You are on page 1of 8

September 2018

Conventional Units
Manufacturer Report for Siemens
Immunology Plus • Lot 27550 • Exp 30−Apr−2020

Albumin Immunonephelometric, fixed time, kinetic g/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 3.19 3.20 6.79 6.69
SD
CV
1
0.155
4.9
0.120
3.8
2 0.209
3.1
0.274
4.1
# Points 12 90 12 84
# Labs 1 1 1 1

Albumin Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 3188 3201 6794 6695
SD
CV
1
155.1
4.9
120.3
3.8
2 208.8
3.1
274.2
4.1
# Points 12 90 12 84
# Labs 1 1 1 1

Alpha−1−Acid Glycoprotein Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 51.00 50.78 138.0 136.8
SD
CV
13.00
5.9
2.11
4.1
2 4.00
2.9
5.29
3.9
# Points 11 86 12 84
# Labs 1 1 1 1

Alpha−1−Antitrypsin Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 90.41 92.34 241.3 247.1
SD
CV
14.41
4.9
6.71
7.3
2 18.95
7.9
23.55
9.5
# Points 41 208 41 200
# Labs 4 5 4 5

Alpha−2−Macroglobulin Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 146.0 135.7 342.0 326.4
SD
CV
16.00
4.1
8.12
6.0
2 8.00
2.3
18.46
5.7
# Points 14 92 13 90
# Labs 1 1 1 1

Antistreptolysin−O (ASO) Immunonephelometric, fixed time, kinetic IU/mL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean −− 71.08 −− 171.1
SD
CV
1 −−
−−
7.03
9.9
2 −−
−−
13.15
7.7
# Points −− 77 −− 78
# Labs 1 1

Siemens BN Series Nephelometers • Siemens BN Liquid ASL (OPBU)


Mean 71.34 74.53 175.2 178.5
SD
CV
13.26
4.6
3.43
4.6
3.12
1.8
5.31
3.0
2
# Points 20 83 20 82
# Labs 1 1 1 1

The statistics in this report are derived from user−supplied data and are provided by Bio−Rad Laboratories as a service to its customers. Such action does not imply support of reported analytes and test methods. Refer to the package
insert for specific analyte claims and stability information. Qualitative percentages may not add up to 100 due to rounding. Peer group statistics contained in this report may not be used without the expressed written consent of Bio−Rad
Laboratories.

Copyright © 2017 Bio−Rad Laboratories, Inc. Generated 16−Oct−2018 12:16:24 Page 1


Manufacturer Report for Siemens September 2018
Immunology Plus • Lot 27550 • Exp 30−Apr−2020 Conventional Units

Antistreptolysin−O (ASO) Immunoturbidimetric IU/mL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 69.13 69.71 181.0 182.6
SD
CV
1 3.94
5.7
4.16
6.0
2 9.31
5.1
10.26
5.6
# Points 65 777 64 814
# Labs 6 8 6 9

Siemens ADVIA Chemistry Systems • Randox


Mean 71.47 65.97 170.4 164.0
SD
CV
1 10.24
14.3
11.64
17.6
2 19.64
11.5
28.76
17.5
# Points 19 151 19 145
# Labs 2 2 2 2

Apolipoprotein A−I Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean −− 76.47 −− 216.0
SD
CV
1 −−
−−
2.42
3.2
2 −−
−−
8.03
3.7
# Points −− 12 −− 12
# Labs 1 1

Apolipoprotein A−II Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean −− 20.00 −− 59.52
SD
CV
1 −−
−−
0.380
1.9
2 −−
−−
2.32
3.9
# Points −− 12 −− 12
# Labs 1 1

Apolipoprotein B Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean −− 29.02 −− 105.2
SD
CV
1 −−
−−
1.45
5.0
2 −−
−−
2.49
2.4
# Points −− 12 −− 12
# Labs 1 1

