LUCEA LED
LUCEA 50 - 100
| 2 |
Copyright
All rights reserved. This document may not be copied, adapted or
translated without prior written permission, except as permitted under
copyright law.
© Copyright Maquet SAS
Contents
8.23 Replacing the top cover and backup unit front panel
subassembly | 41
8.24 Special case: US power supply on a termination box | 42
8.25 Electrical connections: installation diagrams | 48
8.26 Recommended cable cross-sections | 49
8.27 Replacing the power-supply on the wall-mounted
suspension | 53
8.28 Replacing the cover on the wall-mounted suspension | 55
8.29 Replacing the mobile spring arm | 56
8.30 Replacing the power supply on the mobile light and/or
the mains cable | 60
8.31 Assembling the LUCEA 100 mobile stand | 64
8.32 Adjustments | 66
9 Configuring the second spare remote control (optional) | 68
10 Maintenance kits | 69
11 Spare parts and consumables | 73
12 Maintenance tools (OPM) | 74
13 Diagrams and parts lists | 76
14 Wiring diagrams | 148
14.1 Wiring diagram: LUCEA 100 lighthead | 148
14.2 Wiring diagram: LUCEA 50 lighthead | 149
15 Drawings of ceiling covers | 150
16 Connecting and disconnecting - Mobile version | 152
17 Summary of power supply wiring diagrams | 153
18 Installation block diagrams | 155
Symbols Meaning
Absolute requirement
May endanger patient and user safety
F
Recommendation
Risk of damage to device or accessories
CE label
The device complies with the requirements of European directive
93/42/EEC relating to medical devices
Medical Equipment
Classification for electric shocks, mechanical hazards and fire risks in
accordance with UL 60601-1,
IEC 60601-2-41 and CSA C22.2 No. 601-1-M90
Symbols Meaning
Attention
Follow the instructions for use
Manufacturer
~ Alternating current
Direct current
REF.
Product technical description and serial numbers.
SN.
Comply with the applicable handling precautions for products that are
sensitive to electrostatic discharges.
S
Protective earth
Medical Equipment
Classified with respect to electric shock, fire and mechanical hazards
only, in accordance with UL 60601-1, IEC 60601-2-41 and CSA C22.2
No. 601-M90
CE label
The device complies with the requirements of European directive
93/42/EEC relating to medical devices
LNE/G-MED certifies that the quality system developed by MAQUET SAS for design, imple-
mentation, sales, installation and after-sales service of surgical lights complies with the
requirements of the following international standards:
ISO 9001:2008
ISO 13485:2004
3.2 CE Marking
Compliance with the requirements of European Directive 93/42/EEC dated 14 June 1993
relating to medical devices was assessed as described in Annex VII of the Directive. LUCEA
50-100 lights belong to Class I as described in Annex IX of Directive 93/42/EEC.
! WARNING
To be compliant with IEC 60364-7-710, lighting backup systems must function for at least 3
hours.
The external system must therefore have a capacity of at least 3 hours. The MAQUET
backup unit makes it possible to check compliance with this requirement.
! WARNING
MAQUET is not responsible for the customer's external backup system (capacity, charge
time, lifespan, etc.)
! Warning
If the system is installed with a customer backup supply, disconnect the customer battery
charger before starting the battery life test.
Reconnect the customer battery charger at the end of the test.
Main light
cm
Diameter d10 23 (9) ±10% 23 (9) ±10%
(inches)
cm
Diameter d50 12 (4.7) 12 (4.7)
(inches)
cm
Diameter d50/d10 > 0.55 > 0.55
(inches)
cm
Illumination depth at 20% 170 (67) 106 (41.7)
(inches)
cm
Illumination depth at 60% 116 (45.5) 62.5 (24.5)
(inches)
Mechanical
mm 437.1x246.39 675.8x243.8
Lighthead overall dimensions
(inches) (17.21x9.70) (26.61x9.60)
Protection rating against a hazardous
- Normal
inflow of water
Methods of sterilisation
- See User's manual
or disinfection
Description Specifications
Supply voltage 100V-240V/50-60 Hz
Max. power of power supply and LUCEA 50 lighthead without 60 VA
battery pack (mobile LCA 50)
Max. power of power supply and LUCEA 100 lighthead without 120 VA
battery pack (mobile LCA 100)
Max. power of power supply and LUCEA 50 lighthead with battery 145 VA
pack (mobile LCA 50 B)
Max. power of power supply and LUCEA 100 lighthead with 247 VA
battery pack (mobile LCA 100 B)
Weight of mobile LUCEA 50 without battery pack (Mobile LCA 50)
Weight of mobile LUCEA 50 with battery pack (Mobile LCA 100) 11 kg
Weight of mobile LUCEA 100 without battery pack (Mobile LCA 50 21 kg
B) 25 kg
Weight of mobile LUCEA 100 with battery pack (Mobile LCA 100 59 kg
B)
(lightheads included)
Length of mains supply cable 5m
Two antistatic swivel casters with brakes
dia. 125 mm
3 swivel casters
Description Specifications
Users must contact their hospital's sanitary specialists. The recommended products must be
used and the recommended procedures followed. Should there be any doubt concerning the
compatibility of active agents to be used, contact your authorised Maquet dealer.
