Professional Documents
Culture Documents
TO
INSTRUMENTAL METHODS
OF ANALYSIS
7
CLASSIFICATION…
Analytical Techniques
UV/VIS Spec
Electrochemistry Chromatography
IR Spec
Mass Spec o Qualitative—Amount present in sample/mixture
Molecular Rotation
Transition of outer most
Valence Electrons
Spectra originate when atoms, ions, and molecules emit or absorb energy.
Atoms, ions, and molecules emit or absorb characteristically; only certain energies of
these species are possible;
the energy of the photon (quantum of radiant energy) emitted or absorbed
corresponds to the difference between two permitted values of the energy of the
species, or energy levels.
Thus the energy levels may be studied by observing the differences between them.
The absorption of radiant energy is accompanied by the promotion of the species
from a lower to a higher energy level;
The emission of radiant energy is accompanied by falling from a higher to a lower
state; and
If both processes occur together, the condition is called resonance.
INTRODUCTION…
Spectrum—
is, a plot of the intensity of emitted or transmitted radiant energy (or some
function of the intensity) versus the energy of that light
Interpretation of spectra provides
o fundamental information on atomic and molecular energy levels,
o the distribution of species within those levels,
o the nature of processes involving change from one level to another,
o molecular geometries,
o chemical bonding, and
o interaction of molecules in solution.
At the practical level, comparisons of spectra provide
o a basis for the determination of qualitative chemical composition
and chemical structure, and
o for quantitative chemical analysis.
INTRODUCTION…
When light is absorbed by a sample, the irradiance of the beam of light is decreased.
Irradiance, I, is the energy per second per unit area of the light beam.
Light is passed through a monochromator (a prism, a grating, or even a filter) to select one
wavelength.
Light with a very narrow range of wavelength is said to be monochromatic (one color)
The monochromatic light, with irradiance Io, strikes a sample of length b.
The irradiance of the beam emerging from the other side of the sample is I. Some of the
light may be absorbed by the sample, so I ≤ Io.
b, length of path through sample
Light Wavelength Sample Light
Source Selector Io I Detector
(Monochromator)
Transmittance, T, is defined as the fraction of the original light that passes through the sample.
irradiance of beam emerging from sample;
irradiance of beam entering sample;
INTRODUCTION…
Therefore, T has the range 0 to 1.
The percent transmittance is simply 100T and ranges between 0 and 100%.
Absorbance is defined as:
I/Io T% A
1 100 0
0.1 10 1
0.01 0 2
Absorbance is so important because it is directly proportional to the concentration, c,
of the light-absorbing species in the sample
INTRODUCTION…
UV/VIS Spectrophotometry - is technique that uses UV/VIS light to measure analyte
concentrations since Ultraviolet (UV) and visible radiation comprise only a small part of the
electromagnetic spectrum.
The basic principle is that each compound absorbs or transmits light over a
certain range of wavelength.
A procedure based on absorption of visible light is called colorimetry
INTRODUCTION…
The absorption of EMR of wavelengths between 200 and 800 nm by molecules can
be employed for both qualitative and quantitative analysis;.
As a wide variety of pharmaceutical substances absorb radiation in the
near-ultraviolet (200–380 nm) and
visible (380–800 nm) regions of the electromagnetic spectrum,
the technique is widely employed in pharmaceutical analysis.
1. Quantitative:
The relationship between :
the concentration of analyte and
the intensity of light absorbed
is the basis of quantitative applications of spectrophotometry.
2. Qualitative:
Features of absorption spectra such as the
molar absorptivity,
spectral position, and
shape and breadth of the absorption band are
related to molecular structure and environment
can be used for qualitative analysis.
INTRODUCTION…
Important advantages of spectrophotometric methods include:
1- Wide applicability;
large number of organic and inorganic species absorb light in the UV-Visible ranges.
