Professional Documents
Culture Documents
Laboratory Apparatus’
Description of Apparatus
Basic Operation
Installation
Sample Preparation
Results
Compiled by
RDC 141 – CO2 Reactivity
Coke
Pressure Gauge
Air Out
Air In
Programmer
Furnace
Air In
Microprocessor
Furnace
Sample Pre-heater
Water Thermostat
Microprocessor
Measuring Unit
Calibration &
Measurement
Description of Apparatus
- Apparatus consists of a measuring unit, control
box, water thermostat and sample pre-heater
- Measurement cell; bottom section stabilized at
20º C, upper head heated to 60º C
- Sample pre-heater holds approximately 40 samples
at a temperature of 40º C
- Thermostat temperature is maintained at 20º C ±
0.5º C
- Control unit calculates results automatically @ W/mK
- Equipment to be calibrated before a series of measurements
- Replace standard sample at least once every 3 months
- Recommended to have two standards
Basic Operation
- Before calibration and routine sample analysis, warm up
equipment for at least 24 hours
- Do not switch apparatus off
- No sample is to be inserted while apparatus conducts an
internal calibration (When switching on or after power failure)
- Do not mark samples on flat surfaces. This affects results
- Place samples in pre-heater 20 minutes before measuring
- Place one sample in measuring cell with insulation ring and
clamp
- Press measuring button
- Measurement takes a few minutes
- Take results.
Installation
- Support apparatus at the base when removing it out of the crate
- Check apparatus for any visible signs of damage that may have
occurred during transportation
- Power plug not supplied – have apparatus wired by qualified electrician
- Apparatus must be properly grounded
- Apparatus to be placed on a solid, flat working surface
- Do not install below a window, air-conditioner or beside a door
(apparatus is EXTREMELY sensitive to a thermal unstable environment)
- Laboratory temperature to remain constant at 21 – 23 º C
- Supply equipment with constant voltage (install stabiliser if necessary)
- Adjust water flow to 40 l/h on thermostat
- Sample pre-heater temperature must be adjusted to 40 º C
- Use only distilled water in thermostat
Sample Preparation
- Core to be extracted from a larger body and
sampling to be carried out according to ISO 8007
- Core diameter 50 ± 0.4 mm. Height 20 ± 0.1 mm
(use grinding machine RDC 149)
- If sample surfaces are not completely flat and
parallel to each other, wrong results will be obtained
- Wash and dry samples at 120º C ± 5º C for 12
hours and let cool to room temperature before measuring
- Record sample dimensions (diameter and height)
Results
Deflection Transducer
Press Assembly
Microprocessor
Waste Shovel
Printer
Hydraulic Control
Unit (inside)
Description of Apparatus
- The apparatus consists of a press assembly where
the sample is placed and a force applied by a load
cell, hydraulic control unit for a compressive force,
transducer to measure deflection and microprocessor
- Microprocessor calculates results automatically
- Results are printed
- Sample is compressed in press assembly unit and crushed (eye
protection is necessary)
- Both compressive strength and young’s modulus
measurements are done simultaneously
Basic Operation
- Sample is subjected to an applied compressive load
- Deflection of sample under load and load at fracture are measured
- Results are calculated by microprocessor and printed
- Set date, time and sample number before measuring
- Calibrate with steel cylinder (50 mm) before conducting a series of
measurements – press “CAL”
- Small variations in core diameter of ± 0.2 mm are acceptable
- If diameter variations are greater – change diameter reading
- Sample placed in press assembly – press “START”
- Clean sample dust from press assembly after every measurement
- If sample is too weak, young’s modulus cannot be calculated “BAD
SAMPLE”
- If a very strong sample does not fracture under maximum load, value
displayed will be incorrect
Installation
- Support apparatus at the base when removing it out of the crate
- Check apparatus for any visible signs of damage that may have occurred
during transportation
- Power plug not supplied – have apparatus wired by qualified electrician
- Apparatus must be properly grounded
- Apparatus to be placed on a solid, flat working surface
- Once located, remove transducer from protective package and place in
locating hole on top of press assembly
- Attach shovel holder with two screws provided
- Level of hydraulic oil in the hydraulic control unit should reach the mark
line. If it needs topping, use only an approved hydraulic oil
- Check pressure relief valve on hydraulic control – set to 110 bar
