Process Control and Automation For Flow Meter Calibration Rig
Process Control and Automation For Flow Meter Calibration Rig
0 Technology (I4Tech)
Vishwakarma Institute of Technology, Pune, India. Sep 23-24, 2022
Abstract—In flow meter calibration, measurements taken while mentation is required which includes the field instruments
2022 International Conference on Industry 4.0 Technology (I4Tech) | 978-1-6654-7196-1/22/$31.00 ©2022 IEEE | DOI: 10.1109/I4TECH55392.2022.9952957
it is in use are compared to measurements taken by a standard such as control valves, level switches, temperature sensor,
flow measuring device under the same circumstances. The flow pressure sensor, prover tank etc for the construction of a
meter scale is then modified until it nearly resembles the bench-
mark. The large volumetric flow calibration test rig is designed better flow calibration rig. Development and integration of
and developed in which the closed conduit flow meters like Elec- control strategies for PID control loop, Tower calibration and
tromagnetic / Ultrasonic are calibrated by flowing water through Flow meter calibration. To obtain calibration measurement
UUT (unit under test) with direct volume comparison with capability (CMC) of ± 0.2 percent or better, create a control
reference standard. The higher accuracy and lower uncertainty scheme that is controlled and monitored by PLC and SCADA.
is required because flow meter is used for commercial metering
purpose, so calibration process needs to be precise to get accurate
metering results. Conventional comparative methods will not give
better accuracy results, as meter readings are commutative and III. M ETHODOLOGY
possibility that error becomes twice. To overcome this problem
M/S SBEM employs a volumetric primary calibration method in The amount of liquid collected is measured volumetrically
accordance with ISO 8316, which reduces calibration uncertainty.
in the flow meter calibration setup. The standard vessel in the
Index Terms—Electromagnetic flow meter, Volumetric calibra- volumetric method is a container with a calibrated volume.
tion, PLC and SCADA, Process Control, Field Instruments. The method’s basic idea is to guide the water passing through
the flow meter into a tank with precise measurements. When
I. I NTRODUCTION the water is passing through the flow meter, the collected
Calibration is an essential aspect of Instrumentation in a volume is measured. The tower has 22 level switches, each of
broad range of industries that require high-accuracy measure- which location is precisely regulated, and the volume between
ments with a negligible percentage of error. Increased trust the switches is precalibrated using a Transfer Standard Meter
in the reading from the flow meter in use is the goal of of high precision. To reach the appropriate volume, switches
calibration. The viscosity of the fluid, installation effects from can be used in conjunction. Control valves over PID loops
bends and fittings, temperature, and pressure are factors that are used to identify and control the velocity. For adjusting
need to be taken into consideration. The standard should also flow rate, the PID loop makes advantage of the reference flow
have a defined traceability and uncertainty chosen to match meter’s 4-20 mA output. The specific level switch indicates the
the expectation of the final measurement. Standards may be level of the tower when the flow rate stabilises, and from that
static or dynamic, and this choice is made on the basis of the level switch the calibration cycle is initiated. The calibration
output and end use of the flow meter. Some flow meters are cycle is paused and the volume between the level switches is
used to measure quantity and have a fast response time; others measured once the flow rate starts to decrease at that specific
are designed to measure flow rate and have a slow response level switch. This cycle is repeated until the flow rate is stable.
time. Calibration of a flow meter should cover a significant The test rig is set up to calibrate lines with diameters ranging
flow-rate range for the flow meter and establish a performance from DN 350 to DN 1600 and flow rates ranging from 250
across that range. In this flow calibration rig Electromagnetic m3/hr to 4000 m3/hr. Sensors with sizes ranging from 350
flow meter are calibrated with very precise and accurate mm to 750 mm are mounted on Line B, while those with sizes
Instrumentation which is controlled and monitored by PLC ranging from 800 mm to 1600 mm are installed on Line A.
and SCADA to achieve CMC of ± 0.2 percent or better as per The flow laboratory consists of an underground sump of 120m³
the standard ISO 8316:1987 (Volumetric method). capacities to store the required water. The test rig is supplied
with water from Reference Tower with total volume capacity
II. S COPE AND O BJECTIVE 50m³ and useable calibration volume is (Tower 1= 5m³ to
A dedicated calibration and testing facility is necessary to 35m³). The uncertainty depends on volumetric tank calibration
enhance/improve calibration also selection of precise Instru- uncertainty, temperature uncertainty and timing uncertainty.
Air
LS105
Start
Release!+!#
LABOARTORY/153UFM03S
contamination. RV102
VolumePulse
150Liter End!"#$:0.000 0.000
LS115 Pulse!#
Reset
123456789 RS485Modbus 420mAControl# #
The water in the calibration tower and pipe work should
Top *" LS117
•
Pulse# PID
Switch
LS119 Stop 3Nos,#" Controller1
Counters !&#
Temp'! End!+!# System
TS102 Stop LS120 RS485Modbus PID
420mAInput
420
RV101 800to1600mm
1200Liter
Test Feedback( FCV500
FBM500
Temp'!
