You are on page 1of 52

Syringe Pump ARGUS 606 S en

Service Manual

14.543_E_SM_en_A606S_4.8X
Table of Contents

1. Introduction....................................................................... 1
1.1 General................................................................................ 1

2. Configuration and Maintenance Mode........................... 2


2.1. General................................................................................ 2
2.2. Entering the Configuration and Maintenance Mode ............ 2
2.3 Pump Settings...................................................................... 2

3. ARGUSservice Software.................................................... 3
3.1. General................................................................................ 3
3.2. Use of the ARGUSservice Software....................................... 4
3.3. Configuration Indices........................................................... 6
3.3.1 Description of configuration settings with flags.................... 10
3.4. Factory Settings.................................................................... 12
3.4.1. Software Versions................................................................. 12
3.4.2. ARGUS 606 S / Software 4.80.............................................. 12
3.5 Configuration of User-Syringes............................................. 13
3.6 TOOLBOX............................................................................ 15
3.7 History Log........................................................................... 18
3.8 History Example................................................................... 19

4. Serial Communication....................................................... 20
4.1. General................................................................................ 20
4.2. Serial Communication Protocol............................................. 20
4.3. Installation of an optional interface board............................ 20
4.4. Pump connections ............................................................... 21

5. Software Updates............................................................. 22
5.1. Firmware Update Procedure................................................. 22
5.2. Re-import the saved configuration file.................................. 27
5.3. Display Language................................................................. 29
5.4. Safety Aspects...................................................................... 30
5.5. Upgrade Procedure from a BASIC to a TIVA Pump................ 30

6. Maintenance...................................................................... 31
6.1. General................................................................................ 31
6.2. Keypad and Display Test....................................................... 31
6.3. Factory Base Syringe Calibration........................................... 33
6.4. Syringe Calibration............................................................... 34
6.5. Pressure Calibration.............................................................. 36
6.6. Pump Accuracy Measurement.............................................. 38
6.7. Correction of the Battery Capacity ....................................... 38
6.8. Fault Codes.......................................................................... 39
6.9. Message “Safety Check is due”........................................... 40

7. Replacement of Parts........................................................ 41
7.1. General................................................................................ 41

8. Wiring Diagram................................................................. 43

9. Bloc Schematic................................................................... 44

10. Safety Standard Check...................................................... 45

11. Repair Order Form (50.123.A)........................................... 46

14.543_E_SM_en_A606S_4.8X 3
4 14.543_E_SM_en_A606S_4.8X
1. Introduction

1.1 General

Important!
This service manual is intended for the exclusive use of authorized persons who have been trained in maintenance and repair of the
ARGUS 606 S syringe pump by CODAN ARGUS AG.

The service manual is meant to be used together with the user manual.

Important!
CODAN ARGUS AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 606 S
syringe pump by a non-authorized person.

Warning!
The ARGUS 606 S syringe pump may only be used with spare parts, accessories, consumables and syringes with Luer-Lock connec-
tions recommended by CODAN ARGUS AG. The functional safety of the pump is not guaranteed if non approved materials are
used. The safety of the patient may be endangered.
This manual contains the latest data available. It is subject to changes in accordance with technical improvements.

CODAN ARGUS AG
Oberneuhofstrasse 10
CH-6340 Baar
Phone +41 41 785 09 44
Fax +41 41 785 09 40
codan@codanargus.com
www.codanargus.com

14.543_E_SM_en_A606S_4.8X 1
2. Configuration and Maintenance Mode

2.1. General

Warning!
Changes in the configuration mode (with or without the “ARGUS service” PC utility tool), preventive maintenance measurements or
repair constitute a modification of the pump and may only be carried out by authorized persons!

After each change of a configuration, a function check and a control measurement has to be performed!

2.2. Entering the Configuration and Maintenance Mode

To enter the configuration and maintenance mode hold the 3rd numeric Key pressed while pressing the ON/OFF Key:

The factory setting of the PIN code is 0 unless otherwise specified by customized configuration. It is recommended to define a PIN
code to prevent unauthorized users from entering.

Press now the START/STOP key to get access to the submenus; use the MENU key to switch from one menu to the next:

2.3 Pump Settings

The first menu after entering the configuration and maintenance mode is “Pump Settings” in which the following adjustments are
possible:

1.* Display Contrast: 0 – 15 / factory setting varies depending on result (default is 7)


2.* Display Backlight: 0 – 255 / factory setting: 240
3.* Buzzer Volume: 0 – 10 / factory setting: 10
4. Firmware Update: Execute (see chapter 5.1)

*for a temporary change of these settings during normal infusion mode, see chapter “Programmable Options” in the User Manual.

The handling of the other menus “Drive Calibration”, “Pressure Calibration”, “Keypad & Display Test” is described in chapter 6
“Maintenance”.

2 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software

3.1. General

For easy and quick configuration procedures it is recommended to use the PC software ARGUSservice. This utility is user friendly PC
software with various functions working over the PC serial COM port. This Windows based software allows you

– to change configurations
– to do a software upgrade
– to set the pump clock
– to set or change a PIN code
– to read and print out the pump’s history
– to replicate easily pump configurations
– to set the service interval
– to download clear text in a language different from English
– to re-program the serial number
– to download medication libraries

The up-to-date and clearly structured design of this self-describing PC-tool allows a very easy and rapid modification of the
ARGUS 606 S, ARGUS 600 S, the ARGUS 707 V and ARGUS 708 V pumps as well as the ARGUS docking stations. This software is
available from your local distributor or directly from CODAN ARGUS AG.

REMARK:
The ARGUSservice is designed for the use with device software versions greater or equal to 4.00.

Warning!
The syringe pump has to be disconnected from the patient before and while the serial interface cable is connected to the pump.
The connection of the ARGUS 606 S over the serial interface RS232 is done by connecting the interface cable (REF 10.093 / 600951)
to the serial interface outlet of the serial PC-COM port.

Before the software can be started make sure that

– the pump is connected to mains


– the interface cable is connected to the pump’s serial interface RS232 and your PC
– the pump is in the configuration mode (see chapter 2.2)

Warning!
After each change of a configuration, a function check and a control measurement has to be performed!
An active communication between the pump and the PC is not apparent on the pump.

14.543_E_SM_en_A606S_4.8X 3
3. ARGUSservice Software

3.2. Use of the ARGUSservice Software

Please verify that you are in possession of the latest ARGUS ser-
vice version. Updates can be downloaded from the Login section
(password protected) under www.codanargus.com.

Open the software tool and click on the respective product type.

1. Click on the box “Start Configurator”

4 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software

2. Click on the box “Configuration”

The configurable indices are divided in chapters. CODAN ARGUS AG / Switzerland


REPAIR ORDER FORM

Purchase order / Proforma invoice number:


Use the “Edit” box to make configuration changes and enter
“0” if no pin is predefined. Customer name and address:

Use the “Save” box to save the configuration file.


Name of contact person: Tel. number:

Device: ARGUS 717 V ARGUS 718 V

Use the “Report” button to create an overview of the present ARGUS 707 V
ARGUS 708 V
ARGUS 606 S
ARGUS 100 P

configuration. ARGUS 600 S


ARGUS 60 P
ARGUS 100 M
A414

Use the “Email to ARGUS” button to send the “configuration ARGUS 60 M

Accessory:
A404 Serial Number:

Serial Number / Production code:

protocol” for analysis to the support center of CODAN ARGUS. Detailed failure or problem description:

Expected work / repair to be done:


Repair
Warranty repair
Replacement

Other Description:

Date: Signature:

50.123.B RepairOrderForm.xls 3.06.2008 CK 50.123.A

14.543_E_SM_en_A606S_4.8X 5
3. ARGUSservice Software

3.3. Configuration Indices

The configurations shown below may vary from the set-


tings of your pump. Factory settings are listed in chapter 3.4.

1. “Menus”
The menus below must be activated in order to make them
available on the pump.

