Professional Documents
Culture Documents
For safe operation, observe the following precautions. If not observed, the safety may be compromised.
1. Use the unit only for the purpose for which it is intended.
2. Follow the procedures, warnings and precautions described in the mamual.
3. Do not disassemble or modify the unit.
4. For repairs, contact our sales office or service representative.
WARNING Used for situations that may cause death or serious injury.
Used for situations that may cause slight injury or damage to the
CAUTION instrument.
Read the safety information in the instruction manual of the gas chromatograph before using this unit.
I
1. Corrosive gas and dust
The vital section of the measuring circuit of this detector is shielded in a case.
In view of maintaining the service life and accuracy of the detector, avoid installing in a dusty or
corrosive environment.
ECD VENT
ECD VENT
GC-2010
RADIOISOTOPE
NUCLIDE 63Ni
RADIOISOTOPE
NUCLIDE 63Ni
II
Contents
Precautions for Safe Operation
1 Outline
1.1 Principle of the ECD (Electron capture Detector) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Response Characteristics of the Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 Parts List
2.1 Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
3 Basic Operation
3.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
3.2 Gas used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Checking the glass insert and the injection inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
3.4 Mounting the column . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
3.5 Setting the flow rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
3.6 Setting the flow rate of detector gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.7 Starting the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.8 Setting the detector temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.9 Setting the temperatures of the column oven and injection port . . . . . . . . . . . . . . . . . 15
3.10 Checking the baseline . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.11 Setting the current value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.12 Setting the range/attenuation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.13 Caution during analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.14 Stopping the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
4 Key Operation
4.1 Setting the detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2 Selecting the detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
4.3 Selecting the detector gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
4.4 Detector ON/OFF . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
4.5 Setting the range/attenuation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
4.6 Setting the current value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.7 Zero adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
5 Service/Maintenance
5.1 Checking the ECD cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.2 How to Keep the ECD Cell Clean . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.3 Cleaning procedure for ECD Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.4 Actions to Be Taken in Case There Is the Doubt of Break . . . . . . . . . . . . . . . . . . . . . 44
ECD-2010 i
Contents
6 Troubleshooting
6.1 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.2 Error Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7 Disuse
7.1 Cautions at Disuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
8 Maintenance Parts
8.1 Maintenance Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
9 Specifications
9.1 Specifications of the ECD-2010 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
10 Appendix
10.1 Structure of ECD Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
10.2 Calibration Curve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
10.3 Conditioning of Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
ii ECD-2010
1 Outline
11.1
. 1.1 Principle of the ECD (Electron capture Detector)
Radiation (Beta-ray) from the radiation source (63Ni) sealed in the ECD cell ionizes an inert gas (N2). If a
pulse voltage is applied to the electrode in the cell, electrons are captured, allowing current to flow. If
electrophilic molecules enter the cell, the molecules absorb the electrons to be negatively ionized.
Negatively-charged molecules move more slowly than free electrons and take longer to reach the positive
electrode. Additionally, they are more likely to be bonded with positive ions again. Accordingly, the electrons
density in the detector drops so that a fewer number of electrons to be constant by increasing the pulse. The
total number of electrons to be captured in a unit time (and thus the current) can be maintained constant by
increasing the pulse frequency as the number of electrons decreases ; the variation in the number of pulses
is proportional to the density of the electrophilic molecules passing through the detector1).
1) R.J. Maggs et al., Anal. Chem. 43 (1971),1966
∆F = f - f 0 = Ka
f : Pulse frequency
f0 : Pulse frequency when carrier gas alone is allowed to flow (Basic frequency)
K : Constant determined by electron capture rate and other factors
a : Concentration of electron captivating substance
Chromatopac
ECD cell
Pulse shaper
Carrier gas
Amplifier E compares the average pulse current from the ECD cell with reference current IR and outputs a
voltage to the voltage-frequency converter (VFC) to make the two currents equal. The pulse from the VFC is
shaped to have appropriate width and height before being sent to the ECD cell, closing a control loop.
ECD-2010 1
1 Outline
11.2
. 1.2 Response Characteristics of the Detector
The electron capture detector has an extremely peculiar response characteristic. The response varies largely
depending upon the kind of compound. Even with the same kind of compound, the response may change
largely with minor differences in the molecular structure, e.g., with racemates.
For reference, relative sensitivities for various compounds are listed below. For the same component,
response may vary depending upon the structure of the detector, temperature, applied voltage, etc.
Therefore, the figures below should be considered as just relative standards.
