You are on page 1of 46

Troubleshooting in

HPLC analysis

Spinco Biotech Pvt. Ltd.


General Troubleshooting

Troubleshooting
 Mobile phase
 Sample
 Injection
 Column
 Detection
General Troubleshooting
Mobile phase Solvent
Grade and Treatment

Absorptio
deionized
Water grade water
Pure water : Organic and Ion pure

n
Distilled water : Organic water
Deionized water : Ion Wavelength (nm)

Millipore or Barnstead system is ideal as they remove inorganic and


organic contaminants.
Modern pure water systems give TOC values as indication of
total organic content in water.
General Troubleshooting
Solvent Grade and Treatment

Difference of Analytical and HPLC Grade

Methanol Acetonitrile Hexane


General Troubleshooting
Solvent Grade and Treatment

Ghost peak will be appeared in gradient elution if water is


of bad quality!

H2O / MeOH r ve
Cu
gradient nt
ie
r ad
G
ODS Column

1 ml/min, 0-100% MeOH


over 10 mins and hold for
15 mins. Ghost Peaks
General Troubleshooting
Solvent Grade and Treatment

HPLC Grade

HPLC Grade solvents only guarantee low UV


absorbing impurities.
It does not mean it is free of all impurities.

-- All solvents to be used should be filtered through at least


0.45um membrane filters before use.
If the solvent has background absorption,

-- linear range will be smaller,


-- noise will be bigger and
-- baseline stability especially during gradient
runs will be poor.
General Troubleshooting
Solvent Grade and Treatment

Some precautions in handling HPLC Grade organic solvents :

1. THF

Analytical grade THF,


contains butylated hydroxytoluene(BHT) as an oxidation
inhibitor. BHT has strong UV absorption.

HPLC grade THF,


It does not contain BHT.
HPLC grade THF must be stored in cool and dark
places to reduce oxidation to form peroxides.

Accumulation of large amount of organic peroxides may


lead to explosions.
General Troubleshooting
Solvent Grade and Treatment

2. Chloroform

Analytical grade chloroform contains 3% ethanol to


prevent generation of phosgene, a highly toxic gas.
However, ethanol, like water, may deactivate silica gel used
in normal phase separation and thus lead to unstable
retention times.

HPLC Grade chloroform therefore does not contain ethanol


as a stabiliser. Once seal is opened, it should be stored in
dark and cool places to reduce the possibility of generation
of phosgene.
General Troubleshooting
Solvent Grade and Treatment

Use of Buffer Solutions

Solutions must be filtered before use.


Must not be left in system to avoid crystallization :
 effect on pump -- damage plunger and seal
 effect on column -- creation of column voids
 effect on flow-line -- corrosion of stainless steel lines
Possible bacterial growth
especially phosphate buffers -- good medium
for bacterial and fungus growth.

Ideally, solutions should be prepared fresh everyday.


General Troubleshooting
Maintenance and
Troubleshooting Solvent Grade and Treatment

Solvent Filtration Kit


General Troubleshooting
Solvent Grade and Treatment

Changing Over of Mobile Phase

Buffer solution
Water

Aqueous Organic Solvent


( methanol, acetonitrile etc. )
2-propanol

Non-aqueous Organic Solvent


( hexane, chloroform )
General Troubleshooting
Solvent Grade and Treatment

Pipe
Do not

20 cm
touch
directly

Suction Filter
rinse suction filter
General Troubleshooting
Maintenance and
Troubleshooting Solvent Grade and Treatment

Premixing of Solvents
For isocratic systems, we often premixed the solvent mixtures
used for analysis.
E.g. To prepare 100ml of 50:50 mixture of water/methanol,
we may :
1. Measure out 50ml of water and 50ml of methanol separately
then mix them together.
2. Measure 50 ml of water and top up to 100ml mark of
measuring cylinder/volumetric flask with methanol.
3. Measure 50ml of methanol and top up to 100ml with water.

All three methods will give different compositions. Whichever


method selected must be used consistently for repeatable results.
General Troubleshooting
Solvent Reservoir Problems

1. Containers should generally be glass as plastics will


leach plasticers leading to unstable baseline and/or
contamination of columns.
Glass should be avoided when carrying out high sensitivity
analysis if inorganic ions as these could be leached from
glass.

2. Containers should have small covered openings. This is to


prevent rapid evaporation of solvents leading to unstable
compositions. It also serves to reduce the vapours in the
room. Do not make containers airtight as this may lead
to creation of partial vacuum and pump starvation.
General Troubleshooting
Solvent Reservoir Problems

3. When using stabiliser free solvents, either change fresh


solvents daily or provide inert atmosphere for solvent
reservoir such as use of helium degasser.

4. Do not position solvents directly in line of drafts from


ventilation ducts, windows, doors or hoods; do not position
where it may be exposed to direct sunlight. This may lead
to temperature gradients and/or increased evaporation.
Effect will be greatest for detectors like RI detectors but
sometimes may also affect UV detectors.

