Professional Documents
Culture Documents
01 Basic Knowledge
02 Z3 Overview
03 Analysis Principle
04 Installation
www.zybio.com
contents
●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●
05 Maintenance
06 Mechanical structure
08 Trouble Shooting
www.zybio.com
01
●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●
Basic Knowledge
www.zybio.com
Blood Constituent
0-10%
40-50%
Smallest size, large Largest size, clear Granules in cytoplasm, Least common, Granules Granules in cytoplasm,
nucleus/cytoplasm ratio, cytoplasm, involved in stained in dark red, in cytoplasm, stained in stained in neutral pink,
clear cytoplasm, major ingestion of foreign responsible for dark blue, important react within an hour of
constituents of human substances such as combating infection source of histamine and tissue injury and are
immune system. bacterial or parasites. and parasites in cytokine. the hallmark of acute
(No granules) (No granules) vertebrates; inflammation;
25 ~ 30% 3 ~8% 0.5 ~ 3% 0 ~1% 50 ~70%
Middle Cells
Granulocyte
02
●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●
Z3 Overview
www.zybio.com
Z3 Overview —appearance
touch screen
sampling probe
start key [Aspirate]
power indicator
Z3 Overview —appearance
4 USB ports
LAN port
Power socket & switch
Waste sensor connector
Lyse house
Z3 Overview — Report parameter
Red Blood Cell count RBC Platelet larger cell ratio P-LCR
Mean Corpuscular Volume MCV White Blood Cell Histogram WBC Histogram
Mean Corpuscular Hemoglobin MCH Red Blood Cell Histogram RBC Histogram
Parameter Background
WBC ≤0.2×109/L
RBC ≤0.02×1012/L
HCT ≤0.5%
PLT ≤5×109/L
Good precision - stable results
HGB 0 ~ 250g/L
PLT 0 ~ 4000×10⁹/L
HCT 0~67%
Z3 Overview-Main Interface
21 parameters
3 histograms + Flag
Z3 Overview — Specification
Parameters 21parameters
Weight 18Kg
Power < 200VA
Easy operation and maintenance
Bottom is close,
aspiration hole is on aside.
Long reagent expiration date
One button to
troubleshooting
03
●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●
Analysis Principle
www.zybio.com
Analysis Principle-blood cell counting
U=R*I
Analysis Principle-blood cell counting
RBC are hundreds of times more than WBC, and some parts of WBC are in the same size
as RBC, so the purpose of adding lyse, is to dissolve the RBC.
Meanwhile, PLT still exist in WBC chamber, but signal of PLT is very weak, easy to be
affected by broken RBC pieces, that’s why WBC chamber don’t test the result of PLT.
Analysis Principle-blood cell counting
RBC signal
RBC chamber: analog signal Digital signal
When test RBC and PLT, WBC signal is also included. WBC result have been tested in WBC
chamber, so software can easily calculate precise RBC result.
PS: 1. RBC and PLT are tested in the same counting chamber.
2. WBC and HGB are tested in another counting chamber.
Analysis Principle--HGB measurement
Principle : Colorimetry
Lyse HGB
(Dissolved) HGB React
compound LED530nm
Colorimetry: C = F × A
HGB (C) = constant factor × lg (Blank light intensity / sample light intensity)
Provide an isotonic
environment for blood cell
04
●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●
Installation
www.zybio.com
Installation – preparation
• Tools needed :
Diagonal pliers
Cross screwdriver
Calibrator & QC (normal)
Blood sample Diagonal pliers Cross screwdriver
Installation Video
Calibrator & QC Blood sample
Installation - preparation
• Environment :
Good power grounding;
Should be free from dust, mechanical vibrations and electrical interference;
Far from brush-type motors, flickering fluorescent lights;
Do not place the analyzer on a slope;
Do not place the analyzer in direct sunlight.
The environment shall be well ventilated.
Temperature: 15-30℃.
Humidity: 20%-85% (operation status).
Atmospheric pressure: 70-106KPa (operation status).
UPS is suggested if the power usually shutdown.