F Recommendation
Check that the power is switched off and the light has cooled down before starting cleaning.
Wipe the equipment with a cloth moistened with a surface cleaner. Follow the manufacturer's
dilution, application time, and temperature recommendations.
Rinse the unit with a cloth and clean water. Wipe dry.
Wipe evenly with a cloth moistened with disinfectant. Follow the manufacturer's
recommendations.
Use a cloth to rinse with clean water in order to remove residues (in particular products
containing aldehydes, quaternary ammonium or surfactants).
Wipe with a dry cloth.
Make sure no liquid residue is left on the device after cleaning.
Recommended products
Prohibited products
! WARNING
Solutions containing glutaraldehyde, phenol, iodine, bleach, alcohol or chloride ions must
not be used.
Fumigation methods are unsuitable for disinfecting the unit and must not be used.
7 Maintenance
! WARNING
The operation and safety of the device may be affected by the removal of certain compo-
nents during servicing operations. For example: For example:
When servicing the electrical power supply,
When servicing the suspension arms and balance system,
When servicing the lighthead.
Contact your authorised Maquet after-sales service department for this type of servicing.
! WARNING
Clic
Annual checks
(must be performed by an authorised technician)
Check the verticality of the suspension tube (ceiling
model).
! WARNING
Annual checks
(must be performed by an authorised technician)
Check the illumination adjustment by pressing and
releasing or pressing and holding and
5
levels
Comments: Signature:
8 Maintenance procedures
! WARNING
Always disconnect the lighthead from the mains and let it cool before performing
maintenance.
! WARNING
The spring arm must be lowered in order to install the lighthead. Doing so compresses the
spring and results in a safety hazard.
Stay clear of the vertical reach of the spring arm.
Lock the spring arm in a safe position.
! WARNING
! WARNING
Remove the sterilisable handle. Unscrew the two CHc M4x65 screws.
Using a flat-head screwdriver, prise off the Disconnect the cable connecting the top
top cover (3). cover and the control board.
F recommendation
To refit the top cover, follow the above steps in reverse order.
! WARNING
Make sure that the top cover and bottom cover click together correctly.
! WARNING
Make sure that the cable connecting the top cable to the control board is correctly
connected.
F Recommendation
Using a can of dry, non-static-generating compressed air, dust off the collimators and the
inside of the transparent housing.
Remove the four Torx screws (1) holding Remove the housing/interface and housing/
the housing subassembly (2) to the cover plate subassembly.
bottom housing subassembly.
Correctly seated
Incorrectly seated
To install the new subassembly, perform the Make sure that the fitted housing is cor-
above steps in reverse order. rectly seated.
! WARNING
During this procedure, take care not to scratch or damage the collimators and housing.
! WARNING
When replacing the housing, make sure that the end of the structure (handle + LED panel) is
correctly seated in the inner groove of the housing.
2
1
Loosen the two Torx screws by a few turns. remove faceted structure, push on the
To
side then pull on the assembly.
! WARNING
Make sure that the cables on the new LED board-collimator subassembly are correctly
connected.
F Recommendation
Using a can of dry, non-static-generating compressed air, dust off the collimators and the
inside of the transparent housing.
21
3
2
4
Unplug the connector (1) on the capacitive Remove the capacitive keypad from under
keypad (2). the mounting plate (4).
Remove the two Torx screws (3). Install the new arm by following the
aforementioned steps in reverse order.
Remove the sterilisable handle (1) Using a flat-head screwdriver, prise off the
Remove the two CHc M4x12 screws (2). top cover.
Using a Torx screwdriver, remove the Fh
screw (3).