2 ECD-2010
1 Outline
1.2 Response Characteristics of the Detector
ECD-2010 3
1 Outline
1.2 Response Characteristics of the Detector
4 ECD-2010
2 Parts List
22.1
. 2.1 Parts List
ECD-2010 5
2 Parts List
2.1 Parts List
6 ECD-2010
3 Basic Operation
33.1
. 3.1 Introduction
ECD-2010 7
3 Basic Operation
33.2
. 3.2 Gas used
Since the ECD is a detector with a high sensitivity, even slight impurities contained in the carrier gas may
affect the stability of the baseline.
A Gas Filter (option: 221-05619-01) and Oxygen trap (option: 221-46985-91) should be mounted along the
flow route to clear impurities from the carrier gas.
Both the carrier gas and make-up gas should meet the following purity standard:
8 ECD-2010
3 Basic Operation
33.3
. 3.3 Checking the glass insert and the injection inlet
For the procedural details, refer to “12. Injection port” of the instruction manual for the GC-2010.
WARNING
High temperature
You might suffer burns. Prior to start the work, check that the temperature at
each part is below 50°C.
CAUTION
In order to prevent seizure or the like, do not turn the screw when the
temperature is high.
The glass insert for the injection port differs in shape depending on the injection port type and column used.
Check that the proper insert for the sample injection mode is selected, that the silica wool is positioned
properly, that there is no contamination, and that the O-ring for the glass insert has not yet deteriorated in
quality.
Replace the injection inlet septum, in principle, when the number of injections reaches one hundred.
ECD-2010 9
3 Basic Operation
33.4
. 3.4 Mounting the column
For the detailed procedure for mounting the column, refer to “4. Installing the column” in the instruction
manual for the GC-2010. The graphite ferrule shall be placed in the position (37 mm) indicated in the
following figure. This placement is possible using the supplied graphite ferrule adjuster (P/N. 221-48610-03)
for the ECD-2010.
Split/splitless ECD-2010
injection port
34mm 37mm
Graphite ferrule
Capillary column
In order to prevent the ECD-2010 from being contaminated, connect the properly conditioned column.
10 ECD-2010
3 Basic Operation
33.5
. 3.5 Setting the flow rate
Set the flow rate of the carrier gas on the screen that is displayed by pressing the [FLOW] key Also set the
column and purge. For the procedural details, refer to “12. Injection Port” of the instruction manual for the
GC-2010.
Fig. 3.5.1
ECD-2010 11
3 Basic Operation
33.6
. 3.6 Setting the flow rate of detector gas
Set the flow rate of the makeup gas in the following procedure:
Press the [DET] key to display the detector’s condition setting screen. If multiple detectors are mounted,
press the [Next Det] in the PF menu to go to the ECD screen.
Press the [Det Gas] key in the PF menu to go to the Makeup gas flow rate setting screen.
Enter the gas type and its flow rate.
Turn On the flow controller.
In case of the capillary column, set the flow rate of the makeup gas to 15~60 ml/min.
NOTE The detection sensitivity of the ECD-2010 depends upon the sample concentration in the carrier gas
containing the makeup gas. Unless the peak range expands, the sensitivity increases as the flow rate of the
makeup gas decreases. For the quantitative analysis, make sure to set the flow rate as that of the last analy-
sis.
Fig. 3.6.1
12 ECD-2010
3 Basic Operation
33.7
. 3.7 Starting the system
Press the [SYSTEM] key when the GC-2010 stops. The GC Start Sequence screen will appear.
Set the start time to 10 minutes or more.
CAUTION
In order to avoid any damage to the ECD cell, set the start time so that the
detector temperature starts increasing after the air in the ECD cell has
entirely been substituted with the carrier gas and makeup gas. Before starting
the system, set the temperatures of the column oven and injection port to
nearly the room temperature (25°C).
After setting the parameters for starting the GC-2010, press the [Start GC] key in the PF menu. The GC will
start and the control will be performed for the temperatures at various parts and the flow rates of gases.
The control of the ECD control unit will not be started unless the detector is On.
For further details of this procedure, refer to “7. Starting and Stopping GC” in the instruction manual for the
GC-2010.
Fig. 3.7.1
ECD-2010 13
3 Basic Operation
33.8
. 3.8 Setting the detector temperature
To set the temperature of the ECD detector, press the [DET] key to display the DET screen.
CAUTION
In order to prevent the cell from being contaminated, set the detector
temperature to a value higher by at least 20°C than the maximum
temperature of the column oven during the analysis.
Using the cursor keys (UP and DOWN), move the cursor to the temperature row.
Enter a value with numeric keys and then press the [ENTER] key.
