MODEL JF-10

PORTABLE X-RAY UNIT

OPERATION MANUAL
PORTABLE X-RAY UNIT OPERATION MANUAL

INDEX

一、INTRODUCTION……………………………………… 1

二、SPECIFICATIONS…………………………………… 1

三、INSTALLATION AND ADJUSTMENT………………… 5

四、OPERATION METHOD………………………………… 7

五、GENERAL DESCRIPTION OF PRINCIPLE ……… 8

六、ATTENTION CAUSE………………………………… 9
PORTABLE X-RAY UNIT OPERATION MANUAL

INTRODUCTION
PAGE 1

ⅠINTRODUCTION
This unit is a portable diagnostic X-ray unit r-ray-proof with a single

focus. It can be use for medical fluoroscopy and radiography on bone, heart,
skull and foreign body in the body, etc. It is easy to assemble and operate.

Ⅱ SPECIFICATIONS
1. Operation condition of the unit
Environmental temperature: 10-40℃
Relative humidity: 30%-75%
Atmospheric pressure: 70-106kPa
2. Power condition: AC 220V±10%(single phase) ,50Hz/60Hz,
Net resistant ≤2Ω
3. The maximum tube voltage: 75kVp
4. The maximum tube current : 10mA
5. Selection scope of time: 0.2-9.9sec.
6. The focus of X-ray tube:1.5×1.5mm2
7. Horizontal turning angle of the X-ray unit head: 360º
8. X-ray unit head turning angle around the horizontal axis: 270º
9. Stretching & retracting range of the X-ray unit head: 700mm
10. Weight: 25Kg
11. Working current of the equipment: 5A
PORTABLE X-RAY UNIT OPERATION MANUAL

SPECIFICATIONS

PAGE 2

X TUBE

组合式机头外形图
PORTABLE X-RAY UNIT OPERATION MANUAL

SPECIFICATIONS
PAGE 3
PORTABLE X-RAY UNIT OPERATION MANUAL

PAGE 4
PORTABLE X-RAY UNIT OPERATION MANUAL

INSTALLATION AND ADJUSTMENT


PAGE 5

Ⅲ ASSEMBLY

1.Take out everything from the case.


2. Latch the case tight, place it on a sound spot.
3. First insert the horizontal track into the horizontal hole of the fixture
and fix it. Then relax the behind fender and insert it into the hole with a
angle of 45º and screw tight, preventing it from sliding out.

4.Attach the controller to right side of the case for balance. Insert the
vertical frame into the square hole on the case and ensure to put the
frame into the bottom of the case.
PORTABLE X-RAY UNIT OPERATION MANUAL

PAGE 6

5.Unit the angular arm, fit it upon the horizontal assembly and screw
tight.

6.Set up the X-ray generator and the fluorescent screen. Use the
shutter when fluoroscoping and the cone for radiography.

7.Connect all the wirings properly.


PORTABLE X-RAY UNIT OPERATION MANUAL

OPERATION METHOD
PAGE 7
Ⅳ TRAINING AND ADJUSTING
1. TRAINING
A new or laid-off unit should to through a training process for
heightening the degree of vacuum in the x-ray tube, there by prolonging its
years of service.
Begin fluoroscopic exposure from the lowest KV gradually up to its
highest, exposing once on each KV step. If all is in order, training is finished.
If there appears any abnormal case, repeat the process until the desired
result is gained.
2. ADJUSTING
(1) Adjustment of the fluoroscopic current.
Turn the power switch to make sure the voltage is 220V and turn the
KV switch to 65KV. The amperemeter should display 3-5mA when
pressing the footswitch. If it is over, adjust the adjustment ring on the
resistor R1 in the controlling box.
(2) Adjustment of the radiographic current.
Turn the power switch to make sure the voltage is 220V and turn the
KV switch to 75KV. The amperemeter should display 10mA when
pressing the timer. If it is short or over, adjust the another adjustment ring
on the resistor R1 in the controlling box.
Note: Before adjusting the ring of the resistor, loosen the screw (don’t
take it off ) and lightly shift the ring until the sought-for position is reached.
Then tighten the screw back.

1 2

Note:1 stand for fluoroscopic resistance ring


2 stand for radiographic resistance ring
PORTABLE X-RAY UNIT OPERATION MANUAL

GENERAL DESCRIPTION OF PRINCIPLE


PAGE 8
Ⅴ DIRECTIONS FOR OPERATION
Before the current is on , be sure that the unit as a whole is well connected
with the ground.
1.Fluoroscope
(1) Turn the switch to
(2) Turn power switch till it reads 220V.
(3) Select the required KV.
(4) Let the patient stand or sit between the x-ray tube and screen.
(5) Press the foot-switch continuously which means fluoroscope. The
doctor can observe through the screen.
(6) Cut off the source and turn the switch to its lowest position.
2.Radiography
(1) Turn the switch to
(2) Turn power switch till it reads 220V.
(3) Select the required KV and exposure time.
(4) Let the patient stand or lie between the x-ray tube and photo.
(5)Press the timer continuously until it stops automatically. Then
Develop the photo.
(6)Cut off the source and turn the switch to its lowest position.
Ⅵ GENERAL DESCRIPTION OF PRINCIPLE
Referring to the principle diagram of the unit’s circuit. Put AC power on
and turn the power switch on. The auto-transformer T is powered on , adjust
the line voltage regulator S until the voltage display screen is 220V Regard
T1 as power, and primary coil of high voltage transformer T2 as load, after it
is powered on through resistor R2 and the counter-voltage for the X-ray
tube.
PORTABLE X-RAY UNIT OPERATION MANUAL

