Professional Documents
Culture Documents
HFe Service Manual 1 PDF
HFe Service Manual 1 PDF
Service Manual
of the EDITOR HFe X-Ray generator's
Version 06220010
B. List of Tables 2
C. General information 3
a. Description 4
b. Definitions and Notations 5
c. Intended purpose 6
D. Safety Information 7
a. Radiation Safety 8
b. Manufacturer's safety 9
c. Operator's/Service Personnel Safety 10
d. Electrical Safety 11
e. Protection against Electrical Shock 12
E. Compliance 13
1 INTRODUCTION 2-1
2 INSTALLATION 3-1
9 Options 10-15
a. Description
This technical description contains installation and calibration instructions, component part lists and circuit
diagrams. Additional documents may be provided by manufacture. Please contact manufacture.
The manual uses the following references for hazards/warnings and peculiarities:
WARNING
This reference is used if a failure to obey the instructions, notes or procedures or a failure to follow them
precisely can lead to injury or lethal accidents.
CAUTION
This reference is used if a failure to obey the instructions, notes or procedures or a failure to follow them
precisely can lead to damage to the equipment or external accessories.
NOTE
This reference is used to draw attention to peculiarities.
This symbol (yellow background) is used to warn against hazardous electrical voltage
This symbol warns against moving parts (fans, fluoroscopy option only)
c. Intended purpose
The HFe Generator intended to be used in medical X-Ray diagnostic / image guidance systems. It is part of
an X-ray system. It provides power to the x-ray tubes. In addition it provides interfaces for x-ray imaging
system, and interface to DAP, AEC ion chamber, and other accessories.
The operation is only allowed by personnel trained/knowledge in X-ray imaging and after reading and
understanding the user manual.
The generator may only be used within medical x-ray rooms and within the rated ambient parameters (see
Chapter 4.5)
The device may not be operated in areas where combustion may be a risk. It shall not be operated near
flammable liquids, flammable anesthetics, disinfecting agents, aerosols spray, or in oxygen-rich environment.
a. Radiation Safety
X-ray equipment generates ionizing radiation, which is potentially hazardous to the patient, operator and
service personnel. If not properly used, the x-ray equipment may be dangerous. Due to the potentially
adverse effects of ionizing radiation, the following recommendations should be read and understood before
attempting to operate this x-ray equipment.
b. Manufacturer's safety
The x-ray equipment was designed to comply with the recommendations of the United States Center for
devices and Radiological Health (CDRH) as specified in the Code of Federal Regulations Title 21, the
Standards for Medical Devices of the International Electrotechnical Commission IEC 60601-1 and Council
Directive 93/42/EEC of the European Commission.
Although this equipment incorporates protection against other x-radiation than the useful imaging beam, a
practical design does not provide complete protection. The equipment's design does not compel the
operators or assistants to take the necessary precautions, nor does it prevent the possibility of improper use.
d. Electrical Safety
All of the movable assemblies and parts of this equipment should be operated with care and routinely
inspected in accordance with the manufacturer's recommendations contained in the present manual.
The equipment is not serviceable and access to the internal parts is prohibited. Live electrical terminals are
deadly; be sure that line disconnect switches are opened and other appropriate precautions are taken before
attaching accessories.
Do not remove the flexible H. T. cables from the X-ray tube housing or the anode / cathode assemblies H.T.
connector while the generator is on. Wait until the electrical mains have been disconnected for at least 1
minute.
E. Compliance
The EDITOR HFe complies with the following main standards, directives, and regulations:
Europe:
rd nd
EN 60601-1 / IEC 60601-1 (3 Edition and 2 Edition)
EN 60601-1-2 / IEC 60601-1-2
nd
EN 60601-1-4 / IEC 60601-1-4 (2 Edition)
nd
IEC 60601-2-54 and EN 60601-2-7 / IEC 60601-2-7 (2 Edition)
Class I equipment
According to the class of equipment, the Editor HFe X-ray generator must be
connected to an earth ground by a protective earth conductor. Failure to
provide a separate earth ground can result in electrical shock hazard causing
injury or death.
- The applied part of the generator is Type B
FDA Warning:
Warning: This x-ray unit may be dangerous to patient and operator unless safe
exposure factors, operating instructions and maintenance schedules are
observed.
The weight values apply to the standard equipment without any options.
Caution!
Do not block
ventilation screen
WARNING
X-ray Generator cabinet under table HFeU must be used only when it is
installed inside X-Ray table. HFeU generator can’t be used standalone by
itself.
[17.91 in]
Caution!
Always leave
ventilation screen free
[21.65 in]
[6.69 in]
[18.90 in]
[1.57 in] [17.72 in]
Caution!
Always leave
ventilation screen free [21.65 in]
121[266.76
97 55 49
lb]
Control Panel 4 [8.82 lb]
[38.19 in]
CAUTION!!
Always leave
ventilation screen
free!!!
[21.65 in]
[43.31 in]
[2.56 in]
[21.65 in]
[ in]
[25.56 in]
Weight: 4kg [8.82lb]
[17.32 in]
[1.97 in]
Weight: 1kg [2.20 lb]
[10.63 in]
[13.03 in]
[10.55 in]
[2.87 in]
Label position 96cm Cabinet (other cabinets are labeled in the same manner)
or
Cabinet only for pretransformer, without all labels inside of the cabinet
xxxxx S/n: serial number of the corresponding generator
yyy maximal input current depends on generator variant
Output Power 32 kW 40 kW 50 kW 65 kW 80 kW
mA / kW @ 0,1s
at 40 kV 400 / 16 400 / 16 400 / 16 400 / 16 400 / 16
at 60 kV 500 / 30 500 / 30 650 / 39 800 / 48 800 / 48
at 80 kV 400 / 32 500 / 40 500 / 40 800 / 64 800 / 64
at 100 kV 320 / 32 400 / 40 500 / 50 650 / 65 800 / 80
at 125 kV 250 / 31 320 / 40 400 / 50 520 / 65 500 / 62.5
at 150 kV 200 / 30 250 / 37.5 320 / 48 400 / 60 500 / 75
Continuous falling load (with yes yes yes yes yes
AEC)
kV range for exposure 40 – 150 kV 40 – 150 kV 40 – 150 kV 40 – 150 kV 40 – 150 kV
in increments of 1 kV 1 kV 1 kV 1 kV 1kV
or in 27 steps 27 steps 27 steps 27 steps 27 steps
kV Accuracy ±(5%+1kV) ±(5%+1kV) ±(5%+1kV) ±(5%+1kV) ±(5%+1kV)
Pulsation 100 kHz 100 kHz 100 kHz 100 kHz 100 kHz
mA range for exposure 10 – 500 mA 10 – 500 mA 10 – 650 mA 10 – 800 mA 10 – 800 mA
in 18 steps 18 steps 19 steps 20 steps 20 steps
mA Accuracy ±(6%+1mA) ±(6%+1mA) ±(6%+1mA) ±(6%+1mA) ±(6%+1mA)
max. mA @ max. kV ±(10%+1mA) for ±(10%+1mA) for ±(10%+1mA) for ±(10%+1mA) for ±(10%+1mA) for
s≤10ms s≤10ms s≤10ms ≤10ms <10ms
500mA @ 80kV 500mA @ 80kV 650mA @ 76kV 800mA @ 81kV 800mA @ 100kV
ms range for exposure 1 – 6300 ms 1 – 6300 ms 1 – 6300 ms 1 – 6300 ms 1 – 6300 ms
in 38 steps 38 steps 38 steps 38 steps 38 steps
ms Accuracy ±(4% +1ms) ±(4% +1ms) ±(4% +1ms) ±(4% +1ms) ±(4% +1ms)
±(10% +1ms) for ±(10% +1ms) for ±(10% +1ms) for ±(10% +1ms) for ±(10% +1ms) for
ms≤10ms ms≤10ms ms≤10ms ms≤10ms ms≤10ms
mAs range 0.5 – 600 mAs 0.5 – 600 mAs 0.5 – 600 mAs 0.5 – 600 mAs 0.5 – 600 mAs
in 32 steps 32 steps 32 steps 32 steps 32 steps
mAs range (optional) 0.5 – 1000mAs 0.5 – 1000mAs 0.5 – 1000mAs 0.5 – 1000mAs 0.5 – 1000mAs
in 34 steps 34 steps 34 steps 34 steps 34 steps
mAs Accuracy ±(10%+0,2mAs) ±(10%+0,2mAs) ±(10%+0.2mAs) ±(10%+0.2mAs) ±(10%+0.2mAs)
Standard AC Line voltage 400 V AC, 400 V AC, 400 V AC, 400 V AC, 400 V AC,
AC Line frequency 50/60Hz 50/60Hz 50/60Hz 50/60Hz 50/60Hz
AC Line Configuration 3N~ 3N~ 3N~ 3N~ 3N~
Other AC line input options: See below See below See below See below See below
(Paragraph 1.3.1) (Paragraph 1.3.1) (Paragraph 1.3.1) (Paragraph 1.3.1) (Paragraph 1.3.1)
Optional Generator Operating
control Console Yes Yes Yes Yes Yes
Stage 1 2 3 4 5 6 7 8 9 10 11 12 13 14
kV 40 41 42 44 46 48 50 52 55 57 60 63 66 70
Stage 15 16 17 18 19 20 21 22 23 24 25 26 27
kV 73 77 81 85 90 96 102 109 117 125 133 141 150
Loading factor time-current-product in mAs:
To be changed from 0.5mAs to 600mAs in 32 stages. The range of the time can be limited in the generator
settings. This loading factor is only active during the 2-point technique (kV and mAs).
Step 1 2 3 4 5 6 7 8 9 10 11 12 13 14
mAs 0,5 0,63 0,8 1,0 1,3 1,6 2 2,5 3,2 4,0 5,0 6,3 8,0 10
Step 15 16 17 18 19 20 21 22 23 24 25 26 27 28
mAs 13 16 20 25 32 40 50 63 80 100 125 160 200 250
Step 29 30 31 32
mAs 320 400 500 600
Loading factor tube current in mA:
To be changed from 10mA to 800mA in 20 stages. The range of the time can be limited in the generator
settings. This loading factor is only active during the 3-point technique (kV, mA, ms).
Step 1 2 3 4 5 6 7 8 9 10 11 12 13 14
mA 10 13 16 20 25 32 40 50 63 80 100 125 160 200
Step 15 16 17 18 19 20
mA 250 320 400 500 650 800
Step 1 2 3 4 5 6 7 8 9 10 11 12 13 14
ms 1 2 3 4 5 6 7 8 10 11 13 16 20 25
Step 15 16 17 18 19 20 21 22 23 24 25 26 27 28
ms 32 40 50 63 80 100 125 160 200 320 400 500 630 800
Step 29 30 31 32 33 34 35 36 37
ms 1000 1250 1600 2000 2500 3200 4000 5000 6300
Maximum Line Regulation 3 Phase 208V/ 400V /420V /440V/ 480V ±10%
The supply cable diameter of the connection between the supply transformer and the disconnect
switch for power and protective earth has to be of sufficient to guarantee the apparent resistance.
WARNING
Please be sure that hot parts are cooled down before starting maintenance or
repair work.
fuse
mains supply slow blow
A
400V 50
420/440/480V 50
208/220V 100
L1
L2
L3
N
PE
F1
F2
K1
0,03A
N PE
F3
S0 L1 L2 L3 N PE X1
Generator
For example cable C2: 06000316 cable form board UNI1 / X13 to board BA6 / X1
06000316 06000316
UNI1/X13 – BA6/X1 UNI1/X13 – BA6/X1
UNI1/X13 BA6/X1
C28
Tube Housing Small Large Main Shift Low Low High Speed HV RAD/FL
(Stator) Speed Speed max.
