96M12752 O Handling of the sensor head

• The GT2-70 Series and peripheral devices are precision


machines. Do not drop, or cause any other impact to these
devices. Doing so may cause damage or malfunction.
• Do not apply weight greater than what is listed below to
the spindle. Do not apply rotational torque. Otherwise the
High-accuracy Digital Contact Sensor spindle may break.

GT2-70 Series Instruction Manual


• The GT2-P12KL(F)/P12L(F)/PA12K(L)/PA12(L) spindle
can rotate a maximum of about 4.5°, the GT2-S1/S5
spindle can rotate a maximum of about 2°. When using an
offset contact (OP-77683) or similar, if pressure in the
rotational direction is applied to the contact, the
measurement position may change.
• Although the GT2-PA12K/PA12/H12K/H12KF/H12/H12F/
H32/H50/A12K/A12/A32/A50 has a protection rating of
IP67, avoid using it immersed in water or in places where
liquid such as oil may come into contact with it.
• Although the GT2-P12K(F)/P12(F) has a protection rating
of IP67G/NEMA Type 13 and the GT2-S1/S5 has a
protection rating of IP67G, some types of oil may damage
the device.
NOTICE • If dust, metal powder, or a similar substance becomes
attached to the spindle of the GT2-PA12K/PA12,
mechanical responses may become slow. If this happens,
replace the dust seal. Refer to the instruction manual
included with the replacement dust seal (OP-87932) for
instructions on how to replace the dust seal.
• Applying an external magnetic field to the GT2-S1/S5 may
affect the measured value. Avoid installations close to
Q Symbols motors and other such devices that generate strong
The following symbols alert you to matters concerning the prevention of magnetic fields.
human injury and product damage. Be sure to read these messages carefully. • Iron and other such ferromagnets being close to the GT2-
S1/S5 may affect the measured value. In the case of an
Indicates a hazardous situation which, if not avoided, installation next to a ferromagnet, perform calibration after
WARNING
could result in death or serious injury. installation.
Indicates a situation which, if not avoided, could result in For the calibration procedure, see "Calibration setting
NOTICE mode" in the GT2-70 Series User's Manual.
product damage as well as property damage.

Safety Information for GT2-70 Series


30 N 100 N 1N
If the following conditions are encountered, immediately turn
OFF the power. Continuing to use the GT2-70 Series under
these abnormal conditions may cause equipment failure. Q Other precautions
WARNING O Effects of surrounding air temperature
• When water or foreign matter enters the controller
• When the GT2-70 Series is dropped or the housing is damaged To use the GT2-70 Series with high accuracy, do not use the GT2-70 Series in
• When the GT2-70 Series produces smoke or an abnormal smell an environment in which the surrounding air temperature changes sharply.
It will take about 20 minutes when using the 1 mm/5 mm type, 40 minutes for the
• Do not use the GT2-70 Series with a voltage other than 12 mm type, and 60 minutes for the 32 mm/50 mm type for the internal temperature
specified voltage, as this may cause fire, electric shock to stabilize when the surrounding air temperature has changed by 10°C.
or equipment failure. O Warming up
• Do not disassemble or modify the GT2-70 Series. This The circuit is not stable immediately after the power turns ON, which sometimes
may cause fire or electric shock. causes the indicated value to gradually fluctuate. Wait for about 20 minutes when using
WARNING the 1 mm/5 mm type, 5 minutes when using the 12 mm type and about 10 minutes
• Do not use this product for the purpose to protect a
human body or a part of human body. when using the 32 mm/50 mm type after the power turns on before starting operation.
• This product is not intended for use as explosion-proof O GT2-PA12K(L)/PA12(L) (See illustration 1 for the pneumatic circuit)
product. Do not use this product in a hazardous location • This model uses a single-acting cylinder. The internal spring will
and/or potentially explosive atmosphere. return the sensor head to its home position.
• The air supply hole cannot be removed.
• Be sure to turn OFF the power of the GT2-70 Series and any • Use an air filter, mist separator, etc. to provide clean dry air. Empty the
connected devices before connecting or disconnecting the drainage from the filter regularly, before it exceeds the specified line.
cables. Otherwise, there may be a risk of damage. • Before connecting the air tube to the air supply hole, be sure to blow
• Do not turn OFF the power while setting parameters. plenty of air through the pipes (flushing) to remove any foreign matter.
Otherwise, the settings may be partially or completely lost. • Make sure that the air pressure of the supplied air is constant and in the range of 0.24
• At startup and during operation, be sure to monitor the to 0.26 MPa for the GT2-PA12K/PA12 and in the range of 0.05 to 0.07 MPa for the GT2-
function performance of the GT2-70 Series. PA12KL/PA12L. Use a precision regulator to control the air pressure. If the air pressure
• Do not modify the GT2-70 Series or use it in any way is below the minimum pressure given above, the spindle may not extend fully.
• When supplying air (with the spindle extended), up to 3 L/min of air will be emitted from the tip
other than described in the specifications.
• The function performance of products used or modified
for the GT2-PA12(K) and up to 0.4 L/min of air will be emitted from the tip for the GT2-PA12(K)L.
• The measuring ability of the device changes according to the air pressure
in this way cannot be guaranteed. of the air supply. Refer to "Specifications" (page 15) for details.
• When the GT2-70 Series is used in combination with other
instruments, function performance may be degraded, O GT2-A12K(L)/A12(L)/A32/A50 (See illustration 1 for the pneumatic circuit)
depending on operating conditions and the surrounding • This model uses a single-acting cylinder. The internal spring will return the sensor head to
environment. Use the GT2-70 Series after fully studying its home position when air pressure is removed. DO NOT supply air to the exhaust valve.
the effect of combined use with other instruments. • The air cylinder cannot be removed.
• Do not expose the GT2-70 Series and peripheral devices • Use the regulator to supply the stable air pressure to the sensor head.
to sudden temperature change. This may cause • Use the air filter and mist separators to supply dry air. Empty the
condensation, damaging the equipment. drainage from the filter regularly, before it exceeds the specified line.
• Make sure that the air pressure of the supplied air is constant and in the range of 0.25 MPa to 0.5 MPa.
O Environmental conditions • The coupling socket and the exhaust valve that are supplied with the
To use the GT2-70 Series properly and safely, do not air cylinder are dedicated for this product. They cannot be removed.
NOTICE
install the GT2-70 Series in the following locations. Use Mist separator
of this equipment in an improper environment may AIR
cause equipment failure. Pressure
• Locations outside source
Air supply part
• Locations with high humidity, a large amount of dust, Air filter Precision regulator Solenoid Speed
or poor ventilation Illustration 1 - Pneumatic circuit valve controller
• Locations where the temperature rises excessively O Power supply
due to direct sunlight, etc. • Noise superimposed on the power supply may cause malfunction. Be sure to
• Locations near corrosive or flammable gas use the DC stabilized power supply provided with an insulation transformer.
• Locations where the GT2-70 Series is directly • In the case of a commercially available switching regulator, be sure to
subjected to vibration or impact ground the frame ground terminal or the ground terminal.
• Locations where water, oil or chemicals may come O About dust boot
into contact with the GT2-70 Series When the dust boot of the sensor head is damaged, use an optional dust boot sold
• Locations where static electricity may easily occur
separately. (For the GT2-P12K(F)/P12(F)/H12K(F)/H12(F)/A12K/A12: OP-84332 (Material:
O Noise countermeasures NBR (attached when shipped)). For the GT2-P12K(F)/P12(F): OP-87859 (Material:
Installation near a source of electrical noise such as a Fluororubber). For the GT2-H32/A32: OP-84459 (Material: NBR (attached when shipped)).
power source or a power cable may cause malfunction For the GT2-H50/A50: (Material: NBR (attached when shipped)). For the GT2-S1: OP-88063
or failure of the equipment. Adopt appropriate (Material: NBR). For the GT2-S5: OP-88065 (Material: NBR)). Attach the dust boot correctly.
countermeasures against noise by using a noise filter The dust boot may deteriorate, depending on the environment it is
or wiring cables in separate ducts, attaching insulation used in. In this case, replace the dust boot regularly.
to the amplifier or the sensor head, etc. * Do not use the dust boot on the GT2-P12KL/P12L/PA12K(L)/PA12(L)/H12KL(F)/H12L(F)/H32L.

1 GT2-70-M-E
Precautions on Regulations and Standards Q Sensor head parts
Q CE Marking O Sensor head
O EMC Directive (2004/108/EC) GT2-P12K/P12 GT2-P12KL/P12L GT2-P12KF/P12F GT2-PA12K/PA12
• Applicable standard EMI: EN61326-1, Class A
EMS: EN61326-1
• This product is designed for use in industrial environments.
Remarks: This specification does not give any guarantee that the end-product with this product
incorporated complies with the essential requirements of EMC Directive. The manufacturer of the end- Sensor head x 1 Sensor head x 1 Sensor head x 1 Sensor head x 1
Precautions on handling x 1 Precautions on handling x 1 Precautions on handling x 1 Precautions on handling x 1
product is solely responsible for the compliance on the end-product itself according to EMC Directive. Nut x 1
Q UL Certificate Key spanner x 1
UL File No. E120439 (Category: NRNT2/NRNT8) GT2-S1 GT2-S5 GT2-PA12KL/PA12L
• Use the power supply with Class 2 output defined in NFPA70
(NEC: National Electrical Code).
• GT2 series are evaluated as open type devices. These products Sensor head x 1 Sensor head x 1 Sensor head x 1
can not be a part of an enclosure. Precautions on handling x 1 Precautions on handling x 1 Precautions on handling x 1
GT2-H12KF/H12F GT2-H12KLF/H12LF
Package Contents Sensor head x 1 Sensor head x 1
Precautions on handling x 1 Precautions on handling x 1
The product package of each model should include the following items. Nut x 1 Nut x 1
Check that all items are included before use. Key spanner x 1 Key spanner x 1

Q Amplifier parts GT2-H12K/H12 GT2-H12KL/H12L GT2-A12K/A12 GT2-A12KL/A12L


O DIN rail mount type
GT2-71N/71P/71CN/71CP (Main unit) GT2-72N/72P/72CN/72CP Sensor head x 1 Sensor head x 1 Sensor head x 1 Sensor head x 1
GT2-71MCN/71MCP (Main unit analog output) (Expansion unit) Precautions on handling x 1 Precautions on handling x 1 Precautions on handling x 1 Precautions on handling x 1

HI
GO
P-H
B-H
P-P
CALC
HI
GO
P-H
B-H
P-P
CALC
GT2-H32 GT2-H32L GT2-A32
LO LO

SET SET
1 1
2 2
3 3

TIM TIM
SET HI SET
HI PV
PV
LO PRES LO PRES
ET
ET MO
MO DE
DE SER
SER IES
IES

Amplifier x 1 Sensor head x 1 Sensor head x 1 Sensor head x 1


Amplifier x 1 Instruction manual x 1 Precautions on handling x 1 Precautions on handling x 1 Precautions on handling x 1
Head connector label x 1 GT2-H50 GT2-A50
*Used for the calculation only mode.