2. “Display”
Index #366 allows displaying the pressure in units different from
mbar.
Index #524 makes for the entry of a specific text containing a
maximum of 16 characters. This text will appear on the display
at switching on the pump.

3. “Handling”
The indices below allow various settings to specify the handling
of the pump.

Warning!
It is strongly recommended to activate index #5 for pumps be-
ing used with TIVA or dose calculation to ensure a maximum of
safety during a syringe change.

4. “Functionality”
The indices below refer to KVO, bolus and pre-alarm functions.

6 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software

5. “Technical”

Warning!
For safety reasons it is strongly recommended to keep index #19
“Buzzer tone at start of infusion” and #44 “Automatic pressure
release after occlusion” activated.

If index #1614 is set on 100%, #1615 is not effective any longer.

A more detailed description of index #368 can be found in


chapter 6.7 “Correction of the Battery Capacity”.

6. “Communication”
See chapter 4.2 “Serial Communication Protocol”

7. “Calibration Syringe”
The flag of index #508 and #509 must be on “YES”; otherwise
a syringe calibration is requested.

8. “Calibration Pressure”
The flag of index #502 and #503 must be on “YES”; otherwise
a pressure calibration is requested.

The indices #502/503/508/509 cannot be modified.


A calibration flag showing “No” will be indicated as “CtrL”
alarm (“Safety check is due”) on the pump display each time
after the device was switched on.
The device needs to be calibrated

14.543_E_SM_en_A606S_4.8X 7
3. ARGUSservice Software

9. “Calibration Display”
The two indices allow a permanent adjustment of the backlight
brightness and the display contrast.

10. “Statistic / Parameters”


The following parameters provide information about the previ-
ous infusion details.

11. “Statistic / Statistic”


The following parameters summarize operating data.

12. “ARGUS Presets”


Index #396 shows the serial number.
A missing serial number will be indicated as CtrL alarm (“Safety
check is due”) on the pump display each time after the device
was switched on.

13. “Syringe Limits”


The standard limits are defined per syringe size (10 – 50/60 ml)
2 ml and 5 ml syringes are not yet validated for usage.

8 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software

14. “Syringe Data Base”


Please use the “Syringe Button” to get access to this menu.

All validated and approved syringes are listed from position 1 –


29. The dimensions ensure the recognition of the syringe and
the correct display of each brand on the pump.
Enable those syringes which shall be used on the pump.
No 30 – 40 are available to define user specific syringes (see
chapter 3.5)

14.543_E_SM_en_A606S_4.8X 9
3. ARGUSservice Software

3.3.1 Description of configuration settings with flags


Index Setting Description

2 Key ON/OFF only in STOP mode Yes: The key ON/OFF is disabled. Pump can be turned off only in STOP
usable mode.
No: The key ON/OFF is enabled. Pump can be turned off in STOP and
RUN mode.
3 Rate change allowed only in Yes: The infusion rate can be changed only in STOP mode
STOP mode No: The infusion rate can be changed in STOP and RUN mode
5 Ask for new therapy after syringe Yes: Question “Continue Therapy? YES/NO” appears after a syringe
change change.
No: Question “Continue Therapy? YES/NO” will not appear after a syr-
inge change
6 Nurse call settings Yes: A static alarm is activated for nurse call.
No: A single pulse alarm is activated for nurse call.
16 Syringe type acknowledge at start Yes: Syringe type must be acknowledged at first start of an infusion.
No: No syringe type acknowledgement necessary.
17 KVO enabling or disabling Yes: KVO function is enabled
(See also indices 20, 21 and 60)
No: KVO function is disabled
19 Buzzer tone at start of infusion Yes: Audible buzzer tone is emitted at start of infusion
No: No audible buzzer tone is emitted at start of an infusion
20 KVO setting for infusion rates up KVO rate can be set between 0.1 - 5.0ml/h
to 3ml/h (For this setting to be active, index 17 must be set to true.)
21 KVO setting for infusion rates KVO rate can be set between 0.1 - 5.0ml/h
higher than 3ml/h (For this setting to be active, index 17 must be set to true.)
22 One-time standby alarm Yes: Standby alarm is released only after the first start of an infusion (See
also index 80)
No: Standby alarm is released even before the first start of an infusion.
Standby alarm is released in STOP mode and in KVO mode.
23 Menu PRESSURE SETTING Yes: Menu PRESSURE SETTING is enabled on the pump. The user is able
to change the alarm limit temporarily.
(See also index 365 to set the default alarm limit and index 366 to select
other pressure units)
No: Menu PRESSURE SETTING is disabled and not visible on the pump.
24 Menu BATTERY INFO Yes: Menu BATTERY INFO is enabled on the pump. The remaining battery
capacity is displayed in percentage and time (format hh:mm), it is calcu-
lated on the basis of the current infusion rate.
No: Menu BATTERY INFO is disabled and not visible on the pump.
27 Menu KEY LOCK Yes: Menu KEY LOCK is enabled on the pump. The numeric keys of
the key pad can be locked / unlocked on the basis of a PIN code being
defined in index 50.
No: Menu KEY LOCK is disabled and not visible on the pump.
32 Menu BOLUS Yes: Menu BOLUS is enabled on the pump offering a manual and auto-
matic bolus function.
(See also index 38 for disabling the automatic bolus function.)
No: Menu BOLUS is disabled and not visible on the pump. No bolus
function is available.
33 Menu SETTINGS Yes: Menu SETTINGS is enabled on the pump. User can temporarily
change buzzer volume (see also index 363) and standby time (see also
index 80).
No: Menu SETTINGS is disabled and not visible on the pump

10 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software

Index Setting Description

38 Automatic bolus function Yes: Automatic and manual bolus functions are enabled
(For this setting to be active, index 32 must be set to true)
No: Only the manual bolus function is enabled.
44 Automatic pressure release after Yes: The pump stops infusing after the detection of an occlusion and
occlusion reduces the pressure by reversing the infusion flow. Once the pressure is
reduced, the pump stops.
No: The pump stops after detection of an occlusion without reducing
pressure.
47 Standby- and battery pre-alarm Yes: This index automatically reduces the sound volume to the minimum
low sound volume in case of a standby alarm or a battery pre-alarm.
No: The sound volume of the pre-alarms corresponds to the programmed
sound volume.
50 PIN code definition for the KEY Enter a number of maximum 4-digits
LOCK
60 KVO only after infusion com- Yes: KVO function is only activated when the volume to be infused (VTBI)
pleted is reached
(For this setting to be active, index 17 must be set to true.)
No: Standard KVO function is activated
65 Clear infused volume and con- Yes: The infused volume can be cleared (in STOP and in RUN mode).
tinue the infusion No: The clear and continue function is disabled.
80 Time period until standby alarm This index defines the time period after the last operating step until a
is released standby alarm is released (only in STOP mode)
363 Buzzer alarm sound volume The alarm volume is adjustable from 5 to 10 (5 = Minimum, 10 = Max-
imum)
365 Pressure default alarm level A pressure alarm is released as soon as the defined limit is reached. The
range reaches from 200mbar to 1200mbar
366 Pressure display unit The user can chose one of the given pressure units below:
1 = [mbar]
2 = [mmHg]
3 = [kPa]
4 = [cmH2O]
5 = [Psi]
(For this setting to be active, index 23 must be set to true)
370 Target volume nearly reached Yes: A pre-alarm announces that the target volume is nearly reached
pre-alarm (Time setting see index 371)
No: Setting is deactivated. No pre-alarm for “target volume nearly
reached” is released.
371 Time period for the release of the This index defines the time range of the pre-alarm ‘Target volume nearly
pre-alarm target volume nearly reached’. The release time is adjustable from 1 min to 15 min. (For this
reached. setting to be active, index 370 must be set to true)
522 Allow change of TOTAL VOLUME Yes: A total volume modification is allowed during a running infusion.
while infusing No: During a running infusion a total volume modification is not allowed.
1612 Medication selection required Yes: The user is forced to select a medication before starting an infusion.
No: The user is able to start an infusion without selecting a medication.
1636 Menu PURGE Yes: Menu PURGE is enabled.
No: Menu PURGE is disabled and not visible on the pump.

