Professional Documents
Culture Documents
Kystron Power
Supply
VSWR meter/
Klystron tube Variable Frequency Detector Micro ammeter/
Isolator meter mount
with mount Attenuator CRO
Figure 17.1 (a) Block diagram of microwave bench setup for Reflex Kh'stron characterisrics
Source 2imimg com
Figure 17.1 (b) Schematicof micmwve bench setup wih Rejlex Klyston as
micrpe
Initial settings of the experiment:
1 Klvstron power
supply:
Beam voltage OFF. Beam voltage knob fully anti clockwise
Repcller voltage fully clockwise
Mod-switch - AM
Amplitude knob - fully
clockwise
2
Frequency knob -mid position
VSWR meter:
Range dB - 40dB / 50 dB, Input switch - low impedance
Meter switch - normal, Gain (coarse and fine) - mid position
approximately
Note: In the following experimental procedure, steps I to 5
are
Reflex Klystron tube (the procedure differs in the Gunn diode common in all experiments that use the
can useeither VSWR or Micro Ammeter for finding the characteristics experiment). In addition
frequency of Reflex Klystron/Gunn diode oscillat
Experimentalprocedure:
1. Connect the equipment as shown in Figure 17.1 (a).
2. The variable attenuator and the reflector voltage should be kept at their maximum positions anu
Klystron power supply to its minimum.
3. Rotate the knob of the frequency meter to one
side fully.
4. Switch on the Klystron power supply and
5. Rotate the beam voltage knob slowly
cooling fan.
clockwise up to 300 V.
6. Vary the repeller voltage slowly and watch the micro
mum deflection in the meter.
ammeter. Set the repeller voltage for muN
7. Tune the plunger of the Klystron mount to get the
maximum output if necessary.
8. Slowly turn the knob of the frequency meter until a dip is observed in the micro ammeter VS*
meter. The frequency reading between two horizontal lines and the vertical marker in a dirett
reading-type wave meter gives the corresponding frequency.
9. Change the reflector voltage and read the current and frequency for each reflector
voltuge
Klystron mode observations on CRO:
1. Set the mode selector switch of Klystron power supply unit to FM-MOD position. Keep the
amplitude, and FM frequency knobs at mid position. Let the remaining knobs are the sur
pos1tion.
2. Switch on the power supplies to both Klystronpower supply unit and oscilloseopc.
setthe
3. Set the Klystron power supply unit meler selector switch to bcam voltage position and
anode beam voltage to approximately 300 Vusing the beam voltage control knob.
Result: The attenuation of the test attenuator is deternined.
Crystal SWR/
Klystron micro ammeter/
detector
power DC Voltmeter
Figure 17.6 (a) Block diagram for measurement offrequency and wavelengh
Initialsettings of the experiment: (same as in Section 17.2)
Experimental procedure (Cavity wave-meter method):
1. The frequency meter cavity should be of resonance at an unknown frequency.
2. The attenuator should be adjusted to provide about 6-10 dB loss.
3. The IKHzsquare-modulated wave should be applied.
4. Amp/Freq for peak reading in the VSWR meter (0° mark on RHS) should be varied.
5. The size of the cavity should be adjusted until resonance occurs. When a small amount of micro
wave energy enters the cavity, a considerable dip in VSWR meter is shown.
6. The unknown frequency can be read on a frequency meter.
Microwave Experiments 17.7
Vmin |Vmin
d cm dh cm
1 1 1 1
C =3x108
(4.6)² = 10.2GHz
-+
Result: The frequency and wavelength of a rectangular waveguide in the TE0 mode are determined.
Crystal Micro
Klystron power Termination
detector ammeter
supply
3, 3,
0
Klystron tube Variable Frequency Directional Matched
with mount Isolator attenuator meter coupler 2 Load
Port 1
Port 2
420
Isolation in dB =-20log =20log=52.46 dB and
420
Insertion Loss in dB= 20log
-201 400
=0.42 dB
VSWR MEASUREMENT
1.9 0
ratio for VSWR
.ia) To determine the voltage standing-wave VSWR 10
b)To determine the voltage standing-wave ratio for
mount, isolator, frequency meter,
inment required: Klystron power supply, Klystron tube with movahle short/termination, S-S
eattenuator, slotted-linc section, crystal detecto, VSWR meter,
BNC'cable, waveguide sands, and accessories.
Crystal VSWR
Klystron power detector meter
supply
(a)
Crystal VSWR
Klystron detector meter
power
Vnax
V.min
Tme
Observations for high VSWR: Beam Voltage = 300 V, Beam Current = 20 mA,
d, = 10.8 cm and d, =8.4 cm, , = 2(d, - d) = 2(10.8 - 7.4) = 6.8 cm The
6.8
d, = 8.5 cm, and d, = 7.9 cm (Using S-S tuner) VSWR= =3.6
n(d -d,) n(8.5-7.9)
Result: Low VSWR and high VSWR of a microwave bench setup are measured.