Professional Documents
Culture Documents
/Spirit/Scope
Document Version
Siemens Healthcare GmbH reserves the right to change its products and services at any
time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens Healthineers office to order a current version or refer to
our website http://www.siemens-healthineers.com.
Disclaimer
Siemens Healthcare GmbH provides this documentation “as is” without the assumption of
any liability under any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.
Copyright
“© Siemens Healthineers, 2015” refers to the copyright of a Siemens Healthineers entity
such as:
Siemens Healthcare GmbH - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany
Sending film sizes that are not available at the customer site may cause problems at the
camera site. For this reason, it is recommended that you delete these film sizes for por‐
trait and landscape formats.
The Grayscale Standard Display Function (GSDF) is implemented for printing and imag‐
ing. No Presentation LUTs are available. For this reason, the DICOM Presentation LUT is
not released.
The use of the presentation LUT has to be disabled by local service for configuring LUT
files.
The density values for the image quality were measured with a calibrated densitometer
X-Rite 301. The result of the measurement from other densitometers used may differ
from the absolute density values as described in the tables.
Print the 17 graystep test image to check the modality and camera adjustment according
to the description. Verify the results using the reference values for max. density 2.80 lis‐
ted in the table. This is also part of the acceptance test.
With syngo versions VC40A and higher, the connection mode for DICOM print needs to
be changed to “unsecure” because “connection mode - secure” is not supported by all
hardcopy cameras.
1.1.2.1 Modality/system
1.1.3 Modality
1.1.3.2 Parameters
For software configuration, HCC data/parameters and possibly additional film matrix
properties (syngo: filming properties) are required for software configuration.
Application Entity
Display Supplier
Properties
AE Title Network administrator or
K8300
HCC service
Port number 5040 HCC Service
HC Device /
Display Supplier
General settings
Type DICOM printer Default entry
Class (printer spec file
kodak8300 Default entry
name)
DICOM Node dv8300 Default entry
Filming properties are "pixel size" and "media sheet size"; this information is required for
image scaling and cannot be changed.
The shown Software configuration is related to the latest SW-Version released. Any set‐
ting of the previous version may differ.
floppy or CD, these files can be selected on the source drive and copied to the depend‐
ent LUT directories on the local host. Unnecessary LUT files can be removed from the
LUT directories on the local host.
Only the topic CT image and SC image are relevant. Depending on the camera, specific
entries have to be made. If nothing else is mentioned, set the CT image and SC image to
"CTLUT" and "2.80" for maximum density. The entries for BL image and SC image have to
be identical.
Since these two checks are equivalent; it is sufficient to check the “specified density”.
Checking the “HCC linearity” may be necessary if the desired density verification (1)
shows deviations and the HCC setting needs to be changed.
Even if the HCC service does not require a test image of the modality for the HCC setting,
it is easier to check this setting - and possibly further HCC corrections - with the HCC line‐
arity verification (2) than with the desired density verification (1). Only the HCC linearity
LUT is effective. A graphical display would now show the current HCC behavior only. This
could be helpful for HCC Service.
The density difference of the BL LUT (Ref. LUT) between steps 1, 2, and 3 has to be a min‐
imum density value of 0.02. Test image 3 (SMPTE image) can be used for visual verifica‐
tion (the 95% gray square has to be visible).
1 2
3 4
5 6
Correction LUT LUT files with Dmin. 0.2 and Dmax. 2.80
Linear Linear LUT
CT LUT Standard LUT; default for all cameras; adapts Dmin. of 0.2
Dark area = less contrast; darker; softer; white area = more con‐
CT02 vs. CT LUT
trast
Dark area = less contrast; darker; softer; white area = more con‐
CT03 vs. CT02
trast
Dark area = more contrast; darker; sharper; white area: no
CT04 vs. CT LUT
changes
Brighter; no changes in contrast for large window width; more
CT05 vs. CT LUT
contrast for head images (small window width)
CT06 vs. CT LUT Brighter than CT5; white area = less contrast
cthead1 LUT for head images, images smoother
cthead2 vs. cthead1 Images smoother
Adapts Dmin of 0.15 (for cameras which are not supported Dmin.
CT015
command)
Adapts Dmin. of 0.17 (for cameras which are not supported Dmin.
CT017
command)
Correction LUT LUT files with Dmin. 0.2 and Dmax. 2.80
LUT for dry cameras; adapts Dmin. of 0.22 (5% visible in SMPTE
CT LUTdry
test pattern)
CT Lung LUT for lung images
1. Load the test image (water image phantom) into the viewer (e.g. 20 cm water phan‐
tom, generated with the constancy procedure by CT).
2. Insert Crosshair (Viewing Card: Tools) and copy the image to the film sheet (“copy im‐
age to film sheet” button).
3. Select film sheet format 1 X 1.
4. Select "Original Image" in the filming card and print the image.
Fig. 3:
Associations
Maximum number of associations supported: 10
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Time 82 sec.
2.1.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Two edges of the diagnostic area are cut (approx 1 mm) for 8x10, 10x12, 11x14,14x17.
Typical slight line structure of the thermo head is visible in technical image. No further
line structure visible.
Application Entity
Display Changes allowed
Properties
AE Title AXYS Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOMprinter No
Class
agfaDS5503_AXYS No
(Printer spec file name)
DICOM Node DSAXYS Yes
2.1.3.1 Identification
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
WARNING, COVER OPEN Cover is open Camera cover, drawer or
door open
WARNING, BAD SUPPLY MGZ Tray empty Problem with supply maga‐
zine
WARNING, FILM JAM Film jam Film jam
n.a. Configuration mode via IE Camera offline
Deviation ob‐
Film size / Direction Desired (mm) Measured tained from de‐
(mm) sired value in %
Fig. 4:
Associations
Maximum number of associations supported: 10
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Time 68 sec.
2.2.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
All four edges of the diagnostic area are cut (approx 1 mm) for 8x10, 10x12,
11x14,14x17.
Application Entity
Display Changes allowed
Properties
AE Title DRY5503 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOMprinter No
Class
agfaDS5503_AXYS No
(Printer spec file name)
DICOM Node DS5503 Yes
2.2.3.1 Identification
No Polarity Normal
No Film destination output tray n.a.
