Professional Documents
Culture Documents
Service Manual
14.543_E_SM_en_A606S_4.8X
Table of Contents
1. Introduction....................................................................... 1
1.1 General................................................................................ 1
3. ARGUSservice Software.................................................... 3
3.1. General................................................................................ 3
3.2. Use of the ARGUSservice Software....................................... 4
3.3. Configuration Indices........................................................... 6
3.3.1 Description of configuration settings with flags.................... 10
3.4. Factory Settings.................................................................... 12
3.4.1. Software Versions................................................................. 12
3.4.2. ARGUS 606 S / Software 4.80.............................................. 12
3.5 Configuration of User-Syringes............................................. 13
3.6 TOOLBOX............................................................................ 15
3.7 History Log........................................................................... 18
3.8 History Example................................................................... 19
4. Serial Communication....................................................... 20
4.1. General................................................................................ 20
4.2. Serial Communication Protocol............................................. 20
4.3. Installation of an optional interface board............................ 20
4.4. Pump connections ............................................................... 21
5. Software Updates............................................................. 22
5.1. Firmware Update Procedure................................................. 22
5.2. Re-import the saved configuration file.................................. 27
5.3. Display Language................................................................. 29
5.4. Safety Aspects...................................................................... 30
5.5. Upgrade Procedure from a BASIC to a TIVA Pump................ 30
6. Maintenance...................................................................... 31
6.1. General................................................................................ 31
6.2. Keypad and Display Test....................................................... 31
6.3. Factory Base Syringe Calibration........................................... 33
6.4. Syringe Calibration............................................................... 34
6.5. Pressure Calibration.............................................................. 36
6.6. Pump Accuracy Measurement.............................................. 38
6.7. Correction of the Battery Capacity ....................................... 38
6.8. Fault Codes.......................................................................... 39
6.9. Message “Safety Check is due”........................................... 40
7. Replacement of Parts........................................................ 41
7.1. General................................................................................ 41
8. Wiring Diagram................................................................. 43
9. Bloc Schematic................................................................... 44
14.543_E_SM_en_A606S_4.8X 3
4 14.543_E_SM_en_A606S_4.8X
1. Introduction
1.1 General
Important!
This service manual is intended for the exclusive use of authorized persons who have been trained in maintenance and repair of the
ARGUS 606 S syringe pump by CODAN ARGUS AG.
The service manual is meant to be used together with the user manual.
Important!
CODAN ARGUS AG shall not assume any responsibility for any manipulations which have been carried out on the ARGUS 606 S
syringe pump by a non-authorized person.
Warning!
The ARGUS 606 S syringe pump may only be used with spare parts, accessories, consumables and syringes with Luer-Lock connec-
tions recommended by CODAN ARGUS AG. The functional safety of the pump is not guaranteed if non approved materials are
used. The safety of the patient may be endangered.
This manual contains the latest data available. It is subject to changes in accordance with technical improvements.
CODAN ARGUS AG
Oberneuhofstrasse 10
CH-6340 Baar
Phone +41 41 785 09 44
Fax +41 41 785 09 40
codan@codanargus.com
www.codanargus.com
14.543_E_SM_en_A606S_4.8X 1
2. Configuration and Maintenance Mode
2.1. General
Warning!
Changes in the configuration mode (with or without the “ARGUS service” PC utility tool), preventive maintenance measurements or
repair constitute a modification of the pump and may only be carried out by authorized persons!
After each change of a configuration, a function check and a control measurement has to be performed!
To enter the configuration and maintenance mode hold the 3rd numeric Key pressed while pressing the ON/OFF Key:
The factory setting of the PIN code is 0 unless otherwise specified by customized configuration. It is recommended to define a PIN
code to prevent unauthorized users from entering.
Press now the START/STOP key to get access to the submenus; use the MENU key to switch from one menu to the next:
The first menu after entering the configuration and maintenance mode is “Pump Settings” in which the following adjustments are
possible:
*for a temporary change of these settings during normal infusion mode, see chapter “Programmable Options” in the User Manual.
The handling of the other menus “Drive Calibration”, “Pressure Calibration”, “Keypad & Display Test” is described in chapter 6
“Maintenance”.
