You are on page 1of 4

EQUIPMENT FAILURE REPORT LOG APRIL -2024

Time Resulted From


Date & time of Rquipment failure Duration to Managed In Managed as Lack of Resulted From Suggested Preventive
SL No
failure reported Job Number Location Department Equipment Name & Asset code category Reason for failure Details Of Failure Impact of failure Fix the House Outsourced Operator Safety Hazard Corrective Actions Taken Status Actions
Failure Knowledge
Time
Date & time of Duration to Suggested Preventive
SL No
failure reported Job Number Location Department Equipment Name & Asset code Details Of Failure Corrective Actions Taken Status
immediate urgent routine Mech Elec Plumb others None life health service business others Fix the yes No yes No yes No Yes No Actions
Failure Checked the dental power
1 2/04/2024,09:00AM AZ20240401 AZZIZIYA Dental OPD-302 Dental Chair(A-BM-DNT-BM0054-1) ✓ ✓ Not working ✓ 2hr ✓ ✓ ✓ ✓ supply ,Air pressure.Repaired the done Nill
Air flowing valve connetion

2 6/03/2024,09:00AM AZ20240402 AZZIZIYA Dental OPD-302 Dental Chair(A-BM-DNT-BM0054-1) ✓ ✓ Water Leaking ✓ 30mins ✓ ✓ ✓ ✓ changed the 12V solenoid valve done Nill

3 15/03/2024,10:00AM AZ20240403 AZZIZIYA Dental OPD-310 Dental Chair (C-BM-DNT-BM0054-8) ✓ ✓ low air pressure ✓ 30mins ✓ ✓ ✓ ✓ adjusted the connections now done Nill
working fine

4 17/03/2024,09:00AM AZ20240404 AZZIZIYA Dental OPD-302 Dental Chair(A-BM-DNT-BM0054-1) Low Air pressure 1hr cleaned the compressor now done Nill
✓ ✓ ✓ ✓ ✓ ✓ ✓ working fine
Compliant Analisis based on Departments Total
4.5
4
3.5
Department Number Of Complants
3
(empty)
2.5 Numbe
Dental OPD-302 3
Dental OPD-310 1 2
Total Result 4 1.5
1
0.5
0
(empty) Dental Dental Total
OPD-302 OPD-310 Result

Compliant Analisis based on Equipments Total


4
3
Equipment Name &Number of Compliants 2
(empty)
1
Dental Chair(A-B 3
0 Numbe
Dental Chair (C-B 1 y) -1
)
-8
)
ul
t
pt 54 54 es
Total Result 4 m 0 0 R
(e 0 0 al
- BM - BM T ot
NT NT
-D -D
M M
-B -B
i r(A r (C
ha ai
lC l Ch
a a
nt nt
De De

Total
Total

Routine
Urgent
Data Immediate
Routine 4
Urgent
Immediate
4

ompliant Analisis based on Failuare Reason Total

1
Mechanical
Data Electric
Mechanical Plumbing
Electric Others
Plumbing 3
Others 1
3

Failuare Management Total

In-House
Data Out-Sourse
In-House 4
Out-Sourse
Number Of Complants

Number of Compliants

You might also like