2
3
1
Remove the four nuts (1) and the four Pull off the transparent housing subassem-
screws (2). bly (3).
Install a new transparent housing subas-
sembly in its place.
Follow the aforementioned steps in reverse
order to install it.
! WARNING
During this procedure, take care not to scratch or damage the collimators and housing.
! WARNING
Make sure that the end of the faceted structure is correctly seated in the groove of the trans-
parent cover.
Incorrectly seated
F Recommendation
Using a can of dry, non-static-generating compressed air, dust off the collimators and the
inside of the transparent housing.
! WARNING
Make sure that the cables on the new subassembly are correctly connected.
! WARNING
Take care when handling the electronic boards. They are supplied in an electrostatic enve-
lope and must therefore be handled with great care. Certain components on the boards may
be destroyed by static electricity (ESD, electrostatic discharge). An anti-static wrist strap
must be worn to eliminate this risk.
F Recommendation
Using a can of dry, non-static-generating compressed air, dust off the collimators and the
inside of the transparent housing.
! Recommendations
Read the following sections before removing the subassembly: 'Removing the top cover',
'Removing the transparent covers', 'Camera' and 'Faceted structure, collimator and control
board subassembly'.
1
2
B
A
Remove the two Torx screws (1). Discard the defective bottom section (B).
Remove the top frame (2). Install the new subassembly by following
Remove the top section (A). the aforementioned steps in reverse order.
Disconnect all the cables.
Remove the touch control (see section
8.5).
! WARNING
Check that the focus function operates correctly before refitting the covers.
! WARNING
The fuse protects the lighthead's control board in the event the voltage polarity is reversed.
! WARNING
Take care when handling the electronic boards. They are supplied in an electrostatic enve-
lope and must therefore be handled with great care. Certain components on the boards may
be destroyed by static electricity (ESD, electrostatic discharge). An anti-static wrist strap
must be worn to eliminate this risk.
Turn off the lighthead and remove the sterilisable handle and the top cover as instructed in
section 8.6.
Using a flat-blade screwdriver, prise off the Using a flat-blade screwdriver prise the
mount on the lighthead and discard it. camera out of its slot.
Disconnect the camera connector from the
lighthead connector.
Camera
Locking/
Unlocking tab
! WARNING
Checkpoints:
Make sure that the camera locks correctly Check that the camera connectors are
in place on its frame (you should hear a correctly connected.
'click').
! WARNING
AQUET SAS does not guarantee the signal quality for cable lengths longer than those
M
proposed in the price list. To prevent potential interference with the video signal (electrical
power supply, isolation transformer), the cable paths need to be checked.
2 4
1
1 2
Analogue signals
PAL composite analogue output (or NTSC depending on option)
8
5
7
6 4
3
Remove the round cap (7) and the two Remove the circlip (5) then the shouldered
screws (8) supplied with the spring arm. ring and the two washers (3).
Remove the rotating contact (6) from the
shaft.
2
1
! WARNING
When installing the new arm, make sure that the circlip is correctly seated in the throat of the
spring arm.
! WARNING
The electrical connections must be performed by a qualified technician only, with the product
and cables disconnected from the power supply.
! WARNING
Check the polarity of all electrical connections before turning on the power.
! WARNING
Damaged wire insulation may result in the risk of electric shock. To protect against the risk of
electrocution, insert the power cables carefully.
! WARNING
A
ut
O A
A
e
rti
So
! WARNING
When installing the power supply, secure the mains cable and the suspension cable to the
tube to reduce mechanical stress on the power supply terminal blocks.
Output
OUT
To suspension
Input
IN
100V-240V
Flange
! Warning
A - Installation:
Place the ferrite bead on the cable connecting to the hospital network, upline from the
Power Supply LCA 24V.
Ferrite bead
Hospital Power supply LUCEA 50/100
network LCA 24V Configuration
There should be one ferrite bead on each 24 V power supply lead entering the power supply
unit.
Pull the two tabs Wind two loops of cable ass the cable back
P
through the bead through and close the bead
1-D
isconnect the four-wire power supply
connector from the backup switchover
2
board.
Sortie
2 - Disconnect the four-wire connector on
the backup switchover board from the
Sortie terminal block.
Alimentation 3 3 - Disconnect the backup unit connector
Bascule
Entrée secours and the customer backup wires.
4 - Remove the defective switchover board.
5 - Install a new backup switchover board.