- Check that bottom plate is driven upwards with 3 s/mm. Do not take into
account the first 3 mm. Velocity is adjusted by turning the oil flow valve
inside the apparatus
Sample Preparation
- Core to be extracted from a larger body and
sampling to be carried out according to ISO 8007
- Core diameter 50 mm (48 – 52 mm acceptable)
- Length 50 ± 0.1 mm
- Wash sample and dry at 120 ± 5º C for 12 hours and cool to
room temperature
- Measure sample diameter before testing (to a precision of 0.1
mm, e.g. 50.3 mm)
- If diameter on apparatus is not adjusted to core diameter of
sample, it may be necessary to calculate deviation manually
from results obtained
Results
Capillary standard
Microprocessor
Measuring liquid
Measuring and calibration
Measuring head
Ventilation fan
Description of Apparatus
- Results are calculated automatically by a
microprocessor and presented on a display
- The unit consists of a permeability cell, a vacuum
pump, U-tube with sensors and control box
- The permeability cell has been designed to ensure
airtight sealing around the periphery of the sample
Basic Operation
Furnace
Description of Apparatus
- Apparatus consists of 1 or more furnaces each
with an electronically-controlled, movable lid, an
electrical transformer, a control box with a temperature
controller and a microprocessor for controlling the furnace and
gas operations
- Two samples are heated simultaneously in one furnace
- Apparatus Symmetry is important for result accuracy and
should be checked weekly
Basic Operation
- Carboxy reactivity is dependent on two anode properties: a) The
intrinsic CO2 Reactivity and b) The permeability of the anode
- The test exposes the anode to a carbon dioxide atmosphere at a
temperature of 960º C for 7 hours
- After cooling, the samples are mechanically tumbled with steel balls
in the RDC 181 – Tumbling Apparatus
- The total weight loss is divided into 2 components; Loss due to
burning & loss due to dust
- All results are calculated in % loss
- Wear gloves when handling samples
- Measure ONLY baked carbon samples
- Insert weighed samples in sample holder & lower furnace lid
- Press “MAN” and check that 200L CO2 /h is supplied to the furnace
- After test, remove samples from furnace & leave to cool
- Weigh each sample with it’s corresponding dust
- Tumble sample for 18 minutes and weigh residue only
- Calculate results
Installation
Control Unit /
Press Assembly
Microprocessor
Printer
Waste
jar
Description of Apparatus
- The apparatus consists of a press assembly, a
hydraulic control unit, transducer for measuring
the height on the granular column and a control
box (microprocessor)
Basic Operation
- The specific electrical resistance, bulk density and
compressibility of coke is measured
- A fraction of granular coke is pressed in a cylinder and a fixed
and constant current (DC) is applied
- The voltage drop and the height of the granular column under
zero and full load is measured
- Results are calculated automatically by the microprocessor
- Data is printed on paper
- Coke material is fed to the measuring cell through a funnel
- At the end of the measurement, the base plate moves to open
the cell and the contents are emptied by gravity
- Apparatus to be calibrated prior to conducting a series of tests
Installation
- Support apparatus at the base when removing it out of the crate (use
overhead crane if required)
- Check apparatus for any visible signs of damage that may have
occurred during transportation
- Power plug not supplied – have apparatus wired by qualified
electrician
- Apparatus must be properly grounded
- Check the level of hydraulic oil in control unit. Use only approved
hydraulic oil for topping up
- Equipment may be switched off if no sampling is done
- Apparatus to be switched on at least 15 minutes before conducting
any measurement
Sample Preparation
- Representative sample in accordance with ISO
Doc. 6375
- Divide into 3 fractions by sieving (ISO Doc. 2325)
-I > 1.4 mm (II b Crushed to 1 – 1.4 mm)
II a 1 – 1.4 mm
III < 1 mm
- Fractions IIa & IIb are mixed
- Dry at 120º C ± 2º C to constant weight
- If sample material is coated with oil, rinse with Methylene
Chloride before drying (ISO Doc. 8723)
- Coke Sample – Weight 15 ± 0.1g
Results
- Results are calculated automatically by the
microprocessor as follows:
RES = MOHM
BD = Bulk Density in g/cm3 or kg/l
PD = Pressed Density in g/cm3 or kg/l
CON = Contraction factor in %
RCY = Resiliency in % of the coke after reduction of
the pressure
RDC 148 – Diamond
Wheel Saw
Diamond wheel
Motor
Movable saw table
Description of Apparatus
- The machine consist of a motor, diamond wheel
and moveable saw table
- Cores with a diameter of 50 mm can be cut to a
