Professional Documents
Culture Documents
Catalog
Installation
Sample test
QC test
Manual calibration
Use up
Installation
• Requirement
• Connected dataline /power supply /reagent tu
bes
• Remove sampling controller
• Install software and connect computer
back
Sample test
• Open mode
• Auto mode
• Worklist mode
SOP
back
QC test
back
Manual calibration
• Manual calibration procedure is as follows:
• 1. In Sample Test menu, use calibration sample to test several times (at
least three).
• 2. Record the tested data.
• 3. Calculate new calibration coefficient
• Calculate new calibration coefficient as per formula below
back
Fluidic diagram
back
Parts introduce&check mechanic
1.sheath flow syringe
2.sample probe assembly
3.mix assembly
4.sample transfer unit
5.Diff reaction bath assembly
6.WBC chamber
7.RBC chamber
8.Negative pressure chamber
9.Positive pressure chamber
10.
Negative pressure pump
11.
Positive pressure pump
12.
Optical assembly
13.
Auto loader panel unit
14.
Ul(sample ) syringe assembly back
15.
Lys syringe assembly
16.
Diluent syringe assembly
17.
Liquid-level detect assembly
18.
Driver panel assembly
19.
Main control panel assembly
Requirement
• Power: AC100V~240V, Input Power ≤250VA, 50/60HZ.
Note:(1)AC power must be well earthed.
(2) AC power shall be stable,we advice customer equip UPS.
• (External) Computer:
CPU: >1.4G
RAM: >2G
Hard disk space available: >20G
Graphics Card: OpenGL 2.0 or above
return
Display aspect ratio: 10: 6
Connected dataline /power supply /reagent tubes
return
Remove sampling controller
return
Install software and connect computer
• Step 1, install software.
• Step 2, change IP address of the PC to 10.0.0.102
• Step 3, open the software and then power on the instrument, after 10 seconds
if the light which is located on the up right side of the screen become green
that means the PC connected with instrument well.
• Step4, prime all reagent 1 or 2 times.
return
sheath flow syringe
return
sample probe assembly
• consist of sampling needle and swab
(sampling needle cleaning head )
• Swab’s function :clean the outer wall of
the sampling needle .
• How to clean the swab :
After cleaning and reassembling the swab
(and ensuring that the inflow/outflow
interface tubing is connected correctly ),turn
on the device and to to “service ----clean
“,double click the “clean the sample probe
“button to perform the operation .
return
mix assembly
return
sample transfer unit
return
Diff reaction bath assembly
• Provide diff1&2 to
optical assembly
• Incubate diff1&2
• How to PM diff bath
Clean diff bath with cotton
swab .
turn on the device and to to
“service ----clean “,double click
the “clean the DIFF bath
“button to perform the
operation .
return
WBC chamber
return
RBC chamber
return
Negative pressure
chamber
• Through the negative pressure
chamber ,WBC/RBC/PLT travel
through the aperture of the
chamber for being counted .
• any waste that counting and from
cleaning the back chamber of the
WBC and RBC flows into the
negative-pressure chamber.
• Provide power for cleaning both
rear chambers.
return
Positive pressure
chamber
• Generate air bubbles in the
DIFF /WBC/RBC
• Release negative chamber
pressure
return
Negative pressure pump
• PP1:drain WBC/Swab
• PP2:drain DIFF /RBC/negative
chamber
• establish a negative pressure of -
30Kpa for negative chamber
return
Positive pressure
pump
return
Optical assembly
return
Ul(sample ) syringe assembly
• Push the sample which be mixed in DIFF to
flow chamber ( 250ul )
• Aspirate the sample from sample tube/WBC
chamber (100ul)
• Push diluted-sample to RBC chamber (100ul)
• Built a bridge for Cleaning inside &outside of
the probe (100ul)
return
Lys syringe assembly
return
Liquid-level detect assembly
return
Driver panel assembly
return
Main control panel assembly
return