System
DAC beginning with VA20
2 Document Version / Disclaimer / Copyright
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens office to order a current version or refer to our website
http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is” without the assumption of any liability under
any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.
Copyright
“© Siemens, 2012” refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Healthcare Private Ltd. - India
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany
1 Requirements 5
6 Documentation of DAC 29
2.1.1 Abbreviations
Fig. 1:
2.2 Preparation
To ensure smooth installation and configuration procedures, all data of the DICOM or
HIS/RIS node to be configured should be read out first:
n For FLUOROSPOT Compact based systems, start local service und select:
- FL-C Service > Configuration > DICOM Nodes Settings
n For Axiom Aristos systems, read out the data as follows:
- select > Options > Service > Local Service > Configuration > Network Nodes or
HIS/RIS
Make a note of the network addresses of the components involved:
- IP Address
- Port No.
- AE Title
If the DAC has to be deactivated, it is helpful to record the original configurations of the
nodes to be modified.
If the image system configuration is affected by any changes, the image system needs
to be rebooted, otherwise the configuration changes will not be accepted. Always re‐
boot the image system after the message “Reboot required” appears.
n Click on <Save> and <Exit> (on systems running with the VB... software, click on
<Next>).
The "Config Check" starts. When finished, the "Config Check" results are dis‐
played.
n Start the automatic configuration by clicking on <AutoConf>.
n Wait until the Auto Configuration is finished.
n Click on <Exit> to leave the Config Check menu (on systems running with the VB...
software, click on <Finish>).
If the image system configuration is affected by any changes, the image system needs to
be rebooted, otherwise the configuration changes will not be accepted. Always reboot
the image system after the message “You must reboot Image System” appears.
C:\dac
If yes, continue with the configuration on the next page.
C:\dac
If yes, continue with the configuration on the next page.
n Insert the data medium with the configuration file provided (XML file).
n Select <Install> to install the DAC filter (xml file).
An Open application is displayed.
Fig. 3:
Prior to installation, clarify which XML filter is needed to solve the specific problem at
this site.
n Browse to the medium where the DAC filter is located and select the filter type nee‐
ded and select <Open> (Note: only one type of filter can be configured).
The name of filter is displayed ( Fig. 5 Page 14) to the right of the “Install” button, a
check mark is displayed under “License” to show that the license is correct.
n Under “DAC host settings”, type in “26000” and under “Number of Associations” type
in “5”.
=> These values are the defaults. Check and make sure they are filled in.
n Select <Add node> to start the DICOM configuration.
Fig. 4:
n Make sure that the configured maximum PDU size is smaller than the send buffer size
and the receive buffer size.
n Confirm that the RUNNING service status is displayed and select <Cancel> to quit the
DAC configuration.
Close Explorer.
Fig. 7:
n Go to next page and select the Logical Name of the remote application.
n Select “localhost” under “Host”.
n Select the AE Title (here it is STORE_PACS); select localhost in Host and enter the Port
Number 26000 for this AE.
n Save the changes with <Save>.
Fig. 10:
Fig. 11:
n After confirmation of "dicom node properties successfully saved", select the Logical
Name and AE Title again.
n Use Verification to verify the configuration.
Fig. 12:
3.1.2.1 Checks
n Check the functionality of the network and HIS/RIS nodes.
Restore is necessary only after re-installing the system SW to the DAC SW.
C:\dac
If yes, continue with ”Restoring the DAC Settings”.
3.2.3.4 Checks
n Check the functionality of the network and HIS/RIS nodes.
Deactivation is necessary only if the DAC functionality is not to be used any longer and a
reversion to the original settings is requested.
Initial publication.
Fill out the table below. File the page in the system documentation together with the
DAC SW CDs and the license diskette: