FUJI COMPUTED RADIOGRAPHY

CR Console Lite
Product Specifications

5th Edition
June, 2003

Fuji Photo Film Co., Ltd.

010-017-40 06.2003
CR Console Lite Product Specifications

ii 010-017-40 06.2003
CR Console Lite Product Specifications

Export Restrictions (TWA) Applying to the CR Console Lite

Export restrictions (TWA) apply to models in which Service Pack 4/Internet Explorer 4 have
been installed on the hard disk of the CR Console Lite Unit, such as models equipped with
DELL GX50 - Celeron of 1.3 GHz or more, because the encryption key bit length of the Internet
Explorer 4 program exceeds the limit.
For models to which export restrictions (TWA) apply, government approval obtained through
due procedure is required in order to export or otherwise remove such models from Japan.
Note that the same restrictions apply not only to the main unit but also to after-sale parts,
service manuals and service training (for exporting or transferring technology) when such items
are dispatched or shipped to foreign countries and when service training is given for trainees
from foreign countries.

This system is classified as a medical device under EC Directive 93/42/EEC.


It has been designed for operation by medical personnel who have undergone the
proper training and under whose direction patients will come in contact with it.

Process waste correctly, as stipulated by local law or any regulations that apply.

! CAUTION
● The copyright of the application software of this unit belongs to Fuji Photo Film Co., Ltd.,
and is protected by the Copyright Act of Japan and the international treaty.
The application software of this unit may not be reproduced or distributed in part or as a
whole without permission of Fuji Photo Film Co., Ltd.
● This product (software operated on the specified hardware) is designed for the purpose of
displaying medical images. If this product is used for purposes other than those
designated in its functional range (use of other application software, etc.), Fuji Photo Film
Co., Ltd. assumes no responsibilities/compensations for the consequences.

010-017-40 06.2003 iii


CR Console Lite Product Specifications

Contents

1. Overview.................................................................................................................................................. 1

2. Features .................................................................................................................................................. 1

3. System Block Diagram ............................................................................................................................ 3

4. System Components ............................................................................................................................... 4


4.1 Standard Components ..................................................................................................................... 4
4.2 Options............................................................................................................................................. 4
4.3 Other System Components ............................................................................................................. 6

5. Specifications........................................................................................................................................... 7
5.1 FCR CR Console Lite Unit............................................................................................................... 7
5.2 Processing Capacity ........................................................................................................................ 7
5.3 Image Storing................................................................................................................................... 8
5.4 Image Output ................................................................................................................................... 9
5.5 Image Processing ............................................................................................................................ 12
5.6 Connection....................................................................................................................................... 12

6. Functions ................................................................................................................................................. 13
6.1 Standard Functions.......................................................................................................................... 13
6.2 Optional Functions........................................................................................................................... 16

7. Principles of Operation ............................................................................................................................ 17


7.1 Patient Reception ............................................................................................................................ 17
7.2 X-ray Exposure ................................................................................................................................ 17
7.3 ID and Exposure Condition Entries ................................................................................................. 17
7.4 Image Reading – Output.................................................................................................................. 17

8. Power Supply Conditions ........................................................................................................................ 18

9. Environmental Conditions ........................................................................................................................ 18

10. External Dimensions and Weight ............................................................................................................ 19


c PC main unit and keyboard ............................................................................................................... 19
d Touch panel LCD display.................................................................................................................... 19
e Display (CRT) ..................................................................................................................................... 20
f High-resolution Display (CRT)............................................................................................................ 20

iv 010-017-40 06.2003
CR Console Lite Product Specifications

1. Overview Connected to the cassette-type Image Reader (FCR 5000 plus /


FCR XG5000, etc.) via networks, the FCR CR Console Lite uses the
keyboard, magnetic card reader (option) and barcode reader (option)
to enter patient ID information/exposure conditions and register IP
numbers. In addition, the FCR CR Console Lite subjects CR images
handled by the Image Reader to quality assurance and optimization
processing, and outputs those images to devices for CRT and/or
LCD image display, image filing or film output.