Beta−2−Microglobulin Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 0.153 0.153 0.537 0.537
SD
CV
1 0.005
3.5
0.005
3.5
2 0.022
4.1
0.022
4.1
# Points 5 5 5 5
# Labs 1 1 1 1

C3 (Complement) Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 96.16 98.12 259.0 267.8
SD
CV
1 3.77
3.9
5.58
5.7
2 10.05
3.9
17.40
6.5
# Points 33 269 34 263
# Labs 2 4 2 4

C3 (Complement) Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 85.01 85.78 224.1 227.8
SD
CV
1 2.43
2.9
3.36
3.9
2 6.45
2.9
7.67
3.4
# Points 110 1048 108 1068
# Labs 9 12 9 12

Copyright © 2017 Bio−Rad Laboratories, Inc. Generated 16−Oct−2018 12:16:24 Page 2


Manufacturer Report for Siemens September 2018
Immunology Plus • Lot 27550 • Exp 30−Apr−2020 Conventional Units

C3 (Complement) Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens Dimension Series
Mean 91.31 92.24 239.2 236.4
SD
CV
1 5.57
6.1
5.38
5.8
2 11.09
4.6
10.28
4.3
# Points 67 420 70 424
# Labs 5 6 5 6

C4 (Complement) Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 13.68 13.77 50.10 51.57
SD
CV
1 0.684
5.0
0.735
5.3
2 1.58
3.2
3.80
7.4
# Points 33 266 32 247
# Labs 2 4 2 4

C4 (Complement) Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 11.89 11.85 40.23 40.81
SD
CV
1 0.490
4.1
0.457
3.9
2 1.47
3.7
1.53
3.8
# Points 116 1044 115 1084
# Labs 9 12 9 12

Siemens Dimension Series


Mean 12.80 13.10 41.80 41.78
SD
CV
1 0.940
7.3
1.16
8.9
2 1.77
4.2
2.59
6.2
# Points 67 388 71 414
# Labs 5 6 5 6

Ceruloplasmin Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 20.67 19.93 57.00 54.47
SD
CV
1 4.01
19.4
2.78
13.9
2 10.26
18.0
6.71
12.3
# Points 21 132 20 119
# Labs 3 4 3 4

CRP (C−Reactive Protein) Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 0.920 0.970 3.63 3.76
SD
CV
1 0.040
4.3
0.041
4.2
2 0.100
2.8
0.142
3.8
# Points 20 83 20 83
# Labs 1 1 1 1

Siemens Dimension Vista


Mean −− 0.865 −− 3.30
SD
CV
1 −−
−−
0.065
7.5
2 −−
−−
0.119
3.6
# Points −− 75 −− 74
# Labs 1 1

CRP (C−Reactive Protein) Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 1.15 1.16 4.00 4.10
SD
CV
1 0.052
4.5
0.067
5.8
2 0.164
4.1
0.191
4.7
# Points 211 2843 212 2880
# Labs 11 14 11 14

Copyright © 2017 Bio−Rad Laboratories, Inc. Generated 16−Oct−2018 12:16:24 Page 3


Manufacturer Report for Siemens September 2018
Immunology Plus • Lot 27550 • Exp 30−Apr−2020 Conventional Units

CRP (C−Reactive Protein) Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens Dimension Series • Siemens Dimension CCRP (RF434)
Mean 1.27 1.25 4.60 4.54
SD
CV
1 0.040
3.1
0.053
4.2
2
0.080
1.7
0.135
3.0
# Points 22 67 21 59
# Labs 1 1 1 1

Siemens Dimension Series • Siemens Dimension CRP (DF37)


Mean 0.987 1.01 4.71 4.73
SD
CV
1 0.183
18.5
0.154
15.3
2
0.186
3.9
0.188
4.0
# Points 166 1923 168 1912
# Labs 7 11 7 11

Siemens Dimension Series • Siemens Dimension RCRP (DF34)