14.543_E_SM_en_A606S_4.8X 11
3. ARGUSservice Software *********
Type:
ARGUS Pump Configuration *********
ARGUS 606 S
Version: 4.80
Serial: 003 5 040

Infused sum : 0.0 ml


Operating time : 0h 10min
3.4. Factory Settings
Report PC time : 2013-02-07T14:07:52

The following chapters show the configuration reports with factory settings. Please note that most of the pumps are already cus-
======================= Configuration ==========================================
tomized when delivered to the customer. A report of your current pump configuration can be generated by pressing the “Report
Menues -------------------------------------------------------------------------
button” in the ARGUSservice tool when the pump is in the configuration 23 :mode.
Menu PRESSURE SETTING Yes No/Yes (No/Yes)
24 : Menu BATTERY INFO Yes No/Yes (No/Yes)
Each setting in your configuration report which varies from the factory setting will be
27 : Menu KEY LOCK
marked by a “!” on the
Yes
left side of the re-
No/Yes (No/Yes)
spective configuration index. 32 : Menu BOLUS Yes No/Yes (No/Yes)
1636 : Menu PURGE No No/Yes (No/Yes)
33 : Menu SETTINGS Yes No/Yes (No/Yes)
3.4.1. Software Versions 369 : Automatic Menu fall back delay tim 60 5 - 60 (sec)
80 : Time period until standby alarm is 120 60 - 3600 (sec)
The latest firmware version will be used for all new devices and replaces all previous software versions . Thus, the distinction
81 : Number of medication names display 6 0 - 255 ()

between firmware “Basic” and firmware “TIVA/Dose Calculation” is eliminated.


Display ------------------------------------------------------------------------
366 : Pressure display unit 1 1 - 5 (Enum)
524 : Display a department info text max. 16 char. (Text)
The TIVA/Dose calculation applications are available on all devices from firmware v4.70 and up, provided that the device is equipped
with pre-defined medications. The medications are specified by your healthcare professionals by means of the medication database
Handling -----------------------------------------------------------------------
365 : Pressure default alarm level 9 2 - 12 (100 mbar)
ARGUSmedDB. 2 : Key ON/OFF only in STOP mode usabl No No/Yes (No/Yes)
361 : Delay time for pressing the key ON 5 0 - 30 (* 0.1 s)
The database allows the creation of Hex data file for the download onto a device.
364 : Key [STOP] delay time 0 0 - 30 (* 0.1 s)
3 : Rate change allowed only in STOP m No No/Yes (No/Yes)
522 : Allow change of TOTAL VOLUME while No No/Yes (No/Yes)
The new factory settings are shown in the next chapter. 523 : Allow medication selection while i No No/Yes (No/Yes)
1612 : Medication selection required No No/Yes (No/Yes)
16 : Syringe type acknowledge at start No No/Yes (No/Yes)
3.4.2. ARGUS 606 S / Software 4.80 50 : PIN code definition for the KEY LO 0 0 - 9999 (code)
1616 : Default patient weight 0.0 0.0 - 300.0 (kg)
5 : Ask for new therapy after syringe Yes No/Yes (No/Yes)

********* ARGUS Pump Configuration ********* Functionality ------------------------------------------------------------------


