BC-5600 Installation & Notice

Hematology Global Technical Support Department


Service@mindray.com
Version : 1.0
NO.:MXQ-12023-BC-5600

© 2011 Mindray Confidential


OTY-12002 ( 3.0 )
Content

 Preparation
 Configuration
 Installation and Notice

2 © 2011 Mindray Confidential


Preparation

Tools:
 Clipper
 Monkey wrench
 Cross-headed screwdriver
 USB flash disk
 7um standard particle
 Calibrator
 Control
 Fresh blood samples

3 © 2011 Mindray Confidential


Preparation

Space requirement:
 At least 100cm to both left and right sides
 At least 50cm behind
 Enough room on or below the countertop to accommodate the diluent, rinse and waste
containers

4 © 2011 Mindray Confidential


Preparation

Power requirement:

Voltage Frequency

Analyzer AC100 V ~ 240 V 50/60 Hz

AC110/115 V 50/60 Hz
Pneumatic unit
AC220/230 V 50/60 Hz

5 © 2011 Mindray Confidential


Preparation

Environment requirement:
 Optimal operating temperature: 15 ℃-30 ℃
 Running temperature: 10 ℃-40 ℃
 Optimal operating humidity: 30 %-85 %
 Atmospheric pressure: 70 kPa-110 kPa
 Free from dust, mechanical vibrations, loud noises and electrical interference

6 © 2011 Mindray Confidential


Configuration

 Open vial sampling: without auto mixing part, without pierce needle.

7 © 2011 Mindray Confidential


Configuration

 Close tube + Open vial sampling:

8 © 2011 Mindray Confidential


Configuration

 Autoloader + Open vial sampling:

9 © 2011 Mindray Confidential


Installation and Notice

 1. Unpacking
 2. Removal
 3. Connection
 4. Start up and adjustment
 5. Unit test

10 © 2011 Mindray Confidential


Unpacking

 Autoloader Configuration
Four packages: main unit, pneumatic unit, autoloader and accessory

 Close tube or Open vial Configuration


Three packages: main unit, pneumatic unit and accessory

11 © 2011 Mindray Confidential


Unpacking

 Main unit

Packing tapes

Top cover

12 © 2011 Mindray Confidential


Unpacking

 Main unit

Accessories of the main unit

13 © 2011 Mindray Confidential


Unpacking

 Main unit

Main unit

14 © 2011 Mindray Confidential


Unpacking

 Main unit

15 © 2011 Mindray Confidential


Unpacking

 Pneumatic unit

Packing tapes

16 © 2011 Mindray Confidential


Unpacking

 Pneumatic unit

Accessory kit

Protective corners

17 © 2011 Mindray Confidential


Unpacking

 Pneumatic unit

Top foam

Pressure regulator

Cardboard

Bottom foam

18 © 2011 Mindray Confidential


Unpacking

 Autoloader

Packing tapes

19 © 2011 Mindray Confidential


Unpacking

 Autoloader

Cardboard

Tube rack

20 © 2011 Mindray Confidential


Unpacking

 Autoloader

Autoloader stop
bar

Foam

21 © 2011 Mindray Confidential


Unpacking

 Accessory

Packing tapes

22 © 2011 Mindray Confidential


Unpacking

 Accessory

Accessories

23 © 2011 Mindray Confidential


Unpacking

 Accessory

Cardboard Operator’s Data


manual management
software

24 © 2011 Mindray Confidential


Removal

 Main unit
Remove the foam protecting the sample probe

Protecting foam

25 © 2011 Mindray Confidential


Removal

 Main unit
Remove the Adhesive Tapes using to Fix the Communication Cable of the Autoloader

Adhesive tapes

26 © 2011 Mindray Confidential


Removal

 Main unit
Remove the Adhesive Tapes using to Fix the Stop Bar of the Front Cover

Adhesive tapes

27 © 2011 Mindray Confidential


Removal

 Main unit
Remove the Plastic Cable Tie using to Fix the Wipe Block

Plastic cable tie

28 © 2011 Mindray Confidential


Removal

 Main unit
Remove the Plastic Cable Tie using to Fix the Cylinder and Clppier

Plastic
cable tie

29 © 2011 Mindray Confidential


Removal

 Main unit
Remove the spacing blocks in the SRV

Clean SRV

30 © 2011 Mindray Confidential


Removal

 Pneumatic unit
Remove the Fixing Screws on the Bottom of the Pneumatic Unit

Fixing screw

31 © 2011 Mindray Confidential


Removal

 Autoloader
Remove the adhesive tape on the autoloader and cut off the plastic cable ties fixing the
accumulation tray and the stop bars of the autoloader