Mean 1.22 1.18 4.31 4.30
SD
CV
1 0.055
4.6
0.082
6.9
2
0.190
4.4
0.236
5.5
# Points 177 1626 171 1610
# Labs 9 11 9 11

CRP, High Sensitivity EIA mg/dL


Level Mon Cum Level Mon Cum
Siemens Dimension Series • Siemens Dimension CCRP (RF434)
Mean 1.17 1.20 4.13 4.33
SD
CV
1 0.060
5.1
0.065
5.5
2
0.400
9.7
0.344
7.9
# Points 23 66 23 64
# Labs 1 1 1 1

CRP, High Sensitivity Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean −− 0.811 −− 3.44
SD
CV
1 −−
−−
0.068
8.4
2 −−
−−
0.267
7.8
# Points −− 52 −− 49
# Labs 1 1

CRP, High Sensitivity Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 1.17 1.16 4.04 4.03
SD
CV
1 0.045
3.8
0.044
3.8
2 0.099
2.4
0.124
3.1
# Points 63 494 62 488
# Labs 3 4 3 4

Haptoglobin Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 90.87 88.04 219.5 213.4
SD
CV
1 3.55
3.9
6.74
7.7
2 18.15
8.3
17.08
8.0
# Points 10 36 10 22
# Labs 2 3 2 4

IgA Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 149.5 143.8 383.3 375.8
SD
CV
1 5.08
3.4
6.46
4.5
2 13.96
3.6
24.83
6.6
# Points 32 267 33 263
# Labs 2 4 2 4

Copyright © 2017 Bio−Rad Laboratories, Inc. Generated 16−Oct−2018 12:16:24 Page 4


Manufacturer Report for Siemens September 2018
Immunology Plus • Lot 27550 • Exp 30−Apr−2020 Conventional Units

IgA Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 137.1 136.5 359.7 359.5
SD
CV
1 4.38
3.2
3.99
2.9
2 10.75
3.0
9.35
2.6
# Points 79 641 83 640
# Labs 5 8 5 8

Siemens Dimension Series


Mean 125.7 123.1 342.0 325.4
SD
CV
1 3.06
2.4
5.84
4.7
2 15.22
4.5
20.93
6.4
# Points 3 10 5 12
# Labs 1 1 1 1

IgE Immunonephelometric, fixed time, kinetic IU/mL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 59.82 62.43 255.1 262.7
SD
CV
1 2.37
4.0
3.22
5.2
2 6.22
2.4
16.00
6.1
# Points 20 84 20 84
# Labs 1 1 1 1

IgG Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 681.7 662.3 1584 1556
SD
CV
1 19.34
2.8
38.45
5.8
2 64.23
4.1
115.4
7.4
# Points 32 269 32 266
# Labs 2 4 2 4

IgG Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 623.3 624.3 1494 1494
SD
CV
1 18.28
2.9
16.37
2.6
2 43.65
2.9
36.26
2.4
# Points 79 633 83 624
# Labs 5 8 5 8

Siemens Dimension Series


Mean 652.4 629.6 1651 1545
SD
CV
1 28.22
4.3
40.55
6.4
2 81.67
4.9
136.1
8.8
# Points 5 12 5 12
# Labs 1 1 1 1

IgG Subclass 1 Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 398.8 415.2 1010 931.4
SD
CV
1 28.80
7.2
27.34
6.6
2 55.87
5.5
72.67
7.8
# Points 6 17 6 17
# Labs 1 1 1 1

IgG Subclass 2 Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 217.6 211.3 501.5 491.5
SD
CV
1 11.51
5.3
12.87
6.1
2 26.62
5.3
24.38
5.0
# Points 19 107 19 105
# Labs 2 2 2 2

Copyright © 2017 Bio−Rad Laboratories, Inc. Generated 16−Oct−2018 12:16:24 Page 5


Manufacturer Report for Siemens September 2018
Immunology Plus • Lot 27550 • Exp 30−Apr−2020 Conventional Units

