You are on page 1of 12

Training Center

Module 10: Technical Safety Inspection


and Preventive Maintenance Part 4
Setting into Service and Documentation

Technical Support International


DialogP SW9.xx_std_Module 10_TSI and PM Part 4_Presentation
Training Center

Functional Electrical Safety Setting Into


Introduction Visual Inspection Documentation
Inspection Analysis Service

5.1 Applied Accessories/Disposables


 The tubing used during the Setting Into Service has to be documented in the Check List.
 A full AV tubing set with dialyzer shall be used to achieve realistic results.

5.2 Switch on Machine


 The machine has to be switched on and pass all steps of the self test.
 Start the therapy with the following settings:
 Time – 4h
 Ultra Filtration Volume – 2 l
 Conductivity – 14,3 mS/cm
 Temperature – 37°C

5.2.1 Self test passed


 The tubing used during the Setting Into Service has to be documented in the Check List.
The self-test will be explained in the next module
Training Center

Functional Electrical Safety Setting Into


Introduction Visual Inspection Documentation
Inspection Analysis Service

5.3 Power Fail Function

 The Power Fail Function has to be tested before therapy is selected.

 Switch on machine for approx. 5 minutes and then disconnect mains plug.

 5.3.1 The machine must give a buzz from the power supply for 1 minute.
Training Center

Functional Electrical Safety Setting Into


Introduction Visual Inspection Documentation
Inspection Analysis Service

5.4/5 Temperature and Conductivity


 An external reference meter for temperature and
conductivity has to be connected to the blue
dialyzer coupling.
 Enter the Therapy and run stable for 10 minutes.

 Make a comparison with the reference device


and the displayed values on the monitor:
 5.4.1 Comparison measurement {at dialyzer
coupling} at 37ºC (-1.5 ºC; +0.5 ºC)
 5.5.1 Comparison measurement {at dialyzer
coupling} at 14.3 mS/cm (±0,2 mS/cm)
Training Center

Functional Electrical Safety Setting Into


Introduction Visual Inspection Documentation
Inspection Analysis Service

5.6 Safety Air Detector (SAD)

 The Safety Air Detector has to be


tested with air.

 Lower the level in the Venous


Chamber (e.g. with level regulation
pump) until air reaches the sensor.

 5.6.1 The machine has to make a:


 visual alarm
 audible alarm

 The function in noted in the Check


List
Training Center

Functional Electrical Safety Setting Into


Introduction Visual Inspection Documentation
Inspection Analysis Service

5.7 Monitor

 5.7 /.1/.2 All the functions related to the screen


have to be tested:
 Function of the keys
 Display illumination
 OSDs
 Touch screen OK
 Image display OK

 The results are documented in the Check List


Training Center

Functional Electrical Safety Setting Into


Introduction Visual Inspection Documentation
Inspection Analysis Service

5.8 UF Comparison Measurement

 5.8.1 A UF comparison measurement is made with:


 15 min
 500 ml/h UF Rate
 125 ml ±15 ml
 The test has to be completed without any alarms
 The levels in the blood line chambers shall not be changed
 Air bubbles can compromise the result

 The procedure is described in the Service Manual

 All values are documented in the Check List


Training Center

Functional Electrical Safety Setting Into


Introduction Visual Inspection Documentation
Inspection Analysis Service

5.9 /.1 Battery Option

 Has to be tested if the option is present.

 The test is done in therapy mode.

 5.9.1 Disconnect mains plug, ensuring the monitor and blood pump are still working.

 Buzzer from SMPS gives three signaling tones.

 Machine is running in battery mode.


Training Center

Functional Electrical Safety Setting Into


Introduction Visual Inspection Documentation
Inspection Analysis Service

5.10 Disinfection

 The machine has to be disinfected after a TSI and PM.

 5.10.1 This can be performed by responsible organization {user}. In this case


the user has to be informed that the disinfection needs to be done.

 5.10.2 Alternatively the service technician has to perform a disinfection

 This is documented in the Checklist.


Training Center

Functional Electrical Safety Setting Into


Introduction Visual Inspection Documentation
Inspection Analysis Service

7.2 Applied Measurement Equipment

 All measuring equipment used in during the STI and PM have to be


documented in the Check List.

 Measurement equipment is subject to the calibration service.


Training Center

Functional Electrical Safety Setting Into


Introduction Visual Inspection Documentation
Inspection Analysis Service

Check Results
Training Center

End of presentation.
For exit please click
 x
at the right top corner of the window.