5. Always keep reservoir above solvent delivery system as it


ensures good siphon feed to pump and prevents starving
pumping system.
General Troubleshooting
Solvent Reservoir Problems

6. Always use a suction filter. This is course filter for solvent


and helps to prevent particulates from reaching pumps.
Keep these filters clean and free from fungus. If dirty,
they may be cleaned by sonicating in beaker of isopropyl
alcohol or 1N nitric acid.

7. Good practice in the lab to label all bottles with composition


inside as well as date of preparation.
General Troubleshooting
Sample Treatment Problems

1. It is always best to dissolve sample in mobile phase.


-- reduces solvent peak and is important in cases
where peak of interest elute close to solvent peak.
-- ensures solubility of sample in mobile phase so it
does not precipitate in the system, especially in column.
-- ensures miscibility of sample solvent with mobile
phase.

2. Always filter sample with 0.45um filters before injection.


If sample is known to be very dirty, filter through 0.2um
filters before injection.
General Troubleshooting

The cause of quantitative error for


sample pretreatment
 Standard solution
 Measure weight

 Accuracy of balance

 Operation

 Impurity

 Uncertain hydrate number

 Hygroscopic
General Troubleshooting

The cause of quantitative error for


sample pretreatment
 Standard solution
 Measure up by solvent

 Accuracy of apparatuses

 Solubility(temp.)

 Inhomogeneous

 Adsorption( to apparatus)

 E.t.c.
General Troubleshooting

The cause of quantitative error for


sample pretreatment
 Standard solution
 Storage

 Evaporation

 Oxidation

 decomposition
General Troubleshooting

Adsorption of target compounds


on the apparatus
Material Compounds
Glass Cation[Quarternary Decrease pH
ammonium, Metal ions] Add ClO4 ion
Amine

Metal Anion Decrease pH


Chelate compounds

Polymer Hydrophobic
compounds
General Troubleshooting
Injector

Loop type Injector

Two injection methods are possible :

1. Partial Loop Injection Method


-- Inject less than half volume of sample loop.

2. Full Loop Injection Method


-- Inject more than 3 times volume of sample loop.
General Troubleshooting
Injector

Loop type Injector


Relationship between injection volume and detector's response
Response of Detector

Loop injection

Partial
injection
3 times volume
Half volume
of sample loop
of sample loop

Volume of Injection
General Troubleshooting
Injector

Manual Injector -- Rheodyne


Results can be explained in terms of Dispersion and
Dilution effects.
Less than half volume More than half volume
dr dr
a in p a in p
m m
pu p u

co co
lu lu
m m
in n n
r a in
d a
dr
loop loop
Sample Zone
Dilution Zone
Dispersion Zone
General Troubleshooting
Injector

Manual Injector -- Rheodyne

Less than 3 times More than 3 times


dr p dr
a in m a p
pu in m
pu

co co
lu lu
m m
n n
ain a in
dr dr
loop loop
Sample Zone
Dilution Zone
General Troubleshooting
Injector

Auto-Injectors -- General Maintenance

1. Purge flow line at least once a day.

2. Change purge liquid at least once a week.

3. Never use salt solutions as purge liquid.

4. Use only septums recommended by manufacturer.

5. Septums should not be reused.

6. Do not fill vials to the brim.


General Troubleshooting

Auto-injector SIL-10ADVP

27
SIL-10ADvp/LC-2010
General Troubleshooting

Sample Loop Sample Loop


Rinse Rinse
Port Port

Pump Pump

INJ Port INJ Port

Column Column

READY Mode Auto Drain Mode


General Troubleshooting
Column
Initial checks on new columns.
1. Column performance check
Column performance check may be carried out before
column is first used using the test mixture and condition
as stipulated in the column performance sheet.

2. Pressure check
Initial operating pressure of the column under normal
analysis conditions should be noted for reference.

3. Guard Columns
Should be used whenever possible but should not add much
pressure to system.
General Troubleshooting
Column

Precautions During Operation.


1. For new columns, it is best to condition under low
flow rate (0.2-0.3 ml/min) for about 30 mins before operating
under analytical flow rate. Same applies for columns that
have not been used for long time.

2. During shutdown, it is best to gradually reduce flow   rate so as


not to subject column to sudden pressure drop.

3. When using buffer solutions, remember to flush with water


before changing to organic solvents to prevent crystallization.
General Troubleshooting
Column

4. It is best to filter mobile phase and sample before use to


reduce clogging by particulates in frit.

5. Sample is best dissolved in mobile phase or if not possible


then as close to mobile phase composition as possible.
This reduce chances that sample may precipitate in column
bed on mixing with mobile phase.

6. Flush with stronger solvent after daily use. For example,


ODS columns may be flushed with acetonitrile after use
daily to remove strongly absorbed materials from column.
General Troubleshooting
Column
Column Troubles and Solutions

High pressure in system


Poor peak shape (tailing, leading, splitting)
Changes in retention time.

Plugged Frit or Column Packing


Adsorbed Sample & Solvent Impurities
Voids Resulting from Shock or Crystallization
Chemical Attack of Packing Material
Drying up of Packing Material
General Troubleshooting
Column

Column Problems and Solutions -- Plugged Frit or Column Packing

Prevention is by filtering both sample and mobile phase


through 0.45m membrane filter or 0.2m membrane
filter if sample is very dirty.