If the indicated detector is not the ECD-2010, press the [Next Det] key in the PF menu to change the
detector. (Pressing the toggle key allows you to change the PF menu.)
Fig. 3.8.1
NOTE The detector temperature can be set via either the [SET] or [DET] screen. The value set on either screen will
be reflected on the other screen.
To check the baseline, set the range to ×10-1 (×2-3 for attenuation) and the current value to 1 nA. For the
output of digital signals via the GCsolution or other, set the current value only.
To change the current value, move the cursor to the current value on the [DET] screen using the UP or
DOWN cursor keys. Enter a value and then press the [ENTER] key.
To change the range, move the cursor to the range on the [DET] screen. Using the LEFT or RIGHT cursor
keys, change the displayed range value and then press the [ENTER] key. (The attenuation can also be
changed in a similar procedure.)
14 ECD-2010
3 Basic Operation
33.9
. Setting the temperatures of the column oven
3.9
CAUTION
In order to prevent the ECD cell from being contaminated with matter eluting
from the column, maintain the detector at a temperature higher by at least
20°C than the column temperature.
Fig. 3.9.1
ECD-2010 15
3 Basic Operation
33.10
. 3.10 Checking the baseline
1. Set the attenuation of the Chromatopac (analog input) to 3 (2 ) and check that the baseline is
3
stabilized to the extent that becomes almost flat. For digital signals via the GCsolution or other, set the
full-scale value to 6.4×104 µV.
2. If the baseline is not stabilized as shown in Fig.3.10.2 or if it deflects from the range, see “6.1
Troubleshooting”.
Noise
Positive drift
Negative drift
Meander
Others
Zero adjustment
Press the [MONIT] key and also press the [Zero Adj.] in the PF menu. The signal output will become 0 µV.
If the indication on the Chromatopac or other deviates from 0 µV after the zero adjustment has been
completed on the GC, additionally make the zero adjustment on the Chromatopac.
For further details, see “4.7 Zero adjustment”.
16 ECD-2010
3 Basic Operation
33.11
. Setting the current value
3.11
The value for the current sent to the detector can be set in the range of 0.00 to 2.00 nA. As the current value
increases, the signal output becomes larger. Perform the analysis at a current value for which the zero
adjustment is possible.
For the high-sensitivity analysis, the current value may sometimes be set to 2 nA, but should normally be set
to 1 nA maximum. If the current value is too small, the output signal becomes so low that it cannot be
determined in some cases.
To change the current value, move the cursor to the current value on the [DET] screen using the UP or
DOWN cursor key. Enter a value and then press the [ENTER] key.
Fig. 3.11.1
ECD-2010 17
3 Basic Operation
33.12
. 3.12 Setting the range/attenuation
1. Analog output
If the analog output is connected to a data processing unit, two types of analog signal output, wide
range and linear range are available.
To connect the analog output to a data processing unit compatible with the wide range (the analog
input to the C-R7A, C-R7Aplus, or C-R8A), set the analog signal type to “Wide”. To connect it to any
other data processing unit, set the analog signal type to “Linear”.
If Wide is selected, no range selection is required. However, if the attenuation control on the data
processing unit does not work (e.g., an error peak occurs), the attenuation can be changed on the
[DET] screen. There are five attenuation levels from ×1 (2-0) to ×2-4. When the exponential value
increases by 1, the output becomes 1/2. Prior to using this function, calibrate the data processing unit.
(See “2.7 Connecting cables to the extended devices” of GC-2010 instruction manual.)
If Linear is selected, the range (amplification factor of the amplifier) can be set in five levels from ×1
(100) to x10-4. The range ×1 provides the largest amplification factor of the amplifier. The range ×10-1
provides 1/10 that amplification factor, the range ×10-2, 1/100, the range ×10-3, 1/1000, and the range
×10-4, 1/10000.
For the high-sensitivity analysis, therefore, use a small range (×1 or ×10-1). Otherwise, use a large
range (×10-2, ×10-3, or ×10-4).
Set the signal range of the detector and the analog signal type on the screen that is displayed by
pressing the [DET] key.
Fig. 3.12.1
2. Digital output
If the digital output is connected a data processing unit compatible with digital signals (the digital input
to the GCsolution or the C-R8A), the digital signals are always output at ×1, which means that no
range selection is required.
18 ECD-2010
3 Basic Operation
33.13
. 3.13 Caution during analysis
CAUTION
Caution on sample
1. Clean up the sample as far as possible, if it is from a biological source,
and then inject it to the gas chromatograph.
2. Do not use a strong electrophilic compound, e.g., carbon tetrachloride or
chloroform as a solvent. This would cause the baseline to deflect from the
range, requiring longer time before the baseline becomes stable again.