2 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software
3.1. General
For easy and quick configuration procedures it is recommended to use the PC software ARGUSservice. This utility is user friendly PC
software with various functions working over the PC serial COM port. This Windows based software allows you
– to change configurations
– to do a software upgrade
– to set the pump clock
– to set or change a PIN code
– to read and print out the pump’s history
– to replicate easily pump configurations
– to set the service interval
– to download clear text in a language different from English
– to re-program the serial number
– to download medication libraries
The up-to-date and clearly structured design of this self-describing PC-tool allows a very easy and rapid modification of the
ARGUS 606 S, ARGUS 600 S, the ARGUS 707 V and ARGUS 708 V pumps as well as the ARGUS docking stations. This software is
available from your local distributor or directly from CODAN ARGUS AG.
REMARK:
The ARGUSservice is designed for the use with device software versions greater or equal to 4.00.
Warning!
The syringe pump has to be disconnected from the patient before and while the serial interface cable is connected to the pump.
The connection of the ARGUS 606 S over the serial interface RS232 is done by connecting the interface cable (REF 10.093 / 600951)
to the serial interface outlet of the serial PC-COM port.
Warning!
After each change of a configuration, a function check and a control measurement has to be performed!
An active communication between the pump and the PC is not apparent on the pump.
14.543_E_SM_en_A606S_4.8X 3
3. ARGUSservice Software
Please verify that you are in possession of the latest ARGUS ser-
vice version. Updates can be downloaded from the Login section
(password protected) under www.codanargus.com.
Open the software tool and click on the respective product type.
4 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software
Use the “Report” button to create an overview of the present ARGUS 707 V
ARGUS 708 V
ARGUS 606 S
ARGUS 100 P
Accessory:
A404 Serial Number:
protocol” for analysis to the support center of CODAN ARGUS. Detailed failure or problem description:
Other Description:
Date: Signature:
14.543_E_SM_en_A606S_4.8X 5
3. ARGUSservice Software
1. “Menus”
The menus below must be activated in order to make them
available on the pump.
2. “Display”
Index #366 allows displaying the pressure in units different from
mbar.
Index #524 makes for the entry of a specific text containing a
maximum of 16 characters. This text will appear on the display
at switching on the pump.
3. “Handling”
The indices below allow various settings to specify the handling
of the pump.
Warning!
It is strongly recommended to activate index #5 for pumps be-
ing used with TIVA or dose calculation to ensure a maximum of
safety during a syringe change.
4. “Functionality”
The indices below refer to KVO, bolus and pre-alarm functions.
6 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software
5. “Technical”
Warning!
For safety reasons it is strongly recommended to keep index #19
“Buzzer tone at start of infusion” and #44 “Automatic pressure
release after occlusion” activated.
6. “Communication”
See chapter 4.2 “Serial Communication Protocol”
7. “Calibration Syringe”
The flag of index #508 and #509 must be on “YES”; otherwise
a syringe calibration is requested.
8. “Calibration Pressure”
The flag of index #502 and #503 must be on “YES”; otherwise
a pressure calibration is requested.
14.543_E_SM_en_A606S_4.8X 7
3. ARGUSservice Software
9. “Calibration Display”
The two indices allow a permanent adjustment of the backlight
brightness and the display contrast.
8 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software
14.543_E_SM_en_A606S_4.8X 9
3. ARGUSservice Software
2 Key ON/OFF only in STOP mode Yes: The key ON/OFF is disabled. Pump can be turned off only in STOP
usable mode.
No: The key ON/OFF is enabled. Pump can be turned off in STOP and
RUN mode.
3 Rate change allowed only in Yes: The infusion rate can be changed only in STOP mode
STOP mode No: The infusion rate can be changed in STOP and RUN mode
5 Ask for new therapy after syringe Yes: Question “Continue Therapy? YES/NO” appears after a syringe
change change.
No: Question “Continue Therapy? YES/NO” will not appear after a syr-
inge change
6 Nurse call settings Yes: A static alarm is activated for nurse call.
No: A single pulse alarm is activated for nurse call.