2- High sensitivity;
analysis for concentrations in the range from 10-4 to 10-6 M are ordinary in the
Spectrophotometric determinations.
treatments.
4- Good accuracy;
Relative errors in concentration measurement lie in the range of 0.1 to 2 %.
21
CHROMOPHORE-color bearing
The energy of radiation being absorbed during excitation of electrons from ground state to excited
state primarily depends on the nuclei that hold the electrons together in a bond.
Chromophore—
is the group of atoms containing electrons responsible for the absorption and
exhibits a characteristic absorption spectrum in the ultraviolet or visible region.
Most of the simple un-conjugated chromophores give rise to high energy transitions of little use
Ketones, Imine
Shifting of 𝜆max
EFFECTS of Auxochromes and Conjugated Chromophors
1. Bathochromic Shift or Red shift:
A shift of an absorption maximum towards longer wavelength or lower energy.
2. Hypsochromic Shift or Blue Shift:
A shift of an absorption maximum towards shorter wavelength or higher energy.
3. Hypochromic Effect:
An effect that results in decreased absorption intensity.
4. Hyperchromic Effect:
An effect that results in increased absorption intensity.
In general it should be possible to predict the effect of non-polar or weakly polar auxochromes, but the effect
of strongly polar auxochromes is difficult to predict. In addition, the availability of non-bonding electrons
which may enter into transitions also contributes greatly to the effect of an auxochrome
ABSORPTION INTENSITY
Molar absoptivities may be very large for strongly absorbing chromophores (>10,000) and
very small if absorption is weak (10 to 100).
The magnitude of ε reflects both
the size of the chromophore and
the probability that light of a given wavelength will be absorbed when it
strikes the chromophore.
A general equation stating this relationship may be written as follows:
ε = 0.87 x 1020 I x a
A Λmax. At pH 6= 274nm
Λmax. At pH 13= 295nm
2000
2000
250
250 270
270 290
290 310
310
Wavelength(nm).
EFFECTS of pH on ABSORPTION…
saturated hydrocarbons—
do not interact with solute molecules
either in the ground or excited state and
the absorption spectrum of a compound
in these solvents is similar to the one in a
A
pure gaseous state.
Polar solvents such as water, alcohols etc. may
stabilize or destabilize the molecular orbitals Solvents water
In case of π π* transitions—
the excited states are more polar than the ground state and
the dipole-dipole interactions with solvent molecules lower the energy of the
excited state more than that of the ground state.
Therefore a polar solvent decreases the energy of π π* transition and
absorption maximum appears ~10-20 nm red shifted in going from hexane to
ethanol solvent.
SOLVENTS EFFECTS…
n π* Transitions
In case of n π* transitions,
the polar solvents form hydrogen bonds with the ground state of polar molecules
more readily than with their excited states.
Therefore, in polar solvents the energies of electronic transitions are increased.
Example,
the absorption maximum of acetone in hexane appears at 279 nm which in water
is shifted to 264 nm, with a blue shift of 15 nm
ABSORPTION LAW
Beer’s Law
the intensity of a beam of parallel monochromatic radiation decreases exponentially
with the number of absorbing molecules.
More simply, it stated that absorbance is proportional to the concentration (c) of the
solution.
Beer demonstrated the following relationship between absorbance and concentration
of a solution:
Absorbance is defined as the amount of light which is absorbed by the substance and
is calculated as the negative logarithm of transmittance:
ABSORPTION LAW…
Lambert Law
When a beam of light is allowed to pass through a transparent medium, the rate
of decrease of intensity with the thickness (b) of medium is directly proportional
to the intensity of light.
Mathematically:
A = εbc
Specific absorbance is—
the absorbance of a substance in solution of specific concentration in a cell of
specific path length. The most commonly used term for specific absorbance is
A1%1cm, which is the absorbance of a 1 g/100ml (1%) solution in a 1cm cell at a
particular wavelength of light. The beer-lambert equation in such case becomes:
A = A1%1cmbc
ABSORPTION LAW…
Limitations and Deviations of Beer-Lambert Law
Beer-Lambert’s law proves a direct correlation between the absorbance (A) of a
molecule to the concentration (c) and the path length (b) of the sample.