1
6 - Reconnect all the connectors and the
customer backup wires.
Ferrite
bead
Flange
Output
OUT
To
suspension
2
A Backup unit
Input
IN
100V-240V
Customer
backup 1
8.19 Replacing the wall cover and front wall panel on the
backup unit
Cover
Power supply
Ref. 699606024 1
Mains
cable
Cover earth
wire
4
3
To
suspension
Flange
Suspension
! WARNING
Make sure that the earth wire is between the power supply and its stud/mount.
Disconnect the two cables (A) and (B). etach the switchover board.
D
Remove the communication cable (cat. 5). Remove the two screws (ARD600520308)
Disconnect the customer backup input. and the two washers (ARD602420306).
Replace the board and carry out the afore-
mentioned steps in reverse order.
2
1
Power supply
Ref. 699606024
Mains
cable
To
suspension
Ferrite
bead
Customer
Flange
backup Customer backup kit
input (568604949)
RJ45 cord Suspension
Wall-mounted (cat. 5) To backup unit (front
backup unit of cover)
(option)
Fit the cover and secure it in place with its four screws.
Power supply
Ref. 699606024
Mains
cables
To
lighthead 2
To Ferrite
Ferrite lighthead 1 bead
bead
Customer
backup input
Wall-mounted To backup unit Customer
backup unit (front of cover) backup input
(option)
Flange
Fit the cover and secure it in place with its four screws.
Remove the four screws and pull off the Remove the four attachment screws from
battery cover. the cover.
Power
supply
board Charger
board
TOP
Panel penetration
693200289 Optional backup
unit (OR)
{
CAT 5
Cable 568603530
Single capacitive
backup unit Cable 699910089 110V-240V
{
Cable 568603533
Cable 699910089
Charger board
568604510
Fuse
Cable
568603537
Cable 568603534
Power supply board
699606024
Fuse
12 V battery 12 V battery
699600040 699600040
DOWN
Kit of 4 A fuses (Ref. 693600063)
Power cable kit (368617555)
8.24 Replacing the top cover and backup unit front panel
subassembly
Output
+ -
! WARNING
Tighten the wires to their corresponding colours on the terminal blocks (phase/neutral/earth).
367301900
568604946
568604954
568604955
Disconnect the hospital power supply from the terminal block (1).
Remove/install each section as shown in the diagram above.
F NOTA
For LUCEA 100 video, the video cable can be separated from the power wires if necessary.
Flexible Conduit
conduit
Run power wires through the conduit attached to the Raco Box.
Power Supply
Access panel LUCEA location, above the ceiling
Raco Box 2 (max) lightheads that contains the lights.
per Nec per box
14x14x3 Termination Mount Termination box
Conduit Box
Suspension tube with power supply on top
weight : 50 lbs
(by MAQUET) Conduit fitting and within 6’ of suspension
tube.
*No high voltage to be placed inside *Ground wire starts at screw on flange and
light or multimedia arm system. continues to power supply.
250V, 20 Mount disconnect next to
NOTES : AMP 2-pole
switch-
power supply box or feed
1. Customer’s design engineer of record is responsible for the design, including code com- disconnect from 2 pole breaker in isola-
per Nec
pliance, of all facility services up to MAQUET supplied equipment. tion panel in or suite custo-
2. All conduit and wiring external to MAQUET supplied equipment, provided by customer or mer or contractor supplied.
contractor. Recommend 1/2’’ conduit. 120VAC/20A
3. Min. wire size for secondary lightpower between transformer and center mount
connections.
Connect power wires through flexible metal conduit and Raco Box locally supplied.
Wire run lengh Min wire AWG
0’-45’ 12 AWG
45’-70’ 10 AWG
| 45 |
50 50
(4) R7,5
60
25
87
°
5
6 125 125
6
15 22
15
20,5
10
60
267
235
74
60
6 6
357
70
R20
A 6 58
=25=
15
32
10
= 50 =
15 Ø à définir selon REP2
R10 R10 Ø to be defined for REP2
13,5 52,5
180
R10 R10
450
410
30
A-A
R10 R10
180
=120=
10
R10 R10
10
30
25
6 6 4 x 45°
20
A 90
R10
R1
=160=
230
200
R40
R5
97
60
130
120
56
60
R40
114 R5 4 holes
98
Outside
dimensions of 220
blanking plate
Panel cut-out
15 220
15
160
130
80
113
250
4 1 2 5 3
4 1 2 5 3
Remove the assembly by removing the Remove the assembly by removing the
screws (4) (and anchors (5) supplied) screws (4) (and anchors (5) supplied)
Remove the wall-mounted backup unit (1) Remove the wall-mounted backup unit (1)
(2) from the junction box (3). (2) from the junction box (3).