Focus Focus [Ohm] [Ohm] 150/180Hz [KV]
50Hz 60Hz
size size (HSS) [µF]
[µF] [µF]
VARIAN
RAD8 Diamond 1,0 2,0 20 50 43 30 no 125 RAD
RAD14 Diamond 0,6 1,2 20 50 43 30 7,5 150 RAD/FL
RAD14 B100 0,6 1,2 16 50 43 30 no* 150 RAD/FL
RAD21 Sapphire 0,6 1,2 20 50 43 30 7,5 150 RAD/FL
RAD60 Sapphire 0,6 1,2 20 50 43 30 7,5 150 RAD/FL
A132 B100 0,6 1,2 16 50 43 30 no 150 RAD/FL
A142 B100 0,3 1,0 16 50 43 30 no* 150 RAD/FL
A192 B130 0,6 1,2 16 50 43 30 7,5 150 RAD/FL
A292 B130 0,6 1,2 16 50 43 30 7,5 150 RAD/FL
G292 B130 0,6 1,2 16 50 43 30 7,5 150 RAD/FL
SIEMENS
RAY 12S-1 SV 0,6 1,2 20 50 43 30 no 150 RAD
RAY 14S-1 SV 0,6 1,2 20 50 43 30 7,5 150 RAD
DUNLEE
PX1429 DU304 0,6 1,2 20 50 40/43 30 no 150 RAD
DU1505 DA180 0,6 1,2 14 46 30 30 no 150 RAD
DU1750 DA90 0,6 1,2 36 43 30 30 no 150 RAD
DR1817 DA90 0,6 1,2 36 44 30 30 no 150 RAD
DR1833 DA90 0,6 1,2 36 43 30 30 no 150 RAD
PHILIPS
DA1750 ROT350 0,6 1,2 36 43 30 ? no 150 RAD
TOSHIBA
E7252X XH-106V 0,6 1,2 9,4 28,3 44 44 6 150 RAD
(XS-AL)
E7252X XH-106V 0,6 1,2 27,5 58,0 30 30 Y 150 RAD
(XS-R)
E7254X XH-157 0,6 1,2 20,2 38,0 30 30 6 150 RAD/FL
E7269 0,6 1,2 9.4 28.3 44 44 6 150 RAD/FL
E7239 1.0 2.0 24 24 No 125 RAD
KAILONG
RADII H1 0,6 1,2 20 50 43 30 no 150 RAD
H1080X
COMET
DI93H DX93H 0,6 1,5 20 50 30 no no 150 RAD
DX96H DX96 0,6 1,3 20 50 30 no no 150 RAD
IAE
X50H C52 1,0 2,0 20 40 25-40 25-40 no 150 RAD
X50AH C52 1,2 2,0 20 40 25-40 25-40 no 150 RAD
RTM90H C52 0,6 1,3 20 40 25-40 25-40 no 150 RAD
RTM90H C52 0,6 1,5 20 40 25-40 25-40 no 150 RAD
RTM101HS C100 0,6 1,5 20 40 25-40 25-40 no 150 RAD
DR Detectors:
DAP
Vacutec DAP
Collimators:
1. Any manual collimator using 24VAC( with 24V option)
2. Automatic collimator xxxx
Image intensifier/
1. FLXIS System
WARNING
Make sure that the mains voltage is switched off before connecting the mains power cord.
In this chapter, only the short names of the cables are used, as described in the cable overview in Error!
Reference source not found..
Optional light
field transformer
X1 X4/X5 X6 X7 X8 X9
• Before installation make sure that the mains voltage in the room distribution box meets
the specifications.
3 phases, L1, L2, L3, neutral N and the protective earth PE are required.
• If the mains voltage does not meet the specification an additional three phase pre-
autotransformer in an optional cabinet is required.
Connect the main power cable to the Connect the mains cable to the terminal on the
sub distribution terminal X1 and the sub-distribution
protective earth cable to the
grounding bar.
L1 X1-1
L2 X1-2
L3 X1-3
N X1-N
PE PE-bar
• The power cord should be fixed with cable straps to the generator cabinet.
E11
480Vac Connection
Jumper E11 to E12
Primary
Connections
Neutral
Connection
230Vac 24Vac
Primary 60Vac Secondary
Secondary
Move wire select 230Vac
here for 400Vac Secondary
Select 480Vac F2
F1
Before connections of mains are attached to generator, turn OFF mains voltage in the room,
and make sure with multi-meter that the voltage is z
L1
N PE
L2
L3
• The power cord should be fixed with cable straps to the generator cabinet.
• Connect the other side of the 3 phase wires of the mains fuse to the Auto transformer taps that
match closest the voltage measured.( i.e if mains voltage is 487VAC the 3 wires should be
connected to 3 480VAC taps in the Autotransformer)
• The power cord should be fixed with cable straps to the generator cabinet.
mains Gen
Removed nipple
Please pay attention to the High voltage cable manufacturer's instructions. Different kinds of
isolation material are used by manufactures (Silicon oil, silicon grease)
WARNING:
HFe Generator can’t be operated without X-Ray tube Thermal Switch or Pressure Sensor
connected to X6-5 and X6-6.
PE
Picture Description
Position of the 30µF phase shift capacitor
on the backside of the electronic cabinet
PE
Jumper Position
X52 1-2
X59 2-3
X61 open
X62 open
Jumper Position
X52 2-3
X59 1-2
X61 close
X62 close
+
1
3
REV. B[ ]
-
S/S, REV. B
JB4
JB5
POWER
POLE
TB1 PINS:
1 2 3 4
PHASE (U)
COMM (V)
MAIN (W)
GND
TB1: External connection to the X-Ray Tube Anode Motor. Use ring lugs TE Connectivity, PN 320634 or
equivalent.
\\\\\\\\\\
Switches
Switch Function
Number
W1 the microprocessor
Jumpers
Jumper Default Function
JP1 Closed For Factory Programming
JP2 Open I/O Setup. Not Used.
JP3 Closed CAN Bus Termination
Test points
TP Signal Function
TP1 +3.3V Bias Supply
TP2 +15V Bias Supply Control signal of the
Inverter IGBT V2 and V3
TP3 GND Ground
TP4 +5V Bias Supply
TP5 GND Ground
Warning:
The connector X5 must not be reconnected to the ZST2 if the system is
switched on.
4.2 Setup
If the electrical installation of all cables and external components is done the generator is ready for
customization.
The setup is divided into different functional blocks
- „System“
- „Generator“
- „Device“ (with sub dialogs for the „XRAY device settings“)
- „Tube settings“
- „Filament settings“
- „kV control settings“
- “Collimator settings”
- “Image system settings”
- „Statistics“
The generator configuration can be done with a service PC using the service software or directly on the
control desk. The configuration via the service software utility is recommended because the generator
configuration via control panel is limited. Some configuration parameters are not available with the control
panel. For more detailed information about the configuration software see document 06220011 which is
provided with generator's documentation.
Before starting the system configuration a short power on test should be done to check that all cables are
connected properly.
- An adapter cable (06220244) from a 9-pole SUB-D (female) to a 2 x 5-pole plug socket. This
adapter cable is connected to the ZST2 board at delivery.
- Connect the other side of the service cable 06000244 to the serial COM port of the service PC. An
extension cable can be used for larger distances to the service PC.
The service software works only with the first 4 COM port numbers. The number of the available
serial COM ports is shown in the hardware manager of the windows operating system. Larger COM
port numbers are reserved for special applications.
- If an error message appears during program start check the file setup.ini or use another COM port.
Firmware version
Start parameter
- If an error message appears try to use another COM port number. The other parameters in the
setup.ini should not be changed.
The download takes normally less than 1min. The transmitted data will be checked at the end of the
transfer.
4.
- Press button “Reset exposers tube 2” to reset the
exposure counter of tube 2.
5.
- The service software use the system time of the
service PC to setup the internal generator clock. Make
sure the time of the PC is correct, before you set the
time inside of the generator.
6.
- Check the generator time by pressing the button “read
time form gen”
Warning:
The value of the internal x-ray tube heat integrator is also reset.
The tube anode must be cooled down for minimum 20 minutes
before a new exposures is started.
Filament heating to low. Result: Tube Filament heating to high. Result: Tube Filament heating ok. Result: Constant tube
current is too low at exposure start current is too high at exposure start current over the complete exposure
The filament current is also not proportional to the exposure voltage. To illustrate this dependency a typical
filament emission chart from an x-ray tube is shown below.
The generator has got a semi-automatic calibration function for tube
currents in the steps shown here
The tube voltage during calibration is 70kV. At the end of the calibration
process two exposures will follow at 40kV and 100kV to calibrate the
nonlinearity between tube current and tube voltage.
- Check that the anode and cathode connection are not interchanged. The filament will heat up using
a 2.5A standby filament current as soon as the generator is switched on. The heated filament is
visible if the tube window is not covered.
- Control the speed and the direction of the anode rotation. For more information see the tube
datasheet and documentation.
- The tubes need to heat up to room temperature prior to calibration.
Device
buttons
Error
messages
Focus
Calibration buttons
parameters
Prameter
values
This runs in parallel to the main window of the service program. The whole data communication to the
generator is shown in the left window. The curves of tube voltage, tube current and the converter-current
from the last exposure are shown in the oscillogram window after pressing F5 or the left mouse button inside
of the oscillogram field.
The tube calibration process contains 15 steps per focus. Each step is indicated by a parameter number, a
short information text and the range of values.
The tube calibration starts automatically with the large focal spot from parameter number 920 up to 939. The
calibration of the small focus starts from parameter number 900 up to 919. The maximum tube current is
limited by the maximum allowed generator parameters and the tube parameters for the large and the small
focus. Tube currents exceeding these limits do not need any calibration.
The parameter 900 must be inserted manually for the calibration of the small focus. The button “Set”
activates the parameter.
The next figure shows the measuring oscillogram window during the calibration process. The time current
product (mAs) is measured for 10ms starting at 6ms and ending at 16ms.
10ms
1
) Calibration only up to the maximum allowed tube current. No x-ray radiation at higher tube current steps
Exposure parameter
The exposure parameters are preset for each calibration step. The time current product (mAs) is
measured for 10ms starting at 6ms and ending at 16ms. This measured mAs product is the basis for
the half automatic calibration. The proposed filament heating current changes automatically after
each exposure if the measured mAs product value is different to the expected mAs.(The suggested
values for 20mA and below may not be correct for long H.V. cable.(about 30m) In that case it will
require entering values up/down as function measured mA) The Table gives an overview of the
expected mAs and the range of a good calibrated result of the mAs.
At the end of the calibration process two exposures will follow at 40kV and 100kV to calibrate the
nonlinearity between tube current and tube voltage. This is only an adjustment of an internal
correction factor. The generator sets automatically a higher tube current depending on the chosen
tube type. The SVC value is in %.
- End of calibration
The calibration is stopped by the button “EXIT”. The calibrated filament heating current values are
stored automatically to the EEPROM. Close the console window by pressing the hide button.
- Press the button “Fil.” to start the subroutine for the tube
calibration.
- Parameter 920
- No exposure is needed
- After the Svc. value is set press the button M
- Select the next Parameter 921
- Parameter 921
- Default is 0000.
- No exposure is needed.
- After the Svc. value is set press the button M
- Select the next Parameter 922
- Start exposure
Information:
Sometimes the automatic filament heating current calculation may
not find the correct value at the low tube currents 2mA and 5mA. In
this case set the value manually.
-
- Parameter 923
Information:
Sometimes the automatic filament heating current calculation may
not find the correct value at the low tube currents 2mA and 5mA. In
this case set the value manually.
- Parameter 924
- Start exposure
- Start exposure
- Parameter 926
- Start exposure
- Start exposure
- Parameter 928
- Start exposure
- Start exposure
- Parameter 930
- Start exposure
- Parameter 932
- Parameter 938
-
- Parameter 939
- Start exposure
- No exposure is needed
- After the Svc. value is set press the button M
- Select the next Parameter 901
- Default 0000.