Power stabilizer is suggested if power voltage is unstable.
Installation - Appearance checking
Check the appearance of the packaging box, if there are any destroy or deformation for packaging box or not. If
have, please take a photo and send to assigned service engineer of our company. After appearance checking,
then remove the packing string , then open the carton box.
Installation - unpacking
Accessary box
Waste container
Take out the instrument from the cargo box, Place the analyzer to flat workbench by handling left &
right side.
Check the appearance of the analyzer main unit, if there are any destroy or deformation for main unit
or not. If have, please take a photo and send to assigned service engineer of our company.
Installation – Space & Power input requirement
Remove
the buffer
Open the right panel, cut off and remove the fixing belt, which fixes sampling system. (Don’t damage the black
conveyor belt.) Remove the black buffer which between the horizontal motor and pathway. Then, remove the
shielding paper from chambers shielding box.
Installation - unpacking
Remove notice label on the power socket Connect the power cord
Remove the Notice label on the power socket and connect the power cord. Ensure the grounding is good. If the
grounding is not good enough (Grounding wire and Null wire voltage is higher than 5V AC), please connect the
grounding wire for the instrument.
Fix grounding wire
Connect reagent tubes / sensor to the corresponding connector (color). White connector is for Diluent solution,
Red connector is for waste. Metal connector is for Waste sensor.
Installation - connection
Lyse is installed inside of the instrument. Open the door of lyse house
Connect lyse tubing.
Installation – connection
Peripheral devices
LAN Port:
Computer (Data Management Software)
LIS
USB Port:
Barcode scanner
Printer
Mouse
Key board
USB port and LAN port USB flash disk
Installation - Install Printer Paper
Insert a new roll Printer Paper into the compartment as shown below:
Ensure the glaze surface face to printer (Thermal printer, 57mm width Printer paper)
Printer
Installation – Power on
Login software
User: admin
Password: admin
Installation – Starting up
Software initialization : Then the analyzer will boot up and have a hardware self-check, fluidics initialization,
fluidics cleaning. If it passes the checking, initialization and cleaning, it will do a background test
automatically.
Installation – Date, Time setting
Enter Management => Setup => General => Time interface to set up the correct date, time and time format.
Installation – Starting up
Loading Reagent
Before delivery, the hydraulic system is empty, so for the first installation, instrument will give reagent empty
alarm. Please enter Reagent => Replace Reagent interface to load corresponding barcode. After apply the barcode,
analyzer will run priming to make the hydraulic system full of reagent.
Installation – Starting up
Reagent loading
When perform loading the reagent, need input the Lyse/Diluent barcode via barcode scanner or input manually.
And the barcode of Lyse local on the bottle of Lyse, the barcode of diluent stick on the cap of diluent (As figure of
above)
Installation – Starting up
If you get HGB alarm, please adjust HGB (Gain) value as following steps.
Enter Management => Setup => System => Gain interface.
Click + or – button to adjust HGB gain value or click 4.5v button to adjust HGB gain value to get HGB blank
value around 4.5±0.1V automatically. Then click OK to save HGB gain.
Installation – Power on
Do one more background test. Make sure blank test results are within range.
Installation – Calibration
Enter Management => Calibration =>Calibrator Cal. interface. Input calibrator information
and target value, then test at least 5 times to finish calibration.
Installation – Quality Control
Enter QC => Setting => New interface. Input QC information, target and limits values,
then test QC accordingly.
Installation – Sample test
Enter Analysis menu. Select sample mode, edit patient information, then can test
sample accordingly.
Installation - Power off
When power off the instrument, please click Shutdown button, then the analyzer will ask for probe cleanser soak.
Finish probe cleanser soak, then instrument indicates to power off the analyzer.
Installation – Transportation & Storage
If the instrument will be kept for a long time such as more than 3 days, it is necessary to enter Management =>
Service => [Drain All] to operate packing procedure.
After click [Drain All] button, follow instrument indication to do the flushing & draining performance. Then fix
sampling system via fixing belt and cover counting chamber like original condition via shielding paper . Finish all above
operation, you can pack the instrument for storage or sending to other place.