360
60 240
40
240
340
75
Option A:
ceiling-mounted
power supply
Option B:
wall-mounted
power supply
Option A
ceiling-mounted
power supply
RS
input
Or
RS Option A:
remote
with backup
unit
Option B: RS
wall-mounted
power supply
with backup unit
CAT 5
Equipment
room
Backup unit
RS
RS (option)
C1 Power supply
C1 C2 Customer
Power supply battery
Minimum Minimum
C1 length recommended C2 length recommended
cross-section cross-section
Power
C4 supply with
MAQUET
backup
LUCEA 100 3h
! Warning
If the system is installed with a customer backup supply, disconnect the customer battery
charger before starting the battery life test.
Reconnect the customer battery charger at the end of the test.
RJ crimping
RJ45 crimping tool 5 720 71 999
tools
F recommendation
o ensure compliance with wiring regulations, follow the order shown below.
T
Note: the wiring is "straight through".
Hold the RJ45 connector with the clip downwards, the cables being inserted facing the
installer.
1 = White / orange
2 = Orange
3 = White / green
4 = Green
5 = White / blue
6 = Blue
7 = White / brown
8 = Brown
1 2 3 4 5 6 7 8
1
1
Remove the 10 long cross-head screws (1). Tighten the four short cross-head screws
(1).
! WARNING
The electrical connections must be performed by a qualified technician only, with the product
and cables disconnected from the power supply.
1 1
5
2
3
2
3
Remove the wall cover (1) by removing the Disconnect the two connectors from the
two attachment screws (2) and their power supply board.
washers (3). Remove the four attachment screws then
replace the board.
Green
Brown
Blue
F Recommendation
! WARNING
The wall bracket and the power supply may be installed and connected only by an electrician
or a MAQUET authorised service agent.
! WARNING
Reversing the polarities may result in damage to the lighthead's electronic circuit board.
Rated power
60 VA 120 VA
(power supply + lighthead)
2
3
2
3
! WARNING
To ensure the user's safety, make sure that all covers on the wall-mounted version are firmly
in place.
Relax the spring arm. To relax the spring arm, unscrew the two
taper-head hex socket screws (1).
3 6 4
5 7
Remove the lighthead. Loosen the brake screw (7) with an Allen
key.
Remove the Torx screw (6).
Remove the fork (4) from the spring arm
(5).
Separate the connectors (3).
1
3
1
Remove the hex socket screw (1). Lubricate the unpainted portion (2) of the
Remove the spring arm. new spring arm (4) (tube of lubricant
supplied).
Fit the new preassembled spring arm (4) on
the pole (3) such that the oblong hole (5) on
the spring arm is aligned with the hole on
the pole (6).
Insert and tighten the hex socket screw (1).
! WARNING
If the spring arm is not relaxed before the lighthead is removed, it may spring up abruptly
and pose a safety hazard.
Lower the spring arm and lock it in a safe position when installing the lighthead.
Always install the lighthead before loading the arm.
The spring arm must be relaxed before the To relax the spring arm, unscrew the two
lighthead is installed. taper-head hex socket screws (1).
3 6 4
5 7
Reload the spring arm. To load the spring arm, tilt the lighthead and
spring arm upwards until the holes line up
with the tappings (2).
Install the two taper-head hex socket
screws (1).
! WARNING
The electrical connections must be performed by a qualified technician only, with the product
and cables disconnected from the power supply.
Remove the screws and lift off the cover. Replace the power supply board or the
batteries.
Disconnect the cables. Reassemble by following the aforemen-
tioned steps in reverse order.
Remove the screws and lift off the cover. Replace the power supply board.
Reassemble by following the aforemen-
tioned steps in reverse order.
Remove the screws and lift off the cover. Replace the power supply board or the
batteries.
Disconnect the cables. Reassemble by following the aforemen-
tioned steps in reverse order.
! WARNING
Incorrect assembly may cause the light to tip over. To avoid damaging the equipment or
causing accidents, carefully follow the instructions given in each step of assembly.