GT2-71D (Pulse output)


Sensor head x 1 Sensor head x 1
Precautions on handling x 1 Precautions on handling x 1
O Sensor head connection cable
GT2-CH2M/5M/10M/20M GT2-CHL2M/5M/10M/20M GT2-CHP2M/5M/10M
Amplifier x 1 Instruction manual x 1 (M8 straight connector) (M8 L-shaped connector) (M8 oil-resistant straight connector)
Switch cover label x 1
O Amplifier connection cable
Use for connector type amplifiers (GT2-71CN/71CP/72CN/72CP/
71MCN/71MCP).
GT2-CA2M/CA10M
• The M8 L-shaped connector type cable cannot be
used with the GT2-H32(L)/H50/A32/A50.
• When a protection rating of IP67G and NEMA Type 13
enclosure is required for GT2-P12K(F)/P12(F), or a
NOTICE
O Panel mount type protection rating of IP67G is required for GT2-S1/S5
use an M8 oil-resistant straight connector type cable.
GT2-75N/75P (Main unit) GT2-76N/76P (Expansion unit) Other cables do not meet the IP67G protection rating
or NEMA Type 13 enclosure rating.
P-H HI P-H
HI B-H B-H
P-P GO P-P
GO CALC CALC

We have taken all possible precautions in packaging; however, if


LO LO

1
2
3
TIM
Amplifier x 1 1
2
3
TIM
HI
Amplifier x 1
any parts are found to be defective or broken, please contact your
HI RV
RV LO PRESET
LO PRESET

nearest KEYENCE sales office.

Panel mounting bracket x 1 Panel mounting bracket x 1

Front protective cover x 1 Front protective cover x 1

Power cable (2 m) x 1 Expansion cable (50 mm) x 1


(Number of conductors: 12)

Instruction manual x 1

Head connector label x 1 I/O cable (2 m) x 1


*Used for the calculation only mode. (Number of conductors: 10)

GT2-70-M-E 2
Mounting the Amplifier and Wiring O Panel mount type (Main unit)
This section describes how to mount the panel mount type:
Q Mounting the amplifier GT2-75N/75P (Main unit).
O DIN rail mount type (Main unit) 1 Create a panel opening for mounting referring to the dimensions below.
This section describes how to mount the DIN rail mount type: 45

(Main unit: GT2-71(C)N/71(C)P/71MCN/71MCP) (unit: mm)


Panel thickness of mounting area
1 Fit the tab of the lower part of the main unit to (3)
1 to 6 mm

the DIN rail. While inserting the main unit in X=24 x (Number of amplifiers -1)+21

X
the direction of Arrow (1), push the body
down in the direction of Arrow (2). (2) (1)

69.5 min.

2 To detach the amplifier, while pushing the main unit in the direction

+0.4
21 -0
of Arrow (1), pull the body up in the direction of Arrow (3).
When using a fixture (Optional item: OP-76877), mount as shown below.
45 Panel cut dimension

2 Insert the main unit, back end first, into the opening of the panel.
Fixture: OP-76877 3 Insert the panel mounting bracket in the direction shown below into the main
O DIN rail mount type (Expansion unit) unit from the back, then fit the front protective cover to the main unit face.
This section describes how to mount the DIN rail mount type Front protective cover
(Expansion unit: GT2-72N/72P/72CN/72CP). Amplifier
Panel
The expansion unit can only be used in addition to the main unit.
Up to 14 expansion units can be added to one main unit. Panel mounting bracket

• When adding an amplifier (expansion unit), be sure


the power of both main and expansion units is OFF
before operation. Mounting when the power is ON
may damage the equipment.
• Be sure to completely connect an expansion unit to a To detach the panel mounting bracket, pull it while
main unit. Oblique or improper connections may pushing apart the tabs provided on its both ends.
damage the equipment.
• When connecting expansion units, some of the
specifications are restricted depending on the O Panel mount type (Expansion unit)
Tab
NOTICE
number of units to be connected. Refer to page 14,
"Specifications" for details for details. This section describes how to mount the panel mount type:
• An expansion unit of different output type cannot be GT2-76N/76P (Expansion unit).
added (for example, an expansion unit of NPN output The expansion unit can only be used in addition to the main unit.
cannot be added to a main unit of PNP output). Up to 14 expansion units can be added to one main unit.
• No expansion unit can be connected to the pulse • Turn OFF the power before connecting the expansion
output type (GT2-71D). cable. Inserting or pulling the cable when the power
• An expansion unit of the DIN rail mount type cannot is turned ON may damage the equipment.
be added to a main unit of the panel mount type. • Be sure to completely connect the expansion cable.
Oblique or improper connections may damage the
1 Detach the expansion cover of the main unit. equipment.
Main unit • When connecting expansion units, some of the
NOTICE specifications are restricted depending on the
Expansion cover number of units to be connected. Refer to Page 15,
"Specifications" for details.
• An expansion unit of different output type cannot be
2 Mount the expansion unit to be added to the DIN rail. added (for example, an expansion unit of NPN output
cannot be added to a main unit of PNP output).
Refer to above, "DIN rail mount type", for details about how to mount. • An expansion unit of the panel mounting type cannot
3 Push and fix the expansion unit to the connector of the main unit until it clicks. be added to a main unit of the DIN rail mounting type.
Main unit

Expansion unit
1 Create panel openings for mounting according to the number of
expansion units to be mounted (the expansion unit to be added).
Refer to above, "Panel mount type (Main unit)", for details about
Connector
panel cut dimensions.
4 Mount the end units (OP-26751: a set of two pieces) sold 2 Mount the amplifier (the expansion unit to be added) to the panel.
separately on both sides of the amplifiers (the main unit and
the expansion unit), then fix the end units with screws on the Refer to above "Panel mount type (Main unit)", for details about how
upper part of each end unit (2 points x 2 units). to mount.
The mounting method of the end unit is the same as that of the amplifier. 3 Connect the amplifiers (the main unit and the expansion unit) with an
expansion cable.
End unit
Expansion
End unit cable Expansion cable

When installing the amplifier side by side, the 300 mm


expansion cable (OP-35361) is required. If using even
NOTICE
one 300 mm expansion cable, the number of expansion
units is limited to four.

3 GT2-70-M-E
Q Wiring of amplifier Connecting and Mounting the Sensor Head
NOTICE Insulate unused I/O wires individually.
Q Names of parts of the sensor head
O How to connect the power cable to the DIN-mount quick disconnect type ● GT2-P12K(L)/P12(L) ● GT2-P12KF/P12F ● GT2-PA12K/PA12
Quick disconnect type amplifiers must be connected with the power
Air supply hole
cable (I/O cable) GT2-CA2M/CA10M (sold separately). Relay connector Relay connector

Cable Cable Cable


Cable between connector Cable connector Cable between connector
click the sensor between the the sensor
head and relay sensor head head and relay
connector and relay connector
connector
Connect the cables with Operation indicator Operation indicator Operation indicator
the white points aligned
Mounting fixture
Mounting fixture
(When connecting the power cable) (When disconnecting the power cable) Dust boot※
Dust boot
O How to connect the power cable to the panel mount type Dust boot

The accessory power cable (I/O cable) must be connected to the Spindle Spindle
amplifier of the panel mount type. Dust seal
Contact Contact

* Not included in the GT2-P12KL/P12L.


click
O GT2-S1 O GT2-S5
Power cable Cable connector Cable connector
(When connecting the power cable) (When disconnecting the power cable)

• The number of power (I/O cable) conductors is different


Reference
between the main unit and the expansion unit (Main unit:
12 conductors, Expansion unit: 10 conductors).
• The power of the expansion unit is supplied through the
expansion cable being connected to the main unit. Position indicator Position indicator

Operation indicator Operation indicator


O Power cable (I/O cable)
The following illustrates the power cable (I/O cable) (also applicable
to the panel mount type). Mounting fixture
Refer to Page 17 of this manual for details about I/O circuits.
Dust boot
GT2-71N/71P/71CN/71CP (Main unit)
Brown Contact
Blue 10-30 V *
Black O GT2-H12K(L)(F)(LF)/H12(L)(F)(LF)/H32(L)/H50
HIGH output
White Cable connector
LOW output
Gray
GO output Head indicator
Orange Cable connector
HH output
Green Head indicator
LL output
Pink
PRESET input
Violet
BANK A input
Pink/Violet
TIMING input
Yellow
BANK B input
Red
RESET input
Dust boot
mounting fastener
* Supply 20-30V DC power when using expansion units.
Spindle
GT2-72N/72P/72CN/72CP (Expansion unit) mounting fastener* 1
Dust boot*3
Brown
Not connected * Dust boot*2
Blue Contact
Not connected * Contact
Black Contact
HIGH output
White *1 A mounting bracket is attached to *3 Not included in the GT2-H32L.
LOW output GT2-H12KF/H12F/H12KLF/H12LF.
Gray
GO output *2 Not included in the GT2-H12KL/H12L/H12KLF/H12LF.
Orange
HH output
Green
LL output
O GT2-A12K(L)/A12(L)/A32/A50
Pink Coupling socket
PRESET input
Violet
BANK A input
Pink/Violet Air cylinder
TIMING input Coupling socket
Yellow
BANK B input
Red Cable connector
RESET input
Cable connector
* Available only on the GT2-72CN/72CP (Quick disconnect type). Head indicator
Exhaust valve
GT2-71MCN/71MCP (Analog output) Air cylinder
Brown Head indicator
Exhaust valve
Blue 20-30 V
Black
HIGH output
White
LOW output Dust boot
Gray
GO output Mounting fixture Mounting fixture
Orange Spindle
Analog output
Green Dust boot*
Analog GND Dust boot
Pink Contact Contact
PRESET input
Violet Contact
BANK A input
Pink/Violet * Not included in the GT2-A12KL/A12L.
TIMING input
Yellow
Red
BANK B input Q Connecting the sensor head connection cable
RESET input
O For GT2-P***/PA***/S1/S5
GT2-71D (Pulse output)
Brown 1 Insert this device's cable into the cable
Blue 10-30 V connector on the relay connector cable.
Black
A phase
White
B phase
Orange
Z phase
Violet
Gray
-
A phase
-
2 Secure the connector with this device's
B phase cable screw.
Green -
Z phase
Pink
Current position returning input

GT2-70-M-E 4
• When connecting the connector, be sure to insert it Q Mounting the sensor head
straight, and tighten it securely. (Recommended O Mounting directly to the jig
tightening torque: 0.4 to 0.5 N•m*) Before mounting the sensor head directory to the jig, create a hole in the jig. Attach the sensor
If the connection is not tight enough, the connector head using the optional head mounting bracket sold separately. A mounting bracket is provided
may be loosened by vibration or other causes, with the GT2-P12KF/P12F/H12K(L)F/H12(L)F as standard. Attach it with the supplied nut.
leading to a connection failure. O GT2-P12K(L)/P12(L)/PA12K(L)/PA12(L)/ O GT2-H32(L)/H50/A32/A50
Mounting
NOTICE (* After tightening it firmly by hand, use pliers or other illustration S1/S5/H12K(L)/H12(L)/A12K(L)/A12(L) Optional head mounting
tools to rotate it about 30° for further tightening.) bracket D (OP-84327)
• When the head is attached to a moving part, and the Optional head mounting
cable will be repeatedly bent, ensure that the cable bracket A (OP-76874)
between the sensor head and relay connector does
not bend. Instead, bend the sensor head cable
Tightening sleeve Nut
connecting the relay connector and the amplifier. Tightening sleeve Nut

O For units other than GT2-P***/PA***/S1/S5


1 Insert this device's cable into the cable Key wrench Key wrench

connector of the sensor head. Optional head mounting When mounting the GT2-
Arrow bracket C (OP-84396) H32L facing upward, be
sure to use the mounting
holes on the side of the
NOTICE sensor head. If the
2 Secure the connector with this Tightening sleeve Nut sensor head is mounted
device's cable screw. using head mounting
bracket D, the spindle
Key wrench
may not fully extend.