- No exposure is needed.
- After the Svc. value is set press the button M
- Select the next Parameter 902
- Parameter 902
- Start exposure
- Start exposure
Information:
Sometimes the automatic filament heating current calculation will
not find the correct value at the low tube currens 2mA and 5mA. In
this case set the value manually.
-
- Parameter 904
- Start exposure
- Start exposure
-
- Parameter 906
- Start exposure
- Start exposure
- Parameter 908
- Parameter 910
-
- Parameter 911
- Parameter 918
- Start exposure
- Start exposure
927:tube1,foc2(100mA)(0..9999) :4160
- Prior to exposure R 3: 70kV 100mA 16ms
Exposure parameter
Changed parameter value
927:tube1,foc2(100mA)(0..9999) :4169
- After exposure R 3: 70kV 99.7mA 16ms #0.999mAs 30
301:setup - :920 - Select parameter number 920 and press to start the
calibration of the large focus
- Parameter 924
924:tube1,foc2(10mA)(0..9999) :3370 Calibration of the 10mA tube current
R 3: 70kV 10mA 50ms - Normally 3100mA – 3500mA
- Expected measured mAs value 0.100mAs
- Start exposure and press to confirm the value of
the new calculated filament heating current.
- The filament heating current changes automatically after
924:tube1,foc2(10mA)(0..9999) :3376 each exposure until the measured mAs is in the range of
R 3: 70kV 10.2mA 50ms #0.107mAs 30 0.090 – 0.110mAs
- Press and select the next parameter by using button
- Parameter 925
925:tube1,foc2(20mA)(0..9999) :3370 Calibration of the 20mA tube current
R 3: 70kV 20mA 25ms - Normally 3400mA – 3700mA
- Expected measured mAs value 0.200mAs
- Start exposure and press to confirm the value of
the new calculated filament heating current.
- The filament heating current changes automatically after
925:tube1,foc2(20mA)(0..9999) :3376 each exposure until the measured mAs is in the range of
R 3: 70kV 20.3mA 25ms #0.199mAs 30 0.190 – 0.210mAs
- Press and select the next parameter by using button
- Parameter 926
926:tube1,foc2(50mA)(0..9999) :3820 Calibration of the 50mA tube current
R 3: 70kV 50mA 16ms - Normally 3700mA – 4000mA
- Expected measured mAs value 0.500mAs
- Start exposure and press to confirm the value of
the new calculated filament heating current.
- The filament heating current changes automatically after
926:tube1,foc2(50mA)(0..9999) :3831 each exposure until the measured mAs is in the range of
R 3: 70kV 48.8mA 16ms #0.492mAs 30 0.490 – 0.510mAs
- Press and select the next parameter by using button
- Parameter 927
927:tube1,foc2(100mA)(0..9999) :4160 Calibration of the 100mA tube current
R 3: 70kV 100mA 16ms - Normally 3900mA – 4300mA
- Expected measured mAs value 1.000mAs
- Start exposure and press to confirm the value of
the new calculated filament heating current.
- The filament heating current changes automatically after
927:tube1,foc2(100mA)(0..9999) :4169 each exposure until the measured mAs is in the range of
R 3: 70kV 99.7mA 16ms #0.999mAs 30 0.950 – 1.050mAs
- Press and select the next parameter by using button
- Parameter 929
929:tube1,foc2(320mA)(0..9999) :4620 Calibration of the 320mA tube current
R 3: 70kV 320mA 16ms - Normally 4100mA – 4500mA
- Expected measured mAs value 2.000mAs
- Start exposure and press to confirm the value of
the new calculated filament heating current.
- The filament heating current changes automatically after
929:tube1,foc2(320mA)(0..9999) :4610 each exposure until the measured mAs is in the range of
R 3: 70kV 320mA 16ms #3.216mAs 30 3.150 – 3.250mAs
- Press and select the next parameter by using button
- Parameter 930
930:tube1,foc2(500mA)(0..9999) :4800 Calibration of the 500mA tube current
R 3: 70kV 500mA 16ms - Normally 4600mA – 5000mA
- Expected measured mAs value 2.000mAs
- Start exposure and press to confirm the value of
the new calculated filament heating current.
- The filament heating current changes automatically after
930:tube1,foc2(500mA)(0..9999) :4803 each exposure until the measured mAs is in the range of
R 3: 70kV 500mA 16ms #4.945mAs 30 4.950 – 5.050mAs
- Press and select the next parameter by using button
- Parameter 931
Calibration of the 650mA tube current. Tube is limited to
500mA tube current. This tube current will be not
calibrated but the filament current must be set manually
931:tube1,foc2(650mA)(0..9999) :4950 approx. 150mA higher than the last calibrated one.
R 3: 70kV 500mA 16ms
In this case 4803mA + 150mA = 4953mA
Set this value manually without starting an exposure
- Press and select next parameter with button
- Parameter 932
Calibration of the 800mA tube current. The tube is limited
to a tube current of 500mA. This tube current will not be
calibrated but the filament current must be set manually
932:tube1,foc2(800mA)(0..9999) :4950 approx. 150mA larger than the last calibrated one.
R 3: 70kV 500mA 16ms In the shown case 4803mA + 150mA = 4953mA
Set this value manually to 4950mA
- Press and select the next parameter by using button
- Parameter 938
Calibration of the correction factor in percent of the tube
current at low exposure voltage.
938:tube1, focus2 fb(50%..150%) :92 - Normal values are between 88% - 99%.
- Expected measured mAs value 3.200mAs
R 3: 40kV 320mA 16ms
- The semi automatic tube calibration changes the
correction factor in steps of +/-1% until the expected mAs
value of 3.200mAs is reached.
- Start exposure and press to confirm the value of
the new calculated correction factor.
- The semi automatic tube calibration changes the
938:tube1, focus2 fb(50%..150%) :93 correction factor in steps of +/-1% until the expected mAs
R 3: 40kV 320mA 16ms #3.173mAs 30 value of 3.200mAs is reached.
- Press and select the next parameter by using button
301:setup - :900 - Select the parameter number 900 and press to start
the calibration of the small focus
- Parameter 901
Sets the boost current during preparation
901:tube1, boost focus1 mA(0..600) :0000 - Normally not used
R 3: 70kV 2.0mA 250ms - Press and select the next parameter by using button
- Parameter 902
902:tube1,foc1(2mA)(0..9999) :2900 Calibration of the 2mA tube current
R 3: 70kV 2.0mA 250ms - Normally 2800mA – 3300mA
- Expected measured mAs value 0.020mAs
- Start exposure and press to confirm the value of
the new calculated filament heating current.
- The filament heating current changes automatically after
each exposure until the measured mAs is in the range of
902:tube1,foc1(2mA)(0..9999) :2906 0.015 – 0.030mAs
R 3: 70kV 2.0mA 250ms #0.029mAs 30
- Press and select the next parameter by using button
- Parameter 903
903:tube1,foc1(5mA)(0..9999) :3120 Calibration of the 5mA tube current
R 3: 70kV 5.0mA 100ms - Normally 3000mA – 3400mA
- Expected measured mAs value 0.050mAs
- Start exposure and press to confirm the value of
the new calculated filament heating current.
- The filament heating current changes automatically after
903:tube1,foc1(5mA)(0..9999) :3120 each exposure until the measured mAs is in the range of
R 3: 70kV 5.1mA 100ms #0.057mAs 30 0.040 – 0.060mAs
- Press and select the next parameter by using button
- Parameter 904
904:tube1,foc1(10mA)(0..9999) :3280 Calibration of the 10mA tube current
R 3: 70kV 10mA 50ms - Normally 3100mA – 3500mA
- Expected measured mAs value 0.100mAs
- Parameter 905
905:tube1,foc1(20mA)(0..9999) :3470 Calibration of the 20mA tube current
R 3: 70kV 20mA 25ms - Normally 3400mA – 3700mA
- Expected measured mAs value 0.200mAs
- Start exposure and press to confirm the value of
the new calculated filament heating current.
- The filament heating current changes automatically after
905:tube1,foc1(20mA)(0..9999) :3479 each exposure until the measured mAs is in the range of
R 3: 70kV 20.2mA 25ms #0.206mAs 30 0.190 – 0.210mAs
- Press and select the next parameter by using button
- Parameter 906
906:tube1,foc1(50mA)(0..9999) :3740 Calibration of the 50mA tube current
R 3: 70kV 50mA 16ms - Normally 3700mA – 4000mA
- Expected measured mAs value 0.500mAs
- Start exposure and press to confirm the value of
the new calculated filament heating current.
- The filament heating current changes automatically after
906:tube1,foc1(50mA)(0..9999) :3744 each exposure until the measured mAs is in the range of
R 3: 70kV 48.9mA 16ms #0.499mAs 30 0.490 – 0.510mAs
- Press and select the next parameter by using button
Parameter 907
- Parameter 908
Calibration of the 200mA tube current. The tube is limited
to a tube current of 63mA. This tube current will not be
calibrated but the filament current must be set manually
908:tube1,foc1(200mA)(0..9999) :4170 approx. 150mA larger than the last calibrated one.
R 3: 70kV 63mA 16ms In this case 4020mA + 150mA = 4170mA
Set this value manually without starting an exposure
- Press and select the next parameter by using button
- Parameter 909
Calibration of the 320mA tube current. The tube is limited
to a tube current of 63mA. This tube current will not be
calibrated but the filament current must be set manually
909:tube1,foc1(320mA)(0..9999) :4170 approx. 150mA larger than the last calibrated one.
R 3: 70kV 63mA 16ms In this case 4020mA + 150mA = 4170mA
Set this value manually without starting an exposure
- Press and select the next parameter by using button
- Parameter 911
Calibration of the 650mA tube current. The tube is limited
to a tube current of 63mA. This tube current will not be
calibrated but the filament current must be set manually
911:tube1,foc1(650mA)(0..9999) :4170 approx. 150mA larger than the last calibrated one.
R 3: 70kV 63mA 16ms In this case 4020mA + 150mA = 4170mA
Set this value manually without starting an exposure
- Press and select the next parameter by using button
- Parameter 912
Calibration of the 320mA tube current. The tube is limited
to a tube current of 63mA. This tube current will not be
calibrated but the filament current must be set manually
912:tube1,foc1(800mA)(0..9999) :4170 approx. 150mA larger than the last calibrated one.
R 3: 70kV 63mA 16ms In this case 4020mA + 150mA = 4170mA
Set this value manually without starting an exposure
- Press and select the next parameter by using button
- Parameter 918
918:tube1, focus1 fb(50%..150%) :92 Calibration of the correction factor in percent of the tube
current at low exposure voltage.
R 3: 40kV 32mA 16ms - Normal values are between 88% - 99%.
- Expected measured mAs value 0.320mAs
- Start exposure and press to confirm the value of
the new calculated correction factor.
- The semi automatic tube calibration changes the
938:tube1, focus1 fb(50%..150%) :93 correction factor in steps of +/-1% until the expected mAs
R 3: 40kV 32.2mA 16ms #3.173mAs 30 value of 0.320mAs is reached.
- Press and select the next parameter by using button
- Parameter 919
939:tube1, focus1 fb(50%..150%) :106 Calibration of the correction factor in percent of the tube
current at high exposure voltage.
R 3:100kV 50mA 16ms - Normal values are between 101% - 115%.
- Expected measured mAs value 0.500mAs.
- Start exposure and press to confirm the value of
939:tube1, focus1 fb(50%..150%) :107 the new calculated correction factor.
R 3: 40kV 48.8mA 16ms #0.496mAs 30 - The semi automatic tube calibration changes the
correction factor in steps of +/-1% until the expected mAs
value of approx. 0.500mAs is reached
301:setup – data stored :302 - Press button to store the calibration data.