Line %Degree!+
350mm !750mm
FCV350
should be allowed to enter into the laboratory area to BFV53 Reference"!
1.002 BFV51
Drain"*
BFV101
Drain"*
BFV202 BFV201
BFV102
Volume
visitors. BFV52
Drain
2
Authorized licensed use limited to: Institut Teknologi Sepuluh Nopember. Downloaded on May 21,2025 at 04:00:10 UTC from IEEE Xplore. Restrictions apply.
cycle is over all the readings are saved in SCADA database Friction factor of straight line is 0.009
or MS-Office with secured path with limited access. Velocity for line A is,
VI. SYSTEM DESIGN FEASIBILITY 2 ∗ 9.81 ∗ 16.5
V = (11)
Design calculations for determining velocity by considering 1.5 + 4∗0.009∗22.5
0.5
the friction losses in the pipe with the help of Bernoulli’s
equation. V = 10.86m/s (12)
1 1 Velocity for line B is,
p1 + ρv12 + ρgh1 = p2 + ρv22 + ρgh2 (1)
2 2
2 ∗ 9.81 ∗ 16.5
Where, V = (13)
1.5 + 4∗0.009∗17
0.35
P is pressure exerted by fluid
V is velocity of fluid V = 9.98m/s (14)
ρisdensityof f luid Calculation of flow rate,
H is height of tower For line A is
For calculating the velocity, V= 10.86 m/s and D= 0.5m
p1 v1 π 2
H = Z1 + + (2) A= (d )
ρg 2g 4
(15)
p1 v2 p2 v2 3.14
Z1 + + 1 = Z2 + + 2 (3) A= ∗ 0.52 (16)
ρg 2g ρg 2g 4
Add friction loss at output side
A = 0.1962m2 (17)
p1 v12
p2 v22
Z1 + + = Z2 + + + h i + hf (4)
ρg 2g ρg 2g
Q=V ∗A (18)
We have to calculate velocity by considering friction loss
3
Authorized licensed use limited to: Institut Teknologi Sepuluh Nopember. Downloaded on May 21,2025 at 04:00:10 UTC from IEEE Xplore. Restrictions apply.
Convert into m³/hr C. PID Implementation
The feedback meter’s 4-20 mA output current is coupled to
Q = 0.9597 ∗ 3600 (29) a PID controller, and the real line flow rate is calculated using
this current value. In order to manage the flow and reach the
desired set point, the PID controller will continuously monitor
Q = 3454.92m3 /hr (30)
the actual line flow rate and produce a 4-20 mA control output.
The motorised valve positioner or VFD is also provided with
this controlled 4-20 mA output in order to manage the valve
position and maintain a constant flow rate.
4
Authorized licensed use limited to: Institut Teknologi Sepuluh Nopember. Downloaded on May 21,2025 at 04:00:10 UTC from IEEE Xplore. Restrictions apply.
system being utilized for automatic process control operations, XIII. DELIVERABLES
data communications and acquisition, mathematical calcula- • Consistently delivering quality results as per commitment
tions and report generations of flow meter calibration. and in conformance to the regulatory requirements
• Consistently comply with field specific requirements for
X. ADVANTAGES fluid flow discipline to deliver accurate and precise results
with enhanced uncertainty.
• For Fluid flow measuring devices with liquid volume
flow rate by using Electromagnetic Flow meter within
measurement range of 500m³/hr to 3500m³/hr with the
calibration measurement capability (CMC) of ± 0.2 per-
cent
XIV. ACKNOWLEDGEMENT
The author is thankful to M/s SBEM Flow Laboratory,
SBEM Pvt. Ltd, Pune for helping throughout the project and
providing this opportunity and facilities to carry out this work.
R EFERENCES
[1] Abbas, Ali, Nitin Kumar Jhankal, and Arun Kumar. ”Calibration of
Electromagnetic Flowmeter Using Advanced SCADA System.”
[2] Seierstad, Johanne. ”Design system for primary calibration of flow.”
Master’s thesis, Institutt for energi-og prosessteknikk, 2013.
[3] ISO 8316, Measurement of liquid flow in closed conduits - Method by
collection of the liquid in a Volumetric tank
XI. RESULT [4] IS0 5168, Measurement of fluid flow - Estimation of uncertainty of a
flow-rate measurement
XII. CONCLUSION
These unique capabilities are a requirement to carry out
the flow calibration procedure and are provided by the PLC
and SCADA system used for the automated operation of the
test field. The calibration report generated is as expected and
accuracy of flow meter is found within limits. The calibration
measurement capability (CMC) of ± 0.2 percent is found
within measurement range of 500m³/hr to 3500m³/hr.
5
Authorized licensed use limited to: Institut Teknologi Sepuluh Nopember. Downloaded on May 21,2025 at 04:00:10 UTC from IEEE Xplore. Restrictions apply.