Auto-off Clear Calibration Battery Indicator Care and Maintenance Accessories Warranty INSTRUCTION MANUAL

To obtain the highest accuracy pH Buffer Solution Checker®Plus is guaranteed for a period of
one year for meter and six months for probe
HI98100
for measurements it is important
to follow these tips:
after date of purchase against defects in
workmanship and materials when used for
Checker®Plus
Code Description
their intended purpose and maintained
pH 4.01 buffer solution, according to instructions. This warranty
• Calibration is only as good as the HI70004P is limited to repair or replacement free of
20 mL sachets (25)
buffer being used. The pH buffer charge. Damage due to accidents, misuse,
values change over time once the pH 7.01 buffer solution, tampering or lack of prescribed maintenance
HI70007P is not covered. If service is required,
20 mL sachets (25)
sachets are opened. Fresh buffer contact your local Hanna office. If under
should be used for each calibration. pH 10.01 buffer solution, warranty, report the model number, date
HI70010P of purchase, serial number and the nature
20 mL sachets (25)
From measurement mode, press Place meter in calibration mode. The Checker®Plus features a low • The probe should be rinsed of the problem. If the repair is not covered
pH 4.01 & 7.01 buffer solution, by the warranty, you will be notified of the
and hold the ON/OFF button. The Press and hold ON/OFF until “CLr” battery indicator. When the battery with purified water each time HI77400P charges incurred. If the instrument is to be
20 mL sachets (10 pcs., 5 ea.)
meter will cycle through “OFF,” “CAL,” is displayed. The meter will now be is running low, the tag will appear before placing in buffer or sample returned to Hanna Instruments, first obtain
to be tested. pH 10.01 & 7.01 buffer solution, a Return Goods Authorization number from
then current auto-off setting. at default calibration. No tags will on screen. When the battery has HI770710P
20 mL sachets (10 pcs., 5 ea.) the Technical Service department and then
be shown in measurement mode been depleted, the tag will blink. • When the meter is not in use it is send it with shipping costs prepaid. When
shipping any instrument, make sure it is
until calibration is performed. important to add several drops of properly packed for complete protection.
storage solution to the protective Electrode Cleaning Solution
cap to keep the probe hydrated. If
“Err” Message Battery Replacement storage solution is not available,
pH 4 or pH 7 buffer can be used.
Code Description
general purpose cleaning
Recommendations
• For improved accuracy it is
HI700601P
solution, 20 mL sachets (25) for Users
recommended to calibrate a
minimum of two points. It is Before using Hanna products, make sure that
important to use buffers that Electrode Storage Solution they are entirely suitable for your specific
application and for the environment in which
bracket the expected value of the they are used. Operation of these instruments
The default setting is 8 minutes sample to be tested. For example, may cause unacceptable interferences to
Code Description
(“d08”). Press ON/OFF button if the expected value is pH 8, the other electronic equipment, thus requiring the
electrode storage solution, operator to take all necessary steps to correct
to change. “d60” is auto-off Checker®Plus should be calibrated HI70300L such interferences. Any variation introduced
500 mL bottle
after 60 minutes, and “d--” using pH 7 and pH 10 buffers. by the user to the supplied equipment may
disables the auto-off feature. electrode storage solution, degrade the instrument’s EMC performance.
HI70300M
In calibration mode, if the meter To change the CR2032 Li-ion battery, • It is important to calibrate and 230 mL bottle To avoid damages or burns, do not put the
instrument in microwave oven. For yours and
displays an “Err” message when in turn the battery cover located on the measure samples at the same the instrument safety do not use or store
the correct fresh buffer solution then back of the meter counterclockwise temperature. A dramatic change the instrument in hazardous environments.
the probe should be cleaned. Place to unlock. Remove cover and in temperature between buffer
the probe in the HI700601 cleaning replace with + side facing up. solutions and samples to be tested
solution for 15 minutes. Rinse with will give inaccurate readings.
purified water and place in storage All rights are reserved. Reproduction in whole or
in part is prohibited without the written consent
solution for 1 hour before calibrating. of the copyright owner, Hanna Instruments Inc.,
Note: Batteries should only be replaced in a safe area
Woonsocket, Rhode Island, 02895, USA.
If the “Err” message persists then the using the battery type specified in this instruction
manual. Old batteries should be disposed in accordance
HI1271 probe should be replaced. with local regulations. HI98100 INSTMAN 1.0 06/15
Thank You Meter Overview Operation Meter Calibration
A One or Two-Point Calibration with pH 7
Thank you for choosing a Hanna Preparation:
Instruments product. Please read
The pH electrode is shipped dry.
this instruction manual carefully

