Professional Documents
Culture Documents
Unit 5 Introduction To Spectrometry 2013-14
Unit 5 Introduction To Spectrometry 2013-14
It results from fluctuations of electric and magnetic fields that are perpendicular to each
other – waves.
Electromagnetic Spectrum
Comprises of a wide range of radiation energies (therefore, frequencies and
wavelengths).
The range is so large that the scale is logarithmic and the spectrum is continuous over all
wavelengths.
Interactions between radiation and matter take place throughout the entire
electromagnetic spectrum, but the effect of the radiation varies depending on the energy –
see Figure 24.3 in SWHC, 8th Ed.
Absorption Spectrometry
Absorption is the process by which a medium attenuates the intensity of certain
frequencies of electromagnetic radiation.
Every particle has a unique set of energy states; the lowest being the ground state.
The particle (atom/ion/molecule) can absorb energy, such as EMR, and be promoted to or
transition to a higher energy, excited state:
M + h M*
The excited species then loses energy and returns to its original (ground) state by emitting
radiation or transferring its excess energy to other atoms or molecules:
M* M + h or heat or energy transfer
Absorption is possible if the energy of the photon (E) exactly matches the energy
difference between ground state and one of the higher energy states:
E4
energy emitted E3
E2 Excited States
Energy
E1
E0 Ground State
absorption emission
energy absorbed, e.g. EMR
Atomic Absorption
The atomic absorption spectrum is made up of a number of narrow absorption lines, for
example, the atomic spectrum for sodium is a line spectrum.
The electron configuration for Na: 1s2 2s2 2p6 3s1. The electronic (UV/VIS) transitions
involve excitation of the single outer electron from 3s to 3p, 4p and 5p.
These excitations are brought on by absorption of photons of radiation whose energy
exactly match the differences in energies between the excited states and the 3s ground
state.
Below is a diagram of an atomic spectrum (Na atoms) and the corresponding energy
diagram:
Question: The energy difference between 3p and 3s is 2.107 eV. Calculate the
wavelength of radiation needed for excitation. (1 eV = 1.6 x 10-19 J) Hint: E = hc/.
Molecular Absorption
Absorption is more complicated.
Electrons not only move from one molecular orbital to another (electronic transitions),
but atoms in the molecules also undergo vibrations and rotations requiring a discrete
amount of energy to initiate or maintain.
Hence, each molecular energy state is comprised of an electronic, vibrational, rotational
component; E = Eelectronic + Evibrational + Erotational
Vibrational transitions:
o molecules have many quantized energy levels (vibrational states) associated with
the bonds holding the molecules together.
Rotational transitions:
o molecules exhibit rotational motion around its centre of gravity.
Therefore, the spectrum is complicated – continuous spectrum obtained.
Absorption Laws
Absorption of radiation follows definite physical laws.
The amount of light absorbed can give us information about the substance.
When light is absorbed within a medium, the radiant power of the beam decreases:
I0 or P0 = intensity of a beam of monochromatic light
passing through a medium
I1 or P = the intensity of radiation emerging from the
solution
When concentration is given in mol L-1 and b in cm, the constant is called the molar
absorptivity, .
A = bc = L mol-1 cm-1
This is widely used in spectroscopy to calculate solution concentrations from absorption
measurements.
2. Chemical Deviations
These occur when an analyte dissociates, associates or reacts with a solvent
forming a product with different absorption spectrum.
Also happens with some large organic molecules are at low concentrations.
3. Instrumental Deviations
These are caused by polychromatic radiation and stray light.
Because of these deviations, the extrapolation of the absorbance-concentration
relationships from standards with low concentrations to samples with high concentrations
will cause errors.
This difficulty can be overcome by preparing calibration curves that show the
experimental relationship between A and the actual concentration
Errors caused by (1) absorption by interfering molecules, or (2) variations in sample cell
width, may be eliminated if they are constant in the sample being analysed and the
calibration samples.
This requires extra care when making up the standard solutions.