maximum length of 130 mm
Basic Operation
- Keep hands and fingers clear of saw blade while cutting
- Always wear protective glasses during sawing
- Place sample on sawing table and start the motor
- If length required is 50 mm, insert steel cylinder with a
length of 80 mm (not delivered) between saw and sample
- Open tap during sawing
- Hold sample with both hands and push table as far as it will go
- Return the saw table and remove the two parts of the sample
- Clean equipment after a series of cuts in order to move the saw
table easily
Installation
Motor
Diamond wheel
(2 X – inside)
Display
Contact needles
Piston for needle
position adjustment
Sample holder
Description of Apparatus
- The apparatus consists of a sample holder with a
current block at each end
- Two contact needles set 50 mm apart to measure
actual voltage drop
- Voltmeter with digital display panel for showing
results
Basic Operation
- Spec. Elec. Resistance is determined by measuring the
voltage drop on a cylindrical sample at a constant current of
1 Ampere
- Keep hands and fingers clear of the sample holding device
during measurement
- Calibration is done before each series of measurements
- Place sample in holding device and check that needles are
equidistant from each other
- Press “START” and keep it pressed until result is displayed
- Carry out 4 measurements rotating the sample 90º each
time
Installation
- Support apparatus at the base when removing it out of the crate
(use overhead crane if required)
- Check apparatus for any visible signs of damage that may have
occurred during transportation
- Power plug not supplied – have apparatus wired by qualified
electrician
- Apparatus must be properly grounded
- Check level of hydraulic oil. If it needs topping, use only approved
oil
- Machine may be switched off if no sampling is done
Sample Preparation
- Sample core with a length of 130 mm and a
diameter of 50 mm
- Minimum specimen length 90 mm (position of
current block must be adjusted)
- Sampling to be carried out according to procedure
specified in ISO 8007
- Sample required has a length of 130 ± 0.2 mm and
a diameter of 50 mm ± 0.1 mm
- Sample to be washed and dried at 120 ± 5º C for
12 hours and cooled to room temperature
Results
- The measurement result is given in µΩm
- Results are shown on the digital display
- 4 measurements are taken per sample
- Calculate the average of the measurements for
individual results
- Where the results exceeds 20 µΩm, round off the
value to the nearest whole number
RDC 151 – Air Reactivity
Anode
Sample holding
rod
Agitating mechanism
Description of Apparatus
- Apparatus consists of 1 or three furnaces each
with an electrical transformer, a control box with a
temperature controller and a microprocessor for
controlling the furnace and gas operation
- The sample holder is manually controlled and has a
dust collection tray with agitation mechanism
- Symmetry is important for result accuracy and
should be checked weekly
- Measure only baked carbon samples
Basic Operation
Reservoir for
Xylene
Vacuum Container
Controller Pycnometer
Vacuum Pump
Description of Apparatus
- The unit consists of a vacuum pump, a vacuum
controller, container for the pycnometer and a
support
- The container has a removable lid, and the outlet
is connected to the pumping device
- The filling device has a reservoir for the Xylene, a
stopper and a tube that is placed in the pycnometer
Basic Operation
Funnel
Measuring Cylinder
Control Panel
Description of Apparatus
- The apparatus consists of a bulk volumeter for
determining the tapped bulk density of solids
- It also has a vibratory feeder
- Tamping mechanism
- 5 Digit electronic pre-selector
- Measuring Cylinder
- ISO/DIS 10 236
Basic Operation
- The coke bulk density, when considered together with the density, is
one criterion for assessing the porosity of anodes
- Bulk density depends on the size, shape and porosity of grains
- 95 – 105 g of sample is placed into funnel
- Turn on bulk volumeter and vibratory feeder simultaneously at the
on/off switch and left hand dial on the feeder from 0 to 1
- Ensure that test portion is transferred to cylinder evenly in about 45
± 15 seconds
- Switch feeder off
- Volumeter will stop after 1500 tampings
- Read volume to 1 cm3 on cylinder (record as V)
Installation
- Support apparatus at the base when removing it
out of the crate. Be careful with glass ware
- Check apparatus for any visible signs of damage
that may have occurred during transportation
- Power plug not supplied – have apparatus wired
by qualified electrician
- Apparatus must be properly grounded
- The number of ‘taps’ must be set on the control
panel and checked (refer to page 12 of manual)