This relationship is a linear for the most part.
However, under certain circumstances the Beer Lambert relationship breaks down
and gives a non-linear relationship.
These deviations from the Beer Lambert law can be classified into three categories:
1. Real Deviations –
are fundamental deviations due to the limitations of the law itself.
2. Chemical Deviations–
are deviations observed due to specific chemical species of the sample which is being
analyzed.
3. Instrument Deviations –
are deviations which occur due to how the absorbance measurements are made.
ABSORPTION LAW…
Conjugated Dienes….
The presence of conjugate double bond decreases the energy difference between
HOMO and LUMO of resulting diene.
As a result, the radiations of longer wavelength are absorbed.
increases the intensity of absorption.
Therefore, the increase in size of the conjugated system gradually
shifts the absorption maximum (λmax) to longer wavelength and
also increases the absorption.
Example,
Ethylene absorbs at 175 nm (ε = 1000) and the conjugation in butadiene gives a strong
absorption at longer wavelength at 230 nm and with higher intensity (ε = >1000).
The presence of alkyl substituents on double bond also produces
bathochromic shift and hyperchromic effect.
These effects are additive in dienes and up to some extent in trienes.
CONJUGATED DIENES, TRIENS & POLYENES and WOODWARD & FIESER’S RULE..
Conjugated Dienes…
Conjugated Dienes, ….
CONJUGATED DIENES, TRIENS & POLYENES and WOODWARD & FIESER’S RULE..
Conjugated Dienes, …
Example,
Here the absorption maxima for dienes 1 and 2 have been calculated according to Woodward rules.
The comparison of calculated λmax values with observed λmax values highlights the importance of
these rules.
β – Carotene, responsible for red color in carrots is a typical example of polyene with 11
conjugated double bonds and exhibits λmax at 445 nm
INSTRUMENTATION: Internal Parts
INSTRUMENTATION
The instruments used in spectroscopy consist of several common components,
including
1. a source of energy that can be input to the sample,
2. a means for isolating a narrow range of wavelengths,
3. a detector for measuring the signal, and
4. a signal processor to display the signal in a form convenient for the analyst
INSTRUMENTATION…
I. Sources of Energy:
All forms of spectroscopy require a source of energy.
a l yt e
In absorption and scattering spectroscopy this energy is t he an igher
supplied by photons. m ote ble, h
ro ta
Emission and luminescence spectroscopy use to p less s te
oa sta
thermal, radiant (photon), or t y
rg
chemical energy ene
b. an effective bandwidth, and The width of the radiation at half the maximum throughput
c. a maximum throughput of radiation.
Effective bandwidth - defined as the width of the band of radiation passing through a wavelength
selector measured at half the band’s height
The ideal wavelength selector has a
high throughput of radiation and
a narrow effective bandwidth.
1. High throughput is desirable because more photons pass through the wavelength selector,
giving a stronger signal with less background noise.
2. Narrow effective bandwidth provides a higher resolution, with spectral features separated by
more than twice the effective bandwidth being
resolved.
Resolution— the separation between two spectral features, such as absorption or
emission lines.
INSTRUMENTATION…
Generally these two features of a wavelength selector are in a
opposition.
Conditions favoring a higher throughput of radiation usually
provide less resolution.
Decreasing the effective bandwidth b
Radiation from the source enters the monochromator through an entrance slit. The radiation is
collected by a collimating mirror, which reflects a parallel beam of radiation to a diffraction
grating. The diffraction grating is an optically reflecting surface with a large number of parallel
grooves. Diffraction by the grating disperses the radiation in space, where a second mirror
focuses the radiation onto a planar surface containing an exit slit. In some monochromators a
prism is used in place of the diffraction grating.