1 Cut the jig to create a sensor head mounting hole referring to the
• When connecting the connector, be sure to insert it illustration below.
straight, and tighten it securely. (Recommended O GT2-P12K(L)/P12(L)/PA12K(L)/PA12(L)/ O GT2-H32(L)/H50/A32/A50
tightening torque: 0.4 to 0.5 N·m*)
If the connection is not tight enough, the connector S1/S5/H12K(L)/H12(L)/A1K2(L)/A12(L)
0.5 to 1.0*2
may be loosened by vibration or other causes, ±1°
0.5 to 1.0*2
±1°
NOTICE 45° 45°
leading to a connection failure. A A

(* After tightening it firmly by hand, use pliers or other Tightening Nut Tightening
Nut
tools to rotate it about 30o for further tightening.)
sleeve mounting sleeve
1 insertion insertion mounting
*
27 ) side side
• The M8 L-shaped connector (GT2-CHL*M) cannot be
+0.0 05 side .1 6
+00.00 side
8 ( +0.0 4+

φ1 0G φ1 5.5 to 11.3*2
used with the GT2-H32(L)/H50/A32/A50. φ0.025 A
5.5 to 11.3*2
φ0.025 A (Unit: mm)
(Unit: mm)
To disconnect the sensor head connection cable, reverse the steps above. *1 For head mounting bracket C *2 Processing accuracy : ±0.05
φ10 +0.1
+0.005

Q Connecting the amplifier O GT2-P12KF/P12F/H12K(L)F/H12(L)F


O DIN rail mount type 0.5 × 45°
recession
1 Remove the lock cover of the connector of the sensor head Nut
mounting
connection cable, and insert the sensor head connection cable φ10
+0.1 05
+0.0
Sensor head
insertion side
side
into the connector on the side of the amplifier until it clicks. 5.5 ∼ 11
(Unit: mm)
Lock cover

Click
2 Insert the tightening sleeve into the hole (from the Chamfered side),
Unlocked
and loosely tighten with the nut.

Sensor head

Tightening sleeve
To disconnect the sensor head connection cable,
Chamfered side
NOTICE press and hold the lock lever on the side of the
connecter and disconnect the cable. Chamfered side

2 Put the lock cover on the connector, and lock the lock cover. Nut
Lock cover
Nut

Locked
In the illustration on the left above, if the nut is fully tightened when
NOTICE
the sensor head is not yet inserted, the fixture will become deformed.
3 Insert the sensor head in the tightening sleeve. While securing the tightening
sleeve with the included key wrench, tighten the nut with a separate wrench.
O Panel mount type
1 Remove the lock cover of the connector of the sensor head
Wrench
connection cable, and insert the sensor head connection cable into
the connector on the back of the amplifier until it clicks.
Lock cover
Key wrench

Click Dust boot


Unlocked

Tightening the nut while fixing the sensor head may lead to damage.
• The tightening torque for mounting bracket A (OP-76874), mounting
Orient the connector so that the lock lever is at the left side.
bracket C (OP-84396), and the nuts supplied with the GT2-H12K(L)F/
To disconnect the sensor head connection cable, H12(L)F/P12KF/P12F is between 5 and 7 N•m (the recommended
NOTICE press and hold the lock lever on the side of the tightening torque is 5 N•m) for all products other than the GT2-S5
NOTICE
connecter and disconnect the cable. and is between 4 and 5 N•m (the recommended tightening torque is 4
N•m) for the GT2-S5. Apply tightening torque between 15 and 18 N•m
(20 to 23 N•m when using the GT2-A32/A50) to head mounting
2 Put the lock cover on the connector, and lock the lock cover. bracket D (OP-84327). Use the wrench that matches the nut width.
Lock cover
• Care must be taken not to damage the dust boot when tightening.
O Using mounting bracket F
Locked
1 Refer to the illustration below and cut the jig to create a sensor head mounting hole.
• GT2-P12K(L)/P12(L)/PA12K(L)/PA12(L)/S1/S5
8.5
M4
x0
.7

Make sure that the metal plating


mounting bracket F attaches to is made
0
.2 0 of SUS (SUS303 recommended) and is
+ 0 .0 0 5 at least 5 mm thick.
0
8 +

5 GT2-70-M-E
2 Loosen the screw on the side of mounting bracket F using the O Mounting to the mounting holes
supplied hexagonal wrench. The mounting holes on the side of the GT2-H32(L)/H50/A32/A50
unit can be used to install to the side of a table or other surface.
Mounting illustration
M4 screws (Mounting hole dimensions)

23 ± 0.15 mm
M4 tap

3 Align mounting bracket F with the hole made in step 1 and secure it
using the supplied hexagonal wrench.
• Roughness of the mounting surface may cause the
sensor to tilt, and sufficient accuracy may not be
obtained. When performing a high-accuracy
NOTICE
measurement, use a jig to mount the sensor head.
"Mounting directly to the jig" (page 5)
• Apply tightening torque between 1.2 and 1.5 N•m.

Q Installing the Air Tube


4 Insert the sensor head, and tighten the screw loosened in step 2 using O Compatible air tube
the supplied hexagonal wrench to secure. The recommended Use a tube with the following specifications.
tightening torque is 0.6 to 0.8 N•m. Make sure that the dust boot does Item Description
not obstruct the metal plating that mounting bracket F is attached to.
Recommended tubing material Urethane
• Make sure that the dust boot does not become
damaged during the procedure. Outer diameter 4 mm
• When loosening the screw with the hexagonal wrench, Tubing size
NOTICE insert the long end of the hexagonal wrench into the Inner diameter 2.5 mm
hex screw and rotate the small end with your hand.
• Do not forcibly tighten the hex screw. O How to Attach/Detach the Air Tube (GT2-PA12K(L)/PA12(L))
• Cannot be used with GT2-P12KF/P12F/H***/A***. • Attaching the air tube
Insert the air tube into the air supply hole on the relay connector.
O Mounting on the side of a surface
When mounting to the side of a GT2-P12K(L)/P12(L)/PA12K(L)/PA12(L)/S1/S5/
H12K(L)/H12(L)/A12K(L)/A12(L) unit, use the optional head mounting bracket B
(OP-76875) or the optional head mounting bracket E (OP-87220).
Mounting illustration Optional head mounting bracket B Air tube
(OP-76875)
Air supply hole

Tightening sleeve Nut Mounting bracket Relay connector

1 Insert the tightening sleeve into the mounting bracket from the side
with a depression, and loosely tighten with the nut.
Tightening sleeve • For best results, cut the end of the tube at a right angle,
and ensure that the outer perimeter is not damaged,
and that it still maintains a circular cross section.
Mounting bracket NOTICE • If the tube is not properly inserted, air leakage may result.
• After attaching, pull on the tube to make sure it is secure.
• Use a urethane tube. Make sure it also has a bending
radius of at least 50 mm.

Nut Notice the correct angle of the tightening sleeve. • Detaching the air tube
To detach the air tube, pull the air tube in the direction of the arrow,
as indicated in the figure below.
• If the sleeve is inserted from the wrong side, the Relay connector
Air tube
sensor head cannot be secured.
• Strongly tightening the nut without the sensor head
NOTICE inserted may deform the tightening nut.
• Head mounting bracket B cannot be used in
combination with head mounting bracket C (OP-
84396) or head mounting bracket D (OP-84327).

2 Insert the sensor head into the tightening sleeve, and tighten
the nut with a wrench. NOTICE Before detaching the tube, be sure to stop any air flow.

O How to Attach/Detach the Air Tube (GT2-A12K(L)/A12(L)/A32/A50)


Nut • Attaching the air tube
Wrench Feed the tubing into the socket until it bottoms out (about 1/2").
The socket will ensure a tight seal around the tubing.
Dust boot

Tightening the nut while fixing the sensor head may


lead to damage. Air tube
• Apply a tightening torque of 5 to 7 N•m (the
recommended tightening torque is 5 N•m) for all
NOTICE products other than the GT2-S5 and a tightening Coupling socket
torque of 4 to 5 N•m (the recommended tightening
torque is 4 N•m) for the GT2-S5.
Use the wrench that fits the width of the nut.
• Care must be taken not to damage the dust boot when tightening.

3 Secure the sensor head with M4 screws. • For best results, cut the end of the tube at a right
Mount the sensor head with the model label on top. For GT2-P12K (L)/ angle, ensure that the outer perimeter is not damaged,
P12 (L)/PA12K /PA12, mount so that the head indicator is on top. and that it still maintains a circular cross section.
M4 screws • If the tube is not properly inserted, air leakage may
result (see figure below).

NOTICE

• Never apply tightening torque over 1.4 N•m. Tube end


NOTICE
• Care must be taken not to damage the dust boot when tightening. • After attachment, pull on the tube to make sure it is secure.

GT2-70-M-E 6
• Detaching the air tube Q How to mount the lift lever
To detach the air tube, (1) press down on the release ring, (2) pull out the tube.
Mount the lift lever (OP-84397) between the spindle and the contact.
Secure the spindle with the lift lever and attach/detach the contact.

(2)

(1)
Release ring

• Before detaching the tube, be sure to stop any air flow. Amplifier display
NOTICE • Press down on the release ring evenly from both sides, and pull the tube out. Uneven
pressure may result in damage to the tube or damage to the operation of the air cylinder. Q Names of parts of the amplifier
O For GT2-A12K/A12/A32/A50 when an IP67 enclosure rating is required Digital LED display Special output indicator
Status indicator Arrow buttons
To maintain an IP67 enclosure rating, the following must be satisfied: GT2-71
• Firmly connect tubing with 2.5 mm inner diameter to the exhaust
port until it bottoms out. Detection level
Ensure that the tube end is free of dirt or water. indicator

Air tube [PRESET]


button
Preset indicator
Bank indicator Bar LEDs Timing input [SET] button
HIGH position indicator/ indicator
LOW position indicator PV indicator [MODE] button
Screw
Exhaust valve For details about the GT2-71D (pulse output type) amplifier,
Reference
refer to "GT2-71D (pulse output type)" (page 13).