Furthermore, provided with the functions as the Built-in Image


Reader (FCR5501 plus / FCR5502D plus / FCR XU-D1 Reader /
FCR VELOCITY U) console, the FCR CR Console Lite can enter
patient ID information/
exposure conditions and assures and optimizes CR images, on the
Built-in Image Reader. (In this case, no barcode reader is required
because IPs need not be registered.)

2. Features (1) The FCR CR Console Lite enables "patient ID information /


exposure conditions entry" and "image quality
assurance/optimization" by itself.

(2) Images are sent back to the device where patient ID


information/exposure conditions were entered. It is thus
possible for technicians to work around the exposure room.

(3) Study information shared in a system where the FCR CR


Console Lite and CR Console Plus are used mixed at the
same time allows intensive management of output images,
integrated management of reservations for study when
keyboard or magnetic card reader (MCR) is used and an
exposure room moving without the need for re-input of patient
information/exposure menus.

(4) Just one FCR CR Console Lite allows you to use the console
function on the cassette-type Image Reader or the Built-in
Image Reader (up to two units for the latter) by alternately
switching from one to another.

(5) FCR CR Console Lite receives images from any Image


Reader via the barcode system.

(6) The system features worklist management on


FCR CR Console Lite when connected to HIS/RIS.

(7) The FCR CR Console Lite stores entered patient ID


information, which can be searched for by entering the
keyword such as patient ID. Note that this search capability is
not available when the FCR CR Console Lite is connected
online.

010-017-40 06.2003 1
CR Console Lite Product Specifications

(8) Preset image processing parameters are automatically


selected through the anatomical region selection menu.

(9) FCR CR Console Lite makes it possible to network with other


competing manufacturer’s PACS.
• Available for DICOM-conformed open network

(10) Due to the effects of digital image processing, the system


produces radiographs that have a high diagnostic value and
are easy to read. Both standard and high resolution reading
are user selectable.

(11) FCR CR Console Lite converts images to irreversible JPEG


images and saves them on an FD, etc. for example, for
reference usage purposes.

(12) The user authentication setup allows you to enhance the


security function.

(13) Exposure results can be output in text data format.

(14) Adoption of the touch panel type color LCD guarantees you
improved viewability and user-friendly operation environment.

2 010-017-40 06.2003
CR Console Lite Product Specifications

3. System Block Diagram

010-017-40 06.2003 3
CR Console Lite Product Specifications

4. System Components
4.1 Standard Components ‰ FCR CR Console Lite Unit and Touch Panel LCD
• IR CL LITE AX-AXX WITH LCDB E
(For FCR 5000 plus / 5000R plus / 550X plus /
XU-D1 Reader / XG-1 Reader)
Including Network Interface, Keyboard and
AC Power Cord 100–120V.
The code AX and/or AXX represents hardware type and/or
application software version and may be replaced with A1, A2
and/or A01, A02 and so on, according to upgrading in the future.

‰ FCR CR Console Lite Unit


• IR CL LITE AX-AXX E
(For FCR 5000 plus / 5000R plus / 550X / XU-D1 Reader /
XG-1 Reader)
The code AX and/or AXX represents hardware type and/or
application software version and may be replaced with A1, A2
and/or A01, A02 and so on, according to upgrading in the future.

‰ Display Monitor
Select either Touch Panel LCD, Full-color CRT or
High-Definition Monochrome CRT
• IR 346 DISPLAY LCD E or IR CL DISPLAY LCD E
• IR 346 DISPLAY CRT 17 IN CE E

‰ AC Power Cord
Select 200-240V for the FCR CR Console Lite as required.
• HIC 655 POWER CORD EU 200V E
• HIC 655 POWER CORD UK 200V E

4.2 Options ‰ Local Print Interface


Including Interface Board and Key Software for Local Print
• IR 346 LOCAL PRINT E

‰ Magnetic Card Reader (MCR)


Used for inputting patient's information with magnetic card
• IR 346 MCR ISO E (For ISO Type)
• IR 346 MCR JIS E (For JIS Type)

‰ Barcode Reader
 Used to read cassette IP barcode.
(Not required when a built-in type Image Reader,
such as FCR550X plus / FCR XU-D1 Reader, is used as
the connected Image Reader.)
 Used to search patient's information when patient's
information on-line system or DICOM Modality Worklist
Information Model has been constructed
• IR 346 BARCODE E