16 Syringe type acknowledge at start Yes: Syringe type must be acknowledged at first start of an infusion.
No: No syringe type acknowledgement necessary.
17 KVO enabling or disabling Yes: KVO function is enabled
(See also indices 20, 21 and 60)
No: KVO function is disabled
19 Buzzer tone at start of infusion Yes: Audible buzzer tone is emitted at start of infusion
No: No audible buzzer tone is emitted at start of an infusion
20 KVO setting for infusion rates up KVO rate can be set between 0.1 - 5.0ml/h
to 3ml/h (For this setting to be active, index 17 must be set to true.)
21 KVO setting for infusion rates KVO rate can be set between 0.1 - 5.0ml/h
higher than 3ml/h (For this setting to be active, index 17 must be set to true.)
22 One-time standby alarm Yes: Standby alarm is released only after the first start of an infusion (See
also index 80)
No: Standby alarm is released even before the first start of an infusion.
Standby alarm is released in STOP mode and in KVO mode.
23 Menu PRESSURE SETTING Yes: Menu PRESSURE SETTING is enabled on the pump. The user is able
to change the alarm limit temporarily.
(See also index 365 to set the default alarm limit and index 366 to select
other pressure units)
No: Menu PRESSURE SETTING is disabled and not visible on the pump.
24 Menu BATTERY INFO Yes: Menu BATTERY INFO is enabled on the pump. The remaining battery
capacity is displayed in percentage and time (format hh:mm), it is calcu-
lated on the basis of the current infusion rate.
No: Menu BATTERY INFO is disabled and not visible on the pump.
27 Menu KEY LOCK Yes: Menu KEY LOCK is enabled on the pump. The numeric keys of
the key pad can be locked / unlocked on the basis of a PIN code being
defined in index 50.
No: Menu KEY LOCK is disabled and not visible on the pump.
32 Menu BOLUS Yes: Menu BOLUS is enabled on the pump offering a manual and auto-
matic bolus function.
(See also index 38 for disabling the automatic bolus function.)
No: Menu BOLUS is disabled and not visible on the pump. No bolus
function is available.
33 Menu SETTINGS Yes: Menu SETTINGS is enabled on the pump. User can temporarily
change buzzer volume (see also index 363) and standby time (see also
index 80).
No: Menu SETTINGS is disabled and not visible on the pump
10 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software
38 Automatic bolus function Yes: Automatic and manual bolus functions are enabled
(For this setting to be active, index 32 must be set to true)
No: Only the manual bolus function is enabled.
44 Automatic pressure release after Yes: The pump stops infusing after the detection of an occlusion and
occlusion reduces the pressure by reversing the infusion flow. Once the pressure is
reduced, the pump stops.
No: The pump stops after detection of an occlusion without reducing
pressure.
47 Standby- and battery pre-alarm Yes: This index automatically reduces the sound volume to the minimum
low sound volume in case of a standby alarm or a battery pre-alarm.
No: The sound volume of the pre-alarms corresponds to the programmed
sound volume.
50 PIN code definition for the KEY Enter a number of maximum 4-digits
LOCK
60 KVO only after infusion com- Yes: KVO function is only activated when the volume to be infused (VTBI)
pleted is reached
(For this setting to be active, index 17 must be set to true.)
No: Standard KVO function is activated
65 Clear infused volume and con- Yes: The infused volume can be cleared (in STOP and in RUN mode).
tinue the infusion No: The clear and continue function is disabled.
80 Time period until standby alarm This index defines the time period after the last operating step until a
is released standby alarm is released (only in STOP mode)
363 Buzzer alarm sound volume The alarm volume is adjustable from 5 to 10 (5 = Minimum, 10 = Max-
imum)
365 Pressure default alarm level A pressure alarm is released as soon as the defined limit is reached. The
range reaches from 200mbar to 1200mbar
366 Pressure display unit The user can chose one of the given pressure units below:
1 = [mbar]
2 = [mmHg]
3 = [kPa]
4 = [cmH2O]
5 = [Psi]
(For this setting to be active, index 23 must be set to true)
370 Target volume nearly reached Yes: A pre-alarm announces that the target volume is nearly reached
pre-alarm (Time setting see index 371)
No: Setting is deactivated. No pre-alarm for “target volume nearly
reached” is released.