INSTRUMENTATION…
The choice of which wavelength exits the monochromator is determined by rotating the
diffraction grating. 1. smaller effective bandwidth and
A narrower exit slit provides a 2. better resolution, but
3. allows a smaller throughput of radiation.
Monochromators are classified as
1. either fixed-wavelength
2. or scanning.
1. In a fixed-wavelength monochromator, the wavelength is selected by manually rotating the
grating.
Normally, a fixed-wavelength monochromator is only used for quantitative analyses
where measurements are made at one or two wavelengths.
2. A scanning monochromator includes a drive mechanism that continuously rotates the
grating, allowing successive wavelengths to exit from the monochromator.
are used to acquire spectra and,
when operated in a fixed wavelength mode, for quantitative analysis.
INSTRUMENTATION…
3. Interferometers
A device that allows all wavelengths of light to be measured simultaneously, eliminating the need for a
wavelength selector.
Radiation from the source is focused on a beam splitter that transmits half of the radiation to a fixed
mirror, while reflecting the other half to a movable mirror. The radiation recombines at the beam
splitter, where constructive and destructive interference determines, for each wavelength, the intensity
of light reaching the detector. As the moving mirror changes position, the wavelengths of light
experiencing maximum constructive interference and maximum destructive interference also changes.
The signal at the detector shows intensity as a function of the moving mirror’s position, expressed in
units of distance or time.
In comparison with a monochromator, interferometers provide two significant advantages.
results from the higher throughput of source radiation- improved signal-to-noise ratio
savings in the time needed to obtain a spectrum- an entire spectrum can be recorded in approximately 1 s,
as compared to 10–15 min with a scanning monochromator
INSTRUMENTATION…
III. Sampling cells and compartments :
rectangular
Sampling cells
Cuvettes
can be round, square or rectangular in section and
constructed from glass, silica or plastic.
Square or rectangular cuvettes have a constant light path, the most usual being 1 cm in length.
1. For the visible region, compartments composed of simple pyrex glass is sufficient
since pyrex will absorb in the UV
square
but will not absorb in the visible region.
round
are suitable for use between 320 and 950 nm.
2. For UV region, silica (quartz) cuvettes are used below 320 nm and
they must be clean and free of scratches.
This makes the sample compartments very expensive.
Good practice is to fill all cuvettes with distilled water and measure the absorbance for each against a
reference blank over the wavelengths to be used.
This value should be essentially zero.
Moulded polystyrene disposable cuvettes are available for use in the visible range and part of the
near UV.
INSTRUMENTATION…
IV. Detectors
Compared with techniques such as IR spectroscopy, which produces many narrow bands,
UV-visible spectroscopy provides a limited amount of qualitative information.
However, the presence of an absorbance band at a particular wavelength often is a good
indicator of the presence of a chromophore.
Useful information about substance can be obtained via examination of its max
and emax, which could be correlated with the structural features
The spectrum is a physical constant, which along with melting & boiling
characterization of compounds
through
and
b. Absorbance ratio
62
QUALITATIVE …
2- Confirmation of identity…
b. Absorbance ratio
Absorbance ratio of a given drug at two different wavelength is constant, provided
that—
Beer’s Law is obeyed at the selected wavelengths
63
QUALITATIVE …
3- Detection of some functional groups:
Example,
An absorption band at about 280-290 nm, which is displaced toward shorter
wavelength with increasing solvent polarity strongly, indicates the presence of
aromatic carbonyl group.
Confirmation of presence of aromatic amine or phenolic structure may be
obtained by testing the pH effect on their spectra
NH2 +
NH3
+ H+
- H+
OH O O
-
OH
H+
d g b a
C = C – C = C – C = O enones
• a -Alkyl cause red shift about 10 nm and a -OH about 35 nm
• b -Alkyl cause red shift about 12 nm and b -OH about 30 nm
• g -Alkyl cause red shift about 18 nm and g -OH about 50 nm
Enone-
also called an α,β-unsaturated carbonyl, is a type of organic compound consisting of an
alkene conjugated to a ketone.