In this case the SVC value will not change after an exposure is taken. Increase the SVC value manually in
100mA steps until the automatic calculation of the tube heating current will start again. Use the marked
arrow button to increase the value manually in larger steps. Press the button M after changing the value
manually.
2. The tube heating current is too high. Therefore the resulting tube current is also too high. The generator
may stop the exposure displaying the error messages E049 “Tube mA too high”. Reduce the filament
heating current in steps of 100mA until the error message will not appear again and the current calculation
will work.3. Communication between service PC and generator breaks down during calibration.
The error message E012 appears on the display of the control panel.
Switch off the generator and close the service software. Start the generator and the service software again.
Open the backup file and finish the calibration starting from the last calibrated step.
CAUTION
Even if the generator is switched off there are still components, inside of the generator
cabinet, which are electrically live!
Only qualified and trained service personnel may open the cabinets!
Incorrect handling can result in danger to life!
message Explanation *
1 Error tube kV max 2
2 error control A max 2
3 error tube mA max 2
4 Error tube +-kV diff 2
5 Error tube +-mA diff 2
6 Error ROM test 1
7 Error RAM test 1
8 Error unknown 2
9 Error no Tube kV 2
10 Error tube kV too high 2
11 Error inverter overload ( > 150000WS) 2
12 Error comm. send timeout 2
13 Error checksum E²Prom 1
14 Error watchdog 2
15 Error comm. receive timeout 2
16 Error E²Prom wait timeout 1
17 Error filament system 2
18 Error DAP system 1
19 Filament parameter anomaly in the E²Prom
20 Error +-15V low 2
21 Error +5V low (+-5%) 2
22 Error key is on (during start of the generator) 2
23 Error XRAY key is on (during start of the generator) 2
1* = call service
2* = call service if problem appears several times
3* = warning, repairable by the user
005 tube +-mA diff (larger than 200mA, signal 'IRDIFF' on the ZST2). This error may
occur only during radiation.
Possible causes:
- Control the connection between HV-trafo and PCB
BUS. Check also the shielding of this cable. The
shielding must be connected on both sides.
- Defective or old high-voltage cables.
Check the connectors on both sides. The affected
parts must be replaced if sooting is visible.
Check the cables over their entire length
- Defective high-voltage transformer
010 tube kV too high. 3ms after start of exposure, the kV may not be higher than 125% of
the desired value. (In setup mode 150%)
Possible causes:
- Tube current is too low at the start of exposure.
This creates an overshoot during the start of
exposure.
- Check the tube calibration.
The uncalibrated tube current must be set +150mA
larger than the last calibrated tube current.
Example:
..
320mA(IR) 4232mA(IHeat) calibrated
500mA(IR) 4404mA(IHeat) calibrated
650mA(IR) 4554mA(IHeat) uncalibrated
800mA(IR) 4554mA(IHeat) uncalibrated
011 inverter overload ( The maximum permissible energy quantity of the converter has
160000Ws) been exceeded.
A possible cause:
- Too many exposures in a short time.
Switch off. The generator must be cooled down.
013 E²Prom checksum The setup data of the generator are checked after powering on.
Possible causes:
- Faulty EEPROM on the ZST2 board.
- Disconnection to the service PC during data
transfer to the generator.
Try to restart the generator and the service program.
First download all data from the ZST2 of the faulty EEPROM for
later troubleshooting.
Second step: Load an older backup from the service PC or from
the generator CD and transfer the complete file to the generator.
Start the generator again. This data are factory data. The tube
needs to be re-calibrated.
014 watchdog No reaction of the generator software for more than 260ms.
A possible cause:
- Damaged microcontroller
015 Receive timeout” with This error appears if no messages were received 5 seconds after
extension (CAN061 – starting the communication.
CAN066 or COM1).
Possible causes:
- see error number 12
016 E²Prom wait timeout After powering on the EEPROM is checked every 100ms for
readiness.
A possible cause:
- The EPROM D4 on the ZST2 is damaged or not
correctly fixed in the socked.
019 filament parameter. The filament parameters in the EEPROM are faulty (checked by
the software). All filament heat data will be set to 0 and no tube
current will be possible.
Rewrite the heating parameters to the EEPROM
022 key is on : If a button is pressed on the control desk during powering on.
Start the control desk setup and check the functionality of the
buttons.
(Connect the dongle to the control desk and start the generator
without pressing any button, except the “switch on button”. Now it
is possible to control each button on the desk by pressing it. )
023 XRAY key is on : If the x-ray hand switch is pressed during powering on.
Start the control desk setup and check the functionality of the two
x-ray hand switches.
(Connect the dongle to the control desk and start the generator
without pressing any button except the “switch on button”)
024 mAs max The maximum mAs product is reached for systems using pulsed
exposure. The maximum mAs is 1200mAs during one exposure
sequence.
025 exposure too short Currently not used
026 Generator not ready. If the generator works with two tubes this error message appears if
there is no feedback from the switch in the high voltage
transformer.
Possible causes:
- Broken fuse F7 in the main distribution of the
generator.
- Check the cable connection to the transformer
032 door open The signal 'DOOR'. It is requested during the preparation phase.
Possible causes:
- Connector X8 on the sub-distribution is open
although no door contact is installed.
- The flat interface cable between NAG2 and ZST2
is damaged.
033 exp. time > 6.3s If the exposure doesn’t stop after 6.3s the backup timer sets this
error.
Possible causes / solutions:
- Check Jumper X46 on the ZST2 (has to be open)
X46 reduces the time of the backup timer to less
than 100ms (only for factory tests).
- Defective Microcontroller board MP16.
035 exp. time < 2ms This error is produced by the BA6 and transferred to the generator
(ZST2).
Possible causes:
- Wrong exposure parameter selection
- Wrong measuring field selection
- Direct radiation onto measuring field
036 AEC exposure break Exposure stopped by user. The operator released the hand switch
before the exposure time is expired.
037 dose too low after 50ms The BA6 examines after 50ms whether the adjusted %-value of the
switching off dose has been reached. This way it also can be
tested whether the emitter is in the correct position. This error is
produced by the BA6 and transferred to the generator (ZST2).
Possible causes:
- Too low selected photograph parameters
- Check tube calibration
038 pulse delay too long If the delay time during two exposures is too long during the pulsed
exposure mode.
039 exp. prepare timeout The maximum preparation time is adjustable via the setup program
from 2s to 20s.
If the operator presses the preparation button for a long time this
error will be set. This error is used to save the filament of the tube
040 device ready timeout No response from the active exposure device (bucky) if the second
stage of the exposure button is pressed.
The maximum response time is adjustable via the setup program
from 1s to 20s.
041 starter timeout The maximum time for the starter is adjustable via the setup
program from 1000ms to 2500ms.
Possible causes:
- broken fuse F2 on the NAG2 board
- check the stator connection to the tube
- If a high speed starter is installed check the error
LEDs on the HSS1 board.
042 grid is on The error appears if the grid is running or the input signal “bucky”
on the NAG2 is active during the generator is switched on. Check
the external bucky devices. For test purposes remove all interface
cables from the NAG2 board.
Possible causes:
- a faulty Interface board NAG2
- Check the solid state relay U2 on the NAG2 board.
045 no main current If the current in the converter is too low during exposure.
A possible cause:
- Open cable between ZST2 and ANS2. Check the
connection between the ZST2 and the ANS2
board inside the converter
- One of more phase of the main supply is missing.
Check automatic circuit breaker F1
- check fuse F6 on the sub distribution
- check the power supply cable to the converter.
046 exposure stopped by user This error appears, if the operator stops the exposure before the
end of set exposure time.
047 controller – E²prom verify The flash memory of the µC and the E²Prom are checked for
updates after switching on.
048 Wrong tube position Not used
049 Tube mA too high This error appears, if the tube current has reached more than 50%
of the set value during the first 30ms of x-ray radiation.
Possible causes / solutions:
- This error appears sometimes during the
calibration. This is normal. In this case set the
heating current parameter manually.
- Check the tube calibration
051 No BUS-Signal from AEC If the automatic exposure control board (BA6 / BA7) stops the
exposure by the signal BUS immediately and additionally with a
software command via CAN BUS. The error message E051 is set if
the exposure stops only by a software command without the
hardware signal BUS.
The cause can be an hardware defect on the BA6 or the ZST2.
Check the optocoupler U1 on the AEC board BA6 or BA7. The
LED V36 signalizes the status of the signal BUS.
052 FLXIS not ready Only for fluoroscopy generators if a FLXIS system is installed. The
FLXIS system is controlled by serial commands. This message
appears after one minute after switching on if the FLXIS does not
boot correctly
053 !Anode heat content >100% ! Maximum heat storage capacity of the tube anode is reached. The
tube needs time to cool down. Additional exposures will damage
the anode.
This message appears every minute if the calculated load of the
anode is higher than the second entry of the cooling data of the
anode in the service program.
Example: calculated value is higher than 212kWs
064 CAN-System The internal CAN bus communication is temporarily faulty. The
CAN bus controller tries to repeat the disturbed messages or
commands. If there are too many transmission errors this error
message will be set.
Possible causes:
- bad cable connection between the CAN bus
devices. Check all internal RJ45 cables and the
data cable to desk.
- CAN bus termination to low. Check the termination
resistors of the CAN bus devices BE26, BA6, BA7,
HSS1, IFF1. The CAN bus needs a termination
only on two devices. Normally on the ZST2 board
(X42 closed) and the BE26 (X17 closed).
6.3.16 Cut
Copying the APR data + text of the recently selected organ to the cache and deleting this data set (the
remaining APR data sets are set forward by one position).
6.3.17 Copy
Copying the APR data + text of the recently selected organ to the cache
6.3.18 Paste
Copying the APR data from the cache to the momentary position (the remaining APR data sets are set back
by one position).
6.3.19 Accept
The changes have to be confirmed by this function. Otherwise, the settings made are lost if another APR
data set is selected.
6.3.20 Region x
Selecting one of 16 body regions. „x“ = the number of selected region. This text can be changed in the text
field and changes of texts or data has to be confirmed by “Accept”.
6.3.24 Accept
The changes have to be confirmed by this function. Otherwise, the settings made are lost if another region is
selected.
Table 1 Table 2
cm kV increment mAs-increment cm kV increment mAs increment
+10 +1 +10 +1
+9 +1 +9 +1
+8 +1 +8 +1
+7 +1 +7 +1
+6 +1 +6 +1
+5 +1 +5 +1
+4 +1 +4 +1
+3 +1 +3 +1
+2 +1 +2 +1
+1 +1 +1 +1
0 0
-1 -1 -1 -1
-2 -1 -2 -1
-3 -1 -3 -1
-4 -1 -4 -1
-5 -1 -5 -1
-6 -1 -6 -1
-7 -1 -7 -1
-8 -1 -8 -1
-9 -1 -9 -1
-10 -1 -10 -1
Table 3 Table 4
cm kV increment mAs increment cm kV increment mAs increment
+10 +1 +10 +1
+9 +1 +9 +1
+8 +1 +8 +1
+7 +1 +7 +1
+6 +1 +6 +1
+5 +1 +5 +1
+4 +1 +4 +1
+3 +1 +3 +1
+2 +1 +2 +1
+1 +1 +1 +1
0 0
-1 -1 -1 -1
-2 -1 -2 -1
-3 -1 -3 -1
-4 -1 -4 -1
-5 -1 -5 -1
-6 -1 -6 -1
-7 -1 -7 -1
-8 -1 -8 -1
-9 -1 -9 -1
-10 -1 -10 -1
cm kV mAs
+10 +1
+9 +0,5
+8 +0,5
+7 +0,5
+6 +0,5
+5 +1
+4 +0,5
+3 +0,5
+2 +0,5
+1 +0,5
0
-1 -0,5
-2 -0,5
-3 -0,5
-4 -0,5
-5 -1
-6 -0,5
-7 -0,5
-8 -0,5
-9 -0,5
-10 -1
A list of an English APR text file is shown at the end of this chapter.
