You are on page 1of 1

400 VAC 3 PHASES + NEUTRAL

POWER
SUPPLY
MAIN CONTACTOR
230 VAC
400 VAC 3 PHASES 230 VAC 230 VAC
Mx MC CB5
24V OF1 OF2 OF4 C60N TRANSFORMER
3 A 2P/2D PROTECTION
125 A / 4P
CB1 CB2 CB4 CURVE C
C60N C60N C60N
100 A 3P/3D 10 A 2P/2D 10 A 2P/2D Transformer
CURVE D CURVE C CURVE C 230V-24Vac
300 mA 30 mA 30 mA
H07RN-F H07RN-F H07RN-F

100 VA

CB6
CT01 CT02 C60N REMOTE SUPPLY
4 A 2P/2D PROTECTION
CURVE C

24 VAC

EJB

PJB: 01-02-03 PJB: 04-05 PJB: 09-10


JB

SECTION 4G25 mm² SECTION 3G2.5 mm² SECTION 3G2.5 mm²

TO MAGNUM C TO CAMARGUE TO DIGITAL SYSTEM


GENERATOR POWER SUPPLY UPS POWER SUPPLY
POWER SUPPLY 24 VAC

24VAC 101

BNC 09
BNC 01

R01.1 Generator contact


R01.3 R01.4
EMERGENCY "X-Ray ON"
BUTTON

108
106

EMB
BNC 10

107
Electrical cabinet BNC 02
front panel

BNC 03 BNC 05 BNC 06 BNC 08 BNC 11 BNC 13


104

POWER OFF LIGHT

POWER ON LIGHT

X-RAY
102

103

OFF ON ON ON

R01.2

ON OFF Green Red Red White


105

BNC 04 BNC 07 BNC 12 BNC 14

Electrical cabinet Above access doors


front panel of X-RAY room

Mx CT01 CT02 R01

0V 100

FOR CAMARGUE TABLE & MAGNUM C GENERATOR

09 / 06 / 20 CREATION BRUNEAU S. CHAVAUDRET J.C. Ø


DATE MODIFICATIONS DRAWN CHECKED REV.

JC CHAVAUDRET 09/06/2020
TITLE: POWER DIAGRAM CAMARGUE - MAGNUM C - DIGITAL SYSTEM
PRODUCT CODE : 8D
S/ASSY: INTERNAL REFERENCE
SEE : CABINET E8D00005
A BRAND OF DMS IMAGING

SHEET
DOCUMENT REFERENCE : S8D 59 005
1/1