371 Time period for the release of the This index defines the time range of the pre-alarm ‘Target volume nearly
pre-alarm target volume nearly reached’. The release time is adjustable from 1 min to 15 min. (For this
reached. setting to be active, index 370 must be set to true)
522 Allow change of TOTAL VOLUME Yes: A total volume modification is allowed during a running infusion.
while infusing No: During a running infusion a total volume modification is not allowed.
1612 Medication selection required Yes: The user is forced to select a medication before starting an infusion.
No: The user is able to start an infusion without selecting a medication.
1636 Menu PURGE Yes: Menu PURGE is enabled.
No: Menu PURGE is disabled and not visible on the pump.
14.543_E_SM_en_A606S_4.8X 11
3. ARGUSservice Software *********
Type:
ARGUS Pump Configuration *********
ARGUS 606 S
Version: 4.80
Serial: 003 5 040
The following chapters show the configuration reports with factory settings. Please note that most of the pumps are already cus-
======================= Configuration ==========================================
tomized when delivered to the customer. A report of your current pump configuration can be generated by pressing the “Report
Menues -------------------------------------------------------------------------
button” in the ARGUSservice tool when the pump is in the configuration 23 :mode.
Menu PRESSURE SETTING Yes No/Yes (No/Yes)
24 : Menu BATTERY INFO Yes No/Yes (No/Yes)
Each setting in your configuration report which varies from the factory setting will be
27 : Menu KEY LOCK
marked by a “!” on the
Yes
left side of the re-
No/Yes (No/Yes)
spective configuration index. 32 : Menu BOLUS Yes No/Yes (No/Yes)
1636 : Menu PURGE No No/Yes (No/Yes)
33 : Menu SETTINGS Yes No/Yes (No/Yes)
3.4.1. Software Versions 369 : Automatic Menu fall back delay tim 60 5 - 60 (sec)
80 : Time period until standby alarm is 120 60 - 3600 (sec)
The latest firmware version will be used for all new devices and replaces all previous software versions . Thus, the distinction
81 : Number of medication names display 6 0 - 255 ()
Approved syringes:
Syringes from various manufacturers are approved to be used with the ARGUS 606 S (see chapter 3.3, point 14.) A list of validated and
approved syringes can be found on our website www.codanargus.com, (login section, manuals for ARGUS 606 S).
Warning!
Only syringes fulfilling the above requirements may be programmed as „USER“ syringe:
14.543_E_SM_en_A606S_4.8X 13
3. ARGUSservice Software
3. Set the name of the user syringe and enter the measures of
the barrel diameter, the plunger length, the clamp diameter and
the syringe length.
Note:
User syringes 20/30 ml:
To record a 30 ml syringe, the syringe length must be measured
on the graduated scale from 0 – 20 ml.
Warning!
After the configuration of a user syringe, a function check and a
control measurement have to be performed!
14 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software
3.6 TOOLBOX
14.543_E_SM_en_A606S_4.8X 15
3. ARGUSservice Software
Use this feature to set an alarm reminding that the safety check
is due. The released alarm is displayed on the pump at power
up showing the message “Safety Check is Due!” accompanied
by an acoustic signal.
Before activating the “Service Interval”, the configuration
indices #393 (interval in months) and #394 (interval in hours)
have to be filled in accordingly.
16 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software
Note:
Reprogramming a serial number is possible without permission key in case the serial number shows 0 after a mainboard change or
after a software update.
In case that a serial number is still available on the board, a permission key must be requested from Codan Argus.
14.543_E_SM_en_A606S_4.8X 17
3. ARGUSservice Software
18 14.543_E_SM_en_A606S_4.8X
3. ARGUSservice Software
14.543_E_SM_en_A606S_4.8X 19
4. Serial Communication
4.1. General
The ARGUS 606 S pump has one serial interface connector per default being connected to ARGUS Docking Stations.
An additional optional interface board is available for a RS 232 interface on the RJ45 connector including a separation device ac-
cording to EN60601-1-1:2000.
Please refer to chapter 4.3 for installation directions.