NOTE 1. In general, the sensitivity becomes the highest when the column flow rate of the carrie gas plus the flow
rate of the makeup gas is 30~60 ml/min.
2. If helium is used as the carrier gas, the baseline level becomes too high when the wide-bore capillary
column or packed column in which the carrier gas flows at a high flow rate is used. This difficulty in
analysis is attributable to the fact that the ionization efficiency per unit time of nitrogen sent as the
makeup gas to the ECD cell decreases. In this case, set the flow rate of the makeup gas to a slightly
larger value.
ECD-2010 19
3 Basic Operation
33.14
. 3.14 Stopping the system
Fig. 3.14.1
20 ECD-2010
3 Basic Operation
3.14 Stopping the system
NOTE The cell and flow path can be maintained to be clean by sending the makeup gas to the cell even if the ECD-
2010 is not used. The baseline will become stable earlier when the system is restarted, than in the case
where the makeup gas is stopped completely.
ECD-2010 21
3 Basic Operation
3.14 Stopping the system
22 ECD-2010
4 Key Operation
44.1
. 4.1 Setting the detector
Press the [DET] key to display the main screen shown in Fig.14.1.1 and to set the detector
temperature, the detector gas flow rate, etc.
When the detector is configured in the line, the setting of the output signal appears in the lower
half portion of the screen.
DETECTOR CONTROLLER
Selection: On/Off, Default: On
When a detector is set to “On”, its current and voltage are controlled.
When a detector is selected in a line and set to “On” here, the detector is used. However,
even if a detector is set to “Off” here, the gas flow and the temperature rises and reaches the
setup value if the detector is configured in a line.
TEMP
Range: 0.0 ~ 350.0°C, Default: 25.0°C
Set the detector temperature. Set usually the detector temperature to a value a little higher (by
approximately 30°C) than the column oven temperature (final temperature in temperature
program) so that adhesion of substances of high boiling point inside the detector can be
prevented. For setting of the maximum temperature, refer to “16.6.4 Setting of maximum
temperature” in “16.6 GC Configuration” of Instruction Manual for GC-2010.
ECD-2010 23
4 Key Operation
4.1 Setting the detector
Current
Setting range: 0.00~2.00 nA, default value: 0.00 nA
Set the value for the current sent to the detector.
SIGNAL RANGE
Selection: ×1/×10-1/×10-2/×10-3/×10-4, Default: /×10-1
You can multiply the analog signal by the coefficient “10-x” when the analog signal type is
linear. If the peak is saturated in the data processing unit, change the setting from “×1” to “×10-
1”, from “×10-2” to “×10-3”, for example. In case of the digital signal, the signal is always output
with “×1”.
SIGNAL ATTENUATION
Selection: ×1/×2-1/×2-2/×2-3/×2-4, Default: /×2-1
You can multiply the analog signal by the coefficient “2-x” when the analog signal type is wide.
If the peak is saturated in the data processing unit, change the setting from “×1” to “×2-1”, from
“×2-2” to “×2-3”, for example. In case of the digital signal, the signal is always output with “×1”.
24 ECD-2010
4 Key Operation
4.1 Setting the detector
PF menu Description
Det Gas Sets the flow rate of makeup gas, if the advanced Pressure Control (APC) is installed.
Next Det Changes over to setup screen for another installed detector.
You can change over screen by pressing [DET] key on detector screen.
ECD-2010 25
4 Key Operation
4.1 Setting the detector
Pressing the [FUNC] key and selecting “7. Service/Maintenance” in the screen that appears
displays the sub-screen shown in the figures.
Fig. 4.1.2
26 ECD-2010
4 Key Operation
4.1 Setting the detector
(0 = Not installed, 9 = Installed at the position other than those in the figure.)
8 4
7 3
6 2
1
Top view
Fig. 4.1.3
Fig. 4.1.4
Cont. Type
The control unit type is automatically recognized and displayed. When this is not displayed,
the control unit may not be mounted correctly or a failure may have occurred.
ECD-2010 27
4 Key Operation
4.1 Setting the detector
Slot No.
Select the APC slot number connected to the detector.
The DET APC No. and Unit type are automatically displayed when SLOT is set.
Also, “APC (1ch)” is used for ECD.
Heater port
This is the temperature control port of the detector to which flow controller set in the slot is
connected.
Set DET 1 when the DET1 is connected to the ECD detector and DET2 when connected to
DET2.
For the details, refer to “16.7 Service and maintenance” in GC-2010 Instruction Manual (CD-
ROM version).