*********
Type: ARGUS Pump
606 SConfiguration ********* 65 : Clear infused volume and continue No No/Yes (No/Yes)
Type:
Version: ARGUS
4.80 606 S 17 : KVO enabling or disabling Yes No/Yes (No/Yes)
Version:
Serial: 4.80
003 5 040 60 : KVO only after infusion completed No No/Yes (No/Yes)
Serial: 003 5 040 20 : KVO setting for infusion rates up 0.10 0.10 - 5.00 (ml/h)
Infused sum : 0.0 ml 21 : KVO setting for infusion rates hig 1.00 0.10 - 5.00 (ml/h)
Infused sum
Operating time : 0.0 ml
0h 10min 38 : Automatic bolus function Yes No/Yes (No/Yes)
Operating time : 0h 10min 39 : Bolus total volume to be reset aft No No/Yes (No/Yes)
Report PC time : 2013-02-07T14:07:52 367 : Syringe empty - pre-alarm 3 1 - 15 (min)
Report PC time : 2013-02-07T14:07:52 370 : Target volume nearly reached - pre Yes No/Yes (No/Yes)
371 : Time period for the release of the 3 1 - 15 (min)
======================= Configuration ==========================================
======================= Configuration ========================================== Technical ----------------------------------------------------------------------
Menues ------------------------------------------------------------------------- 19 : Buzzer tone at start of infusion Yes No/Yes (No/Yes)
Menues
23 : -------------------------------------------------------------------------
Menu PRESSURE SETTING Yes No/Yes (No/Yes) 363 : Buzzer alarm sound volume 10 5 - 10 (volume)
23
24 : Menu PRESSURE SETTING
BATTERY INFO Yes No/Yes (No/Yes) 22 : One-time standby alarm No No/Yes (No/Yes)
24
27 : Menu BATTERY
KEY LOCKINFO Yes No/Yes (No/Yes) 47 : Standby- and battery pre-alarm low No No/Yes (No/Yes)
27
32 : Menu KEY
BOLUSLOCK Yes No/Yes (No/Yes) 6 : Nurse call settings No No/Yes (No/Yes)
32 : Menu BOLUS
1636 : Menu PURGE Yes
No No/Yes
No/Yes (No/Yes)
(No/Yes) 1614 : Energy-saving mode 75 0 - 100 (%)
1636
33 ::Menu
MenuSETTINGS
PURGE No
Yes No/Yes (No/Yes)
No/Yes (No/Yes) 1615 : Delay time for energy-saving mode 10 0 - 120 (sec)
33 : Automatic
369 Menu SETTINGS
Menu fall back delay tim Yes
60 No/Yes
5 (No/Yes)
- 60 (sec) 44 : Automatic pressure release after o Yes No/Yes (No/Yes)
369 Automatic
80 : Time periodMenu fall
until back delay
standby alarmtim
is 60
120 5 3600
60 - - 60 (sec) 368 : Adjustment of the displayed batter 100 25 - 100 (%percent)
80
81 : Time period
Number until standby
of medication namesalarm is
display 120
6 600--3600
255 (sec)
() 393 : Service interval in months 0 0 - 24 (months)
81 : Number of medication names display 6 0 - 255 () 394 : Service interval in hours of opera 0 0 - 10000 (hours)
Display ------------------------------------------------------------------------
Display
366 : ------------------------------------------------------------------------
Pressure display unit 1 1 - 5 (Enum) Communication ------------------------------------------------------------------
366
524 : Pressure
Display adisplay unitinfo text
department 1 max. 16 1 - 5 (Enum)
char. (Text) 518 : Permissions for communication 2 0 - 3 (Enum)
524 : Display a department info text max. 16 char. (Text) 1617 : Interface/Ethernet board 0 0 - 2 (Enum)
Handling -----------------------------------------------------------------------
Handling -----------------------------------------------------------------------
365 : Pressure default alarm level 9 2 - 12 (100 mbar) ======================= Calibration ============================================
365
2 : Pressure
Key ON/OFFdefault
only inalarm
STOPlevel
mode usabl 9
No 2 - 12 (100
No/Yes mbar)
(No/Yes)
2 : Key
361 ON/OFF
Delay time only in STOP mode
for pressing usabl
the key ON No
5 No/Yes
0 - 30 (No/Yes)
(* 0.1 s) Syringe ------------------------------------------------------------------------
361
364 : Delay time for
Key [STOP] delaypressing
time the key ON 5
0 0 - 30 (* 0.1 s) ! 508 : Drive calibrated with part 1 Yes No/Yes (No/Yes)
364
3 : Key
Rate[STOP]
changedelay time
allowed only in STOP m 0
No 0 - 30 (*
No/Yes 0.1 s)
(No/Yes) ! 509 : Drive calibrated with part 2 Yes No/Yes (No/Yes)
3 : Allow
522 Rate change
changeallowed
of TOTALonly in STOP
VOLUME whilem No No/Yes (No/Yes)
522 : Allow medication
523 change of TOTAL VOLUME
selection while
while i No No/Yes (No/Yes) Pressure -----------------------------------------------------------------------
523 ::Allow
1612 medication
Medication selection
selection while i
required No
No No/Yes
No/Yes (No/Yes)
(No/Yes) ! 502 : Pressure calibrated flag Yes No/Yes (No/Yes)
1612
16 ::Syringe
Medication
typeselection
acknowledgerequired
at start No
No No/Yes (No/Yes)
No/Yes (No/Yes) ! 503 : Pressure 0bar calibrated Yes No/Yes (No/Yes)
16 : PIN
50 Syringe
codetype acknowledge
definition at KEY
for the start
LO No
0 No/Yes
0 - (No/Yes)
9999 (code)
50 ::PIN
1616 code patient
Default definition for the KEY LO
weight 0
0.0 0 - 300.0
0.0 9999 (code)
(kg) Display ------------------------------------------------------------------------
1616
5 ::Ask
Default patient
for new weight
therapy after syringe 0.0
Yes 0.0 No/Yes
- 300.0 (No/Yes)
(kg) 362 : Display backlight brightness 240 0 - 255 (steps)
5 : Ask for new therapy after syringe Yes No/Yes (No/Yes) 1613 : Display contrast 7 1 - 15 (level)
Functionality ------------------------------------------------------------------
Functionality ------------------------------------------------------------------
65 : Clear infused volume and continue No No/Yes (No/Yes) ==================================================================================
65
17 : Clear infusedor
KVO enabling volume and continue
disabling No
Yes No/Yes (No/Yes)
17 : KVO only
60 enabling
afterorinfusion
disablingcompleted Yes
No No/Yes (No/Yes)
60 : KVO setting
20 only after
forinfusion
infusioncompleted
rates up No
0.10 No/Yes
0.10 - (No/Yes)
5.00 (ml/h)
20 : KVO setting for infusion rates hig
21 up 0.10
1.00 0.10 - 5.00 (ml/h)
21 : Automatic
38 KVO setting for function
bolus infusion rates hig 1.00
Yes 0.10 No/Yes
- 5.00 (No/Yes)
(ml/h)
38
39 : Automatic
Bolus totalbolus function
volume to be reset aft Yes
No No/Yes (No/Yes)
39 : Syringe
367 Bolus total
emptyvolume to be reset aft
- pre-alarm No
3 No/Yes
1 (No/Yes)
- 15 (min)
367 : Target
370 Syringevolume
empty nearly
- pre-alarm
reached - pre 3
Yes 1 - 15 (No/Yes)
No/Yes (min)
370 : Time
371 Target volume
period nearly
for reachedof
the release - the
pre Yes
3 No/Yes
1 (No/Yes)
- 15 (min)
371 : Time period for the release of the 3 1 - 15 (min)
Technical ----------------------------------------------------------------------
Technical ----------------------------------------------------------------------
19 : Buzzer tone at start of infusion Yes No/Yes (No/Yes)
19 : Buzzer tone
363 alarmat start
sound of infusion
volume Yes
10 No/Yes
5 - 10 (No/Yes)
(volume)
363
22 : Buzzer alarm sound
One-time standby alarmvolume 10
No 5 - 10 (volume)
No/Yes (No/Yes)
22
47 : One-time
Standby- standby alarm
and battery pre-alarm low No No/Yes (No/Yes)
47
6 : Standby-
Nurse calland battery pre-alarm low
settings No No/Yes (No/Yes)
6 ::Nurse
1614 call settings
Energy-saving mode No
75 No/Yes
0 - 100 (No/Yes)
(%)
1614
1615 : Energy-saving
Delay time formode
energy-saving mode 75
10 0 - 100
120 (%)
(sec)
1615
44 ::Automatic
Delay time for energy-saving
pressure mode
release after o 10
Yes 0 - 120 (No/Yes)
No/Yes (sec)
44 : Adjustment
368 Automatic pressure release after
of the displayed o
batter Yes
100 25No/Yes (No/Yes)
- 100 (%percent)
368 : Service
393 Adjustment of thein
interval displayed
months batter 100
0 250--100 (%percent)
24 (months)
393
394 : Service interval in months
hours of opera 0 0 -010000
- 24 (months)
(hours)
394 : Service interval in hours of opera 0 0 - 10000 (hours)
Communication ------------------------------------------------------------------
12 Communication ------------------------------------------------------------------
518 : Permissions for communication 2 0 - 3 (Enum) 14.543_E_SM_en_A606S_4.8X
518
1617::Permissions for communication
Interface/Ethernet board 20 00--32 (Enum)
(Enum)
1617 : Interface/Ethernet board 0 0 - 2 (Enum)
======================= Calibration ============================================
3. ARGUSservice Software

3.5 Configuration of User-Syringes

Approved syringes:
Syringes from various manufacturers are approved to be used with the ARGUS 606 S (see chapter 3.3, point 14.) A list of validated and
approved syringes can be found on our website www.codanargus.com, (login section, manuals for ARGUS 606 S).

User defined syringes:


If any other brand shall be used, the following characteristics must be given:
– The syringe must be CE marked
– It must be pressure resistant and safe for the use with syringe pumps
– It must be three-part: barrel, plunger, sealing
– The connections must be a Luer-Lock connection (the same applies to extension lines).

Warning!
Only syringes fulfilling the above requirements may be programmed as „USER“ syringe:

1. Enter the Syringe Selection Menu

14.543_E_SM_en_A606S_4.8X 13
3. ARGUSservice Software

2. Mark the size of the syringe to be defined

Click the PLUS button to open the syringe setup.

3. Set the name of the user syringe and enter the measures of
the barrel diameter, the plunger length, the clamp diameter and
the syringe length.

Note:
User syringes 20/30 ml:
To record a 30 ml syringe, the syringe length must be measured
on the graduated scale from 0 – 20 ml.

User syringes 50/60 ml


To record a 60 ml syringe, the syringe length must be measured
on the graduated scale from 0 – 50 ml.

Warning!
After the configuration of a user syringe, a function check and a
control measurement have to be performed!

We recommend using syringes approved by CODAN ARGUS AG


as user defined syringes may not comply with the stated flow
rate accuracy due to mechanical tolerances!

14 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software

3.6 TOOLBOX

Several applications are available for maintenance:

Click on the box “Start Configurator” and enter by clicking on


“Toolbox”

The ARGUSservice Toolbox offers various additional features to


adapt your pump to specific needs.

Use this function to synchronize the pump internal clock with


the clock of your PC.
Note: The factory setting bases on Central European Time
(GMT + 1.00 h). All history logs are recorded accordingly. There
is no automatic switch between summer and winter time.

A PIN code prevents unauthorized persons to change config-


urations by means of the ARGUSservice tool. The default PIN
code of the factory setting is “0”.

14.543_E_SM_en_A606S_4.8X 15
3. ARGUSservice Software

The last 1’000 events are registered in real time.


For details see chapter 3.7 “History Log”.

Use this feature to set an alarm reminding that the safety check
is due. The released alarm is displayed on the pump at power
up showing the message “Safety Check is Due!” accompanied
by an acoustic signal.
Before activating the “Service Interval”, the configuration
indices #393 (interval in months) and #394 (interval in hours)
have to be filled in accordingly.

