1 ALARM (CONTINUE)
ALARM MESSAGE DESCRIPTION
Code H0.1 ROM inside of main MCU fails. At this time, the heater will stop working ,press
Silence/Reset Key to clear alarming for 4min, please refer to theservice part.
Code H0.2 Internal system inside of main MCU fails. At this time, the heater will stopworking,
press Silence/Reset Key to clear alarming for 4min, please refer to the service part.
Code H0.3 Communication of main and auxiliary MCU A fails, At this time, the heater will
stop working, press Silence/Reset Key to clear alarming for 4min,please refer to the service
part.
Code H0.4 ROM inside of auxiliary MCU fails, At this time, the heater will
stopw o r k i n g , pr e s s S i l e n c e / R e s e t K e y t o c l e a r a l a r m i n g f o r 4 m i n , p l e a s e refer
to the service part.
Code H0.5 Internal system inside of auxiliary MCU fails, At this time, the heater willstop
working, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.
Code H0.6 E2ROM inside main MCU fails, At this time, the heater will stop working ,press
Silence/Reset Key to clear alarming for 4min, please refer to theservice part.
Code H0.7 SRAM failure (including system cord failure). At this time, the heater willstop
working, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.
Code H0.8 Main ADC failure(e.g., auxiliary system TLV2548, including wrong sample,chip
damage, no response and so on). At this time, the heater will stopworking, press
Silence/Reset Key to clear alarming for 4min, please refer to the service part.
Code H1.0 H e a t c i r c u i t f a i l ur e ( e . g . , h e a t e r, s o l i d r e l a y, m e c h a n i c a l r e l a y a n d
a l l control circuit fails). At this time, the heater will stop working, pressSilence/Reset
Key to clear alarming for 4min, please refer to the servicepart.
Code H1.1 The timer RTC inside the system fails or is not accurate, At this time, theheater will
stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service
part.
Code H1.2 If the keyboard circuit is broken, At this time, the heater will stop working,press
Silence/Reset Key to clear alarming for 4min, please refer to theservice part.
The system indicator is on
Code H1.3 If the battery inside of the controller is error, At this time, the heater willstop
working, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.
PROCEDURE
At first, connected power supply, press and hold Set key for 2s as soon as switch on until
Airtemprature indicator displays “ --.-“, and no display on Skin and Set temprature indicator.
And then,press Up or Down until Air temprature indicator displays “78.6”, after then, press
Set key to enter intothe system set screen, or press Silence/Reset key to exit.After entering into
the set screen, Set temprature indicator displays set code, Air temprature indicator displays the
specific set data.Turn on the power switch5-3 I2
Press and hold the for 3sEnter passwordPress key, EXIT Press or key to enter password
78.6,Then press key to confirm Error Password checkYES
System setting screenPress key or Key, set temperature indicator displaysthe set codePress or
Key can change current set value ENDPress key Press and together SAVE and EXIT Return to
original set Brightness set of indicator
S E RV I C E M A N U A L F O R I N F A N T I N C U B AT O R
EDITION/REVISION B/0
CODE 00.1
LED Lightness setIndicate the lightness , and set range is 00.1~01.5. The larger the lighter.
CODE 01.7 Set value of air flow Over temperature alarm A (37
Set method of alarm value for air flow blocked (set temperatuer
37, and set range is38.5~48.0. For example, 45.0 means alarm value for air flow blocked 45.0
CODE 01.8
Set value of air flow Over temperature alarm B (37)Set method of alarm value for air flow
blocked (set temperatuer >37
, and set range is 40.5~52.0.For example, 45.0 means alarm value for air flow blocked 45.0
E0.1 Air temperature sensor failure alarm Refer to Flow chart 5.1
E0.2 Isolated temperature sensor alarm Refer to Flow chart 5.2
E0.3 Air temperature deviation failure alarm Refer to Flow chart 5.3
E0.4 Air flow sensor alarm Refer to Flow chart 5.4
E0.5 Skin temperature sensor failure alarm Refer to Flow chart 5.5
E0.6 Over temperature failure alarm Refer to Flow chart 5.6
E0.7 Air circulation failure alarm Refer to Flow chart 5.7
E0.8 Wrong position of skin sensor alarm Check the position of skin sensor
E0.9 Fan motor stop alarm Refer to Flow chart 5.8
E1.0 High deviation temperature failure alarmCheck whether there is heat sourcenearby, if yes,
please leave far awayfrom the heat source.
E1.1 Low deviation temperature failure alarmCheck the close condition of all doorsand panels
H0.1 ROM failure alarm inside of Main MCU Replace Main MCU
H0.2 Internal system failure alarm inside of Main MCUReplace Main MCU
H0.3 communication failure alarm :
1. Check the connection of circuit between the main MUC and auxiliaryMCU
2. Replace main MUC and auxiliaryMCU
H 0 . 4 R O M f ai l u r e a l a r m i n s i d e a u x i l i a r y M C U R e p l a c e a u x i l i a r y M C U
H0.5 Internal system failure alarm insideauxiliary MCUReplace auxiliary MCU
H0.6 E2PROM failure alarm Replace Main MCU
H0.7 SRAM failure alarm Replace U13 on main board
H0.8 Main ADC failure alarm Refer to Flow chart 5.9
H0.9 Auxiliary ADC failure alarm Refer to Flow chart 5.10
H1.0 Heat system failure alarm Refer to Flow chart 5.11
H1.1 Real timer RTC failure alarm Replace U20 on mian board
H1.2 Keyboard fault alarmWhen the key board of the control panelon controller short-circuit
alarms, pleasereplace the control panel.
H1.3 Battery failure alarm Inner storage battery failure alarm:1. Check the wire connection
betweenthe inner storage battery and the power board.2. Replace the inner storage
battery.Power failure alarm Refer to Flow chart 5.12