TABLE 3.

1 ALARM (CONTINUE)
ALARM MESSAGE DESCRIPTION

Code H0.1 ROM inside of main MCU fails. At this time, the heater will stop working ,press
Silence/Reset Key to clear alarming for 4min, please refer to theservice part.

Code H0.2 Internal system inside of main MCU fails. At this time, the heater will stopworking,
press Silence/Reset Key to clear alarming for 4min, please refer to the service part.

Code H0.3 Communication of main and auxiliary MCU A fails, At this time, the heater will
stop working, press Silence/Reset Key to clear alarming for 4min,please refer to the service
part.

Code H0.4 ROM inside of auxiliary MCU fails, At this time, the heater will
stopw o r k i n g , pr e s s S i l e n c e / R e s e t K e y t o c l e a r a l a r m i n g f o r 4 m i n , p l e a s e refer
to the service part.

Code H0.5 Internal system inside of auxiliary MCU fails, At this time, the heater willstop
working, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.

Code H0.6 E2ROM inside main MCU fails, At this time, the heater will stop working ,press
Silence/Reset Key to clear alarming for 4min, please refer to theservice part.

Code H0.7 SRAM failure (including system cord failure). At this time, the heater willstop
working, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.

Code H0.8 Main ADC failure(e.g., auxiliary system TLV2548, including wrong sample,chip
damage, no response and so on). At this time, the heater will stopworking, press
Silence/Reset Key to clear alarming for 4min, please refer to the service part.

Code H0.9 Auxiliary ADC failure(e.g., auxiliary system TLV2544, including


wrongsample, chip damage, no response and so on). At this time, the heater will stop
working, press Silence/Reset Key to clear alarming for 4min, please refer to the service
part.

Code H1.0 H e a t c i r c u i t f a i l ur e ( e . g . , h e a t e r, s o l i d r e l a y, m e c h a n i c a l r e l a y a n d
a l l control circuit fails). At this time, the heater will stop working, pressSilence/Reset
Key to clear alarming for 4min, please refer to the servicepart.

Code H1.1 The timer RTC inside the system fails or is not accurate, At this time, theheater will
stop working, press Silence/Reset Key to clear alarming for 4min, please refer to the service
part.

Code H1.2 If the keyboard circuit is broken, At this time, the heater will stop working,press
Silence/Reset Key to clear alarming for 4min, please refer to theservice part.
The system indicator is on

Code H1.3 If the battery inside of the controller is error, At this time, the heater willstop
working, press Silence/Reset Key to clear alarming for 4min, pleaserefer to the service part.

PROCEDURE
At first, connected power supply, press and hold Set key for 2s as soon as switch on until
Airtemprature indicator displays “ --.-“, and no display on Skin and Set temprature indicator.
And then,press Up or Down until Air temprature indicator displays “78.6”, after then, press
Set key to enter intothe system set screen, or press Silence/Reset key to exit.After entering into
the set screen, Set temprature indicator displays set code, Air temprature indicator displays the
specific set data.Turn on the power switch5-3 I2
Press and hold the for 3sEnter passwordPress key, EXIT Press or key to enter password
78.6,Then press key to confirm Error Password checkYES

System setting screenPress key or Key, set temperature indicator displaysthe set codePress or
Key can change current set value ENDPress key Press and together SAVE and EXIT Return to
original set Brightness set of indicator

S E RV I C E M A N U A L F O R I N F A N T I N C U B AT O R
EDITION/REVISION B/0

CODE 00.1
LED Lightness setIndicate the lightness , and set range is 00.1~01.5. The larger the lighter.

CODE 00.2 Air temperature sensor compensation


Set value of temperature compensation for air temperature sensor , and set range is
-5.0~05.0 ,the first figure of set value means the increasing and decreasing value. For Example,
00.2
Means t h a t t h e a i r t e m p e r a t u r e a d d s 0 . 2 a s compensation. -1.3 means that the air
temperature reduces as compensation.

CODE 00.3 Air flow temperature sensor compensation


temperature compensation for air flow temperature sensor , and set range is -5.0~05.0 , thefirst
figure of set value means the increasing and decreasing value. For Example, 00.2
means thatt h e a i r f l o w t e m p e r a t u r e a d d s 0 . 2 a s compensation. -1.3
means that the air flow temperaturer e d u c e s 1 . 3 a s compensation.

CODE 00.4 Skin temperature sensor compensation


Set value of temperature compensation for skin temperature sensor , and set range is-5.0~05.0 ,
the first figure of set value means the increasing and decreasing value. For Example,00.2
m e a n s t h a t t h e s k i n t e m p e r a t u r e a d d s 0 . 2 a s compensation. -1.3
means that the skint e m p e r a t u r e r e d u c e s 1 . 3 a s compensation.

CODE 00.5 Machine Number


Set method of machine number, and set range is 00.0~02.0. For example, 00.1 measn No.
1machine, 01.5 means No. 15.

CODE 00.6 Motor speed alarm value


Set method of fan speed, and set range is 00.0~25.5. Every 0.1 means 20 rpm. For
example,05.0 means 1000 rpm. Alarm will be active if the fan speed is lower than this value.