Sample Preparation
- Take a representative sample of the coke according to ISO 6375
- Dry uncrushed sample in an oven at 120 ± 5º C
- Separate sample into classes by screening through wire sieves
- For the test grain sizes 3.15 to 4 mm and/or 1 to 2 mm are used
- If the granular range 0.25 to 8 mm is to be examined, the following
classes are recommended:
4 to 8 mm
2 to 4 mm
1 to 2 mm
0.5 to 1 mm
0.25 to 0.5 mm
- Weigh 95 to 105 g to nearest 0.1 g (m) and place in funnel
Results
- Bulk density is determined from the mass and volume of the
granular material after tamping
- (TPD) Tapped bulk density is expressed in grams per cubic
centimeter
m
TPD = -----
V
M = mass of test portion (g)
V = Volume after test portion tamping (cm3)
RDC 155 – Dust Fineness
BLAINE
Load Piston
Display
U-Tube Operation Buttons
Printing Unit
Sample Holder
Control Panel
Electrical Motor
Motor
Control Switches
Control Box
Control Switches
Furnace
Length Measuring
Temperature Controller Unit
Air Pressure Control
Description of Apparatus
- Measurement of mechanical properties for carbon
anodes is important due to their influence on thermal
shock behaviour
- Thermal shock fracture results from the mechanical
stresses in an anode due to heat wave penetration
- The unit consists of a furnace, temperature controller,
Sylvac P10L Probe, a length measurement unit and a
pneumatic double acting cylinder for the insertion and
removal of samples
Basic Operation
- A sample is heated in a furnace at a constant
temperature of 300º C and the expansion is
measured after 3 hours
- Insert samples to be measured. 3-6 seconds after
sample is in the top position, pre-set length
measurement unit to zero
- After 3 hours read the length difference on each
display
Installation
- Support apparatus at the base when removing it out of the crate. May
be lifted with overhead crane
- Check apparatus for any visible signs of damage that may have
occurred during transportation
- Power plug not supplied – have apparatus wired by qualified electrician
- Apparatus must be properly grounded
- Apparatus is extremely sensitive to ambient temperature and draught
(temp variations below 2º C)
- Before connecting to air supply, ensure that valve is open (turn to the
left)
- Air should be free of water (less than 100 ppm)
- Increase gas pressure slowly. Regulate to 2 bar
- Machine to be turned on two days prior to commissioning
- Insert a blank sample into the furnace during heating
Sample Preparation
- Sampling to be carried out according to procedure
specified in ISO 8007
- Maximum core diameter 50.8 mm
- Length 50 ± 0.2 mm
- Dry samples at 120º C for 12 hours and let to cool
to room temperature
Results
- The average coefficient of thermal expansion is
measured (CTE) from 25 - 300º C
___ Δl 1000 x Δl
CTE (25-300º C) = -----------=--------------x 10-6 K-1
l x (300-25)K 13.75
Control Box
Pneumatically operated
furnace cover
Air Connection
Furnace
Description of Apparatus
Thermometer
Heating Mantle
3 Glass beakers;
not included
Bunsen burner (100ml)
Description of Apparatus
- Distillation unit for heating pitch consisting of:
retort stand, collar clamp, mounting arm,
thermometer, bunsen burner (not included) and
electronically controlled heating mantle
- 3 (100 ml) beakers are required (not included)
Basic Operation
(F1+F2)
PD360 = ------ X 100%
M0
RDC 171 Quinoline Insoluble
Temperature Regulator
Condenser
Flask
Heating Mantle
Description of Apparatus
- The quinoline insoluble fraction of binder pitches
are measured by digesting an amount of binder
sample in quinoline at 80º C for a given time
- The solution is then filtered and the filter cake dried
- This is weighed and expressed as a % on the initial
material
- The apparatus consists of a Glassware-unit and a
filtration unit for extracting the insolubles
- Vacuum grease should be smeared on all the joints
and seals
Basic Operation
- The quinoline has not been supplied with the
equipment
- Filter paper must be dried and placed in desiccator
- Before measuring, make sure that the hot-plate
and appropriate glassware is available
- The binder of pitch is weighed, digested with
quinoline at 80º C and filtered to extract the
insoluble material as a filter cake
- The cake is dried and weighed
Installation
- Install in an efficiently vented HOOD and check that hood is operating
properly. Conduct all measurements in a fume cupboard
- The mantle must not be energised without a flask containing liquid
- If the heating mantle gets wet, unplug from power supply
immediately
- Avoid exposure to excessive levels of gas
- Support apparatus at the base when removing it out of the crate
- Check apparatus for any visible signs of damage that may have