n.a. Tray 1 Selected
n.a. Tray 2 Deselected
n.a. Tray 3 Deselected
n.a. Tray 4 Deselected
No Film orientation Portrait
No Trim No
No Film Size ID (empty)
No Medium type (empty)
n.a. Print even if Film Size ID / Medium No
Type is not supported
n.a. Print even if Application Type is not No
supported
No Initially allocated memory size per 100
print job (Mbytes):
No Kernel site (*) Interpolation type CubicHighRes; smoothFactor -0.7
No Magnification Magn. max
No Decimate / Crop behavior: disabled
No Pixel Size ratio 1.0 / 1.0
No Perception LUT Linear
n.a. Kanamori-like value 0
n.a. Custom Value (empty)
No Illumination (cd/m) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density 20
No Maximum density size (*) 320
Header annotation
No Contents None
n.a. Font site (*) Courier 12
Footer annotation
No Contents None
n.a. Font site Courier 12
n.a. Window (%)
n.a. Level (%) (empty)
n.a. Image orientation r0
n.a. Conformance level 1
n.a. N-event enabled Off
n.a. Mammo modality No
n.a. Mammo CWR Rotation No
n.a. Association time-out(s) 0
n.a. Image time-out(s) 0
n.a. ALLOW IMPLICIT VR LITTLE ENDIAN Selected
n.a. ALLOW EXPLICIT VR LITTLE ENDIAN Selected
n.a. ALLOW EXPLICIT BIG ENDIAN Selected
n.a. Resolution 508
n.a. Supported abstract syntax Presentation LUT SOP class
Print job SOP class
Printer SOP class
Basic annotation box SOP class
Verification service class
Basic film box SOP class
Basic film session SOP class
Basic color PM Meta SOP class
Print Queue Management SOP IN‐
STANCE
Basic Grey Image Box SOP Class
Basic Color Image Box SOP Class
Basic gray PM Meta SOP class
Print queue management SOP class
n.a. Early processing allowed Yes
n.a. LR like off
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
PRINTER DOWN Printer is down Connection to camera is not
available
BAD SUPPLY MGZ Supply magazine is open Problem with supply maga‐
zine
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 5:
Associations
Maximum number of associations supported: 10
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Time 70 sec.
2.3.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
All four edges of the diagnostic area are cut (approx 1 mm) for 8x10, 10x12,
11x14,14x17.
Application Entity
Display Changes allowed
Properties
AE Title DRY5500 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
agfadrystar5500 No
(Printer spec file name)
DICOM Node DS5500 Yes
2.3.3.1 Identification
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
PRINTER DOWN Camera is down Connection to camera s not
available
FILM JAM Film jam Film jam
COVER OPEN Cover is open Camera cover, drawer or
door open
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 6:
Associations
Maximum number of associations supported: 10
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Time 68 sec.
2.4.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Sporadic problems occurred with MDM calibration. The films had different LUTs after
MDM calibration and were out of tolerance. Difficulties to find a calibration which
matches for all film sizes together. After several retries this was successful. It is recom‐
mended not to use 14x17-film for MDM calibration, with 10x12 this was successful.
The first film after film calibration has a higher optical density and can be out of toler‐
ance.
Although the support for the Basic Color Print Management SOP Class is specified, color‐
printing is not useable.
Application Entity
Display Changes allowed
Properties
AE Title DRY5302 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
agfadrystar5302 No
(Printer spec. file name)
DICOM Node DS5302 Yes
2.4.3.1 Identification
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
WARNING, SUPPLY EMPTY Film tray empty Camera out of film
WARNING, COVER OPEN Cover was opened Camera cover, drawer or
door open
Deviation ob‐
Film size / Direction Desired (mm) Measured tained from de‐
(mm) sired value in %
Fig. 7:
Associations
Maximum number of associations supported: 10
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Time 70 sec.
2.5.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Application Entity
Display Changes allowed
Properties
AE Title DRY5301 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
HC Device /
Display Changes allowed
General settings
Class
agfadrystar53021 No
(Printer spec. file name)
DICOM Node DS5301 Yes
1. Hint: Spec file agfadrystar5302 is basic released also for DryStar 5301.
2.5.3.1 Identification
No Polarity Normal
No Film orientation Portrait
No Trim No
No Film Size ID (empty)
No Medium type (empty)
n.a. Print even if Film Size ID / Medium Type is No
not supported
n.a. Print even if Application Type is not suppor‐ No
ted:
No Initial allocated memory size per print job 100
(Mbytes):
No Kernel site (*) Interpolation type: cubicHighRes; smoothing fac‐
tor -0.7
No Magnification Max
No Decimate Crop Behaviour disabled
No Pixel Size Ratio: 1.0 / 1.0
Yes Perception LUT Linear
n.a. Kanamori-like value 0
n.a. Custom Value (empty)
No Illumination (cd/m2) 2000
No Reflected ambient light (cd/m2) 10
No Border density (OD x 100) Black
No Empty image density (OD x 100) Black
No Minimum density (empty)
No Maximum density Site 320
n.a. Header annotation
n.a. Contents none
n.a. Font site (*) Courier 12
n.a. Footer annotation
n.a. Contents none
n.a. Font site (*) Courier 12
n.a. Window (%) (empty)
n.a. Level (%) (empty)
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
WARNING, SUPPLY EMPTY Film tray empty Camera out of film
WARNING, COVER OPEN Cover was opened Camera cover, drawer or
door open
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 8:
Associations
Maximum number of associations supported: 11
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Time 83 sec.
2.6.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
The background command does not work properly. The background is always white.
Application Entity
Display Changes allowed
Properties
AE Title DRY5300 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOMprinter No
Class
agfadrystar5300 No
(Printer spec file name)
DICOM Node DS5300 Yes
2.6.3.1 Identification
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 9:
Associations
Maximum number of associations supported: More than 30
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Time 83 sec
3.1.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Film size 8” x 10” is not released for subdivision 1x1 with linear LUT.
Film size 8” x 10” is not released also for subdivisions smaller than 6 on 1 (4 on 1; 2 on 1;
etc) with CT LUT.
The background command does not work properly. The background is always white.
The film size 11INX14IN needs to be deleted from the Somaris service configuration.
Application Entity
Display Changes allowed
Properties
AE Title HORIZON Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
codonicshorizon No
(Printer spec file name)
DICOM Node Horizon Yes
3.1.3.1 Identification
Content:
DV FILM BLUE Gamma: 1.0
Contrast: 0
Dmax: 300
Dmin: 0
Film View: Front
Illumination: 2000
Reflected Light: 10
LUT: linear
Polarity: Positive
Background: Black
Scaling: Bilinear
Rotate: Auto
Image Warnings: On
Captions: Off
Border Fill: On
Crop Anchor: Center
Scale Factor: 0.0
Coverage: Standard
Content:
Step in Service Mode (with "<" and "Menu")
select Job Settings
Select: Name: PR_hostname
PR_hostname Media DV FILM BLUE 14X17
(AE title of mo‐ Receive Tray: Default
dality)
Priority: Default
Gamma: 1.0
Contrast: 0
Dmax: Default
Dmin: Default
Film View: Front
Illumination: Default
Reflected Light: Default
LUT: linear
Polarity: Positive
Saturation: Default
MCM: Default
Background: Black
Scaling: Bilinear
Rotate: Auto
Image Warnings: On
Captions: Off
Border Fill: Default
Crop Anchor: Center
Scale Factor: Default
Coverage: Standard
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 10:
Associations
Maximum number of associations supported: 9
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive im‐
age, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection was used (point-to-point).
4.1.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Very small line structur is visible in technical images. This is a general effect of thermal
print head.
Some pixels are cuted at right bottom edge of film size 10"x12" portrait.
The Linear LUT with Dmax 3,0 printed on 10"x12" film, was out of tolerance for Dmin,
only. The measured value 0,16 was just below 0,17 lower border and is a minor prob‐
lem, because for image quality reason it is even good.