Load an ARGUS language file in order to change the display


language. English is included in the pump’s firmware; any other
language must be downloaded.
To activate the English language again, the downloaded lan-
guage has to be erased (see menu “Extras” below).
Only one language can be activated on the pump.

This feature supports fast and easy configuration procedures


including the released syringes. Use your specific configuration
files previously saved on your PC or the configuration settings
of the connected pump. The configuration settings have to be
in accordance with the software version of your pump; com-
patibility is given when the two first digits, e.g. 4.40 and 4.43,
are identical.
Note: Settings of pump clock, safety check reminder and calib-
ration data cannot be replicated from one pump to another.

The medication library must have the format .xml in order to be


downloaded by mean of the ARGUSservice tool.
The file is created by the responsible health care professional.

16 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software

Enter the menu Extra:

a) The function “Reprogramming serial number” enables you


to re-program a serial number which may in rare cases go lost
during an update procedure or after the replacement of a main
board.
b) The function “Erase configuration areas” can be used after
having received a permission Key from CODAN ARGUS.
c) The function “Erase language” will revert the display inform-
ation wording into the English base version.
d) The function “Erase medications…” will delete the complete
medication database.

Note:
Reprogramming a serial number is possible without permission key in case the serial number shows 0 after a mainboard change or
after a software update.
In case that a serial number is still available on the board, a permission key must be requested from Codan Argus.

14.543_E_SM_en_A606S_4.8X 17
3. ARGUSservice Software

3.7 History Log

18 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software

3.8 History Example

The pump registers the following events:


Logs are recorded in real time.

Use the button “Load remaining…” to download all registered


data.

Use the “Report” button to save the log as pdf file.

14.543_E_SM_en_A606S_4.8X 19
4. Serial Communication

4.1. General

The ARGUS 606 S pump has one serial interface connector per default being connected to ARGUS Docking Stations.

An additional optional interface board is available for a RS 232 interface on the RJ45 connector including a separation device ac-
cording to EN60601-1-1:2000.
Please refer to chapter 4.3 for installation directions.

Other optional interface types (Ethernet, USB, WLAN) will be provided on request.

Warning!
The serial interface designed by CODAN ARGUS enables the communication with ARGUS 100 M and ARGUS 60 M docking sta-
tions. In such a case the separation device according to EN60601-1-1:2000 is given by the docking station itself!
It is prohibited to connect the interface cable (REF. 600951/10.093) to the docking interface of the pump while the pump is connec-
ted to a patient!

An active communication between the pump and the PC is not apparent on the pump!

4.2. Serial Communication Protocol

The following characteristics are valid for all ARGUS devices (volumetric pumps, syringe pumps, docking stations with V4.xx and
PCs) which are intended to communicate with the device mentioned in this service manual.
• Full-duplex RS232, currently 4800Baud for single pumps, 9600 Baud for docking stations (also on master/slave-link).
• Simple master (host/PC) – slave (device) communication (host does polling).
• The host has to repeat the request if there is no valid response.
• Uses a checksum (CRC-8).
• Binary data transmission, thus no ASCII/text parsing.
• Fast & direct communication with pumps on ARGUS docking station.
• Specified timeouts during remote mode.
• Basic framing technique used as in the Serial Infrared Link Access Protocol (IrLAP) version 1.1.
Please contact your local distributor or CODAN ARGUS AG for the complete serial communication protocol description.

4.3. Installation of an optional interface board

An optional interface board can be installed on the main board of the ARGUS 606 S.
Remove any covers on the main board connectors (if any).
Please plug-in the interface vertically and do not bend it. Carefully close the device and ensure a proper positioning of the interface
board in the mechanical guide included in the device cover.
To activate the interface board it is necessary to adjust the respective indices in the pump configuration (tree “Communication”).

We recommend performing a safety standard check after the installation of an optional interface board.

20 14.543_E_SM_en_A606S_4.8X
4. Serial Communication

4.4. Pump connections

The pump connections have the following functions:

The schematic to the right shows the nurse call connection.

14.543_E_SM_en_A606S_4.8X 21
5. Software Updates

Requirements for a software update


To update a CODAN ARGUS device, the following items are needed:
• PC with Microsoft® Windows™ 2000 or newer, .NET Framework must be installed!
• RS-232 serial interface cable (REF. 600951/10.093)
• Configuration tool “ARGUSservice”
• Latest firmware included in a text file named “A606_xxx.txt” (“xxx” is the placeholder for the firmware version) or embedded
firmware to be found in the ARGUSservice software.

Those items are available from your local distributor or from CODAN ARGUS AG.

Warning!
After a software update, a safety standard check has to be performed!
An active communication between the pump and the PC is not apparent on the pump.

5.1. Firmware Update Procedure

Prearrangements

Make sure that the pump is plugged to mains

Make sure that the pump is connected to your PC by means of the ARGUS interface cable
(REF 600951/10.093)

Select the correct serial COM port

Set the pump into the configuration mode

22 14.543_E_SM_en_A606S_4.8X
5. Software Updates

Start the ARGUSservice and select “Pumps ARGUS….”

Check the selected syringe and the syringe limits and note
them down manually.

Save the current configuration before updating the firmware.

Warning!
It is strongly recommended to save the actual configuration in
order to maintain the serial number and the calibration para-
meters. In addition it is advisable to create a report in order to
crosscheck the configuration after the update.

Use the button “Save” for this procedure.

Quit the program and close ARGUSservice.

14.543_E_SM_en_A606S_4.8X 23
5. Software Updates

Restart the ARGUSservice

Use the button “Update-Center” for the required product.

This window will be implemented with Argus Service Version


4.71.11145

If all measures have been taken press yes; if not, press no and
make the required saves.

24 14.543_E_SM_en_A606S_4.8X
5. Software Updates

Make sure that “OK” is displayed next to the “Ping” button


which confirms the SELECTION of the correct serial COM port.

The pump will automatically display the following information


on the LCD:
ARGUS 606 S
BOOTLOAD MODE
SW-REL.: X.XX
Warning!
Please note that the Software Release X.XX displayed here
above corresponds with the boot loader version, not with the
installed firmware!

Click on “Load HEX firmware”

14.543_E_SM_en_A606S_4.8X 25
5. Software Updates

Set the flag o “Embedded version”

Warning!
Only the latest version of the ARGUSservice will contain the
latest pump firmware embedded. Compare with the version
under www.codanargus.com in the download section (the
necessary password can be ordered under codan@codanargus.
com).

Select the correct embedded versions and click the “Flash”


button,

Wait until the message “Flash completed successfully” shows


up and confirm with “OK”.

Select “OK”

Should the pump start in regular mode, turn it off and start it
again in configuration mode by

pressing the 3rd numeric key while pressing the ON/OFF Key.

Once the pump starts in configuration mode, it is ready for the


reimport.

26 14.543_E_SM_en_A606S_4.8X
5. Software Updates

5.2. Re-import the saved configuration file

Start the Argus Service - Configuration

Edit – Pin 0

Use the button “Import”:

Open the previously saved configuration file:

14.543_E_SM_en_A606S_4.8X 27
5. Software Updates

Ò Activate the flags on „ Full configuration / Calibration area / Serial Number“, only if requested by the program

Ò Don’t activate the flag “syringe”

and press OK for download.

If the download procedure failed, the pump will prompt the “CrtL” alarm (“Safety Check is due”):
a) check the serial number and the calibration flags on indices 502/503 508/509
b) execute a safety standard check incl. syringe and pressure calibration (see also 6.4 and 6.5)

28 14.543_E_SM_en_A606S_4.8X
5. Software Updates

5.3. Display Language

If you need another display language than English it’s necessary to load the respective language file into the pump. The latest lan-
guage files are available under www.codanargus.com (Login section, Software, ARGUS 606 S).

Start the ARGUSservice:

Open “Toolbox”

OpenClick on “Language”

Select the language file (type *. HEX). Your language file can be downloaded from codanargus.com\Login

After the download, the pump returns automatically into the normal operating mode.