occurred during transportation
- Power plug not supplied – have apparatus wired by qualified
electrician
- Apparatus must be properly grounded
Sample Preparation
- Sampling to be done according to ISO 6257
- Parent sample is crushed to pass a 0.25 mm
screen and a sub-sample of 1 ± 0.5 g is taken for
analysis
Results
- The quantity of quinoline-insoluble material,
expressed as a percentage by mass, is given by the
formula:
(M1 – M2)
Q1 = ---------------- x 100
M0
M0 = mass of test portion
M1 = Mass of filter paper and filter cake after drying
M2 = Mass of dried filter paper
RDC 172 Toluene Insoluble
Condenser
Temperature Regulator
Flask
Heating Mantle
Description of Apparatus
- The apparatus consists of a reflux unit for
boiling the pitch and toluene reagents as
well as a filtration unit for extracting the
insolubles
- Vacuum grease should be smeared on all
the joints and seals
Basic Operation
- The toluene has not been supplied with the equipment
- Filter paper must be dried and placed in desiccator
- Cooling water must be available and flowing freely
- A vacuum source for the filter flask must be available (e.g. a water jet
or electric vacuum pump)
- Before measuring, make sure that the hot-plate and appropriate
glassware is available
- A sample of binder pitch is weighed, boiled under reflux with toluene
and filtered
- Place test portion with 150 ml toluene in 250 ml glass flask
- Start condenser water circulation and turn on heating mantle
- Bring contents to a steady boil and ensure that a reflux action occurs
(vapours should condense and drop back into flask) for 15 minutes
- Agitate the flask from time to time
- Heat 100 ml toluene to 100 º C on a hot plate. Insert a dried and
weighed filter paper into the filtration unit and moisten with hot
toluene (refer Ops Manual)
- When filtration is complete, remove filter paper and insolubles , dry
for two hours at 140º C - Weigh to the nearest 0.001 g
Installation
- Install in an efficiently vented HOOD and check that hood is
operating properly. Conduct all measurements in a fume cupboard
- The mantle must not be energised without a flask containing liquid
- If the heating mantle gets wet, unplug from power supply
immediately
- Avoid exposure to excessive levels of gas
- Support apparatus at the base when removing it out of the crate
- Check apparatus for any visible signs of damage that may have
occurred during transportation
- Power plug not supplied – have apparatus wired by qualified
electrician
- Apparatus must be properly grounded
Sample Preparation
- Sampling to be done according to ISO 6257
- Parent sample is crushed to pass a 0.25 mm
screen and a sub-sample of 1 ± 0.5 g is taken for
analysis
Results
- The toluene insoluble fraction is also known as the sum of α and β
resins
- The quantity of toluene-insoluble material,
expressed as a percentage by mass, is given by the
formula:
(M1 – M2)
T1 = ---------------- x 100
M0
M0 = mass of test portion
M1 = Mass of filter paper and filter cake after drying
M2 = Mass of dried filter paper
RDC 173 Water Content
Water Out
Condenser
Water In
Heating Mantle
Description of Apparatus
- The apparatus consists of a distillation unit for
heating the pitch
- Xylene is used for the measurements but has not
been included with the apparatus
Basic Operation
- Perfect cleanliness of the apparatus is essential
- Remove all traces of contamination from the graduated receiver and from the
interior of the reflux condenser by washing with chromic-sulphuric acid, distilled
water and acetone. Dry
- Put ~2 ml water in receiver and fill with xylene. Open stop clock and close when
water level reaches 0
- Add 150 ml of xylene to distillation flask with test portion
- Heat gently until test portion dissolves and then adjust heat so that condensate
falls from the end of the reflux condenser at a rate of 2 – 5 drops / sec
- Continue until water level remains constant for 5 minutes in graduate receiver
- Allow to cool and read the volume of water to nearest scale division; correct
with any certified error of receiver
Installation
- Install in an efficiently vented HOOD and check that hood is operating properly. - Conduct all
measurements in a fume cupboard
- The mantle must not be energised without a flask containing liquid
- If the heating mantle gets wet, unplug from power supply immediately
- Avoid exposure to excessive levels of gas
- Support apparatus at the base when removing it out of the crate
- Check apparatus for any visible signs of damage that may have occurred during
transportation
- Power plug not supplied – have apparatus wired by qualified electrician
- Apparatus must be properly grounded