Other optional interface types (Ethernet, USB, WLAN) will be provided on request.
Warning!
The serial interface designed by CODAN ARGUS enables the communication with ARGUS 100 M and ARGUS 60 M docking sta-
tions. In such a case the separation device according to EN60601-1-1:2000 is given by the docking station itself!
It is prohibited to connect the interface cable (REF. 600951/10.093) to the docking interface of the pump while the pump is connec-
ted to a patient!
An active communication between the pump and the PC is not apparent on the pump!
The following characteristics are valid for all ARGUS devices (volumetric pumps, syringe pumps, docking stations with V4.xx and
PCs) which are intended to communicate with the device mentioned in this service manual.
• Full-duplex RS232, currently 4800Baud for single pumps, 9600 Baud for docking stations (also on master/slave-link).
• Simple master (host/PC) – slave (device) communication (host does polling).
• The host has to repeat the request if there is no valid response.
• Uses a checksum (CRC-8).
• Binary data transmission, thus no ASCII/text parsing.
• Fast & direct communication with pumps on ARGUS docking station.
• Specified timeouts during remote mode.
• Basic framing technique used as in the Serial Infrared Link Access Protocol (IrLAP) version 1.1.
Please contact your local distributor or CODAN ARGUS AG for the complete serial communication protocol description.
An optional interface board can be installed on the main board of the ARGUS 606 S.
Remove any covers on the main board connectors (if any).
Please plug-in the interface vertically and do not bend it. Carefully close the device and ensure a proper positioning of the interface
board in the mechanical guide included in the device cover.
To activate the interface board it is necessary to adjust the respective indices in the pump configuration (tree “Communication”).
We recommend performing a safety standard check after the installation of an optional interface board.
20 14.543_E_SM_en_A606S_4.8X
4. Serial Communication
14.543_E_SM_en_A606S_4.8X 21
5. Software Updates
Those items are available from your local distributor or from CODAN ARGUS AG.
Warning!
After a software update, a safety standard check has to be performed!
An active communication between the pump and the PC is not apparent on the pump.
Prearrangements
Make sure that the pump is connected to your PC by means of the ARGUS interface cable
(REF 600951/10.093)
22 14.543_E_SM_en_A606S_4.8X
5. Software Updates
Check the selected syringe and the syringe limits and note
them down manually.
Warning!
It is strongly recommended to save the actual configuration in
order to maintain the serial number and the calibration para-
meters. In addition it is advisable to create a report in order to
crosscheck the configuration after the update.
14.543_E_SM_en_A606S_4.8X 23
5. Software Updates
If all measures have been taken press yes; if not, press no and
make the required saves.
24 14.543_E_SM_en_A606S_4.8X
5. Software Updates
14.543_E_SM_en_A606S_4.8X 25
5. Software Updates
Warning!
Only the latest version of the ARGUSservice will contain the
latest pump firmware embedded. Compare with the version
under www.codanargus.com in the download section (the
necessary password can be ordered under codan@codanargus.
com).
Select “OK”
Should the pump start in regular mode, turn it off and start it
again in configuration mode by
pressing the 3rd numeric key while pressing the ON/OFF Key.
26 14.543_E_SM_en_A606S_4.8X
5. Software Updates
Edit – Pin 0
14.543_E_SM_en_A606S_4.8X 27
5. Software Updates
Ò Activate the flags on „ Full configuration / Calibration area / Serial Number“, only if requested by the program
If the download procedure failed, the pump will prompt the “CrtL” alarm (“Safety Check is due”):
a) check the serial number and the calibration flags on indices 502/503 508/509
b) execute a safety standard check incl. syringe and pressure calibration (see also 6.4 and 6.5)
28 14.543_E_SM_en_A606S_4.8X
5. Software Updates
If you need another display language than English it’s necessary to load the respective language file into the pump. The latest lan-
guage files are available under www.codanargus.com (Login section, Software, ARGUS 606 S).
Open “Toolbox”
OpenClick on “Language”
Select the language file (type *. HEX). Your language file can be downloaded from codanargus.com\Login
After the download, the pump returns automatically into the normal operating mode.
Warning!
Respect the safety aspects as described in the following chapter.