2
3
1
2 1
4 3
Remove the five screws and washers (1) Insert the tube (1) in the socket on the
with the Allen key supplied. metal star base, aligning its notch (2) with
• Fit the four legs with casters on the base. the notch on the base (3). Press firmly (the
Place the short legs (2) at the rear and the diameters of the tube and socket are similar
long legs (3) at the front. to avoid any play).
• Put the four screws and washers back in • Secure the tube in place with the screw and
place. star washer (4).
Slide the unit (1) onto the stand (2), aligning Fasten the unit to the leg with the M8x25
the notch with the short leg in the middle. screw (3), washer, external tooth lock
washer, and nut (4).
F Recommendation
5
4 7
1
Slide the unit (1) onto the stand (2) and turn Remove the screw and washer holding the
it until it locks into place in the stand. bumper (7) in place.
Fasten the unit to the stand by tightening
the M8x25 mm screw and the star
washer (3).
Unscrew the ten M6x10 screws (5) and
remove the cover from the unit (4).
Fasten the unit (1) with the two M6x10
screws and external tooth lock washers
inside the unit (6).
! WARNING
Flammable gases may be released and explode if Maquet original batteries are not used.
Always replace these batteries with Maquet batteries (see section 10 'Spare parts and
consumables').
! WARNING
Flammable gases may be released and explode if the poles on either or both batteries are
reversed. Insert the batteries with the terminals facing outward and follow the colour codes
below when wiring the terminals.
To power To
supply lighthead
red black
White cable on
black terminal
Black cable on
black terminal
Insert the two batteries (8) in their compartments with the terminals (9) facing outward.
Fasten the connectors to the terminals (9) by tightening the screws and nuts with a 10 mm
spanner. Make sure that the colours correspond as indicated above.
Fully slide the batteries into their compartments then fasten the bumper in place with the
screw and washer.
Secure the cover by tightening the 10 screws.
8.33 Adjustments
Pull out the silicone cap. lace the silicone cap to the side to keep it
P
out of the way.
Press and hold the two buttons , for 5 seconds to access the configuration menu
as indicated by the right LED being lit
Configuration menu :
Adress 0x60
Adress 0x61
Backlight menu :
The system returns to normal mode and the lighthead number backlight is as configured.
In configuration mode, the module does not respond to messages from the master.
! WARNING
IMPORTANT:
Two different addresses must be used for the two spare remote controls.
If this is not the case, the second spare remote control cannot perform the mains/battery
changeover test and the battery life test.
10 Maintenance kits
F Recommendation
IMPORTANT:
Before placing an order for an LED regulator board and collimator subassembly, check
the code specified on the label, located on the underside of the lighthead.
If no information is given, please order a subassembly with colour temperature K6.
ARD56010513 Connector 1
References Description
Part
No. Description Qty
number
Diagram 1 LUCEA 50
1 lighthead
6
2
10
11
12
13
7
9
8
3
14
15
16
4
32 19 30
17
19
33
34
35
18
20
36
19
22
21
23
A
24
25
26
27
28
F Recommendation
IMPORTANT:
Before placing an order for an LED regulator board and collimator subassembly, check
the code specified on the label, located on the underside of the lighthead.
If no information is given, please order a subassembly with colour temperature K6.
Part
No. Description Qty
number
6
2
10
11
12
13
7
9
8
3
14
15
16
4
32 19 30
17
19
33
34
35
18
20
36
19
22
21
23
A
24
25
26
27
28
F Recommendation
IMPORTANT:
Before placing an order for an LED regulator board and collimator subassembly, check
the code specified on the label, located on the underside of the lighthead.
If no information is given, please order a subassembly with colour temperature K6.