Qualitative …
6- Empirical Rules for Absorption Wavelengths of Conjugated Systems
Carbonyl Compounds
Aromatic Compounds
The simplest aromatic compound is benzene.
It shows
two primary bands at 184 (ε = 47,000) and 202 (ε = 7400) nm and
a secondary fine structure band at 255 nm (ε = 230 in cyclohexane).
Substituents on the benzene ring also cause-
bathochromic and hypsochromic shifts of various peaks.
Unlike dienes and unsaturated ketones, the effects of various substituents on the
benzene ring are not predictable.
However, qualitative understanding of the effects of substituents on the
characteristics of UV-Vis spectrum can be considered by classifying the substituents
into
electron-donating and electron-withdrawing groups
6. Empirical Rules for Absorption Wavelengths of Conjugated Systems…
Aromatic Compounds…
The non-bonding electrons increase the length of π-system through resonance and
shift the primary and secondary absorption bands to longer wavelength.
More is the availability of these non-bonding electrons, greater the shift will be.
In addition, the presence of non-bonding electrons introduces the possibility of n
π* transitions.
If non-bonding electron is excited into the extended π*chromophore, the atom from
which it is removed becomes electron-deficient and the π-system of aromatic ring
becomes electron rich.
This situation causes a separation of charge in the molecule and such excited state is
called a charge-transfer or an electron-transfer excited state
6. Empirical Rules for Absorption Wavelengths of Conjugated Systems…
On the other hand, in the case of anilinium cation, there are no n electrons for
interaction and absorption properties are quite close to benzene.
But in aniline, the primary band is shifted to 232 nm from 204 nm in anilinium cation
and the secondary band is shifted to 285 nm from 254 nm
6. Empirical Rules for Absorption Wavelengths of Conjugated Systems…
Aromatic Compounds…
Effect of π Conjugation
Conjugation of the benzene ring also shifts the primary band at 203.5 nm more effectively to
longer wavelength and secondary band at 254 nm is shifted to longer wavelength to lesser extent.
In some cases, the primary band overtakes the secondary band.
For example, in benzaldehyde, the secondary band appears at 282 nm and primary band at
242 nm but in case of cinnamaldehyde, primary band appears at 281 nm and remains
merged with secondary band (figure 10).
The hyperchromic effect arising due to extended conjugation is also visible
6. Empirical Rules for Absorption Wavelengths of Conjugated Systems…
Aromatic Compounds…
Aromatic Compounds…
naphthalene
The primary band at 184 nm in benzene shifts to
220 nm in case of naphthalene and
260 nm in case of anthracene.
Similarly, the structured secondary band which appears as broad band around 255 nm
in benzene is shifted to
270 nm and in case of naphthalene
340 nm in case of anthracene molecules anthracene
QUANTITATIVE
SPECTROPHOTOMETRY
UV/VIS spectrophotometry is perhaps the most widely used spectrophotometric
technique for the quantitative analysis of chemical substances.
It has found increasing usefulness as a means of assaying pharmaceutical
substances described in the pharmacopeias.
In general, about 90% of all the quantitative determinations are performed by
spectroscopic techniques.
In the field of health alone, 95 % of all quantitative determinations are
performed by UV-Visible spectrophotometer and similar techniques
The use of UV-visible spectrophotometry in quantitative analysis is centered
around the Beer-Lambert law, which relates the absorbance of an analyte to its
concentration.
QUANTITATIVE …
A. Assay of single components
The assay of an absorbing substance may be quickly carried out by preparing a solution
in a transparent solvent and measuring its absorbance at a suitable wavelength.
The concentration of the absorbing species is then calculated from the measured
absorbance using the following procedures.
