TECHNEWS

TechNews: TN#603 Distribution: Internal External

Product: CS 8200 3D Neo Validity: NA

Date: July 26th , 2023 Writer: Service engineering

Document Reference: TN#603-20230726-EN Contact : Carestream Dental - 4, rue F. Pelloutier -


Croissy-Beaubourg - 77435 Marne-la-Vallée Cedex 2 (France)

Revision 26/09/2023: This TN is obsolete. Please update to CS 8200 3D V2 1.0.374.0 see TN595.
Column Calibration issue with CS 8200 Neo Driver 1.0.372.0

Problem description:

The error “MOT_65280 Up / Down column movement position out of range” occurs at the end of the
column calibration ID2307 and the calibration is not effective. The calibration parameter saved are the
factory default value.

Solutions:

If the installation required a specific column calibration (limit of ceiling):


- You must downgrade to the version of Driver 1.0.351 and run the column calibration again.

If the unit driver has been upgraded after the installation:


- You have nothing to do.

If the unit require a column calibration for service reason:


- You must downgrade to the version of Driver 1.0.351 and run the column calibration again.

A new release of driver with correction will be available soon in Infotec.