4 010-017-40 06.2003
CR Console Lite Product Specifications

‰ LCD Stand
A stand used to retain the Touch Panel LCD.
• IR 346 LCD STAND E

‰ Wall Bracket
A bracket used to fasten the Touch Panel LCD or the Magnetic
Card Reader or Barcode Reader on the wall.
• IR 346 LCD BRACKET E (for LCD)
• IR 346 MCR BRACKET E (for MCR)
• IR CL BCR BRACKET E (for Barcode)

‰ Serial Interface Board


Required for performing the optional function of patient's
information on-line system and/or X-ray control unit on-line
• IDT-4 IF BD FOR PCI E

‰ Various types of cables


 Cable between FCR CR Console Lite and Image Recorder
• ODF/PS IF CABLE *M E (For Local Print)
(* : 5, 15, 30, 45, 60, 100, 150)
 Connector conversion cable
This cable is used for conversion from full-pitch 50-pin
connector to the half-pitch 50 pin connector
• MF-PS 667/DMS IF CBL E

‰ Various Optional Software


 DICOM
• Modality Worklist Information Model
– FIND SOP Class (SCU)
(IR CL DICOM ID MWM E)
(IR CL DICOM ORDER MWM E)
• Modality Performed Procedure Step SOP Class (SCU)
(IR CL DICOM MPPS E)
• CR Image Storage SOP Class (SCU)
(IR 346 DICOM CR STORAGE E)
(IR 346 DICOM IM PROCESSED E)
• Storage Commitment Push Model SOP Class (SCU)
(IR CL DICOM COMMITMENT E)
• Basic Grayscale Print Management Meta SOP Class (SCU)
(IR 346 DICOM PRINT E)
 DMS Network
(IR 346 DMS NET-WORK IN-OUT E)
 Patient's information on-line
(IR-346 ID ONLINE E)
 X-ray control unit on-line
(IR CL X ONLINE E)
 Detailed QA
(IR 346 DELUX-QA E)
 Electronic Shutter
(IR CL SHUTTER-PROC E)

010-017-40 06.2003 5
CR Console Lite Product Specifications

 Image composition function


(CL IM COMPOSITION E)
 MFP
(IR 346 MFP E)
 GPR
(CL GPR E)
 ES
(IR CL ES E)
 Image magnification/full screen display
(CL IM-MAGNIFY FOR LITE E)
 LUT adjustment
(CL LUT ADJUSTMENT FOR LITE E)

4.3 Other System Components (1) Image Reader


FCR XG5000
FCR XG-1 Reader
FCR 5000 plus / 5000R plus (IDT type)
FCR VELOCITY U*
FCR XU-D1 Reader
FCR 5501 plus / 5501D plus / 5502D plus
* When the CR Console Plus is connected, achievable
processing capacity will be different.
Note: FCR 5000R plus (CSL type) cannot be connected.

(2) Image Recorder


FM-DP L
DRYPIX 3000 (DICOM I/F specification)
DRYPIX 1000 (DICOM I/F specification)
DRYPIX 7000 (DICOM I/F specification)
FM-DP 3543 / FM-DP 2636 (CR I/F specification)
FL-IM D / FL-IM DM
FL-IM 3543 II / FL-IM 2636 II
FL-IM 3543N / FL-IM 3543MN / FL-IM 2636N
FN-PS 551

(3) DMS (Via Network I/F)


QA-WS 771
HI-C 655
HI-C 654
Synapse Server
OD-F 624 / 624L

(4) ID Card Writer: IC-W701


Used for issuing card for when patient attribute information is
needed to be input with a card (for ISO format).

* See the individual product specifications for details.

6 010-017-40 06.2003
CR Console Lite Product Specifications

5. Specifications
5.1 FCR CR Console Lite Unit (1) PC main unit : DELL Optiplex GX50
• CPU : Celeron, 1.3GHz or equivalent processing capacity
• Memory : 512MB or more
• HDD : E-IDE 40GB or more

(2) Display Monitor


• LCD : TOUCH PANEL SYSTEM 9415TD15/G2
15-inch LCD with touch panel (1024×768)
• CRT : DELL P780
17-inch Full-color CRT (1024×768)
* The product names described above are registered trademarks or
trademarks of manufacturers/sales companies, in general.
Since general purpose personal computers are used for the unit
control section, the appearance and specifications are subject to
change or improvement without notice.