14.543_E_SM_en_A606S_4.8X 29
5. Software Updates
Warning!
For medical device traceability your local distributor or CODAN ARGUS AG needs to be informed about every device update (serial
number) you performed!
Warning!
Do not make any software updates as long as the device is in use and/or connected to a patient!
Warning!
An active communication between the pump and the PC is not apparent on the pump!
All previous Basic or Tiva versions can be upgraded to the new version
Please take note of the following remarks before you start the upgrade:
– There is no additional hardware modification required,
– The upgrade is subject to an additional charge.
– The upgrade procedure requires direct contact with CODAN ARGUS AG, Switzerland to se-
cure traceability and correct labeling. You don’t need any upgrade key.
– Unauthorized upgrades may lead to discontinued function of the previously in-
stalled standard firmware and to the termination of any warranty claims!
– A safety standard check has to be performed on each upgraded pump!
30 14.543_E_SM_en_A606S_4.8X
6. Maintenance
6.1. General
Warning!
Only authorized persons who have been trained by CODAN ARGUS AG or by the local distributor are allowed to service the ARGUS
606 S Syringe pump. In case of a request for repair, send the device together with the filled out “repair order form” (see chapter
11.) to the local distributor. Further information is available from:
CODAN ARGUS AG
CH-6340 Baar / Switzerland
E-mail: codan@codanargus.com
Warning!
The Safety Standard Check (SSC) has to be performed at least every 24 month or after
10’000 hrs of operation. The check has to be done in accordance with chapters 6.2 – 6.9.
No special maintenance of the ARGUS 606 S syringe pump is necessary. There are no
wear and tear parts.
Warning!
It is recommended to define a PIN code to prevent unauthorized users from entering the configuration mode.
Enter the configuration mode (refer to chapter 2.2). Press the MENU Key several times to enter the described menus here after
Display Test
This test enables the detection of pixel errors. The control of the
lines must be done twice. The correct functioning of the display
is confirmed by a black display at the end of this test.
1. Press the MENU Key several times until „KEYPAD & DISPLAY
TEST“ is displayed.
2. Use the SELECTION Key to choose „DISPLAY TEST“
3. Press the START Key 2x
14.543_E_SM_en_A606S_4.8X 31
6. Maintenance
Keypad Test
This test checks the correct functioning of all Keys ex-
cept the ON/OFF Key. The ON/OFF Key is automatically
checked at any switching on and off of the pump. The test
was successful after every Key has been pressed. To fin-
ish the test press the ON/OFF Key to turn the pump off.
32 14.543_E_SM_en_A606S_4.8X
6. Maintenance
The factory base calibration is an integral part of the manufacturing process done by CODAN ARGUS AG. This process must be
repeated after the replacement of an assembly, e.g. main board, complete the syringe drive or parts of it, syringe barrel holder,
potentiometer, etc., as well as after a downfall or a heavy impact.
Please remove any syringe before you start the calibration process.
2. Press the START Key to get access to the different signal voltages.
Do not press this key again as this will only be done during the final calibration with a closed device!
3. Make sure that the red syringe barrel holder stands in completely closed position; adjust the white potentiometer to 360mV ±
40mV (BARREL DIAM.); tighten the set screw on the white cog wheel and verify the declared value again.
4. Pull out the syringe barrel holder completely; the signal of the barrel diameter must be within 2600mV and 3180mV. If this
range is exceeded, you have to repeat step 3.
5. Push the syringe drive towards the left. Adjust the black potentiometer on the drive unit to 300mV ± 10mV (PLUNGER POS.);
tighten the set screw and verify the declared value again.
6. Push the syringe drive towards the right; the signal of the plunger position must read between 2900mV and 3000mV. If this
range is exceeded, step 5 has to be repeated.
9. Make sure that the piston clamps are fully closed and that the clamp lever does not abut against the table; the signal of the
CLAMP DIAM. must read 460mV ± 50mV. If this value is not reached, steps 7 to 8 have to be repeated.
10. Fully open the piston clamps; the signal of the CLAMP DIAM. must read between 2400mV and 3000mV. If this range is not
reached, steps 7 to 9 have to be repeated.
