You are on page 1of 1

Calibration Record

GAS01ANL01CO

Kiln Inlet CO Analyzer

Location Preheater Frequency of Calibration 6 Months

Measuring Unit % AcceptanceCriteria ±1.5%

Low Range 0
High Range 3

Calibration Data

Calibration Date 17-May-2022 Next Due date for Calibration 17-Nov-2022

Before Zero After Zero Before Span After Span

0.21% 0 0.0274 0.03

Kamran & Saravanan


Elect.Engg HOD (E & I)
Calibrated By Checked By Approved By