Q Digital LED display


Air tube
The main display during detection can be switched to the displays
• Tighten the screw of the exhaust valve with a torque of 0.5 to 0.6 N•m. as shown below by pressing the left/right Arrow buttons.
If the screw is loose, the IP67 rating cannot be guaranteed. Refer to "Detection Functions" in the "GT2-70 Series User's
Manual" for details about the setting mode screen in which various
By loosening the screw of the exhaust valve, you can
Reference settings are performed.
change the exhaust port angle.

O Adjustment of Spindle Movement Speed 1. P.V. value display


To adjust the spindle movement speed, install a speed controller between
the cylinder and the air supply. Using OP-87970 (for the GT2-PA12K/
PA12/A12K(L)/A12(L)/A32/A50) is recommended.
8. Preset value setup display 2. Calculated value display*1
Air supply

Speed controller Indication symbol


7. LL setting value display 3. R.V. value display
while using 5-output function*2

4. HH setting value display


6. LOW setting value display while using 5-output function*2

Adjusting the speed.

• To further decrease spindle movement speed, use a coil-shaped 5. HIGH setting value display
tube (OP-87986) or similar to increase the distance between the air *1 This display appears only while the calculation function is used
NOTICE
supply hole and the speed controller. and an expansion unit is connected.
• The speed controller will not operate if installed in the reverse direction. *2 • This display appears only when [5out] is selected for [16.
Special output setting] in the basic setting mode.
Contact • This display does not appear on GT2-71MCN/71MCP.
1. P.V. value display ([P.V. = Present Value] Criterion value display)
This section describes how to replace the contact, and how to mount the lift lever. Displays a value to be used for output judgment.
• Detach the sensor head from the device or fixture (metal plating, 2. Calculated value display
etc.) before replacing the contact. Displays a calculated value such as a maximum or minimum value
• When applying pliers to the contact, be sure not to rotate the main of several detection points created when adding an expansion
part and cover the contact with a cloth. unit(s) (Displayed only when an expansion unit(s) is added).
• Never apply tightening torque over 0.2 N•m when attaching a contact. 3. R.V. value display ([R.V. = Raw Value] Raw value display)
NOTICE • Fix the roller contact (OP-77680) or the offset contact (OP-77683) Displays an actual detection value of the detection target.
with the fixing nut in the same direction as in actual use. 4. HH setting value display while using 5-output function
• Applying an adhesive, a thread locking agent, is recommended to Displays/Sets the setting value to set above the HIGH setting value.
prevent the nut from getting loose. 5. HIGH setting value display
• Position the roller contact carefully in the proper direction. Care Displays/Sets an upper limit value of the range of the detection target.
must be taken not to apply misdirected force to the shaft. 6. LOW setting value display
Displays/Sets a lower limit value of the range of the detection target.
Q How to replace the contact 7. LL setting value display while using 5-output function
Displays/Sets the setting value to set below the LOW setting value.
While securing the spindle with the accessory key wrench, detach the 8. Preset value setup display
contact with pliers. Then attach a new contact. Displays/Sets an arbitrary value to be added to or subtracted from
You cannot use the key wrench with the GT2-S1/S5, so use your the display value.
hands to secure the head.
Spindle
Zero-Point Correction
When you use this equipment for the first time or after the sensor head
Pliers is changed, be sure to correct the reference zero point.
Contact • When "------" is displayed, zero-point correction is disabled.
Reference
• Zero-point correction can be performed approx. 1 million
times. When zero-point calibration is frequently used, you
Cover the contact can perform setting so that the zero-point calibration will not
Key wrench with a cloth be written into the memory. Refer to "Origin Alignment" in the
"GT2-70 Series User's Manual" for details.

7 GT2-70-M-E
Q Setup using buttons 2 Detect the good workpiece and press the [SET] button to capture the
detected value.
1 Set any main display and perform the "SEt" and the detected value blink alternately, and the value of the
detection of the detection target (master good workpiece is fixed.
workpiece) to serve as a reference for zero-
point correction. Master workpiece 3 Detect the HIGH defective workpiece and press the [SET] button to
capture the detected value.
2 Press the [PRESET] button (white button) with the master workpiece The median value of this detected value and the one captured in step 2 blinks.
HIGH setting value is fixed.
being detected.
After "PrESEt" has blinked several times on the digital LED display 4 Press the right Arrow button until the LOW
of the amplifier, "0.000.0"* will appear. setting value display of the main display
The zero-correction is completed. appears.
LOW setting value display
5 Detect the good workpiece again and press the
[SET] button to capture the detected value.
6 Detect the LOW defective workpiece and press the [SET] button to
capture the detected value.
* When the preset function is set, the preset value will appear. The median value of this detected value HIGH
and the one captured in step 5 blinks. setting value
Q Setup using external input (Pink wire) LOW setting value is fixed. LOW
Two-point tuning is completed. setting value
1 Perform detection of the target (master workpiece) to serve as a
Press the left/right Arrow buttons to HIGH defective Good LOW defective
reference for zero-point correction. return to the P.V. value display. workpiece workpiece workpiece
2 Connect the pink wire to the appropriate terminal. Q Preset
Refer to page 15 of this manual for details about external input circuit diagrams. This function displays the value obtained by adding or subtracting a
preset value to or from the detected value.
Setup of Limit Values By using this function and the zero point correction function together,
The limit values are an upper limit value (HIGH setting value) and a you can set an arbitrary value as a reference point of the work.
lower limit value (LOW setting value). Setting these values enables Refer to "Preset Function" in the "GT2-70 Series User's Manual" for
three types of judgment (display/output): above the upper limit (HIGH), details about the preset function.
below the lower limit (LOW), and within the range (GO). Q Bank switching
Q Manual setup of limit values By using the bank function, the setting values of the HH side, HIGH
The following shows how to set manually an upper limit value (HIGH side, LOW side, LL side, preset point, and the preset value can be
setting value) and a lower limit value (LOW setting value). registered up to four patterns.
The bank switching is useful when there are multiple detection targets
1 When in the main display, press the left/ since the pre-registered settings (four patterns) can be easily switched.
right Arrow buttons until the HIGH Refer to "Bank Function" in the "GT2-70 Series User's Manual" for
setting value display appears. details about the bank function.
2 Enter an upper limit value (HIGH setting HIGH setting value display Q Limit output function
value) with the top/bottom Arrow buttons. This function can detect and output the mechanical movement of the spindle
3 Press the right Arrow button to display the on the contraction and extension sides. The stroke end of both sides can be
detected, so this function can be used for interlocking the device.
LOW setting value display.
The limit output function cannot be used on GT2-71MCN/
4 Enter a lower limit value (LOW setting Reference
71MCP (analog output type).
value) with the top/bottom Arrow buttons. LOW setting value display
The setup of range criteria values is Q 5-output
completed. This function can increase the output levels to 5 levels: HH, HIGH,
To return to the P.V. value display, press the left/right Arrow buttons. GO, LOW, and LL. HH/LL can be set in the same way as HIGH/LOW.
Refer to "Setting 5-output" in the "GT2-70 Series User's Manual" for
details about the 5-output function.
Tips for Convenient Functions The 5-output function cannot be used on GT2-71MCN/
Reference
Q Automatic setup of limit values 71MCP (analog output type).
This function automatically sets an upper limit value (HIGH setting Q Analog output function
value) and a lower limit value (LOW setting value).
For master workpieces, set by "tolerance tuning"; and for actual On GT2-71MCN/71MCP, the analog current (4 to 20 mA) is output
works (good or defective), set by "two-point tuning". based on the measurement value (R.V.). The upper limit value and
the lower limit value can be set to a desired value.
O Tolerance tuning Refer to "Analog Output Settings" in the "GT2-70 Series User's
This is a method for setting a range based on Manual" for details about the analog output function.
the detection value of a master workpiece Current output
when the master workpiece is available.
HIGH setting value
20mA
Tolerance tuning setting range
LOW setting value

When "------" is displayed, tolerance tuning is disabled.


Reference
When "-FFFF" or "FFFF" is displayed, the settings cannot
be set correctly.

1 When in the main display, press the left/ 4mA


right Arrow buttons until the P.V. value 0 Lower limit value Upper limit value
Measurement value
display appears, then perform detection of
The analog output function can only be used on GT2-
the master workpiece. Reference
71MCN/71MCP (analog output type).
Tolerance tuning is enabled only when in the P.V. value
Reference
display. Q Calculation with expansion unit
The GT2-70 Series can calculate various values using the detected
2 Press the [SET] button while the master workpiece is being detected values of multiple detection points such as the maximum value or the
to capture the detected value. minimum value when an expansion unit(s) is added (up to 14 units).
3 Press the upper/bottom Arrow buttons to enter a tolerance tuning Refer to pages 2 to 3 of this manual for details about how to add an
expansion unit.
setting range.
The outline of the calculation function is shown below.
4 Press the [SET] button to fix the tolerance tuning setting range. Number of units connected
After [SEt] blinks several times on the digital LED display of the Calculation
No. Description Calculation Calculation
amplifier, the P.V. value display automatically appears. function
mode only mode
Tolerance tuning is completed.
O Two-point tuning C1 Max.
value
Displays the maximum value of the values
of the main unit and expansion unit(s)
This is the method of setting the median values of the detected
good and/or defective workpieces as a range when the good C2 Min.
value
Displays the minimum value of the values
of the main unit and expansion unit(s)
workpiece and HIGH/LOW defective workpieces are available. Displays a difference between the maximum 2 to 14 units
Two-point tuning is disabled when "------" is displayed as a C3 Flatness value and the minimum value of the values
Reference
R.V. value. When "- FFFF" or "FFFF" is displayed, setting of the main unit and expansion unit(s) 1 to 14 units
cannot be done accurately. Displays the average obtained by dividing the
C4 Average sum of the values of the main unit and expansion
1 When in the main display, press the left/ unit(s) by the number of units connected
right Arrow buttons until the HIGH setting Displays a difference obtained by
value display appears. C5 Reference
difference subtracting the display value of the main Not selectable
HIGH setting value display unit from that of each expansion unit(s)

