Professional Documents
Culture Documents
5 Diff Principle
Optical System Structure
Optical System Adjustment
Troubleshooting
Neutro
Eos RBC
Mono
PLT
Baso
Lymph
After Lysing
6 © 2011 Mindray Confidential
What is 5-part diff?
Neutrophil 50 ~70%
Basophil 0 ~1%
Eosinophil 0.5 ~ 3%
Hydrodynamic Focusing
3-diff analyzer
• Sheath flow surrounds and forces blood cells suspended in dilution to pass through the
center of flow cell one by one.
• Blood cells pass through the Laser spot of flow cell with high speed.
20~30 um
11 © 2011 Mindray Confidential
Multi-angle Laser Scatter
3 4
Low Angle Scatter 2~5
reflects the cell volume
1 Sample
2 Laser light
3 Lens
1
12 © 2011 Mindray Confidential
Laser Scatter: Mie Scatter
Mie scatter:
Cells are bigger than the laser wave length.
Laser
LAS
The low angle scatter signal (2~5) along the direction of incidence light is called
forward scatter.
It reflects the volume of cell.
Laser
MAS
The high angle (8~20) scatter signal along the direction of incidence light is called side
scatter.
It is related to cellular contents (granularity, nuclear content, and so forth ). It reflects
the internal complexity of cell.
17 © 2011 Mindray Confidential
Flow Cytometry + Laser Scatter
Blood
LEO(I) RBC
LEO(II)
Morphological treatment
SIZING STAIN
Eos
Neu
Blood
LBA RBC
Morphological treatment
SIZING
Eos
Neu
Baso
Horizontal axis: High angle scatter signal reflects the internal complexity of cell.
Vertical axis: Low angle scatter signal reflects the volume of cell.
Neu
Baso
Mono
Lymph
Total
WBC
Eos
Lymph
Neutro
Eos
Mono
Small size Shrink greatly.
Then get
Lymph#, Mon#, Eos#, Neu#
board
Laser control
MAS PD Assembly
Assembly
LAS PD
LAS Beam Stop
MAS Beam Stop
Flow Cell
Assembly
small
Semiconductor – more stable, low quantity
of heat
long life and low cost for maintenance
Generate stable sample flow; make blood cell in the sample flow pass trough laser spot
(irradiation spot) one by one.
First collimate the scatter light, then use prisms to separate the light into two directions,
and use beam stop in each direction to collect the light at certain angle.
• Block the scatter light at certain angle, and then focus onto the PD.
1. Install the back lens assembly. Rotate the forward assembly to let it be parallel with
back lens assembly
Pitch Pin
1
44 © 2011 Mindray Confidential
Adjust Alignment
2. Adjust flow cell assembly (1): Rotate the flow cell to let it be parallel with the forward
assembly.
Symmetric
3
3
4. Move back lens assembly to let the light spot in the middle of the beam stop
5. Move flow cell into optical path, until you see two clear lines and make them in the
middle as below
7. Adjust PD assembly
Tools:
FLUKE124
Test points:
LAS TP43---LASIN
MAS TP44---MASIN
Group 1
Group 2
Eos
Lymph Baso
4 Parts
Eos
Lymph Baso
DIFF BASO
Leucopoenia Eosinophilia
ATL LIC
The target value of LAS/MAS CG Position is different according to the lot no.. Please contact
Mindray service engineer to get the right one.
If the normal control is not available, use the normal fresh blood to adjust the gain
generally.
Forward optical
assembly First loosen these two screws
Flow cell
First,
Check reagents, Clean the WBC counting channel with software
Second,
Identify the situation according to scattergram
Hardware failed?
Dirty/Clog?
Improper gain?
To be continued in the Liquid System