42
9
28
7 24 23 22
43
A 6 21
4
38 1 41 14 15 16 13 1718 44
39 40 3
34 20 19 11 12 26 10
32
33
2 45 5
31
25
29
30
lighthead
LUCEA 100
42
9
28
7 24 23 22
43
A 6 21
4
38 1 41 14 15 16 13 1718 44
39 40 3
34 20 19 11 12 26 10
32
33
2 45 5
31
25
29
30
LUCEA 100 lighthead
Diagram 3 LUCEA 50
Diagram 4 LUCEA 50
5 3
Diagram 7
Diagram 8 LUCEA 50
Spring arm
1
16
2
8
17
18 3
4
7 5
6
11
13
12 10
14
15
Spring arm
16
17
18 1
8 4
5
6
7
10
11
13
15
12
14 9
1
5
2
13
14
6
12
7
9
8
1
3
4
10
11
=
=
7
1
119,5
8
9
5
1
10
2
4
7
3
5
6
11
=
=
12
119,5
8
2
9
4
10
5
6
11
=
=
12
119,5
8
1
7
6
2
8
9
3
4
10
11
=
=
119,5
14
13
12
7
1
9
8
Diagram 14 LUCEA 50
1
2
4
9
11
13
5
10
12
14
Diagram 15 LUCEA 50
22
21
19
23
20
16
17
20 19 18 15
6 3 4 2 1
8
13
9
11
5 14 12 10 7
11 3
8
7 10 1
14
6 13
12
5
9
2
1
5 690 72 999
6
4
3
12
7
5
13
10
9
11
A
Two-position button:
LUCEA 50 position and
LUCEA 100 position
32 29 28
34 2 8 3 4 5
6
21 20 14 19 9 23 22
9
10
31
11
13
22
12
15
30 13 33
9
18 16
17
10
24 25 26 27
LUCEA 50 Rolling base
5 690 14 200
Part
No. Description Qty
number
1 CHc screw M4x8 DIN 7984 1
2 5 690 14 027 Stand tube 1
4 5 690 14 025 Plastic star section 1
5 M8 nut DIN 934 5
6 5 690 14 022 Blue 30 - LUCEA 50 short stand section 3
7 5 690 14 021 Blue 30 - LUCEA 50 long stand section 2
8 5 690 14 024 Blue 30 - LUCEA 50 cap 5
7
5 3
12
6
8
10
13
11
5 690 75 999
Part
No. Description Qty
number
1 5 690 14 000 Mobile stand straight tube 1
2 5 690 14 040 Mobile 5-star stand 1
3 5 690 14 032 Mobile transformer unit 1
4 5 690 14 031 PSU bracket 1
5 5 665 07 33 M3x8 screw DIN 966 - EN ISO 7047 8
6 6 951 10 310 M3 x 10 stainless steel cylindrical spacer 4
7 5 690 03 021 MPS 120-228 power supply 1
8 5 665 30 06 Flat washer, A-shape, 3.2 mm DIN 125 4
9 6 005 30 306 CHc screw M3x6 A2 stainless steel 4
10 5 665 07 62 CHc screw, low profile M5x10 DIN 6912-8.8-A4E 3
11 5 665 30 75 Star washer, A-shape 5.3 mm DIN 6797-A4E 10
3 690 10 997 Blue 80 - Mobile LUCEA 100 mains cable loom S/A
17 5 690 03 031 Wiring 1
18 6 914 00 095 KEC mains connector 1
19 6 890 04 520 2-pole fuse drawer without voltage selector 1
20 5 601 89 21 Sleeve 1
16
14
15
26
27
4 24 5 23 22 6 17 3 20 18
2 24
21
3
19
5
5
13 12 25 11 10
29
5 686 04 524 Charger board
1
A
Diagram 21
28
2 4 13 14 5 35 Two-position button:
3
LUCEA 50 position and
6 34 LUCEA 100 position
7
3 17 9
8
30
19
9
10
11 20 21
32 31 33 12
26
23
22 24
38
37
36
24
23
39
27
18
13 16
25
26
LUCEA 100 Rolling base
2 1 6 14
16 8
20
4
15
7 5
17
22
19 18
11 21
3
9
10
12
13
2 6 14 1 16 20
8 15
18 4
5
7
17
19 22
21
3 11
13 12 10
1 6
14
8
2
16
7
5
4
18 17
11 19
3
15
13 12 10
1 6 14 16
15
4
7 17
5 18
19
3
11
9
13 12 10
15
14
9
13
10
6
4
7
3
5
8
16 18
17
19 11
12
1
15
14
9
13 10
6 8
11
7
5 18
4
16
17 3
19
12
15 27 11 28 25 22
17
23
16
24
3
30 29 26 12 14 13
5
31
21 18
20 19
10
4 6
9 2 33 32 34 7 1
1
2 5
4 1
2
5
6
6
3
4
5
8
Diagram 33 LUCEA 50
Wall-mounted suspension
4
3
2
1
Wall-mounted suspension
4
2
3
1
2
1
3
1
2
10
3
4
5
4
5
11
7
8
9
4
12
11
6
10
14 Wiring diagrams
Fork cable
Ref. 568603550
Control
board
Keypad cable
Ref. 568604513
Capacitive keypad
568604501
Switch cable
Ref. 568603540
Control
board
On/off switch
Fork cable
Ref. 568603527
Ferrite bead
Control board
Keypad cable
Ref. 568604513
Switch cable
Ref. 568603540
9 ±0,2 Ø478 ±1
2x Ø5,5 8
2
115 +0
R600
83 ±0,5
83 ±0,5
R13
A Ø77 ±0,2
126 ±1
A (1:1)
2 ±0,5
16
2 ±0,5
R8 R4
17 ±1
47±1
142 ±0,5
2
250 +0
R13
83 ±0,5
83 ±0,5
A Ø77±0,2
Ø126 ±1
A (1:1)
2 ±0,5
2±0,5
16
R8 R4
17±1
47±1
Ø500 ±0,3
Ø480
45° 45°
6- 00,1
6- 00,1
Ø77±0,2
1,5
4 100 ±0,2
xØ
7 165 ±0,2
! WARNING
Example with
a mobile LUCEA 50
! WARNING
Do not pull on the power supply lead to disconnect it from the power outlet.