The Presentation LUT with Dmax 3,0 printed on 10"x12" and 11"x14" film, was out of tol‐
erance for Dmin, only. The measured value 0,16 was just below 0,17 lower border and is
a minor problem, because for image quality reason it is even good.
The CTLUT with Dmax 2,8 printed on 8”x10”, 10"x12" and 11"x14" film, was out of toler‐
ance for the lowest Dmin only. The measured value 0,15 resp. 0,16 is less than the mi‐
nus tolerance (= 0,17), but it’s a minor problem, because for image quality reason it is
even good.
The AUTO calibration process failed. The internal densitometer must be calibrated with
the external densitometer values (24step image). After several calibration cycles the cali‐
bration have been passed.
If the printing queue of modality was stopped because of insufficient memory of cam‐
era, the printing could be continued by pressing the CONTINUE button of FILM TASK
STATUS window.
The measured density values have been converted via a conversion table as shown in
Drypix LITE service manual. This conversion table is not available in operation manual.
FUJI recommends the X-Rite 301 densitometer with or without filter. For both cases the
conversion tables are available in service manual.
Application Entity
Display Changes allowed
Properties
AE Title DRYLITE Yes
Port number 104 No
HC device/
Display Changes allowed
General Settings
Type DICOM printer No
Class
(printer-specific file fujidrypixLITE No
name)
DICOM node DPLITE Yes
4.1.3.1 Identification
Precede Picking Up 0: No
Use max./min. density 1: Modality
Extension format ID: (empty)
Use default annotation 0:No
Change base color (disable:110H[F]) 1: Enable
Define film size of 11” x 14’’ 1: 25.7 x 36.4 cm
Select DICOM configuration Magnify
AE Title PR_hostname
Smoothing type 1: Sharp
Default requested decimate / crop behavior 1: Decimate
Magnification type 1: CUBIC
Requested image size 0
Default magnification type 1: modality
Default smoothing type 1: modality
Default requested Image Size 0: Printer
Default decimate / crop behavior 0: Printer
Procedure 0: A-VR
Sharp edge detection 1: ON
Sharp detection level 600
Sharp sharpness 130
Sharp character color 0: white
Smooth edge detection 1: ON
Smooth detection level 800
Smooth sharpness 30
Smooth character color 0: white
Medium edge detection 1: ON
Medium detection level 600
Medium sharpness 80
Medium character color 0: white
Select DICOM configuration LUT Common
AE Title PR_Hostname
Default LUT 1
Illumination 2000
Ambient light 10
Configuration information (Default LUT#) 0:printer
Default illumination 1:modality
Default ambient light 1:modality
Procedure 0:SAR
Select DICOM configuration LUT1
AE Title PR_Hostname
Max. density 300
Min. density 20
r table number 17
Contrast 100
Number of tuning points 0
P1_Density - P5_Density 120
P1_Shift - P5_ Shift 0
P1_Contrast - P5_ Contrast 100
Select DICOM configuration Output Format
AE Title PR_Hostname
Film size 4:14IN X 17IN
Medium type 1: Blue
Film orientation 1:Portrait
Border density 280
Polarity 1:Normal
Trim 0:OFF
Trim width 1
Trim density 0
Number of copies 1
Print priority 1: HIGH
Default number of copies 1:modality
Default print priority 1:modality
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
BAD SUPPLY MGZ Camera in config. mode Printer offline
FILM JAM Film jam Film jam
Difference ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Difference ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Fig. 11:
Associations
Maximum number of associations supported: 10
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive im‐
age, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection was used (point-to-point).
4.2.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
If the associations exceed the maximum of 10, the printer will abort and not reject the
association.
The 100% modulation within pixel test image (full matrix without interpolation) is not
completely visible for all film sizes.
Some edges of the exposed area of film (within printed range/diagnostic area) of film
size 11INX14IN (portrait) and 14INX17IN (landscape) were cut off a little bit - because of
rounded corners of film.
The radiation distortion for Drypix Smart is class A. It can be installed only in industrial
and hospital areas.
If the printing queue of modality was stopped because of insufficient memory of cam‐
era, the printing could be continued by pressing the CONTINUE button of FILM TASK
STATUS window.
Film size 14INX14IN is not available for Fuji Europe, so this film size is not released.
Application Entity
Display Changes allowed
Properties
AE Title DPSMART Yes
Port number 104 No
HC device/
Display Changes allowed
General Settings
Type DICOM printer No
Class
(printer-specific file fujidrypixSMART No
name)
DICOM node DPSMART Yes
4.2.3.1 Identification
Illumination 2000
Ambient light 10
Configuration information (Default LUT#) 0:printer
Default illumination 1:modality
Default ambient light 1: Modality
Procedure 0:SAR
Select DICOM configuration LUT1
AE Title PR_Hostname
Max. density 300
Min. density 20
r table number 17
Contrast 100
Number of tuning points 0
Select DICOM configuration Output format
AE Title PR_Hostname
Film size 4:14IN X 17IN
Medium type 1: Blue
Film orientation 1:Portrait
Border density 280
Polarity 1:Normal
Trim 0:OFF
Trim width 1
Trim density 0
Number of copies 1
Print priority 1: HIGH
Resolution 0: Standard
Default number of copies 1:modality
Default print priority 1:modality
Default medium type 1:modality
Default film orientation 1:modality
Default film size 1:modality
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
WARNING; PRINTER OFFLINE Camera in config. mode Camera offline
WARNING, FILM JAM Film jam Film jam
Difference ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Difference ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Fig. 12:
Associations
Maximum number of associations supported: 10
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Time 85 sec.
4.3.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
The CE certificate does not declare EMC classification. Class A, hospital use only.
For all trays the automatic calibration was not successful. Use the internal printer densi‐
tometer calibration first and then the manual calibration mode of printer.
Magnification type "REPLICATE" and "BILINEAR" is not specified for this printer.
14IN x 14IN landscape is not specified by FUJI DCS, but with a special workaround for
syngo classic (implemented in printer specification file V1.1), 14INX14IN landscape
printing will be handled like printing with orientation portrait.
Film size 14IN x 14IN is not available for Fuji Europe, so this film size is not released.
Application Entity
Display Changes allowed
Properties
AE Title DRYPLUS Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
fujidrypixPLUS No
(Printer spec. file name)
DICOM node DPPLUS Yes
4.3.3.1 Identification
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
PRINTER OFFLINE Printer in config. mode Camera is offline
FILM JAM Film jam Film jam
SUPPLY EMPTY Supply empty Camera out of film
Deviation ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Fig. 13:
Associations
Maximum number of associations supported: 10
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive im‐
age, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point-to-point) was used.
4.4.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
If one film magazine shall be used with different film sizes, the FUJI technician must
modify the magazine. It is possible to order for each film size a separate magazin that
the customer can change.
If the printing queue of modality was stopped because of insufficient memory of cam‐
era, the printing could be continued by pressing the CONTINUE button of FILM TASK
STATUS window.
