Professional Documents
Culture Documents
SESSION 10-11
SPECTROSCOPY
• The science that deals with light and its absorption and emission
by solutions and other material substances is called spectroscopy
or Spectrometry
• The broad term for the instrument used is spectrometer , while a slightly
more specific term (when a light sensor known as a phototube is used) is
spectrophotometer
BASIC PRINCIPLES OF SPECTROSCOPY
• The speed with which electromagnetic waves move is called the speed of
light and is given the symbol c. The speed of light in a vacuum is
approximately 3.00 ×10 10cm/sec
• The number of the moving electromagnetic waves that pass a fixed point in
1 sec is called the frequency of the light. Its symbol is the lowercase Greek
letter nu, ν. It is expressed in waves (most often called cycles) per second,
or hertz (Hz)
• Mathematically, wavelength, speed, and frequency are related by the
formula
BASIC PRINCIPLES OF SPECTROSCOPY
BASIC PRINCIPLES OF SPECTROSCOPY
BASIC PRINCIPLES OF SPECTROSCOPY
• The movement of an
electron between electron
energy levels is called an
electronic energy ransition
• Atoms in which no
electrons are in the higher
levels are said to be in the
ground state (E0)
• Transmittance is symbolized as T.
BASIC PRINCIPLES OF SPECTROSCOPY
• Similarly, molecules and complex ions may also emit light under certain
conditions, a phenomenon known as fluorescence
BASIC PRINCIPLES OF SPECTROSCOPY
BASIC PRINCIPLES OF SPECTROSCOPY
BASIC PRINCIPLES OF SPECTROSCOPY
The equation of the straight-line A vs. C standard curve is known as the the
Beer-Lambert law, or simply as Beer’s law.
A statement of Beer’s law is
in which A : Absorbance;
a : Absorptivity, or extinction coefficient: the inherent ability
of a chemical species to absorb light and is constant at a given
wavelength, pathlength, and concentration
b : Pathlength : is the distance the light travels through the
measured solution. It is the inside diameter of the sample
container placed in the light path
c : Concentration.
• This is because of the high intensity of the xenon arc lamp. The
cuvette is held snugly in a spring-loaded holder.
ULTRAVIOLET–VISIBLE (UV-VIS)
2. Photodiodes
• A photodiode is a p–n jun ction constructed with the top p layer so thin
that it is transparent to light.
• Light shining through the p layer creates additional free electrons in the n
layer that can diffuse to the p layer, thus creating an electrical current that
depends on the intensity of the light.
7. IR spectra are characterized by rather sharp absorption bands and each such
band is characteristic of a particular covalent bond in the sample molecule.
Sampling
• Glass and plastic are undesirable materials for the cells because glass and
plastic are molecular (covalent) materials and would absorb IR light and
interfere with reading the sample.
• Inorganic salts, such as NaCl and KBr, are ionic materials and do not absorb
IR light because ionic bonds cannot undergo vibrational energy transitions.
Covalent bonds can vibrate while ionic bonds cannot.
IR SPECTROMETRY
Liquid Sampling
Pure liquids (often referred to as neat liquids) and liquid solutions, sandwiching a
thin layer of liquid between two large NaCl or KBr crystals (the windows) is the
classic procedure for mounting the sample in the path of the light
Sampling
IR SPECTROMETRY
Solid Sampling
The most straightforward method for analyzing a solid material by infrared
spectrometry is to dissolve it in a suitable solvent and then to measure this
solution using a liquid sampling cell
IR SPECTROMETRY
• absorption spectra for atoms are due to atoms in the gas phase
absorbing UV-VIS light from a light source and are characterized
by very narrow wavelength absorption bands called spectral
lines.
INTRODUCTION
6. Since the analytes for atomic spectroscopy are severely limited (elements
only), compared to the large number of molecular and complex ion
analytes for UV-VIS molecular absorption spectrometry the wavelengths
used for quantitation are well known and do not require the analyst to ever
first measure the absorption or emission spectra
INTRODUCTION
• For those atoms that are easily excited under these conditions, the emitted
wavelengths are in the visible region of the spectrum with characteristic
color because each element has its own characteristic line spectrum, the
atomic fingerprint, resulting from the particular energy transitions that
element has.