11. Press the EXIT Key to abandon the calibration menu. Do not press the START key!
12. Close the device and go through the final calibration (chapter 6.5 – 6.6)
The disregard of the above given range of values may provoke a technical alarm when the pump will be started up.
14.543_E_SM_en_A606S_4.8X 33
6. Maintenance
The DRIVE CALIBRATION menu has two submenus, one for cal-
ibration part 1 and one for part 2.
The submenus allow the calibration of the barrel, the clamp dia-
meter and plunger position.
Calib. Part 1
34 14.543_E_SM_en_A606S_4.8X
6. Maintenance
Calib. Part 2
3.
Insert the short calibration part ( 31, l=20mm,
REF.600584/11.194)
14.543_E_SM_en_A606S_4.8X 35
6. Maintenance
Before starting this procedure, make sure that the pump casing is closed and screwed together to ensure a stable force distribution.
2. Syringes or calibration parts must not be inserted at this point; the driving head has to be positioned midway.
4. The pump runs backward to reach a stable offset value. A target value of 600mV is pre-programmed. As soon as the offset value
has stabilized, press the second numeric Key several times until approximately 600mV are reached and remain stable.
5. Press the START Key twice. Caution: The pump does not stop automatically at the mechanical end stop!
Calib. Sensitivity
36 14.543_E_SM_en_A606S_4.8X
6. Maintenance
Control Measurement
1
Use a manometer and a 50 ml syringe for the calibration and the control measurement if there is no spring balance available.
14.543_E_SM_en_A606S_4.8X 37
6. Maintenance
The capacity of the battery will decrease over time like all
batteries do. Therefore the capacity status (percentage and
remaining time) displayed in the menu “BATTERY INFO” may
differ from the real battery capacity.
The configuration menu in the ARGUSservice tool allows ad-
justing the battery capacity to real capacity values. The default
value of a new battery is 100% (see index 368). The battery
capacity can be reduced down to a minimum of 50 %.
Warning!
The battery charging system included in the ARGUS 606 S re-
quires additional electronics (temperature and current sensors)
to provide a high current to accelerate charging time. Therefore
the battery pack must be obtained from CODAN ARGUS AG.
The charging time may depend on the current pump status
(infusing rate, stand-by).
38 14.543_E_SM_en_A606S_4.8X
6. Maintenance
A technical failure will be indicated on the pump by a continuous acoustic alarm and/or a continuous red indicator lamp. During a
failure state, the fault code on the LC display allows the analysis of the cause:
Example of a technical failure
If the pump was switched OFF after a detected failure, the fault code will be stored in the history log and also in the configuration
file of the pump (index 380 - 385).