- Cooling water must be available and flowing freely
- Smear all joints and seals with vacuum grease
Sample Preparation
- Sampling to be carried out according to ISO 6375
- Parent sample is crushed to pass a 8 mm screen
and a sub-sample of 100 ± 0.1 g taken for
analysis
Results
- The water content, expressed as a percentage by mass, is
given by the formula:
M1 x 100
--------------
M0
Cooler
Extractor
Thimble
Flask
Heater
Description of Apparatus
- This test is useful for the control of green paste
composition for green electrodes, Söederberg and
ramming paste
- The method is based on the extraction of the
binder in an apparatus using hot quinoline
- The apparatus consists of an Extractor Tube with
a volume of 1000 ml
- Flask Ø 48 mm, h = 200 mm
- Cooler
Basic Operation
- 200 g sample weighed into extractor thimble
- Thimble placed in extractor and 750 ml fresh quinoline is added into
flask
- Cooler is then fitted on top of extractor and flask is heated up slowly
- When boiling, adjust water flow rate to obtain complete condensation
of quinoline vapours
- Apparatus is let under reflux until quinoline in extractor is clear (24 –
36 h)
- Quinoline is replaced with 750 ml methanol and the rest of the
quinoline is removed after 4 hours of reflux
- Apparatus is cooled and placed in a 400 ml beaker and dried in a
cabinet for 4 hours
- After cooling in desiccator, thimble is weighed
Sample Preparation
- Sample weight 200 g ± 0.1 g
- Granulometry : < 8 mm dried at 106º C for 2
hours
Results
- Results are calculated as follows:
(I-F) X 100
Binder Content (%) = -----------------------
I
I = Initial Weight
F = Final Weight
RDC 175 Volatile Matter
Muffle Furnace
Temperature
Crucible stand
Controller
with Crucibles
Description of Apparatus
- The apparatus is used to measure the volatile matter
of green coke produced in delayed coking process
- The method used is a conventional heating gravimetric
method, where volatile matter is calculated in terms of
loss from a test portion heated in a crucible at a fixed
temperature of 900º C for 7 minutes
- The apparatus consists of quartz crucibles with a
crucible stand, electric muffle furnace and temperature
controller
Basic Operation
- Heat 6 quartz crucibles and lids for ± 1 hour at 900º C
- Allow crucibles to cool in dessicator to room temperature
- Weigh each crucible (without lid) to nearest 0.001g (Wc)
- Weigh ± 1g of test portion into each crucible and record weight to
nearest 0.001g (Wa)
- Put lids on and insert into crucible stand
- Insert into pre-heated furnace as quickly as possible
- Wait 7 minutes ± 5 seconds
- Remove crucibles from furnace and place on a cold base for 3 minutes
- Leave crucibles to cool in desiccator for 30 minutes
- Remove lids from crucible
- Weigh crucibles and samples to nearest 0.001g (Wb)
- Crucible stand must have ‘all’ crucibles placed in holder
- Double analysis for each sample is required
Installation
- Keep apparatus away from any door or building structure
- Well ventilated location or use fume hood where fresh air is
continuously recycled
- Support apparatus at the base when removing it out of the crate
- Check apparatus for any visible signs of damage that may have
occurred during transportation
- Power plug not supplied – have apparatus wired by qualified
electrician
- Apparatus must be properly grounded
- Connect the exhaust outlet to a piping system that disperses the
volatiles outside the building
- Program the temperature controller
- Press “start” button and ensure that temperature reaches 900 ± 10º C
within 30-40 minutes
- If the furnace is not used, turn off the rocker switch on the
temperature controller
Sample Preparation
- A laboratory sample of green coke is collected
according to the procedure specified in ISO 6375
- ± 200g of sample is crushed in a mortar so that it
passes through a sieve of 0.25 mm apertures
- The crushed and sieved product is dried in an
electric oven at 120 ± 2º C, to constant weight
Results
- The volatile matter of green coke, expressed as a
percentage by mass, is given by the following
formula:
(Wb - Wa) x 100
----------------------
Wb – Wc
Wa = Weight of crucible w/o lid + sample after test
Wb = Weight of crucible w/o lid + sample before test
Wc = Weight of crucible without lid
RDC 178 Coking Value
Temperature
Measuring Electric Furnace
Device
Crucibles and
Support
Description of Apparatus
- The coking value is the amount of residue that is
obtained by heating pitch to 550º C
- The coking value of pitch is related to the density
of the binder matrix within the anode structure and
responsible for the following properties: apparent
density, permeability, thermal conductivity and other
mechanical properties
