1.0 PURPOSE:
To describe the standard operating procedure for calibration of micropipettes.
2.0 SCOPE:
This procedure is applicable for calibration of micropipettes used in the microbiology
laboratory.
3.0 RESPONSIBILITY:
3.1 Senior Executive-Microbiologist/ Designee
4.0 ACCOUNTABILITY:
4.1 Head-QC/ Designee
4.2 Head-QA/ Designee
5.0 PROCEDURE:
5.1 Receipt of micropipette:
5.1.1 Whenever a new micropipette is received it should be checked for its make, range that
is specified in the purchase order.
5.1.2 Verify if the supplier’s certificate of analysis is present.
5.1.3 Internally calibrate the micropipette given below the point 5.2 as follows.
5.1.4 After calibration that gives satisfactory result, assign the micropipette a serial number
and other details in the format: IM-II/QM/007/F2.
5.1.5 Assign the micropipette serial number as follows:-QM/MP/XX.
Where,
QM-Quality Control Microbiology,
MP-Micropipette,
XX-Sequential number starting from 01
Prepared By Reviewed By
Approved By
Signature
Date
Designation
Format number: IM-II/QA/001/A2-00
INTERMED Page 2 of 4
SOP No.: IM-II/QM/007
STANDARD OPERATING PROCEDURE
Title: CALIBRATION PROCEDURE FOR
Revision No.: 00
MICROPIPETTES
Department Microbiology Supersedes SOP: NA
Effective Date: Next Review Date:
Three point calibrations shall be performed for the variable range pipette. E.g. 0.01ml, 0 .05 ml,
0.1ml range pipette perform calibration at set point 0.1ml, 0.5ml & 1ml.
5.15 Example calculation
Weight of Water = 100.1 mg
Density of Water at 25ºC = 0.997044 g
Weight of water
Volume of purified water =
Density of water
100.1 mg
Volume of purified water at 25ºC =
997.044 mg
Volume of purified water at 25ºC = 100.40 µl
5.16 Recorded the observation in the format: IM-II/QM/007/F1
5.17 Acceptance Criteria: the volume variation should be within ±2% of designated set
volume.
5.18 If any deviation observed during calibration of micro pipette the analyst shall be inform
to supervisor for necessary corrective action.
5.19 Frequency of Calibration: Once in three month or after each breakdown.
5.20 Acceptance Criteria:
Pipettes and Desired Calibration Volume
1 10 to 100 µl 10 µl 50 µl 100 µl
Prepared By Reviewed By
Approved By
Signature
Date
Designation
Format number: IM-II/QA/001/A2-00
INTERMED Page 4 of 4
SOP No.: IM-II/QM/007
STANDARD OPERATING PROCEDURE
Title: CALIBRATION PROCEDURE FOR
Revision No.: 00
MICROPIPETTES
Department Microbiology Supersedes SOP: NA
Effective Date: Next Review Date:
7.0 DEFINITIONS:
7.1 Micropipette is a laboratory instruments designed for measuring small quantities of
liquid for routine analysis.
Prepared By Reviewed By
Approved By
Signature
Date
Designation
Format number: IM-II/QA/001/A2-00