Shanghai Changji SYD-0165A Vacuum Distillation Tester
2
Shanghai Changji SYD-0165A Vacuum Distillation Tester
temperature controlling system for air bath of receiving flask, vacuum system, vacuum manometer,
illumination system, and cooling device for heater, etc.
The instrument is shown as Fig.1. The function of each parts of the instrument is described as
follows:
(1) (2) (3) (4) (5) (6) (7) (8) (9) (10)
(11) (12) (13) (14) (15) (16) (17) (18) (19) (20)
Fig.1
(1) Vacuum tube and joints for vacuum pump: To connect to a vacuum pump;
(2) Vacuum adjustment valve: To adjust vacuum degree of the instrument;
(3) Digital display pressure meter: To display the vacuum pressure value;
(4) Illumination light: Illumination light in the working chamber; It is convenient to read
reading of distillate;
(5) Glass door: The door of the air bath of receiving flask; Open the door and you can make the
determination;
3
Shanghai Changji SYD-0165A Vacuum Distillation Tester
(6) Heating furnace: Electric heating furnace; It is used to heat distillation flask;
(7) Distillation flask: To hold sample; It is placed on heating furnace;
(8) Thermometer: It is a glass thermometer, and it is inserted into the distillation flask;
(9) Fixed board for distillation flask: It is used to fix the distillation flask together with the
fixed strip, so the distillation flask will not fall down when you lower down the furnace;
(10) Up and down knob for heating furnace (Furnace height adjustment knob): It is to adjust
the height of heating furnace; It is on the right of the instrument;
(11) Vacuum pump: To create vacuum environment during tests;
(12) Joints for vacuum pump: To be connected with the vacuum system of the instrument;
(13) Temperature controller: To control the temperature in the air bath of receiving flask;
(14) Temperature sensor: To measure the temperature in the air bath of receiving flask;
(15) Receiving flask: 100 ml; it is to receive distillate;
(16) Voltmeter: It shows the value of heating voltage. It indicates the value of heating power;
(17) Light switch: Turn on the switch and the illumination light in the working chamber is on;
(18) Cool switch: Turn on the switch and the fan for the heating furnace starts working;
(19) Power switch: Turn on the switch and the instrument gets power;
(20) Heat adjustment knob: To adjust the working voltage of the heating furnace.
4
Shanghai Changji SYD-0165A Vacuum Distillation Tester
5
Shanghai Changji SYD-0165A Vacuum Distillation Tester
Fig.2
The methods to preset temperature: Pull the measurement switch of temperature controller
upwards to the temperature setting position. Adjust the SET knob using a screwdriver to preset
controlling temperature. Then pull the measurement switch downwards to the temperature
measuring position. The temperature shown on the temperature controller at this time is the real
temperature of air bath and the temperature controller will automatically control the temperature of
air bath to the controlling temperature.
8. Rotate the heat adjustment knob to heat the distillation flask and make determination as per
relative standards.
9. Turn off the heat adjustment knob after a test. Turn on the cool switch, lower the heater to
the lowest position, and open the glass door to let the distillation flask cool naturally.
10. When the reading of thermometer is lower than 100 ℃, please turn on the vacuum
adjustment valve slowly and turn off the vacuum pump at the same time.
(III) Attention
1. Keep the instrument clean and protect it from corrosiveness of acid, alkali, oil dirty, and
moisture, etc. Especially keep the electric parts away from moisture.
2. For the operation safety, please check the grounding end of the instrument to make sure it is
fine grounded.
3. Please turn off the heater and lower down the heater after test. Turn on the cool switch at the
same time to cool the heater.
4. Keep the instrument away from moisture and corrosive gas.
5. If the instrument is in trouble, please ask a professional to repair it or contact us immediately.
Do not disassemble the instrument by yourself.
Warning: If the electric parts of the instrument are in trouble, please turn off the power supply
6
Shanghai Changji SYD-0165A Vacuum Distillation Tester
switch and cut off the power supply at first, and then ask a professional to check and repair it.
6. When you use the pump for the first time, please fill oil to the pump until the surface of oil
reaches to the scale line in the window.
7. In order to ensure operation safety of the user, please do no touch the heater during tests.
Please keep the heater away from oils.
9 6 (Outer diameter)×4 (Inner diameter) Meter 2 Connect the pump and the
PU tube instrument
10 Fine port sleeve Piece 1
2. Technical documents
(1) Operation Manual 1 piece
(2) Quality Certificate 1 piece
(3) Repair Guarantee 1 piece
3.Optional parts
7
Shanghai Changji SYD-0165A Vacuum Distillation Tester
Packager: Checker:
Date: Date:
2. Please connect the vacuum tube and the vacuum degree calibrator
3. Please connect the vacuum pump.
Attention: No leak in all vacuum tubes.
4. Calibration Points: 2mmHg,15mmHg,35mmHg and 50mmHg.
5. Vacuumizing, until the readings of vacuum degree calibrator is 50mmHg and make the
readings stable.
6. Observe the readings of digital display vacuum meter, if it is not 50mmHg, please adjust it,
the adjustment method is:
Press the return key of the digital display vacuum meter. decimal places begins to flicker,it
means that the system starts calibration. So you can use upper and lower keys to adjust the value of
8
Shanghai Changji SYD-0165A Vacuum Distillation Tester
vacuum degree, until the same as 50mmHg of the vacuum degree calibrator.Then press the return
key again, it returns to the test state. If there is no operation on the meter in 10s, the digital display
vacuum meter will enter the test state automatically.
7. Continue to vacummize, until the readings of vacuum degree calibrator is 35mmHg,
15mmHg and 2mmHg, to make the reading stable.
8. Observe of the readings of the digital display vacuum meter separately, if they are not
35mmHg, 15mmHg and 2mmHg, please follow the fifth method to adjust, until the readings is the
same as the calibrator.
9. The readings of the "vacuum calibrator" and “Digital display vacuum meter” are within the
error range of the 0.5 mmHg. If some of readings are not in the range, please follow the 2mmHg to
judge.
Attention: When other high precision vacuum meters are used, the method of calibration is same as
above.