- The apparatus consists of an electric muffle furnace,
temperature measuring device, nickle and porcelain
crucibles and a support
Basic Operation
- Heat 2 porcelain crucibles with lid for ± 2 hours at 560 ± 10º C
- Allow to cool to ambient temperature in desiccator and weigh
to nearest 0.001g
- Test portion of 1 ± 0.05g
- Make a bed of 10 ± 1 mm thick on base of nickel crucible. Place one of
porcelain crucibles with test portion & close lid
- Fill space between the two crucibles with more petroleum coke so that
the porcelain crucible is completely embedded
- Close nickel crucible with lid
- Repeat procedure with the other crucible and place on support in
furnace at 550 ± 10º C
- After 2,5 h, remove from furnace and allow to cool
- Weigh porcelain crucibles and contents to nearest 0.001g
Installation
- Keep apparatus away from any door or building structure
- Well ventilated location or use fume hood where fresh air is
continuously recycled
- Support apparatus at the base when removing it out of the crate
- Check apparatus for any visible signs of damage that may have
occurred during transportation
- Power plug not supplied – have apparatus wired by qualified
electrician
- Apparatus must be properly grounded
- Connect the exhaust outlet to a piping system that disperses the
volatiles outside the building
- Program the temperature controller
- Press “start” button and ensure that temperature reaches 550 º C and
that the temperature remains constant
- If the furnace is not used, turn off the key on the temperature
controller
Sample Preparation
- Sampling to be done according to ISO 6257 method
- If the sample is sufficiently hard, crush with small
jaw-crusher and grind in mortar to pass 300 µm
mesh sieve, and if possible, a 212 µm sieve
Results
- The coking value, expressed as a percentage by
mass, is given by the formula:
M2 – M1
---------------- X 100
M0
M0 = Mass of test portion
M1 = Mass of empty porcelain crucible
M2 = Mass of porcelain crucible and residue
Report the mean value of the duplicate determinations
RDC 180 Bench Drilling Machine
On / Off Switch
Drill Alignment Drill Insert Handle
Handle Speed Adjustment
Steel Cylinder
Motor
Sieve and Pan
Pressure Gauge
Mould Assembly
Press
Description of Apparatus
- The apparatus is used to produce the compressed
sample which is inserted into the Mettler Furnace
to in order to establish the softening point of the
binder
- The apparatus consists of a press (hydraulic driver
piston that applies a pressure of 100 bar to a
mould assembly mounted underneath)
Basic Operation
- Place the test portion into the Top Ram Holder and insert the Top Ram
- Compress the pitch into the cup with 100 bar pressure by pressing
“PRESS”
- Sample is pressed for 15 seconds
- Remove mould assembly and rotate the upper half while keeping the
sample cup holder stationary
- This causes the pitch to shear horizontally at the top of the sample cup
- Separate the different parts of the mould assembly
- Clean the mould of all residual pitch to avoid misalignment or binding
- The sample cup can now be used for the determination of the softening
point
Installation
- Support apparatus at the base when removing it out of the crate
- Check apparatus for any visible signs of damage that may have
occurred during transportation
- Power plug not supplied; have apparatus wired by qualified
electrician
- Apparatus must be properly grounded
- Assemble the Mould and place the sample cup with the smaller
diameter mounted downwards
- Place mould under the press and push “press” . Check that 100 bar is
applied. The pressure may have to be adjusted
Sample Preparation
- Sampling to be carried out in accordance with ISO
6257 method
- Parent sample is crushed to pass a 2 mm screen
and a sub-sample of 1 ± 0.01g is taken for
analysis
RDC 186 Density in Water
Basic Operation
- Use same procedure as for density in Xylene
RDC 187 Flexural Strength
Printer
Description of Apparatus
- Mechanical properties for carbon anodes such as
flexural strength are influenced by mixing
conditions
- Poor values may result in problems during anode
handling and decreased thermal cracking
resistance
- During the measurement, an increased force is
applied to a cylindrical shaped sample and the
value at which the sample fractures is measured
Basic Operation
- Set date, time and sample identification number
- Measure sample diameter
- Set core diameter (small variations, up to ± 0.2 mm
are acceptable)
- Insert carbon sample core in the press assembly
- Push “Start”
- Clean press assembly of residue with a brush
- Results are calculated by microprocessor and
printed
Installation
- Support apparatus at the base when removing it out of the crate.