One-Point
Before using the Checker®Plus,
before using the instrument.
remove the protective cap and
For additional information, visit condition the electrode by soaking
www.hannainst.com or e-mail the tip (bottom 4 cm (1.5”)) in pH 7.01
us at tech@hannainst.com. buffer solution for several hours. Then
follow the calibration procedure. pH 4 will then blink on the display. Press the button to exit and save Meter will exit to measurement mode
calibration. “Sto” will then be displayed. and the calibration tag will be displayed.
• Do not be alarmed if white crystals Press the ON/OFF button to turn the From measurement mode, press If pH 7 buffer solution is used as the
appear around the cap. This is meter on. All tags will be displayed. and hold the ON/OFF button first point the buffer is recognized
normal with pH electrodes and they until “CAL” is displayed. with the blinking stability indicator.
dissolve when rinsed with water. The meter will go into measurement
mode: current reading and When the reading is stable, the

Two-Point
• Connect the electrode to the meter.
calibrated buffers are displayed. stability indicator will disappear
• Turn the Checker®Plus on by and pH 7 will be calibrated. If pH 7
pressing ON/OFF button. is the only calibration point, finish
• Remove the protective cap and one-point procedure at right.
immerse the tip of the electrode Liquid
If using pH 4 or pH 10 as a second point, Use pH 4 or 10 to perform a two-point When the reading is stable, the stability Meter will exit to measurement mode and
Specifications in the sample to be tested. Crystal
Display continue two-point procedure at right.
calibration. The value is automatically
recognized and displayed with the
indicator will disappear. “Sto” will be
displayed when the calibration is saved.
the calibration tags will be displayed.
• Stir gently and wait for a
blinking stability indicator.
Range 0.00 to 14.00 pH stable reading. ON/OFF
Button
Resolution 0.01 pH
NEVER IMMERSE THE ELECTRODE
B
When “7.01” blinks on the display,
Accuracy
±0.2 pH OVER THE MAXIMUM IMMERSION
place the tip of the probe into a
One-Point Calibration with pH 4 or pH 10
(@25°C/77°F)
LEVEL. THE CONNECTOR MUST
pH 4, 7, or 10 buffer solution.
Calibration automatic, one or two-point ALWAYS BE CLEAN AND DRY.
Maximum
Electrode HI1271 (included) Immersion
• After use, rinse the electrode with Level
Battery Type CR2032 Li-ion
water and store it with a few drops
approximately 1000 hours of HI70300 storage solution in the
Battery Life
of continuous use pH Probe
protective cap.
8 minutes, 60 minutes,
• Replace protective cap after For one or two-point
A
Auto-off
or can be disabled
each use. calibration using pH 7
0 to 50°C (32 to 122°F); buffer go to procedure A
Environment DO NOT USE DISTILLED OR
RH 95% max
DEIONIZED WATER FOR STORAGE If pH 4 or 10 buffer solution is used When the reading is stable, the stability Meter will exit to measurement mode and
50 x 174 x 21 mm (2 x 6.8 x 0.9“) Recommended For one-point calibration as the first point the value of the indicator will disappear. “Sto” will be the calibration tag will be displayed.
B
Dimensions PURPOSES.
(excluding probe) Immersion displayed when the calibration is saved.
using pH 4 or pH 10 buffer buffer is recognized and displayed
Level 4 cm (1.5”)
Weight 50 g (1.8 oz.) go to procedure B with the blinking stability indicator.