Film size 14IN X 14IN is not available for Fuji Europe, so this film size is not released.
Application Entity
Display Changes allowed
Properties
AE Title PRIMA Yes
Port number 104 No
HC Device /
Display Changes allowed
General Settings
Type DICOM printer No
Class
(printer-specific file fujidrypixPRIMA No
name)
DICOM node DPPRIMA Yes
4.4.3.1 Identification
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
WARNING, SUPPLY EMPTY Film tray empty Camera out of film
WARNING, COVER OPEN Cover was opened Camera cover, drawer or
door open.
n.a. Camera in config. mode Camera offline
Difference ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Difference ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Fig. 14:
Associations
Maximum number of associations supported: 10
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
Time 85 sec.
4.5.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Not all image pixels visible on film, because of rounded corners. This only affects a very
few pixel.
Magnification Types “BILINEAR” and “REPLICATE “ are not supported (this is specified), the
Status-Code 0x0116 should be switched off in Image-Mode modalities which use these
Magnification Types.
The Drypix 5000 (only 110V) is only introduced to the USA Market, the Drypix 7000 has
a power supply for 220V (Europe).
Film size 14IN x 14IN is not available for Fuji Europe, so this film size is not released.
Application Entity
Display Changes allowed
Properties
AE Title DRY7000 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
Fujidrypix7000 No
(Printer spec file name)
DICOM Node DP7000 Yes
4.5.3.1 Identification
Content Magnify:
Procedure 0:A-VR
Default Smoothing Type 1: SHARP
Adjustment Param. SHARP
Edge Detection 1:on
Detection level 600
Sharpness 130
Char. Mode 0:White
Adjustment param. MEDIUM
Edge Detection 1:On
Detection level 600
Sharpness 80
Char. Mode 0:White
Adjustment param. SMOOTH
Edge Detection 1:On
Detection level 800
Sharpness 30
Char. Mode 0:White
Default Requested Decimate / Crop 1:Decimate
Behavior:
Content LUT:
Procedure 0:SAR
Default LUT 1
Adjustment parameter -> LUT1
-> r table number 17
-> Max. density 300
-> Min. density 20
-> Contrast 100
-> Number of tuning points 0
Dmax 3.6 0:Standard
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
PRINTER OFFLINE Printer in config. mode Camera offline
COVER OPEN Cover opened Camera cover, drawer or
door open
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 15:
Associations
Maximum number of associations supported: 10
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
4.6.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
The Magnification Types “REPLICATE” and “BILINEAR” are not specified in the DICOM Con‐
formance Statement. (Important for image mode modalities). Modalities which send “BI‐
LINEAR” or “REPLICATE” get back a Status 0x0116 (Attribute out of Range). This can be
prevented with the printer setting “PROTOCOL -> Attribute Value Out of Range (0116H
Warning)” configured to “0:not indicate” (See 7.6). The Magnification Type “CUBIC” is
used as default.
10 associations are specified to work at one time. The 11th association is canceled with
an A-ABORT. An A-ASSOCIATE-RJ was expected due to DICOM.
Filmsize 8” x 10”, orientation landscape: some edges of the exposed area of film (within
printed range/diagnostic area) were cut off a little bit - because of rounded corners of
film.
14IN x 14IN LANDSCAPE is not specified in the DCS of the printer but it works if using the
same matrix as in PORTRAIT
Film size 14IN x 14IN is not available for Fuji Europe, so this film size is not released.
Application Entity
Display Changes allowed
Properties
AE Title DRY4000 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
fujidrypix4000 No
(Printer spec. file name)
DICOM Node DP4000 Yes
4.6.3.1 Identification
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
PRINTER OFFLINE Printer in config. mode Connection not available
FILM JAM Film jam Film jam
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 16:
Associations
Maximum number of associations supported: 12
Performance
Use medical image 4 of patient's "Reference Images", maximum film size, positive image,
4x5 format, Copies=1, default center/width, magnification type none.
Use optimum connection (point to point).
Time 66 sec.
5.1.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
The radiation distortion for DryView 6950 is class A. It can only be installed in industrial
and hospital areas.
Some pixels of serrated borderline are cut due to rounded corners of film. Image not
centered on film, maybe solved during next Carestream maintenance.
Application Entity
Display Changes allowed
Properties
AE Title DV6950 Yes
Port number 5040 No
HC device/
Display Changes allowed
General settings
Type DICOM printer No
Class
(Printer-specific file carestream_6950 No
name)
DICOM node DV6950 Yes
5.1.3.1 Identification
Setup
Content
tion,
DICOM SCP,
Communica‐
tions
Content
tion, Name: 6950
DICOM SCP, Manufacturer: Carestream Health
Destination Model: 6950
Media
(parameter identical for all types: Blue, Clear, Mammo BLue):
Media Type Page Ext. Image Ext. Img. Origin.
(x,y) (x,y) (x,y)
8x10 5332 x 6632 4924 x 6224 204,204
10x12 6632 x 7934 6224 x 7526 204,204
11x14 7282 x 9232 6874 x 8824 204,204
14x14 9232 x 9232 8824 x 8824 204,204
14x17 9232 x 11182 8824 x 10774 204,204
Page and Image Options:
Min Vert Sep=2 Min Horiz Sep=2
Max Band Page=10 Max Img Band=10
Pixel Pitch=25,59 Border DMax=4.15
Image Dmin=0.00 Image DMax=4.15
Lines Image Annot=1 Lines Page Annot=2
Image Annot Height=84 Page Annot Height=84
Max Magnif.Factor=50 Max Minif.Factor=0.1
Trim Width=2 Accept Annotations=Yes
Accept Color Image=Yes Accept BW Image=Yes
Mix Aspect Ratio Page=Yes Mix Aspect Ratio Row=Yes
Mix Image Size Std Fmt=Yes Mix Image Size Row=Yes
Mix Req.Size Page=Yes Minify Image To Fit=Yes
Scale Image To Fit=Yes Crop Image To Fit=Yes
Req.Size OK=Yes
Configuration Notes=empty
Content
tion,
- Negotiate PLUT=No - Negotiate Color=No
DICOM SCP,
- Smoothing=5 (No) - Interpolation=Replicate (No)
SCP Config.,
- TFT=LINODS.w87 (Always - Perception LUT Contrast=9 (Al‐
select specif‐ Apply) ways Apply)
ic modality
- Image Polarity=Normal (No) - Tonescaling Method = Dry(No)
- Image Dmin=0 (No) - Image Dmax=3.1 (No)
- Illumination=2000 (No) - Reflected Ambient Light=10 (No)
Film and Image Options:
- Film Size=14”x17” (No) - Sorter Bin =1 No
- Film Base= Blue Film (No) - Film Orientation=Portrait (No)
- Border Density=3.1 (No) - Trim=No (No)
- Method=Minify/Scale (No)
Other Options:
- Report warnings=Yes (Al‐ - send N-Events to SCU=No (Al‐
ways Apply) ways Apply)
- Capture Date=No - Capture Time=No
- Hosp.Name=No - Study Page Nbr=No
- Modality ID=No - Film Session Label Override=No
- Print Priority=Normal (No)
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
COVER OPEN Front door opened and a Camera cover, drawer or
print job was send. door open.
