<CHECKING LIST>
1.Check the “E V E N T V I W E R ” whether some error is there or not.
(Refer to the SERVICE MANUAL TSFE-U02002)
4.If your machine have 10GB HDD, check the partition like next.
C:¥→6GB, D:¥→4GB. If partition is not like this way, try to divide the partition.
Refer to the SERVICE TIPS “TSU-E02051”
5.Confirm the normal working when you reboot the machine at least 5 times.
And then, check the normal working of probe exchanging.
8.Check the Color signal and Doppler sound by seeing the CAROTID with L040
PROBE.
Normally, red color signal can be heard from right side speaker.
And blue color signal can be heard from left side speaker.
Check the “color contrast” and “sound” .