Press the [SET] key and display the SET key screen.
Press [Line Confg] of the PF menu and display the sub-screen of the line configuration.
Fig. 4.1.5
28 ECD-2010
4 Key Operation
44.2
. 4.2 Selecting the detector
The key operation and the screen display when selecting the detector are as follows:
Pressing the [DET] key displays the setting screen of the detector. When more than one detector is loaded,
the setting screen of the currently selected detector will be displayed.
Press the toggle key to select the PF menu. Pressing the [Next Det] can select the other detectors.
When no detector is connected, pressing the [DET] displays “Detector is not installed”.
When changing from FID of DET#1 to ECD of DET#2, the screen is changed as follows:
Fig. 4.2.1
ECD-2010 29
4 Key Operation
44.3
. 4.3 Selecting the detector gases
Press the [DET] key to display the [DET] screen. Select Detector Gas from the PF menu. The [DET Gas]
screen will appear.
Set the type of gas to “N2”.
Set the flow rate of the makeup gas using the following values as guidelines:
Fig. 4.3.1
PF key
Return: Returns to the detector setting screen.
Program: Goes to the detector gas flow rate program screen.
On/Off: Toggles between turning ON and OFF the control for the flow rates of detector gases
NOTE 1. The detection sensitivity of the ECD-2010 depends upon the concentration of the sample in the carrier
gas (including the makeup gas). Therefore, the sensitivity increases, as the flow rate of the makeup gas
becomes lower, unless the peak range expands.
2. To purge the detector connecting part of the column, send the makeup gas to even the packed column.
3. If the detector gas is stopped, it will take longer time before the baseline becomes stable when the
system is restarted. It is recommended that the makeup gas be continuously sent even if the GC stops.
30 ECD-2010
4 Key Operation
4.3 Selecting the detector gases
Fig. 4.3.2
Under the general analytical conditions, the flow rate program for the detector gases is not necessary to use.
ECD-2010 31
4 Key Operation
44.4
. 4.4 Detector ON/OFF
Using the UP and DOWN cursor keys, position the cursor on the desired item. Select ON or OFF by
horizontally moving the cursor, and then press the [ENTER] key.
Fig. 4.4.1
32 ECD-2010
4 Key Operation
44.5
. 4.5 Setting the range/attenuation
Position the cursor on the indicated item with the UP and DOWN cursor, change the range with the LEFT and
RIGHT cursors, and then press the [ENTER] key. Normally, use ×10-1. For the high-sensitivity analysis, use
×1. If the analog signal type is Wide, use ×2-1.
Digital signals are output as ×1 regardless of the above setting.
Fig. 4.5.1
ECD-2010 33
4 Key Operation
44.6
. 4.6 Setting the current value
The value for the current sent to the detector can be set in units of 0.01 nA in the range of 0.00 to 2.00 nA.
The default value is 0.00 nA. The output signal increases as the current value becomes larger. However, the
sensitivity remains almost unchanged because the noise level simultaneously becomes higher.
Move the cursor to the row for the current value using the UP or DOWN cursor key. Enter a value with
numeric keys and then press the [ENTER] key.
Fig. 4.6.1
34 ECD-2010
4 Key Operation
44.7
. 4.7 Zero adjustment
Fig. 4.7.1
ECD-2010 35
4 Key Operation
4.7 Zero adjustment
Fig. 4.7.2
[Zero adj.]
Z key on the
Chromatopac
Zero point on
the Chromatopac
36 ECD-2010
4 Key Operation
4.7 Zero adjustment
Fig. 4.7.3
[Down]
Z key on the
[Up] Chromatopac
Zero point on
the Chromatopac
The baseline has decreased to the The zero point is If the baseline is stable,
negative side since the [Down] key adjusted for the the analysis is possible.
has been held down too long. Chromatopac.
ECD-2010 37
4 Key Operation
4.7 Zero adjustment
Fig. 4.7.4
When the level does not move with key operation, it is exceeding upper or lower limit. In such a case, refer
to “6.1 Troubleshooting”.
38 ECD-2010
5 Service/Maintenance
55.1
. 5.1 Checking the ECD cell
If the sample or other accumulates on the inside of the ECD cell, the baseline level becomes higher. Even in
this case, the analysis can be performed so far as the baseline does not deflect from the range when the
[Zero Adj] key is pressed. If the current value is decreased, the baseline can be lowered. In this case,
however, the output signal (peak) becomes smaller.
For the daily service, check the ECD cell for contamination in the following steps:
1. Set the temperatures of the column mounted and other parts, the flow rate, and the current value as
during the analysis.