GT2-70-M-E 8
Number of units connected Setting item Initial value
Calculation
No. Description Calculation Calculation Static hold delay stability criterion dEFALt
function
mode only mode
Displays a degree of twist obtained from 3 units only 4 units only Static hold delay stability width 0.0100
C6 Twist the values of four detection points Calculation mode setting noCALc
C7 Warpage Displays a degree of warpage obtained
from the values of three detection points 2 units only 3 units only Calculation method C1.mAX
Displays a thickness obtained by *1 When using GT2-P12K (L)(F)/P12 (L)(F)/PA12K /PA12.
C8 Thickness sandwiching the detection target between 1 unit only 2 units only Default values of basic settings
the main unit and the expansion unit Setting item Initial value
O Calculation mode Measurement Direction change nor
When "Calculation mode" is selected in "A1. Calculation mode", the main unit
outputs the calculation result, while the expansion unit does not. Only the head Multiplier 1.0
indicator operates in the same way as the one on the main unit. (Except when Output mode no
"C5. Reference difference" is selected for the calculation method) Displayed number of digits 0.0001
Refer to "Calculation with expansion unit" in the "GT2-70 Series User's Manual"
for details about the calculation function. Hysteresis 0.0030
The calculation mode can be selected only on the main unit Batch input setting onE
Reference
with one or more expansion units connected.
Special output setting noUSE
O Calculation only mode
When calculation only mode is selected in [A1. Calculation mode], Limit output switch (HH) position 0.5000
only the main unit outputs the calculation result, and the expansion Limit output switch (LL) position 0.5000
unit outputs the judgment according to each setting.
Refer to "Calculation with expansion unit" in the "GT2-70 Series * GT2-71MCN/71MCP (analog output type) has the following
User's Manual" for details about calculation only mode. setting items and initial values for items 16 to 18.
• The calculation only mode is available only on the main Setting item Initial value
unit with two or more expansion units connected.
• When the calculation only mode is Analog range setting dEFALt
selected, "C5. Reference difference" Free range settings (Hi side) (Head stroke)*
cannot be selected for the
NOTICE calculation method. Free range settings (Lo side) 0.0000
• When the calculation only mode is * Ex.: When the sensor head is GT2-H12, the initial value is "12.0000".
selected, the sensor head cannot be Initial values of added function settings
connected to the main unit. Attach
the included "head connector label" Setting item Initial value
to the head connector of the main unit. Select preset data r.v.
Q Keylock Preset memory yES
The keylock allows you to avoid erroneous button operation during detection. Preset point setting onE
When the keylock is active, setup operations other than changing the display are disabled. Power-saving function (ECO) oFF
Reference The keylock can be set on the main display only.
Jam detection function oFF
O Set the keylock Jam detection position 0.5000
When keylock is active, setup operations other than main display
switching and the [PRESET] button operation are disabled. Initial values of calibration function settings
When in the main display, while pressing Setting item Initial value
the [MODE] button, press the upper Arrow Calibration function setting (Not initialized)
button for at least two seconds.
The keylock display will appear for several 2nd point target value 5.0000
seconds and then change to the main display.
MODE
+ (Press for at least two seconds) Keylock display
Q Switching the display unit
The display unit can be switched from the initialization screen.
O Set full keylock This section explains how to switch from "mm" display to "inch" display.
When full keylock is active, setup operations other than main display
switching are disabled.
1 Display "rSt.YES" from the initialization screen.
Press the down Arrow button for at least two Refer to "Initialization" for more details on how to display "rSt.YES".
seconds while pressing the [MODE] button in 2 Pressing the [MODE] button, press the [SET] button for 2 seconds or more.
the main display.
MODE
3 Use the up and down Arrow button to select "inch".
+ (Press for at least two seconds)
Full keylock display 4 Press the [MODE] button.
O Cancel the keylock or full keylock "rSt.End" will appear.
Press "up or down" Arrow button for at least Switching the display unit is completed.
two seconds while pressing the [MODE] button
in the main display when the keylock is active. Setting Mode Display
The keylock cancel display will appear for several
seconds and then change to the main display.
MODE
Keylock cancel display Q Function setting mode
+ or (Press for at least two seconds) You can set the following functions in the function setting mode.
Q Initialize (Reset to the default state) No. Item Description
• Initialization can be set on the main display only. Detection mode Select the detection mode.
Reference
• Initialization cannot cancel "span adjustment results". Hold update When "P-H", "b-H", or "P-P" is selected in , select "Timing
• Refer to "Calibration" in the "GT2-70 Series User's Manual" for details method input" or "Regular update" for the update timing of the hold value.
about how to set "Span adjustment" to the factory default state. Response Longer response time makes the average
• During initialization, performing other steps than the following ones
stops initialization, and the display returns to the main display.
time data time longer, which stabilizes the value.
For the timing input, select "External timing"
Timing type
1 When in the main display, while pressing the or "Self timing (internal timing)".
[MODE] button, press the [SET] button five times. Self timing When self timing is selected in , set the
"rSt.no" will appear. level threshold of the timing input.
2 Press the top/bottom Arrow buttons to select "rSt.YES".
Self timing
When "Std" is selected in , select the timing to fix the
criterion value after the value rises above (or falls below) the
3 Press the [MODE] button. delay type self timing level: "When the setting time is elapsed (delay
"rSt.End" will appear. timer)" or "After the display value is stabilized (static hold)".
Initialization is completed. User specified Set the delay time when "When the setting time is
O Default value list delay time elapsed (delay timer)" is selected in .
Default values of function settings Static hold delay Set the reference of stability criterion when "After the
Setting item Initial value stability criterion display value is stabilized (static hold)" is selected in .
Detection mode Std Static hold delay Enter a desired number when USEr (user
Hold update method tim stability width specification) is selected in .
Set the calculation mode when an expansion unit(s) is added.
Response time 100 (400)*1 Calculation
*The calculation mode will appear only when
Timing type t-in mode setting
an expansion unit(s) is added.
Self timing level 0.5000 Calculation Set the calculation method when "noCALc" is
method selected in .
Self timing delay type Stb.d
User specified delay time t=1000

9 GT2-70-M-E
O How to enter the function setting mode
 Static hold delay stability criterion
Enter the function setting mode by the following procedure.
• In the main display, press the [MODE] button for at least two seconds. Only available when "Stb.d" is selected in [ Self timing delay type].
GT2-71

MODE

[MODE] button
Default

Main screen User

 Detection mode
 Static hold delay stability width
MODE
Only available when "USEr" is selected in [ Static hold delay stability criterion].

MODE
Standard

NG hold Setting value


input screen
Peak hold
Bottom hold  Calculation mode setting
Only available when an expansion unit is connected.
Peak to peak


MODE

Hold update method


Only available when either "P-H", "b-H", or "P-P" is selected in [ Detection mode]. Calculation
not used

MODE Calculation mode

Calculation only mode*


Timing input

Regular update * Displayed only when two or more expansion units are connected.

 Calculation method
 Response time Only available when an expansion unit is connected.
An item other than "noCALc" is selected for [ Calculation mode].

MODE

When "nG-H" or "P-P" is selected


Setting value in 01. Detection mode. MODE

When "CAL.noH" is selected in


input screen A1. Calculation mode setting.

"400" is displayed on the amplifier used with GT2-P12K(L)(F)/P12 Max. value


(L)(F)/PA12K(L)/PA12(L). (default when shipped)
Min. value
 Timing type
Only available when either "Std", "P-H", or "b-H" is selected in [ Detection mode]. Flatness
Average
Reference
MODE difference*
Twist

External timing input Warpage

Rising edge Thickness


self-timing
Falling edge
self-timing
* This is not displayed when "CAL.noH" is selected for [ Calculation mode].
Q Basic setting mode
 Self timing level You can set the following functions in the basic setting mode.
Only available when item other than "t-in" is selected in [ Timing type]. No. Item Description
Measurement Set the "Normal" or "Reverse" to set the measurement
direction change value to increase or decrease based on position.
MODE

Multiplier Set the ratio of the detected value.


Setting value Output mode Set "ON (N.O.)" or "OFF (N.C.)" for output judgment.
input screen
Displayed
Set the number of display digits.
 Self timing delay type number of digits
Only available when the following conditions are provided. Hysteresis Set the hysteresis value.
• "Std" is selected in [ Detection mode]. Batch input Set to make input to all added expansion units when
• Item other than "t-in" is selected in [ Timing type]. setting an input is made to the input line of the main unit.
The settings for items numbered 16 and higher are different
MODE
depending on the amplifier unit being used.
When using a model other than GT2-71MCN/MCP (analog output type)
Static hold No. Item Description
Special output
Select the function of the special output lines.
Delay timer setting
Set the reference position where the limit
Limit output HH
- output turns on toward the spindle contraction
 User specified delay time side teaching
side when the "Lt.USEr" is selected in .
Only available when "t.d" is selected in [ Self timing delay type].
Limit output HH With the reference position being 0, set the
side criterion position where the limit output of the
MODE
position setting contraction side turns on.
Set the reference position where the limit
Limit output LL output turns on toward the spindle extension
Setting value -
side teaching
input screen side when the "Lt.USEr" is selected in .

GT2-70-M-E 10
No. Item Description <<When using a model other than GT2-71MCN/MCP (analog output type)>>
Limit output LL With the reference position being 0, set the  Special output setting
side criterion position where the limit output extension side
position setting turns on.
When using GT2-71MCN/MCP (analog output type)
No. Item Description
Analog range Set the analog output range to "dEFALt (head Not used
setting stroke)" or "FrEE (free range)".
Free range Set the Hi side for the free range setting 5-output
settings (Hi side) when "FrEE" is selected in 16. Limit output
Free range Set the Lo side for the free range setting Limit output user setting
settings (Lo side) when "FrEE" is selected in 16.
All GO
O How to enter the basic setting mode
Enter the basic setting mode by the following procedure. All limit output
• In the main display, press the [MODE] button and the [SET] button
at the same time for at least two seconds.
The items that can be selected (displayed) are different
Reference
GT2-71 depending on the settings in " Calculation mode". (When
[SET] button the expansion unit is not added, the items that can be
selected (displayed) are the same as for "noCALc".
[MODE] button Special output

Calculation mode setting *1 *1
O O O O O -
Main screen O O*2 O O O*3 -
O O - - O O
 Measurement direction change *1 This can be set only on the main unit that an expansion
Only available when a value other than "CAL.noH" is selected in [ Calculation mode]. unit is added.
*2 On the expansion unit, this can be set only when "C5.rEF
(Reference difference) is selected in "A2. Calculation Mode".
*3 This can be set only when "C5.rEF (Reference
difference)" is selected in "A2. Calculation method".
MODE

Normal Limit output HH side teaching


Only available when "Lt.USEr" is selected in [ Special output setting].
Reverse

 Multiplier
Only available when a value other than "CAL.noH" is selected in [ Calculation mode]. Reference
position set

MODE  Limit output HH side criterion position setting


Only available when "Lt.USEr" is selected in [ Special output setting].
Setting value
input screen
MODE

 Output mode
Setting value
input screen
MODE

Limit output LL side teaching


Only available when "Lt.USEr" is selected in [ Special output setting].
N.O.

N.C.

 Displayed number of digits Reference


position set

MODE  Limit output LL side criterion position setting


Only available when "Lt.USEr" is selected in [ Special output setting].
Setting value
input screen
MODE

 Hysteresis
Setting value
input screen
MODE
<<When using GT2-71MCN/MCP (analog output type)>>

Setting value
 Analog range setting
input screen

 Batch input setting


Available to set only on the main unit to which an expansion unit is added.
Default

MODE
Free range settings

Separate input
 Free range settings (Hi)
Concurrent input Only available when "FrEE" is selected in " Analog range setting".

MODE

Setting value
input screen

11 GT2-70-M-E
 Free range settings (Lo)  Jam detection function
Only available when "FrEE" is selected in " Analog range setting". Not available to set on the main unit that selects "CAL.noH" in [ Calculation mode], and
the main unit and expansion unit that selects values other than "t-in" in [ Timing type].