EMPTY(film size) BLUE/CLR/ Empty tray while printing EMPTY 14x14 BLUE
MAMMO
Difference ob‐
Film size/orientation Desired (mm) Measured tained from de‐
(mm) sired value in %
Fig. 17:
Associations
Maximum number of associations supported: 6
Performance
Use medical image 4 of the "Reference Images" patient, film size 14" x 17", positive im‐
age, 4 x 5 format, Copies=1, default center/width, magnification type replicate.
The optimum connection was used (point-to-point).
5.2.1.4 Restrictions
For restrictions not related to the HCC, see ( Restrictions and known effects / Page 8)
EMC classification EN55011, class A (only for use in hospitals and industrial areas).
Application entity
Display Changes allowed
Properties
AE title DV5950 Yes
Port number 5040 No
HC device/
Display Changes allowed
General settings
Type DICOM printer No
Class
(Printer-specific file carestream_5950 No
name)
DICOM node DV5950 Yes
5.2.3.1 Identification
Setup
Content
DICOM SCP,
Communica‐
tions:
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
WARNING, PRINTER OFFLINE Printer is offline Camera is not available
WARNING, COVER OPEN Cover was opened Camera cover, drawer or
door open.
Difference ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Fig. 18:
Associations
Maximum Number of Associations supported: 4
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive im‐
age, 4x5 format, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
5.3.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Application Entity
Display Changes allowed
Properties
AE Title DV5700 Yes
Port number 5040 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
(Printer-specific file carestream_5700 No
name)
DICOM node DV5700 Yes
5.3.3.1 Identification
Image Quality
- Smoothing Type/ 0 - NO
- Interpolation / CUBIC - NO
- TFT Set / LINODS.W87 - Always Apply
- Perception LUT Contrast / 11 - Always Apply
- Image Polarity / NORMAL - NO
- Image Dmin / 0 – NO
- Tonscaling method / Dry - NO
- Image Dmax / 3 - NO
- Illumination / 2000 - NO
- Reflected Ambient Light / 10 - NO
Image Resizing
- Method / MINIFY-Scale – NO
Content SCP Cofiguration, Other Options
Status Reporting
- Report Warnings / YES - Always Apply
- Send N-Events to SCU / NO - Always Apply
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
WARNING, PROC INIT Camera offline Camera not available
WARNING, COVER OPEN Cover opened Camera cover, drawer or
door open
Difference ob‐
Measured
Film size / direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 19:
Associations
Maximum number of associations supported: 12
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive im‐
age, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Time 80 sec.
5.4.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
Application Entity
Display Changes allowed
Properties
AE Title DV6850 Yes
Port number 5040 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
carestream_6850 No
(Printer spec file name)
DICOM node DV6850 Yes
5.4.3.1 Identification
Image Quality
- Tonescaling Method / Perception LUT - Always Apply
- TFT Set Name / LINODS.W87 - Always Apply
- Polarity / NORMAL - NO
- Interpolation / REPLICATE - Always Apply
- Density Min / 0.23 - NO
- Density Max / 3.10 - NO
- Smoothing / 5 - NO
- Perception LUT Contrast / 11 - Always Apply
- Curve Shape Contrast / 0 - NO
- Curve Shape Value / 429 - NO
- Wet Pivot Density / 1.2 - NO
- Illumination / 2000 - NO
- Reflected Ambient Light / 10 - NO
Content SCP Configuration, Film and Image Options
Image Resizing
- Method / MINIFY-Scale – NO
- Text Box Location / NONE – Always Apply
Status Reporting
- Send Warnings / YES - Always Apply
- Send N-Events to SCU / NO - Always Apply
Annotation Elements
- Print Date / NO - Always Apply
- Print Hospital Name / NO - Always Apply
- Print Modality ID / NO - Always Apply
- Print Time / NO - Always Apply
- Print Page Number / NO - Always Apply
- Common Text Override / NO - Always Apply
- Film Session Label / “empty”/ NO
Modality Identification
- Source ID / 3.2
- Modality ID / SOM56850124.16
- Modality Type / Other - NO
Modality Defaults
- Manufacturer / (empty)
- Type / (empty)
- Model / (empty)
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
EMPTY(film size) BLUE/CLR/ Empty tray while printing EMPTY <8x10> BLUE
MAMMO
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
PRINTER OFFLINE Camera offline Connection to camera is not
available
PRINTER INIT Camera initilze while boot‐ Processor initializing
ing
Delete restriction of basic release: The restriction of printer from basic release: “A supply
empty error message will not be send to modality as specified. Always NORMAL/NORMAL
is the N-GET response” is not vailid for SOM/5 VC50A because the N_GET response is
“EMPTY <8x10> BLUE” as specified.
Difference ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Difference ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 20:
Associations
Maximum number of associations supported: 12
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive im‐
age, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point-to-point) was used.
Time 94 sec.
5.5.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Use of the device in a hospital (industrial) area only because of EN55011, Class A.
Application entity
Display Changes Allowed
Properties
AE title DV5850 Yes
Port number 5040 No
HC device/
Display Changes allowed
General settings
Type DICOM printer No
Class
carestream_5850 No
(printer-spec. file name)
DICOM node DV5850 Yes
5.5.3.1 Identification
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
WARNING, PRINTER OFFLINE Camera offline Connection to camera not
available
WARNING, COVER OPEN Cover was opened Camera cover, drawer or
door open
Difference ob‐
Film size/orientation Desired (mm) Measured tained from de‐
(mm) sired value in %
Fig. 21:
A3 and A4 glossy paper is automatically cut from 297 mm x 75 m roll paper. The size will
be correct only if the roll paper is set to have a border of 3 mm.
Associations
Maximum number of associations supported: n.a.
Performance
Use medical image 4 of the "Reference Images" patient, film size A4, positive image, 4x5
format, Copies=1, default center/width, magnification type none.
The optimum connection was used (point-to-point).
5.6.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Too little contrast in white areas for glossy and matte paper
A3 and A4 glossy paper is automatically cut from 297 mm x 75 m roll paper with a bor‐
der set to 3 mm. The size will be correct only if the roll paper is set to have a border of 3
mm.
For pure color images or pure black and white images to be printed on a paper sheet,
the camera must be configured twice in the local service tool: One camera configuration
(8-bit grayscale) is needed for pure black and white images with color appearance, and a
second camera configuration (24-bit color) bit is needed for color images with color ap‐
pearance.
For color images, more contrast in black and white areas is attainable if the color bright‐
ness value is set to -10 in the DICOM server configuration of Carestream.
If you require the calcium scoring report (*.pdf file) to be printed, you will need to install
Windows driver "Chroma DICOM Printer Driver 64 bit2.0.0.23.exe" for chroma.
Additional information:
n Color displacements are visible as thin lines in 2 of the 42 primary color fields.