Warning!
Respect the safety aspects as described in the following chapter.

14.543_E_SM_en_A606S_4.8X 29
5. Software Updates

5.4. Safety Aspects

Warning!
For medical device traceability your local distributor or CODAN ARGUS AG needs to be informed about every device update (serial
number) you performed!

Warning!
Do not make any software updates as long as the device is in use and/or connected to a patient!

Warning!
An active communication between the pump and the PC is not apparent on the pump!

A standard safety check has to be performed after every software update!


Please verify the software version installed (step 1 of the standard safety check) by entering the configuration mode
(chapter 2.2).

5.5. Upgrade Procedure from a BASIC to a TIVA Pump

General information about the new version

The new software version can be used as basic device.


To use it with dose calculation/TIVA a medication library must be downloaded.

All previous Basic or Tiva versions can be upgraded to the new version

Please take note of the following remarks before you start the upgrade:
– There is no additional hardware modification required,
– The upgrade is subject to an additional charge.
– The upgrade procedure requires direct contact with CODAN ARGUS AG, Switzerland to se-
cure traceability and correct labeling. You don’t need any upgrade key.
– Unauthorized upgrades may lead to discontinued function of the previously in-
stalled standard firmware and to the termination of any warranty claims!
– A safety standard check has to be performed on each upgraded pump!

30 14.543_E_SM_en_A606S_4.8X
6. Maintenance

6.1. General

Warning!
Only authorized persons who have been trained by CODAN ARGUS AG or by the local distributor are allowed to service the ARGUS
606 S Syringe pump. In case of a request for repair, send the device together with the filled out “repair order form” (see chapter
11.) to the local distributor. Further information is available from:

CODAN ARGUS AG
CH-6340 Baar / Switzerland
E-mail: codan@codanargus.com

Warning!
The Safety Standard Check (SSC) has to be performed at least every 24 month or after
10’000 hrs of operation. The check has to be done in accordance with chapters 6.2 – 6.9.
No special maintenance of the ARGUS 606 S syringe pump is necessary. There are no
wear and tear parts.

Warning!
It is recommended to define a PIN code to prevent unauthorized users from entering the configuration mode.

Enter the configuration mode (refer to chapter 2.2). Press the MENU Key several times to enter the described menus here after

6.2. Keypad and Display Test

This menu is designed to assist technical staff during the safety


standard check. The aim is to test the correct functioning of the
user interface including Key pad and display.

Display Test
This test enables the detection of pixel errors. The control of the
lines must be done twice. The correct functioning of the display
is confirmed by a black display at the end of this test.

1. Press the MENU Key several times until „KEYPAD & DISPLAY
TEST“ is displayed.
2. Use the SELECTION Key to choose „DISPLAY TEST“
3. Press the START Key 2x

4. The vertical lines must show a regular pattern; press the


START Key
5. The vertical lines must show a regular pattern; press the
START Key
6. All pixels are black; press the START Key

14.543_E_SM_en_A606S_4.8X 31
6. Maintenance

Keypad Test
This test checks the correct functioning of all Keys ex-
cept the ON/OFF Key. The ON/OFF Key is automatically
checked at any switching on and off of the pump. The test
was successful after every Key has been pressed. To fin-
ish the test press the ON/OFF Key to turn the pump off.

1. Use the SELECTION Key to choose „KEYPAD TEST“

2. Press the START Key

3. Press each Key on the pump. The proper functioning is con-


firmed by .

Led und Backlight Test


This test enables to check the proper functioning of
– mains LED
– indicator lamp LED (green, amber and red)
– brightness of the backlight
Please notice the display messages accompanying each test
step.

1. Use the SELECTION Key to select “LED & BACKLIGHT”

Follow all the steps indicated on the display screen.

32 14.543_E_SM_en_A606S_4.8X
6. Maintenance

6.3. Factory Base Syringe Calibration

The factory base calibration is an integral part of the manufacturing process done by CODAN ARGUS AG. This process must be
repeated after the replacement of an assembly, e.g. main board, complete the syringe drive or parts of it, syringe barrel holder,
potentiometer, etc., as well as after a downfall or a heavy impact.

Please remove any syringe before you start the calibration process.

When spare parts shall be replaced, please go through chapter 7 first.


Open the device but leave all cables connected, then enter the configuration mode (refer to chapter 2.2). Press the MENU Key to
enter the calibration menu.

  1. Use the SELECTION Key to select „CALIB. PART 1“

  2. Press the START Key to get access to the different signal voltages.

Do not press this key again as this will only be done during the final calibration with a closed device!

  3. Make sure that the red syringe barrel holder stands in completely closed position; adjust the white potentiometer to 360mV ±
40mV (BARREL DIAM.); tighten the set screw on the white cog wheel and verify the declared value again.

  4. Pull out the syringe barrel holder completely; the signal of the barrel diameter must be within 2600mV and 3180mV. If this
range is exceeded, you have to repeat step 3.

  5. Push the syringe drive towards the left. Adjust the black potentiometer on the drive unit to 300mV ± 10mV (PLUNGER POS.);
tighten the set screw and verify the declared value again.

  6. Push the syringe drive towards the right; the signal of the plunger position must read between 2900mV and 3000mV. If this
range is exceeded, step 5 has to be repeated.

  7. Remove the red cover of the driving head according chapter 7.


Please verify that the piston clamps are fully closed and adjust the signal of the potentiometer to
460mV ± 50mV (CLAMP DIAM.); tighten the set screw and verify the declared value again.

  8. Assemble the red driving unit again.

  9. Make sure that the piston clamps are fully closed and that the clamp lever does not abut against the table; the signal of the
CLAMP DIAM. must read 460mV ± 50mV. If this value is not reached, steps 7 to 8 have to be repeated.

10. Fully open the piston clamps; the signal of the CLAMP DIAM. must read between 2400mV and 3000mV. If this range is not
reached, steps 7 to 9 have to be repeated.

11. Press the EXIT Key to abandon the calibration menu. Do not press the START key!

12. Close the device and go through the final calibration (chapter 6.5 – 6.6)

The disregard of the above given range of values may provoke a technical alarm when the pump will be started up.

14.543_E_SM_en_A606S_4.8X 33
6. Maintenance

6.4. Syringe Calibration

The DRIVE CALIBRATION menu has two submenus, one for cal-
ibration part 1 and one for part 2.
The submenus allow the calibration of the barrel, the clamp dia-
meter and plunger position.

Calib. Part 1

1. Use the SELECTION Key to select CALIB. PART 1


2. Press the START Key

3. Insert the long calibration part ( 17, l=120mm, REF.


600148/10.153) and make sure that the piston clamps cap-
ture firmly without leaving a gap between drive unit and cal-
ibration part!

4. Slightly press the calibration part against the drive unit.

5. Check whether the measurement readings are within the


specified range. If not, please repeat the factory base calibra-
tion (chapter 6.3.)

6. Press the START Key twice.

34 14.543_E_SM_en_A606S_4.8X
6. Maintenance

Calib. Part 2

1. Use the SELECTION Key to select „CALIB. PART 2“

2. Press the START Key

3.
Insert the short calibration part ( 31, l=20mm,
REF.600584/11.194)

4. Check whether the measurement readings are within the


specified range. If not, please repeat the factory base calibra-
tion (chapter 6.4).

5. Press the START Key twice.

14.543_E_SM_en_A606S_4.8X 35
6. Maintenance

6.5. Pressure Calibration

The PRESSURE CALABRATION menu has two sub menus:


The first makes for the offset calibration (0 bar absolute pressure calibration by adjusting the digital potentiometer). The second one
makes for the calibration of the pressure sensitivity using two landmarks at 0.2 and 1.2 bars to interpolate the pressure measure-
ment.

Before starting this procedure, make sure that the pump casing is closed and screwed together to ensure a stable force distribution.