IgG Subclass 3 Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 21.18 19.78 52.32 47.68
SD
CV
1 2.53
11.9
1.52
7.7
2 4.95
9.5
3.82
8.0
# Points 19 108 18 101
# Labs 2 2 2 2

IgG Subclass 4 Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 37.69 37.11 98.12 94.42
SD
CV
1 3.72
9.9
2.86
7.7
2 12.72
13.0
7.98
8.5
# Points 18 106 18 101
# Labs 2 2 2 2

IgM Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 62.42 64.06 272.5 274.3
SD
CV
1 2.18
3.5
3.74
5.8
2 10.45
3.8
17.42
6.4
# Points 33 269 32 259
# Labs 2 4 2 4

IgM Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 53.75 54.57 232.3 235.0
SD
CV
1 1.98
3.7
2.01
3.7
2 6.06
2.6
6.97
3.0
# Points 70 521 73 535
# Labs 5 8 5 8

Siemens ADVIA Chemistry Systems • Siemens ADVIA Chem, IgM_2, Rgt Lots 221871+
Mean 55.10 55.33 231.8 236.4
SD
CV
1 0.900
1.6
2.02
3.7
4.50
1.9
2 5.43
2.3
# Points 12 104 12 104
# Labs 1 1 1 1

Siemens Dimension Series


Mean 59.60 58.58 277.4 269.1
SD
CV
1 3.44
5.8
3.63
6.2
2 7.99
2.9
12.41
4.6
# Points 5 12 5 12
# Labs 1 1 1 1

Kappa Light Chain, Total Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 166.2 167.8 399.3 403.9
SD
CV
1 13.69
8.2
10.92
6.5
2 22.88
5.7
18.59
4.6
# Points 17 106 18 101
# Labs 2 2 2 2

Lambda Light Chain, Total Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 88.53 88.24 203.3 207.8
SD
CV
1 4.71
5.3
3.95
4.5
2 13.31
6.5
9.17
4.4
# Points 17 105 18 101
# Labs 2 2 2 2

Copyright © 2017 Bio−Rad Laboratories, Inc. Generated 16−Oct−2018 12:16:24 Page 6


Manufacturer Report for Siemens September 2018
Immunology Plus • Lot 27550 • Exp 30−Apr−2020 Conventional Units

Prealbumin Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 14.93 14.40 42.25 40.95
SD
CV
1 0.409
2.7
1.02
7.1
2 1.65
3.9
2.02
4.9
# Points 50 391 46 419
# Labs 3 4 3 4

Siemens Dimension Series


Mean 17.11 16.23 49.94 46.35
SD
CV
1 0.410
2.4
1.44
8.9
2 1.29
2.6
5.49
11.9
# Points 27 197 27 171
# Labs 1 1 1 1

Rheumatoid Factor Agglutination IU/mL


Level Mon Cum Level Mon Cum
Siemens Dimension Series
Mean 12.67 13.24 108.8 101.7
SD
CV
1 0.520
4.1
1.27
9.6
2 1.72
1.6
15.44
15.2
# Points 6 40 6 37
# Labs 1 1 1 1

Rheumatoid Factor Agglutination U/mL


Level Mon Cum Level Mon Cum
Siemens Dimension Series
Mean 12.67 13.24 108.8 101.7
SD
CV
1 0.520
4.1
1.27
9.6
2 1.72
1.6
15.44
15.2
# Points 6 40 6 37
# Labs 1 1 1 1

Rheumatoid Factor Immunonephelometric, fixed time, kinetic IU/mL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 10.80 10.68 102.8 99.08
SD
CV
1 0.000
0.0
0.342
3.2
2 2.70
2.6
16.78
16.9
# Points 11 146 11 137
# Labs 1 2 1 2

Siemens BN Series Nephelometers • Siemens BN Liquid RF (OPCE)