4. Press the [Zero Free] key in the PF menu and then read the detector signal level indicated.
5. In principle, the signal level (µV) in the above step 4. falls below the following value against the set
current value, the cell contamination is regarded as within the acceptable level:
Set current value (nA) × 90000 (e.g., the set current value is 1 nA, the limit is 90000 µA).
However, even if this limit is exceeded, the analysis can be performed so far as the satisfactory peak
can be obtained when the current is set to a smaller value.
5. Press the [Return] key in the PF menu. You will return to the Main screen for the standard self-
diagnosis. Press the [StartDiag] key in the PF menu to start the diagnosis.
6. When the diagnosis is completed, the comprehensive diagnostic result and the number of anomalies
will be displayed.
7. If the diagnostic result for the ECD frequency is Not Good, it is recommended that the cell be cleaned
(the radiation source be replaced).
(see “5.3 Cleaning Procedure for ECD Cell.”)
For the details of the self-diagnosis, refer to Chapter “14. Diagnosis” in the instruction manual for the
GC-2010 (CD-ROM version).
NOTE In the standard self-diagnosis, the default value for the threshold of the ECD frequency is 40 kHz at 2 nA.
This value can be changed if the standard value for the system check is selected in the system configuration
of the GCsolution for the ECD-2010. However, it cannot be changed on the GC-2010 alone.
The ECD frequency is in proportion to the detector signal level, and the output frequency of 40 kHz is
equivalent to the signal level of 2,500,000 µV.
ECD-2010 39
5 Service/Maintenance
55.2
. 5.2 How to Keep the ECD Cell Clean
The following cautions should be exercised in order to keep the ECD cell clean:
When a packed column with a low boiling point of the liquid phase is used at a high temperature,
conditioning the ECD cell immediately after use is an effective method for extending the operational
life of the ECD cell.
7. Set the detector temperature to a value higher than that of the column oven.
The liquid phase of the column does not accumulate on the ECD cell if the detector temperature is
higher than the column oven temperature. Set the detector tremperature to a value 20~50°C higher
than that of the column oven. However, beware that the max. operation temperature of the ECD is
350°C. (Refer to “3.8 Setting the detector temperature”.)
40 ECD-2010
5 Service/Maintenance
5.2 How to Keep the ECD Cell Clean
ECD-2010 41
5 Service/Maintenance
55.3
. 5.3 Cleaning procedure for ECD Cell
The ECD cell used for gas chromatography may through long service, be contaminated with samples,
column packing or liquidphase, causing sensitivity to drop, abnormal peck shapes or inverted peaks, etc. In
such a case contact our company for cleaning (repair) of the ECD cell in the following procedure:
CAUTION
For the mounting and removing the ECD unit, contact our service engineer.
1. Ask our Sales office for a “Request for a Cleaning ( Repairing ) ECD Cell” slip.
2. Write the necessary information including your special request on the slip and send it to us (Sales
office).
3. Remove the ECD unit from the gas chromatograph main body.
4. The ECD unit must be sent by the user himself directly to our office by mail according to the law.
(For the mailing method and the address, please contact our Sales office.)
5. The ECD unit sent from you will be cleaned (or repaired)
by us according to the information on the slip.
6. After having been cleaned or repaired, the ECD unit will be sent back to you by mail.
7. Mount the ECD unit on the main body.
NOTE In cleaning the ECD cell, it is unavoidable to handle the nonsealed source. Therefore, if the user wants to
clean the cell on his side, it is essential that he observe the applicable national, state and/or local laws and
regulations regarding the handling of radioactive materials.
42 ECD-2010
5 Service/Maintenance
5.3 Cleaning procedure for ECD Cell
Call us.
Branch office or agent
Request
GC-2010 slip
Remove the ECD unit. Use the box the ECD unit originally
came in for mailling.
Pack
Repair
GC-2010
Completed.
ECD-2010 43
5 Service/Maintenance
55.4
. Actions to Be Taken in Case There Is the
5.4
Doubt of Break
If there is a doubt of break of the ECD, the measurement with the survey meter must be carried out. If there
is any abnormality such as the disuse of the ECD etc.
44 ECD-2010
6 Troubleshooting
66.1
. 6.1 Troubleshooting
This chapter covers causes of and countermeasures against troubles encountered when using the ECD.
Troubles which are attributable to the gas chromatograph main unit are covered in “Troubleshooting” in the
instruction manual of the gas chromatograph.
If a trouble cannot be cleared by the countermeasure given hereafter, or if a troubles other than outlined in
this chapter occurs, contact Shimadzu office or agent which is nearest to you.