Calib. Absol. 0.0bar

1. Use the SELECTION Key to select „CALIB. ABSOL. 0.0ba“

2. Syringes or calibration parts must not be inserted at this point; the driving head has to be positioned midway.

3. Press the START Key.

4. The pump runs backward to reach a stable offset value. A target value of 600mV is pre-programmed. As soon as the offset value
has stabilized, press the second numeric Key several times until approximately 600mV are reached and remain stable.

5. Press the START Key twice. Caution: The pump does not stop automatically at the mechanical end stop!

Calib. Sensitivity

1. Use the SELECTION Key to select „CALIB. SENSITIVITY“.

2. Insert the tension spring balance1.

3. Press the START Key.

4. The delivery speed can be gradually increased or decreased


pressing the 4th and the 1st numeric Key respectively.

5. Reaching 0.2bar, press the START Key.

6. Reaching 1.2bar, press the START Key.

7. Press the START Key again.

36 14.543_E_SM_en_A606S_4.8X
6. Maintenance

Control Measurement

1. Switch the pump off and on again.

2. Control the pressure survey by means of the spring balance1 :


– Adjustment of the occlusion alarm limit to 900 mbar
– Initiation of the pressure build-up with 200 ml/h
– Release of an occlusion alarm at 0.9 bar ± 200 mbar and release pressure by reverse delivery

1
Use a manometer and a 50 ml syringe for the calibration and the control measurement if there is no spring balance available.

14.543_E_SM_en_A606S_4.8X 37
6. Maintenance

6.6. Pump Accuracy Measurement

Insert a 50 ml syringe (e.g. Codan Perfusion) filled with distilled


water. Purge the extension line and start to administer the
water into a cup placed on a balance which has been zeroed
before. Please verify that the inserted syringe type corresponds
with the syringe displayed in the pump screen.

The pump settings must be:


Rate: 200 ml/h
Volume: 20 ml

The resulting net weight shall read 20 g ±2% (refers to a rate


range from 1 ml/h – 1’100 ml/h).

6.7. Correction of the Battery Capacity

The capacity of the battery will decrease over time like all
batteries do. Therefore the capacity status (percentage and
remaining time) displayed in the menu “BATTERY INFO” may
differ from the real battery capacity.
The configuration menu in the ARGUSservice tool allows ad-
justing the battery capacity to real capacity values. The default
value of a new battery is 100% (see index 368). The battery
capacity can be reduced down to a minimum of 50 %.

Procedure to evaluate the correction factor:


1. Discharge the fully charged battery @ 5ml/h
2. Read out the time being used to discharge the bat-
tery (use the recall function in the INFO Menu
of the pump or read the syringe scale or look
for the time stamp in the history log)
3. Battery lifetime correction factor = infu-
sion time / specified runtime * 100
4. Select index 368 in the ARGUSservice tool
and enter the calculated factor

Warning!
The battery charging system included in the ARGUS 606 S re-
quires additional electronics (temperature and current sensors)
to provide a high current to accelerate charging time. Therefore
the battery pack must be obtained from CODAN ARGUS AG.
The charging time may depend on the current pump status
(infusing rate, stand-by).

38 14.543_E_SM_en_A606S_4.8X
6. Maintenance

6.8. Fault Codes

A technical failure will be indicated on the pump by a continuous acoustic alarm and/or a continuous red indicator lamp. During a
failure state, the fault code on the LC display allows the analysis of the cause:
Example of a technical failure

If the pump was switched OFF after a detected failure, the fault code will be stored in the history log and also in the configuration
file of the pump (index 380 - 385).

Fault
Code Possible Failure cause
00 Reset test failed
901 Check all cable connections
10xx Software error or memory error
11xx Software error
12xx Software error
13xx Software error
14xx Hardware error
1404 Check all cable connections
1405 Unconnected stepper motor
1406 Check all cable connections
1413 Unconnected stepper motor
1414 Short circuit in stepper motor or in its connection
1415 Change the mainboard
1419 Battery voltage out of range
1421 Input voltage out of range
1425 Voltage reference on main board out of range
1426 Unconnected DMS-Board or
voltage reference on DMS board out of range
1428 Internal temperature sensor defective or
internal device temperature too high
1429 Battery temperature sensor defective or
battery temperature too high
1434 Bad calibration of size potentiometer?
1435 Bad calibration of size potentiometer?
1436 Bad calibration of clamp potentiometer?
1437 Bad calibration of clamp potentiometer?
Unconnected flex cable to drive movement control
1438 Bad calibration of pressure potentiometer?
1439 Bad calibration of pressure potentiometer?
1440 Unconnected display cable on display board
15xx Software error or internal bus error
16xx Pressure out of range
1601 Pressure signal out of range, DMS defective
17xx Problems with the drive

14.543_E_SM_en_A606S_4.8X 39
6. Maintenance

Fault
Code Possible Failure cause
1704 Syringe drive not correct installed ( Check the holder )
1710 Plunger calibration out of range
1711 Plunger signal out of range
18xx Problems with the Keypad
1800 Key pressed too long or Key blocked
1804 Check display board connection, then all other cables
1832 Check all cable connections
19xx Problems with the battery
1902 Unconnected battery
1904 Unconnected battery
1908 SMD Fuse 0.5AF defective
20xx Software error
21xx Software error
22xx Syringe dimension invalid
2201 Barrel calibration diameter out of range
2202 Clamp calibration diameter out of range
23xx Software error
24xx Communication error
25xx Software error
26xx Memory error (hardware failure)
27xx Software error
28xx Memory error (hardware failure)
29xx Internal bus error (hardware failure)
30xx Internal bus error (hardware failure)

6.9. Message “Safety Check is due”

After the time of use has elapsed and a safety standard check is due, the reminder alarm “Safety Check is Due” will be displayed
after power up of the pump; See chapter 3.3 (“Statistics” indices 390-392).

The same message lights up as well under the following circumstances:


– Calibration is faulty (pressure or mechanic); see indices 502, 503, 508 and 509 in the configuration menu “Calibration”
– Invalid serial number verify index 396 in the configuration menu “ARGUS Presets”
– Empty coin cell on the main board
– Check Date Time

40 14.543_E_SM_en_A606S_4.8X
7. Replacement of Parts

7.1. General

Warning!
The ARGUS 606 S may only be used with accessories which are approved by CODAN ARGUS AG. Please refer to the CODAN AR-
GUS “Spare Part Catalogue” containing the respective reference numbers!

Repair shall only be performed by trained specialists being in possession of a manufacturer’s authorization!

Repair shall only be done with origin spare parts which shall not be modified without prior manufacturer’s authorization!

Disconnect the power cord and all interface connections prior to disassembly!

The internal mains connectors under the transformer aren’t insulated!

A complete recalibration is mandatory after:


1. The configuration of new syringe
2. The replacement of a pressure sensor
3. The replacement of a complete syringe drive
4. The replacement of a side wall or the housing
5. The replacement of the main board

Warning!
If the complete syringe drive or parts of it, the side wall with the pressure sensor, the pump housing or the main board was re-
placed, a full calibration according to chapters 6.4 – 6.6 is required. Also a factory base calibration might be required.

Remove the battery connector (X15) prior to replacing any parts!

Disassembly of the housing:


Remove pole clamp at the rear side. Remove the screws on the casing base (5 • M4) and the screw at the gray syringe guide
(1•M3).
Place the casing cover behind the casing base.

How to remove the main board (REF.601254 or 601255):


Remove the battery connector first and then all other cables of the main board.
Remove the drive completely prior to the replacement of the main board.
After the replacement or after the re-insertion, a basic calibration according to chapter 6.4 and a final calibration according to
chapters 6.5 - 6.6 are in any case required.

How to remove the syringe drive (REF. 601069/10.182):


Move the drive unit fully to the right and remove the black fixing plate down at side wall of the drive. Remove the plugs X3, X10
and X14 (motor cable, flex cable and cable to the sidewall). Remove the syringe drive out from the housing by fully pressed clamp &
clutch levers.

