Has electrical safety

2 years

Visual check – Check that the following parts of the device are in a good and safe condition: complete
and legible labelling.

Visual check – Check that the following parts of the device are in a good and safe condition: housing
without damage.

Visual check – Check that the following parts of the device are in a good and safe condition: front panel
(since the front panel prevents fluid from entering it is important that the front panel is in a good
condition and that its full surface adheres to the housing).

Visual check – Check that the following parts of the device are in a good and safe condition: power
supply cord and mains connector insulation in a defect-free state, contacts are clean and show no
corrosion.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Display accuracy – Check the temperatures displayed: Upper target indication: 43.7°C to 44.3°C.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Display accuracy – Check the temperatures displayed: Lower target indication: 41.2°C to 41.8°C.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Temperature Control – (select 41°C) Check: The warmer must be able to be started and it must not
change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Temperature Control – (S1 to 40.5°C; S2 to 40.5°C) Check: The warmer must change over to the alarm
status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Temperature Control – (S1 to 40.5°C; S2 to 40.5°C) Check: After pressing the ‘arrow up’ key, the display
must show E26 alternatingly.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Temperature Control – (S1 and S2 to 41.5°C) Check: The warmer must be able to be started and it must
not change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Temperature Control – (S2 to 40.5°C; S1 to 40.5°C) Check: The warmer must change over to the alarm
status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Temperature Control – (S2 to 40.5°C; S1 to 40.5°C) Check: After pressing the ‘arrow up’ key, the display
must show E26 alternatingly.
Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Excessive temperature alarms – (S1 to 44°C; S2 to 41.5°C; S3 to 40.5°C) Check: The warmer must be able
to be started and it must not change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Excessive temperature alarms – (S1 to 45.7°C) Check: The warmer must change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Excessive temperature alarms – (S1 to 45.7°C) Check: After pressing the ‘arrow up’ key, the display must
show E16 alternatingly.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Excessive temperature alarms – (S1 back to 44°C) Check: The warmer must be able to be started and it
must not change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Excessive temperature alarms – (S2 to 45.7°C) Check: The warmer must change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Excessive temperature alarms – (S2 to 45.7°C) Check: After pressing the ‘arrow up’ key, the display must
show E16 alternatingly.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Excessive temperature alarms – (S2 back to 41.5°C) Check: The warmer must be able to be started and it
must not change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Excessive temperature alarms – (S3 to 45.7°C) Check: The warmer must change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms –
Excessive temperature alarms – (S3 to 45.7°C) Check: After pressing the ‘arrow up’ key, the display must
show E16 alternatingly.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms – NTC
cable break alarms – (S1, S2 and S3 to 40.5°C) Check: The warmer must be able to be started and it must
not change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms – NTC
cable break alarms – (Set the switch S1 to the ‘disconnected symbol key’) Check: The warmer must
change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms – NTC
cable break alarms – (Set the switch S1 to the ‘disconnected symbol key’) Check: After pressing the
‘arrow up’ key, the display must show E16 alternatingly.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms – NTC
cable break alarms – (S1 back to 40.5°C and restart the warmer) Check: The warmer must be able to be
started and it must not change over to the alarm status.
Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms – NTC
cable break alarms – (Set the switch S2 to the ‘disconnected symbol key’) Check: The warmer must
change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms – NTC
cable break alarms – (Set the switch S2 to the ‘disconnected symbol key’) Check: After pressing the
‘arrow up’ key, the display must show E16 alternatingly.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms – NTC
cable break alarms – (Set the switch S2 back to 40.5°C and restart the warmer) Check: The warmer must
be able to be started and it must not change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms – NTC
cable break alarms – (Set the switch S3 to the ‘disconnected symbol key’) Check: The warmer must
change over to the alarm status.

Display accuracy, temperature control, excessive temperature alarms and NTC cable break alarms – NTC
cable break alarms – (Set the switch S3 to the ‘disconnected symbol key’) Check: After pressing the
‘arrow up’ key, the display must show E16 alternatingly.

Inspection of the heating profile – Visual check – Check if connecting socket, connecting housing and the
silicone rubber jacket* of the heating profile are in good condition and are not damaged.

Inspection of the heating profile – Check the applicability of the used mutimeter – Check that the
resistance value is within 2000 +/- 10 .

Inspection of the heating profile – Carry out the measurement – Record the resistance values for NTC11.

Inspection of the heating profile – Carry out the measurement – Record the resistance values for NTC12.

Inspection of the heating profile – Carry out the measurement – Record the resistance values for NTC2.

Inspection of the heating profile – Carry out the measurement – Record the room temperature.

Inspection of the heating profile – Carry out the measurement – Check that the resistance values
measured and the resistance value of the room temperature is within a 200 Ohm range.

Inspection of the heating profile – Carry out the measurement – Check that the resistance value of the
heating wire is within 6.5 +/- 0.5 .

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 1 (H1) – Check that the following takes place: the display shows "H1".

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 1 (H1) – Check that the following takes place: the power LED flashes.

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 1 (H1) – Check that the following takes place: the alarm LED flashes.

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 1 (H1) – Check that the following takes place: the acoustic alarm sounds.
Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 2 (H2) – Check that the following takes place: the display shows "H2".

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 2 (H2) – Check that the following takes place: the power LED flashes.

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 2 (H2) – Check that the following takes place: the alarm LED flashes.

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 2 (H2) – Check that the following takes place: the acoustic alarm sounds.

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 3 (H3) – Check that the following takes place: the display shows "H3".

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 3 (H3) – Check that the following takes place: the power LED flashes.

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 3 (H3) – Check that the following takes place: the alarm LED flashes.

Inspection of combination control unit/heating profile – Operation of the excessive temperature cut-offs
– Cut-off 3 (H3) – Check that the following takes place: the acoustic alarm sounds.

Inspection of combination control unit/heating profile – Function of Auto restart after power supply
interruption – Check that the warmer automatically switches on again with the same setting and starts
without you having to press any keys.

Calibration of the System Testbox – Measuring the reference resistance value – Check that the
resistance value is within 2000 +/- 5.

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: ‘disconnected symbol’: 13369 +/- 150 . (S1)

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: 40.5°C: 2740 +/- 10 . (S1)

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: 41.5°C: 2647 +/- 10 . (S1)

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: 44°C: 2428 +/- 10 . (S1)

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: 45.7°C: 2292 +/- 10 . (S1)

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: ‘disconnected symbol’: 13369 +/- 150 . (S2)

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: 40.5°C: 2740 +/- 10 . (S2)
Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: 41.5°C: 2647 +/- 10 . (S2)

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: 45.7°C: 2292 +/- 10 . (S2)

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: ‘disconnected symbol’: 13369 +/- 150 . (S3)

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: 40.5°C: 2740 +/- 10 . (S3)

Calibration of the System Testbox – Measuring the resistance values of the temperature presets – Check
the resistance value of the according position is within: 45.7°C: 2292 +/- 10 . (S3)