1. Abnormal baseline
■ Significant noise occurs in the baseline.
Noise
Cause Countermeasures
The electrical connector is not securely connected. Check the connector and connect securely.
There is a large quantity of impurities in the carrier • Purge the air inside the flow line completely. (Refer
gas or make-up gas. to “10.3 Conditioning of Equipment”)
• Replace the filter (molecular sieve filter, gas filter
or oxygen trap).
• Replace the plumbing parts.
• Replace the gas container, changing to a gas of
higher purity.
Gas leakage Check for gas leakage and correct any joints from
which gas is leaking.
The cell is contaminated. Check the cell for contamination. If necessary, send
the ECD unit to our Company for cleaning. (Refer to
“5. Service/Maintenance”.)
ECD-2010 45
6 Troubleshooting
6.1 Troubleshooting
Cause Countermeasures
Air is trapped in the flow line. Purge the air completely. (Refer to “10.3 Conditioning
of Equipment”.)
The carrier gas or make-up gas is not flowing and air Correct the flow conditions of the carrier gas or make-
is mixed in. up gas.
Injection port (glass insert) or column is contaminated. • Clean the glass insert or replace it. (Refer to “3.3
Checking the glass insert and the injection inlet”.)
• Condition the column. (Refer to the instruction
manual of the gas chromatograph.)
Temperature of each section is unstable. Monitor the temperature of each section on the
monitor screen ([MONIT] key).
Contact our branch office or agent if the temperature
will not stabilize.
The flow rate is unstable. Monitor the pressure and flow rate on the monitor
screen ([MONIT] key).
Contact our branch office or agent if the temperature
will not stabilize.
The cell is contaminated. Check the cell for contamination. If necessary, send
the ECD unit to our company for cleaning. (Refer to
“5. Service/Maintenance”.)
Cause Countermeasure
Air is trapped in the flow route. Purge the air completely. (Refer to “10.3 Conditioning
of Equipment”.)
The carrier gas or make-up gas is not flowing and air Correct the flow conditions of the carrier gas or make-
is mixed in. up gas.
The cell is contaminated. Check the cell for contamination. If necessary, send
the EDC unit to our company for cleaning. (Refer to
“5. Service/Maintenance”.)
46 ECD-2010
6 Troubleshooting
6.1 Troubleshooting
Cause Countermeasures
The carrier gas is leaking. Check for gas leakage and correct any piping joints
from which gas is leaking.
The split ratio is too large. (in case of split analysis) Reduce the split ratio.
The quantity of the sample injected is too small, or the Increase quantity or concentration of injected sample.
sample concentration is too low.
The value of current, range or attenuation is inappro- Set the range or attenuation to ×1 or current.
priate.
The capillary column has been connected in the Mount the column in the correct way.
wrong way. (Refer to “3.4 Mounting the column”.)
An incorrect column is used. Replace the column in the correct one, taking the
polarity into consideration.
There is a large quantity of impurities in the carrier • Purge the air inside the flow route completely.
gas or make-up gas. (Refer to “10.3 Conditioning of Equipment”.)
• Replace the filter (molecular sieve filter, gas filter
or oxygen trap).
• Replace the plumbing parts.
• Replace the gas container, changing to a gas of
higher purity.
The cell is contaminated. Check the cell for contamination. If necessary, send
the ECD unit to our company for cleaning. (Refer to
“5. Service/Maintenance”.)
ECD-2010 47
6 Troubleshooting
6.1 Troubleshooting
4. The linearity of the calibration curve has become significantly poor. (Refer to “10.2 Calibration
Curve”.)
Calibration curve
Cause Countermeasures
There is a large quantity of impurities in the carrier • Purge the air inside the flow route completely.
gas or make-up gas . (Refer to “10.3 Conditioning of Equipment”.)
• Replace the filter (molecular sieve filter, gas filter
or oxygen trap).
• Replace the plumbing parts.
• Replace the gas container. (changing to a gas of
higher purity.)
The column is contaminated . • Condition the column. (Refer to the instruction
manual of the gas chromatograph.)
The cell is contaminated. Check the cell for contamination. If necessary, send
the ECD unit to our company for cleaning. (Refer to
“5. Service/Maintenance”.)
48 ECD-2010
6 Troubleshooting
66.2
. 6.2 Error Message
Error messages specific to the ECD are explained in this section. As for the error messages relating to the
gas chromatograph main unit , refer to “19.2 Error Messages” in the instruction manual of the gas
chromatograph.
If the unit for which the temperature is controlled exceeds the upper limit for its operating temperature, the
confirmation screen appears prompting you to reset the error with the above message displayed.