(3) OS
Windows 2000 Professional (English)

5.2 Processing Capacity (1) Time required for film output


(Requirements: MFP applied and DRYPIX 7000)

 FCR XG5000
(10pix/mm high-density image reading
in the 14"×14" two-image output format)
Network print via the FCR CR Console Lite
Approx. 110 sec.

 FCR 5501 plus / XU-D1 Reader


(10pix/mm high-density image reading
in the 14"×14" two-image output format)
Network print via the FCR CR Console Lite
Approx. 210 sec.

 FCR 5502D plus (10pix/mm high-density image reading


in the 14"×14" two-image output format)
Network print via the FCR CR Console Lite
Approx. 215 sec.

 FCR VELOCITY U (10pix/mm high-density image reading


in the 14"×14" two-image output format))
Network print via the FCR CR Console Lite
Approx. 90 sec.

010-017-40 06.2003 7
CR Console Lite Product Specifications

(2) Time required for image display on the FCR CR Console Lite
<For 14"×14">
Image display start on Image display completion
the FCR CR Console Lite on the FCR CR Console Lite
FCR 5000 plus/
Approx. 70 sec. Approx. 75 sec.
FCR 5000R plus
FCR 5501 plus/
Approx. 52 sec. Approx. 56 sec.
FCR XU-D1
FCR 5502D plus Approx. 60 sec. Approx. 64 sec.
FCR XG5000 Approx. 11sec. Approx. 36sec.
FCR VELOCITY U Approx. 2 sec. Approx. 11 sec.

5.3 Image Storing Images input with the Image are automatically stored in the hard disk
of the FCR CR Console Lite. Note that images are stored temporarily
in the hard disk, the images exceeding the number allowable for
storage will be erased consecutively from the older ones.

The following shows the numbers of images that the FCR CR


Console Lite can store in the hard disk:

Number of images: Approximately 2000 images


Notes: • The number of images stored is calculated based on the
assumption that high-density 14"×17", 14"×14", 10"×12" and
8"×10" images are registered with a ratio of 1:4:4:1.
• For ES images, the number of images that can be stored will be
smaller than 2000 images.

8 010-017-40 06.2003
CR Console Lite Product Specifications

5.4 Image Output Hard copies can be obtained via connection to the Image Recorder.
In a configuration where the HQ is effective, one of any of the
following image reduction ratios and formats can be selected by
specifying the pixel density for each menu at the
FCR CR Console Lite.
z For standard pixel-density images
Reduction Ratio Two-image Output /
One-image Output
Reading Size Four-image Output
17” × 17” 40% 58%
14" × 17" (35 × 43cm) 50% 67%
14" × 14" (35 × 35cm) 50% 67%
10"× 12" 67% 100%
8" × 10" 86% 100%
24 × 30cm 67% 100%
18 × 24cm 86% 100%
18 × 43cm 67% 67%
24 × 30cmHR 67% 100%
18 × 24cmHR 100% / 86%* 100%
* Alternately changeable.

(a) One-Image Output


i) All Image Outputs ii) All Image Outputs iii) 18×43cm Image Output

Image ID Image ID information


information

(20×25cm film)

(26×36cm film) (26x36cm film)

(b) Two-Image Output


i) L/R Image Output ii) 2on1(A/B) Image Output iii) 18×43cm Image Output

Image Image
(26×36cm film) (26×36cm film)

* Images A and B should belong (26x36cm film)


to the same study.

(c) Four-Image Output


Image

ID ID
information information

* Images A, B, C and D should


Image belong to the same study.
(35×43cm film)

010-017-40 06.2003 9
CR Console Lite Product Specifications

z For high pixel-density images


Reduction Ratio Two-image Output/
One-image Output
Reading Size Four-image Output
17” × 17” 40% 80%
14" × 17" (35 × 43cm) 50% 100%
14" × 14" (35 × 35cm) 50% 100%
10"× 12" 67% 100%
8" × 10" 86% 100%
24 × 30cm 67% 100%
18 × 24cm 86% 100%
18 × 43cm 100% 100%
24 × 30cmHR 67% 100%
18 × 24cmHR 100% / 86%* 100%
* Alternately changeable.