However, sometimes, precipitation may occur or some dirt


may get into system due to exposure of mobile phase or sample
and this will result in clogging.
General Troubleshooting
Column

Column Problems and Solutions -- Plugged Frit or Column Packing

Normal flushing with string solvent may dissolve and remove


some of these particles but in cases where the particulates are
insoluble, flushing the column in reverse may help.

Note : Do not connect to detector in this case and do not use


this method for routine flushing as it may lead to loosening of
packing material.

Finally, replace filter frit if it cannot be washed. Be careful


not to disturb column bed or performance may drop and
in some cases, the stainless steel frits are fused to the columns
and these cannot be replaced.
General Troubleshooting
Column
Column Problems and Solutions --
Adsorbed Sample & Solvent Impurities

Impurities will cause a tailing peak.


Wash a column using high solubility solvent.

Remove this parts and


repack the packing material.
only 1 cm will
column

be polluted
LC
General Troubleshooting
Column
Column Problems and Solutions --
Voids Resulting from Shock or Crystallization

As HPLC columns are pressure-packed, subjecting them to


pressure shock such as dropping the column or opening of
drain valve during operation may result in formation of voids.

In the event of crystallization, even though the crystals may be


Re-dissolved and was away, the packing pushed aside by their
presence may not return to original place resulting in voids.

Do not open drain valve


LC-10AD :::::::
during operation
000 0 000 0

xxxx xxxx
7 8 9

xxxx xxxx 4 5 6

xxxx xxxx
1 2 3

xxxx xxxx
0 ¥ xxxx
General Troubleshooting
Column

Column Problems and Solutions --


Voids Resulting from Shock or Crystallization

Voids will cause split peaks.

void
Am I seeing
Every peak will be split. double?

Hard to repair!!!
General Troubleshooting
Column
Column Problems and Solutions --
Chemical Attack of Packing Material

Silica gel Polymer


pH range 2 - 7.5 widely use
Organic solvent all solvent limitation
Pressure less than 250 kgf/cm2 low pressure
Temperature better to set possible to set
at less than 60C at high temperature

Caution : Must check the value in terms of


every item of polymer based column
General Troubleshooting
Column

Column Problems and Solutions --


Drying up of Packing Material

Frequently, especially for polymer based columns, swelling


characteristics (shape) of packing depends very much on
solvent. When column is dry, these may be deformed
permanently.

Especially for GPC type columns, the choice of solvent is


used is very specific. Not only must it be kept wet, it must
not be used with wrong type of solvent or whole characteristic
of polymer (shape and pore size) will be changed permanently.
General Troubleshooting
Waste Bottle

Make sure that waste line is not immersed in liquid waste


in
the waste bottle as this may create back pressure on cell.

Detector Detector
General Troubleshooting

Detector cell
Back pressure
0.3mmI.D. x 2mL.
General Troubleshooting
Maintenance and
Troubleshooting
Other Trouble

Peak Shape Distortion -- A. Asymmetry (Leading / Tailing)

Case 1 : development of method


Possible causes : 1. Improper solvent selection (may also result in
peak splitting
Methanol as a sample solvent Ethanol as a sample solvent

20 L Caffeine 20 L Caffeine

Ethanol as a sample solvent Better inject small


amount of sample.
10 L Caffeine
General Troubleshooting
Maintenance and
Troubleshooting
Other Trouble

Peak Shape Distortion -- A. Asymmetry (Leading / Tailing)


SOLUBILITY EFFECT

Low soluble solvent High soluble solvent


General Troubleshooting
Maintenance and
Troubleshooting
Other Trouble
Peak Shape Distortion -- A. Asymmetry (Leading / Tailing)
2. Secondary retention effects : due to heavy metals in silica gel or
residual silanol groups in modified silica gel.
C18
OH Silanol group
C18
silica core
-
OC18 negative charge
C18
C18
C18
OH O-TMS
silica core C18 TMS treatment C18
OH silica core O-TMS
C18 TMS : trimethylsilyl group
C18
C18
C18 [End capping type]
[Non-End capping type]
General Troubleshooting
Maintenance and
Troubleshooting
Other Trouble

Peak Shape Distortion -- A. Asymmetry (Leading / Tailing)

C18
+ O-TMS High pure type of
silicaMcore C18 O silica gel are available.
O-TMS
M+ C18 O
C18

3. Overloading of sample
For every column, no. of binding sites is limited, if sample loading grossly
exceeds the columns capacity, leading or tailing may occur (most probably
tailing).

4. Adjustment of pH of mobile phase.


General Troubleshooting
Maintenance and
Troubleshooting
Other Trouble
Peak Shape Distortion -- A. Asymmetry (Leading / Tailing)

Case 2 : Sudden peak shape distortions in existing methods

1. Column (Guard/Main) failure. -- clean or replace column.


2. Column dissolution may occur for example if high pH (>7) is used with
silica based column.
3. Problems with secondary retention effects may occur if different make
of column is used.
4. In particular, broadening may be caused by dead volumes.

Good

Dead Volume

You might also like