05
●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●●
Maintenance
www.zybio.com
Maintenance
Routine Maintenance:
Wipe the appearance of the instrument with neutral detergent, distilled water
or alcohol.
Caution: Avoid using corrosive acids, alkaline, and volatile organic solvent. Only
neutral detergent can be used.
Caution: Avoid wiping the inner of the instrument.
Daily Maintenance:
1. When the instrument shutdown, it will run daily shutdown probe cleanser
cleaning procedure.
2. If we need keep the instrument power on more than 24H, we can set daily
auto cleanser soak time in Management→ Setup→ System → Maintenance
interface. After the first 24H, when system time reach the time we have set,
the instrument will remind user to do cleanser soak.
3. Use soft cotton to clean sample probe with alcohol .
Maintenance
Weekly Maintenance:
Weekly Maintenance:
Monthly Maintenance:
Monthly Maintenance:
1. Check and clean sample probe, incase of it is blocked inside, try dispense the diluent to
check if it is blocked.
2. Clean sample probe swab, in case of crystal affect instrument performance.
3. Clean chambers shielding box, in case of crystal affect instrument performance.
4. Check and clean WBC and RBC chambers and back bath tubing, if necessary replace
back bath tubing.
5. Replace WBC and RBC apertures’ O ring, if performance is not good even finish step 4.
6. Check Vacuum Pressure, clean Vacuum Tank and Vacuum Pump membrane inside.
7. Check Valve function, if necessary, disassemble it and clean valve inside.
8. Check tubing system, if any tube loose or dirty, maintain or replace them.
Maintenance
Mechanical structure
www.zybio.com
Mechanical structure
4 USB ports
LAN port
Power socket & switch
Waste sensor connector
Lyse house
Sampling system
Syringe
Wash set
Sampling system
Vacuum tank
Valves
Sampling system
Motor Driver board
Lyse house
Core board
Main control board
www.zybio.com
Hardware-Main control board
3 - 8: Indicator
SV05 Connects to RBC back bath, sample probe wash set (in) and Valve 2
Big syringe 10ml Aspirate and dispense diluent and lyse, sample, probe cleanser.
(diluent/lyse) probe wash set cleaning and back flush.
Hydraulic system– liquid pump
Function:
Generate vacuum pressure for vacuum tank, support sample probe wash
set(sample probe cleaning), WBC/RBC chamber washing, blood cell counting.
Hydraulic system– air pump
Function:
Mixing WBC & RBC chamber and assist chamber washing.
Hydraulic system– sample probe
Function:
Aspirate & dispense sample or probe cleanser, dispense diluent.
Hydraulic system– sample probe wash set
Function:
Wash sample probe exterior.
Hydraulic system–vacuum pressure sensor
Function:
Monitor vacuum tank pressure. Give alarm if vacuum pressure is out of range.
Test Flow chart
Test Flow chart
Hydraulic system–time sequence
Trouble Shooting
www.zybio.com
Trouble Shooting-Aperture clog
Reagent alarm :
A. Reagent empty
1. Check whether the corresponding reagent bottle is empty.
2. If yes, replace new reagent and load the barcode.
3. If reagent is enough and alarm still exist, enter Management => Status => Volt./Current
interface to check the reagent sensor voltage(0~2.5V means empty, 3.5V~5V means full.).
4. Enter Management => Service => Prime interface, run prime function to avoid bubble in tube.
B. Reagent expired
1. Enter Management => Setup => Reagent menu.
2. Select “One key elimination” to activate clear reagent expired alarm function. Dis-select
“Reagent expired” option and click OK to save.
3. Then we can click the alarm area, on the pop-up dialog box have one key clear alarm function.
Click clear to cancel the alarm and close the dialog box, then we can continue use the reagent.
C. Reagent run out
Reagent use up, need to load a new reagent.
HGB blank abnormal
HGB adjustment:
1. Enter Management=> Setup => System=> Gain.
2. Change HGB gain value to get HGB blank value around
4.5±0.1V, then save HGB gain.
Counting chamber overflow issue