Calibration Report of Korea UV Tester

Procedure: 33K3-4-1810-1 Work Order # : VN-18010028

MPC Control #: CA6993 Serial Number : CA6993

Asset ID : NA Calibration Date : January 19, 2018

Time Accuracy Test


(Test Timer chamber 1)
Lower Upper
Nominal As Found As Left Result Uncertainty(±)
Limit Limit
01:50:00 01:47:00 01:51:39 01:51:41 01:53:00 PASS 00:00:26

(Test Timer chamber 2)


Lower Upper
Nominal As Found As Left Result Uncertainty(±)
Limit Limit
01:50:00 01:47:00 01:51:36 01:51:42 01:53:00 PASS 00:00:26

End of Calibration Report