MODE

MODE

Setting value
input screen
OFF

Q Additional function setting mode


ON
You can set the following functions in the additional function setting mode.
No. Item Description User
Select preset data Select the data to use when performing the preset.
Store preset value Set whether or not to save "zero-point correction". Jam detection teaching
Only available when "USEr" is selected in [ Jam detection function].
Set the memory method for "zero-point
Preset point
correction" when using the bank function.
Power-save To decrease the power consumption or not to
function (ECO) display the numerical value, set this function.
Jam detection function When the spindle does not fully extend. Reference
position fixed
- Jam detection teaching Set the reference position for the jam detection.
Jam detection Set the position to perform the jam detection  Jam detection check point setting
check point setting function with the reference position being 0. Only available when "USEr" is selected in [ Jam detection function].
O How to enter the added function setting mode
Enter the added function setting mode by the following procedure.
• In the main display, press the [MODE] button and the left Arrow MODE

button at the same time for at least two seconds.


GT2-71 Left Arrow button Setting value
input screen

Q Calibration setting mode


[MODE] button
You can set the following functions in the calibration setting mode.
No. Item Description
Main screen Calibration
Set the calibration function.
function setting
 Select preset data -
1st point
Set the zero point (1st point).
detection/capture
Not available to set on the main unit for which "CAL.noH" is selected in [ Calculation mode].
2nd point target value Set the target value of the span adjustment (2nd point).
2nd point Set the detection position for the span
-
MODE detection/capture adjustment.
O How to enter the calibration setting mode
R.V. value Enter the calibration setting mode by the following procedure.
• In the main display, press the [MODE] button and the right Arrow
P.V. value button at the same time for at least two seconds.
GT2-71 Right Arrow button
 Store preset value
Not available to set on the main unit for which "CAL.noH" is selected in [ Calculation mode]. [MODE] button

MODE
Main screen

YES  Calibration function setting

NO
MODE

 Preset point setting


Default
Not available to set on the main unit for which "CAL.noH" is selected in [ Calculation mode].
Adjustment

MODE

 1st point detection/capture


Common for Only available when "AdJ" is selected in [ Calibration function setting].
all banks
Save for
each bank
Blink alternately
 Power-save function
Detect 1st point

Capture the
MODE
detection value

OFF *Target value for the first point is the preset value.

ECO Half  2nd point target value setting

ECO All
MODE

Setting value
input screen

GT2-70-M-E 12
 2nd point detection/capture GT2-71D (pulse output type)
Q Names of parts of the amplifier
Advanced setting switches
Blink alternately

Detect 2nd point

Capture the
detection value Input indicator
Pulse delivery indicator
Power/alarm indicator

Error message Item Description


Error display Cause Corrective action Power/alarm Lights up green when the power is turned on.
indicator Flashes red when there is an error*.
The sensor head cable is Connect the sensor head to the
disconnected. amplifier. Pulse delivery
indicator Flashes green when delivering a pulse.
• The sensor head cable is broken.
Change the cable or sensor Lights green when there is current position
head.
• The sensor head is damaged. Input indicator
request input.
An overcurrent is present • Check that the load is within the rating. Advanced setting
Sets the advanced information for pulse output.
in the output line. • Check that the output line is not switches
contacting other lines and frames.
* An error occurs in the following situations.
Recycle power to the equipment • For 700 ms after power turns on (power on reset time)
Failed to write/read data.
and perform initialization. *1 • Head is not connected or becomes disconnected
• If something is caught in the spindle, causing the • Overcurrent flows through the output
spindle not to move, change the sensor head.
Spindle movement is not detected
• Check the external timing input. (Press the Q Output signal
SET button to reset.) The signals output by GT2-71D are A, B, and Z phase signals,
The value fell below (rose Change the delay timer time which are voltage differential format line driver output that is
above) the timing level setting. (Press the SET button compliant with EIA-422.
during the delay time. to reset.)
Phase A
When setting the internal The value should not fall below (rise
timing, detection was above) the timing level before the Phase B
completed when the value fell criterion value is fixed. Alternatively,
Phase Z
below (rose above) the timing lengthen the static hold delay
level before the criterion value stabilizing amplitude so that the
(P.V. value) is fixed. criterion value can be easily fixed.
Contact between the After turning OFF the power, check the Z phase turns on when the sensor head
spindle is detected at home position. Home position
amplifiers is lost. connection between the amplifiers. (Set as a factory default;
• In the case of using the it cannot be changed.)
calculation function, the number
of the expansion unit when the • Reset the calculation mode or Q Advanced setting switch
power is turned ON differs from perform the initial reset of the Perform pulse output settings with the advanced setting switch.
the number of the expansion main unit. Setting item Switch No. Description
unit in the memory of the main • After turning OFF the power,
unit when the calculation is set. check the connections Set A phase or B phase (A phase or B phase)
• When using the calculation between the amplifiers. as the minimum phase difference.
function, a model other than the 1
Minimum 0.5µs 2.5µs* 5µs 25µs
GT2 Series is added.
ON ON ON ON
• When using the reference difference phase
(rEF) calculation function, an error difference
occurred in the main unit. Check each main/expansion 2 1 2 1 2 1 2 1 2
• When using other than the reference unit for the cause of the error.
difference (rEF) calculation function, an * This is the factory default setting.
error occurred in an expansion unit. Set the pulse resolution.
The sensor head is connected Remove the sensor head from the 3 0.1µm 0.5µm* 1µm 10µm
to the main unit in which the main unit or set the mode other
calculation only mode is set. than the calculation only mode. Pulse ON ON ON ON

The read/write setting switch of resolution


the connected communication
unit DL-RS1A is set to RW.*2 Set 4 3 4 3 4 3 4 3 4

the read/write setting switch to R. * This is the factory default setting.


Refer to "RS-232 Communication Sets the pulse fluctuation direction over the
"com.Loc" is displayed Unit DL-RS1A User's Manual" operation direction for the sensor head spindle.
and cannot change the Displayed when this equipment is
Pulse expands when the Pulse shrinks when the
setting. accessed (operation command/ Pulse spindle is pushed in* spindle is pushed in
settings and status writing/keylock) via
<DL-CL1>, <DL-DN1> or <DL-EP1>. fluctuation 5 ON ON
The buttons are locked, but after direction
approximately 10 seconds, "com.unLoc"
is displayed on the amplifier and the 5 5
buttons are unlocked.*2 * This is the factory default setting.
The calculation mode is Check the calculation mode settings
used, so the function again.The function setting mode can be (Not used) 6 ------------------------
setting mode cannot be changed only when the calculation mode of
changed the main unit is set to C0. oFF or C5. rEF. When using a sensor head other than GT2-
The criterion value (P.V. Perform the timing input and fix Reference
P12K(L)(F)/PA12K(L)/S1/S5/H12K(L)(F)(LF)/
value) becomes "------" the criterion value, and then A12K(L), even if the pulse resolution is set to 0.1 µm,
during the tolerance tuning. perform the tolerance tuning. a resolution of 0.5 µm is used.
The hold detection is selected or the Phase A /phase B Allowable
timing input is not performed, so the Perform the timing input. minimum phase Phase A cycle frequency of the
criterion value (P.V. value) is not fixed. difference counter
The operations of the expansion unit 0.5µs 2µs 500kHz
are locked while the simultaneous Deactivate the keylock on the
input is selected and the keylock is main unit. 2.5µs 10µs 100kHz
activated on the main unit. 5µs 20µs 50kHz
• If the number of expansion units 25µs 100µs 10kHz
has been changed, set the special
After ALL GO is selected in
the special output setting, output setting again.
the number of expansion • If the number of expansion units has
units changed. not been changed, check if each
expansion unit operates correctly,
then turn on the power again.
*1 If not recovered by this action, there may be a possibility of a failure
or permanent damage. In this case, change the amplifier unit.
*2 Bank switching, reset input, and preset input are available even
during "com.Loc".

13 GT2-70-M-E
O Minimum phase difference
Set the minimum phase difference of phase A and phase B (phase
A and phase B ).

Phase A

Phase B

Minimum phase difference: 2.5 μs

O Pulse fluctuation Direction


Direction where pulse increases when the spindle contracts
When advanced setting switch 5 is turned off (factory default
setting), the pulse increases when the spindle contracts, and the
pulse decreases when the spindle extends.

ON
Phase A (Black wire)
OFF
ON
Phase B (White wire)
OFF

ON
Phase A (Black wire)
OFF
ON
Phase B (White wire)
OFF

Direction where pulse increases when the spindle extends


When advanced setting switch 5 is turned on, the pulse increases
when the spindle extends, and the pulse decreases when the
spindle contracts.

ON
Phase A (Black wire)
OFF
ON
Phase B (White wire)
OFF

ON
Phase A (Black wire)
OFF
ON
Phase B (White wire)
OFF

Q Current position notification output function


When the automatic current position request input line (pink wire) is
turned off from on, a pulse is output from the home to the current
mechanical position of the spindle.
This function can be used when there is a difference between the
counter value and the measurement position, such as cases where
the power is turned off during measurement or other troubles occur.
"Power cable (I/O cable)" (page 4)

Current position ON
request input OFF

Phase A ON
(Black line) OFF
Phase B ON
(White line) OFF

Phase Z ON
(Orange line) OFF
(1) (2) (3)

(1) When the automatic current position request input (pink wire) is
turned off, the pulse of the spindle is output according to the
amount of movement.
(2) When the automatic current position request input (pink wire) is
turned on, all pulse outputs are turned off.
(3) When the automatic current position request input (pink wire) is
turned from on to off, the number of pulses from the home to the
current position is output.

GT2-70-M-E 14
Specifications
Q Sensor head
 Pencil type (short type)
Model GT2-S1 GT2-S5
Detection system Differential transformer
Measuring range*1 1 mm 5 mm
Operation range*1 2 mm 7 mm
Resolution 0.1 μm
Indication accuracy*2, *4 1 μm (P-P)
When installed facing down 1.12 N 1.0 N
Measuring
When installed sideways 1.1 N 0.95 N
force*3
When installed facing up 1.08 N 0.9 N
Sampling interval 1ms
Mechanical response*4 40 Hz 20 Hz
Operation indicator 2-color LED (red, green)
Enclosure rating*5 IP67G (JIS) and IP67 (IEC)
Surrounding air temperature -10 to +55°C (No freezing)
Environmental
resistance Relative humidity 10 to 85%RH (No condensation)
Vibration resistance 10 to 55 Hz, Compound amplitude 1.5 mm, 2 hours each in X, Y, and Z axis
Shock resistance 1000 m/s2 (IEC60068-2-27)
Main unit Main unit cover: SUS440C (GT2-S1), SUS430F (GT2-S5)
Cable between the sensor head and relay connector: PUR, Relay connector: PPSU
Material Dust boot NBR
Contact*6 SUS304, tungsten carbide
Head cable Sold separately (connected to relay connector)
Weight (excluding cable)*7 Approx. 30 g Approx. 40 g
*1 The measuring range indicates the range over which the measured value is displayed. The operation range indicates the range over which the spindle can actually move.
*2 For the GT2-S1, the indication accuracy is 1 μm for an arbitrary width of 0.1 mm within the area of measuring range center ± 0.15 mm and is 2 μm over this whole area.
For the GT2-S5, the indication accuracy is 1 μm for an arbitrary width of 0.2 mm within the area of measuring range center ± 0.3 mm and is 2 μm over this whole area.
Over the whole measuring range, the indication accuracy is linearity ± 0.15% of F.S. (F.S. is 1 mm for the GT2-S1 and 5 mm for the GT2-S5.)
*3 Typical value in the vicinity of the measured center value. Note that the measuring force varies depending on the installation status of the dust boot.
*4 When the surrounding temperature is 20°C.
*5 When an M8 oil-resistant cable (GT2-CHP2M/CHP5M/CHP10M) is used for the sensor head cable.
*6 The contact is a separately sold option.
*7 Including the relay connector.
Note: Cannot be used with amplifier units purchased before April 15th, 2014.