• All flames require both a fuel and an oxidant in order to exist. Bunsen
burners and Meker (Fisher) burners utilize natural gas for the fuel and air
for the oxidant.
Flame Atomic Absorption
• The light sources used in atomic absorption instruments are sources that
emit spectral lines. Specifically, the spectral lines used are the lines in the
line spectrum of the analyte being measured.
The internal excitation and emission process occurs inside this cup when the
lamp is on and the anode (positive electrode) and cathode are connected
to a high voltage
Flame Atomic Absorption
3. Premix Burner
• The burner used for flame AA is a premix burner because all the
components of the flame (fuel, oxidant, and sample solution) are
premixed, as they take a common path to the flame
3. Premix Burner
Flame Atomic Absorption
4. Optical Path
The optical path for flame AA is arranged in this order:
1. light source
2. Flame (sample container)
3. Monochromator
4. Detector
4. Optical Path
Graphite Furnace Atomic
Absorption
• Unlike flame AA and graphite furnace AA, the ICP technique measures the
emissions from an atomization/ionization/excitation source rather than the
absorption of a light beam passing through an atomizer
• ICP is the most important technique because the excitation source is much
hotter, resulting not only in atomizaton, but also in ionization and many more
emissions, and more intense emissions, than from the flame.
• The ICP torch resembles a flame as the plasma emerges from the magnetic
field. The temperature of the ICP torch is in excess of 6000 K.
Inductively Coupled Plasma
Inductively Coupled Plasma
Inductively Coupled Plasma
• Not only does this make for a very broad application for ICP, but it
also means that a given sample may undergo very rapid and
simultaneous multielement analysis.
• The other option is comparable to the diode array design used with UV-VIS
molecular absorption instruments. This design is known as the simultaneous
direct reading polychromator design, in which there are a number of exit slits
and phototubes for measuring a number of lines at once
• Flame Photometry
• Cold Vapor Mercury
• Hydride Generation
• Spark Emission
• Atomic Fluorescence
Summary
BASIC PRINCIPLE
• Unlike all the others, this technique does not use light at
all
N2 MH+
++
+
+ + + + + +
++ + +
+ ++ +
MH2+
++ +
++ + + +
++
Sample in solution ++
+ + + +
+
++ ++ ++ +
+ +
N2 gas ++ +
MH3+
Sample plate
Laser
hn
+/- 20 kV Grid (0 V)
ION SOURCES
• Advantages :
1. high ion current ,thus giving good sensitivities.
2. Fragmentation unambiguous identification of analyte possible.
• Disadvantages :
3. Disappearance of molecular ion peak (sometime)
4. The need to volatize the sample (degradation before ionization can
occur )
5. Limited to analyte having MW smaller than 103 Daltons.
ION SOURCES
ION SOURCES
• One of the most common reagent is methane which react with high-energy
electrons to give several ions such as CH4+,CH3+and CH2+ .The first two
predominate and represent about 90%.This ions react rapidly with additional
methane molecules .
ION SOURCES
ION SOURCES
A high electric voltage (potential) that gives rise to lines of equipotential result in
an electric field crowd around the needle tip. The electric field is the most
intense at the surface (point) of the tip and is where ionization occurs.
R = m/Δm
Δm : the mass difference between two adjacent peaks that are just resolved,
m : the nominal mass of the first peak.
detector
+
+ +
V
• Ions are formed in pulses.
• The drift region is field free.
• Measures the time for ions to reach the detector.
• Small ions reach the detector before large ones.
MASS ANALYZER
2. Quadrupole
2. Quadrupole
• The most common mass analyzers in use today.
• More compact , less expensive, and more rugged than most other type of
MS.
• It offers the advantage of low scan times, which is particularly useful for real-
time scanning of chromatographic peaks.
• The heart of a quadrupole instrument is the four parallel cylindrical rods that
serve as electrodes.