CODE 00.7 Set value of Low deviation for air temperature


Set method of low deviation alarm in air mode, and set range is 00.0~05.0. For example,
00.8means 0.8 .

CODE 00.8 Set value of High deviation for air temperature


Set method of high deviation alarm in air mode, and set range is 00.0~05.0. For example,
00.8means 0.8 .
CODE 00.9 Set value of low deviation for skin temperature
Set method of low deviation alarm in skin mode, and set range is 00.0~05.0. For example,
00.8means 0.8 .

CODE 01.0 Set value of high deviation for skin temperature


Set method of high deviation alarm in skin mode, and set range is 00.0~05.0. For example,
00.8means 0.8 .

CODE 01.1 Minimum set value of air temperature


Set method of Minimum set value of air temperature, and set range is 20.0~34.0. For
example,25.0 means Minimum set value 25.0 .

CODE 01. 2 Maximum set value of air temperature


Set method of Maximum set value of air temperature in mode of>37.0
, and set range is37.0~39.0. For example, 39.0 means Maximum set value 39.0 .

CODE 01.3 Minimum set value of skin temperature


Set method of Minimum set value of skin temperature, and set range is 30.0~34.0. For
example,34.0 means Minimum set value 25.0 .

CODE 01.4 Maximum set value of skin temperature


Set method of Maximum set value of skin temperature in mode of>37.0
, and set range is37.0~39.0. For example, 39.0 means Maximum set value 39.0 .

CODE 01.5 Set value of Over temperature alarm A 37


Set method of Over temperature alarm in air mode (set temperatuer 37
, and set range is37.5~39.0. For example, 38.0 means over-temp alarm value 38.0 .

CODE 01.6 Set value of Over temperature alarm B ( 37 )


Set method of Over temperature alarm in air mode (set temperatuer 37
, and set range is38.5~40.0. For example, 40.0 means over-temp alarm value 40.0 .

CODE 01.7 Set value of air flow Over temperature alarm A (37
Set method of alarm value for air flow blocked (set temperatuer
37, and set range is38.5~48.0. For example, 45.0 means alarm value for air flow blocked 45.0

CODE 01.8
Set value of air flow Over temperature alarm B (37)Set method of alarm value for air flow
blocked (set temperatuer >37
, and set range is 40.5~52.0.For example, 45.0 means alarm value for air flow blocked 45.0

5.4.2 TEST EQUIPMENT REQUIRED


The test equipment listed below is required for troubleshooting the Controller. Equivalent
testequipment may be substituted.
Digital VOM, FLUKE 8060A
31.0
Air Flow Temperature Simulation

Resistor R=23166 Ω / 0.25W/0.5%

TABLE 5.1 ERROR CODES


ALARM CODE TYPE EXPLANATION AND CORRECTIVE ACTION

E0.1 Air temperature sensor failure alarm Refer to Flow chart 5.1
E0.2 Isolated temperature sensor alarm Refer to Flow chart 5.2
E0.3 Air temperature deviation failure alarm Refer to Flow chart 5.3
E0.4 Air flow sensor alarm Refer to Flow chart 5.4
E0.5 Skin temperature sensor failure alarm Refer to Flow chart 5.5
E0.6 Over temperature failure alarm Refer to Flow chart 5.6
E0.7 Air circulation failure alarm Refer to Flow chart 5.7
E0.8 Wrong position of skin sensor alarm Check the position of skin sensor
E0.9 Fan motor stop alarm Refer to Flow chart 5.8
E1.0 High deviation temperature failure alarmCheck whether there is heat sourcenearby, if yes,
please leave far awayfrom the heat source.
E1.1 Low deviation temperature failure alarmCheck the close condition of all doorsand panels

H0.1 ROM failure alarm inside of Main MCU Replace Main MCU
H0.2 Internal system failure alarm inside of Main MCUReplace Main MCU
H0.3 communication failure alarm :
1. Check the connection of circuit between the main MUC and auxiliaryMCU
2. Replace main MUC and auxiliaryMCU
H 0 . 4 R O M f ai l u r e a l a r m i n s i d e a u x i l i a r y M C U R e p l a c e a u x i l i a r y M C U
H0.5 Internal system failure alarm insideauxiliary MCUReplace auxiliary MCU
H0.6 E2PROM failure alarm Replace Main MCU
H0.7 SRAM failure alarm Replace U13 on main board
H0.8 Main ADC failure alarm Refer to Flow chart 5.9
H0.9 Auxiliary ADC failure alarm Refer to Flow chart 5.10
H1.0 Heat system failure alarm Refer to Flow chart 5.11
H1.1 Real timer RTC failure alarm Replace U20 on mian board
H1.2 Keyboard fault alarmWhen the key board of the control panelon controller short-circuit
alarms, pleasereplace the control panel.
H1.3 Battery failure alarm Inner storage battery failure alarm:1. Check the wire connection
betweenthe inner storage battery and the power board.2. Replace the inner storage
battery.Power failure alarm Refer to Flow chart 5.12