When checking the color bars, be aware that not all the fields can be distinguished.
n
n Glossy paper can attain a maximum density of 2.2; the maximum density of matte
paper is only 2.0.
Application entity
Display Changes allowed
Properties
AE title CHROMA Yes
Port number 5040 No
HC device/
Display Changes allowed
General settings
Type DICOM printer No
Class
carestream_Chro‐
(Printer-specific file No
ma_Paper
name)
DICOM node DVCHROMA Yes
Color appearance: = "8-bit grayscale" to print black and white images
Color appearance: = "24-bit color" to print color images
Known effect: For pure color images or pure black and white images to be printed on a
paper sheet, the camera must be configured twice in the local service tool: One camera
configuration (8-bit grayscale) is needed for pure black and white images with color ap‐
pearance, and a second camera configuration (24-bit color) bit is needed for color im‐
ages with color appearance.
5.6.3.1 Identification
(*) Color Brightness Value: For B&W images = 0, for color images = -10
It is also recommended to add some additional white paper below the printout to im‐
prove the results during the measurement.
Before testing the image quality, calibrate the densitometer. Afterward, check the white
field of the densitometer in the "paper" mode (dot area). Note the values for V.
The following table describes the measuring density values of the gray scale CT test im‐
age of reference paper that contains only CT images.
Difference ob‐
Measured
Paper size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Difference ob‐
Measured
Paper size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Fig. 23:
Associations
Maximum number of associations supported: 12
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive im‐
age, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point-to-point) was used.
Time 65 sec.
5.7.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
The serrated border line and some pixels of the diagnostic area are cut because of roun‐
ded corners and a very small border.
The standard EN55011 (EMC) is supported with Class A (Hospital use) only.
Application Entity
Display Changes allowed
Properties
AE Title DV6800 Yes
Port number 5040 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
carestream_6800 No
(Printer spec file name)
DICOM node DV6800 Yes
5.7.3.1 Identification
Image Quality
- Tonescaling Method / Perception LUT – Always Apply
- TFT Set Name / LINODS.W87 - Always Apply
- Polarity / NORMAL - NO
- Interpolation / REPLICATE – Always Apply
- Density Min / 0.20 - NO
- Density Max / 3.00 - NO
- Smoothing / 5 - NO
- Perception LUT Contrast / 10 - Always Apply
- Curve Shape Contrast / 0 - NO
- Curve Shape Value / 429 - NO
- Wet Pivot Density / 1.2 - NO
- Illumination / 2000 - NO
- Reflected Ambient Light / 10 - NO
Content SCP Configuration, Film and Image Options
Image Resizing
- Method / MINIFY-Scale – NO
- Text Box Location / NONE – Always Apply
Status Reporting
- Send Warnings / YES - Always Apply
- Send N-Events to SCU / NO(YES in case of N-Events) - Always Apply
Annotation Elements
- Print Date / NO - Always Apply
- Print Hospital Name / NO - Always Apply
- Print Modality ID / NO - Always Apply
- Print Time / NO - Always Apply
- Print Page Number / NO - Always Apply
- Common Text Override / NO - Always Apply
- Film Session Label / (emtpy) / NO
- Source ID / 3.2
- Modality ID / SOM56800124.16
- Modality Type / Other
- NO
Modality Defaults
- Manufacturer / (empty)
- Type / (empty)
- Model / (empty)
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
COVER OPEN Cover open Camera cover, drawer or
door open
FILM JAM Camera offline Camera not available
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
EMPTY <film size> BLUE Empty tray (no message occurred)
n.a. Camera initailze while boot‐ Processor initializing
ing
Difference ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Difference ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 24:
Associations
Maximum number of associations supported: 12
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive im‐
age, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
5.8.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Application Entity
Display Changes allowed
Properties
AE Title DV5800 Yes
Port number 5040 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
carestream_5800 No
(Printer spec file name)
DICOM node DV5800 Yes
5.8.3.1 Identification
Other Options:
- Warnings=Yes (Always Apply), N-Events=No (Always Apply)
- Capture Date=No (Always), Capture Time=No (Always)
- Hosp.Name=No (Always), Study Page No.=No (Always)
- Modality ID= No (Always), Film Session Label Override=No (Always)
- Print Priority=Normal (No)
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
COVER OPEN Cover opened Camera cover, drawer or
door open
FILM JAM Film jam Film jam
CALIBRATION ERR Calibration error Problem in calibration. Film
quality may not be optimal.
Difference ob‐
Film size / Direction Desired (mm) Measured tained from de‐
(mm) sired value in %
Fig. 25:
Associations
Maximum number of associations supported: 12
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
5.9.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
The minimum density command is not supported. The minimum density value is the val‐
ue of the fog on the film.
The round edge of the film sheet cuts the diagnostic area.
Application Entity
Display Changes allowed
Properties
AE Title DV89001 Yes
Port number 5040 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
HC Device /
Display Changes allowed
General settings
Class
kodak8900_smallB No
(Printer spec file name)
DICOM Node DV8900 Yes
5.9.3.1 Identification
Content
8900 setup: (Provided by Kodak)
Content
8900 setup: (Provided by Kodak)
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 26:
Associations
Maximum number of associations supported: 13
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
5.10.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
The background command does not work properly. Camera doesn't support background
white.
Application Entity
Display Changes allowed
Properties
AE Title DV87001 Yes
Port number 5040 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
HC Device /
Display Changes allowed
General settings
Class
(Printer-specific file imation8700 No
name)
DICOM Node DV8700 Yes
5.10.3.1 Identification
Content: Configuration
MIM 200 setup
Content: Configuration
General:
Log.Name=8700
General:
Database Name=8700, Model=8700, Memory=32
Media info.:
Media=35x43cm, 4096x5223, 5133, blue
Output parameters.:
Cmd=SuperSet, Baud=19200, Data=8,Stop=1, Parity=None
Output parameters.:
Pixel=12, Header/Line, Transfer=2000 KHz, Acq=60
Supported item:
Collation=On, Text=On, Tone Scaling=On, Rotation=On
Interpolation:
NoMag=On, Bilin=On, Repl=On,Cubic Spline=On
Format types:
Custom=On, Slide=On, Row Symmetric=On
Format types:
1, 2, 4, 6, 8, 9, 12, 15, 16, 20, 24, 30, 35=On
Content: Configuration
Formatting:
Aspect Ratio Mix on Page=On, Image Size Mix in Std Fmt=On
Formatting:
DICOM Annotation Box=On, Req.Size Scale=On
Formatting:
Req.Size Mix in Row=On, Trim=On
Formatting:
Aspect Ratio Mix in Row=On, Image Size Mix in Row=On
Formatting:
Req.Image Size=On, Req.Size Mix on Page=On
Formatting:
True Landscape=On
Formatting:
Max Img per Band=20, Max Band per Page=20, Max Img Col=0
Formatting:
Max Img Row=0, Min Horiz Sep=2, Min Vert Sep=2
Formatting:
Trim Width Enabled=2, Trim Width Disabled=0
Formatting:
Pixel Pitch=12.8205
Image processing:
Gray scale=On, Minimize=On, Max Image Dmax=3.1
M952 usage:
Min H/V Mag=1, Max H/V Mag=64, Min Dmax 952=1.7
Under Configure, SCP Services, select the specific modality, modify:
Content: Configuration
Image resizing
Resizing <M/S>=Minify/Scale, Always Apply=N
Relocate Text Box, Location = none
N-EVENT Reporting, Reporting=N, Always Apply=Y
Warning Status=Y
PLUT=N
Basic Color=N
Tone Scaling=PL (PerceptionLUT), Always Apply=Y
Curve Shape/TFT, TFT/ULUT= Wrkstn2a.w87, Always Apply=Y
Contrast (dry)=01, Always Apply=Y
Interp. /Smoothing, Smoothing Type=0, Always Apply=N
Image Dmin=0.00, Always Apply=N,
Image Dmax=3.00, Always Apply=N
Polarity=Normal, Always Apply=N
Border=3.00, always apply=N
Size=14"x17", Always Apply=N
Base=Blue Film, Always Apply=N
Priority=Normal, Always Apply=N
Page Annotation, Annotation Setup, Always Apply=N, Modality ID=Off, Hos‐
pital Name=Off, Capture Date=Off, Capture Time=Off, Page Number=Off
8700 Plus setup
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 27:
Associations
Maximum number of associations supported: 12
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
5.11.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
The minimum density command is not supported. The minimum density value is the val‐
ue of the fog on the film.
