USER'S MANUAL
1
Contenido
I. Generation..............................................................................................................................................3
II. Main Technical Data...............................................................................................................................3
III. Structure Characteristics....................................................................................................................4
IV. Operation procedure..........................................................................................................................4
(I) The preparation and inspection before use........................................................................................4
(II) Operation in course........................................................................................................................5
(III) Data setting & checking..................................................................................................................6
(IV) Sterilizing processing......................................................................................................................6
V. Attention and Maintenance..................................................................................................................10
VI. WARNING.........................................................................................................................................10
Page 2
I. Generation
Model LS-B75L Vertical Cylindrical Pressure Steam Sterilizer is controlled by microcomputer, installed with
an electrical heater, a timer, an automatically pressure-temperature controller, a safety valve, a releasing
valve, a pressure-temperature indicator, an alarm bell for finished sterilization and an instrument to
automatically cut off the power of heater. It has the advantages .of effective sterilization, convenient
operation, safe, less electricity consuming and cheap. It is ideal equipment for clinics, scientific research
institutes and other organizations to sterilizer surgical instruments, fabrics, glasses, culture media etc.
Sketch diagram
1 Housing
2 Drainage Knob
3 Overload protection
switch
4 Control board
5 Spanner
6 handle
7 Ring screws
8 Releasing Valve
9 Lid of the Container
10 Safs Valve
11 Gasket
12 Flange
13 Tightening Bolt
14 Brand
15 Drum
16 Container
17 water level pipe
18 power wire (Fig. 1 ) structure
19 easy adapter for release
Page 3
III. Structure Characteristics
1. The sterilizer component by container, lid, pipe. control system, heating system and safe systems
etc.
2. The structure of the unit is a single layer cylindrical body. Its design, manufacture and inspection
are strictly operated in accordance with the State Standards of Utilization Safety of Pressure
Vessel (see the Quality Certificate for the details).
3. The lid is pressed once, the lock type is by screw, safe and reliable.
4. The pipe structured by electric valve and manual water release valve, the valves reliable on
movement.
5. The outline of the unit is quadrate. The displays and the control switches are centered on the
controller board in the front of the unit. It is convenient.
6. The control system adopted microcomputer full automatically program processing , digital display
the program.:
l) Touch panel: sterile temperature, time, exchange temperature be adjustable and with
memory function.
2) A steam pressure stabilizer is used, the range is o134°C to sterilize different objects.
3) Sterile time range 0 99minond be selected , when the chamber 's temperature reached the
setting value, the timing auto in back way, after sterilizing, it cut off the power and alarm
automatically.
4) Sterile exchange temperature range 0 110°C , for better release the cool air in chamber,
and make sure the sterilizing more complete and more safety.
7. The electrical-heated elements of the unit are immersion tubes AC220v with high heating effect. It
should be grounded reliably.
8. with over temperature , over pressure and lack water protection device :
1) Open type safe valve, normal it is close, when the pressure inner go over the limited, the safe
valve auto open and release the steam, until the pressure back to normal it will close again
for safety use.
2) With water level device, if no water or water lack occur, the alarm sound.
3) With over load protector, it will auto cut off power when short circuit, over voltage.
9. The main body is made of good quality stainless steel, and has the features of anti-corrosion ,
easy maintenance and long service life.
Page 4
(II) Operation in course
1, Key
set --- set sterile temperature/ sterile time/ exchange temperature and data check in the course
Set exchange temperature → (0 - 110 °C ) I sterile temperature → (0 - 134 °C ) and sterile time → (0 -
99min.)
▲ ------ increase. For data increase & reset in the course.
▼ ------- decrease
For data decrease & reset in the course
Start ------start work. For start sterile cycle.
Screen display
2 display
top nixie indicator------ temp. screen & data set Display real chamber temperature and
data set
Low nixie indicator --------- time screen & data set Display real sterile time and data set
value
Include prepare pilot, temperature rise pilot, sterile pilot and end pilot.
water level indicator------ indicate instant water level
Include high level and low level pilot.