32 © 2011 Mindray Confidential


Removal

 Autoloader
Remove the plastic cable tie fixing the longitudinal feeding pusher

Autoloader Cut the plastic


cable tie

33 © 2011 Mindray Confidential


Removal

 Autoloader
Remove the two screws fixing the protective cover

Remove and
keep them for
use later

34 © 2011 Mindray Confidential


Connection

 Installing the autoloader


Install the Stop Bars

Location hole

Put the barrier behind the


location hole

Insert the two stop bars into the slots on the bottom plate of the main
unit. Then fix the bars onto the bottom plate with two cross-pan head
screws

35 © 2011 Mindray Confidential


Connection

 Installing the autoloader


Connect the Communication Cable

The cables should be put into the gap in


case of pressed

36 © 2011 Mindray Confidential


Connection

 Installing the autoloader


Fix the autoloader

Do not tightly fix the screws, since the autoloader may need
further adjustment

37 © 2011 Mindray Confidential


Connection

 Connect the Reagents and Waste Container

38 © 2011 Mindray Confidential


Connection

 Connect the Reagents and Waste Container

Installing the supporting board for


diluent container

39 © 2011 Mindray Confidential


Connection

 Connect the Pneumatic Unit

40 © 2011 Mindray Confidential


Connection

 Connect the Optional Equipment

41 © 2011 Mindray Confidential


Connection

 Connect to the Power Source


Main unit: AC100-240V
Pneumatic unit: AC110/220V

42 © 2011 Mindray Confidential


Start up and adjustment

 Check before Starting the Analyzer

Check that the tubes of the wipe block move


smoothly and are not worn.

Check that the SRV tubes do not get loose


and are not worn.

43 © 2011 Mindray Confidential


Start up and adjustment

 Check before Starting the Analyzer

Check that the tubes of the autoloader


are not worn. They do not get loose;
interfere with the front cover or with the
moving parts.

44 © 2011 Mindray Confidential


Start up and adjustment

 Turn on
Turn on pneumatic unit first, then main unit; press Power Button to start up

45 © 2011 Mindray Confidential


Start up and adjustment

 Turn on
If error occurs during startup, the analyzer will not run the fluidics initialization, and a dialog box will pop
up. Click "No", and then go to the “Service>Log" screen to check the error Log.

Service engineer ID: service


Password: Se s700

46 © 2011 Mindray Confidential


Start up and adjustment

 Adjust the Position of Autoloader

Loosen the two screws to adjust the position

47 © 2011 Mindray Confidential


Start up and adjustment
 Adjust the Position of Autoloader

Put an empty tube rack on the autoloader


platform, and access Service->Debug menu;
first “Initialize”, then click “Load”, then
“Feed” to let the rack go step by step.
Don’t feed the tube rack manually.

48 © 2011 Mindray Confidential


Start up and adjustment

 Adjust the Position of Autoloader

Click the "Pinch cylndr fwd" button in the


"Manipulater" area to move the manipulator to
the pinching position and check if it is right above
the tube position; move the autoloader left and
right on the stop bar to adjust

49 © 2011 Mindray Confidential


Start up and adjustment

 Autoloader Test

Check whether the tube could be caught by the


clipper properly.

Check whether the piecing needle is in the center of


tube cap.

Check whether the tube could be put back into the


rack smoothly.

Check whether it is normal when unloading the


tube rack.

50 © 2011 Mindray Confidential


Start up and adjustment

 Fix the Protective Cover of the Autoloader

Fix the protective cover of the autoloader


by using the two screws removed
previously

51 © 2011 Mindray Confidential


Start up and adjustment

 Adjust Inner barcode scanner (possible)

52 © 2011 Mindray Confidential


Start up and adjustment

 Adjust Inner barcode scanner (possible)

Horizontal position
Vertical position

53 © 2011 Mindray Confidential


Start up and adjustment

 Adjust Pierce depth (possible)

Air hole

Sampling hole

54 © 2011 Mindray Confidential


First loosen
this screw; then
fasten after
adjustment

Moving outside, piecing needle


goes inside

Sample out
Air hole connector
55 © 2011 Mindray Confidential
Start up and adjustment

