Water trap and MP40 with Philips
Philips Intellivue Return gas tubing
MX550 with Draeger Philips Intellivue outlet to flurabsorb
water trap for sample line MP40 water trap & exiting the back of
connection sample line port Intellivue MP40