You are on page 1of 5

MODEL Brilliance 40-slice

X-RAY TUBE

X-RAY TUBE ANODE

Heat storage, hu (X-ray tube 8,0,000 (MRC technology)


anode)

Heat dissipation rate, hu/min (X- 1610000


RAY TUBE)

Tube cooling (X-ray tube anode) Oil/air

Tube focal spot, mm (X-ray tube 0.5 x 1 small, 1 x 1 large


anode)

POWER NEEDED 200-500 VAC, 50/60 Hz, 100 kVA, 3-phase

Type (CONFIGURATION) Multislice

# of slices (X-ray tube anode) 40

Image storage

Hd capacity, gb (IMAGE 292


STORAGE)

No. online images (IMAGE 514,242 uncompressed (512 x 512)


STORAGE)

Archive (IMAGE STORAGE) 9.1 GB rewritable EOD, 620 MB CD

HELICAL SCANNING (DISPLAY) Yes

Max scan time, sec (DISPLAY) 100

Max scan volume, cm (DISPLAY) 162


Spatial resolution, lp/cm 24
(DISPLAY)

Pitch (DISPLAY) 0 to 1.5, user selectable

Reconstruction time per image, 0.025, 3-D cone beam


sec (DISPLAY)

DICOM 3.0 COMPATIBLE (POWER Yes


NEEDED, VAC)

IMAGE PROCESSING (MAXIMUM


DISPLAY)

Cpu (NETWORK/DATABASE) Intel, Windows OS

Scan fovs, cm (IMAGE Up to 50


PROCESSING)

Image reconstruction matrixes 256 x 256, 512 x 512; optional 768 x 768, 1024 x 1024
(IMAGE PROCESSING)

Reconstruction time

Per slice, sec (Reconstruction Up to 40 images/sec with 3-D cone beam


time)

For localization scan, sec 5


(Reconstruction time)

PERFORMANCE

High-contrast spatial resolution 24


0% mtf, lp/cm (PERFORMANCE)

Low-contrast resolution, mm at 4 at 0.3%


% at <=4 rads (PERFORMANCE)
Sound, % at <= 2.5 rads 0.27
(PERFORMANCE)

DISPLAY

MONITOR 18" flat LCD 2

Matrixes, pixels (DISPLAY) 1280 x 1024

Range of ct numbers (DISPLAY) -1,000 to +3,095

Image enlarging scale (DISPLAY) Up to 10x; real time

TABLE FEATURES

Range of movement

Vertical, cm (Range of 52-104


movement)

Longitudinal, cm (Range of 190


movement)

Scanning range,cm (Range of 162


movement)

Max. patient weight, (precision), 204 (±0.25 mm)


kg (Range of movement)

SCORING OF CORONARY ARTERY Yes


CALCIFICATION (IMAGE
STORAGE)

GENERATOR

Output, kw (X-RAY 60
GENERATORS)

Kvp range (GENERATOR) 80,120, 140


Ma range (IMAGING SYSTEM) 20in 1 mA increments

FLOOR SPACE NEEDED

Size, m2 (FLOOR SPACE 25.9


NEEDED)

GANTRY

Geometry (GANTRY) Rotate-rotate

DETECTOR (SCATTERED LIGHT) 40 mm z-axis coverage, solid-state GOS

Rows (GANTRY) 52

Elements/row (GANTRY) 672

# detection channels (GANTRY) 34,944 elements 1

Rotation times, sec 360 0.5,0.75,1,1.5, 2; optional 0.4


(GANTRY)

Partial (GANTRY) 0.33; opt 0.28

Slice thickness, mm (GANTRY) 0.5-10 variable

X-ray fan beam angle, ° 54.4


(GANTRY)

Gantry angle deg (GANTRY) ±30

Gantry size, hxwxd,cm (GANTRY) 203 x 239 x 94

Gantry weight, kg (GANTRY) 1764

Gantry opening, cm (GANTRY) 70

Scan localizer (GANTRY) Laser

OTHER ATTRIBUTES Spiral AutoStart, isotropic imaging, functional imaging (perfusion),


(Interference compensation) calcium scoring, cardiac imaging, Dental CT, bone mineral analysis
(Q-BMAP), CT endoscopy, lung nodule assessment, stenosis
analysis, stent planning, true 40channel with 40-mm coverage,

FDA CLEARANCE (Interference Yes


compensation)

CE MARK (MDD) (Interference Yes


compensation)

MARKETING REGION
(Interference compensation)

This unit upgrade to 64 from the f actory with cardiac kit

You might also like