Fault
Code Possible Failure cause
00 Reset test failed
901 Check all cable connections
10xx Software error or memory error
11xx Software error
12xx Software error
13xx Software error
14xx Hardware error
1404 Check all cable connections
1405 Unconnected stepper motor
1406 Check all cable connections
1413 Unconnected stepper motor
1414 Short circuit in stepper motor or in its connection
1415 Change the mainboard
1419 Battery voltage out of range
1421 Input voltage out of range
1425 Voltage reference on main board out of range
1426 Unconnected DMS-Board or
voltage reference on DMS board out of range
1428 Internal temperature sensor defective or
internal device temperature too high
1429 Battery temperature sensor defective or
battery temperature too high
1434 Bad calibration of size potentiometer?
1435 Bad calibration of size potentiometer?
1436 Bad calibration of clamp potentiometer?
1437 Bad calibration of clamp potentiometer?
Unconnected flex cable to drive movement control
1438 Bad calibration of pressure potentiometer?
1439 Bad calibration of pressure potentiometer?
1440 Unconnected display cable on display board
15xx Software error or internal bus error
16xx Pressure out of range
1601 Pressure signal out of range, DMS defective
17xx Problems with the drive
14.543_E_SM_en_A606S_4.8X 39
6. Maintenance
Fault
Code Possible Failure cause
1704 Syringe drive not correct installed ( Check the holder )
1710 Plunger calibration out of range
1711 Plunger signal out of range
18xx Problems with the Keypad
1800 Key pressed too long or Key blocked
1804 Check display board connection, then all other cables
1832 Check all cable connections
19xx Problems with the battery
1902 Unconnected battery
1904 Unconnected battery
1908 SMD Fuse 0.5AF defective
20xx Software error
21xx Software error
22xx Syringe dimension invalid
2201 Barrel calibration diameter out of range
2202 Clamp calibration diameter out of range
23xx Software error
24xx Communication error
25xx Software error
26xx Memory error (hardware failure)
27xx Software error
28xx Memory error (hardware failure)
29xx Internal bus error (hardware failure)
30xx Internal bus error (hardware failure)
After the time of use has elapsed and a safety standard check is due, the reminder alarm “Safety Check is Due” will be displayed
after power up of the pump; See chapter 3.3 (“Statistics” indices 390-392).
40 14.543_E_SM_en_A606S_4.8X
7. Replacement of Parts
7.1. General
Warning!
The ARGUS 606 S may only be used with accessories which are approved by CODAN ARGUS AG. Please refer to the CODAN AR-
GUS “Spare Part Catalogue” containing the respective reference numbers!
Repair shall only be performed by trained specialists being in possession of a manufacturer’s authorization!
Repair shall only be done with origin spare parts which shall not be modified without prior manufacturer’s authorization!
Disconnect the power cord and all interface connections prior to disassembly!
Warning!
If the complete syringe drive or parts of it, the side wall with the pressure sensor, the pump housing or the main board was re-
placed, a full calibration according to chapters 6.4 – 6.6 is required. Also a factory base calibration might be required.
How to remove the red cover (REF. 600936/10.151) of the driving head (REF.600937/10.152):
Remove the 3 screws on the cover.
Important: To disassemble the unit, open the beaks by hand one third (or put a coin between the beaks) then pull the cover with
the levers out of the housing.
14.543_E_SM_en_A606S_4.8X 41
7. Replacement of Parts
Warning!
The shrink-wrapped PCB (Printed Circuit Board) includes a temperature sensor which has to be positioned in direction of the con-
nector (avoid a positioning towards the transformer)!
Charging procedure:
1. Discharge the battery @ 5ml/h until the pump turns automatically off. This first step is ne-
cessary in order to synchronize the battery and the firmware!
2. Recharge the battery
3. Execute battery test:
Discharge the fully charged battery @ 5ml/h
Read out the time being used to discharge the battery (chapter 6.6)
Compare it with the reference value of a new battery.
Repeat the charge/discharge cycle, if necessary, until the specified runtime is reached.
Replacement of the casing cover including LC display, keypad and display board
(REF. 601060/10.185)
Perform the Keypad and Display Test according to chapter 6.2.
Check the display contrast setting in the maintenance mode of the pump after replacement. The display contrast may vary.
Spare parts and its reference numbers are listed separately in the “Spare Part List”.
42 14.543_E_SM_en_A606S_4.8X
8. Wiring Diagram
14.543_E_SM_en_A606S_4.8X 43
9. Bloc Schematic
44 14.543_E_SM_en_A606S_4.8X
10. Safety Standard Check
The SSC has to be performed at least every 24 months or after 10’000 hours of operation. The check has to be done in accordance
with the user- and the service manual.
14.543_E_SM_en_A606S_4.8X 45
11. Repair Order Form
Purchase Order
Proforma Invoice Number:
Customer Name:
Address:
Contact Person:
Tel. Number:
Device Identification :
ARGUS 717 V ARGUS 707 V ARGUS 600 S ARGUS 404
ARGUS 718 V ARGUS 708 V ARGUS 606 S ARGUS 414
ARGUS 300 P ARGUS 500 P ARGUS 600 P ARGUS 60 P ARGUS 100 P
ARGUS 300 M ARGUS 500 M ARGUS 600 M ARGUS 60 M ARGUS 100 M
Accessory:
46 14.543_E_SM_en_A606S_4.8X
14.543_E_SM_en_A606S_4.8X
CODAN ARGUS AG
Oberneuhofstrasse 10
CH-6340 Baar
Phone +41 41 785 09 44
Fax +41 41 785 09 40
info@codanargus.com
www.codanargus.com