How to remove the red cover (REF. 600936/10.151) of the driving head (REF.600937/10.152):
Remove the 3 screws on the cover.
Important: To disassemble the unit, open the beaks by hand one third (or put a coin between the beaks) then pull the cover with
the levers out of the housing.

Replacement of the battery (REF. 601074):


Turn the pump off and disconnect the power cord. Open the battery cover (REF. 600701/10.183) and replace the battery. Check all
cable connections carefully and close the battery cover.

14.543_E_SM_en_A606S_4.8X 41
7. Replacement of Parts

Warning!
The shrink-wrapped PCB (Printed Circuit Board) includes a temperature sensor which has to be positioned in direction of the con-
nector (avoid a positioning towards the transformer)!

Charging procedure:
1. Discharge the battery @ 5ml/h until the pump turns automatically off. This first step is ne-
cessary in order to synchronize the battery and the firmware!
2. Recharge the battery
3. Execute battery test:
Discharge the fully charged battery @ 5ml/h
Read out the time being used to discharge the battery (chapter 6.6)
Compare it with the reference value of a new battery.
Repeat the charge/discharge cycle, if necessary, until the specified runtime is reached.

Replacements of the complete drive (REF. 601069/10.182):


After each disassembly or replacement of the syringe drive, the basic calibration and the final calibration (refer to chapters 6.4 –
6.6) must be executed to guarantee a perfect pump operation and pressure monitoring.

Replacement of the main board (REF. 601271)


Remove the main board according the description above. Insert the new main board; ensure the cog wheel on the potentiometer is
unlocked!
A basic calibration according chapter (6.4) and a final calibration according chapters (6.5 – 6.6) is required. During the basic calibra-
tion, the cog wheel will be locked again.

Replacement of the casing cover including LC display, keypad and display board
(REF. 601060/10.185)
Perform the Keypad and Display Test according to chapter 6.2.
Check the display contrast setting in the maintenance mode of the pump after replacement. The display contrast may vary.

Replacement of the casing base (REF. 601061/10.186, 230V or REF.601147, 115V))


Remove the main board and syringe drive according the description above. Remove the transformer (601062/10.187 or
601103/10.188), holder with edge board (601064/10.189) and battery cover (600701/10.183) from the old casing base.
After the re-insertion a basic calibration according to chapter 6.4 and a final calibration according to chapters 6.5 – 6.6 is required.

Replacement of the syringe clamp (REF. 600750/11.168)


Remove the main board and loose the cog wheel on the white potentiometer.
Change the syringe barrel holder by loosening the snap ring at the end of the gear.
After the re-insertion a basic calibration of the barrel diameter according to chapter 6.4 and a complete final calibration according
to chapters 6.5 – 6.6 is required. In the basic calibration, the cog wheel will be locked again.

For repair purpose only: Red housing of the drive unit


Remove the soldered connection to the drive unit on the pump’s inside. Open the red cover of the drive unit according the descrip-
tion above. Replace the housing drive unit and establish the solder connection again.
After the replacement, a basic calibration according chapter 6.4 and a final calibration according chapters 6.5 – 6.6 is required.

Replacement of motor and gear (REF. 601066/10.180))


Remove the syringe drive and replace the motor and gear.
After the re-insertion, a final calibration according to chapters 6.5 – 6.6 is required.

Replacement of coin cell on the main board


Remove the existing coin cell by pushing out the cell with an insulated object.
Be aware of local short circuits to the electronics.
Any standard BR1632 coin cell (3V, 120mAh) can be used.
After replacement switch the pump ON (normal user mode), the internal real time clock will be re-initialized. Switch off the pump.
Go into the configuration and maintenance mode and set the clock by means of the ARGUSservice tool (see chapter 3.6).
The alarm message disappears at next power-up cycle (see chapter 6.9).

Spare parts and its reference numbers are listed separately in the “Spare Part List”.

42 14.543_E_SM_en_A606S_4.8X
8. Wiring Diagram

14.543_E_SM_en_A606S_4.8X 43
9. Bloc Schematic

44 14.543_E_SM_en_A606S_4.8X
10. Safety Standard Check

The SSC has to be performed at least every 24 months or after 10’000 hours of operation. The check has to be done in accordance
with the user- and the service manual.

SAFETY STANDARD CHECK (ssc) CODAN ARGUS AG / ARGUS 606 S


SERIAL NO: DISTRIBUTOR / HOSPITAL:
INVENTORY NO:
SOFTWARE VERSION: DEPARTMENT:
1 Check the software version • Update is required; new version no: ..............................
• Software is up-to-date
2 Visual check for damage, completeness and • Housing, Iabeis, connectors, power cable, accessories, etc. 
cleanness • Pull out syringe holder completely (without syringe) 
• Turnsyringe holder to left/right position; verify the proper \J func- 
tioning of the mechanical stop; the distance from handhold 0 stop
to duct stop must measure > 18 mm (otherwise the guidance duct
is worn out)
• Piston clamp closes completely (without syringe) 

3 Execute keypad and display test • See chapter 6.2 


4 Check switches and potentiometers • Start to infuse and test general activity status by verifying alarm 
and buzzer functions plus red blinking status lamp. Repeat this
procedure until all the tests are done:
• Press clutch Iever down: verify alarm indicators, display picture and 
text information
• Lift the syringe a little from the syringe switch: verify alarm indicat- 
ors, display picture and text information
• Press the clamp Iever up: verify alarm indicators, display picture 
and text information
• Pull the syringe holder: verify alarm indicators, display picture and 
text information
5 Insert a 50 ml syringe • Verify whether the pump runs smooth at its maximumrate 
6 Check the occlusion alarm pressure (see section • Preset Ievei: 900 mbar; measured Ievei: ........... mbar 
6.5) • Temporarily set the preset Ievei to the maximum and make sure to 
reach a pressure 2! 1.2 bar
7 Check accuracy • See chapter 6.6; measured volume: ........................ ml 
8 Check the "nurse-call" connection • Without alarm 
• By provoking analarm 
9 Check the docking station interface (insert • The indicator for mains operation must light green 
pump into a docking station)
10 Check time and date • Use the ARGUSservice: "Toolbox/Set clock" 
11 Check battery runtime (see chapter 6.7) • Charge the battery 
• Switch the pump on (to reset the infusion time) and discharge the 
battery at 2 mllh until the pump switches off automatically
• Recall the infusion time/runtime, comapre it with the value (see 
ARGUSservice index 368)
• Charge the battery again 
For internal use by Codan Argus only Electrical test Use a 2.5 m power cable and attach the measurement results 
(IEC 62353)
The ARGUS 606 S has passed the SSC and is safe for use; reset now the service interval!
Date
Name Signature

14.543_E_SM_en_A606S_4.8X 45
11. Repair Order Form

Purchase Order
Proforma Invoice Number:
Customer Name:

Address:

Contact Person:

Tel. Number:

Device Identification :
ARGUS 717 V ARGUS 707 V ARGUS 600 S ARGUS 404
ARGUS 718 V ARGUS 708 V ARGUS 606 S ARGUS 414
ARGUS 300 P ARGUS 500 P ARGUS 600 P ARGUS 60 P ARGUS 100 P
ARGUS 300 M ARGUS 500 M ARGUS 600 M ARGUS 60 M ARGUS 100 M
Accessory:

Serial-No.: Production Code:

Detailed failure or problem description:

Excepted work / repair to be done:


Repair Description:
Warranty repair
Replacement
Other

Date: Name: Signature:

46 14.543_E_SM_en_A606S_4.8X
14.543_E_SM_en_A606S_4.8X

CODAN ARGUS AG
Oberneuhofstrasse 10
CH-6340 Baar
Phone +41 41 785 09 44
Fax +41 41 785 09 40
info@codanargus.com
www.codanargus.com

You might also like