Example organ nr. 7:
:#007 Foot lateral:EXP 0,1,2,55,8.0,100,100:AEC 0,1:OBJ 07
6.4.1 Declaration
“#007” Number of the organ. This number is shown in the first display line.
“Organ name“ (Foot lateral) maximum 34 characters
thickness
:OBJ 01
02
...
EDITOR HFe
04 35
32 39 23 11 10 09 46 47
14 07 05 12 34
205 203 201 240 241 220 221 231 232 233 210 211 212
206 204 202 242 243 222 223 235 234 236 251 230 250
XRAY
page 7-5
key 1,3
PB No Designation / Function
Push-button 001 Store push-button for the anatomic program
Push-button 002 Selection of the organ group 0 or numeric key 0
Push-button 003 Information query
Push-button with
004 Select right measuring field
LED
Push-button with
005 Select Film foil system
LED
Push-button 006 Select exposure mode (R1, R2, R3)
Push-button 007 New patient
Push-button with
008 Select Film foil system
LED
Push-button 009 Selection of the organ group 3 or numeric key 3
Push-button 010 Selection of the organ group 2 or numeric key 2
Push-button 011 Selection of the organ group 1 or numeric key 1
Push-button with
012 Select centered measuring field
LED
Push-button 013 Correction, higher density
Push-button with
014 Select small focal spot
LED
Push-button with
015 Select large focal spot
LED
Push-button 016 Correction, lower density
Push-button 017 Selection of the organ group 6 or numeric key 6
Push-button 018 Selection of the organ group 5 or numeric key 5
Push-button 019 Selection of the organ group 4 or numeric key 4
Push-button with
020 Select left measuring field (disabled)
LED
Push-button 021 Select mA for radiography, lower values
Push-button 022 Select mAs for radiography, lower values
Push-button with
023 mode 3 / x-ray device 3, radiography/fluoroscopy(enabled)
LED
Push-button with
024 mode 4 / x-ray device 4, (disabled)
LED
Push-button 025 Selection of the organ group 9 or numeric key 9
Push-button 026 Selection of the organ group 8 or numeric key 8
Push-button 027 Selection of the organ group 7 or numeric key 7
Push-button with
028 Error signalization and confirmation
LED
Push-button 029 Select mA for radiography, higher values
Push-button 030 Select mAs for radiography, higher values
Push-button 031 Select kV for radiography, higher values
Push-button with
032 mode 1 / x-ray device 1, only fluoroscopy (enabled)
LED
Push-button 033 Function button 4
7. Buzzer test
The buzzer goes on for 1s.
Buzzer test: one second
8. Display settings 1
- Brightness LED: 20 – 100 default 60
- Brightness display: 20 – 100 default 100
- Baud rate COM1: 06 – 14 default 7 (9600) (external COM port)
0 = disable
4 = 1200
5 = 2400
6 = 4800
7 = 9600
8 = 19200
9 = 38400
10 = 57600
11 = 76800
12 = 100000
13 = 115200
14 = 125000
To save press M
Key 8
Key 14 Key 9
Key 13 Key 17
Key 15 Key 18
Key 16 Key 10
Key 19 Key 11
Key 20 Key 23
Key 21 Key 24
Key 22 Key 12
Figure 6-4: Touch Panel
2. Insert new battery by pushing it in with slight force.(Positive polarity on top) ). See replacement parts
for more information about the used battery
3. Dispose battery according to local code
4. Set clock time according to procedure chapter 3.2.3 Setup of the internal time
Note: When battery is replaced the X-Ray tube KHU calculated value is lost. If tube is hot wait until it cools
before starting to take exposures.
V1, V2, Green Indication of an active Gate The IGBTs are driven during X-ray. Only
V8, V9, signal on during the exposure
V13
8.5.3 Attention
The fans push hot air through the ventilation grill on the top of the generator.
The three fans are fixed on a metal mesh. The diameter of the holes is only 5mm, so the fans are not
reachable by human body parts from outside the inverter. Only trained service personnel are allowed to open
the front door of the inverter by using tools to access the fans.
8.5.4 Connections
Part Connection
Thermo-switch (50°C) ANS2 X12
Thermo-switch (80°C) ANS2 X11
Fans (1 / 2 / 3) ANS2 X13 (PIN: 1,2 / 3,4 / 5,6)
Table below provides information about calibration needed to perform when specific board/device is
replaced. For additional information about jumpers setting on the boards refer to chapter 7 or chapter 9 for
options. Depending on replacement part that was replaced test described below may need to below may
need to be checked again. (Such as generator accuracy, reproducibility, or linearity)
Remove screw of the front cover and unscrew from the front the rear panel locks in order at access parts
inside the generator.
- Jumper settings
(see chapter 9.4.1.5)
- Generator settings via service
software
(see document 06220011)
460319-001 Configuration:
IFC1 – Standard Interface Board
Contact Spellman
X-Ray Tube Configuration:
Replacement tube of the same tube
type:
- Stator connection
(see chapter 2.4.6)
- Tube calibration
(see chapter 3.2.4)
Replacement tube of another tube type:
- Phase shift capacitor
configuration
(see chapter 2.4.6)
- Stator connection
(see chapter 2.4.6)
- Generator with High speed
option HSS1
(see configuration HSS1)
- Generator settings via service
software
(see document 06220011)
- Tube calibration
(see chapter 3.2.4)
Contact Spellman Configuration for analog ionization
AEC Ion Chamber (AEC option)
chambers:
- Jumper settings
(see chapter 9.4.1.4)
- Generator settings via service
software
(see document 06220011)
- AEC calibration
(see chapter 9.4.2.4)
Configuration for digital ionization
chambers:
- Jumper settings
(see chapter 9.5.1.4)
- Generator settings via service
software
(see document 06220011)
- AEC calibration
(see chapter 9.5.2.3.1)
Contact Spellman High Voltage GmbH for appropriate revision of the replacement board/module in case the
marking of revision is not clear on the existing board/module.
Warning All inspections are done after the power is OFF, and circuit breaker if OFF
Operating console:
1. Check that symbols and markings on the console didn’t wear out to the point that they can’t be
recognized
2. Check functionality of the console keys/switches
3. Check Exposure switch and cable functionality. (Optional)
H.V Cables:
Calibration Verification:
Verify calibration of generator according to the tests below during installation and annual scheduled
maintenance. If any tests are out of spec recalibrate generator according calibration procedures on Chapter
3.2.4.
Measurements Setup:
• The measurements shall be taken with SID=100 cm or 40”.
• The HVL value should be in accordance to IEC /FDA requirement.
RAD Reproducibility:
The reproducibility test shall be performed twice for 3 point and 2 point technique with test conditions
below. Compliance shall be determined based on 10 consecutive measurements within an hour.
The reproducibility test shall be performed at 2 point technique with test conditions below. Compliance
shall be determined based on 10 consecutive measurements within an hour.
Test condition:
2 point technique 80kV attenuation material thickness shall be adjusted to have exposure time of about
100ms-150ms
n
xi [mR] x [uGy/mR] xi − x ( xi − x ) 2 CV
1
2
3
4 x
5 =(X1+X2+X3+ CV less than
6 X4+X5+X6+X 0.045
7+X8+X9+X1
7 0)/10
8
9
10
S= [ ∑ ( xi − x ) 2 / (n-1)2 ]1/2
CV=S/ x
Where:
S= Standard Deviation
n
y j − yk yi
LIN = yj = ∑
y j + yk i =1 n
where yj and yk are each the average value of consecutive 10 reading groups of yi [mR/mAs] values
which is the measured dose xi divided by the selected current time product CTP [mAs].
xi
yi =
CTP
Enter LIN on the Compliance Data Sheet and verify that all values of LIN are equal or less 0.090.
Answer the question on the Linearity on the Compliance Data Sheet.
M = Measured Parameter
M-1
Set 2 60 56.1 160 151 200 190 250 237 320 304 400 380 0.500 .507
63.9 169 210 263 336 420 .493
M-2
Set-4 80 75.3 400 380 500 475 630 559 800 761 mA* 5% 0.063 .066
84.7 420 525 661 839 (max) ±1mA .060
M-4
Set-5 90 84.9 63 58.6 80 74.8 100 94 125 118 160 151 0.250 .253
95.1 67.4 85.2 106 132 169 .247
M-5
Set-6 100 94.5 100 94 125 118 160 151 200 190 mA* 5% 0.100 .105
105.5 106 132 169 210 (max) ±1mA .095
M-6
Set-8 140 113.7 250 237 320 304 400 380 500 475 0.100 .105
126.3 263 336 420 525 .095
M-8
Set-9 130 123.3 10 8.1 13 11.9 16 13.8 20 17.6 25 22.4 0.320 .324
136.7 11.9 15.1 18.2 22.4 27.6 .316
M-9
Set- 100 94.5 10 8.1 13 11.9 16 13.8 20 17.6 25 22.4 1.000 1.014
10 105.5 11.9 15.1 18.2 22.4 27.6 .986
M-10
Set
M = Measured Parameter
Test # kVp mAs mAs mAs mAs mAs mAs Limits mAs mAs
Limits Limits Limits
Set-1 50 250 240 320 307 400 384 500 480
260 333 416 520
M-1
Set-3 70 50 45 63 57 80 73 100 91
55 69 87 109
M-3
Set-4 80 40 36.2 50 45 63 57 80 73
43.8 55 69 87
M-4
*) The fuse F3 on the PCB NAG2 is soldered. Replace the board NAG2 if this fuse is broken.
10.1.1 Introduction
There are two different construction stages available. The standard version has connections for 2 Bucky
devices and the optional version for 4 Bucky devices.
The configuration can be done by wire jumpers.
Optional connectors
Internal interface
connection to the ZST2
J1 wire jumper
24V closed
230V opened
The green framed parts are not assembled with the standard interface board NAG2.
JP3 and JP4 must not be closed together at the same time. The output of the optocoupler OK3 can be
damaged in this case.
The interface board NAG2 provides an interface voltage of 24V DC (max 80mA) on connector X13. This can
be used for bucky devices with 24V interfaces. The configuration must be changed if the bucky interface
needs another interface voltage.
The following three pictures show the configurations for different interface voltages.
The 24V DC (max 80mA) interface voltage is provided on the connector X13.
X13.1 (lower connector) = 24Vdc
X13.2 (upper connector) = 0V
J1_1 – J1_2 closed
JP1 opened
The 24Vac interface voltage is not provided by the generator in the standard configuration. An optional
24Vac power supply can be ordered.
J1_1 – J1_2 closed
JP1 closed
A separate delivered insulated cable can be used to connect the NAG2 connector X9 to the 230Vac
connector X5 on the sub-distribution in the case that the bucky devices need a 230Vac interface voltage. X5
on the sub-distribution is fused with 6,3A.
J1_1 – J1_2 opened
JP1 closed
WARNING:
A connection failure, a modification, an upgrade or disconnecting any part of the involved items
from the network/ data coupling may produce the following hazardous situations:
• The automatic brightness control may not reach the correct value
• The automatic exposure control may not terminate the exposure at the correct time
• The obtained image may be corrupted, not stored properly or not been triggered properly
• The communication with the imaging system may be corrupted
• Communication errors lead to error messages either on the generator or the imaging system,
which will block the further operation
WARNING:
The integrator / assembler is responsible to identify, analyze and control all additional risks arising
from the integration. Especially changes to the data coupling, any data coupling configuration
change, any connection or disconnection of additional items to the data coupling and any update of
connected equipment may introduce new risks and require a new analysis by the composer.
The integrator / assembler has to verify the correct integration according to the fluoroscopy imaging
system's service manual provided by its manufacturer.