For one-image output using 14" × 17" or 14" × 14" IPs and/or
four-image output using all size's IPs, images are output on 14"
× 17" size film. In other cases, images are output on 26 × 36cm
size film.

(a) One-Image Output


i) 17"×17" or 14"×17" ii) 14"×14" Image Output

ID information

Image

Image
ID
information
(35×43cm film) (35×43cm film)

iii) 8"×10"(ST) or 18×24cm(ST) iv) 18×43cm Image Output


One-image Output

ID
Image ID information
information Image

(20×25cm film)

(35×43cm film)

v) Image Output other than vi) 18×43cm Image Output


17"×17", 14"×17", 14"×14"
or 18×43cm

ID
information

ID Image
Image
information

(26×36cm film) (14"×17" film)

10 010-017-40 06.2003
CR Console Lite Product Specifications

(b) Two-Image Output


i) L/R Image Output ii) 2on1(A/B) Image Output

ID ID
information information

Image Image
(26×36cm film) (26×36cm film)
* Images A and B should belong
iii) L/R Image Output to the same study.

ID

(14"×17" film)

(c) Four-Image Output


Image

ID ID
information information

* Images A, B, C and D should


Image belong to the same study.
(35×43cm film)

(d) 17”×14” one-image output

ID information

Image

(14"×17" film)

The following output formats are available depending on the


connected printer.

• 14”×14” 100%-size image output on a 14”×14” film


• 8”×10”/18×24cm 100-size image output on an 8”×10” film and
reduced image output on films of other sizes
• 10”×12”/24×30cm 100%-size image output on a 10”×12” film

010-017-40 06.2003 11
CR Console Lite Product Specifications

5.5 Image Processing Spatial Frequency processing, Tonal Conversion processing,


Dynamic Range Control (DRC) processing,
Tomographic Artifacts Suppression (TAS) processing,
Flexible Noise Control (FNC ) processing,
Multi-objective Frequency processing (MFP option),
Energy Subtraction processing (ES option),
Grid Pattern Removal processing (GPR option)

5.6 Connection • Direct connection


LP connection ..................................... LP interface
• Network connection
CR-IR / DICOM / DMS
Network connection ............................ 100BASE-TX/10BASE-T

12 010-017-40 06.2003
CR Console Lite Product Specifications

6. Functions
6.1 Standard Functions (1) Patient ID information input function
Patient ID information can be input from the keyboard.

(2) Patient ID formation local database function


This function stores input patient ID information (a maximum of
50,000 items) on the hard disk, making search for the information
thus stored possible thereafter according to the patient ID or patient
name.

Note: However, this function cannot be used shared between


FCR CR Console Lite Units. (Note also that this function is not
available for online connection.)

(3) Parameters setting function


• Exposure menu/study menu
This function selects and sets exposure menu/study menu, such
as a set menu consisting of chest frontal and lateral images, which
determines image processing parameters to be used on the Image
Reader.
Exposure menu/study menu can be selected by selecting an
exposure menu/study menu after an exposure region has been
selected.
The Auto Repeat function enables IP registration/built-in exposure
processing repeatedly for the same exposure menu.
• Film output
Specifies film format of an image and the number of printouts
required.
• EDR mode
Specifies image reading mode (AUTO/SEMI/FIX/MANUAL) for
each menu.
• Image reversal
Specifies up-side-down or right-and-left reversal of an image when
it is printed on film or displayed on the monitor.
• Film mark (The film mark cannot be positioned within the effective
image area.)
Specifies film mark (character strings) to be displayed on film.
• Image density adjustment
Alternately changes image density preset for each menu between
high quality and standard.
• Radiographer name entry
Displays a selected radiographer name on film.

(4) IP number registration function


Using the barcode reader that is an optional item, the IP number can
be read and registered, allowing thus patient ID relevant to the IP or
parameters to be notified to the Image Reader.
(Note that this function is not available when a built-in Image Reader
unit is connected.)

010-017-40 06.2003 13
CR Console Lite Product Specifications

(5) Image display function


Images sent from the Image Reader to the FCR CR Console Lite will
be listed on the study list window.