 Pencil type (GT2-PA12K(L)/PA12(L) are air cylinder types)


Model GT2-P12K GT2-P12KF GT2-P12KL GT2-P12 GT2-P12F GT2-P12L GT2-PA12K GT2-PA12 GT2-PA12K(L) GT2-PA12(L)
Detection system Scale shot system II Absolute type (without tracking error)
Measuring range 12 mm
Resolution 0.1 μm 0.5 μm 0.1 μm 0.5 μm 0.1 μm 0.5 μm
Indication accuracy *1 1 μm (P-P) 2 μm (P-P) 1 μm (P-P) 2 μm (P-P) 1 μm (P-P) 2 μm (P-P)
When installed facing down 1.0 N 0.2 N 1.0 N 0.2 N 1.2 N 0.4 N
Measuring
When installed sideways 0.95 N 0.15 N 0.95 N 0.15 N 1.15 N 0.35 N
force *2
When installed facing up 0.9 N 0.1 N 0.9 N 0.1 N 1.1 N 0.3 N
Sampling interval 4 ms
Mechanical response *1 10 Hz 4 Hz 10 Hz 4 Hz -
Pressure range - 0.24 to 0.26 MPa 0.05 to 0.07 MPa
Max. pressure - 0.5 MPa
Fluid to use - Clean dry air
Operation indicator 2-color LED (red, green)
IP67G (JIS)*3 IP67G (JIS)*3
Enclosure rating IP67 (IEC) - IP67 (IEC) - IP67 (IEC) -
NEMA Type 13*3 NEMA Type 13*3
Environmental Surrounding air temperature -10 to +55°C (No freezing) 0 to +55°C (No freezing)
resistance
Relative humidity 35 to 85% RH (No condensation)
Vibration resistance 10 to 55 Hz, Compound amplitude 1.5 mm, 2 hours each in X, Y, and Z axis
Shock resistance 1000 m/s2 (IEC60068-2-27)
Main unit cover: SUS303, Spindle for GT2-PA12K/PA12: SUS430 (Fluorine-based coating),
Main unit Indicator: PET, Sensor head - Connection cable: PUR, Connector: PBT,
Dust seal (GT2-PA12K/PA12 only): SUS303/SU304/aluminium (alumite treated)/special polyester fiber
Material Dust boot NBR - NBR -
SUS304, SUS304,
Contact point *4 SUS304, SUS304,
SUS304, tungsten carbide SUS304, SUS440C tungsten tungsten
SUS440C SUS440C
carbide carbide
Head cable Sold separately (connected to relay connector)
Weight (excluding cable) *5 Approx. 35 g Approx. 45 g Approx. 35 g Approx. 45 g Approx. 35 g
*1 When the surrounding temperature is 20°C. Entire measuring range.
*2 This is a typical value at the center of the measuring range. The measuring force varies depending on the mounted condition of the dust boot.
This is a representative value when using the GT2-PA12K/PA12 with a pressure of 0.25 MPa and when using the GT2-PA12KL/PA12L with a pressure of 0.06 MPa. The measuring force varies
depending on the air pressure used. See table 1 for more details.
There is no dust boot for GT2-PA12K(L)/PA12(L).
*3 When using an M8 oil resistant cable (GT2-CHP2M/5M/10M) as the sensor head cable.
*4 The contact is a separately sold option.
*5 Including the relay connector.
Note: Cannot be used with amplifier units purchased before April 15th, 2014.

Table 1. Relationship between spindle position and measuring force grouped according to used air pressure.

GT2-PA12K/PA12 GT2-PA12KL/PA12L
Measuring force (N) Measuring force (N)
1.7 0.55

0.50
Applied pressure (MPa) Applied pressure (MPa)
0.45
0.26 MPa 0.07 MPa
1.2 0.40
0.25 MPa 0.06 MPa

0.24 MPa 0.35 0.05 MPa

0.30

0.7 0.25
0 1 2 3 4 5 6 7 8 9 10 11 12 Spindle position (mm) 0 1 2 3 4 5 6 7 8 9 10 11 12 Spindle position (mm)
Spindle extended Spindle retracted Spindle extended Spindle retracted
(When installing side mounting, the measuring force is -0.05 N. When installing upward mounting, the measuring force is -0.1 N.)

15 GT2-70-M-E
O Box type
Model GT2-H12K GT2-H12KL GT2-H12KF
GT2-H12KLF GT2-H12 GT2-H12L GT2-H12F GT2-H12LF GT2-H32 GT2-H32L GT2-H50
Detection system Scale shot system II Absolute type (without tracking error)
Measuring range 12 mm 32 mm 50 mm
Resolution 0.1 µm 0.5 µm
Indication accuracy *1 1 µm (P-P) 2 µm (P-P) 3 µm (P-P) 3.5 µm (P-P)
When installed facing down 1.0 N 0.4 N 1.0 N 0.4 N 1.0 N 0.4 N 1.0 N 0.4 N 2.1 N 1.2 N 3.2 N
Measuring force *2 When installed sideways 0.9 N 0.3 N 0.9 N 0.3 N 0.9 N 0.3 N 0.9 N 0.3 N 1.8 N 0.9 N 2.8 N
When installed facing up 0.8 N 0.2 N 0.8 N 0.2 N 0.8 N 0.2 N 0.8 N 0.2 N 1.5 N 0.6 N 2.4 N
Sampling interval 1 ms
Mechanical response *1 10 Hz 4 Hz 10 Hz 4 Hz 10 Hz 4 Hz 10 Hz 4 Hz 6 Hz 5 Hz 7 Hz
Operation indicator 2-color LED (red, green)
Enclosure rating IP67 - IP67 - IP67 - IP67 - IP67 - IP67
Surrounding air temperature -10 to +55°C (No freezing)
Environmental Relative humidity 35 to 85% RH (No condensation)
resistance
Vibration resistance 10 to 55 Hz, Compound amplitude 1.5 mm, 2 hours each in X, Y, and Z axis
Shock resistance 1000 m/s2 (IEC60068-2-27)
Main unit Main unit cover: zinc die-cast, Indicator: polyarylate (PAR)
Material Dust boot NBR - NBR - NBR - NBR - NBR - NBR
Contact point *3 SUS304, tungsten carbide SUS304, SUS440C
Cable Sold separately (connected to M8 connector)
Weight (excluding cable) Approx. 95 g Approx. 95 g Approx. 100 g Approx. 100 g Approx. 95 g Approx. 95 g Approx. 100 g Approx. 100 g Approx. 270 g Approx. 320 g
*1 When the surrounding temperature is 20°C. Entire measuring range. *3 The contact is a separately sold option.
*2 This is a typical value at the center of the measuring range. The measuring force varies depending
on the mounted condition of the dust boot.

O Air cylinder type


Model GT2-A12K GT2-A12KL GT2-A12 GT2-A12L GT2-A32 GT2-A50
Detection system Scale shot system II Absolute type (without tracking error)
Measuring range 12 mm 32 mm 50 mm
Resolution 0.1 µm 0.5 µm
Indication accuracy *1 1 µm (P-P) 2 µm (P-P) 3 µm (P-P) 3.5 µm (P-P)
When installed facing down 1.2 N 0.4 N 1.2 N 0.4 N 2.1 N 3.2 N
*2 When installed sideways 1.1 N 0.3 N 1.1 N 0.3 N 1.8 N 2.8 N
Measuring force
When installed facing up 1.0 N 0.2 N 1.0 N 0.2 N 1.5 N 2.4 N
Sampling interval 1 ms
Operation indicator 2-color LED (red, green)
Pressure range 0.25 to 0.5 MPa
Max. pressure 1.0 MPa
Fluid to use Dry air
Enclosure rating IP67*3 - IP67*3 - IP67*3
Surrounding air temperature 0 to +55°C (No freezing)
Environmental
resistance Relative humidity 35 to 85% RH (No condensation)
Vibration *4 10 to 55 Hz, Compound amplitude 1.5 mm, 2 hours each in X, Y, and Z axis
Shock resistance 2
1000 m/s (IEC60068-2-27) -
Main unit Main unit cover: zinc die-cast, Cylinder: aluminum alloy, Air coupling socket (resin part): polyacetal, Air coupling socket (metal part): nickel plated brass, Indicator: polyarylate (PAR)
Material Dust boot FKM - NBR - NBR
Contact point *5 SUS304, tungsten carbide SUS304, SUS440C SUS304, SUS440C
Sensor head cable Sold separately (connected to M8 connector)
Weight (excluding sensor head cable) Approx. 145 g Approx. 340 g Approx. 405 g
*1 When the surrounding temperature is 20°C. Entire measuring range. *4 When using mounting bracket D (OP-84327) with GT2-A32/A50, vibration is 0.75 mm.
*2 This is a typical value at the center of the measuring range. The measuring force varies depending *5 The contact is a separately sold option.
on the mounted condition of the dust boot.
*3 Connect an air tube to the exhaust valve so that no foreign materials enter from the valve.