28 V output
Mains 4 mm²
100/240V - recommended
50/60Hz up to 80 m
Power supply
28 V output
Mains 4 mm²
100/240V - recommended
50/60Hz Power supply up to 80 m
+
Switchover board
Customer backup
4 mm²
28 V output
Mains 4 mm²
100/240V - recommended
50/60Hz Power supply up to 80 m
with Maquet backup
24V customer
input, 4 mm²
recommended 24 V DC output
(AC/DC) (insulated cable)
24 V
power supply
LUCEA
LUCEA 50 simple plafonnier : ARDLCA209012C LUCEA 100 + LUCA 50 plafonnier : ARDLCA20902C LUCEA100 simple plafonnier : ARDLCA219000C
Installation block diagrams
Alimentation 90-240V :
Murale simple : ARD568604954 Pack batteries + Alimentation/Chargeur (secours) : Options :
Murale double : ARD568604955 Caméra : ARD568604971
Plafonnière simple : ARD568604956 Power Module LCA 1W BM : ARD568604959 Télécommande : ARD 568604970
Plafonnière double : ARD568604957
Installation block diagram: ceiling-mounted
515053999 515053999
Alimentation murale :
Simple : ARD568604954
Double : ARD568604955
LCA 50 : ARD568604987
LCA100 : ARD568603987
Contre-plaque :
ARD569079999
Kit alimentation
LCA 50 et LCA100 : ARD568604958
Bras ressort MOB LCA50 : ARD 568604940 Bras ressort LCA100 : ARD 568603940
Bras ressort MOB BAT : ARD 568604942
OU
OU
Simple ref. 568604952 Double ref. 568604953 Simple ref. 568604956 Double ref. 568604957 Simple Double Alimentation 1 voie Simple Double
1 voie 2 voies 1 voie 2 voies ref. 568604954 ref. 568604955 ref. 568604959 ref. 568604954 ref. 568604955
1 voie 2 voies 1 voie 2 voies
Kit bascule secours Kit bascule secours Kit bascule secours Kit bascule secours Termination Box Termination Box
(option) (option) (option) (option) ref. 568604946 ref. 568604946
1 voie 2 x 1 voie 1 voie 2 x 1 voie
ref. 568604949 ref. 568604949 ref. 568604949 ref. 568604949
Kit report mural Kit report mural Kit report mural Kit report mural Kit report mural
1 voie ref 568604950 2 voies ref 568604951 1 voie ref 568604950 2 voies ref 568604951 1 voie ref 568604950
(option) (option) (option) (option) (option)
MAQUET SAS
Parc de Limère
Avenue de la Pomme de Pin The Getinge Group is a leading global provider of equipment and
CS 10008 ARDON systems that contribute to quality enhancement and cost efficiency
within healthcare and life sciences. Equipment, service and tech-
45074 ORLÉANS CEDEX 2, France nologies are supplied under the brands ArjoHuntleigh for patient
Telephone: +33 (0) 2 38 25 88 88 handling and hygiene, disinfection, DVT prevention, medical beds,
Fax: +33 (0) 2 38 25 88 00 therapeutic surfaces and diagnostics, Getinge for infection control
and prevention within Healthcare and Life Science and Maquet for
Surgical Workplaces, Cardiovascular and Critical Care.
To find your closest representative:
Please go to
www.maquet.com