• Adjust Pierce depth (possible)

First loosen
this screw; then
fasten after
adjustment

Moving outside, piecing


needle goes inside

56 © 2011 Mindray Confidential


Start up and adjustment

• Adjust Pierce depth (possible)

Make sure the


sampling hole could
be seen

57 © 2011 Mindray Confidential


Start up and adjustment

• Keep this box to prevent any leaking

58 © 2011 Mindray Confidential


Unit test

 Background Test
Requirements:
WBC ≤0.3
HGB ≤1
RBC ≤0.03
HCT ≤0.5
PLT ≤10

59 © 2011 Mindray Confidential


Unit test

 Confirm the optical system


Calibrator: used to check the LAS/MAS CG position
At the "Count" screen, select the mode "OV-WB" and "CBC+5DIFF", and then run the calibrator three times (recommended). Select the
first analysis record of calibrator in the “Review”, and click "OpAdjust".

60 © 2011 Mindray Confidential


Unit test

 Confirm the optical system


Calibrator: used to check the LAS/MAS CG position
Select "Calibrator" mode and click "Calculate". Record the "LAS/MAS CG Position" of the particle with larger "LAS 0.1max Number",
and then check the "LAS/MAS CG Position" of the other 2 analyses .

61 © 2011 Mindray Confidential


Unit test

 Confirm the optical system


Calibrator: used to check the LAS/MAS CG position
Calculate the mean values of "LAS/MAS CG Position" of the 3 analyses, and make sure: LAS CG position mean within ±2.80 from the
target; MAS CG position mean within ±3.20 from the target.

62 © 2011 Mindray Confidential


Unit test

 Confirm the optical system


Calibrator: used to check the LAS/MAS CG position
If the mean values are out of the range, enter the "LAS CG Target" and "MAS CG Target" and click ">>>" to get the gain values for the
1st calibrator analysis (note down the new gains).

Select the right Particle


group as the previous pages

If you set the new gains just


by 1 calibrator test, not
three times tests as
recommended, you could
directly click this button to
apply the new gains

63 © 2011 Mindray Confidential


Unit test

 Confirm the optical system


Calibrator: used to check the LAS/MAS CG position
Configure gain of the other 2 analyses in the same way. Calculate the mean values of the LAS gain and MAS gain of the 3 analyses as
the final gains. Click "Setup"→"Gain", and enter the final gains in the "WBC_LAS(DIFF)" and "WBC_MAS(DIFF)" fields. In the
"Corresponding Factor of DIFF and BASO" area, enter 1.87 in the LAS field and 1.48 in the MAS field.

After setting DIFF gains,


use the factors to set the
BASO gains automatically

64 © 2011 Mindray Confidential


Unit test

 Confirm the optical system


Control: If the calibrator is not available, use the normal control instead; the operation is the
same, while just having different target values.

The target value of LAS/MAS CG Position is different according to the lot no.. Please contact Mindray service engineer
to get the right one.

65 © 2011 Mindray Confidential


Unit test

 Confirm the optical system


7um standard particle: use to check the LAS/MAS 0.1max width
Go to the "Gain" setup screen. Record the gain of WBC_MAS(DIFF), and enter "20" in the WBC_MAS(DIFF) field.

66 © 2011 Mindray Confidential


Unit test

 Confirm the optical system


7um standard particle: use to check the LAS/MAS 0.1max width
At the "Count" screen, run the 7um standard particles in the OV-WB-CBC+5DIFF mode, and then go to the "Review" screen. Select the
analysis record and click "OpAdjust". Click "Calculate", and check the particle with larger "LAS 0.1max Number“.

The requirements: LAS


0.1max Width≤9;
MAS 0.1max Width≤13

67 © 2011 Mindray Confidential


Unit test

 Confirm the optical system


7um standard particle: use to check the LAS/MAS 0.1max width
After check, go to the gain setup screen, and enter the gain value you've just recorded in the WBC_MAS(DIFF) field.

68 © 2011 Mindray Confidential


Unit test

 Normal Fresh Sample test


Select Normal Fresh (collected no more than 4 hours before) anticoagulated venous blood
samples, and run the samples in the OV-WB mode. The normal Scattergrams shall be like
below:

69 © 2011 Mindray Confidential