Q Amplifier (GT2-71(C)N/71(C)P/72(C)N/72(C)P/75N/75P/76N/76P/71MCN/71MCP)
Mount type *1 DIN rail mount Panel mount DIN rail mount
Type
Main unit/Expansion unit *2 Main unit Expansion unit Main unit Expansion unit Main unit
NPN output GT2-71(C)N GT2-72(C)N GT2-75N GT2-76N GT2-71MCN
Model
PNP output GT2-71(C)P GT2-72(C)P GT2-75P GT2-76P GT2-71MCP
Number of units that can be added *2 Up to 14 expansion units for 1 main unit
Supply voltage *2 10 to 30 VDC, including 10% ripple (P-P) Class 2 20 to 30 VDC, including 10% ripple (P-P) Class 2
Display range -199.9999 to 199.9999
Display resolution 0.1 µm
Normal 2200 mW max. (73.3 mA max. at 30 V) 2700 mW max. (90 mA max. at 30 V)
Current consumption Power saving (eco half) 1800 mW max. (60.0 mA max. at 30 V) 2300 mW max. (76.7 mA max. at 30 V)
Power saving (eco) 1700 mW max. (56.7 mA max. at 30 V) 2200 mW max. (73.3 mA max. at 30 V)
hsp (3 ms), 5 ms, 10 ms, 100 ms, 500 ms, 1000 ms
Response time
(hsp (12ms), 20ms, 40ms, 400ms, 2000ms, 4000ms when using GT2-P***/PA*** with a sensor head)
Control output NPN output NPN open collector 40 V 50 mA max. Residual voltage: 1 V max.*2
(HH/HIGH/GO/LOW/LL) *3 PNP output PNP open collector 30 V 50 mA max. Residual voltage: 1 V max.*2
Timing input Input time: 2 ms min.
Preset input
Control input
Reset input Input time: 20 ms min.
Bank input
Output range ––––––––––––––– 4 to 20 mA, Maximum load resistance 350 
Analog output (Can be set anywhere within the analog output range)
Response time ––––––––––––––– Specified response time + 1 ms
Surrounding air temperature*2 -10 to +50°C (No freezing) *2
Environmental resistance Relative humidity 35 to 85%RH (No condensation)
Vibration 10 to 55 Hz, Horizontal amplitude: 1.5 mm, 2 hours each in X, Y, and Z axis
Material Main unit cover/Front cover: Polycarbonate (PC), Keytop: Polyacetal (POM), Front sheet: Polyethylene terephthalate (PET), Cable: Polyvinyl chloride (PVC)
GT2-71N/71P/72N/72P Approx. 140 g (Including a power cable)
Weight GT2-75N/75P/76N/76P Approx. 140 g (Including a panel mounting bracket, front protective cover, power cable)
GT2-71(M)CN/71(M)CP/72CN/72CP Approx. 70 g
*1 When using the DIN rail mount type unit, be sure to mount it to a DIN rail (fixed to a metal ◆ When connecting 9 to 15 units including the main unit
plate). Use the end unit (OP-26751) when connecting expansion units. • Supply voltage: 20 to 30 VDC
*2 When adding an expansion unit(s), the following restrictions apply depending on the • Control output current: 10 mA max. (including the output current of the DL-RB1A)
number of connected units. • Residual voltage: 1.5 V max.
◆ When connecting 2 to 8 units including the main unit • (GT2-71MCN/71MCP only) Surrounding air temperature: -10 to +45°C
• Supply voltage: 20 to 30 VDC *3 HH and LL are not available for GT2-71MCN/71MCP.
• Control output current: 20 mA max.
• (GT2-71MCN/71MCP only) Surrounding air temperature: -10 to +45°C

Q Amplifier (GT2-71D)
Mount type DIN rail mount
Type
Main unit/Expansion unit Main unit
Model GT2-71D
Number of units that can be added Expansion not possible
Supply voltage 10 to 30 VDC, including 10% ripple (P-P) Class 2
Indicators Power indicator (green), alarm indicator (red), pulse output indicator (green), input indicator (green)
Pulse resolution Select from 0.1 µm, 0.5 µm, 1 µm, 10 µm
Power consumption 1600 mW or less (53.3 mA or less with 30 V)
Minimum phase difference Select from 0.5 µs, 2.5 µs, 5 µs, 25 µs
Control input (homing) No-voltage input (with contact, without contact), input time 20 ms or more
Output signal Phase A/B/Z 90° phase difference, differential square-wave (EIA-422 compliant), x 4
Output signal level +5 V
Surrounding air temperature -10 to +50°C (No freezing)
Environmental
Relative humidity 35 to 85%RH (No condensation)
resistance
Vibration 10 to 55 Hz, Horizontal amplitude: 1.5 mm, 2 hours each in X, Y, and Z axis
Material Main unit cover Polycarbonate (PC)
Weight Approx. 110 g (including a power cable)

GT2-70-M-E 16
Circuit diagram Q External input circuit diagram
GT2-71D
Q I/O circuit diagram DC5V

GT2-71(M)(C)N/72(C)N/75N/76N
Brown *1
DC10-30V*3 Pink
Black (HI output)
protection circuit Load DC5-40V

circuit
Overcurrent

Main
White (LO output)
Main circuit

Load DC5-40V
Blue
Gray (GO output)
Load DC5-40V
Orange (HH output)*2
Load DC5-40V (Short circuit current: 1 mA max.)
Green (LL output)*2
Load DC5-40V
Q Recommended input device
1
Blue *
0V
Pink, Violet, Pink/Violet, Yellow, Red *4
Input circuit External input Line receiver AM26LS32 or equivalent
*1 Brown and blue are available for main units (GT2-71(M)(C)N/75N) only.
Not available for expansion units (GT2-72N/76N).
The expansion unit of the connector type (GT2-72CN) is not connected.
*2 In the analog type (GT2-71MCN), the orange and green lines are analog output lines.
Refer to the analog output circuit diagram for details. 390Ω
*3 When an expansion unit is connected or for the analog type (GT2-71MCN), the voltage is DC20-30V.
*4 Refer to external input circuit diagram for details about external input.

GT2-71(M)(C)P/72(C)P/75P/76P
Brown *1
DC10-30V*3
Black (HI output) WARRANTIES AND DISCLAIMERS
protection circuit

Load
Overcurrent

White (LO output)


Main circuit

Gray (GO output) Load (1) KEYENCE warrants the Products to be free of defects in materials and workmanship for a period of one (1)
Load year from the date of shipment. If any models or samples were shown to Buyer, such models or samples were
Orange (HH output)*2
used merely to illustrate the general type and quality of the Products and not to represent that the Products
Load
Green (LL output)*2 would necessarily conform to said models or samples. Any Products found to be defective must be shipped to
Blue *1 Load KEYENCE with all shipping costs paid by Buyer or offered to KEYENCE for inspection and examination. Upon
0V
Pink, Violet, Pink/Violet, Yellow, Red *4 examination by KEYENCE, KEYENCE, at its sole option, will refund the purchase price of, or repair or replace
Input circuit External input at no charge any Products found to be defective. This warranty does not apply to any defects resulting from
*1 Brown and blue are available for main units (GT2-71(M)(C)P/75P) only. any action of Buyer, including but not limited to improper installation, improper interfacing, improper repair,
Not available for expansion units (GT2-72P/76P).
The expansion unit of the connector type (GT2-72CP) is not connected. unauthorized modification, misapplication and mishandling, such as exposure to excessive current, heat,
*2 In the analog type (GT2-71MCP), the orange and green lines are analog output lines. coldness, moisture, vibration or outdoors air. Components which wear are not warranted.
Refer to the analog output circuit diagram for details. (2) KEYENCE is pleased to offer suggestions on the use of its various Products. They are only suggestions,
*3 When an expansion unit is connected or for the analog type (GT2-71MCP), the voltage is DC20-30V.
*4 Refer to external input circuit diagram for details about external input. and it is Buyer's responsibility to ascertain the fitness of the Products for Buyer’s intended use.
KEYENCE will not be responsible for any damages that may result from the use of the Products.
Q Analog output circuit diagram (3) The Products and any samples ("Products/Samples") supplied to Buyer are not to be used
internally in humans, for human transportation, as safety devices or fail-safe systems, unless
GT2-71MCN/71MCP their written specifications state otherwise. Should any Products/Samples be used in such a
manner or misused in any way, KEYENCE assumes no responsibility, and additionally Buyer
Orange will indemnify KEYENCE and hold KEYENCE harmless from any liability or damage
Main circuit

Analog output (4 to 20 mA) whatsoever arising out of any misuse of the Products/Samples.
(4) OTHER THAN AS STATED HEREIN, THE PRODUCTS/SAMPLES ARE PROVIDED WITH NO
OTHER WARRANTIES WHATSOEVER. ALL EXPRESS, IMPLIED, AND STATUTORY
WARRANTIES, INCLUDING, WITHOUT LIMITATION, THE WARRANTIES OF MERCHANTABILITY,
Green*
0V FITNESS FOR A PARTICULAR PURPOSE, AND NON-INFRINGEMENT OF PROPRIETARY
RIGHTS, ARE EXPRESSLY DISCLAIMED.
* The green line is internally common with the blue line. IN NO EVENT SHALL KEYENCE AND ITS AFFILIATED ENTITIES BE LIABLE TO ANY PERSON
OR ENTITY FOR ANY DIRECT, INDIRECT, INCIDENTAL, PUNITIVE, SPECIAL OR
Q External input circuit diagram CONSEQUENTIAL DAMAGES (INCLUDING, WITHOUT LIMITATION, ANY DAMAGES RESULTING
GT2-71(M)(C)N/72(C)N/75N/76N FROM LOSS OF USE, BUSINESS INTERRUPTION, LOSS OF INFORMATION, LOSS OR
INACCURACY OF DATA, LOSS OF PROFITS, LOSS OF SAVINGS, THE COST OF
Pink: Preset input
DC5V
Violet: Bank A input PROCUREMENT OF SUBSTITUTED GOODS, SERVICES OR TECHNOLOGIES, OR FOR ANY
Pink/Violet: Timing input MATTER ARISING OUT OF OR IN CONNECTION WITH THE USE OR INABILITY TO USE THE
Yellow: Bank B input PRODUCTS, EVEN IF KEYENCE OR ONE OF ITS AFFILIATED ENTITIES WAS ADVISED OF A
Red: Reset input
POSSIBLE THIRD PARTY’S CLAIM FOR DAMAGES OR ANY OTHER CLAIM AGAINST BUYER. In
some jurisdictions, some of the foregoing warranty disclaimers or damage limitations may not apply.
circuit
Main

Blue*1
BUYER'S TRANSFER OBLIGATIONS:
(Short circuit current: 1 mA max.) If the Products/Samples purchased by Buyer are to be resold or delivered
*1 Blue is available for main units (GT2-71(M)(C)N/75N) only. to a third party, Buyer must provide such third party with a copy of this
Not available for expansion units (GT2-72(C)N/76N). document, all specifications, manuals, catalogs, leaflets and written
GT2-71(M)(C)P/72(C)P/75P/76P information provided to Buyer pertaining to the Products/Samples.
DC10-30V E 1101-3
Brown*1
Main circuit

Pink: Preset input


Violet: Bank A input
Pink/Violet: Timing input
Yellow: Bank B input
(Short circuit current: 2 mA max.) Red: Reset input
*1 Brown is available for main units (GT2-71(M)(C)P/75P) only.
Not available for expansion units (GT2-72(C)P/76P).

Pulse Output Type Circuit Diagram


Q I/O circuit diagram
GT2-71D
Brown
DC5V Black: A phase
Main circuit

White: B phase
33W Black, White, Orange
Orange: Z phase
33W _
Violet, Gray, Green Violet:_A phase
Gray: B_phase
Blue Green: Z phase
Pink* Pink: Input
Input circuit

* Refer to external input circuit diagram for details about external input.

17 GT2-70-M-E
GT2-70-M-E 18
19 GT2-70-M-E
Copyright (c) 2014 KEYENCE CORPORATION. All rights reserved.