- IFF1 (Interface Fluoroscopy): A programmable Interface with galvanic isolated inputs and outputs to
the imaging system.
- UNI1 (UNIversal adapter): An additional board to adapt the galvanic isolated inputs and outputs of
the IFF1 to the inputs and outputs of the TV system and the x-ray device.
The interface has many special in- and outputs to control a wide range of different TV-chains provided by
different manufacturers. It is possible to select presettings for special TV-chains in the generator setup.
All three modes work with manual and automatic dosage control.
10.2.2.4 Potentiometers
Potentiometer Function
R90 Adjustment of the analog optocoupler for continuous fluoroscopy. Adjusted in factory
R91 Adjustment of the analog optocoupler for continuous fluoroscopy. Adjusted in factory
10.2.3.5 Connectors
Connector Function
X10 Input for special applications
X11 Input for special applications
X12 Input for special applications
X13 Connection to the BA6 for TV-systems without an own automatic exposure control
X14 Film focus distance input
X15 Connection to motor driven collimator. Provided by optional control panel.
X21 RS232 communication port
X22 Connection for footswitch or to the x-ray device.
X23 Connection to the TV- system (parallel inputs/outputs)
X24 Connection to the TV- system (parallel inputs/outputs)
The jumper X40 must be closed if the TV-system delivers only one ABC signal for continuous and pulsed
fluoroscopy.
The output of the impedance converter on the UNI1 goes to the input of the isolated amplifier. The voltage
divider, consisting of R1, R2, R3, R118 and the two jumpers X3 and X4, must be adapted to the analog
output signal of the TV-system. To get a good and fast control of the kV in the automatic fluoroscopy mode,
the input voltage on test point X8 must be in a middle range between 1,5V and 3,5V if the dose rate is
correct. The voltage at test point X5 (GND/E = X18) is nearly the same as the voltage on test point X8 (GND
= X9).
The voltage divider is designed for a wide range of input voltages. The table below shows some samples
and the resulting jumper positions.
On the next pages the standard default hardware settings, connections to a TV-chain with x-ray device and
the timing diagrams are shown.
A foot switch or the x-ray device can be connected to connector X22 to give the start signal for exposures
and fluoroscopy.
X24 contains signals for special applications and for seldom systems. Most of the pins are reserved inputs
and outputs.
The jumper X28 must be set if the x-ray device sends only an exposure start signal (no preparation signal).
With a closed jumper X28 the signals for preparation and exposure are combined to one.
The exposure stops upon releasing the hand switch or when the exposure time has expired.
The information in this document refer to the “device type” “fluoro” = “fluoro”.
In the device type “no grid” it is possible to set a device to a device using fluoroscopy only without any
radiography functions. For this operation the checkbox "Enable device" has to be deactivated.
On the left side are the radiography settings for the fluoroscopy device and on the right side are the
fluoroscopy settings for the fluoroscopy device.
It is possible to work with an analog dose error signal from the TV-image system for a fast automatic
brightness control or with digital signals to change the high voltage in 1kV steps.
10.2.6.3.1 Automatic brightness control via the analog inputs “ABC_C_1” and “ABC_C_2”
The ABC values in the window on the left side (1) are for continuous fluoroscopy and on
the right site (2) are for pulsed fluoroscopy.
The regulation speed can be changed by the kp factor. 100 is the default value.
The “ABC set point” must be set to the values, measured on test point X8 for continuous fluoroscopy and on
test point X14 for pulsed fluoroscopy on the board IFF1, when the dose rate has reached the correct value. It
is also possible to measure the voltage with the service program in the test window of the IFF1(see Figure ).
The dose rate on the input of the image intensifier must be in an appropiate range. (for example 305nGµ/s at
70kV, normal zoom on an 11” image intensifier and a phantom of 25mm AL)
The IFF1 is preset for a analog dose error input signal in a range from +10 .. –10V. The IFF1 shifts the input
voltage in a positive range between 0 to +5V. This means, 0V from the TV chain results in an analog input
signal ABC_C of approx +3,2V.
If the input range of this signal is different the jumpers X3 and X4 must be changed.
For example:
During fluoroscopy at 70kV and a dose rate of approximately 305nGy the measured ABC value using the
test window is 2,84V. In this case set the “ABC set point” = 2.90. If the measured values vary a little bit
between two measurements the hysteresis has to be increased.
Warning!
The high-speed starter works with dangerous high voltages. The capacitors are still
charged, after turning off the generator
The discharge of the capacitors on the HSS1 module needs at least. 150s after switch-
off.
The capacitors must be discharged before the hardware settings can be changed.
It is absolutely essential to ensure that the tube is not rotating on high speed, if the
generator is turned off.
An uncontrolled deceleration of the anode can cause damage to the bearings of the
anode plate and thereby damaging the tube.
Turn on the generator immediately, when it is switched off during the tube is running
on high speed. The tube will be braked automatically after switching on.
- The connection of non-listed X-ray tubes is not permitted or require prior consultation with the
Spellman GmbH Bochum
- For the connection of two X-ray tubes, make sure that both can be operated with the same stator
voltage. A combination of X-ray tubes with different stator voltage is not possible
- The high-speed starts are limited to 4 per minute, to protect the connected X-ray tube.
- All hardware settings on the High speed starter module and the tube settings in the Generator setup
must be set to comply with the connected x-ray tube before switching on.
Verify:
o Stator voltage
o Phase shift capacitor
o Start time
o Brake time
o Anode data at 150Hz
o Maximum tube current
Types of operation:
- Anode braking
o 3s – 6s deceleration of the anode rotation below 3000 U/min
The anode will be accelerated with reduced energy on each exposure if the tube is running on high speed.
The high speed starter will reduce the rotor speed below 3000 r/min after the end of the continuous running
interval has been reached or if the preparation is stopped during the high speed start time.
The high-speed starter starts a safety-braking each time the generator is switched on.
10.3.3.2.1 X6 Tube 1
10.3.4.4 Potentiometer
Poti-Number Function
R44 Adjustment of the maximum allowed converter current
10.3.4.6 Jumpers
Connection Function
X11, X12, X13, X14 See chapter 9.3.7.1.4
Connection Function
X20, X21, X22, X23 See chapter 9.3.7.1.4
Connection Function
X3.1, X3.2, X3.3 See chapter 9.3.7.1.4
Connection Function
X16.1 – X16.2 and Wired jumper, disconnects the mains from the control
X16.3 – X16.4 unit and the power unit.
Disconnect only for factory setup.
TP Signal Function
TP1 INVFRQA\ Control signal of the Inverter IGBT V1 and V4
TP2 INVFRQB\ Control signal of the Inverter IGBT V2 and V3
TP3 / IBR IBR Active value of the inverter current
TP4 / +5V +5V Power supply +5V DC microcontroller
TP5 / GND GND Ground microcontroller
TP6 / ISOLL ISOLL Maximum valid inverter current (default 1,8V)
TP7 +UZK positive intermediate circuit voltage
TP9 / 12V +12V_EXT Power supply +12V (external) of the IGBT driver
TP10 / GND/12 GND_12 Power supply GND (external) of the IGBT driver
For a later assembly of the module the following parts must be built into the main distribution:
- Stator voltage
- Frequency 160/180Hz
- Continuous running interval
- Shift capacitor
- Start frequency of the high speed start
- End frequency of the high speed start
- Pulse width of the brake
- Timing of the continuous running
Settings:
- Stator voltage
- Frequency 160/180Hz
- Continuous running interval
- Shift capacitor
- Start frequency of the high speed start
- End frequency of the high speed start
- Pulse width of the break
- Timing of the continuous running
Important information:
- The DIP switches S3.1, S3.2, S3.7 and S3.8 are only for factory use. Other settings of these
switches can damage the tube.
The DIP switches S3.1, S3.2, S3.7 and S3.8 are only for factory use. Other settings of
these switches can damage the tube.
10.3.7.2.1 Dip switch S1 (tube1) and S2 (tube2) (separate settings for each tube)
The tube braking works using an AC current with a sloping frequency from 160Hz to 5Hz.
After a braking time between 3 to 6s the tube will rotate with less than 3000U/min. A deceleration to 0U/min.
is not possible.
The pulse width during the brake time may vary between 3ms to 10ms and can be set by the DIP switches
S1 for tube 1 and S2 for tube 2.
The brake time for each tube must be set in the generator setup.
The proper settings for different tube types are listed in the compatibility chart.
S1 S2 Function
Tube1 Tube2
S1.1 S2.1 Bit 0, Pulse width
S1.2 S2.2 Bit1, Pulse width
S1.3 S2.3 Bit 2, Pulse width
S1.4 S2.4 Not used
S1.5 S2.5 Bit 0, Start frequency of the high speed starter
S1.6 S2.6 Bit 1, Start frequency of the high speed starter
S1.7 S2.7 Bit 0, End frequency of the high speed starter
S1.8 S2.8 Bit 1, End frequency of the high speed starter
10.3.7.2.5 Settings of the end frequency of the high speed start after 1s
Values are valid for both versions, Standard and Option 1
S1 or S2 S1 or S2
Sx.7 Sx.8 End frequency / Hz
ON ON 160
OFF ON 166
ON OFF 174
OFF OFF 180
- Stator voltage
- Shift capacitor
- Start frequency of the high speed start
- End frequency of the high speed start
- Pulse width of the braking
- Timing of the continuous running
Next, the software settings in the generator service software must be set:
Use the service software to change the settings in the generator software. Use the data cable Ident-Nr.
06220244 for the connection between the board ZST2 and the service PC. All software settings are made on
the ZST2 of the generator.
Make a backup copy of the generator settings before changing the settings of the High-speed starter.
Accept the changes by pressing the button „Accept”, then close the window using the „OK“ button.
Set the power data of the anode in the window for the small focus using button „Focus 1“ and for the large
focus using button „Focus 2“
The power ratings must be in compliance with the original datasheet of the tube. The default settings set by
Spellman GmbH may vary from the ones of the used tube.
For example:
During preparation the tube starts every 500ms lasting 50ms with the maximum frequency but with low pulse
width and low energy.
The time of the restart is a product of the parameter “Anode cycle on time” and a factor depending of the
continuous running time:
For example:
During the continuous running the tube will now restart for 150ms before starting the second exposure. The
restart works with the maximum frequency but with low pulse width and low energy.
Possible causes:
- The hardware settings do not comply with the connected tube. Check the documentation of the tube
stator.
o Check all hardware settings on the high-speed starter.
o Measure the resistance of the stator windings at the cable side at the connector to the high
speed starter
example:
On the cable side at the connector Resistance
W – V (between Pin 2 and 3) 21 Ohm
U – V (between Pin 1 and 3) 51 Ohm
U – W (between Pin 1 and 2) 72 Ohm
- Jumper JP1 is closed (only closed during electric tests in the factory)
Possible causes:
- Tube with high stator impedance. Unknown tube and stator type
o Confer with Spellman GmbH Bochum
- Tubes have a thermo contact inside of the stator windings which open the stator at a temperature
above 70°C.
o Wait until the tube is cooled down.
o The most tube manufacturers allow only 2 high-speed starts per minute.
Possible causes:
- The hardware settings of the stator voltage are incorrect.
o Control the settings of DIP switch S3.3 and the wire junction on connector X16.
Possible causes:
- Broken IGBT in the inverter of the HSS1
o Change the complete module
10.4.1.1.1 Type A
- Measuring chamber type A: non-integrating preamplifier (VV20 or VV21):
10.4.1.1.2 Type B
- Measuring chamber type B: integrating preamplifier:
10.4.1.1.3 Type C
- Measuring chamber type C: integrating preamplifier.
- This chamber use low active control signals.
Attention: This chamber has a pin assignment other than VACUTEC chambers of type B.
The Spellman measuring chamber cables doesn’t work with this chamber!
10.4.1.1.4 Type D
- Measuring chamber type D: non-integrating preamplifier without adjustment of the amplification
by field-combination.