PAGE 9
..Ⅶ TROUBLES AND BEMEDIES
If the equipment has some problems, many of these problems can be
corrected using the following problem guide.
1. Power on lame will not light ----First be sure the key switch is turned on,
then check to be sure it’s plugged in completely and check the building
circuit breaker or fuse.
2. Lamp of the timer and audible signal are not energized by pressing, the
timer----check that the power on lamp is on (If not ,see item 1).
3. Light or blank films doesn’t work-----Be sure the power on and audible
signal operate normally (If not , see item 1,2). When an exposure is
made, be sure that the timer is depressed during the entire exposure. Do
not release it until the exposure e terminates automatically.
4. X-ray doesn’t stop after the selected time -----The lamp of the timer or
audible signal does not terminate automatically, immediately , release the
timer, turn power switch “OFF’ and disconnect power.
The above steps are taken, the problems have been not soluted ,
require the technicians to repair it.
Ⅷ ATTENTION CAUSE
1. In order to be safe during operation, a good ground connection wire
should be fixed.
2. Because the heat capacity of the small model X-ray unit is somewhat
limited, it is suggested that operated with intervals.
3. During disassembling the machine , first disassemble the unit head,
then the other parts.
4. After operation , cut off the power supply , put every selector on the
controlling box to its lowest position, put the X-ray tube head into balance
position.
5. Attention to keep the unit well, and take all kinds of the parts slightly.
S4 S6

4 24V

XP3
PAGE 10

3 12V XP4
T2
80 82
2 5V P
16
H1
0V
S29
1 7 1 7
OFF
~K
S1 ON 15 240V 1 7
S19 NAME TYPE
14 235V 14 XS3,4
13 230V 13
R1 RESISTANT RXQ-T-10-1.5K
12 220V 12
11 210V 11 R2 RESISTANT RXQ-T-10-300Ω
10 202V 10
地 83
XP1-3 V2 VARIABLE
9 193V
R3 -RESISTANT
BC-25W-500Ω
S2
8 182V 8 R2 XP1-2 XP3-11 XP3-12
R4 M 84 R4 RESISTANT RXYC-20-24Ω
7 168V 7
6 154V
XP1-1 SWITCH KS30-1-8
24Ω 300Ω S1
K-1 J1-3 J1-2
S2 SWITCH KS30-1-7
K2
PV K-2.3 V1 D1
1 C1 S3 SWITCH KN1
W 81 2CZ5A/200V
S3 拍片 V1 DIODE
29V
XP1-5 0 86

7805
5 9 C4 85 88
3 6 D2
T3 BA1.BA2 FUSE
0V 2 5 8
1 V2
XP1-4 5V X-RAY TUBE XD2-1/85
F
R3 1 4 7 J1 J2-3

XP2-8
C3 87 89
BA1 BA2 500Ω K4 K RELAY LY4AC/24V
透视 XP2-6 XP2-7
R1 POWER LAMP
H1 XDX-1-6.3V
XP2-9
1.5K H2 J2-2 (GREEN)
POWER LAMP XDX-1-6.3V
LB J1-6 H2 (YELLOW)
XP1
Q1
5V XP.XS PLUGSOCKET
3 2 1 J2
D3 PCB CIRCUIT BOARD
R6 XP5
PA D4
1 2 3 4 5
3 4 PV VOLTMETER 85L1 0-250V
3 2 1 1 555 1 2 3 4 5
XS1 XS2 8 XS5 PA AMPEREMETER 85C1 0-20mA
2 IC1 6
5 4 3 2 1 R5

5 4 3 2 1 S4 TIMER
XP6
PORTABLE X-RAY UNIT OPERATION MANUAL

W1 C2
XP2 S5 FOOT-SWITCH

W HEAT FUSE 250v-10A-85 C


地 L N

JF10 ELECTRICAL PRINCIPLE DIAGRAM


PORTABLE X-RAY UNIT OPERATION MANUAL

PAGE 11

The encadement listing of model JF10 portable X-Ray Unit

Serial number name number


1 controller 1
2 The head of X-Ray 1
3 The collective light canister of 1
Photography
4 Calculagraph 1
5 Foot brake 1
6 Power line 1
7 The line of the head and controller 1
8 The line of the ground 1
9 The level trajectory 1
10 Syphon(long,short) 2
11 The pieces of the stand pole 1
12 Coalition axes 1
13 Leaden door 1
14 Fluorescent screen 1
15 Knighthead 2
16 Drier 1
17 Operation manual 1
18 Eligible certification 1
19 Fuse 4
20 Aluminium piece 0.5mm/2mm 1/piece