Optical Instruments
Instruments for measuring absorption are comprised of five basic components:
Source
Wavelength selectors
Sample
Radiation detector
Readout Device
Radiation Sources
A good source should have a stable, high intensity output that covers a wide spectral
range.
No single source is suitable for all of the spectral regions and an instrument designed to
make measurements over a wide range of wavelengths may contain two or even three
different sources. Examples of sources:
Wavelength Selectors
Since radiation sources are continuous, some means is needed for selecting the particular
wavelength at which a measurement is to be made.
It is impossible to isolate a single wavelength from a continuum, so a very narrow band
of wavelengths has to suffice.
A narrow bandwidth improves sensitivity and selectivity.
Filters (UV/VIS):
Usually limited to one wavelength.
Absorption filters are relatively inexpensive but limited to VIS region.
Interference filters use optical interference to provide narrow bands.
These output smaller bandwidth than absorption filters
Monochromators (UV/VIS/IR):
Monochromators are comprised of:
o Entrance & exit slits
o Collimator to produce a parallel beam of light.
o Prism or grating to disperse radiation
Dr. D. Gordon-Smith, UTech
10
Gratings:
o Modern, less expensive than prisms
o However, prone to stray radiation and higher order spectra.
o Consist of a number of parallel grooves (blazes) on a polished surface.
UV-VIS gratings (300–2000 grooves/mm)
IR gratings (10–200 grooves/mm)
*Consult text for the diagram and operation of the Czerney-Turner grating
monochromator
Prisms:
o Larger, more expensive
o Unstable in laboratory atmosphere
Sample
Containers are called cells or cuvettes
They have a variety of shapes and sizes appropriate for each instrument
The cell/cuvette material must be transparent in the spectral region of interest
o UV – quartz
o VIS – quartz or fused silica glass
o IR – NaCl, AgCl or KBr
Square or rectangular cell offer constant path length and are therefore free of distortions
(remember Beer‟s law assumptions!)
Oval cells are marked in front for proper and constant light path orientation
Also, the detector electrical signal must be directly proportional to the power of the beam
striking it.
UV/VIS detectors are Photon Transducers (150 – 3000 nm):
o Respond to photons
o Based on interaction of radiation with a reactive surface.
IR transducers are Heat Transducers (600 – 40,000 nm):
o Respond to heat
*Consult the text for more information on the phototube and photomultiplier tube
detectors
Calibration Methods
Instruments only measure a property of the analyte (X) and therefore cannot tell the user
the concentration of the analyte without more information.
Hence, instrumental analysis requires calibration.
Calibration is the process of determining the relationship between a measured signal and
the concentration of analyte that generates the signal.
NOTE: calibration curves are based on Beer‟s law and there tends to be a deviation from
linearity at high concentrations. Therefore, one needs to know the linear and dynamic
ranges of the method:
4. Both solutions are made up to the same volume and their absorbances are measured
5. The concentration of the analyte in the sample is determined using the following
equation:
𝑨 𝑪 𝑽
𝑪𝒙 = 𝑨 𝑽𝒙 −𝑨
𝒔 𝒔
𝒕 𝒕 𝑽
𝒙 𝒙
where:
Cx = concentration of Ax = absorbance of the Vx = volume of
the analyte analyte solution analyte solution
Cs = concentration of At = total absorbance Vt = volume of Vs = volume of
the added standard of the analyte and volumetric flasks standard solution
standard added
NOTE:
i. The standard added is a small volume of a concentrated solution so that the total
solution volume and ionic strength do not change appreciably
ii. The standard addition method has the advantage that it measures analyte
concentration in the natural environment of the sample and compensates for
variations caused by physical and chemical interference in the sample solution
iii. Both standards and samples are exposed to the same instrument fluctuations since
both are measured at the same time
iv. The method is slow and the sample preparation involved may be more intricate.
NOTE:
The IS must be absent from the sample matrix so the only source of it is the added
amount.