Attention: This chamber has a pin assignment other than VACUTEC chambers of type B.
10.4.1.2 Jumper settings for Measuring Chamber VACUTEC BAK70145 / 14500 42,
43,04
For this measuring chamber it is important to set the measuring signal to a positive gradient. Therefore
make sure that jumper S4 on the measuring chamber is set to ON.
- Amplification
Amplification S1 S2 S3
0,1V/µGy (10V=100µGy) OFF OFF OFF
0,5V/µGy (10V=20µGy) ON OFF OFF default
1V/µGy (10V=10µGy) OFF ON OFF
2V/µGy (10V=5µGy) OFF OFF ON
Densitometer, Dosimeter
Figure 9-7 shows the measurement set-up to adjust the AEC using a dosimeter.
Phantom 4, 1mm Cu
Phantom 1, 25 mm Al
Phantom 3, 50 mm Wasser
Phantom 2, 100 mm Wasser
Meßsignal
Belichtungsautomat
Measuring signal AEC
To adjust the AEC use the service mode of the AEC. To set this mode follow this procedure:
3. Switch the generator on and press the key at the same time.
item function
1 AEC-CHAMBER 1 adjustment of measuring chamber 1
2 AEC-CHAMBER 2 adjustment of measuring chamber 2
3 AEC-CHAMBER 3 adjustment of measuring chamber 3
4 AEC-CHAMBER 4 adjustment of measuring chamber 4
5 KV COMPENS., SCREEN 0 kV-compensation for film-screen-system 0,
9 START PHASE CORRECTION Correction factor for short exposure times, measuring chamber with
external preamplifier
The message AEC-CHAMBER 1, SCREEN 0(1-80):045 is displayed. The values in brackets designate the
parameter range.
item function
1 SCREEN 0 (1-80): adjustment of film-screen- system 0, e.g. 400 FFS
2 SCREEN 1 (1-80): adjustment of film-screen- system 1, e.g. 800 FFS
3 SCREEN 2 (1-80): adjustment of film-screen- system 2, e.g. 200 FFS
4 FIELD CORR. R(0-15): correction of right hand measuring field
5 FIELD CORR. L(0-15): correction of left hand measuring field
6 FIELD SWAP(0,1): Orientation of the measuring fields with respect to the X-ray beam
direction
7 PULSED EXP. (0,1): 1 = pulsed exposure
0 = single exposure
8 INTEGR. CH(0,1) 1 = measuring chamber type B, C
0 = measuring chamber type A, D
9 GRAYLEVEL (1-7) fine adjustment in 1/8 exposure points (7=+3/8, 4= normal, 1= -3/8)
Table 9-6: Context menu: adjustment of measuring chamber.
Important:
In the first place the type of the measuring chamber must be set in menu "INTEGR. CH(0,1)", because
otherwise all adjustment steps would lead to malfunction of the AEC.
M
1. Select "AEC-CHAMBER 1" with and press .
M
6. To store values press again.
The display shows: AEC SETUP DATA SAVED
Message AEC-CHAMBER 1, SCREEN 0(1-80):045 is displayed. Values in brackets define the parameter
range.
If the measured dose founded in menu item “SCREEN X (1-80) differs from the value given in the table,
the AEC must be fine adjusted in the menu item “GRAYLEVEL (1-7)”.
The default parameter value 4 is normal (+/- 0/8 exposure points). 7 means +3/8 exposure points from
the value you founded under AEC-CHAMBER X, SCREEN 0(1-80):...) and 1 means –3/8 exposure
points.
3. Repeat steps 1 and 2 until the measured dose complies with the values listed in the table.
4. Remove the measuring chamber of the dosimeter.
M
11. To store the parameters press again.
After successfully storing the parameter values AEC SETUP DATA SAVED is displayed
Repeat the whole procedure for film-screen-system 1 and 2 (SCREEN1, SCREEN2), if provided.
1. Put either the dosemeter measuring chamber or a cassette with film into the bucky.
2. Select the central measuring field only.
3. Set X-ray parameters to 1-point technique, 80 kV.
4. Take an exposure.
5. Log the measured dose or the density, respectively.
II III
1
3 linkes Meßfeld
linkes Meßfeld rechtes Meßfeld
left-hand field rechtes Meßfeld left-hand field
right-hand field right-hand field
2 Fußende
mittleres Meßfeld foot end mittleres Meßfeld
central field central field
Figure 9-8: PHILIPS measuring chamber, Orientation Figure 9-9: VACUTEC measuring chamber, Orientation
of the measuring fields in the direction of of the measuring fields in the direction of
the beam the beam
measuring
chamber orientation parameter "FIELD SWAP (0,1):"
field
left 1
0
right 3
PHILIPS
left 3
1
right 1
left II
0
right III
VACUTEC
left III
1
right II
Table 9-8: Spatial orientation and measuring fields
M
1. Select "AEC-CHAMBER 1" by and press .
2. Select context menu "FIELD SWAP(0,1):".
3. Set parameters according to Table .
M
6. To store values press again.
The display shows: AEC SETUP DATA SAVED
M
1. Select "AEC-CHAMBER 1" with and press .
M
6. To store values press again.
The display shows: AEC SETUP DATA SAVED
To adjust the next measuring chamber repeat all steps of chapter 9.4.2.4.
For each film-screen-system film density depends on tube voltage in a different way. This influence must be
compensated to get sufficient images. Compensation is done in the menu "KV COMPENS., SCREEN n" for
a tube voltage from 40 kV to 150 kV using 8 steps. Please keep in mind the minimum and maximum values
for tube voltage which has been set during installation.
The compensation is given in % of the parameter set at 80 kV. This implies automatically that the value set
at 80 kV must be 100%.
Set the parameters for tube voltages below and above 80 kV according to the procedure described below.
10.4.3.1 Preparations
1. Remove the AL- and CU-phantoms off the beam.
2. Place the phantoms 2 and 3 (100 mm water, 50 mm water) in patient's position.
3. Select 1-point-technique.
4. Select central measuring field.
5. Select standard film-screen-system.
M
6. Select menu "KV COMPENS., SCREEN 0" using and press .
Note:
It is recommended to start the procedure at 80 kV. The parameter value is assumed to be 100 because all
previous adjustments were done at 80 kV.
If films are available, skip chapter 9.4.3.2, because there is a non-linear dependence of tube voltage, dose
and film density.
At 40 kV you may get many faulty exposures. In this case it is recommended to extrapolate the values
instead of doing measurements.
Repeat the steps of this chapter for film-screen-systems 1 and 2 (SCREEN 1, SCREEN2).
M
1. Select "DOSE TEST AFTER 50MS" using and press .
2. Set parameter to a value between 1% (01) and 50% (50).
10.4.3.5 Startphase
In case of very short exposure times, the quality of the images may not be sufficient, especially when using
measuring chambers with external preamplifiers. If this happens a correction factor may be set to
compensate effects at the beginning of the exposure in the menu START PHASE CORRECTION.
M
1. Select "START PHASE CORRECTION" using and press .
2. Set exposure parameters at the control desk: 3-point-technique, 80 kV, 5 ms.
3. Select central measurement field.
4. Insert the film, take an exposure and process the film.
5. Measure the density with the densitometer. Density must fit into the interval of 1.6 to 1.8.
cm
6. If the density does not lie within this interval correct the parameter value by using buttons and
cm
, respectively, or enter the value directly.
7. Repeat steps 2 to 5 until density fits into the interval.
If all necessary steps are done to adjust the measuring chambers, the generator is set back to normal
operation mode.
The generator starts the WARM UP and is ready for normal operation.
In case errors occur during operation the messages "E034" to "E038" are displayed .
The status of the AEC is indicated by the LEDs V33-V35 and V57 on the BA6 board. Also hardware errors
are shown on the automatic exposure control board.
14 o o o Watch dog
15 o o o o not used
Table 9-10: Error messages indicated by LEDs on the BA6 (Binary Coding).
Remaining LEDs
V49 V50
Serial datex BA6 – EDITOR
* * for CAN Bus Interface the LEDs are not used.
V53 V54
* * Serial datex BA6 – extern interface for service
V111 ° - 15V OK
V110 ° + 15V OK
V48 ° + 5V OK
Measuring chamber 1
Measuring chamber 2
Measuring chamber 3
Measuring chamber 4
Jumper X32
VV20 - closed - 300V activated
VV21 - opened - 300V deactivated
Up to 5 measuring chambers may be connected to the AEC using RJ45 patch cable (BA7 type A).
*Up to two 5-field chambers (BAK70139) may be connected to the BA7 type B using a 15-pole sub-D
connector.
- Amplification
Amplification S1 S2 S3 S4
100 pulses = 100µGy OFF OFF OFF OFF
100 pulses = 20µGy OFF OFF OFF ON
100 pulses = 10µGy OFF OFF ON OFF
100 pulses = 5µGy OFF ON OFF OFF
100 pulses = 2,5µGy ON OFF OFF OFF default
Pin
1 0V
2 +15V
3 signal pulses RS485 A
4 right-hand field selection
5 Chamber select / Start
6 signal pulses RS485 B
7 left-hand field selection
8 central field selection
Cover shielding
Table 9-11: Pin assignment AEC measuring chamber connector, 8-pole RJ45
Pin
1 RS422-A
2 Enable left measuring field, low active
3 Enable center measuring field, low active
4 Reset / run, low active
5 RS422-B
6 Enable right measuring field, low active
7 Not used
8 Supply voltage (+11.5...16V)
9 GND
10 Supply voltage (+11.5...16V)
11 Portrait, low active
12 GND
13 Inverted, low active
14 Supply voltage (+11.5...16V)
15 GND
Cover shielding
Table 9-12: Pin assignment AEC measuring chamber connector, 15-pole sub-D
Jumper
X33 1-2 CAN BUS
Phantom 1: Aluminium block, 100 cm², 25 mm thickness; placed in front of the collimator
Phantom 2: Water, 100 mm thickness; placed in patient's position
Phantom 3: Water, 50 mm thickness; placed in patient's position
Phantom 4: Copper plate, 100 cm², 1 mm thickness; placed in front of the collimator
Densitometer, dosemeter
Figure shows the measurement set-up to adjust the AEC using a dosemeter.
Phantom 4, 1mm Cu
Phantom 1, 25 mm Al
Phantom 3, 50 mm Wasser
Phantom 2, 100 mm Wasser
Meßsignal Belichtungsautomat
Measuring signal AEC
To adjust the AEC use the service mode of the AEC. To set this mode follow the procedure:
3. Switch the generator on and press the key at the same time.
4. enter the “key code” in code number 300 (default is 0000)
Item function
1 AEC-CHAMBER 1 adjustment of measuring chamber 1
2 AEC-CHAMBER 2 adjustment of measuring chamber 2
3 AEC-CHAMBER 3 adjustment of measuring chamber 3
4 AEC-CHAMBER 4 adjustment of measuring chamber 4
5 KV COMPENS., SCREEN 0 kV-compensation for film-screen-system 0,
9 START PHASE CORRECTION Correction factor for short exposure times, measuring chamber with
external preamplifier
The message AEC-CHAMBER 1, SCREEN 0(1-80):045 is displayed. The values in brackets define the
parameter range.
Item function
1 SCREEN 0 (1-80): adjustment of film-screen- system 0, e.g. 400 FFS
2 SCREEN 1 (1-80): adjustment of film-screen- system 1, e.g. 800 FFS
3 SCREEN 2 (1-80): adjustment of film-screen- system 2, e.g. 200 FFS
4 FIELD CORR. R(0-15): correction of right hand measuring field, *not to be used for 5-field chambers
5 FIELD CORR. L(0-15): correction of left hand measuring field, *not to be used for 5-field chambers
6 FIELD SWAP(0,1): Orientation of the measuring fields according to X-ray beam direction
7 PULSED EXP. (0,1): 1 = pulsed exposure
0 = single exposure
8 INTEGR. CH(0,1) 0 or 1 not used with BA7
9 GRAYLEVEL (1-7) fine adjustment in 1/8 exposure points (7=+3/8, 4= normal, 1= -3/8)
Table 9-17: Context menu: adjustment of measuring chamber.