An overhead crane may be used (Weight 100 kg)
- Check apparatus for any visible signs of damage that may have
occurred during transportation
- Power plug not supplied – have apparatus wired by qualified
electrician
- Apparatus must be properly grounded
- Check the level of oil in pressure control unit (mark line) and use only
RDC approved oil if topping is required
- Pressure relief valve on the hydraulic control to be set to 45 bar
- Check that the bottom plate lifts at 5 s/mm. Do not take into account
the first 3 mm. Adjust by turning the oil flow valve inside the
apparatus
Sample Preparation
- The test specimen is to be extracted from a larger
body according to ISO 8007 procedure
- Core Ø 50 ± 1.0 mm, length 130 mm
- Sample to be washed and dried at 120 ± 5º C for
12 hours and cooled to room temperature
Results
- Flexural Strength is given in Mpa
RDC 188 Hardness Tester
Turn Screw
Base with
penetrating pins
Description of Apparatus
- The apparatus is used to determine the degree of
hardness of the pre-baked anode butts
- The measurement is calculated by the depth of
penetration of two pins into the butt sample, by
rotating the screw mechanism
Basic Operation
- A load of 20 kg with a maximum variation of 5 kg is to be
applied by the user
- Hold the instrument firmly with one hand on the oxidised
anode surface
- Slowly turn the handle until the pins touch and slightly
penetrate the carbon surface
- This is the starting position
- Turn the handle and count the number of half revolution
- The amount of revolutions including the last where instrument
motion is observed are counted
- This corresponds to the depth in mm
Results
- Do 3 measurements on three different locations
and calculate the mean value
- Due to the scatter of anode butts, it is
recommended to measure at least 30 butts to
characterize a given population
Criterion for soft and hard butts
Hard : 1 to 3 mm
Medium : 4 to 5 mm
Soft : above 6 mm
RDC 192 Sodium Vapour Test
Manometer
Temperature Controller
Furnace
Vacuum Pump
Description of Apparatus
- The apparatus is used to measure the chemical resistance of
carbon towards sodium
- The test specimen is heated together with sodium to 700º C for
2 hours
- After this treatment, the changed mechanical properties can be
determined by measuring the compressive strength
- The apparatus consists of a steel vessel where the sample is
placed in a furnace, a vacuum pump, a transformer and a
control box
- The control box has a temperature controller to regulate the
heat-up rate and two valves to switch between vacuum pump
and argon supply
Basic Operation
- Sodium must never be exposed to water
- The sodium is not supplied with the apparatus
- Weigh the metal test tube without sample holder or lid
- Weigh ~ 13g ± 0.5g of sodium (dried with paper) and insert into test tube
- Insert sample holder with dried sample in test tube and close with lid
- Place in heater and connect test tube with the plastic tube on right side of
control box
- Press “START’ and start vacuum pump
- Open test tube valve until pressure has reached 1.0 bar and switch vacuum off
- Never start analysis before test tube is airtight at -1.0 bar for at least 10
minutes
- Break the vacuum with Argon
- Test tube pressure will increase from -0.8 to ~0.2 bar while temperature
reaches 800º C
- Test lasts around 12 hours. Heater will stop and cools down
- Follow procedure in Operational Manual carefully
Installation
- Quality of gas is important: Ar better than 99.99%, H 2O ≤ 10 ppm, O2 ≤ 5 ppm
- Inlet pressure not to exceed 0.2 bar & use appropriate tubing
- Apparatus to be kept in a proper water free fume cupboard. The cupboard has
to be free of any chemical material or working utensils
- Use overhead crane (use support structures provided on equipment) – Weight
180 kg
- Check apparatus for any visible signs of damage that may have occurred during
transportation
- Power plug not supplied – have apparatus wired by qualified electrician
- Apparatus must be properly grounded
- Connect argon supply
- Conduct air-tightness test until vacuum pressure remains stable at -1.0 bar for at
least 20 minutes (refer to Installation Section of Manual)
Sample Preparation
- Sampling to be carried out as per ISO 8007
- Core Ø 50 mm, although values in the range of
48-52 mm are acceptable
- Length must be 50 ± 1 mm and the sample is to
be dried at 120 ± 5º C for 24 hours, cooled and
stored in desiccator
Results
- Calculate sodium resistance as follows:
CSA
-------
CSB