Professional Documents
Culture Documents
Supplier Customer
QC QC PDI
IGI QC
check check check
check
QC
check
Reference International standards
Example-
Example- Example- 1. ACSR106/25 SUURSAVO
1. ACSR34/6 SPARROW 1. ACSR89/52 DOTTEREL
2. ACSR54/9 RAVEN 2. ACSR42/25 SAVO
Cross sectional view and Configuration
Switch on the instrument by ON/OFF switch and allow 10 minutes warm up time.
For electrical connection, make all connections as per circuit diagram (Displayed on
workstation) Four terminal resistance measurement methods.
Example-
+I +V-------------------------- -V -I
1m
Take test piece of 1.5 meter long. Straight it.
Fix the Wire sample in terminal fixture/Hand. Before performing the test
ensure that the Wire attains room temp.
Take the reading on room temperature and convert to 20*c as per below
conversion factor:-
1 .
R20*c = Rt X
1 + (t)
Where -
R20* - Resistance at 20 deg C
Rt- Resistance at room temp.
α - temp. coefficient for aluminium - 0.004
α - temp.coefficient for Al alloy – 0.0036
Δt- (room temp.-20*c)
Testing Procedures