The European Standard EN55011 is supported only with Class A (hospital use).
Application Entity
Display Changes allowed
Properties
AE Title DV81501 Yes
Port number 5040 No
HC Device /
Display Changes allowed
General settings
Type DICOM Printer No
HC Device /
Display Changes allowed
General settings
Class
kodak8150 No
(Printer spec file name)
DICOM node DV8150 Yes
5.11.3.1 Identification
Deviation ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Fig. 28:
Associations
Maximum number of associations supported: 8
Performance
Use medical image 4 of the "Reference Images" patient, maximum film size, positive im‐
age, 4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Time 59 sec.
6.1.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
If a film is not inserted, but configured in camera server, the queue is stopped with a
camera software problem. This job must be deleted from the camera queue.
The DICOM server of the camera has to be configured with magnification type “Cubic7”.
(magnification type "NONE" is not supported by the camera
The EMC classification is not declared in CE 93/42/EEC. Therefore, this device can be in‐
stalled only in hospital areas.
Application entity
Display Changes allowed
Properties
AE title DPRO873 Yes
Port number 104 No
HC device/
Display Changes allowed
General settings
Type DICOM printer No
HC device/
Display Changes allowed
General settings
Class
konicaDryPro873_LB No
(Printer-spec. file name)
DICOM Node DPRO873 Yes
6.1.3.1 Identification
DICOM PRINT IN
N-event port no. 0
Empty notice LV Failure
Error level M/M error receive (dimmed)
Max PDU 64
Upper stamp use OFF
If a job is sent with a film size that is not inserted in the camera, the job aborts (response
of film box reports status 262) and must be deleted so that it is possible to continue with
filming.
Deviation ob‐
Film size/orientation Desired (mm) Measured tained from de‐
(mm) sired value in %
Fig. 29:
Associations
Maximum number of associations supported: 9
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Time 77 sec.
6.2.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Magnification "NONE" not supported therefore Konica Server needs a fixed setting:
Magnification "CUBIC" with smoothing type 7.
If “Camera cover open”, no message is displayed. If the camera is ready, jobs will be
queued but not printed.
If a film is not inserted, but configured in camera server, the queue is stopped with a
camera software problem. This job must be deleted from the camera queue.
Application Entity
Display Changes allowed
Properties
AE Title DPRO832 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
konicaDryPro832_largeB No
(Printer spec. file name)
DICOM Node Konica832 Yes
6.2.3.1 Identification
Configuration LUT:
select input channel in "LUT Setup Screen" window:
DICOM PRINT IN
N-Event Port No. 0
Empty Notice LV Standard (checkbox selected)
Error Level M/M Abort
Max PDU 64
Deviation ob‐
Film size / Direction Desired (mm) Measured tained from de‐
(mm) sired value in %
Fig. 30:
Associations
Maximum number of associations supported: 32
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
6.3.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
The minimum density command is not supported. The minimum density value is the val‐
ue of the fog on the film.
Application Entity
Display Changes allowed
Properties
AE Title DPRO793 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
konicaDryPro793_largeB No
(Printer spec. file name)
DICOM Node Konica793 Yes
6.3.3.1 Identification
14x17 Magazine EMPTY 14x17 BLUE EMPTY 14x17 BLUE 14x17 blue film supply
empty empty
Front door open CHECK PRINTER CHECK PRINTER Check camera
(*1) refer to all messages in the DCS of the hardcopy camera
(*2) refer to all messages in the DCS of the modality
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 31:
Associations
Maximum number of associations supported: 16
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
6.4.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Application Entity
Display Changes allowed
Properties
AE Title DP771 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
konicaDyPro771_largeB No
(Printer spec file name)
DICOM Node KonicaDP771 Yes
6.4.3.1 Identification
Exit
Select U90 - Password
Enter
INPUT PASSWORD
Enter
Select U80 Service Mode
Enter
Exit
Select S30 DICOM PRIORITY
Select CH3 "specific modality name"
Enter
Exit
Select S20 Print Cond.
Select CH3 "specific modality name"
Enter
S21 STAMP SELECT
Exit
Data was changed
Save the data? Yes
Enter
Exit
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
14" x 14" X No No No
Landscape Y No No No
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Fig. 32:
Associations
Maximum number of associations supported: 8
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
Time 87 sec.
7.1.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
Horizontal lines visible in the grey field image are due to the MTF pattern, with horizon‐
tal black and white lines.
Application entity
Display Changes allowed
Properties
AE title UPDF750 Yes
Port number 104 No
HC device/
Display Changes allowed
General settings
Type DICOM printer No
HC device/
Display Changes allowed
General settings
Class
SonyUPDF750 No
(printer-spec. file name)
DICOM Node UPDF750 Yes
7.1.3.1 Identification
Content AE Settings
LOAD UPDF750
DELETE AE Title
SAVE AE Title
Call Type: Called (selected)
Calling (deselected)
Partial Matching (deselected)
Always use Mammo Media (deselected)
Replace CUBIC with (empty) (deselected)
CUBIC_SH Parameter (empty) 0-100 (5)
CUBIC_VR Parameter (empty) 0-100 (70)
Content Basic Film Session SOP Class N-CREATE/N-SET Default
Medium Type: BLUE FILM
Number of Copies: 1
Content Warning
0116H Warning
Film Size ID (selected)
Medium Type (selected)
Number of Copies (selected)
Image Display Format (selected)
Annotation Display Format ID (selected)
Configuration Information (selected)
Image Position (selected) Bits Allocated (selected)
Bits Stored (selected)
Requested Decimate / Crop Behaviour (deselected)
B605H Warning
Max Density (selected)
Min Density (selected)
B604H/B609H Warning
B604H Warning (selected)
B609H Warning (selected)
Print Behavior for unsupported/unavailable Film Size ID/Medium Type
Supported Film Size ID/Medium Type
8x10 NORMAL (selected)
8x10 MAMMO (deselected)
10x12 NORMAL (selected)
10x12 MAMMO (deselected)
11x14 NORMAL (selected)
14x17 NORMAL (selected)
Accept unsupported/unavailable
Accept unsupported Film Size ID/Medium Type (deselected) Accept unavaila‐
ble Film Size ID/Medium Type (selected)
Process: QUEUE
Priorities:
Film Size ID or Medium Type FILM SIZE ID
Film Size ID BIGGER
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
WARNING / EMPTY 14x17 Empty film supply 14x17 blue film supply emp‐
BLUE ty
WARNING / NO SUPPLY MGZ no supply magazine inside Film supply not available
Deviation ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size/orientation Desired (mm) tained from de‐
(mm) sired value in %
Fig. 33:
Associations
Maximum number of associations supported: 8
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
format 4x5, Copies=1, default center/width, magnification type none.