On Low level, the low pilot light; high level the high pilot light, medium, the light flash.
Page 5
Break Circuit Switch
On the side lower, for standby state when the power on.
Setting Method:
1.1 Point set key: First time press set key top line nixie tube display 1111, means enters cool air release
setting, the second line nixie tube display cool air release setting last time. Press t the value increase, t
decrease, the range 0-110 °C.Keep press ▲ or ▼ ,the value rapid increase or decrease.
1.2 Press again setting key, top line nixie tube display 2222, means enters sterilizing setting, the second
line nixie tube display sterile temperature last time.
Press t the value increase, t decrease, the range 0-134 °C
Keep press t or t ,the value rapid increase or decrease.
1.3 Third time press setting key, top line nixie tube display 3333, means enters sterilizing time setting, the
second line nixie tube display sterile hours last time. Press t the value increase, t decrease, the range
0-99 min
Keep press t or t ,the value rapid increase or decrease.
1.4 Fourth time press setting key, top line display real temperature value, second line display the content
now you check. If you don't return to the state before check, after 30sec.auto return to the state of
before check
1.5 Store the data. Top line inner temperature, second line display 0.
Page 6
The customer can preset the required sterilizing time (see form 1) according to following table.
Reference table to sterilization time and temperature (form)
Objects Required heat preservation Steam Pressure Mpa Relative steam Temperature °C
time (min)
2. temperature rise
Press start key, enters processing of temperature rise.
2.1 Temperature rise pilot light.(note: only high level pilot light can be heating)
2.2 Top line nixie tube display real temperature , second display cool air release temperature.
2.3 Heat tube get electric and start heating, meanwhile cool air be release by valve.
2.4 When the temperature inner reached the setting temperature, release valve close, the heating tube
heating going on, Top line nixie tube display time value, second display front 2-bit hour, rear 2-bit min.
3, sterilizing
When the temperature inner reached the setting temperature, start the timing, one by one minute reduce.
At the end the time reduce to 0, tum to end, release start and the temperature go down, until reached to
102°C ,the buzzer sound to tell you end, gauge arm back to 0, after minute you can take the object be
sterilized .
Page 7
2. normal trouble and dealing
2. Circuit
Page 8
Fig.4 output circuit board High voltage, please cut off the power then repair.
Page 9
V. Attention and Maintenance
The operator of the sterilizer should have the operation knowledge and responsibility to regularly
maintain the unit to prevent any accident.
1. Please don't sterilizing different kinds of objects at the same time, such as surgical dressings and
solution, rubber goods and instruments, otherwise, it will not be result in effective sterilization. Use
a recording thermometer, sterilization indicative agent or other biological methods to check the
required temperature and time, to fit a reliable sterilization effect.
2. If the trouble happened in the course, please dealing it and what for 5 sec. then restart.
3. The solution which will be sterilized should be filled in heat resistant flasks, but don't fill too full.
1/2 – 3/4 of the volume are good. The flasks should be plugged with absorbent gauze, in no wise
with the non-pore plugs, such as rubber of cork stoppers. The flasks should be placed in a tray
and then in the container in case the flasks may break up and solution will remain in the tray and
will not contaminate other articles.
4. Every day after sterilization, drain the water from the sterilizer by opening the draining valve on
the right bottom of the unit, dry the sterilizer and scrub the water stain, to improve the quality of
sterilization and prolong the service life of the sterilizer.
5. The gasket of the lid may become hardened after a long time used, change with a new one.
6. Please inspection the pressure gauge every year, any problems please change it in time.
7. The temperature and the pressure meter must be standardization every year. The temperature
and the pressure are related data.
VI. WARNING
When the pressure exceeds 34 Psi, the sterilizer should automatically exhaust to drop the pressure. If
safety valve fails and pressure keeps increasing, please cut off the power as soon as possible. The
warning marks on the above of the power switch.