(6) Study pre-registration function


If contents of a study that include patient information, exposure
menu, etc. are available ahead of time, such study contents can be
registered together.

(7) Simplified QA function


This function subjects a study image that has been selected and
displayed to quality assurance and optimization processing
(changing various information items).
Information items to be changed can be replaced.
• Changing image processing information
- Exposure menus:
If an exposure menu is changed, image processing (e.g.
standardization processing, gradation processing, spatial
frequency processing, etc.) parameter values are changed to
those that have been preset for each exposure menu.
- Density/contrast:
GA (contrast) and GS (density) are changed.
- Standardization:
S value (sensitivity) and L value (latitude) are changed.
- Rotation/reversal:
Images are rotated or reversed by 90°.
- Film mark:
The film mark replaces the lead characters.
Note that if the film mark is imbedded, artifacts may appear
around it as a result of frequency processing.
• Changing image information
- Patient information:
Patient name, patient ID No., sex and date of birth are
changed.
- Study information:
The study ID is changed.
- Series information:
The anatomical region name is changed.
- Image information:
Exposure menu name, requesting department, study date
and order of images are changed.

• Changing image distribution information


- Target device for image distribution:
Device targeted for image distribution is changed.
A maximum of eight destination devices can be specified for
one image.
Note that the output capacity varies depending on the
number of devices specified for image distribution.

14 010-017-40 06.2003
CR Console Lite Product Specifications

- Pixel density conversion:


Depending on the device targeted for image destination,
images can be changed from high-pixel density to standard
pixel density.
Note: However, high-pixel density images cannot be distributed to
a device that is unable to handle such images. It is also not
possible to convert standard pixel density images to
high-pixel density images.

• Changing the film format


The film output format specified on the FCR Image Reader is
changed.
• Applying/releasing image write protection
To prevent image data from being deleted, the write protection can
be applied or released accordingly.
This write protection capability is applied for each study unit.

(8) Study information sharing function


This function allows you to share study information between among
multiple CR Console Lite units, making the following workflow
possible.
• Centralized study reservation using the keyboard/MCR.
• Moving a study from one X-ray room to another.
• Integrated output management.

(9) Simplified image (JPEG) storage function


This function records onto an FD the JPEG-formatted data of the
image that is now being displayed, which can be used for reference
purposes.

(10) Comment entry function


This function enters comments, such as precautions, for each study
or image.

(11) Image move function


A specified image can be moved into a different study. When an
image is moved, information incidental to such image will be
rewritten to that of the study to which the image was moved.

(12) Additional exposures specified to an already completed study


Additional exposures can be specified to a study that has been
completed.
Those additional exposures will be regarded included in the already
completed study, being thus handled as the one same study.

(13) Optimized image displayed on the monitor


An image optimized taking into consideration the gradation
characteristics can be displayed on the monitor.

(14) Anatomical region name setup function


An anatomical region name can be set up for the DICOM CR Storage
function, according to the MPM code.

010-017-40 06.2003 15
CR Console Lite Product Specifications

6.2 Optional Functions (1) Detailed QA function


On the QA screen, Spatial Frequency processing, Tonal Conversion
processing, Dynamic Range Control processing, Tomographic
Artifact Suppression processing or Multi-objective Frequency
processing or Pattern Enhancement processing for Mammography
can be selected.

(2) HIS/RIS connection function


• Connecting to the HIS/RIS, patient attribute information and
exposure-related information can be communicated with each
other. (DICOM MWLM, MPPS)
• Connecting to the HIS/RIS, patient attribute information can be
communicated with each other. (Patient’s information on-line)

To enable this connection, optional software is required for each


CR Console Lite unit.

(3) Film output function


? Image recorder connection function
Image data signals can be transmitted directly to the image
recorder using the serial interface.
@ DICOM printer connection function
Image data signals can be transmitted to the image recorder via
networks according to the DICOM PRINT function.

(4) DICOM network connection function


Image data signals can be transmitted to the DICOM archiver as
DICOM storage or storage commitment.

(5) DMS network


Image data signals can be transmitted to the FCR Data Management
System.

(6) Electronic shutter


Areas other than images can be painted black.