Mean −− −− 96.18 97.82
SD
CV
1 −−
−−
−−
−−
2.62
2.7
2 4.33
4.4
# Points −− −− 20 81
# Labs 1 1

Rheumatoid Factor Immunonephelometric, fixed time, kinetic U/mL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 10.80 10.68 102.8 99.08
SD
CV
1 0.000
0.0
0.342
3.2
2 2.70
2.6
16.78
16.9
# Points 11 146 11 137
# Labs 1 2 1 2

Siemens BN Series Nephelometers • Siemens BN Liquid RF (OPCE)


Mean −− −− 96.18 97.82
SD
CV
1 −−
−−
−−
−−
2.62
2.7
2 4.33
4.4
# Points −− −− 20 81
# Labs 1 1

Copyright © 2017 Bio−Rad Laboratories, Inc. Generated 16−Oct−2018 12:16:24 Page 7


Manufacturer Report for Siemens September 2018
Immunology Plus • Lot 27550 • Exp 30−Apr−2020 Conventional Units

Rheumatoid Factor Immunoturbidimetric IU/mL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 8.35 8.06 153.7 158.5
SD
CV
1 0.763
9.1
1.02
12.6
2 15.43
10.0
19.95
12.6
# Points 96 855 56 548
# Labs 8 11 5 10

Siemens ADVIA Chemistry Systems • Siemens ADVIA Chem, RF, Cal Lots 180409R/193238 and later
Mean 7.68 8.31 179.2 185.7
SD
CV
1 0.940
12.2
1.01
12.2
7.26
4.1
2 7.42
4.0
# Points 12 104 12 104
# Labs 1 1 1 1

Siemens Dimension Series • Sentinel


Mean 12.85 12.72 118.1 111.6
SD
CV
1 1.14
8.9
1.15
9.0
2 5.25
4.4
10.13
9.1
# Points 13 21 14 24
# Labs 1 1 1 1

Rheumatoid Factor Immunoturbidimetric U/mL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 8.35 8.06 153.7 158.5
SD
CV
1 0.763
9.1
1.02
12.6
2 15.43
10.0
19.95
12.6
# Points 96 855 56 548
# Labs 8 11 5 10

Siemens ADVIA Chemistry Systems • Siemens ADVIA Chem, RF, Cal Lots 180409R/193238 and later
Mean 7.68 8.31 179.2 185.7
SD
CV
1 0.940
12.2
1.01
12.2
7.26
4.1
2 7.42
4.0
# Points 12 104 12 104
# Labs 1 1 1 1

Siemens Dimension Series • Sentinel


Mean 12.85 12.72 118.1 111.6
SD
CV
1 1.14
8.9
1.15
9.0
2 5.25
4.4
10.13
9.1
# Points 13 21 14 24
# Labs 1 1 1 1

Transferrin Immunonephelometric, fixed time, kinetic mg/dL


Level Mon Cum Level Mon Cum
Siemens BN Series Nephelometers
Mean 164.0 167.3 422.0 453.1
SD
CV
1 4.00
2.4
9.75
5.8
2 12.00
2.8
34.22
7.6
# Points 20 151 20 126
# Labs 1 2 1 2

Transferrin Immunoturbidimetric mg/dL


Level Mon Cum Level Mon Cum
Siemens ADVIA Chemistry Systems
Mean 170.0 175.1 415.9 428.6
SD
CV
1 4.78
2.8
6.53
3.7
2 8.90
2.1
14.53
3.4
# Points 51 437 52 433
# Labs 3 7 2 6

Siemens Dimension Series


Mean 164.0 161.8 426.1 418.2
SD
CV
1 10.80
6.6
9.07
5.6
2 17.47
4.1
25.41
6.1
# Points 40 299 41 302
# Labs 2 2 2 2

Copyright © 2017 Bio−Rad Laboratories, Inc. Generated 16−Oct−2018 12:16:24 Page 8