Example 1:
1. Calculate the concentration of methyl testosterone in an ethanolic solution of which the
absorbance in a 1 cm cell at its λmax, 241 nm, was found to be 0.890. The Specific
Absorptivity in the B.P. is 540 at 241 nm. (0.00165 g/100ml)
It is the procedure adopted in many spectrophotometric assays of the BP and for the
majority of spectrophotometric assays of USP.
Occasionally, a linear but non proportional r/nship b/n concentration and absorbance
occurs, which is indicated by a significant negative or positive intercept in a beer’s law
plot.
Quantitative …
A. Assay of single components…
Astd1-Astd2
Example:
The λmax of ephedrine HCl and Chlorocresol are 257 nm and 279 nm respectively and the
specific absorptivity values in 0.1M HCl solution are:
Ephedrine HCl at 257 nm = 9.0 Chlorocresol at 257 nm = 20
Ephedrine HCl at 279 nm = 0 Chlorocresol at 279 nm = 105
Calculate the concentration of ephedrine HCl and Chlorocresol in a batch of Ephedrine
HCl injection, diluted 1 to 25 with water, giving the following absorbance values in 1 cm
cells.
= (Aalk+Aint ) - (Aacid+Aint)
= Aalk - Aacid
∆A = ∆ε .b. C
• If the substance is not affected by pH, it can be quantitatively converted by means of a suitable reagent
to a chemical species that has d/t spectral properties to its un-reacted parent species. 84
QUANTITATIVE …
B. Assay of Substances in Multi-component…
4. Derivative spectroscopy
Derivative spectroscopy uses first or higher derivatives of absorbance with respect to
wavelength for qualitative analysis and for quantification.
If a spectrum is expressed as absorbance,
85
QUANTITATIVE …
B. Assay of Substances in Multi-component…
4. Derivative spectroscopy
A first-order derivative is the rate of change of absorbance with respect to
wavelength.
It passes through zero at the same wavelength as λmax of the absorbance band.
86
QUANTITATIVE …
B. Assay of Substances in Multi-component…
4. Derivative spectroscopy
• Note that the number of bands observed is equal to the derivative order plus one.
Advantages
Derivative spectrum shows better resolution of overlapping bands the fundamental
spectrum and may permit the accurate determination of the λ max of the individual
bands.
It permits discrimination against broad band interferences, arising from turbidity
or non-specific matrix absorption.
Thus, the information content of a spectrum is presented in a potentially more
useful form, offering a convenient solution to a number of analytical problems, such
as resolution of multi-component systems, removal of sample turbidity, matrix
background and enhancement of spectral details.
87
QUANTITATIVE …
B. Assay of Substances in Multi-component…
4. Derivative spectroscopy
Discrimination
88
QUANTITATIVE …
B. Assay of Substances in Multi-component…
4. Derivative spectroscopy
It is possible to measure the absolute value of the derivative at an abscissa value
(wavelength) corresponding to a zero-crossing of one of the components in the
mixture.
• This is termed a zero-crossing measurement.
89
OTHER APPLICATION
E. Spectrophotometeric titrations
One or more of the reactants or products absorb radiation.
They are carried out in a vessel for which the light path is constant.
The absorbance is directly proportional to concentration.
Titration Curves
Plot of absorbance as a function of titrant volume and consists of two straight-line
regions with different slopes.
90
OTHER APPLICATION..
E. Spectrophotometric titrations…
Advantages
More accurate results than direct titrimetric analysis are obtained.
Can be used for the titration of very dilute solutions (Sensitive)
Do not need favorable equilibrium constants as those required for titration
that depends upon observations near the end point.
Can be used for all types of reactions (Redox, acid-base, complexometric ,
pptmetry…etc).
91
COLORIMETRY
Is a technique which involves measurement of absorbance in the visible region is
known as colorimetry.
Involves measurement of color intensity of compounds.