(7) X-ray control unit connecting online to another X-ray control unit
Voltage and current values preset for each menu will be transferred
to the X-ray control unit.

(8) Image composition function


Two or three long-view images can be composed together.
Auto and Manual modes are available for image composition
processing.

(9) Image magnification/full screen display


This function item magnifies a specified position of an image or an
entire image and displays it on the monitor.
However, the result of such image magnification will not be reflected
on output film or transferred image.

(10) LUT adjustment function


Log linear-adjusted images can be displayed for each CRT.

16 010-017-40 06.2003
CR Console Lite Product Specifications

7. Principles of Operation
7.1 Patient Reception Patient ID information is entered into hospital HIS/RIS.
Without HIS/RIS, patient ID information can lie, requisition issued onto
magnetic ID card. Or patient ID information can be entered on the
keyboard.

7.2 X-ray Exposure The IP cassette containing an erased IP is set in the X-ray exposure unit
to make an exposure.

7.3 ID and Exposure FCR CR Console Lite receives patient ID information from HIS/RIS, or
Condition Entries reads the magnetic ID card so that the patient’s ID information is input.
The exposure conditions such as anatomical region are sent from
HIS/RIS or selected on the FCR CR Console Lite.

7.4 Image Reading – The cassette containing the exposed IP is inserted into the cassette
Output setting unit at the Image Reader.

After the IP has been read by the Image Reader, the image data will then
be subjected to final image processing so that it is displayed on the FCR
CR Console Lite.

After confirming the transmitted image data at the FCR CR Console Lite,
the image will be handled as follows.

? Image data will be sent to the image recorder as image signals.


Modulated laser beams will then be used to print these signals on
film.
@ Image data will be sent to the image storage device or image
displaying device.

010-017-40 06.2003 17
CR Console Lite Product Specifications

8. Power Supply Conditions


Power voltage Current Notes
AC100V to 120V or
Main unit 2.0A or 1.0A Supplied from the AC outlet.
AC180V to 265V
Touch panel AC100V to 120V or
0.6A or 0.33A Supplied from the AC outlet.
LCD display AC200V to 240V
AC100V to 120V or
CRT display 1.7A or 0.9A Supplied from the AC outlet.
AC200V to 240V
High-resolution AC100V to 120V or
1.6A or 0.8A Supplied from the AC outlet.
CRT display AC200V to 240V
Specifications may be changed without prior notice.

9. Environmental Conditions

Humidity Maximum heat


Condition Temperature
(With no dew condensation) value
When in
10°C – 35°C 20% – 80%RH 110Wh
Controller operation
main unit When not in
-40°C – 65°C 20% – 80%RH 110Wh
operation
When in
5°C – 40°C 10% – 80%RH 32Wh
Touch panel operation
LCD display When not in
-40°C – 60°C 5% – 95%RH 32Wh
operation
When in
5°C – 40°C 10% – 80%RH 120Wh
operation
CRT display
When not in
-40°C – 60°C 5% – 95%RH 120Wh
operation
When in
0°C – 40°C 20% – 80%RH 150Wh
High-resolution operation
CRT display When not in
-20°C – 60°C 5% – 90%RH 150Wh
operation
Specifications may be changed without prior notice.

18 010-017-40 06.2003
CR Console Lite Product Specifications

10. External Dimensions and Weight


Classification Width Depth Height Weight
Main unit Approx. 390mm Approx. 435mm Approx. 110mm Approx. 10kg
Touch panel LCD display Approx. 400mm Approx. 210mm Approx. 410mm Approx. 8kg
CRT display Approx. 400mm Approx. 430mm Approx. 450mm Approx. 20kg
High-resolution CRT display Approx. 480mm Approx. 440mm Approx. 470mm Approx. 29kg
Keyboard Approx. 460mm Approx. 170mm Approx. 30mm Approx. 1kg
Hardware specifications, external dimensions and weight are subject to change without prior notice.

? PC main unit and keyboard (DELL GX50)


Unit : mm
390 435

460 170

@ Touch panel LCD display


Unit : mm
385

175

010-017-40 06.2003 19
CR Console Lite Product Specifications

A Display (CRT)
Unit : mm
400 430

B High-resolution Display (CRT)


Unit : mm
480 440

20 010-017-40 06.2003