• Opposite rods are connected electrically, one pair being attached to the
positive side of a variable dc source and the other pair to the negative
terminal.
• At any given moment ,all of the ions except those having a certain m/z
values strike the rods and are converted to neutral molecules .Thus only ions
having limited range of m/z value reach the transducer.
MASS ANALYZER
2. Quadrupole
MASS ANALYZER
2. Quadrupole
Quadrupoles have variable ion transmission modes
m2 m1
m4 m3 m2 m1
m4 m3
m2 m1
m2 m2 m2 m2
m4 m3
3. Ion Trap
• An ion trap is a device in which gaseous anions or cations can be
formed and confined for extended periods by electric and/or
magnetic fields.
• Several types of ion traps have been developed, and two are
currently used in commercial MS.
1- The ion cyclotron resonance trap.
2- The Simple ion trap.
MASS ANALYZER
4. Magnetic Sector
• In this instrument, the magnet is a powerful, variable-field
electromagnet, the poles of which are shaped to cause a bending
of the path of the fragments through a specific angle, such as 90 °
• It is possible to vary the field strength of the magnet
• in such a way as to scan the magnetic field and to focus the ion
fragments of variable mass-to-charge ratio onto the detector slit
one at a time.
• In this way, specific fragments created at the electron beam can be
separated from other fragments and detected individually.
MASS ANALYZER
4. Magnetic Sector
MASS ANALYZER
4. Magnetic Sector
• It is employ a permanent magnet or an electromagnet to cause the
beam from the ion source to travel in a circular path, most
commonly of 180,90 or 60 deg.
• Will KE= z e V=1/2mv2 ( V= voltage between A and B , and e=
electronic charge =1.60 x 10-19) all ions have the same number of
charges z are assumed to have the same kinetic energy after
acceleration regardless of their mass.
• The heavier ions must travel through the magnetic sector at lower
velocities.
MASS ANALYZER
• Ion Trajectery
Ions drift into a spatially uniform static magnetic field of strength (B) that causes the
motion to become circular in a plane perpendicular to the direction of the magnetic
field. Within this ion-trap, the angular frequency (ωc) is inversly proportional to
the m/z value
MASS ANALYZER
The presence of ions between a pair of detector electrode (in the trapping
cell) will not actually produce any measurable signal. It is necessary to excite
the ions of a given m/z as a coherent package to a larger orbital radius, by
applying an RF sweep of a few milliseconds across the cell. One frequency
will excite one particular (Fourier transformation allows for all frequencies to
measure simultaneously)
Measurement of the angular frequency leads to values for m/z and thus to the
mass spectrum. Because frequency can be measured more accurately than
any other physical property, the technique has a very high mass resolution
MASS ANALYZER
1. Electron Multiplier
2. Faraday Cup
• Ion travelling at high speed strike the inside of the metal (faraday) cup and cause
secondary electron to be ejected. This production of electrons constitutes a
temporary flow of electric current until the electrons have been captured. The
faraday cup detector is simple and robust and is used in situations in which high
sensitivity is not required
• The transducer is aligned so that ions exiting the analyzer strike the collector
electrode.
• This electrode is surrounded by a cage that prevents the escape of reflected ions
and ejected secondary electrons.
• The collector electrode is inclined with respect to the path of the entering ions
so that particles striking or leaving the electrode are reflected a way from the
entrance to the cup.
DETECTOR
2. Faraday Cup
DETECTOR
3. Scintillator detector
• A fast ion causes electrons to be emitted and these are accelerated towards a
second dynode. In this case, the dynode consist of a substance (a scintillator)
wich emits photons (light). The emitted light is detected by a photomultiplier
and is converted into an electric current. Since photon multipliers are very
sensitive, high gain amplification of the arrival of a single ion is achieved. This
detectors are also important in studies on metastable ions.
MASS SPECTRA
• In most cases there will be more than one analyte ion for a given
element (due to the presence of more than one naturally
occurring isotope) and therefore more than one signal for that
element.
ICP-MS
• Nearly all elements, however, have at least one isotope mass that
no other element has.