The optimal connection (point to point) was used.
7.2.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
After sending a print job with a film size which is not inserted in the camera trays, no
warning message appears in printer status report. The print job is stored in camera and
will be printed after the right film size was inserted.
Horizontal line structure visible in technical gray field image, because of strong black and
white contrast changes printed with thermal print head.
For accurate measurement of UPT-51xBL films a X-Rite 301X densitometer with 301-34-
SO special conversion optics has to be used.
For magnification type cubic there is no value for smoothing type supported, printouts
with different values have the same impression - value 0 is working fine.
There are 3 gamma curves available for Siemens Healthcare: FLC300N, AX300N and
LIN_300N: The selection of the gamma curve is possible by setting the called AE title to
UPDF550_G6, UPDF550_G7 or UPDF550: Called AE UPDF550_G6 for FLC uses gamma ta‐
ble 6 with FLC300N LUT. Called AE UPDF550_G7 for AX uses gamma table 7 with
AX300N LUT. Called AE UPDF550 for linear uses gamma table 8 with LIN_300N LUT (de‐
fault).
The setting of sharpness as a gamma curve parameter is “OFF” by default, but it can be
increased up to 14 (edge enhancement).
Application Entity
Display Changes allowed
Properties
AE Title UPDF550 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
SonyUPDF550 No
(Printer spec file name)
DICOM Node UPDF550 Yes
7.2.3.1 Identification
SW Revision: MA APP:V1.52,
MA Loader: V1.04,MA Boot: V0.20,CT:1.10,ICTAG:
V1.00,DICOM 06.01
Physical Interface: RJ45, Ethernet 10/100 base T/Tx
Network protocol: TCP/IP
Image memory: 3.5 GB
Hard disk: None
No file available.
File name
Configuration via printers display and via IE browser.
Content
No file: Information>
Printer keypad settings: IP ADDRESS xxx.xxx.xxx.xx Network address
Press “Menu” TOTAL PRINTS ***
PRINT Quality>
Select Gamma
1: SONY
2-5: USER x
6: FLC300N > FLC LUT
7: AX300N > AX LUT
8: LIN_300N > Linear LUT till 3.0
9-20: USER xx >
SHARPNESS OFF
PATTERN PRINT
1: 14x17 [EXEC]
2: 11x14 [EXEC]
SELECT POINT 95% [25]
SELECT POINT 5% [75]
Save Gamma > [EXEC]
CALIBRATION>
Full Range CAL >
1: 14x17 [EXEC]
2: 11x14 [EXEC]
Auto Calibration [ON/off]
CAL Reset >
1: 14x17 [EXEC]
2: 11x14 [EXEC]
SPOT Measure > [EXEC]
CAL Print Check >
1: 14x17 [EXEC]
2: 11x14 [EXEC]
CAL Undo > [EXEC]
GAMMA SEL>
14x17 /LIN_300N/ /LIN_300N/:
11x14 /LIN_300N/ Default LUT till 3.0
10x12 /LIN_300N/
8x10 /LIN_300N/
PRINT OPERATION>
REPRINT [EXEC]
PRINT QTY 1to125
SYSTEM SETTINGS>
NETWORK SETTINGS>
IP ADDRESS: xxx.xxx.xxx.xx
SUBNET MASK: xxx.xxx.xxx.x
DEFAULT Gateway: xxx.xxx.xxx.xxx
DITHER [ON/off]
MIRROR [on/OFF]
L/R MARK [on/OFF]
STRICT SIZE [on/OFF]
FRAME TYPE [Hk/CAP2/c1/wd]
CAPTION > UPPER [ON/off]
LOWER [ON/off]
DEFAULT SETTING [EXEC]
HEAD offset ADJ [ON/off]
MEMORY STICK > load …
EDGE MODE [ON/off]
LINEAR 20 [on/OFF]
MAX DENSITY [320]
FILM EMPTY BUZZ. [on/OFF]
VERTICAL Setting [L/r] Side up
OUTPUT POSITION [front/NORMAL]
OUTPUT TRAY [ON/off]
DEFAULT SIZE 2 [8x10]
GAMMA ADJ System [NATIVE/compat]
MAINTENANCE > ...
TEST PRINT > ...
MANUAL MECHA > ...
ERROR INFO > ...
ERR No. …
ERR RESET [EXEC]
ADJUST > …
SELF CHECK > …
MEDIA (TAG) MENU > …
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
COVER OPEN Door opened and a print job Camera cover, drawer or
was send. door open.
FILM JAM Film jam Film jam
EMPTY 14x17 BLUE Empty supply (11x14 BLUE) 11x14 blue film supply emp‐
ty
Deviation ob‐
Film size / Direction Desired (mm) Measured tained from de‐
(mm) sired value in %
Fig. 35:
Associations
Maximum number of associations supported: 8
Performance
Use medical image 4 of patient "Reference Images", maximum film size, positive image,
4x5 format, Copies=1, default center/width, magnification type none.
The optimum connection (point to point) was used.
7.3.1.4 Restrictions
For restrictions independent of the HCC, refer to ( Restrictions and known ef‐
fects / Page 8).
For accurate measurement of UPT-517BL film a X-Rite 301X densitometer with 301-34-
SO special conversion optics has to be used.
The gamma curves fixed for Siemens Healthcare are FLC300H, AX_DI300G, SL32_14C.
Application Entity
Display Changes allowed
Properties
AE Title UPDF500 Yes
Port number 104 No
HC Device /
Display Changes allowed
General settings
Type DICOM printer No
Class
SonyUPDF500 No
(Printer spec file name)
DICOM Node UPDF500 Yes
7.3.3.1 Identification
Possible Printer error mes‐ Action to provide error Displayed message on mo‐
sages message of printer dality
COVER OPEN Cover was opened. Camera cover, drawer or
door open.
NO SUPPLY MGZ No supply magazine. film supply not available.
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
Deviation ob‐
Measured
Film size / Direction Desired (mm) tained from de‐
(mm) sired value in %
siemens-healthineers.com