The message AEC-CHAMBER 1, SCREEN 0(1-80):045 is displayed. Values in brackets define the range of
the parameter.
3. Repeat steps 1 and 2 until the measured dose is in agreement with the values listed in the table.
4. Remove the detector of the dosemeter.
M
10. Use to select menu "SAVE AEC SETUP" and press .
The display shows the message: SAVE AEC SETUP, TO SAVE PUSH -> 'M':
M
11. To store the parameters press again.
After successfully storing the parameter values AEC SETUP DATA SAVED is displayed
Repeat the whole procedure for film-screen-system 1 and 2 (SCREEN1, SCREEN2), if necessary.
1. Put either the dosemeter detector or a cassette containing a film into the Bucky.
2. Select the central measuring field only.
3. Set X-ray parameters to 1-point technique, 80 kV.
4. Take an exposure.
5. Note the measured dose or the density, respectively.
M
15. Select "SAVE AEC SETUP" using and press .
16. The display shows the message: SAVE AEC SETUP, TO SAVE PUSH -> 'M':
M
17. Press again to save the values.
The message AEC SETUP DATA SAVED appears in the display.
II III
linkes Meßfeld
rechtes Meßfeld
left-hand field
right-hand field
mittleres Meßfeld
central field
M
1. Select "AEC-CHAMBER 1" by and press .
2. Select context menu "FIELD SWAP(0,1):".
3. Set parameter according to Figure 9-13:.
M
5. Select "SAVE AEC SETUP" using and press .
Message SAVE AEC SETUP, TO SAVE PUSH -> 'M': is displayed
M
6. To store values press again.
The display shows: AEC SETUP DATA SAVED
M
1. Select "AEC-CHAMBER 1" by using and press .
M
6. To store values press again.
The display shows: AEC SETUP DATA SAVED
To adjust the next measuring chamber repeat all steps of chapter 9.5.2.3.1.
For each film-screen-system the film density depends on tube voltage in a different way. This influence must
be compensated to get sufficient images. Compensation is done in the menu "KV COMPENS., SCREEN n"
in the tube voltage interval from 40 kV to 150 kV within 8 steps. Please keep in mind the minimum and
maximum values for tube voltage which has been set during installation.
The compensation is given in % of the parameter set at 80 kV. This implies automatically that the value set
at 80 kV must be 100%.
Set the parameter for the tube voltage below and above 80 kV using the procedure described below.
10.5.3.1 Preparations
1. Remove the AL and Cu phantoms off the beam.
2. Place the phantoms 2 and 3 (100 mm water, 50 mm water) at the patient's position.
3. Select 1-point-technique.
4. Select central measuring field.
5. Select standard film-screen-system.
M
6. Select menu "KV COMPENS., SCREEN 0" using and press .
Note:
It is recommended to start the procedure at 80 kV. The parameter value is assumed to be 100% because all
previous adjustments were done at 80 kV.
If films are available, skip chapter 9.4.3.2, because there is a non-linear dependence of tube voltage, dose
and film density.
At 40 kV there may occur a lot of faulty exposures. In this case it is recommended to extrapolate the values
instead of doing measurements.
M
17. To store values press again.
The display shows: AEC SETUP DATA SAVED
Repeat the steps of this chapter for film-screen-systems 1 and 2 (SCREEN 1, SCREEN2).
M
1. Select "DOSE TEST AFTER 50MS" using and press .
2. Set parameter to a value between 1% (01) and 50% (50).
M
4. Select "SAVE AEC SETUP" using and press .
Message SAVE AEC SETUP, TO SAVE PUSH -> 'M': is displayed
M
5. To store values press again.
The display shows: AEC SETUP DATA SAVED
In case of very short exposure times the quality of the images may not be sufficient, especially if measuring
chambers with external preamplifier are used. In this case a correction factor may be set to compensate
effects at the beginning of the exposure in the menu START PHASE CORRECTION.
M
1. Select "START PHASE CORRECTION" using and press .
2. Set exposure parameters on the control panel: 3-point-technique, 80 kV, 5 ms.
3. Select central measurement field.
4. Insert the film, take an exposure and process the film.
5. Measure the density with the densitometer. Density must fit into the interval of 1.6 to 1.8.
cm
6. If the density does not lie within this interval correct the parameter value by using the buttons and
cm
, respectively, or enter the value directly.
7. Repeat steps 2 to 5 until density fits into the interval.
M
9. Select "SAVE AEC SETUP" using and press .
Message SAVE AEC SETUP, TO SAVE PUSH -> 'M': is displayed
M
10. To store values press again.
The display shows: AEC SETUP DATA SAVED
M
2. Select "SAVE AEC SETUP" using and press .
Message SAVE AEC SETUP, TO SAVE PUSH -> 'M': is displayed
M
3. To store values press again.
The display shows: AEC SETUP DATA SAVED
If all necessary steps are done to adjust the measuring chambers, the generator can be set back to normal
operation mode.
The generator starts the WARM UP and is ready for normal mode operation.
If errors occur during operation the messages "E034" to "E038" are displayed .
E035 Exposure too short exposure time <2ms Dose too high
E036 Exposure terminated The exposure has been e.g. user released the exposure
interrupted by an external button, the door to examination
cause room has been opened, dose too
low, exposure parameters in 2-
(mAs) or 3-point technique (ms)
chosen are too small ...
E037 Dose too low Safety termination at the collimator is closed, exposure
beginning of an exposure parameters (kV) too small, not
measuring anything
E038 delay between two series of exposures have
exposures too long been terminated (>3s)
Table 9-19: Error codes displayed on the control panel
14 o o o Watchdog
15 o o o o Not used
V53 V54
* * Transmitting data from BA7 to MP15/16 (not used / *not available on BA7 type B)
V110 ° + 15V OK
V48 ° + 5V OK
Introduction:
The interface board provides GUI control of the system via either Ethernet or Serial communications.
For Ethernet the default IP Address of the board is 192.168.1.4.
The computer initially needs the IP Address set to 192.168.1.10:
• Control Panel
• Device Manager
• Ethernet: Change IP Address to 192.168.1.10
The interface also allows the use of an optional HSS2 control module.
TOP VIEW
JB9 14 Optional Configurable 14 position dual in line connector, 0.1” center. This connector has
Interface. unused I/O’s which can be configured as required.
Pin Signal description
NOTE: This is not in use.
1 Relay 1: N.O. contactor
2 Relay 1: Common
3 Relay 1: N.C. contactor
4 Spare
5 Open Drain Output
6 Open Drain Output
7/12/13 Not Used
/14
8 Control Input
9 Control Input
10 Control Input
11 Open Drain Output
JB10 8 Ethernet Connector 8 position RJ-45 connector, Standard Ethernet connection
The default IP Address of the board is 192.168.1.4
JB13 10 Mini-Console Interface 10 position dual in line connector, 4.20mm center. “Mini-Fir Jr.” series
for generator control.
Pin Signal description
1 Power On Command
2 Prep On Indicator, LED Driver
3 Exposure On Indicator, LED Driver
4 Power on Indicator, LED Driver
5 Ground
6 Power Off Command
7 Prep On Command
8 Exposure On Command
9 +12V Standby
10 Buzzer
SIDE VIEW
Interface Board Setup: The unit will have 2, 3, 4 or 5 cables connected depending on the options and
configuration. If the interface needs to be replaced or moved, note which cables are installed prior to
removing them. When the interface is reinstalled connect the required cables back in.
1. JB2: RS232 interface. Optional.
2. JB6: Power and communication for interface board. This connector is always present.
3. JB10: Ethernet. Optional.
4. JB12: HSS. Optional.
5. JB13: Mini Console Interface. Optional.
JB12
JB10
JB6
JB2
JB6
Connection to
Operating console
JB10: Ethernet
Option (OEM)
JB12: HSS2
Option
Generator can be controlled from the GUI interface with the Mini Console Control Interface.
The Connection Select dialog box is set for Ethernet when set as above. Click the connect button when
finished.
After the communication interface has been established the GUI control need to be set for Black Box mode
or Control Desk mode. This is set by clicking the “Oper” box as shown below.
A Service Password is required to access the Operation Mode Change function. After the “Oper” box is
clicked the Password request box will open as follows:
The Operations Settings dialog box will open and the operation mode can be selected.(Black Box Mode
emulates Interface Box)
Select and click the “Done” box when finished.(Control Desk Mode is not in use)
3. Preparation 4. Exposure
(PREP)Button Button
1. Power On
3. X-rays On
2. Prep On
Optional Hand
Held Exposure
Switch
Mini Console:
1. Push and hold down the Prep button. The Prep indicator will flash on/off. When Prep is completed,
the indicator will stay on in a steady state.
2. While still holding the Prep button down, push down on the exposure button. The exposure indicator
will illuminate and an audible noise sounds. After exposure is completed both indicators will turn off.
2 Once the Prep is completed, push the switch completely down. The exposure indicator will
illuminate. After exposure is completed both indicators will turn off
Warning!
The high-speed starter works with dangerous high voltages. The capacitors are still
charged, after turning off the generator
The discharge of the capacitors on the HSS 2 module needs at least 150s after switch-
off. NOTE: There is a Neon Lamp, DS1 located on the power board. Never service the
unit if this lamp is illuminated.
The capacitors must be discharged before the hardware settings can be changed.
It is absolutely essential to ensure that the tube is not rotating on high speed, if the
generator is turned off.
An uncontrolled deceleration of the anode can cause damage to the bearings or glass
of insert thereby damaging the tube.
- The high-speed starter HSS2 works only with tubes that are listed in the compatibility list.
- The connection of non-listed X-ray tubes is not permitted or require prior consultation with the
Spellman GmbH Bochum
- The operator should limit the high-speed starts to 2 per minute, to protect the connected X-ray tube.
- All hardware settings on the High speed starter module and the tube settings in the Generator setup
must be set to comply with the connected x-ray tube before switching on.
Verify:
o Stator voltage
o Start time
o Brake time
o Anode data at 180Hz
o Maximum tube current
Types of operation:
The high speed starter will reduce the rotor speed to below 3000 RPM after the end of the continuous
running interval has been reached or if the preparation is stopped during the high speed start time.
Top View
+
1
3
REV. B[ ]
JB4
-
S/S, REV. B
JB5
POWER
POLE
CONNECTO
RS
TB1 PINS:
1 2 3 4
TB1: External connection to the X-Ray Tube Anode Motor. Use ring lugs TE Connectivity, PN 320634 or
equivalent.
10.3.3.2 Switches
Switch Function
Number
W1 the microprocessor
10.3.3.3 Jumpers
TP Signal Function
TP1 +3.3V Bias Supply
TP2 +15V Supply Control signal of the Inverter IGBT V2 and V3
TP3 GND Ground
TP4 +5V Bias Supply
The HSS2 now needs to be configured for the connected X-Ray tube.
Turn on MAINS. Enable power to the system via the the Mini-Console. In ≈10 seconds the GUI will read
“System Functional”. The Status indicator will change from Blue to Green.
The unit will finish the startup sequence in ≈6 seconds. The GUI will then be displayed as shown.
The “Success” dialog box should open. Click on the “OK” button.
- The anode motor is not connected properly. One or more of the leads is not connected.
o Measure the resistance of the stator windings at the cable side at the connector to the high
speed starter
example:
On the cable side at the connector Resistance
W – V (between Pin 2 and 3) 21 Ohm
U – V (between Pin 1 and 3) 51 Ohm
U – W (between Pin 1 and 2) 72 Ohm