When the ECD is used, the max. temperature of the detector is set to 350°C or less. For the ECD, if the
detector temperature exceeds 350°C, the safety function operates automatically and it becomes OVER
TEMP. error.
Therefore, if an attempt to set the detector temperature to over 350°C, this OVER TEMP error returns.
Beware that, even when the detector temperature has been set to less than 350°C, the temperature may
exceed 350°C momentary on the way to the set value.
If this error occurs, move the cursor to “Ignore Error” and then press the [ENTER] key. Eliminate the cause
of the error by changing the settings or taking any other action. Several 10 seconds after the [ENTER] key
has been pressed, the confirmation screen appears. Position the cursor on “Reset Error” and then press
the [ENTER] key.
If you return to the confirmation screen when you are changing the settings, select “Ignore Error” again.
ECD-2010 49
6 Troubleshooting
6.2 Error Message
50 ECD-2010
7 Disuse
77.1
. 7.1 Cautions at Disuse
When disusing the used ECD etc., you must take legal administrative procedures at the same time. Also, in
the case of disposing the gas chromatograph equipped with the ECD, be sure to remove the ECD and take
the measures mentioned above.
NOTE When removing the ECD from the gas chromatograph, contact our service engineer.
* When removing the ECD unit, plug the hole with the heat insulating material which was cut off at
installation.
ECD-2010 51
7 Disuse
7.1 Cautions at Disuse
52 ECD-2010
8 Maintenance Parts
88.1
. 8.1 Maintenance Parts List
ECD-2010 53
8 Maintenance Parts
8.1 Maintenance Parts List
54 ECD-2010
9 Specifications
99.1
. 9.1 Specifications of the ECD-2010
ECD-2010 55
9 Specifications
9.1 Specifications of the ECD-2010
56 ECD-2010
10 Appendix
1010.1
. Structure of ECD Cell
10.1
Electrode
Radioactivity mark
mechanism confirmation mark
To VENT
Radiation source
Heating block
ECD cell body
Capillary adaptor
ECD-2010 57
10 Appendix
1010.2
. 10.2 Calibration Curve
This chapter covers the explanation of linearity of calibration curve which has been described in 6.1
“Troubleshooting”.
Calibration curve
The “Calibration curve” refers to the graph shown in Fig.10.2.1. this graph shows the relationship between
the sample quantity and the height of the peak.
Dynamic range
The “linearity of calibration curve” refers to in what range the graph can be regarded as the straight line.
This range is called as the dynamic range. For instance in Fig.10.2.1, the graph between the sample
quantity of 10-1 pg to the sample quantity 103 can be regarded as the straight line. Therefore, the dynamic
range in this case is 103 / 10-1 = 104.
Quantativity
In the range in which the calibration curve appears as straight line, the peak height is proportional to the
sample quantity. It means that the quantitative analysis cad be performed within this range. Note that the
quantitative analysis cannot be performed out of this range.
Dynamic range
6
10
5
10
Peak height
4
10
3
10
2
10
-1 0 1 2 3 4
10 10 10 10 10 10 (Pg)
Sample quantity
Detection limit
58 ECD-2010
10 Appendix
1010.3
. Conditioning of Equipment
10.3
After the installation of equipment or after the replacement of pipings, air is trapped inside the flow line. If air
is mixed in the gas inside the flow line, the base line, the base line connot be stabilized and analysis cannot
be performed correctly. Equipment must be conditioned prior to starting analysis. Given in this section is an
example of settings of equipment conditioning . As for the details of the operational procedure, refer to “3.
Basic Operation”.
It is recommended to condition the equipment in the following settings :
2. Column
Use of an empty column (column without liquid phase) is recommended.
(to prevent the liquid phase of the column from being accumulated on the ECD cell)
When a normal column is connected, use a column with an excellent heat resistance Also, a packed
column should not be used for the conditioning of equipment since its liquid phase quantity is large.
3. Temperature setting
Column over temperature : 340°C (empty column)
Injection port temperature : 340°C
Detector temperature : 340°C
NOTE When a normal column is connected, the column oven temperature should be set lower than max.
temperature of the column by 30°C and more.
NOTE When the FQ tube (0.2 mmID×2 m) mentioned above is connected, conditioning can be performed if the flow
rate is within the control range of the flow controller and if the carrier gas is flowing to the column.
5. Detector settings
Range=1 , Current=1 nA
(Check the baseline by changing the values if necessary.)
6. Conditioning time
~half a day
(This time depends on the period during which the ECD has not been used.)
ECD-2010 59
10 Appendix
10.3 Conditioning of Equipment
60 ECD-2010